Technology White Paper | CARESTREAM Touch Prime Ultrasound System
Touch Prime: Advanced ultrasound imaging with an
extraordinary user experience
Author: Ajay Anand, PhD
Introduction
The CARESTREAM Touch Prime Ultrasound System is based on
an all new computing architecture equipped to provide precise
advanced front-end beam forming, novel imaging modes, and
automated image optimization capabilities. Advanced
ultrasound beamforming and post-processing technologies
greatly benefit from access to increased computing power.
Medical ultrasound imaging continues to seamlessly leverage
advances from the consumer space, and the use of Graphic
Processing Units (GPUs) in combination with highly-integrated
front-end hardware is a prime example of this trend.
Embracing this, the Touch Prime system is based on a new
hardware architecture built from the ground up and equipped
with game-changing ultrasound imaging technologies. When
combined with an extraordinary user experience, it makes the
Touch Prime the next revolution in Ultrasound.
The Touch Prime is targeted for Radiology, OB/GYN,
Musculoskeletal (MSK), Vascular and general cardiac imaging
applications. This whitepaper highlights some of the key
underlying technologies available on the Touch Prime
Ultrasound System.
SynTek Architecture: Next Generation Smart
Beamforming
In conventional ultrasound systems, the time required to
acquire ultrasound data is physically constrained by the sound
propagation speed in the body. In soft tissue, this averages
1540 m/s. Conventional ultrasound systems have to wait for
this sound propagation, during both transmit and receive, due
to the serial nature of line-by-line data acquisition. As a result,
these systems require the sacrifice of frame rate and overall
speed, often forcing the clinician to choose between frame
rate and image quality. For example, temporal resolution may
be sacrificed for improved spatial resolution.
The acquisition speed limitation of conventional systems also
causes a degradation of either frame rate or image quality
when duplex modes such as color flow (Doppler imaging) is
employed. This is because color Doppler techniques can
require multiple pulses per scan line, and therefore frame rates
are lower than those for comparable depth grayscale anatomic
imaging. The problem, as noted earlier, is that conventional
beamforming is physically constrained by the serial data
acquisition method.
Another common implementation in modern conventional
ultrasound scanners equipped with digital beamformers is the
use of dynamic receive focusing but a fixed static transmit
focus. With digital beamformers, the echo signals are
continually focused as they are received by the transducer and
thus the received beams are maintained in constant focus. This
is done by constantly updating the beamformer focus delays as
the echoes return from increasing depths. However, since
image quality is a product of the transmit-and-receive focusing
profile, the best image quality is typically obtained around the
transmit focus where the beam width is narrowest.
In recent years, multizone imaging has been introduced and is
widely available on commercial ultrasound systems. This
approach is based on using consecutive transmissions along a
given direction with the transmit focus placed at different
depths, and blending the received beams from each of these
transmissions together to form an image scan line. However,
this approach compromises frame rate for improved
image quality.
Technology White Paper | CARESTREAM Touch Prime Ultrasound System
2
The CARESTREAM Touch Prime Ultrasound System presents a
paradigm shift from these conventional approaches by
employing an all new system architecture that leverages
integrated advanced GPU processing power and proprietary
parallel beamforming hardware. The inherent data parallelism
and high throughput offered by this combination is the basis
for advanced beamforming algorithms. Together, the
hardware and software forms the beamforming architecture
introduced by Carestream as SynTek architecture.
SynTek architecture is a significant departure from the serial
line-by-line acquisition approaches. With Syntek, a given tissue
region is insonified in multiple directions from independent
transmit firings. The echoes received by the transducer are
coherently summed together, taking into account the
difference in round-trip travel time from the transducer to the
tissue location and back for each of the firings. Thus, by
combining the information independently obtained from many
such transmit events throughout the imaging region of
interest, the SynTek architecture in effect synthesizes a
transmit beam that is narrow not only at a single point or
region in the image (around the transmit focus in conventional
imaging), but over the entire spatial extent, leading to
improved image quality.
In a conventional system, using such a transmit scheme with
multiple transmit firings would have led to a reduction in
frame rate. However, the parallel acquisition and real-time
processing capabilities of the SynTek architecture leads to
minimal compromise in frame rate at the cost of image quality.
Summary: SynTek Architecture
The performance advantages provided by the SynTek
architecture are multiple:
• The image resolution and penetration is enhanced,
since multiple beams at every depth location are
coherently combined, leading to a better Signal-to-
Noise ratio.
• Frame rates are increased due to a reduced number of
transmits being used to create the image.
• Since multiple overlapping transmit beams are used to
reconstruct a given scanline, the beam is focused over
multiple depths simultaneously, in turn making the
transmit focus position less critical as well. In color
flow and Doppler imaging, SynTek architecture allows
for more consistent visualization of subtle tissue
contrast differences, while simultaneously improving
the ability to see small structures. There is more
information at depth as well as increased frame rates
for improved visualization of moving structures.
The image below is an abdominal scan of the liver acquired
with the 6C2 transducer on the Touch Prime Ultrasound
System.
This image illustrates superb spatial and contrast resolution
throughout the entire image with uncompromised
clinical detail.
Clinical Benefits of SynTek Architecture
• Uniform lateral resolution over the entire depth,
making position of transmit focus less critical
• High frame rate while maintaining the spatial
resolution
• Improved penetration (e.g. for deep abdominal
imaging)
Technology White Paper | CARESTREAM Touch Prime Ultrasound System
3
Smart System Control (SSC)
Another path-breaking technology aimed at ultrasound
workflow enhancement available on the Touch Prime system
that harnesses the power of GPUs and parallel computing is
Smart System Control (SSC). It is a real-time optimization
technique implemented on the Touch Prime system to
automatically adjust over 25 different imaging parameters
(including some that are not under user control) to provide the
user with optimal image quality. It essentially provides
automatic system tuning beyond traditional keyboard/user
controls. SSC technology has detailed insight and control of
extended sets of parameters and their dependencies.
SSC technology can be used in B-mode, Color mode and
Doppler imaging modes. To determine the optimal settings,
the algorithm takes as input the user preference for frame rate
or resolution and optimizes other system parameters
automatically. The optimization is performed continuously in
the background in real-time, and does not require the user to
initiate it from the user console for each scan.
The following are examples of system parameters that SSC
automatically adjusts during the system operation:
Ultrasound Mode Parameters
B-mode • Line density
• Number of focal zones
• Number of transmit beams
• Number of Compounding angles
B-mode + Color (Duplex) • B-mode & Color line density
• Number of pulses transmitted in
each waveform packet (shots per
estimate)
SSC technology ensures constantly optimized images with
limited user interaction, which results in improved workflow
and diagnostic confidence. It has the potential to reduce exam
times because there is no need to adjust the imaging
parameters manually. The result is an optimized image with
