SlideShare a Scribd company logo
1 of 32
METHOD 
VALIDATION 
BY:- 
DR. HINAL SHAH
INTRODUCTION 
 Method selection and validation – Key steps in the process of 
implementing new methods in the clinical laboratory. 
 In Clinical Chemistry Laboratory- influenced strongly by guidelines.
 The Clinical & Laboratory Standards Institute (formerly National Committee 
for Clinical Laboratory Standards) and International Organization for 
Standardization have published a series of protocols and documents to be 
followed by CCL and manufacturers for method validation. 
 Also, laboratory accreditation requirements have to be met.
DEFINITIONS 
 “The suite of procedures to which an analytical method is subjected to provide 
objective evidence that the method, if used in the manner specified, will 
produce results that conform to the statement of the method validation 
parameters” 
 “The process used to confirm that the analytical procedure employed for a 
specific test is suitable for its intended use” 
 ISO/IEC 17025 (ISO/IEC 2005, section 5.4) states that method validation is 
“confirmation by examination and provision of objective evidence that the 
particular requirements for a specified intended use are fulfilled”
When Should Methods Be Validated? 
 No method should be used without validation 
 ISO/IEC 17025 encourages the use of methods published in international, 
regional, or national standards and implies that if these methods are used 
without deviation, then the validation requirements have been satisfied. 
 What does need to be validated are-i. 
Nonstandard methods, designed and developed by individual laboratories 
ii. Standard methods used outside their scope 
iii. Amplifications and modifications of standard methods.
Method selection/ 
development and 
validation cycle
VERIFICATION 
 Synonymous with Single Laboratory Validation 
 To be done when a laboratory uses a method for the first time (with 
standard materials) 
 Also done when a particular aspect of the method or its implementation is 
changed, for example-a) 
New analyst 
b) New equipment or equipment part 
c) New batch of reagent 
d) Changes in the laboratory premises
 Minimum verification is to analyze a material before and after the change 
and check for consistency of the results, both in terms of mean and 
standard deviation. 
 Basic experiments that should be performed to verify the use of a standard 
method before first use in a laboratory: 
 Bias/recovery 
 Precision 
 Measurement uncertainty 
 Calibration model (linearity) 
 Limit of detection
METHOD SELECTION- CRITERIA 
1. Medical Need 
2. Analytical Performance 
3. Practical Criteria- depend on whether it is a Laboratory Developed 
Test or a Commercial kit method.
Method Validation- 
Requirements 
I. Calibration 
II. Precision 
III. Accuracy and Trueness 
IV. Linearity 
V. Limit of Detection (LOD) 
VI. Analytical/Measuring Range 
VII. Analytical Sensitivity 
VIII. Analytical Specificity and Interference 
IX. Ruggedness and Robustness 
X. Documentation
I. CALIBRATION 
o Established by measurement of samples with known quantities of the analyte 
(calibrators). [Only certified reference materials with known traceability should 
be used] 
o Done by- Chemical standards or samples with known quantities of analyte 
present in the typical matrix. 
o Calibration functions- may be linear or curved. 
o If the assumed calibration function does not correctly reflect the true 
relationship between instrument response & analyte concentration, a systemic 
error or bias is likely to be associated with the analytical method.
II. PRECISION 
 Precision has been defined as the closeness of agreement between 
independent results of measurements obtained under stipulated 
conditions. 
 Depends on – the stability of the instrument response for a given quantity 
of analyte. 
 The degree of precision is usually expressed on the basis of statistical 
measures of imprecision, such as Standard Deviation(SD) OR Coefficient 
of Variation(CV)
 If calibration function is linear and imprecision remains same over 
analytical measurement range, SD also tends to be constant over 
analytical measurement range. If imprecision increases proportionally, SD 
also tends to increase proportionally to the concentration and relative 
imprecision(i.e. CV) remains constant. 
 Imprecision of measurements is solely related to the random error. 
 Precision is specified as follows:-
1) Repeatability: Corresponds to with-in run precision. It is closeness of 
agreement between results of successive measurements carried out 
under the same conditions. 
2) Reproducibility: Closeness of agreement between results of 
measurements performed under changed conditions of measurements. 
2 specifications of reproducibility – 
 Total or between-run precision (intermediate precision):- its method of 
assessment depends on usage of method and variation in 
analyst/equipment. 
 Inter-laboratory precision.
 An important aspect of a full method validation is estimating bias 
components attributable to the method itself and to the laboratory carrying 
out the analysis. It can be evaluated by inter-laboratory studies.
III. TRUENESS & ACCURACY 
 Trueness:- 
• Closeness of agreement between the mean value (obtained from a large 
series of results of measurements) & the true value. 
• Bias – Difference between the mean and the true value. 
Inversely related to the trueness. 
• Qualitative term. 
• True value – in practice is the accepted reference value. 
• Evaluated by comparison of measurements by a given routine method & a 
reference measurement procedure.
 Accuracy :- 
• Closeness of agreement between the result of a measurement and true 
value. 
• Qualitative term. 
• Influenced by- BIAS and IMPRECISION. Reflects total error. 
• Total error= bias + 2SD 
Concepts of recovery, drift and carryover are related to trueness.
RECOVERY 
 It means - the fraction of a test material in a matrix that can be quantified. 
target conc. – conc. of blank 
 % recovery= -------------------------------------------- X 100 
measured conc. 
 A recovery close to 100% is a prerequisite for a high degree of trueness, 
but does not ensure unbiased results.
DRIFT & CARRYOVER 
 DRIFT- caused by instrument instability over time, so that calibration 
becomes biased. 
 CARRYOVER- must be close to ‘Zero’ to ensure unbiased results.
IV. LINEARITY 
 It refers to the relationship between measured and expected values over 
the analytical measurement range. 
 Linearity of the analytical procedure is its ability to obtain test results which 
are directly proportional to the concentration of analyte in sample. 
 Most of the times, a dilution series of a sample is examined. E.g., linearity 
of glucose estimation is found by measuring serial dilutions of a stock 
solution of 1000mg/dl concentration.
A response curve of an electrochemical measurement of copper with concentration of analyte 
y is the measured current, x is the concentration of copper
V. LIMIT OF DETECTION 
 Defined as the lowest value that exceeds the measurements of a blank 
sample. 
 It is the “minimum detectable net concentration”. 
 It presumably is the smallest concentration of analyte that can be 
determined to be actually present, even if the quantification has large 
uncertainty 
 “Limit of detection”  a clearly defined cut off above which the analyte is 
measured and below which it is not
VI. ANALYTICAL / MEASURING 
RANGE 
 It is the analyte concentration range over which measurements are within 
the declared tolerances for imprecision and bias of the method. 
 This range extends from the lowest concentration (i.e. lower limit of 
quantification {LloQ}) to the highest concentration (i.e. upper limit of 
quantification {UloQ}) for which the performance specifications are fulfilled.
VII. ANALYTICAL SENSITIVITY 
 Defined as the ability of an analytical method to assess small differences in 
the concentration of the analyte. 
 The smaller the random variation of the instrument response and the 
steeper the slope of the calibration function at a given point, the better is 
the ability to distinguish small differences in analyte concentration. 
 It depends on the PRECISION of the method.
VIII. ANALYTICAL SPECIFICITY & 
INTERFERENCE 
 Analytical specificity is the ability of an assay procedure to determine the 
concentration of the target analyte without influence of potentially interfering 
substances or factors in the sample matrix. 
 Interfering substances/factors:- 
i. Hyperlipemia 
ii. Hemolysis 
iii. Bilirubin 
iv. Antibodies 
v. Degradation products of the analyte 
vi. Exogenous substances 
vii. Anticoagulants
 Interferences from hyperlipemia, hemolysis and bilirubin – concentration 
dependent &can be quantified as a function of the concentration of the 
interfering compound. 
 CLSI has published a detailed protocol for evaluation of interference.
IX. RUGGEDNESS & ROBUSTNESS 
 A method is classed as rugged if its results remain sufficiently unaffected 
as designated environmental and operational conditions change. 
 Practicality of the method ultimately depends on how rugged it is. 
 A robustness study addresses two areas of concern: the need for a 
method to be able to yield acceptable results under the normal variation of 
conditions expected during routine operation, and the portability of the 
method between laboratories with changes in instrumentation, people, and 
chemical reagents.
X. DOCUMENTATION 
o It is the most important step in method validation. 
o ‘If you have not documented, you have not done’ 
o Proper documentation is proof of validity. 
o Method validation is only as good as its documentation, which describes 
the method, the parameters investigated during the method validation 
study, the results of the study, and the Method Validation conclusions 
concerning the validation.
Reasons for Documentation 
1. to allow a user of the method to understand the scope of validation and 
how to perform experiments within the scope of the validation 
2. the method is not validated until it is documented. No number of 
experiments is of any use if they are not documented for future users and 
clients to consult 
As the documentation will become part of the laboratory’s quality 
assurance system, proper documentary traceability (as opposed to 
metrological traceability of a result) must be maintained, with revision 
versions and appropriate authorizations for updates.
When all the above steps are 
complete, the method is said to be 
validated and can be now put into use.
References 
1. Tietz Textbook of Clinical Chemistry and Molecular Diagnosis; Edition; 
Chapter 2; pg 
2. Pictures and graphs from m.authorstream.com

