MRI ARTIFACTS
BY:MS.AALIA ABDULLAH
ASSISTANT PROFESSOR RADIOLOGY AND IMAGING
MEWAR UNIVERSITY RAJASTHAN
Artifacts
• Artifacts are features that appear on a reconstructed image but are
not part of the original anatomy.
• They may make the image unclear and harder to interpret, or they
can mimic the appearance of pathology.
• The radiographer’s aim is to minimize artifacts. However, some
artifacts have clinical/diagnostic uses.
Movement/Physiological Artifacts
• Motion artifact due to physical movements
• Motion artifact in MRI refers to the image degradation caused by
patient or object motion during image acquisition. It can result in
blurred, distorted, or ghosted images, and can significantly affect
image quality and diagnostic accuracy.
CORRECTION
• Give clear instructions to patient before the scan
• Constantly communicate to claustrophobic patients
• Make the patient comfortable
• Use immobilization equipment or paddings
• Feed and wrap the infants
• Use oral or IV sedation
• Perform the scans under GA
• reduce scan time using fast scan techniques
• Adjust phase and frequency encoding directions to project artifact
away from anatomy
• Use compressed sensing techniques
Aliasing / Wrap around artifact
• Aliasing, also known as wrap-around artifact, is a common artifact
that can occur in magnetic resonance imaging (MRI) when the field of
view (FOV) is smaller than the actual object being imaged.
• Spatial localisation is determined by the frequency of the spins
within a selected sample of tissue.
• In 3D imaging phase encoding is used to select sections of tissue. If
the anatomy extends beyond the FOV, a similar process may cause
wrap-over in the slab select direction between slices at either end of
the volume.
CORRECTION
• Increase field of view (FOV):
• using surface coils
• applying saturation bands.
MRI Zipper Artifact
• Zipper artifact is a type of artifact in MRI images that appears as a
series of bright and dark stripes that run parallel to the phase-
encoding direction.
• It is caused by a timing error in the gradient system that causes the
gradient to switch on too early or too late, leading to phase errors in
the acquired signals.
• There can be a range of hardware and software problems that could
contribute to the zipper artifact in MRI scans
• One common factor is inadequate RF shielding within the scanner
room, which could result from factors such as a door being left open,
a faulty door seal, or interference from electrical equipment like a
satellite monitor. Additionally, the zipper artifact could arise from
faults in the MRI system itself or issues related to the mains electricity
supply.
• Effect on image: Zipper artifact is characterized by the presence of
stripes or bands across the image that appear as bright and dark lines
that resemble the teeth of a zipper.
• These stripes or bands can appear parallel or oblique to the imaging
plane and can affect image quality and diagnostic accuracy.
CORRECTION
• Check door is closed, check door seal is intact, remove electrical
equipment, check for blown bulbs in scanner room
• Use MRI compatible monitoring equipment
• Call engineer if no obvious solution is found
• Regularly perform quality control tests to identify and address any
issues with the scanner’s hardware.
Blooming Artifact
• Blooming artifact is a type of artifact that occurs in magnetic
resonance imaging (MRI) when there are regions of high magnetic
susceptibility differences.
• It appears as signal spreading or blooming beyond the actual
boundaries of an object
• Susceptibility Weighted Imaging (SWI) is a sequence that is
particularly sensitive to magnetic susceptibility differences, making it
useful for visualizing blood products, including hemorrhages.
However, SWI can be susceptible to blooming artifact, which occurs
when the signal from blood products spreads beyond their actual
boundaries, leading to signal loss and distortion in surrounding
structures.
CORRECTION
• Sequence selection: Choosing the appropriate MRI sequence is
crucial. Gradient-echo sequences, such as susceptibility-weighted
imaging (SWI) or T2*-weighted sequences, are commonly used to
detect hemorrhages due to their sensitivity to blood products.
However, they are also more susceptible to blooming artifact. Spin-
echo sequences, such as T1-weighted or T2-weighted sequences, may
be preferred when blooming artifact is a concern.
Flow Artifacts
• Flow artifacts in MRI refer to distortions or signal changes in the images
caused by the movement of fluids, such as cerebrospinal fluid (CSF) and
blood flow.
• These artifacts can result in blurring, ghosting, or signal loss in the images,
making it challenging to accurately interpret the anatomical structures.
CORRECTION
• Lower TE can also be used, which will reduce the time between the
90° and 180° pulses and decrease the likelihood of spins moving in
and out of the slice, thereby reducing the occurrence of flow void.
• NEX (Number of Excitations): NEX refers to the number of times the
MRI scan is repeated to acquire multiple signal averages. Increasing
the NEX can help improve the signal-to-noise ratio (SNR) by reducing
random noise. This, in turn, can enhance image quality and reduce
flow artifacts.
MRI ARTIFACTS.pptx