fewer touch-panel hard and soft key presses by the user.
Taking it further, the intent is that for routine applications, SSC
technology can lead to out-of-the-box optimized images with
minimal manual changes needed from the user.
Beyond Imaging Performance
With the power of SynTek and GPU processing, the
CARESETREAM Touch Prime Ultrasound System is only using a
small fraction of its potential power at first release. In
conjunction with its programmable touch user interface, the
system is an ideal platform for upgrades without the need for
major hardware changes. As Carestream continues to innovate
with enhancements to imaging performance and new
features, the system lends itself to easy upgradeability and has
power for further technology-driven innovative applications.
The following are examples of such innovative applications
available at first release.
Smart Flow Imaging
To assess blood flow in arteries and veins, Doppler ultrasound
uses the frequency shift that occurs when sound waves
bounce off a moving object. Such images are used to uncover
blockages and clots, the narrowing of blood vessels and
congenital vascular malformations. In color Doppler imaging,
echo measurements are displayed as colors that indicate the
speed and direction of blood flow. They are used to identify
narrowed blood vessels, for example, and the tiny “jets” of
blood associated with vascular anomalies. Power Doppler
imaging is even more sensitive than color Doppler, visualizing
blood flow through tiny vessels such as those that feed tumors
in the thyroid and scrotum, as well as lesions just below the
skin. Spectral Doppler technique calculates and then graphs
the velocity of blood flow according to the distance blood
travels over time. Ordinary color and spectral Doppler
ultrasound measures the velocity of flow components toward
or away from the transducer. It uses the angle of the
ultrasound beam relative to the flow direction to calculate the
actual flow velocity through the vessel.
In current conventional implementations, the accuracy of
Doppler computations depends on precise knowledge of the
direction of the ultrasound beam and direction of the flow in
the vessel (and the angle α between them).
Technology White Paper | CARESTREAM Touch Prime Ultrasound System
4
As the following Doppler equation indicates, when the
ultrasound beam is perpendicular to the vessel (90°), this
computation is impossible, because there is no flow
component in the direction of the beam. In effect, the
measurement is impossible when the insonation angle is over
60°, because then small errors in measuring the two directions
lead to large discrepancies in the results.
Smart Flow imaging is a ground-breaking technology available
on the Touch Prime Ultrasound System that has the potential
to revolutionize the workflow for many Doppler ultrasound
applications. In addition to revealing flow pattern details, it
saves valuable time and makes the exam much easier. The new
proprietary Smart Flow method visualizes blood flow in all
directions, independent of imaging angle. It uses a
technique called transverse oscillation to overcome the angle
limitations of ordinary Doppler ultrasound. It creates an
effective component of the ultrasound oscillation
perpendicular to the transmitting beam – an oscillation in the
transverse direction (Jensen JA 2001; Jensen JA, Munk P
1998). Smart Flow technology generates a 2D interference
pattern in the received ultrasound signal. This allows the
system to calculate not only the axial component of the
velocity (as traditional color Doppler), but also the transverse
component. Therefore, Smart Flow technology eliminates the
angle dependence and enables both the detection and
visualization of complex flows.
To display the information from Smart Flow technology on the
Touch Prime platform, color coding and/or arrows are used.
The length of the arrow, in addition to the color, indicates the
magnitude, and the orientation of the arrow indicates the flow
direction. The figures below illustrate a side-by-side
comparison between conventional color flow imaging and the
new Smart Flow technology in a carotid artery scan. The
conventional color flow (left) produces drop-outs (regions with
no flow) at locations within the vessel where the flow is
oriented perpendicular to the acoustic beam (with no steering
or angling of the color box). For the same anatomy, the Smart
Flow image (right) displays continuous and robust flow
information throughout the entire lumen, even when the flow
is perpendicular to the acoustic beam.
Comparison between conventional color flow (left) vs. Smart Flow (right) imaging in a carotid artery scan
Technology White Paper | CARESTREAM Touch Prime Ultrasound System
5
The figure below illustrates a Smart Flow image of the jugular
vein (top) and carotid artery (bottom) in a normal volunteer
acquired with an 8L2 transducer. The arrows indicate opposite
flow directions, with longer arrow lengths towards the middle
of the carotid artery lumen representing higher velocities.
Smart Flow image of a jugular vein and carotid artery with the
arrows representing direction and velocity magnitude
In conclusion, Smart Flow technology provides an intuitive
visual representation of flow in all directions, making it well
suited to visualize transverse flow and complex flow patterns
including turbulence. The measurements are angle-
independent. It enables a faster workflow since the overall
hemodynamic view is available without having to steer. The
results are more consistent with less clinician dependence.
Smart Flow Assist
Spectral Doppler imaging is commonly performed to quantify
the flow profiles in the blood vessels. The measurement is
displayed as a waveform trace, depicting the velocity
distribution (profile) at the given spatial location as a function
of time. The typical steps involved in obtaining a velocity
profile include: 1) Turn on Color Flow mode to obtain the
orientation of the vessel, 2) Turn on Spectral Doppler mode, 3)
Update the beam steering direction, 4) Move the gate to
obtain the highest velocity, and 5) Angle-correct the gate. The
above steps require manual effort by the user, and moreover
steps 3 to 5 have to be repeated with movement of the
transducer. Contrary to this conventional mode of operation,
the Smart Flow Assist technology eliminates the need for
repeated manual adjustments. With the proprietary
functions in Smart Flow Assist technology, the software
automatically updates the beam steering, gate position and
angle correction. It is updated at all times, even when the
transducer is moved.
The benefits of Smart Flow Assist technology are even more
pronounced when measuring volume flow (VF). In this
situation, the workflow to obtain VF is reduced from 10 steps
(for a VF on a frozen image - and 10 more for every new
location) to 2 steps (and one for every new location).
Technology White Paper | CARESTREAM Touch Prime Ultrasound System
carestream.com
©Carestream Health, Inc., 2015. CARESTREAM is a trademark of
Carestream Health. CAT 200 0114 07/15
Conclusion
The CARESTREAM Touch Prime Ultrasound System provides
several advanced ultrasound technologies, leveraging
advanced GPU processing power for enhanced imaging
performance. It incorporates a next-generation beamforming
approach, leading to simultaneous improvement in frame rate
and spatial resolution compared to conventional systems and
automation of ultrasound scan settings that results in
improved workflow efficiency. These novel technologies, when
combined with exceptional user experience features such as an
all-Touch control panel, a thoughtful system design around
ergonomics, and one-touch transducer activation, makes the
Touch Prime a system of choice for Radiology, OB/GYN,
Musculoskeletal (MSK), Vascular and general cardiac imaging
applications.
Ajay Anand is a member of the ultrasound R&D team at Carestream Health. He has more than 10 years’ experience leading the development of
novel ultrasound technologies, and is a co-inventor of more than 20 patent filings in the field of medical ultrasound.
References
Jensen JA, “A new estimator for vector velocity estimation” IEEE Trans Ultrason Ferroelectr Freq Control 2001; 48 (4): 886-894.
Jensen JA, Munk P “A new method for estimation of velocity vectors”, IEEE Trans Ultrason Ferroelectr Freq Control 1998; 45 (3):
837-51.