More Related Content

What's hot

Quality control in clinical laboratories
Quality control in clinical laboratoriesQuality control in clinical laboratories
Quality control in clinical laboratoriesAshish Jawarkar
 
Laboratory Internal Quality Control presentation master revision, 2014
Laboratory Internal Quality Control presentation master revision, 2014Laboratory Internal Quality Control presentation master revision, 2014
Laboratory Internal Quality Control presentation master revision, 2014Adel Elazab Elged
 
Validation of lab instruments and quantitative test methods
Validation of lab instruments and quantitative test methods Validation of lab instruments and quantitative test methods
Validation of lab instruments and quantitative test methods Mostafa Mahmoud
 
Analytical method validation
Analytical method validationAnalytical method validation
Analytical method validationSai Praveen Reddy
 
Laboratory Quality Control .ppt
Laboratory Quality Control .pptLaboratory Quality Control .ppt
Laboratory Quality Control .pptABRARAHMED767665
 
M10 bioanalytical method validation
M10 bioanalytical method validationM10 bioanalytical method validation
M10 bioanalytical method validationsawantanil
 
Quality control in clinical laboratory
Quality control in clinical laboratoryQuality control in clinical laboratory
Quality control in clinical laboratorydrgomi basar
 
Qualification of analytical instruments
Qualification of analytical instrumentsQualification of analytical instruments
Qualification of analytical instrumentsFaris ameen
 
ANALYTICAL VARIABLES IN QUALITY CONTROL.pptx
ANALYTICAL VARIABLES IN QUALITY CONTROL.pptxANALYTICAL VARIABLES IN QUALITY CONTROL.pptx
ANALYTICAL VARIABLES IN QUALITY CONTROL.pptxDr. Jagroop Singh
 
Quality control in the medical laboratory
Quality control in the medical laboratoryQuality control in the medical laboratory
Quality control in the medical laboratoryAdnan Jaran
 
Biological variation as an uncertainty component
Biological variation as an uncertainty componentBiological variation as an uncertainty component
Biological variation as an uncertainty componentGH Yeoh
 
Method validation and Verification .pdf
Method validation and Verification .pdfMethod validation and Verification .pdf
Method validation and Verification .pdfMMathaiyanMani
 
Quality assurance in the department of clinical biochemistry
Quality assurance in the department of clinical biochemistryQuality assurance in the department of clinical biochemistry
Quality assurance in the department of clinical biochemistryDipesh Tamrakar
 
Method Validation - Linearity
Method Validation - LinearityMethod Validation - Linearity
Method Validation - Linearitylabgo
 

What's hot (20)

Quality control in clinical laboratories
Quality control in clinical laboratoriesQuality control in clinical laboratories
Quality control in clinical laboratories
 
Laboratory Internal Quality Control presentation master revision, 2014
Laboratory Internal Quality Control presentation master revision, 2014Laboratory Internal Quality Control presentation master revision, 2014
Laboratory Internal Quality Control presentation master revision, 2014
 
Validation of lab instruments and quantitative test methods
Validation of lab instruments and quantitative test methods Validation of lab instruments and quantitative test methods
Validation of lab instruments and quantitative test methods
 
