Masahisa Yamane - Recognizing the importance of guiding catheters for successful PCI
1. 4th European Live Summit on Retrograde CTO Revascularization, May 8-9,2015 : Zurich Heart House, University Hospital Zurich, Swiss
Recognizing the importance of
guiding catheters for successful PCI
Masahisa Yamane , MD, FACC
St Luke‘s International Hospital ,Tokyo, Japan
Saitama Sekishinkai Hospital, Saitama, Japan
2. 4th European Live Summit on Retrograde CTO Revascularization, May 8-9,2015 : Zurich Heart House, University Hospital Zurich, Swiss
Disclosure Statement of Financial Interest
I, Masahisa Yamane MD,DO NOT have a financial interest/arrangement
or affiliation with one or more organizations that could be perceived as a
real or apparent conflict of interest in the context of the subject of this
presentation.
3. 4th European Live Summit on Retrograde CTO Revascularization, May 8-9,2015 : Zurich Heart House, University Hospital Zurich, Swiss
Importance of GC
Special
Devices
Micro Catheter
Guide Wire
Guiding Catheter
4. 4th European Live Summit on Retrograde CTO Revascularization, May 8-9,2015 : Zurich Heart House, University Hospital Zurich, Swiss
Approach to
the coronary
Back up
support
Device
Deliverability
Safety
Shaft Linearity
Kink Resistance
Torque Performance
Shape
(Coaxial engagement)
Shaft Support
Shape (Supporting
Point)
Shape Memory
Anti Heat
Deformation
Large Inner Lumen
Lumen Retention
Smooth Inner
Layer
Tip Flexibility
Visibility
Round Processed
Tip
Importance of GC
5. 4th European Live Summit on Retrograde CTO Revascularization, May 8-9,2015 : Zurich Heart House, University Hospital Zurich, Swiss
Most Important Factors
Back up support
Anti Heat Deformation
Safety
Maneuverability
These factors are essential for stable procedures, especially in long,
complex cases.
6. 4th European Live Summit on Retrograde CTO Revascularization, May 8-9,2015 : Zurich Heart House, University Hospital Zurich, Swiss
Current trend
Catheter with stiff shaft
Strong back up support due to its rigidity
Concern about safety due to tip stiffness
Catheter with soft shaft
Safe due to soft tip
Extra attention needed to get better back up support
Backup Safety
Catheter with stiff shaft ○ △
Catheter with soft shaft △ ○
7. 4th European Live Summit on Retrograde CTO Revascularization, May 8-9,2015 : Zurich Heart House, University Hospital Zurich, Swiss
Ideal
Relieve the operator’s stress during treatment by
facilitating a stable procedure with strong back up
support and soft tip
Back up Safety Co-axiality
Ideal GC ◎ ◎ ◎
8. 4th European Live Summit on Retrograde CTO Revascularization, May 8-9,2015 : Zurich Heart House, University Hospital Zurich, Swiss
Birth
9. 4th European Live Summit on Retrograde CTO Revascularization, May 8-9,2015 : Zurich Heart House, University Hospital Zurich, Swiss
Product Concept
- Realizes all the essential characteristics for a
guiding catheter – both back up support and safety.
Creating a stable environment for the doctors to
concentrate on PCI treatment
*Most conventional guiding catheters can be divided into two types: those
that have a rigid shaft and tip which undermines safety, and those that have
a flexible shaft and tip which sacrifices the stability of the procedure.
However, the Hyperion realizes both safety and backup support.
10. 4th European Live Summit on Retrograde CTO Revascularization, May 8-9,2015 : Zurich Heart House, University Hospital Zurich, Swiss
Backup Support
Maneuverability
Flexible Tip
Sufficient Specification
11. 4th European Live Summit on Retrograde CTO Revascularization, May 8-9,2015 : Zurich Heart House, University Hospital Zurich, Swiss
Backup Support - Hyper shaft
Focusing on flexibility of the tip, soft resin material is applied in
Hyperion. While resin normally gets deformed due to heat,
ASAHI’s unique technology enables the catheter to have flexibility,
back up support, and anti-heat deformation.
Low← Water absorption→ High
Image: original polymer
Polyamide
Unique
processing
Soft←flexibility→Hard
Image: polymer processed
Polyamide
Low← Water absorption→ High
Soft←flexibility→Hard
12. 4th European Live Summit on Retrograde CTO Revascularization, May 8-9,2015 : Zurich Heart House, University Hospital Zurich, Swiss
Backup Support ~anti- heat deformation~
Test Method:
The curve is extended in warm water to a shape as if it is engaged, and the
catheter tip is fixed onto the axis of the force gauge. The back-up support
is evaluated by measuring the change of force in certain time intervals (0-
120 mins)
60%
65%
70%
75%
80%
85%
90%
95%
100%
0 0,5 1 5 10 30 60 90 120
Competitor A
Competitor B
Competitor C
Competitor D
Hyperion
[Time:min]
reduction
13. 4th European Live Summit on Retrograde CTO Revascularization, May 8-9,2015 : Zurich Heart House, University Hospital Zurich, Swiss
Backup Support ~2nd curve~
2nd curve with high rigidity provides good
retention of the curve.