More Related Content

What's hot

3D Spherical Harmonics Deformation Visualization of Intraoperative Organ
3D Spherical Harmonics Deformation Visualization of Intraoperative Organ3D Spherical Harmonics Deformation Visualization of Intraoperative Organ
3D Spherical Harmonics Deformation Visualization of Intraoperative OrganIOSR Journals
 
TREATMENT OF TGN WITH CYBERKNIFE FRAMELESS RADIOSURGERY SYSTEM
TREATMENT OF TGN WITH CYBERKNIFE FRAMELESS RADIOSURGERY SYSTEMTREATMENT OF TGN WITH CYBERKNIFE FRAMELESS RADIOSURGERY SYSTEM
TREATMENT OF TGN WITH CYBERKNIFE FRAMELESS RADIOSURGERY SYSTEMSubrata Roy
 
AN OVERVIEW ON CYBERKNIFE
AN OVERVIEW ON CYBERKNIFEAN OVERVIEW ON CYBERKNIFE
AN OVERVIEW ON CYBERKNIFESubrata Roy
 
Significance of Robust Motion Management & Approaches In Radiotherapy
Significance of  Robust Motion Management & Approaches In RadiotherapySignificance of  Robust Motion Management & Approaches In Radiotherapy
Significance of Robust Motion Management & Approaches In RadiotherapySubrata Roy
 
Portfolio Featuring AnyScan SPECT/PET/CT
Portfolio Featuring AnyScan SPECT/PET/CTPortfolio Featuring AnyScan SPECT/PET/CT
Portfolio Featuring AnyScan SPECT/PET/CTjohnseawright
 
Cyberknife At Saint Raphael’s Campus_revB
Cyberknife At Saint Raphael’s Campus_revBCyberknife At Saint Raphael’s Campus_revB
Cyberknife At Saint Raphael’s Campus_revBJustin Vinci
 
Digital Fluoroscopy Imaging System
Digital Fluoroscopy Imaging SystemDigital Fluoroscopy Imaging System
Digital Fluoroscopy Imaging SystemMuhammad Arif Afridi
 
DSA DEEPAK GUPTA
DSA  DEEPAK GUPTADSA  DEEPAK GUPTA
DSA DEEPAK GUPTADEEPAK
 
Introduction to Biomedical Imaging
Introduction to Biomedical ImagingIntroduction to Biomedical Imaging
Introduction to Biomedical Imagingu.surgery
 

What's hot (20)