Analytical method validation
Analytical method validationAnalytical method validation
Analytical method validation
 
Quality control
Quality controlQuality control
Quality control
 
Laboratory Quality Control .ppt
Laboratory Quality Control .pptLaboratory Quality Control .ppt
Laboratory Quality Control .ppt
 
M10 bioanalytical method validation
M10 bioanalytical method validationM10 bioanalytical method validation
M10 bioanalytical method validation
 
Troubleshooting IQC / EQAS
Troubleshooting IQC / EQASTroubleshooting IQC / EQAS
Troubleshooting IQC / EQAS
 
Calibration
CalibrationCalibration
Calibration
 
Quality assurance
Quality assuranceQuality assurance
Quality assurance
 
Quality control in clinical laboratory
Quality control in clinical laboratoryQuality control in clinical laboratory
Quality control in clinical laboratory
 
Analytical Method Validation
Analytical Method ValidationAnalytical Method Validation
Analytical Method Validation
 
Qualification of analytical instruments
Qualification of analytical instrumentsQualification of analytical instruments
Qualification of analytical instruments
 
ANALYTICAL VARIABLES IN QUALITY CONTROL.pptx
ANALYTICAL VARIABLES IN QUALITY CONTROL.pptxANALYTICAL VARIABLES IN QUALITY CONTROL.pptx
ANALYTICAL VARIABLES IN QUALITY CONTROL.pptx
 
Quality control in the medical laboratory
Quality control in the medical laboratoryQuality control in the medical laboratory
Quality control in the medical laboratory
 
Biological variation as an uncertainty component
Biological variation as an uncertainty componentBiological variation as an uncertainty component
Biological variation as an uncertainty component
 
Analytical method validation
Analytical method validation Analytical method validation
Analytical method validation
 
Method validation and Verification .pdf
Method validation and Verification .pdfMethod validation and Verification .pdf
Method validation and Verification .pdf
 
Quality assurance in the department of clinical biochemistry
Quality assurance in the department of clinical biochemistryQuality assurance in the department of clinical biochemistry
Quality assurance in the department of clinical biochemistry
 
Method Validation - Linearity
Method Validation - LinearityMethod Validation - Linearity
Method Validation - Linearity
 

Viewers also liked

Analytical method validation by manoj ingale(best ppts)
Analytical method validation by manoj ingale(best ppts)Analytical method validation by manoj ingale(best ppts)
Analytical method validation by manoj ingale(best ppts)Indus Biotech Pvt.Ltd.
 
analytical method validation
analytical method validationanalytical method validation
analytical method validationDr. Raja Abhilash
 
Sampling analysis-reporting
Sampling analysis-reportingSampling analysis-reporting
Sampling analysis-reportingdikasolanki
 
Current Concepts in Laboratory Testing to Guide Antimicrobial Therapy
Current Concepts in Laboratory Testing to Guide Antimicrobial TherapyCurrent Concepts in Laboratory Testing to Guide Antimicrobial Therapy
Current Concepts in Laboratory Testing to Guide Antimicrobial TherapyPathogens Outlook
 
Analytical method validation
Analytical method validationAnalytical method validation
Analytical method validationceutics1315
 
Analytical method validation workshop
Analytical method validation workshopAnalytical method validation workshop
Analytical method validation workshopSiham Abdallaha
 
Quality control of antimicrobial susceptibility tests
Quality control of antimicrobial susceptibility testsQuality control of antimicrobial susceptibility tests
Quality control of antimicrobial susceptibility testsAmr Eldakroury
 
Analytical method validation
Analytical method validationAnalytical method validation
Analytical method validationSUBHASISH DAS
 
polyclonal antibodies
polyclonal antibodiespolyclonal antibodies
polyclonal antibodiesnida rehman
 
ACT antimicrobial susceptibility testing, inoculation and drug sensitivity
ACT antimicrobial susceptibility testing, inoculation and drug sensitivity ACT antimicrobial susceptibility testing, inoculation and drug sensitivity
ACT antimicrobial susceptibility testing, inoculation and drug sensitivity Arsh Gull
 
Analytical method validation raaj gprac [compatibility mode]
Analytical method validation raaj gprac [compatibility mode]Analytical method validation raaj gprac [compatibility mode]
Analytical method validation raaj gprac [compatibility mode]Rajashri Survase Ojha
 
Pkqa iso 17025 requirements
Pkqa iso 17025 requirementsPkqa iso 17025 requirements
Pkqa iso 17025 requirementsPat Kilbane
 
Methods of Assessing Bioavailability
Methods of Assessing BioavailabilityMethods of Assessing Bioavailability
Methods of Assessing BioavailabilityDibrugarh University
 
Semen analysis as per WHO and clinical implications
Semen analysis as per WHO and clinical implicationsSemen analysis as per WHO and clinical implications
Semen analysis as per WHO and clinical implicationsSandro Esteves
 
ICH Q2 Analytical Method Validation
ICH Q2  Analytical Method ValidationICH Q2  Analytical Method Validation
ICH Q2 Analytical Method ValidationNaila Kanwal
 
Measurement And Error
Measurement And ErrorMeasurement And Error
Measurement And Errorwilsone
 
Susceptibility Testing Review
Susceptibility Testing ReviewSusceptibility Testing Review
Susceptibility Testing ReviewMargie Morgan
 

Viewers also liked (20)

Analytical method validation by manoj ingale(best ppts)
Analytical method validation by manoj ingale(best ppts)Analytical method validation by manoj ingale(best ppts)
Analytical method validation by manoj ingale(best ppts)
 
analytical method validation
analytical method validationanalytical method validation
analytical method validation
 
Sampling analysis-reporting
Sampling analysis-reportingSampling analysis-reporting
Sampling analysis-reporting
 
Current Concepts in Laboratory Testing to Guide Antimicrobial Therapy
Current Concepts in Laboratory Testing to Guide Antimicrobial TherapyCurrent Concepts in Laboratory Testing to Guide Antimicrobial Therapy
Current Concepts in Laboratory Testing to Guide Antimicrobial Therapy
 
ASTM D19 Method Validation Procedures
ASTM D19 Method Validation ProceduresASTM D19 Method Validation Procedures
ASTM D19 Method Validation Procedures
 
Analytical method validation
Analytical method validationAnalytical method validation
Analytical method validation
 
Analytical method validation workshop
Analytical method validation workshopAnalytical method validation workshop
Analytical method validation workshop
 
Production of Monoclonal and Polyclonal Antibodies for Detection of Geminivir...
Production of Monoclonal and Polyclonal Antibodies for Detection of Geminivir...Production of Monoclonal and Polyclonal Antibodies for Detection of Geminivir...
Production of Monoclonal and Polyclonal Antibodies for Detection of Geminivir...
 