More Related Content

What's hot

Sensitometry3
Sensitometry3Sensitometry3
Sensitometry3
mr_koky
 
ultrasound transducers and resolution
ultrasound transducers and resolutionultrasound transducers and resolution
ultrasound transducers and resolution
Vallabhaneni Bhupal
 

What's hot (20)

Elastography
ElastographyElastography
Elastography
 
Mri components
Mri componentsMri components
Mri components
 
MRI Safety Basics
MRI Safety BasicsMRI Safety Basics
MRI Safety Basics
 
Basic Pulse Sequences In MRI
Basic Pulse Sequences In MRIBasic Pulse Sequences In MRI
Basic Pulse Sequences In MRI
 
MRI Parameters for Imaging
MRI Parameters for ImagingMRI Parameters for Imaging
MRI Parameters for Imaging
 
teleradiology
teleradiologyteleradiology
teleradiology
 
Chapter 4 image display
Chapter 4 image displayChapter 4 image display
Chapter 4 image display
 
Interactions of ultrasound with matter
Interactions of ultrasound with matterInteractions of ultrasound with matter
Interactions of ultrasound with matter
 
K space and parallel imaging
K space and parallel imagingK space and parallel imaging
K space and parallel imaging
 
Sensitometry3
Sensitometry3Sensitometry3
Sensitometry3
 
ultrasound transducers and resolution
ultrasound transducers and resolutionultrasound transducers and resolution
ultrasound transducers and resolution
 
Gamma Camera
Gamma CameraGamma Camera
Gamma Camera
 
Mri gradient coils
Mri gradient coilsMri gradient coils
Mri gradient coils
 
MRI Artefacts
MRI ArtefactsMRI Artefacts
MRI Artefacts
 
Ct image artifacts
Ct image artifactsCt image artifacts
Ct image artifacts
 
Image Quality, Artifacts and it's Remedies in CT-Avinesh Shrestha
Image Quality, Artifacts and it's Remedies in CT-Avinesh ShresthaImage Quality, Artifacts and it's Remedies in CT-Avinesh Shrestha
Image Quality, Artifacts and it's Remedies in CT-Avinesh Shrestha
 
Teleradiology
TeleradiologyTeleradiology
Teleradiology
 
MR IMAGE QUALITY PARAMETERS.pptx
MR IMAGE QUALITY PARAMETERS.pptxMR IMAGE QUALITY PARAMETERS.pptx
MR IMAGE QUALITY PARAMETERS.pptx
 
Chapter 6 image quality in ct
Chapter 6 image quality in ct Chapter 6 image quality in ct
Chapter 6 image quality in ct
 
CT Angiography Head and Neck
CT Angiography Head and NeckCT Angiography Head and Neck
CT Angiography Head and Neck
 

Similar to MRI ARTIFACTS.pptx

artifactsusgctmriradiologypk-190830131406.pptx
artifactsusgctmriradiologypk-190830131406.pptxartifactsusgctmriradiologypk-190830131406.pptx
artifactsusgctmriradiologypk-190830131406.pptx
Shubham661884
 
