Challenges in QC Medical Device Testing 
Instron® Biomedical Applications & Solutions 
September 2014
2 
Instron® Breadth of Knowledge to Evaluate Biomedical Product Lifecycles 
Orthopaedics 
•Research of materials for bone and joint replacement 
•Implants of all types (fatigue, wear and simulation) 
Biomaterials 
•Soft and hard native tissues, artificial materials, tissue engineering 
•Heavily university and research based 
Dentistry 
•Restorative and prosthodontic materials 
•Implants (fatigue and wear) 
Implants 
•Stents, heart valves, delivery systems, catheters, guide wires, pacemakers, ocular and auditory implants 
Consumables and Equipment 
•Hospital and surgical supplies, needles, sutures, tubing, packaging 
•Quality control and product development 
Pharmaceutical 
•Drug delivery systems, tablets, packaging 
•Quality control and product development
3 
Example: Cardiovascular Stents 
•Product Development 
•Prototype Evaluation 
•Manufacturing 
•Delivery
4 
Component Selection & Testing 
•Shape Memory Alloys (SMAs) are known for their large recoverable strains and are an ideal material for biomedical applications 
•Nitinol (NiTi), the most popular SMA, employed primarily for vascular stents, are studied extensively in tension 
•ASTM F2516 requires the materials to be loaded and unloaded in a cyclic test (as shown in the graph to the lower right) 
•Solutions: 
•Contacting or non-contacting extensometry to measure specimen strain (point to point) 
•Digital Image Correlation (DIC) to develop a full field strain map 
ASTM F2516 test method for NiTi in tension
5 
Component Selection & Testing 
•Radial compression and relaxation (crimping and uncrimping) of stents using RX fixture as an optional accessories on standard Instron® frames to simulate forces on stents during use 
RX fixture is an accessory on Instron frames
6 
Aortic Heart Valve 
Prototype Physiologic Testing 
A key factor for medical device success is biologic compatibility and physiologic function. 
Instron® bioreactor systems allow R&D teams to evaluate this criteria for new devices and replacement tissues on the bench before initiating expensive animal testing.
7 
•Fatigue testing 
(Simulate 10 years in vivo) 
•Tension 
•Compression 
•Torsion 
•Stent testing to match FEA Modeling 
•Tension, Torsion, Bending 
•Multiple Specimen Testing 
•Instron® actively involved in 
ASTM F04.30 committee 
•Fatigue systems that utilize linear electric technology (versus hydraulic oil) are the preferred solution for medical device labs: no oil, low noise, small footprint 
Final Product Durability Testing 
E10000 with 12-head accessory for durability testing of stents over millions of cycles
8 
Final Product Delivery Testing 
•Tortuous path represents vasculature 
•Usually glass or plastic fixture 
•Test stent deployment devices 
•Test friction of catheters (coatings) in a physiological environment 
•Machine can be used vertical or horizontal 
Tortuosity tester with custom grip control accessory to customize pressure (friction) at device opening
9 
3 Challenges in Biomedical QC Testing 
•Standards to follow and what others are doing in the market—what you test today may not be what you test tomorrow 
1) Knowing How to Test 
•Best practices to ensure a highest level of product quality 
2) Ensuring Quality of Results 
•Small investments to protect you for the future 
3) Regulatory Compliance
10 
Knowing How to Test 
Standards to follow and what others are doing in the market—what you test today may not be what you test tomorrow 
1
11 
Biomedical Applications: Orthopaedics 
Applications & Standards 
•Spinal Implants: ASTM F1717 
•Intervertebral Discs: ASTM F2346 
•Intervertebral Fusion: ASTM F2077 
•Hip Replacements: ISO 7208 
•Knee Replacements: ISO 14879 
•Hip Screws: ASTM F384 
Recommended solution for spinal implants testing 
Automated tuning wizard optimizes the system for a range of specimen stiffness. This allows for testing anything from hydrogels to bone on the same machine with optimal performance. 
Testing Tip
12 
Biomedical Applications: Dental 
•Dental implant fatigue to ISO 14801 
•Fixture for testing endosseous dental implants 
•2-15 Hz depending on materials 
Testing Tip 
A 30º angle of the dental implant during testing is considered to simulate the “worse case” scenario of loading conditions. 
Testing to ISO 14801 in a BioBath 
3- and 4-point bend testing on ceramic implants
13 
Biomedical Applications: Medical Consumables 
•Syringes, needles: ISO 7886-1 
•Sutures, staples, catheters: ISO 10334 
•Gloves, masks, shields: ASTM D3578, ASTM D3577 ASTM D6319, & ISO 11193 
•Bandages, tapes, pads 
•Machines (EKG, Glucose, DO, etc.) 
ISO 7886-1 specifies test conditions for aspiration and injection of fluids from hypodermic needles 
BioBox allows for testing at and above body temperature 
Testing Tip 
Testing at body temperature vs. room temperature can drastically effect results. See results from our demo lab:
14 
Biomedical Applications: Pharmaceuticals 
•Packaging: ASTM F88, ASTM D882, & ASTM D1894 
•Caps for bottles, syringes, vials 