Quality control of antimicrobial susceptibility tests
Quality control of antimicrobial susceptibility testsQuality control of antimicrobial susceptibility tests
Quality control of antimicrobial susceptibility tests
 
Analytical method validation
Analytical method validationAnalytical method validation
Analytical method validation
 
polyclonal antibodies
polyclonal antibodiespolyclonal antibodies
polyclonal antibodies
 
ACT antimicrobial susceptibility testing, inoculation and drug sensitivity
ACT antimicrobial susceptibility testing, inoculation and drug sensitivity ACT antimicrobial susceptibility testing, inoculation and drug sensitivity
ACT antimicrobial susceptibility testing, inoculation and drug sensitivity
 
Analytical method validation raaj gprac [compatibility mode]
Analytical method validation raaj gprac [compatibility mode]Analytical method validation raaj gprac [compatibility mode]
Analytical method validation raaj gprac [compatibility mode]
 
Pkqa iso 17025 requirements
Pkqa iso 17025 requirementsPkqa iso 17025 requirements
Pkqa iso 17025 requirements
 
Methods of Assessing Bioavailability
Methods of Assessing BioavailabilityMethods of Assessing Bioavailability
Methods of Assessing Bioavailability
 
Semen analysis as per WHO and clinical implications
Semen analysis as per WHO and clinical implicationsSemen analysis as per WHO and clinical implications
Semen analysis as per WHO and clinical implications
 
ICH Q2 Analytical Method Validation
ICH Q2  Analytical Method ValidationICH Q2  Analytical Method Validation
ICH Q2 Analytical Method Validation
 
Measurement And Error
Measurement And ErrorMeasurement And Error
Measurement And Error
 
Measurement & Error
Measurement & ErrorMeasurement & Error
Measurement & Error
 
Susceptibility Testing Review
Susceptibility Testing ReviewSusceptibility Testing Review
Susceptibility Testing Review
 

Similar to METHOD VALIDATION STEPS

Analytical method validation
Analytical method validationAnalytical method validation
Analytical method validationroshankhetade2
 
ANALYTICAL METHOD VALIDATION (PHARMACEUTICAL INDUSTRY).pdf
ANALYTICAL METHOD VALIDATION (PHARMACEUTICAL INDUSTRY).pdfANALYTICAL METHOD VALIDATION (PHARMACEUTICAL INDUSTRY).pdf
ANALYTICAL METHOD VALIDATION (PHARMACEUTICAL INDUSTRY).pdfBALASUNDARESAN M
 
Seminar on introduction to validation
Seminar on introduction to validationSeminar on introduction to validation
Seminar on introduction to validationvasantipatil5
 
THE CONCEPT OF TRACEABILITY IN LABORATORY MEDICINE - A TOOL FOR STANDARDISATION
THE CONCEPT OF TRACEABILITY IN LABORATORY MEDICINE - A TOOL FOR STANDARDISATIONTHE CONCEPT OF TRACEABILITY IN LABORATORY MEDICINE - A TOOL FOR STANDARDISATION
THE CONCEPT OF TRACEABILITY IN LABORATORY MEDICINE - A TOOL FOR STANDARDISATIONMoustafa Rezk
 
Analytical Method Validation.pdf
Analytical Method Validation.pdfAnalytical Method Validation.pdf
Analytical Method Validation.pdfHiteshGajjar7
 
Analytical methods validation as per ich & usp
Analytical methods validation as per ich & uspAnalytical methods validation as per ich & usp
Analytical methods validation as per ich & uspGANESH NIGADE
 
Analytical method validation as per ich and usp
Analytical method validation as per ich and usp Analytical method validation as per ich and usp
Analytical method validation as per ich and usp shreyas B R
 
Pharmacy Slideshow XL by Slidesgo.pptx
Pharmacy Slideshow XL by Slidesgo.pptxPharmacy Slideshow XL by Slidesgo.pptx
Pharmacy Slideshow XL by Slidesgo.pptxMahmoudNaguipabdelha
 
Analytical Method Validation basics by Dr. A. Amsavel
Analytical Method Validation  basics by Dr. A. AmsavelAnalytical Method Validation  basics by Dr. A. Amsavel
Analytical Method Validation basics by Dr. A. AmsavelDr. Amsavel A
 
All you know About Analytical method validation
All you  know About Analytical method validation All you  know About Analytical method validation
All you know About Analytical method validation CynthiaCynthiaPLegro
 
Analytical Method Validations & Detailed Method Validation Parameters
Analytical Method Validations & Detailed Method Validation ParametersAnalytical Method Validations & Detailed Method Validation Parameters
Analytical Method Validations & Detailed Method Validation ParametersBussaSravanKumar
 
Analytical method validation 2020
Analytical method validation 2020Analytical method validation 2020
Analytical method validation 2020PRANJAY PATIL
 
International conference on harmonisation validation of analytical procedures
International conference on harmonisation validation of analytical proceduresInternational conference on harmonisation validation of analytical procedures
International conference on harmonisation validation of analytical proceduresnoushin javidvand
 
QA QC Program for Waste Water Analysis ppt
QA QC Program for Waste Water Analysis pptQA QC Program for Waste Water Analysis ppt
QA QC Program for Waste Water Analysis pptKrisnaBagtasos1
 
2. Quality Control Notes
2. Quality Control Notes2. Quality Control Notes
2. Quality Control NotesLeah Molai
 

Similar to METHOD VALIDATION STEPS (20)

Analytical method validation
Analytical method validationAnalytical method validation
Analytical method validation
 