New Technology: 核磁共振相容節律器 - ”MRI Standard of Care for Pacemaker”)_20130914中區
New Technology:  核磁共振相容節律器 - ”MRI Standard of Care for Pacemaker”)_20130914中區New Technology:  核磁共振相容節律器 - ”MRI Standard of Care for Pacemaker”)_20130914中區
New Technology: 核磁共振相容節律器 - ”MRI Standard of Care for Pacemaker”)_20130914中區
Taiwan Heart Rhythm Society
 
X -RADIATION or RONTGEN RADIATION (X-RAY)
X -RADIATION or RONTGEN RADIATION(X-RAY)X -RADIATION or RONTGEN RADIATION(X-RAY)
X -RADIATION or RONTGEN RADIATION (X-RAY)
Ma Wady
 

Similar to MRI ARTIFACTS.pptx (20)

MRI Artifact Radiology
MRI Artifact RadiologyMRI Artifact Radiology
MRI Artifact Radiology
 
MRI BRAIN
MRI BRAINMRI BRAIN
MRI BRAIN
 
The science of refractive surgery by Dr. Iddi.pptx
The science of refractive surgery by Dr. Iddi.pptxThe science of refractive surgery by Dr. Iddi.pptx
The science of refractive surgery by Dr. Iddi.pptx
 
artifactsusgctmriradiologypk-190830131406.pptx
artifactsusgctmriradiologypk-190830131406.pptxartifactsusgctmriradiologypk-190830131406.pptx
artifactsusgctmriradiologypk-190830131406.pptx
 
MRI basics
MRI basicsMRI basics
MRI basics
 
IVUS
IVUSIVUS
IVUS
 
MRI Procedure and Preparation
MRI Procedure and PreparationMRI Procedure and Preparation
MRI Procedure and Preparation
 
Basics of MRI
Basics of MRIBasics of MRI
Basics of MRI
 
Radiographic system clinical engineering
Radiographic system clinical engineeringRadiographic system clinical engineering
Radiographic system clinical engineering
 
Artifacts in conventional radiology.pptx
Artifacts in conventional radiology.pptxArtifacts in conventional radiology.pptx
Artifacts in conventional radiology.pptx
 
Radiographic Considerations in Implants
Radiographic Considerations in ImplantsRadiographic Considerations in Implants
Radiographic Considerations in Implants
 
New Technology: 核磁共振相容節律器 - ”MRI Standard of Care for Pacemaker”)_20130914中區
New Technology:  核磁共振相容節律器 - ”MRI Standard of Care for Pacemaker”)_20130914中區New Technology:  核磁共振相容節律器 - ”MRI Standard of Care for Pacemaker”)_20130914中區
New Technology: 核磁共振相容節律器 - ”MRI Standard of Care for Pacemaker”)_20130914中區
 
Radiology
RadiologyRadiology
Radiology
 
Power Doppler,MRS,SWI.pptx
Power Doppler,MRS,SWI.pptxPower Doppler,MRS,SWI.pptx
Power Doppler,MRS,SWI.pptx
 
Ct artiact
Ct artiactCt artiact
Ct artiact
 
X -RADIATION or RONTGEN RADIATION (X-RAY)
X -RADIATION or RONTGEN RADIATION(X-RAY)X -RADIATION or RONTGEN RADIATION(X-RAY)
X -RADIATION or RONTGEN RADIATION (X-RAY)
 
Magnetic Resonance Imaging (MRI)
Magnetic Resonance Imaging (MRI)Magnetic Resonance Imaging (MRI)
Magnetic Resonance Imaging (MRI)
 