•Residual seal force 
Mean Max Load 
Testing Tip 
Testing packaging often requires a fast test and detection of quick events (e.g. first and average peaks). Setting the system data rate and bandwidth appropriately are important. 
Different Data Rate and Bandwidth Effect Results
15 
Biomedical Applications: Biomaterials 
•Biological Materials 
•Bone, tendons, ligaments, muscles, skin, etc. 
•Synthetic Materials to Mimic Biomaterials 
•Hydrogels, polymer scaffolds, etc. 
Tensile testing tissue with pneumatic grips 
Testing Tip 
Biomaterials are inherently slippery and difficult to grip. To get a better grip, without damaging the material, add friction. This can be done with surfalloy jaw faces or by simply using sandpaper between the grip faces. Also, using grips that can allow you to modify grip pressure, consistently (like a pneumatic style) can help significantly improve repeatability of results.
16 
Common Medical Device QC Systems 
3300 Series 
An affordable line of electromechanical instruments designed to provide Instron quality, reliability, and repeatability at an affordable price. 
The premier family of electromechanical instruments designed to deliver superior accuracy and reliability, ergonomics, and upholds the Instron name as the most trusted in mechanical testing. 
5900 Series 
2300 Series 
A family of electromechanical systems designed specially for the quality control industry in China.
17 
Ensuring Quality of Your Results 
Best practices to ensure a highest level of product quality 
2
18 
Why is this Important? 
•Most common issue in global medical device QC labs 
•“Why aren’t my results the same?” 
•Quality, repeatability & reliability of test results are a MUST 
•Multiple test systems in labs 
•Multiple labs on site 
•Multiple geographical locations of labs 
•Multiple products being manufactured 
•Challenges? 
•Every sample/specimen/application is different 
•Every QC lab has unique testing processes & procedures 
•Every QC lab has a different set of critical results
19 
Qualitatively … Sources of Error 
Testing Tip 
Perform an audit of your lab, equipment ,and operating procedures to see which of these variables could be causing variability in your results!
20 
Quantitatively … GR&R 
•Once you have completed a lab audit you can measure the quality of the results using a GR&R study 
•GR&R = Gage Repeatability & Reliability 
•Non-destructive testing (typically) 
•Can be performed destructively, but BE CAREFUL! 
•Used to qualify and/or validate equipment 
•“Quantifies” total system error based on three components 
•Part variation (PV) 
•Equipment variation (EV) 
•Appraiser (operator) variation (AV) 
•GR&R threshold values (by AIAG) 
•R&R < 10% = good (can identify process changes) 
•10% < R&R < 30% = okay (can identify part variability) 
•R&R > 30% = bad (can’t identify part variability)
21 
Regulatory Compliance 
Small investments to protect you for the future 
3
22 
Traceability & Data Security 
•Instron® has been partnered with Xybion Corporation, experts in FDA compliance, for more than 10 years 
•Together, we created ComplianceBuilder™ 
•Instron Bluehill® 3, WaveMatrix™, and Partner™ Software with ComplianceBuilder allows for 21 CFR § 11 Compliance 
All methods & data are secured, with user sign-in and traceability of user, changes & all revisions of data and method files can be restored through the ComplianceBuilder file management system
23 
How does ComplianceBuilder™ Help? 
•Allows use of electronic records; without an electronic file management system – you need to store paper files in a secured location (paper cost, space cost) 
•ComplianceBuilder monitors data / methods used for Instron® systems and can produce a simple report to prove traceability for audit (required by FDA) 
•Relevant Files: Sample, Method, Raw Data & Report Files 
•Relevant Actions: Additions, Modifications & Deletions 
•If any change is made to a relevant Instron file, ComplianceBuilder will capture it in a secure database, ready to be retrieved for an FDA audit 
•Can use e-signature feature to sign-off on changes and modifications 
When a change is made to an Instron sample or method file, the user can log in with a unique ID and password, and also explain the reason for the change; all changes (who, when, what) are captured
24 
Equipment Validation and Documentation 
•21 CFR 820.72: Inspection, measuring, and testing equipment 
•This requires IQ/OQ 
IQ – Installation Qualification 
•Installation Conditions 
•Operation & Safety features 
•Provided system manuals and documentation 
•Environmental conditions 
•Appropriate services & utilities 
OQ – Operational Qualification 
•Validation of correct results 
•Transducer validation 
•Speed and displacement validation 
•Software functionality check 
As part of our standard service offerings, Instron® can provide all IQ/OQ documentation needed for IQ/OQ – saving you time and ensuring full compliance with standards.
25 
THANK YOU FOR YOUR TIME! 
Please contact your local 
Instron® Sales Representative 
with any questions.