ANALYTICAL METHOD VALIDATION (PHARMACEUTICAL INDUSTRY).pdf
ANALYTICAL METHOD VALIDATION (PHARMACEUTICAL INDUSTRY).pdfANALYTICAL METHOD VALIDATION (PHARMACEUTICAL INDUSTRY).pdf
ANALYTICAL METHOD VALIDATION (PHARMACEUTICAL INDUSTRY).pdf
 
Seminar on introduction to validation
Seminar on introduction to validationSeminar on introduction to validation
Seminar on introduction to validation
 
THE CONCEPT OF TRACEABILITY IN LABORATORY MEDICINE - A TOOL FOR STANDARDISATION
THE CONCEPT OF TRACEABILITY IN LABORATORY MEDICINE - A TOOL FOR STANDARDISATIONTHE CONCEPT OF TRACEABILITY IN LABORATORY MEDICINE - A TOOL FOR STANDARDISATION
THE CONCEPT OF TRACEABILITY IN LABORATORY MEDICINE - A TOOL FOR STANDARDISATION
 
Q2 r1 guidelines
Q2 r1 guidelinesQ2 r1 guidelines
Q2 r1 guidelines
 
Analytical Method Validation.pdf
Analytical Method Validation.pdfAnalytical Method Validation.pdf
Analytical Method Validation.pdf
 
Analytical method validation
Analytical method validationAnalytical method validation
Analytical method validation
 
Analytical methods validation as per ich & usp
Analytical methods validation as per ich & uspAnalytical methods validation as per ich & usp
Analytical methods validation as per ich & usp
 
Validation methods.pptx
Validation methods.pptxValidation methods.pptx
Validation methods.pptx
 
Analytical method validation as per ich and usp
Analytical method validation as per ich and usp Analytical method validation as per ich and usp
Analytical method validation as per ich and usp
 
Pharmacy Slideshow XL by Slidesgo.pptx
Pharmacy Slideshow XL by Slidesgo.pptxPharmacy Slideshow XL by Slidesgo.pptx
Pharmacy Slideshow XL by Slidesgo.pptx
 
Analytical Method Validation basics by Dr. A. Amsavel
Analytical Method Validation  basics by Dr. A. AmsavelAnalytical Method Validation  basics by Dr. A. Amsavel
Analytical Method Validation basics by Dr. A. Amsavel
 
Validation
ValidationValidation
Validation
 
Validation my ppt
Validation my pptValidation my ppt
Validation my ppt
 
All you know About Analytical method validation
All you  know About Analytical method validation All you  know About Analytical method validation
All you know About Analytical method validation
 
Analytical Method Validations & Detailed Method Validation Parameters
Analytical Method Validations & Detailed Method Validation ParametersAnalytical Method Validations & Detailed Method Validation Parameters
Analytical Method Validations & Detailed Method Validation Parameters
 
Analytical method validation 2020
Analytical method validation 2020Analytical method validation 2020
Analytical method validation 2020
 
International conference on harmonisation validation of analytical procedures
International conference on harmonisation validation of analytical proceduresInternational conference on harmonisation validation of analytical procedures
International conference on harmonisation validation of analytical procedures
 
QA QC Program for Waste Water Analysis ppt
QA QC Program for Waste Water Analysis pptQA QC Program for Waste Water Analysis ppt
QA QC Program for Waste Water Analysis ppt
 
2. Quality Control Notes
2. Quality Control Notes2. Quality Control Notes
2. Quality Control Notes
 

Recently uploaded

Radiation Dosimetry Parameters and Isodose Curves.pptx
Radiation Dosimetry Parameters and Isodose Curves.pptxRadiation Dosimetry Parameters and Isodose Curves.pptx
Radiation Dosimetry Parameters and Isodose Curves.pptxDr. Dheeraj Kumar
 
Apiculture Chapter 1. Introduction 2.ppt
Apiculture Chapter 1. Introduction 2.pptApiculture Chapter 1. Introduction 2.ppt
Apiculture Chapter 1. Introduction 2.pptkedirjemalharun
 
call girls in Dwarka Sector 21 Metro DELHI 🔝 >༒9540349809 🔝 genuine Escort Se...
call girls in Dwarka Sector 21 Metro DELHI 🔝 >༒9540349809 🔝 genuine Escort Se...call girls in Dwarka Sector 21 Metro DELHI 🔝 >༒9540349809 🔝 genuine Escort Se...
call girls in Dwarka Sector 21 Metro DELHI 🔝 >༒9540349809 🔝 genuine Escort Se...saminamagar
 
Lippincott Microcards_ Microbiology Flash Cards-LWW (2015).pdf
Lippincott Microcards_ Microbiology Flash Cards-LWW (2015).pdfLippincott Microcards_ Microbiology Flash Cards-LWW (2015).pdf
Lippincott Microcards_ Microbiology Flash Cards-LWW (2015).pdfSreeja Cherukuru
 
Wessex Health Partners Wessex Integrated Care, Population Health, Research & ...
Wessex Health Partners Wessex Integrated Care, Population Health, Research & ...Wessex Health Partners Wessex Integrated Care, Population Health, Research & ...
Wessex Health Partners Wessex Integrated Care, Population Health, Research & ...Wessex Health Partners
 
LUNG TUMORS AND ITS CLASSIFICATIONS.pdf
LUNG TUMORS AND ITS  CLASSIFICATIONS.pdfLUNG TUMORS AND ITS  CLASSIFICATIONS.pdf
LUNG TUMORS AND ITS CLASSIFICATIONS.pdfDolisha Warbi
 
Presentation on Parasympathetic Nervous System
Presentation on Parasympathetic Nervous SystemPresentation on Parasympathetic Nervous System
Presentation on Parasympathetic Nervous SystemPrerana Jadhav
 
METHODS OF ACQUIRING KNOWLEDGE IN NURSING.pptx by navdeep kaur
METHODS OF ACQUIRING KNOWLEDGE IN NURSING.pptx by navdeep kaurMETHODS OF ACQUIRING KNOWLEDGE IN NURSING.pptx by navdeep kaur
METHODS OF ACQUIRING KNOWLEDGE IN NURSING.pptx by navdeep kaurNavdeep Kaur
 
call girls in munirka DELHI 🔝 >༒9540349809 🔝 genuine Escort Service 🔝✔️✔️
call girls in munirka  DELHI 🔝 >༒9540349809 🔝 genuine Escort Service 🔝✔️✔️call girls in munirka  DELHI 🔝 >༒9540349809 🔝 genuine Escort Service 🔝✔️✔️
call girls in munirka DELHI 🔝 >༒9540349809 🔝 genuine Escort Service 🔝✔️✔️saminamagar
 