MRI PROCEDURE OF BRAIN
MRI PROCEDURE OF BRAIN MRI PROCEDURE OF BRAIN
MRI PROCEDURE OF BRAIN
 
Mri
MriMri
Mri
 
Imaging in implants
Imaging in implantsImaging in implants
Imaging in implants
 

More from AALIA ABDULLAH

More from AALIA ABDULLAH (12)

cardiac multislice CT.prospective ECG ,retrospective ECG Gating.pptx
cardiac multislice CT.prospective ECG ,retrospective ECG Gating.pptxcardiac multislice CT.prospective ECG ,retrospective ECG Gating.pptx
cardiac multislice CT.prospective ECG ,retrospective ECG Gating.pptx
 
Transformer.pptx
Transformer.pptxTransformer.pptx
Transformer.pptx
 
RADIOLOGY PROTECTION.pptx
RADIOLOGY PROTECTION.pptxRADIOLOGY PROTECTION.pptx
RADIOLOGY PROTECTION.pptx
 
Abdomen ultrasound.pptx
Abdomen ultrasound.pptxAbdomen ultrasound.pptx
Abdomen ultrasound.pptx
 
ULTRASOUND ARTIFACTS.pptx
ULTRASOUND ARTIFACTS.pptxULTRASOUND ARTIFACTS.pptx
ULTRASOUND ARTIFACTS.pptx
 
ULtrasound.pptx
ULtrasound.pptxULtrasound.pptx
ULtrasound.pptx
 
MRI Contrast And Plain.pptx
MRI Contrast And Plain.pptxMRI Contrast And Plain.pptx
MRI Contrast And Plain.pptx
 
MObile and portable radiography.pptx
MObile and portable radiography.pptxMObile and portable radiography.pptx
MObile and portable radiography.pptx
 
ATOMIC ENERGY REGULATORY BOARD (AERB).pptx
ATOMIC ENERGY REGULATORY BOARD (AERB).pptxATOMIC ENERGY REGULATORY BOARD (AERB).pptx
ATOMIC ENERGY REGULATORY BOARD (AERB).pptx
 
ULTRASOUND EQUIPMENT
ULTRASOUND EQUIPMENTULTRASOUND EQUIPMENT
ULTRASOUND EQUIPMENT
 
Contrast media OF MRI.pptx
Contrast media OF MRI.pptxContrast media OF MRI.pptx
Contrast media OF MRI.pptx
 
Physical And Physiological Basis Of Magnetic Relaxation, by AALIA
Physical And Physiological Basis Of Magnetic Relaxation, by AALIAPhysical And Physiological Basis Of Magnetic Relaxation, by AALIA
Physical And Physiological Basis Of Magnetic Relaxation, by AALIA
 

Recently uploaded

Cara Menggugurkan Kandungan Dengan Cepat Selesai Dalam 24 Jam Secara Alami Bu...
Cara Menggugurkan Kandungan Dengan Cepat Selesai Dalam 24 Jam Secara Alami Bu...Cara Menggugurkan Kandungan Dengan Cepat Selesai Dalam 24 Jam Secara Alami Bu...
Cara Menggugurkan Kandungan Dengan Cepat Selesai Dalam 24 Jam Secara Alami Bu...
Cara Menggugurkan Kandungan 087776558899
 
Physiologic Anatomy of Heart_AntiCopy.pdf
Physiologic Anatomy of Heart_AntiCopy.pdfPhysiologic Anatomy of Heart_AntiCopy.pdf
Physiologic Anatomy of Heart_AntiCopy.pdf
MedicoseAcademics
 
Difference Between Skeletal Smooth and Cardiac Muscles
Difference Between Skeletal Smooth and Cardiac MusclesDifference Between Skeletal Smooth and Cardiac Muscles
Difference Between Skeletal Smooth and Cardiac Muscles
MedicoseAcademics
 