14. 4th European Live Summit on Retrograde CTO Revascularization, May 8-9,2015 : Zurich Heart House, University Hospital Zurich, Swiss
Backup Support ~3rd curve~
Unique curve proximal to 2nd curve provides stable
backup support by increasing the area of the catheter
in contact with the aorta wall.
Conventional
product
Hyperion
(with3rdcurve)
Conventional product Hyperion
JL/AL shape
15. 4th European Live Summit on Retrograde CTO Revascularization, May 8-9,2015 : Zurich Heart House, University Hospital Zurich, Swiss
Maneuverability –HENKA-Braid
Unlike conventional guiding catheters, Hyperion’s shaft
is braided differently at the proximal and the distal area.
⇒ - enables transmission of the push force
- prevents backing out of the catheter
16. 4th European Live Summit on Retrograde CTO Revascularization, May 8-9,2015 : Zurich Heart House, University Hospital Zurich, Swiss
CompetitorG CompetitorF CompetitorH CompetitorI CompetitorZ
Competitor A CompetitorB CompetitorC CompetitorD CompetitorE
Braid appearance
未 未
HENKA-Braid
17. 4th European Live Summit on Retrograde CTO Revascularization, May 8-9,2015 : Zurich Heart House, University Hospital Zurich, Swiss
The curves behave
similarly, but
Competitor A, can only
withstand the force in a
shorter period of time,
and backs out sooner.
Hyperion 7Fr AL1.0
Backed out in 34 sec.
Competitor A 7Fr AL1.0
Backed out in 9 sec.
HENKA-Braid Persistent shaft
Observing the behavior of the tip
when applying torque to the
catheter while the tip is anchored
onto a stick.
18. 4th European Live Summit on Retrograde CTO Revascularization, May 8-9,2015 : Zurich Heart House, University Hospital Zurich, Swiss
Hyperion(34 sec.)
Shaft shows its
flexibility until right
before backing out.
Competitor B(30 sec.)
Stayed straight without changing its
form and then backed.
Competitor C(9 sec.)
Less persistency.
Bending curve just
before jumping out
Competitor A(11 sec.)
Less capacity even the shaft has
bended a little.
Competitor D(13 sec.)
Shaft has bended but couldn’t stay in its
place.
HENKA-Braid Persistent shaft
19. 4th European Live Summit on Retrograde CTO Revascularization, May 8-9,2015 : Zurich Heart House, University Hospital Zurich, Swiss
0,00
0,50
1,00
1,50
2,00
2,50
6Fr 7Fr 8Fr
Competitor A
Competitor B
Competitor C
Competitor D
1st Curve Flexibility 19
Force (Hyperion as default)
1.00
20. 4th European Live Summit on Retrograde CTO Revascularization, May 8-9,2015 : Zurich Heart House, University Hospital Zurich, Swiss
Flexible Tip -Material
Soft material & tungsten powder – same as that of the
ASAHI Corsair
Facilitates safe engagement and precise positioning due
to its flexibility and visibility.
21. 4th European Live Summit on Retrograde CTO Revascularization, May 8-9,2015 : Zurich Heart House, University Hospital Zurich, Swiss
All Round Processed Tip
All Round Processed Tip
Without sharp edges on the tip, the Hyperion can be
maneuvered safely.
Round processed tip of the
conventional products
All round processed Tip of
Hyperion
22. 4th European Live Summit on Retrograde CTO Revascularization, May 8-9,2015 : Zurich Heart House, University Hospital Zurich, Swiss
Features at a glance
Backup Support
Maneuverability
Flexible Tip
Industry Standard Inner/Outer Diameter
I.D. O.D.
6Fr
1.80mm
(0.071inch)
2.09mm
7Fr
2.05mm
(0.081inch)
2.40mm
8Fr
2.28mm
(0.090inch)
2.70mm
23. 4th European Live Summit on Retrograde CTO Revascularization, May 8-9,2015 : Zurich Heart House, University Hospital Zurich, Swiss
Lineup
A variety of shapes and sizes
are available!
24. 4th European Live Summit on Retrograde CTO Revascularization, May 8-9,2015 : Zurich Heart House, University Hospital Zurich, Swiss
Case presentation
25. 4th European Live Summit on Retrograde CTO Revascularization, May 8-9,2015 : Zurich Heart House, University Hospital Zurich, Swiss
Case 1 LAD CTO
Hyperion 7F PB3.5SH and 7F SAL1.0SHGaia1stable engagement until the end
26. 4th European Live Summit on Retrograde CTO Revascularization, May 8-9,2015 : Zurich Heart House, University Hospital Zurich, Swiss
Case 2 LAD CTO
Hyperion 7F SPB 35SHGaia 2Injection through Caravel over 7F SAL1 0SHFielder XT-RDirect crossing by Gaia2Clear visualizationSitting well even at the endSoft tip ensures safty
27. 4th European Live Summit on Retrograde CTO Revascularization, May 8-9,2015 : Zurich Heart House, University Hospital Zurich, Swiss
Conclusion
The Hyperion is a guiding catheter which
allows for safe, effective coaxial engagement
with superior back-up support
The Hyperion is a safe and effective guiding
catheter for all PCI operators.