3D Spherical Harmonics Deformation Visualization of Intraoperative Organ
3D Spherical Harmonics Deformation Visualization of Intraoperative Organ3D Spherical Harmonics Deformation Visualization of Intraoperative Organ
3D Spherical Harmonics Deformation Visualization of Intraoperative Organ
 
TREATMENT OF TGN WITH CYBERKNIFE FRAMELESS RADIOSURGERY SYSTEM
TREATMENT OF TGN WITH CYBERKNIFE FRAMELESS RADIOSURGERY SYSTEMTREATMENT OF TGN WITH CYBERKNIFE FRAMELESS RADIOSURGERY SYSTEM
TREATMENT OF TGN WITH CYBERKNIFE FRAMELESS RADIOSURGERY SYSTEM
 
AN OVERVIEW ON CYBERKNIFE
AN OVERVIEW ON CYBERKNIFEAN OVERVIEW ON CYBERKNIFE
AN OVERVIEW ON CYBERKNIFE
 
Significance of Robust Motion Management & Approaches In Radiotherapy
Significance of  Robust Motion Management & Approaches In RadiotherapySignificance of  Robust Motion Management & Approaches In Radiotherapy
Significance of Robust Motion Management & Approaches In Radiotherapy
 
CT definition flash - SIEMENS
CT definition flash - SIEMENSCT definition flash - SIEMENS
CT definition flash - SIEMENS
 
Spect technology
Spect technologySpect technology
Spect technology
 
Portfolio Featuring AnyScan SPECT/PET/CT
Portfolio Featuring AnyScan SPECT/PET/CTPortfolio Featuring AnyScan SPECT/PET/CT
Portfolio Featuring AnyScan SPECT/PET/CT
 
Cyberknife At Saint Raphael’s Campus_revB
Cyberknife At Saint Raphael’s Campus_revBCyberknife At Saint Raphael’s Campus_revB
Cyberknife At Saint Raphael’s Campus_revB
 
Basic principles of CT scanning
Basic principles of CT scanningBasic principles of CT scanning
Basic principles of CT scanning
 
1. principles & instruments of CT
1. principles & instruments of CT1. principles & instruments of CT
1. principles & instruments of CT
 
TomoTherapy
TomoTherapyTomoTherapy
TomoTherapy
 
Computed tomography
Computed tomography Computed tomography
Computed tomography
 
CT Scan Vs MRI Scan
CT Scan Vs MRI ScanCT Scan Vs MRI Scan
CT Scan Vs MRI Scan
 
SIEMENS - MAGNETOM Aera 1.5T
SIEMENS - MAGNETOM Aera 1.5TSIEMENS - MAGNETOM Aera 1.5T
SIEMENS - MAGNETOM Aera 1.5T
 
Digital Fluoroscopy Imaging System
Digital Fluoroscopy Imaging SystemDigital Fluoroscopy Imaging System
Digital Fluoroscopy Imaging System
 
4. system components
4. system components4. system components
4. system components
 
Radiology ppt
Radiology pptRadiology ppt
Radiology ppt
 
Somatom Definition Edge -SIEMENS
Somatom Definition Edge -SIEMENSSomatom Definition Edge -SIEMENS
Somatom Definition Edge -SIEMENS
 
DSA DEEPAK GUPTA
DSA  DEEPAK GUPTADSA  DEEPAK GUPTA
DSA DEEPAK GUPTA
 
Introduction to Biomedical Imaging
Introduction to Biomedical ImagingIntroduction to Biomedical Imaging
Introduction to Biomedical Imaging
 

Viewers also liked

Principles of Doppler ultrasound
Principles of Doppler ultrasoundPrinciples of Doppler ultrasound
Principles of Doppler ultrasoundSamir Haffar
 
Qi carotid study 5 10-12 -cme
Qi carotid study 5 10-12 -cmeQi carotid study 5 10-12 -cme
Qi carotid study 5 10-12 -cmeinertia182
 
Doppler Effect with math
Doppler Effect with mathDoppler Effect with math
Doppler Effect with mathAlexander Burt
 
Genetic principles in paediatric surgery
Genetic principles in paediatric surgeryGenetic principles in paediatric surgery
Genetic principles in paediatric surgeryFariha Hussain Nijhum
 
Basic physics of ultrasound.JH
Basic physics of ultrasound.JHBasic physics of ultrasound.JH
Basic physics of ultrasound.JHhari baskar
 
Physical Principles Of Ultrasound
Physical Principles Of UltrasoundPhysical Principles Of Ultrasound
Physical Principles Of Ultrasoundu.surgery
 
Doppler ultrasound of the portal system - Normal findings
Doppler ultrasound of the portal system - Normal findingsDoppler ultrasound of the portal system - Normal findings
Doppler ultrasound of the portal system - Normal findingsSamir Haffar
 

Viewers also liked (7)

Principles of Doppler ultrasound
Principles of Doppler ultrasoundPrinciples of Doppler ultrasound
Principles of Doppler ultrasound
 
Qi carotid study 5 10-12 -cme
Qi carotid study 5 10-12 -cmeQi carotid study 5 10-12 -cme
Qi carotid study 5 10-12 -cme
 
Doppler Effect with math
Doppler Effect with mathDoppler Effect with math
Doppler Effect with math
 
Genetic principles in paediatric surgery
Genetic principles in paediatric surgeryGenetic principles in paediatric surgery
Genetic principles in paediatric surgery
 
Basic physics of ultrasound.JH
Basic physics of ultrasound.JHBasic physics of ultrasound.JH
Basic physics of ultrasound.JH
 