VarSeq 2.6.0: Advancing Pharmacogenomics and Genomic Analysis
VarSeq 2.6.0: Advancing Pharmacogenomics and Genomic AnalysisVarSeq 2.6.0: Advancing Pharmacogenomics and Genomic Analysis
VarSeq 2.6.0: Advancing Pharmacogenomics and Genomic AnalysisGolden Helix
 
COVID-19 (NOVEL CORONA VIRUS DISEASE PANDEMIC ).pptx
COVID-19  (NOVEL CORONA  VIRUS DISEASE PANDEMIC ).pptxCOVID-19  (NOVEL CORONA  VIRUS DISEASE PANDEMIC ).pptx
COVID-19 (NOVEL CORONA VIRUS DISEASE PANDEMIC ).pptxBibekananda shah
 
Hematology and Immunology - Leukocytes Functions
Hematology and Immunology - Leukocytes FunctionsHematology and Immunology - Leukocytes Functions
Hematology and Immunology - Leukocytes FunctionsMedicoseAcademics
 
SWD (Short wave diathermy)- Physiotherapy.ppt
SWD (Short wave diathermy)- Physiotherapy.pptSWD (Short wave diathermy)- Physiotherapy.ppt
SWD (Short wave diathermy)- Physiotherapy.pptMumux Mirani
 
epilepsy and status epilepticus for undergraduate.pptx
epilepsy and status epilepticus  for undergraduate.pptxepilepsy and status epilepticus  for undergraduate.pptx
epilepsy and status epilepticus for undergraduate.pptxMohamed Rizk Khodair
 
call girls in paharganj DELHI 🔝 >༒9540349809 🔝 genuine Escort Service 🔝✔️✔️
call girls in paharganj DELHI 🔝 >༒9540349809 🔝 genuine Escort Service 🔝✔️✔️call girls in paharganj DELHI 🔝 >༒9540349809 🔝 genuine Escort Service 🔝✔️✔️
call girls in paharganj DELHI 🔝 >༒9540349809 🔝 genuine Escort Service 🔝✔️✔️saminamagar
 
PULMONARY EDEMA AND ITS MANAGEMENT.pdf
PULMONARY EDEMA AND  ITS  MANAGEMENT.pdfPULMONARY EDEMA AND  ITS  MANAGEMENT.pdf
PULMONARY EDEMA AND ITS MANAGEMENT.pdfDolisha Warbi
 
Presentation on General Anesthetics pdf.
Presentation on General Anesthetics pdf.Presentation on General Anesthetics pdf.
Presentation on General Anesthetics pdf.Prerana Jadhav
 
POST NATAL EXERCISES AND ITS IMPACT.pptx
POST NATAL EXERCISES AND ITS IMPACT.pptxPOST NATAL EXERCISES AND ITS IMPACT.pptx
POST NATAL EXERCISES AND ITS IMPACT.pptxvirengeeta
 
Let's Talk About It: To Disclose or Not to Disclose?
Let's Talk About It: To Disclose or Not to Disclose?Let's Talk About It: To Disclose or Not to Disclose?
Let's Talk About It: To Disclose or Not to Disclose?bkling
 
History and Development of Pharmacovigilence.pdf
History and Development of Pharmacovigilence.pdfHistory and Development of Pharmacovigilence.pdf
History and Development of Pharmacovigilence.pdfSasikiranMarri
 

Recently uploaded (20)

Radiation Dosimetry Parameters and Isodose Curves.pptx
Radiation Dosimetry Parameters and Isodose Curves.pptxRadiation Dosimetry Parameters and Isodose Curves.pptx
Radiation Dosimetry Parameters and Isodose Curves.pptx
 
Apiculture Chapter 1. Introduction 2.ppt
Apiculture Chapter 1. Introduction 2.pptApiculture Chapter 1. Introduction 2.ppt
Apiculture Chapter 1. Introduction 2.ppt
 
call girls in Dwarka Sector 21 Metro DELHI 🔝 >༒9540349809 🔝 genuine Escort Se...
call girls in Dwarka Sector 21 Metro DELHI 🔝 >༒9540349809 🔝 genuine Escort Se...call girls in Dwarka Sector 21 Metro DELHI 🔝 >༒9540349809 🔝 genuine Escort Se...
call girls in Dwarka Sector 21 Metro DELHI 🔝 >༒9540349809 🔝 genuine Escort Se...
 
Lippincott Microcards_ Microbiology Flash Cards-LWW (2015).pdf
Lippincott Microcards_ Microbiology Flash Cards-LWW (2015).pdfLippincott Microcards_ Microbiology Flash Cards-LWW (2015).pdf
Lippincott Microcards_ Microbiology Flash Cards-LWW (2015).pdf
 
Wessex Health Partners Wessex Integrated Care, Population Health, Research & ...
Wessex Health Partners Wessex Integrated Care, Population Health, Research & ...Wessex Health Partners Wessex Integrated Care, Population Health, Research & ...
Wessex Health Partners Wessex Integrated Care, Population Health, Research & ...
 