More Related Content

More from Instron

Sheet Metal Testing Challenges
Sheet Metal Testing ChallengesSheet Metal Testing Challenges
Sheet Metal Testing ChallengesInstron
 
3 Challenges in Plastics Testing: Melt Flow, Heat Deflection Temperature, & I...
3 Challenges in Plastics Testing: Melt Flow, Heat Deflection Temperature, & I...3 Challenges in Plastics Testing: Melt Flow, Heat Deflection Temperature, & I...
3 Challenges in Plastics Testing: Melt Flow, Heat Deflection Temperature, & I...Instron
 
Introduction to Digital Image Correlation (DIC)
Introduction to Digital Image Correlation (DIC)Introduction to Digital Image Correlation (DIC)
Introduction to Digital Image Correlation (DIC)Instron
 
Composites Testing: Challenges & Solutions
Composites Testing: Challenges & SolutionsComposites Testing: Challenges & Solutions
Composites Testing: Challenges & SolutionsInstron
 
The Fundamentals of Rheology
The Fundamentals of RheologyThe Fundamentals of Rheology
The Fundamentals of RheologyInstron
 
Fundamentals of Melt Flow Testing
Fundamentals of Melt Flow TestingFundamentals of Melt Flow Testing
Fundamentals of Melt Flow TestingInstron
 
Understanding Changes to Key Plastics Testing Standards
Understanding Changes to Key Plastics Testing Standards Understanding Changes to Key Plastics Testing Standards
Understanding Changes to Key Plastics Testing Standards Instron
 
Challenges of Rebar Testing
Challenges of Rebar TestingChallenges of Rebar Testing
Challenges of Rebar TestingInstron
 
Metals Tensile Testing Standards: ISO 6892-1 ASTM E8/8M for Strain Control
Metals Tensile Testing Standards: ISO 6892-1 ASTM E8/8M for Strain ControlMetals Tensile Testing Standards: ISO 6892-1 ASTM E8/8M for Strain Control
Metals Tensile Testing Standards: ISO 6892-1 ASTM E8/8M for Strain ControlInstron
 
Challenges of Pipe & Tube Testing
Challenges of Pipe & Tube TestingChallenges of Pipe & Tube Testing
Challenges of Pipe & Tube TestingInstron
 

More from Instron (10)

Sheet Metal Testing Challenges
Sheet Metal Testing ChallengesSheet Metal Testing Challenges
Sheet Metal Testing Challenges
 
3 Challenges in Plastics Testing: Melt Flow, Heat Deflection Temperature, & I...
3 Challenges in Plastics Testing: Melt Flow, Heat Deflection Temperature, & I...3 Challenges in Plastics Testing: Melt Flow, Heat Deflection Temperature, & I...
3 Challenges in Plastics Testing: Melt Flow, Heat Deflection Temperature, & I...
 