Physical Principles Of Ultrasound
Physical Principles Of UltrasoundPhysical Principles Of Ultrasound
Physical Principles Of Ultrasound
 
Doppler ultrasound of the portal system - Normal findings
Doppler ultrasound of the portal system - Normal findingsDoppler ultrasound of the portal system - Normal findings
Doppler ultrasound of the portal system - Normal findings
 

Similar to whitepaper-advanced-ultrasound-imaging-201507

Digital Tomosynthesis: Theory of Operation
Digital Tomosynthesis: Theory of OperationDigital Tomosynthesis: Theory of Operation
Digital Tomosynthesis: Theory of OperationCarestream
 
Go 2 d3d ceph_cataloge
Go 2 d3d ceph_catalogeGo 2 d3d ceph_cataloge
Go 2 d3d ceph_catalogeYenDoan19
 
Tube and Grid Alignment System
Tube and Grid Alignment System Tube and Grid Alignment System
Tube and Grid Alignment System Carestream
 
A VLSI Architecture Realisation of an Wireless Endoscopy System
A VLSI Architecture Realisation of an Wireless Endoscopy SystemA VLSI Architecture Realisation of an Wireless Endoscopy System
A VLSI Architecture Realisation of an Wireless Endoscopy Systemijesajournal
 
COMPUTER APPLICATIONS IN DENTISTRY.pptx
COMPUTER APPLICATIONS IN DENTISTRY.pptxCOMPUTER APPLICATIONS IN DENTISTRY.pptx
COMPUTER APPLICATIONS IN DENTISTRY.pptxParkhiBhatngar
 
IRJET - Change Detection in Satellite Images using Convolutional Neural N...
IRJET -  	  Change Detection in Satellite Images using Convolutional Neural N...IRJET -  	  Change Detection in Satellite Images using Convolutional Neural N...
IRJET - Change Detection in Satellite Images using Convolutional Neural N...IRJET Journal
 
Noise reduction based on partial reference, dual-tree complex wavelet transfo...
Noise reduction based on partial reference, dual-tree complex wavelet transfo...Noise reduction based on partial reference, dual-tree complex wavelet transfo...
Noise reduction based on partial reference, dual-tree complex wavelet transfo...IEEEFINALYEARPROJECTS
 
Noise reduction based on partial reference, dual-tree complex wavelet transfo...
Noise reduction based on partial reference, dual-tree complex wavelet transfo...Noise reduction based on partial reference, dual-tree complex wavelet transfo...
Noise reduction based on partial reference, dual-tree complex wavelet transfo...JPINFOTECH JAYAPRAKASH
 
A Novel Framework For Preprocessing Of Breast Ultra Sound Images By Combining...
A Novel Framework For Preprocessing Of Breast Ultra Sound Images By Combining...A Novel Framework For Preprocessing Of Breast Ultra Sound Images By Combining...
A Novel Framework For Preprocessing Of Breast Ultra Sound Images By Combining...IRJET Journal
 
Advances in digital imaging
Advances in digital imagingAdvances in digital imaging
Advances in digital imagingDrRupanjanRoy
 
CT based Image Guided Radiotherapy - Physics & QA
CT based Image Guided Radiotherapy - Physics & QACT based Image Guided Radiotherapy - Physics & QA
CT based Image Guided Radiotherapy - Physics & QASambasivaselli R
 
Better Imaging equals better patient outcomes
Better Imaging equals better patient outcomesBetter Imaging equals better patient outcomes
Better Imaging equals better patient outcomesA-dec Australia
 
The Role of Computers in Medical Physics
The Role of Computers in Medical PhysicsThe Role of Computers in Medical Physics
The Role of Computers in Medical PhysicsVictor Ekpo
 
CBCT in dental practice
CBCT in dental practiceCBCT in dental practice
CBCT in dental practiceZana Hussein
 
ctfluoroscopy11-230414043252-3728b13a (1).pdf
ctfluoroscopy11-230414043252-3728b13a (1).pdfctfluoroscopy11-230414043252-3728b13a (1).pdf
ctfluoroscopy11-230414043252-3728b13a (1).pdfNiteshChandraDas1
 
Numerical simulation of flow modeling in ducted axial fan using simpson’s 13r...
Numerical simulation of flow modeling in ducted axial fan using simpson’s 13r...Numerical simulation of flow modeling in ducted axial fan using simpson’s 13r...
Numerical simulation of flow modeling in ducted axial fan using simpson’s 13r...iaemedu
 

Similar to whitepaper-advanced-ultrasound-imaging-201507 (20)

Digital Tomosynthesis: Theory of Operation
Digital Tomosynthesis: Theory of OperationDigital Tomosynthesis: Theory of Operation
Digital Tomosynthesis: Theory of Operation
 
Go 2 d3d ceph_cataloge
Go 2 d3d ceph_catalogeGo 2 d3d ceph_cataloge
Go 2 d3d ceph_cataloge
 
Tube and Grid Alignment System
Tube and Grid Alignment System Tube and Grid Alignment System
Tube and Grid Alignment System
 
CT dose reduction
CT dose reductionCT dose reduction
CT dose reduction
 
A VLSI Architecture Realisation of an Wireless Endoscopy System
A VLSI Architecture Realisation of an Wireless Endoscopy SystemA VLSI Architecture Realisation of an Wireless Endoscopy System
A VLSI Architecture Realisation of an Wireless Endoscopy System
 
COMPUTER APPLICATIONS IN DENTISTRY.pptx
COMPUTER APPLICATIONS IN DENTISTRY.pptxCOMPUTER APPLICATIONS IN DENTISTRY.pptx
COMPUTER APPLICATIONS IN DENTISTRY.pptx
 
IRJET - Change Detection in Satellite Images using Convolutional Neural N...
IRJET -  	  Change Detection in Satellite Images using Convolutional Neural N...IRJET -  	  Change Detection in Satellite Images using Convolutional Neural N...
IRJET - Change Detection in Satellite Images using Convolutional Neural N...
 