Obat Aborsi Ampuh Usia 1,2,3,4,5,6,7 Bulan 081901222272 Obat Penggugur Kandu...
Obat Aborsi Ampuh Usia 1,2,3,4,5,6,7 Bulan  081901222272 Obat Penggugur Kandu...Obat Aborsi Ampuh Usia 1,2,3,4,5,6,7 Bulan  081901222272 Obat Penggugur Kandu...
Obat Aborsi Ampuh Usia 1,2,3,4,5,6,7 Bulan 081901222272 Obat Penggugur Kandu...
Halo Docter
 

Recently uploaded (20)

VIP ℂall Girls Arekere Bangalore 6378878445 WhatsApp: Me All Time Serviℂe Ava...
VIP ℂall Girls Arekere Bangalore 6378878445 WhatsApp: Me All Time Serviℂe Ava...VIP ℂall Girls Arekere Bangalore 6378878445 WhatsApp: Me All Time Serviℂe Ava...
VIP ℂall Girls Arekere Bangalore 6378878445 WhatsApp: Me All Time Serviℂe Ava...
 
Cara Menggugurkan Kandungan Dengan Cepat Selesai Dalam 24 Jam Secara Alami Bu...
Cara Menggugurkan Kandungan Dengan Cepat Selesai Dalam 24 Jam Secara Alami Bu...Cara Menggugurkan Kandungan Dengan Cepat Selesai Dalam 24 Jam Secara Alami Bu...
Cara Menggugurkan Kandungan Dengan Cepat Selesai Dalam 24 Jam Secara Alami Bu...
 
Physicochemical properties (descriptors) in QSAR.pdf
Physicochemical properties (descriptors) in QSAR.pdfPhysicochemical properties (descriptors) in QSAR.pdf
Physicochemical properties (descriptors) in QSAR.pdf
 
ABO Blood grouping in-compatibility in pregnancy
ABO Blood grouping in-compatibility in pregnancyABO Blood grouping in-compatibility in pregnancy
ABO Blood grouping in-compatibility in pregnancy
 
Part I - Anticipatory Grief: Experiencing grief before the loss has happened
Part I - Anticipatory Grief: Experiencing grief before the loss has happenedPart I - Anticipatory Grief: Experiencing grief before the loss has happened
Part I - Anticipatory Grief: Experiencing grief before the loss has happened
 
Physiologic Anatomy of Heart_AntiCopy.pdf
Physiologic Anatomy of Heart_AntiCopy.pdfPhysiologic Anatomy of Heart_AntiCopy.pdf
Physiologic Anatomy of Heart_AntiCopy.pdf
 
7 steps How to prevent Thalassemia : Dr Sharda Jain & Vandana Gupta
7 steps How to prevent Thalassemia : Dr Sharda Jain & Vandana Gupta7 steps How to prevent Thalassemia : Dr Sharda Jain & Vandana Gupta
7 steps How to prevent Thalassemia : Dr Sharda Jain & Vandana Gupta
 
Intro to disinformation and public health
Intro to disinformation and public healthIntro to disinformation and public health
Intro to disinformation and public health
 
TEST BANK For Guyton and Hall Textbook of Medical Physiology, 14th Edition by...
TEST BANK For Guyton and Hall Textbook of Medical Physiology, 14th Edition by...TEST BANK For Guyton and Hall Textbook of Medical Physiology, 14th Edition by...
TEST BANK For Guyton and Hall Textbook of Medical Physiology, 14th Edition by...
 
Top 10 Most Beautiful Chinese Pornstars List 2024
Top 10 Most Beautiful Chinese Pornstars List 2024Top 10 Most Beautiful Chinese Pornstars List 2024
Top 10 Most Beautiful Chinese Pornstars List 2024
 
Difference Between Skeletal Smooth and Cardiac Muscles
Difference Between Skeletal Smooth and Cardiac MusclesDifference Between Skeletal Smooth and Cardiac Muscles
Difference Between Skeletal Smooth and Cardiac Muscles
 
Face and Muscles of facial expression.pptx
Face and Muscles of facial expression.pptxFace and Muscles of facial expression.pptx
Face and Muscles of facial expression.pptx
 