Introduction to Digital Image Correlation (DIC)
Introduction to Digital Image Correlation (DIC)Introduction to Digital Image Correlation (DIC)
Introduction to Digital Image Correlation (DIC)
 
Composites Testing: Challenges & Solutions
Composites Testing: Challenges & SolutionsComposites Testing: Challenges & Solutions
Composites Testing: Challenges & Solutions
 
The Fundamentals of Rheology
The Fundamentals of RheologyThe Fundamentals of Rheology
The Fundamentals of Rheology
 
Fundamentals of Melt Flow Testing
Fundamentals of Melt Flow TestingFundamentals of Melt Flow Testing
Fundamentals of Melt Flow Testing
 
Understanding Changes to Key Plastics Testing Standards
Understanding Changes to Key Plastics Testing Standards Understanding Changes to Key Plastics Testing Standards
Understanding Changes to Key Plastics Testing Standards
 
Challenges of Rebar Testing
Challenges of Rebar TestingChallenges of Rebar Testing
Challenges of Rebar Testing
 
Metals Tensile Testing Standards: ISO 6892-1 ASTM E8/8M for Strain Control
Metals Tensile Testing Standards: ISO 6892-1 ASTM E8/8M for Strain ControlMetals Tensile Testing Standards: ISO 6892-1 ASTM E8/8M for Strain Control
Metals Tensile Testing Standards: ISO 6892-1 ASTM E8/8M for Strain Control
 
Challenges of Pipe & Tube Testing
Challenges of Pipe & Tube TestingChallenges of Pipe & Tube Testing
Challenges of Pipe & Tube Testing
 

Recently uploaded

Unit 1 - Soil Classification and Compaction.pdf
Unit 1 - Soil Classification and Compaction.pdfUnit 1 - Soil Classification and Compaction.pdf
Unit 1 - Soil Classification and Compaction.pdfRagavanV2
 
BSides Seattle 2024 - Stopping Ethan Hunt From Taking Your Data.pptx
BSides Seattle 2024 - Stopping Ethan Hunt From Taking Your Data.pptxBSides Seattle 2024 - Stopping Ethan Hunt From Taking Your Data.pptx
BSides Seattle 2024 - Stopping Ethan Hunt From Taking Your Data.pptxfenichawla
 
data_management_and _data_science_cheat_sheet.pdf
data_management_and _data_science_cheat_sheet.pdfdata_management_and _data_science_cheat_sheet.pdf
data_management_and _data_science_cheat_sheet.pdfJiananWang21
 
AKTU Computer Networks notes --- Unit 3.pdf
AKTU Computer Networks notes ---  Unit 3.pdfAKTU Computer Networks notes ---  Unit 3.pdf
AKTU Computer Networks notes --- Unit 3.pdfankushspencer015
 
Structural Analysis and Design of Foundations: A Comprehensive Handbook for S...
Structural Analysis and Design of Foundations: A Comprehensive Handbook for S...Structural Analysis and Design of Foundations: A Comprehensive Handbook for S...
Structural Analysis and Design of Foundations: A Comprehensive Handbook for S...Dr.Costas Sachpazis
 
VIP Model Call Girls Kothrud ( Pune ) Call ON 8005736733 Starting From 5K to ...
VIP Model Call Girls Kothrud ( Pune ) Call ON 8005736733 Starting From 5K to ...VIP Model Call Girls Kothrud ( Pune ) Call ON 8005736733 Starting From 5K to ...
VIP Model Call Girls Kothrud ( Pune ) Call ON 8005736733 Starting From 5K to ...SUHANI PANDEY
 
Intze Overhead Water Tank Design by Working Stress - IS Method.pdf
Intze Overhead Water Tank  Design by Working Stress - IS Method.pdfIntze Overhead Water Tank  Design by Working Stress - IS Method.pdf
Intze Overhead Water Tank Design by Working Stress - IS Method.pdfSuman Jyoti
 