Noise reduction based on partial reference, dual-tree complex wavelet transfo...
Noise reduction based on partial reference, dual-tree complex wavelet transfo...Noise reduction based on partial reference, dual-tree complex wavelet transfo...
Noise reduction based on partial reference, dual-tree complex wavelet transfo...
 
Noise reduction based on partial reference, dual-tree complex wavelet transfo...
Noise reduction based on partial reference, dual-tree complex wavelet transfo...Noise reduction based on partial reference, dual-tree complex wavelet transfo...
Noise reduction based on partial reference, dual-tree complex wavelet transfo...
 
CT Scanner Trends
CT Scanner TrendsCT Scanner Trends
CT Scanner Trends
 
A Novel Framework For Preprocessing Of Breast Ultra Sound Images By Combining...
A Novel Framework For Preprocessing Of Breast Ultra Sound Images By Combining...A Novel Framework For Preprocessing Of Breast Ultra Sound Images By Combining...
A Novel Framework For Preprocessing Of Breast Ultra Sound Images By Combining...
 
Advances in digital imaging
Advances in digital imagingAdvances in digital imaging
Advances in digital imaging
 
CT based Image Guided Radiotherapy - Physics & QA
CT based Image Guided Radiotherapy - Physics & QACT based Image Guided Radiotherapy - Physics & QA
CT based Image Guided Radiotherapy - Physics & QA
 
Better Imaging equals better patient outcomes
Better Imaging equals better patient outcomesBetter Imaging equals better patient outcomes
Better Imaging equals better patient outcomes
 
Tomo modulation
Tomo modulationTomo modulation
Tomo modulation
 
The Role of Computers in Medical Physics
The Role of Computers in Medical PhysicsThe Role of Computers in Medical Physics
The Role of Computers in Medical Physics
 
CBCT in dental practice
CBCT in dental practiceCBCT in dental practice
CBCT in dental practice
 
ctfluoroscopy11-230414043252-3728b13a (1).pdf
ctfluoroscopy11-230414043252-3728b13a (1).pdfctfluoroscopy11-230414043252-3728b13a (1).pdf
ctfluoroscopy11-230414043252-3728b13a (1).pdf
 
CT fluoroscopy1 (1).pptx
CT fluoroscopy1 (1).pptxCT fluoroscopy1 (1).pptx
CT fluoroscopy1 (1).pptx
 
Numerical simulation of flow modeling in ducted axial fan using simpson’s 13r...
Numerical simulation of flow modeling in ducted axial fan using simpson’s 13r...Numerical simulation of flow modeling in ducted axial fan using simpson’s 13r...
Numerical simulation of flow modeling in ducted axial fan using simpson’s 13r...
 