Obat Aborsi Ampuh Usia 1,2,3,4,5,6,7 Bulan 081901222272 Obat Penggugur Kandu...
Obat Aborsi Ampuh Usia 1,2,3,4,5,6,7 Bulan  081901222272 Obat Penggugur Kandu...Obat Aborsi Ampuh Usia 1,2,3,4,5,6,7 Bulan  081901222272 Obat Penggugur Kandu...
Obat Aborsi Ampuh Usia 1,2,3,4,5,6,7 Bulan 081901222272 Obat Penggugur Kandu...
 
Top 10 Most Beautiful Russian Pornstars List 2024
Top 10 Most Beautiful Russian Pornstars List 2024Top 10 Most Beautiful Russian Pornstars List 2024
Top 10 Most Beautiful Russian Pornstars List 2024
 
TEST BANK For Porth's Essentials of Pathophysiology, 5th Edition by Tommie L ...
TEST BANK For Porth's Essentials of Pathophysiology, 5th Edition by Tommie L ...TEST BANK For Porth's Essentials of Pathophysiology, 5th Edition by Tommie L ...
TEST BANK For Porth's Essentials of Pathophysiology, 5th Edition by Tommie L ...
 
Creeping Stroke - Venous thrombosis presenting with pc-stroke.pptx
Creeping Stroke - Venous thrombosis presenting with pc-stroke.pptxCreeping Stroke - Venous thrombosis presenting with pc-stroke.pptx
Creeping Stroke - Venous thrombosis presenting with pc-stroke.pptx
 
SEMESTER-V CHILD HEALTH NURSING-UNIT-1-INTRODUCTION.pdf
SEMESTER-V CHILD HEALTH NURSING-UNIT-1-INTRODUCTION.pdfSEMESTER-V CHILD HEALTH NURSING-UNIT-1-INTRODUCTION.pdf
SEMESTER-V CHILD HEALTH NURSING-UNIT-1-INTRODUCTION.pdf
 
Test bank for critical care nursing a holistic approach 11th edition morton f...
Test bank for critical care nursing a holistic approach 11th edition morton f...Test bank for critical care nursing a holistic approach 11th edition morton f...
Test bank for critical care nursing a holistic approach 11th edition morton f...
 
spinal cord disorders and paraplegia .
spinal cord disorders  and  paraplegia .spinal cord disorders  and  paraplegia .
spinal cord disorders and paraplegia .
 
VIP ℂall Girls Thane West Mumbai 9930245274 WhatsApp: Me All Time Serviℂe Ava...
VIP ℂall Girls Thane West Mumbai 9930245274 WhatsApp: Me All Time Serviℂe Ava...VIP ℂall Girls Thane West Mumbai 9930245274 WhatsApp: Me All Time Serviℂe Ava...
VIP ℂall Girls Thane West Mumbai 9930245274 WhatsApp: Me All Time Serviℂe Ava...
 