LUNG TUMORS AND ITS CLASSIFICATIONS.pdf
LUNG TUMORS AND ITS  CLASSIFICATIONS.pdfLUNG TUMORS AND ITS  CLASSIFICATIONS.pdf
LUNG TUMORS AND ITS CLASSIFICATIONS.pdf
 
Presentation on Parasympathetic Nervous System
Presentation on Parasympathetic Nervous SystemPresentation on Parasympathetic Nervous System
Presentation on Parasympathetic Nervous System
 
METHODS OF ACQUIRING KNOWLEDGE IN NURSING.pptx by navdeep kaur
METHODS OF ACQUIRING KNOWLEDGE IN NURSING.pptx by navdeep kaurMETHODS OF ACQUIRING KNOWLEDGE IN NURSING.pptx by navdeep kaur
METHODS OF ACQUIRING KNOWLEDGE IN NURSING.pptx by navdeep kaur
 
call girls in munirka DELHI 🔝 >༒9540349809 🔝 genuine Escort Service 🔝✔️✔️
call girls in munirka  DELHI 🔝 >༒9540349809 🔝 genuine Escort Service 🔝✔️✔️call girls in munirka  DELHI 🔝 >༒9540349809 🔝 genuine Escort Service 🔝✔️✔️
call girls in munirka DELHI 🔝 >༒9540349809 🔝 genuine Escort Service 🔝✔️✔️
 
VarSeq 2.6.0: Advancing Pharmacogenomics and Genomic Analysis
VarSeq 2.6.0: Advancing Pharmacogenomics and Genomic AnalysisVarSeq 2.6.0: Advancing Pharmacogenomics and Genomic Analysis
VarSeq 2.6.0: Advancing Pharmacogenomics and Genomic Analysis
 
COVID-19 (NOVEL CORONA VIRUS DISEASE PANDEMIC ).pptx
COVID-19  (NOVEL CORONA  VIRUS DISEASE PANDEMIC ).pptxCOVID-19  (NOVEL CORONA  VIRUS DISEASE PANDEMIC ).pptx
COVID-19 (NOVEL CORONA VIRUS DISEASE PANDEMIC ).pptx
 
Hematology and Immunology - Leukocytes Functions
Hematology and Immunology - Leukocytes FunctionsHematology and Immunology - Leukocytes Functions
Hematology and Immunology - Leukocytes Functions
 
SWD (Short wave diathermy)- Physiotherapy.ppt
SWD (Short wave diathermy)- Physiotherapy.pptSWD (Short wave diathermy)- Physiotherapy.ppt
SWD (Short wave diathermy)- Physiotherapy.ppt
 
epilepsy and status epilepticus for undergraduate.pptx
epilepsy and status epilepticus  for undergraduate.pptxepilepsy and status epilepticus  for undergraduate.pptx
epilepsy and status epilepticus for undergraduate.pptx
 
call girls in paharganj DELHI 🔝 >༒9540349809 🔝 genuine Escort Service 🔝✔️✔️
call girls in paharganj DELHI 🔝 >༒9540349809 🔝 genuine Escort Service 🔝✔️✔️call girls in paharganj DELHI 🔝 >༒9540349809 🔝 genuine Escort Service 🔝✔️✔️
call girls in paharganj DELHI 🔝 >༒9540349809 🔝 genuine Escort Service 🔝✔️✔️
 
PULMONARY EDEMA AND ITS MANAGEMENT.pdf
PULMONARY EDEMA AND  ITS  MANAGEMENT.pdfPULMONARY EDEMA AND  ITS  MANAGEMENT.pdf
PULMONARY EDEMA AND ITS MANAGEMENT.pdf
 
Presentation on General Anesthetics pdf.
Presentation on General Anesthetics pdf.Presentation on General Anesthetics pdf.
Presentation on General Anesthetics pdf.
 
POST NATAL EXERCISES AND ITS IMPACT.pptx
POST NATAL EXERCISES AND ITS IMPACT.pptxPOST NATAL EXERCISES AND ITS IMPACT.pptx
POST NATAL EXERCISES AND ITS IMPACT.pptx
 
Let's Talk About It: To Disclose or Not to Disclose?
Let's Talk About It: To Disclose or Not to Disclose?Let's Talk About It: To Disclose or Not to Disclose?
Let's Talk About It: To Disclose or Not to Disclose?
 
History and Development of Pharmacovigilence.pdf
History and Development of Pharmacovigilence.pdfHistory and Development of Pharmacovigilence.pdf
History and Development of Pharmacovigilence.pdf
 