whitepaper-advanced-ultrasound-imaging-201507

  • 1. Technology White Paper | CARESTREAM Touch Prime Ultrasound System Touch Prime: Advanced ultrasound imaging with an extraordinary user experience Author: Ajay Anand, PhD Introduction The CARESTREAM Touch Prime Ultrasound System is based on an all new computing architecture equipped to provide precise advanced front-end beam forming, novel imaging modes, and automated image optimization capabilities. Advanced ultrasound beamforming and post-processing technologies greatly benefit from access to increased computing power. Medical ultrasound imaging continues to seamlessly leverage advances from the consumer space, and the use of Graphic Processing Units (GPUs) in combination with highly-integrated front-end hardware is a prime example of this trend. Embracing this, the Touch Prime system is based on a new hardware architecture built from the ground up and equipped with game-changing ultrasound imaging technologies. When combined with an extraordinary user experience, it makes the Touch Prime the next revolution in Ultrasound. The Touch Prime is targeted for Radiology, OB/GYN, Musculoskeletal (MSK), Vascular and general cardiac imaging applications. This whitepaper highlights some of the key underlying technologies available on the Touch Prime Ultrasound System. SynTek Architecture: Next Generation Smart Beamforming In conventional ultrasound systems, the time required to acquire ultrasound data is physically constrained by the sound propagation speed in the body. In soft tissue, this averages 1540 m/s. Conventional ultrasound systems have to wait for this sound propagation, during both transmit and receive, due to the serial nature of line-by-line data acquisition. As a result, these systems require the sacrifice of frame rate and overall speed, often forcing the clinician to choose between frame rate and image quality. For example, temporal resolution may be sacrificed for improved spatial resolution. The acquisition speed limitation of conventional systems also causes a degradation of either frame rate or image quality when duplex modes such as color flow (Doppler imaging) is employed. This is because color Doppler techniques can require multiple pulses per scan line, and therefore frame rates are lower than those for comparable depth grayscale anatomic imaging. The problem, as noted earlier, is that conventional beamforming is physically constrained by the serial data acquisition method. Another common implementation in modern conventional ultrasound scanners equipped with digital beamformers is the use of dynamic receive focusing but a fixed static transmit focus. With digital beamformers, the echo signals are continually focused as they are received by the transducer and thus the received beams are maintained in constant focus. This is done by constantly updating the beamformer focus delays as the echoes return from increasing depths. However, since image quality is a product of the transmit-and-receive focusing profile, the best image quality is typically obtained around the transmit focus where the beam width is narrowest. In recent years, multizone imaging has been introduced and is widely available on commercial ultrasound systems. This approach is based on using consecutive transmissions along a given direction with the transmit focus placed at different depths, and blending the received beams from each of these transmissions together to form an image scan line. However, this approach compromises frame rate for improved image quality.
  • 2. Technology White Paper | CARESTREAM Touch Prime Ultrasound System 2 The CARESTREAM Touch Prime Ultrasound System presents a paradigm shift from these conventional approaches by employing an all new system architecture that leverages integrated advanced GPU processing power and proprietary parallel beamforming hardware. The inherent data parallelism and high throughput offered by this combination is the basis for advanced beamforming algorithms. Together, the hardware and software forms the beamforming architecture introduced by Carestream as SynTek architecture. SynTek architecture is a significant departure from the serial line-by-line acquisition approaches. With Syntek, a given tissue region is insonified in multiple directions from independent transmit firings. The echoes received by the transducer are coherently summed together, taking into account the difference in round-trip travel time from the transducer to the tissue location and back for each of the firings. Thus, by combining the information independently obtained from many such transmit events throughout the imaging region of interest, the SynTek architecture in effect synthesizes a transmit beam that is narrow not only at a single point or region in the image (around the transmit focus in conventional imaging), but over the entire spatial extent, leading to improved image quality. In a conventional system, using such a transmit scheme with multiple transmit firings would have led to a reduction in frame rate. However, the parallel acquisition and real-time processing capabilities of the SynTek architecture leads to minimal compromise in frame rate at the cost of image quality. Summary: SynTek Architecture The performance advantages provided by the SynTek architecture are multiple: • The image resolution and penetration is enhanced, since multiple beams at every depth location are coherently combined, leading to a better Signal-to- Noise ratio. • Frame rates are increased due to a reduced number of transmits being used to create the image. • Since multiple overlapping transmit beams are used to reconstruct a given scanline, the beam is focused over multiple depths simultaneously, in turn making the transmit focus position less critical as well. In color flow and Doppler imaging, SynTek architecture allows for more consistent visualization of subtle tissue contrast differences, while simultaneously improving the ability to see small structures. There is more information at depth as well as increased frame rates for improved visualization of moving structures. The image below is an abdominal scan of the liver acquired with the 6C2 transducer on the Touch Prime Ultrasound System. This image illustrates superb spatial and contrast resolution throughout the entire image with uncompromised clinical detail. Clinical Benefits of SynTek Architecture • Uniform lateral resolution over the entire depth, making position of transmit focus less critical • High frame rate while maintaining the spatial resolution • Improved penetration (e.g. for deep abdominal imaging)
  • 3. Technology White Paper | CARESTREAM Touch Prime Ultrasound System 3 Smart System Control (SSC) Another path-breaking technology aimed at ultrasound workflow enhancement available on the Touch Prime system that harnesses the power of GPUs and parallel computing is Smart System Control (SSC). It is a real-time optimization technique implemented on the Touch Prime system to automatically adjust over 25 different imaging parameters (including some that are not under user control) to provide the user with optimal image quality. It essentially provides automatic system tuning beyond traditional keyboard/user controls. SSC technology has detailed insight and control of extended sets of parameters and their dependencies. SSC technology can be used in B-mode, Color mode and Doppler imaging modes. To determine the optimal settings, the algorithm takes as input the user preference for frame rate or resolution and optimizes other system parameters automatically. The optimization is performed continuously in the background in real-time, and does not require the user to initiate it from the user console for each scan. The following are examples of system parameters that SSC automatically adjusts during the system operation: Ultrasound Mode Parameters B-mode • Line density • Number of focal zones • Number of transmit beams • Number of Compounding angles B-mode + Color (Duplex) • B-mode & Color line density • Number of pulses transmitted in each waveform packet (shots per estimate) SSC technology ensures constantly optimized images with limited user interaction, which results in improved workflow and diagnostic confidence. It has the potential to reduce exam times because there is no need to adjust the imaging parameters manually. The result is an optimized