MRI ARTIFACTS.pptx

  • 1. MRI ARTIFACTS BY:MS.AALIA ABDULLAH ASSISTANT PROFESSOR RADIOLOGY AND IMAGING MEWAR UNIVERSITY RAJASTHAN
  • 2. Artifacts • Artifacts are features that appear on a reconstructed image but are not part of the original anatomy. • They may make the image unclear and harder to interpret, or they can mimic the appearance of pathology. • The radiographer’s aim is to minimize artifacts. However, some artifacts have clinical/diagnostic uses.
  • 3. Movement/Physiological Artifacts • Motion artifact due to physical movements • Motion artifact in MRI refers to the image degradation caused by patient or object motion during image acquisition. It can result in blurred, distorted, or ghosted images, and can significantly affect image quality and diagnostic accuracy.
  • 4.
  • 5. CORRECTION • Give clear instructions to patient before the scan • Constantly communicate to claustrophobic patients • Make the patient comfortable • Use immobilization equipment or paddings • Feed and wrap the infants • Use oral or IV sedation • Perform the scans under GA • reduce scan time using fast scan techniques • Adjust phase and frequency encoding directions to project artifact away from anatomy • Use compressed sensing techniques
  • 6. Aliasing / Wrap around artifact • Aliasing, also known as wrap-around artifact, is a common artifact that can occur in magnetic resonance imaging (MRI) when the field of view (FOV) is smaller than the actual object being imaged. • Spatial localisation is determined by the frequency of the spins within a selected sample of tissue. • In 3D imaging phase encoding is used to select sections of tissue. If the anatomy extends beyond the FOV, a similar process may cause wrap-over in the slab select direction between slices at either end of the volume.
  • 7.
  • 8. CORRECTION • Increase field of view (FOV): • using surface coils • applying saturation bands.
  • 9. MRI Zipper Artifact • Zipper artifact is a type of artifact in MRI images that appears as a series of bright and dark stripes that run parallel to the phase- encoding direction. • It is caused by a timing error in the gradient system that causes the gradient to switch on too early or too late, leading to phase errors in the acquired signals. • There can be a range of hardware and software problems that could contribute to the zipper artifact in MRI scans
  • 10. • One common factor is inadequate RF shielding within the scanner room, which could result from factors such as a door being left open, a faulty door seal, or interference from electrical equipment like a satellite monitor. Additionally, the zipper artifact could arise from faults in the MRI system itself or issues related to the mains electricity supply.
  • 11. • Effect on image: Zipper artifact is characterized by the presence of stripes or bands across the image that appear as bright and dark lines that resemble the teeth of a zipper. • These stripes or bands can appear parallel or oblique to the imaging plane and can affect image quality and diagnostic accuracy.
  • 12.
  • 13. CORRECTION • Check door is closed, check door seal is intact, remove electrical equipment, check for blown bulbs in scanner room • Use MRI compatible monitoring equipment • Call engineer if no obvious solution is found • Regularly perform quality control tests to identify and address any issues with the scanner’s hardware.
  • 14. Blooming Artifact • Blooming artifact is a type of artifact that occurs in magnetic resonance imaging (MRI) when there are regions of high magnetic susceptibility differences. • It appears as signal spreading or blooming beyond the actual boundaries of an object
  • 15.
  • 16. • Susceptibility Weighted Imaging (SWI) is a sequence that is particularly sensitive to magnetic susceptibility differences, making it useful for visualizing blood products, including hemorrhages. However, SWI can be susceptible to blooming artifact, which occurs when the signal from blood products spreads beyond their actual boundaries, leading to signal loss and distortion in surrounding structures.
  • 17. CORRECTION • Sequence selection: Choosing the appropriate MRI sequence is crucial. Gradient-echo sequences, such as susceptibility-weighted imaging (SWI) or T2*-weighted sequences, are commonly used to detect hemorrhages due to their sensitivity to blood products. However, they are also more susceptible to blooming artifact. Spin- echo sequences, such as T1-weighted or T2-weighted sequences, may be preferred when blooming artifact is a concern.
  • 18. Flow Artifacts • Flow artifacts in MRI refer to distortions or signal changes in the images caused by the movement of fluids, such as cerebrospinal fluid (CSF) and blood flow. • These artifacts can result in blurring, ghosting, or signal loss in the images, making it challenging to accurately interpret the anatomical structures.
  • 19.
  • 20. CORRECTION • Lower TE can also be used, which will reduce the time between the 90° and 180° pulses and decrease the likelihood of spins moving in and out of the slice, thereby reducing the occurrence of flow void. • NEX (Number of Excitations): NEX refers to the number of times the MRI scan is repeated to acquire multiple signal averages. Increasing the NEX can help improve the signal-to-noise ratio (SNR) by reducing random noise. This, in turn, can enhance image quality and reduce flow artifacts.