Top Rated Pune Call Girls Budhwar Peth ⟟ 6297143586 ⟟ Call Me For Genuine Se...
Top Rated  Pune Call Girls Budhwar Peth ⟟ 6297143586 ⟟ Call Me For Genuine Se...Top Rated  Pune Call Girls Budhwar Peth ⟟ 6297143586 ⟟ Call Me For Genuine Se...
Top Rated Pune Call Girls Budhwar Peth ⟟ 6297143586 ⟟ Call Me For Genuine Se...Call Girls in Nagpur High Profile
 
Thermal Engineering Unit - I & II . ppt
Thermal Engineering  Unit - I & II . pptThermal Engineering  Unit - I & II . ppt
Thermal Engineering Unit - I & II . pptDineshKumar4165
 
Call for Papers - International Journal of Intelligent Systems and Applicatio...
Call for Papers - International Journal of Intelligent Systems and Applicatio...Call for Papers - International Journal of Intelligent Systems and Applicatio...
Call for Papers - International Journal of Intelligent Systems and Applicatio...Christo Ananth
 
Booking open Available Pune Call Girls Koregaon Park 6297143586 Call Hot Ind...
Booking open Available Pune Call Girls Koregaon Park  6297143586 Call Hot Ind...Booking open Available Pune Call Girls Koregaon Park  6297143586 Call Hot Ind...
Booking open Available Pune Call Girls Koregaon Park 6297143586 Call Hot Ind...Call Girls in Nagpur High Profile
 
UNIT-V FMM.HYDRAULIC TURBINE - Construction and working
UNIT-V FMM.HYDRAULIC TURBINE - Construction and workingUNIT-V FMM.HYDRAULIC TURBINE - Construction and working
UNIT-V FMM.HYDRAULIC TURBINE - Construction and workingrknatarajan
 
chapter 5.pptx: drainage and irrigation engineering
chapter 5.pptx: drainage and irrigation engineeringchapter 5.pptx: drainage and irrigation engineering
chapter 5.pptx: drainage and irrigation engineeringmulugeta48
 
Thermal Engineering-R & A / C - unit - V
Thermal Engineering-R & A / C - unit - VThermal Engineering-R & A / C - unit - V
Thermal Engineering-R & A / C - unit - VDineshKumar4165
 
FULL ENJOY Call Girls In Mahipalpur Delhi Contact Us 8377877756
FULL ENJOY Call Girls In Mahipalpur Delhi Contact Us 8377877756FULL ENJOY Call Girls In Mahipalpur Delhi Contact Us 8377877756
FULL ENJOY Call Girls In Mahipalpur Delhi Contact Us 8377877756dollysharma2066
 
Call for Papers - Educational Administration: Theory and Practice, E-ISSN: 21...
Call for Papers - Educational Administration: Theory and Practice, E-ISSN: 21...Call for Papers - Educational Administration: Theory and Practice, E-ISSN: 21...
Call for Papers - Educational Administration: Theory and Practice, E-ISSN: 21...Christo Ananth
 
The Most Attractive Pune Call Girls Manchar 8250192130 Will You Miss This Cha...
The Most Attractive Pune Call Girls Manchar 8250192130 Will You Miss This Cha...The Most Attractive Pune Call Girls Manchar 8250192130 Will You Miss This Cha...
The Most Attractive Pune Call Girls Manchar 8250192130 Will You Miss This Cha...ranjana rawat
 

Recently uploaded (20)

(INDIRA) Call Girl Bhosari Call Now 8617697112 Bhosari Escorts 24x7
(INDIRA) Call Girl Bhosari Call Now 8617697112 Bhosari Escorts 24x7(INDIRA) Call Girl Bhosari Call Now 8617697112 Bhosari Escorts 24x7
(INDIRA) Call Girl Bhosari Call Now 8617697112 Bhosari Escorts 24x7
 
Unit 1 - Soil Classification and Compaction.pdf
Unit 1 - Soil Classification and Compaction.pdfUnit 1 - Soil Classification and Compaction.pdf
Unit 1 - Soil Classification and Compaction.pdf
 