METHOD VALIDATION STEPS

  • 1. METHOD VALIDATION BY:- DR. HINAL SHAH
  • 2. INTRODUCTION  Method selection and validation – Key steps in the process of implementing new methods in the clinical laboratory.  In Clinical Chemistry Laboratory- influenced strongly by guidelines.
  • 3.  The Clinical & Laboratory Standards Institute (formerly National Committee for Clinical Laboratory Standards) and International Organization for Standardization have published a series of protocols and documents to be followed by CCL and manufacturers for method validation.  Also, laboratory accreditation requirements have to be met.
  • 4. DEFINITIONS  “The suite of procedures to which an analytical method is subjected to provide objective evidence that the method, if used in the manner specified, will produce results that conform to the statement of the method validation parameters”  “The process used to confirm that the analytical procedure employed for a specific test is suitable for its intended use”  ISO/IEC 17025 (ISO/IEC 2005, section 5.4) states that method validation is “confirmation by examination and provision of objective evidence that the particular requirements for a specified intended use are fulfilled”
  • 5. When Should Methods Be Validated?  No method should be used without validation  ISO/IEC 17025 encourages the use of methods published in international, regional, or national standards and implies that if these methods are used without deviation, then the validation requirements have been satisfied.  What does need to be validated are-i. Nonstandard methods, designed and developed by individual laboratories ii. Standard methods used outside their scope iii. Amplifications and modifications of standard methods.
  • 6. Method selection/ development and validation cycle
  • 7. VERIFICATION  Synonymous with Single Laboratory Validation  To be done when a laboratory uses a method for the first time (with standard materials)  Also done when a particular aspect of the method or its implementation is changed, for example-a) New analyst b) New equipment or equipment part c) New batch of reagent d) Changes in the laboratory premises
  • 8.  Minimum verification is to analyze a material before and after the change and check for consistency of the results, both in terms of mean and standard deviation.  Basic experiments that should be performed to verify the use of a standard method before first use in a laboratory:  Bias/recovery  Precision  Measurement uncertainty  Calibration model (linearity)  Limit of detection
  • 9. METHOD SELECTION- CRITERIA 1. Medical Need 2. Analytical Performance 3. Practical Criteria- depend on whether it is a Laboratory Developed Test or a Commercial kit method.
  • 10. Method Validation- Requirements I. Calibration II. Precision III. Accuracy and Trueness IV. Linearity V. Limit of Detection (LOD) VI. Analytical/Measuring Range VII. Analytical Sensitivity VIII. Analytical Specificity and Interference IX. Ruggedness and Robustness X. Documentation
  • 11. I. CALIBRATION o Established by measurement of samples with known quantities of the analyte (calibrators). [Only certified reference materials with known traceability should be used] o Done by- Chemical standards or samples with known quantities of analyte present in the typical matrix. o Calibration functions- may be linear or curved. o If the assumed calibration function does not correctly reflect the true relationship between instrument response & analyte concentration, a systemic error or bias is likely to be associated with the analytical method.
  • 12. II. PRECISION  Precision has been defined as the closeness of agreement between independent results of measurements obtained under stipulated conditions.  Depends on – the stability of the instrument response for a given quantity of analyte.  The degree of precision is usually expressed on the basis of statistical measures of imprecision, such as Standard Deviation(SD) OR Coefficient of Variation(CV)
  • 13.  If calibration function is linear and imprecision remains same over analytical measurement range, SD also tends to be constant over analytical measurement range. If imprecision increases proportionally, SD also tends to increase proportionally to the concentration and relative imprecision(i.e. CV) remains constant.  Imprecision of measurements is solely related to the random error.  Precision is specified as follows:-
  • 14. 1) Repeatability: Corresponds to with-in run precision. It is closeness of agreement between results of successive measurements carried out under the same conditions. 2) Reproducibility: Closeness of agreement between results of measurements performed under changed conditions of measurements. 2 specifications of reproducibility –  Total or between-run precision (intermediate precision):- its method of assessment depends on usage of method and variation in analyst/equipment.  Inter-laboratory precision.
  • 15.  An important aspect of a full method validation is estimating bias components attributable to the method itself and to the laboratory carrying out the analysis. It can be evaluated by inter-laboratory studies.
  • 16. III. TRUENESS & ACCURACY  Trueness:- • Closeness of agreement between the mean value (obtained from a large series of results of measurements) & the true value. • Bias – Difference between the mean and the true value. Inversely related to the trueness. • Qualitative term. • True value – in practice is the accepted reference value. • Evaluated by comparison of measurements by a given routine method & a reference measurement procedure.
  • 17.  Accuracy :- • Closeness of agreement between the result of a measurement and true value. • Qualitative term. • Influenced by- BIAS and IMPRECISION. Reflects total error. • Total error= bias + 2SD Concepts of recovery, drift and carryover are related to trueness.
  • 18. RECOVERY  It means - the fraction of a test material in a matrix that can be quantified. target conc. – conc. of blank  % recovery= -------------------------------------------- X 100 measured conc.  A recovery close to 100% is a prerequisite for a high degree of trueness, but does not ensure unbiased results.
  • 19. DRIFT & CARRYOVER  DRIFT- caused by instrument instability over time, so that calibration becomes biased.  CARRYOVER- must be close to ‘Zero’ to ensure unbiased results.
  • 20.
  • 21. IV. LINEARITY  It refers to the relationship between measured and expected values over the analytical measurement range.  Linearity of the analytical procedure is its ability to obtain test results which are directly proportional to the concentration of analyte in sample.  Most of the times, a dilution series of a sample is examined. E.g., linearity of glucose estimation is found by measuring serial dilutions of a stock solution of 1000mg/dl concentration.
  • 22. A response curve of an electrochemical measurement of copper with concentration of analyte y is the measured current, x is the concentration of copper
  • 23. V. LIMIT OF DETECTION  Defined as the lowest value that exceeds the measurements of a blank sample.  It is the “minimum detectable net concentration”.  It presumably is the smallest concentration of analyte that can be determined to be actually present, even if the quantification has large uncertainty  “Limit of detection”  a clearly defined cut off above which the analyte is measured and below which it is not
  • 24. VI. ANALYTICAL / MEASURING RANGE  It is the analyte concentration range over which measurements are within the declared tolerances for imprecision and bias of the method.  This range extends from the lowest concentration (i.e. lower limit of quantification {LloQ}) to the highest concentration (i.e. upper limit of quantification {UloQ}) for which the performance specifications are fulfilled.
  • 25. VII. ANALYTICAL SENSITIVITY  Defined as the ability of an analytical method to assess small differences in the concentration of the analyte.  The smaller the random variation of the instrument response and the steeper the slope of the calibration function at a given point, the better is the ability to distinguish small differences in analyte concentration.  It depends on the PRECISION of the method.
  • 26. VIII. ANALYTICAL SPECIFICITY & INTERFERENCE  Analytical specificity is the ability of an assay procedure to determine the concentration of the target analyte without influence of potentially interfering substances or factors in the sample matrix.  Interfering substances/factors:- i. Hyperlipemia ii. Hemolysis iii. Bilirubin iv. Antibodies v. Degradation products of the analyte vi. Exogenous substances vii. Anticoagulants
  • 27.  Interferences from hyperlipemia, hemolysis and bilirubin – concentration dependent &can be quantified as a function of the concentration of the interfering compound.  CLSI has published a detailed protocol for evaluation of interference.
  • 28. IX. RUGGEDNESS & ROBUSTNESS  A method is classed as rugged if its results remain sufficiently unaffected as designated environmental and operational conditions change.  Practicality of the method ultimately depends on how rugged it is.  A robustness study addresses two areas of concern: the need for a method to be able to yield acceptable results under the normal variation of conditions expected during routine operation, and the portability of the method between laboratories with changes in instrumentation, people, and chemical reagents.
  • 29. X. DOCUMENTATION o It is the most important step in method validation. o ‘If you have not documented, you have not done’ o Proper documentation is proof of validity. o Method validation is only as good as its documentation, which describes the method, the parameters investigated during the method validation study, the results of the study, and the Method Validation conclusions concerning the validation.
  • 30. Reasons for Documentation 1. to allow a user of the method to understand the scope of validation and how to perform experiments within the scope of the validation 2. the method is not validated until it is documented. No number of experiments is of any use if they are not documented for future users and clients to consult As the documentation will become part of the laboratory’s quality assurance system, proper documentary traceability (as opposed to metrological traceability of a result) must be maintained, with revision versions and appropriate authorizations for updates.
  • 31. When all the above steps are complete, the method is said to be validated and can be now put into use.
  • 32. References 1. Tietz Textbook of Clinical Chemistry and Molecular Diagnosis; Edition; Chapter 2; pg 2. Pictures and graphs from m.authorstream.com