image with fewer touch-panel hard and soft key presses by the user. Taking it further, the intent is that for routine applications, SSC technology can lead to out-of-the-box optimized images with minimal manual changes needed from the user. Beyond Imaging Performance With the power of SynTek and GPU processing, the CARESETREAM Touch Prime Ultrasound System is only using a small fraction of its potential power at first release. In conjunction with its programmable touch user interface, the system is an ideal platform for upgrades without the need for major hardware changes. As Carestream continues to innovate with enhancements to imaging performance and new features, the system lends itself to easy upgradeability and has power for further technology-driven innovative applications. The following are examples of such innovative applications available at first release. Smart Flow Imaging To assess blood flow in arteries and veins, Doppler ultrasound uses the frequency shift that occurs when sound waves bounce off a moving object. Such images are used to uncover blockages and clots, the narrowing of blood vessels and congenital vascular malformations. In color Doppler imaging, echo measurements are displayed as colors that indicate the speed and direction of blood flow. They are used to identify narrowed blood vessels, for example, and the tiny “jets” of blood associated with vascular anomalies. Power Doppler imaging is even more sensitive than color Doppler, visualizing blood flow through tiny vessels such as those that feed tumors in the thyroid and scrotum, as well as lesions just below the skin. Spectral Doppler technique calculates and then graphs the velocity of blood flow according to the distance blood travels over time. Ordinary color and spectral Doppler ultrasound measures the velocity of flow components toward or away from the transducer. It uses the angle of the ultrasound beam relative to the flow direction to calculate the actual flow velocity through the vessel. In current conventional implementations, the accuracy of Doppler computations depends on precise knowledge of the direction of the ultrasound beam and direction of the flow in the vessel (and the angle α between them).
  • 4. Technology White Paper | CARESTREAM Touch Prime Ultrasound System 4 As the following Doppler equation indicates, when the ultrasound beam is perpendicular to the vessel (90°), this computation is impossible, because there is no flow component in the direction of the beam. In effect, the measurement is impossible when the insonation angle is over 60°, because then small errors in measuring the two directions lead to large discrepancies in the results. Smart Flow imaging is a ground-breaking technology available on the Touch Prime Ultrasound System that has the potential to revolutionize the workflow for many Doppler ultrasound applications. In addition to revealing flow pattern details, it saves valuable time and makes the exam much easier. The new proprietary Smart Flow method visualizes blood flow in all directions, independent of imaging angle. It uses a technique called transverse oscillation to overcome the angle limitations of ordinary Doppler ultrasound. It creates an effective component of the ultrasound oscillation perpendicular to the transmitting beam – an oscillation in the transverse direction (Jensen JA 2001; Jensen JA, Munk P 1998). Smart Flow technology generates a 2D interference pattern in the received ultrasound signal. This allows the system to calculate not only the axial component of the velocity (as traditional color Doppler), but also the transverse component. Therefore, Smart Flow technology eliminates the angle dependence and enables both the detection and visualization of complex flows. To display the information from Smart Flow technology on the Touch Prime platform, color coding and/or arrows are used. The length of the arrow, in addition to the color, indicates the magnitude, and the orientation of the arrow indicates the flow direction. The figures below illustrate a side-by-side comparison between conventional color flow imaging and the new Smart Flow technology in a carotid artery scan. The conventional color flow (left) produces drop-outs (regions with no flow) at locations within the vessel where the flow is oriented perpendicular to the acoustic beam (with no steering or angling of the color box). For the same anatomy, the Smart Flow image (right) displays continuous and robust flow information throughout the entire lumen, even when the flow is perpendicular to the acoustic beam. Comparison between conventional color flow (left) vs. Smart Flow (right) imaging in a carotid artery scan
  • 5. Technology White Paper | CARESTREAM Touch Prime Ultrasound System 5 The figure below illustrates a Smart Flow image of the jugular vein (top) and carotid artery (bottom) in a normal volunteer acquired with an 8L2 transducer. The arrows indicate opposite flow directions, with longer arrow lengths towards the middle of the carotid artery lumen representing higher velocities. Smart Flow image of a jugular vein and carotid artery with the arrows representing direction and velocity magnitude In conclusion, Smart Flow technology provides an intuitive visual representation of flow in all directions, making it well suited to visualize transverse flow and complex flow patterns including turbulence. The measurements are angle- independent. It enables a faster workflow since the overall hemodynamic view is available without having to steer. The results are more consistent with less clinician dependence. Smart Flow Assist Spectral Doppler imaging is commonly performed to quantify the flow profiles in the blood vessels. The measurement is displayed as a waveform trace, depicting the velocity distribution (profile) at the given spatial location as a function of time. The typical steps involved in obtaining a velocity profile include: 1) Turn on Color Flow mode to obtain the orientation of the vessel, 2) Turn on Spectral Doppler mode, 3) Update the beam steering direction, 4) Move the gate to obtain the highest velocity, and 5) Angle-correct the gate. The above steps require manual effort by the user, and moreover steps 3 to 5 have to be repeated with movement of the transducer. Contrary to this conventional mode of operation, the Smart Flow Assist technology eliminates the need for repeated manual adjustments. With the proprietary functions in Smart Flow Assist technology, the software automatically updates the beam steering, gate position and angle correction. It is updated at all times, even when the transducer is moved. The benefits of Smart Flow Assist technology are even more pronounced when measuring volume flow (VF). In this situation, the workflow to obtain VF is reduced from 10 steps (for a VF on a frozen image - and 10 more for every new location) to 2 steps (and one for every new location).
  • 6. Technology White Paper | CARESTREAM Touch Prime Ultrasound System carestream.com ©Carestream Health, Inc., 2015. CARESTREAM is a trademark of Carestream Health. CAT 200 0114 07/15 Conclusion The CARESTREAM Touch Prime Ultrasound System provides several advanced ultrasound technologies, leveraging advanced GPU processing power for enhanced imaging performance. It incorporates a next-generation beamforming approach, leading to simultaneous improvement in frame rate and spatial resolution compared to conventional systems and automation of ultrasound scan settings that results in improved workflow efficiency. These novel technologies, when combined with exceptional user experience features such as an all-Touch control panel, a thoughtful system design around ergonomics, and one-touch transducer activation, makes the Touch Prime a system of choice for Radiology, OB/GYN, Musculoskeletal (MSK), Vascular and general cardiac imaging applications. Ajay Anand is a member of the ultrasound R&D team at Carestream Health. He has more than 10 years’ experience leading the development of novel ultrasound technologies, and is a co-inventor of more than 20 patent filings in the field of medical ultrasound. References Jensen JA, “A new estimator for vector velocity estimation” IEEE Trans Ultrason Ferroelectr Freq Control 2001; 48 (4): 886-894. Jensen JA, Munk P “A new method for estimation of velocity vectors”, IEEE Trans Ultrason Ferroelectr Freq Control 1998; 45 (3): 837-51.