Water Industry Process Automation & Control Monthly - April 2024
Water Industry Process Automation & Control Monthly - April 2024Water Industry Process Automation & Control Monthly - April 2024
Water Industry Process Automation & Control Monthly - April 2024
 
BSides Seattle 2024 - Stopping Ethan Hunt From Taking Your Data.pptx
BSides Seattle 2024 - Stopping Ethan Hunt From Taking Your Data.pptxBSides Seattle 2024 - Stopping Ethan Hunt From Taking Your Data.pptx
BSides Seattle 2024 - Stopping Ethan Hunt From Taking Your Data.pptx
 
data_management_and _data_science_cheat_sheet.pdf
data_management_and _data_science_cheat_sheet.pdfdata_management_and _data_science_cheat_sheet.pdf
data_management_and _data_science_cheat_sheet.pdf
 
AKTU Computer Networks notes --- Unit 3.pdf
AKTU Computer Networks notes ---  Unit 3.pdfAKTU Computer Networks notes ---  Unit 3.pdf
AKTU Computer Networks notes --- Unit 3.pdf
 
Structural Analysis and Design of Foundations: A Comprehensive Handbook for S...
Structural Analysis and Design of Foundations: A Comprehensive Handbook for S...Structural Analysis and Design of Foundations: A Comprehensive Handbook for S...
Structural Analysis and Design of Foundations: A Comprehensive Handbook for S...
 
VIP Model Call Girls Kothrud ( Pune ) Call ON 8005736733 Starting From 5K to ...
VIP Model Call Girls Kothrud ( Pune ) Call ON 8005736733 Starting From 5K to ...VIP Model Call Girls Kothrud ( Pune ) Call ON 8005736733 Starting From 5K to ...
VIP Model Call Girls Kothrud ( Pune ) Call ON 8005736733 Starting From 5K to ...
 
Intze Overhead Water Tank Design by Working Stress - IS Method.pdf
Intze Overhead Water Tank  Design by Working Stress - IS Method.pdfIntze Overhead Water Tank  Design by Working Stress - IS Method.pdf
Intze Overhead Water Tank Design by Working Stress - IS Method.pdf
 
Top Rated Pune Call Girls Budhwar Peth ⟟ 6297143586 ⟟ Call Me For Genuine Se...
Top Rated  Pune Call Girls Budhwar Peth ⟟ 6297143586 ⟟ Call Me For Genuine Se...Top Rated  Pune Call Girls Budhwar Peth ⟟ 6297143586 ⟟ Call Me For Genuine Se...
Top Rated Pune Call Girls Budhwar Peth ⟟ 6297143586 ⟟ Call Me For Genuine Se...
 
Thermal Engineering Unit - I & II . ppt
Thermal Engineering  Unit - I & II . pptThermal Engineering  Unit - I & II . ppt
Thermal Engineering Unit - I & II . ppt
 
Call for Papers - International Journal of Intelligent Systems and Applicatio...
Call for Papers - International Journal of Intelligent Systems and Applicatio...Call for Papers - International Journal of Intelligent Systems and Applicatio...
Call for Papers - International Journal of Intelligent Systems and Applicatio...
 
(INDIRA) Call Girl Aurangabad Call Now 8617697112 Aurangabad Escorts 24x7
(INDIRA) Call Girl Aurangabad Call Now 8617697112 Aurangabad Escorts 24x7(INDIRA) Call Girl Aurangabad Call Now 8617697112 Aurangabad Escorts 24x7
(INDIRA) Call Girl Aurangabad Call Now 8617697112 Aurangabad Escorts 24x7
 
Booking open Available Pune Call Girls Koregaon Park 6297143586 Call Hot Ind...
Booking open Available Pune Call Girls Koregaon Park  6297143586 Call Hot Ind...Booking open Available Pune Call Girls Koregaon Park  6297143586 Call Hot Ind...
Booking open Available Pune Call Girls Koregaon Park 6297143586 Call Hot Ind...
 
UNIT-V FMM.HYDRAULIC TURBINE - Construction and working
UNIT-V FMM.HYDRAULIC TURBINE - Construction and workingUNIT-V FMM.HYDRAULIC TURBINE - Construction and working
UNIT-V FMM.HYDRAULIC TURBINE - Construction and working
 
chapter 5.pptx: drainage and irrigation engineering
chapter 5.pptx: drainage and irrigation engineeringchapter 5.pptx: drainage and irrigation engineering
chapter 5.pptx: drainage and irrigation engineering
 
Thermal Engineering-R & A / C - unit - V
Thermal Engineering-R & A / C - unit - VThermal Engineering-R & A / C - unit - V
Thermal Engineering-R & A / C - unit - V
 
FULL ENJOY Call Girls In Mahipalpur Delhi Contact Us 8377877756
FULL ENJOY Call Girls In Mahipalpur Delhi Contact Us 8377877756FULL ENJOY Call Girls In Mahipalpur Delhi Contact Us 8377877756
FULL ENJOY Call Girls In Mahipalpur Delhi Contact Us 8377877756
 
Call for Papers - Educational Administration: Theory and Practice, E-ISSN: 21...
Call for Papers - Educational Administration: Theory and Practice, E-ISSN: 21...Call for Papers - Educational Administration: Theory and Practice, E-ISSN: 21...
Call for Papers - Educational Administration: Theory and Practice, E-ISSN: 21...
 
The Most Attractive Pune Call Girls Manchar 8250192130 Will You Miss This Cha...
The Most Attractive Pune Call Girls Manchar 8250192130 Will You Miss This Cha...The Most Attractive Pune Call Girls Manchar 8250192130 Will You Miss This Cha...
The Most Attractive Pune Call Girls Manchar 8250192130 Will You Miss This Cha...
 

ChallengesQCMedicalDeviceTesting_Sept2014

  • 1. Challenges in QC Medical Device Testing Instron® Biomedical Applications & Solutions September 2014
  • 2. 2 Instron® Breadth of Knowledge to Evaluate Biomedical Product Lifecycles Orthopaedics •Research of materials for bone and joint replacement •Implants of all types (fatigue, wear and simulation) Biomaterials •Soft and hard native tissues, artificial materials, tissue engineering •Heavily university and research based Dentistry •Restorative and prosthodontic materials •Implants (fatigue and wear) Implants •Stents, heart valves, delivery systems, catheters, guide wires, pacemakers, ocular and auditory implants Consumables and Equipment •Hospital and surgical supplies, needles, sutures, tubing, packaging •Quality control and product development Pharmaceutical •Drug delivery systems, tablets, packaging •Quality control and product development
  • 3. 3 Example: Cardiovascular Stents •Product Development •Prototype Evaluation •Manufacturing •Delivery
  • 4. 4 Component Selection & Testing •Shape Memory Alloys (SMAs) are known for their large recoverable strains and are an ideal material for biomedical applications •Nitinol (NiTi), the most popular SMA, employed primarily for vascular stents, are studied extensively in tension •ASTM F2516 requires the materials to be loaded and unloaded in a cyclic test (as shown in the graph to the lower right) •Solutions: •Contacting or non-contacting extensometry to measure specimen strain (point to point) •Digital Image Correlation (DIC) to develop a full field strain map ASTM F2516 test method for NiTi in tension
  • 5. 5 Component Selection & Testing •Radial compression and relaxation (crimping and uncrimping) of stents using RX fixture as an optional accessories on standard Instron® frames to simulate forces on stents during use RX fixture is an accessory on Instron frames
  • 6. 6 Aortic Heart Valve Prototype Physiologic Testing A key factor for medical device success is biologic compatibility and physiologic function. Instron® bioreactor systems allow R&D teams to evaluate this criteria for new devices and replacement tissues on the bench before initiating expensive animal testing.
  • 7. 7 •Fatigue testing (Simulate 10 years in vivo) •Tension •Compression •Torsion •Stent testing to match FEA Modeling •Tension, Torsion, Bending •Multiple Specimen Testing •Instron® actively involved in ASTM F04.30 committee •Fatigue systems that utilize linear electric technology (versus hydraulic oil) are the preferred solution for medical device labs: no oil, low noise, small footprint Final Product Durability Testing E10000 with 12-head accessory for durability testing of stents over millions of cycles
  • 8. 8 Final Product Delivery Testing •Tortuous path represents vasculature •Usually glass or plastic fixture •Test stent deployment devices •Test friction of catheters (coatings) in a physiological environment •Machine can be used vertical or horizontal Tortuosity tester with custom grip control accessory to customize pressure (friction) at device opening
  • 9. 9 3 Challenges in Biomedical QC Testing •Standards to follow and what others are doing in the market—what you test today may not be what you test tomorrow 1) Knowing How to Test •Best practices to ensure a highest level of product quality 2) Ensuring Quality of Results •Small investments to protect you for the future 3) Regulatory Compliance
  • 10. 10 Knowing How to Test Standards to follow and what others are doing in the market—what you test today may not be what you test tomorrow 1
  • 11. 11 Biomedical Applications: Orthopaedics Applications & Standards •Spinal Implants: ASTM F1717 •Intervertebral Discs: ASTM F2346 •Intervertebral Fusion: ASTM F2077 •Hip Replacements: ISO 7208 •Knee Replacements: ISO 14879 •Hip Screws: ASTM F384 Recommended solution for spinal implants testing Automated tuning wizard optimizes the system for a range of specimen stiffness. This allows for testing anything from hydrogels to bone on the same machine with optimal performance. Testing Tip
  • 12. 12 Biomedical Applications: Dental •Dental implant fatigue to ISO 14801 •Fixture for testing endosseous dental implants •2-15 Hz depending on materials Testing Tip A 30º angle of the dental implant during testing is considered to simulate the “worse case” scenario of loading conditions. Testing to ISO 14801 in a BioBath 3- and 4-point bend testing on ceramic implants
  • 13. 13 Biomedical Applications: Medical Consumables •Syringes, needles: ISO 7886-1 •Sutures, staples, catheters: ISO 10334 •Gloves, masks, shields: ASTM D3578, ASTM D3577 ASTM D6319, & ISO 11193 •Bandages, tapes, pads •Machines (EKG, Glucose, DO, etc.) ISO 7886-1 specifies test conditions for aspiration and injection of fluids from hypodermic needles BioBox allows for testing at and above body temperature Testing Tip Testing at body temperature vs. room temperature can drastically effect results. See results from our demo lab:
  • 14. 14 Biomedical Applications: Pharmaceuticals •Packaging: ASTM F88, ASTM D882, & ASTM D1894 •Caps for bottles, syringes, vials •Residual seal force Mean Max Load Testing Tip Testing packaging often requires a fast test and detection of quick events (e.g. first and average peaks). Setting the system data rate and bandwidth appropriately are important. Different Data Rate and Bandwidth Effect Results
  • 15. 15 Biomedical Applications: Biomaterials •Biological Materials •Bone, tendons, ligaments, muscles, skin, etc. •Synthetic Materials to Mimic Biomaterials •Hydrogels, polymer scaffolds, etc. Tensile testing tissue with pneumatic grips Testing Tip Biomaterials are inherently slippery and difficult to grip. To get a better grip, without damaging the material, add friction. This can be done with surfalloy jaw faces or by simply using sandpaper between the grip faces. Also, using grips that can allow you to modify grip pressure, consistently (like a pneumatic style) can help significantly improve repeatability of results.
  • 16. 16 Common Medical Device QC Systems 3300 Series An affordable line of electromechanical instruments designed to provide Instron quality, reliability, and repeatability at an affordable price. The premier family of electromechanical instruments designed to deliver superior accuracy and reliability, ergonomics, and upholds the Instron name as the most trusted in mechanical testing. 5900 Series 2300 Series A family of electromechanical systems designed specially for the quality control industry in China.
  • 17. 17 Ensuring Quality of Your Results Best practices to ensure a highest level of product quality 2
  • 18. 18 Why is this Important? •Most common issue in global medical device QC labs •“Why aren’t my results the same?” •Quality, repeatability & reliability of test results are a MUST •Multiple test systems in labs •Multiple labs on site •Multiple geographical locations of labs •Multiple products being manufactured •Challenges? •Every sample/specimen/application is different •Every QC lab has unique testing processes & procedures •Every QC lab has a different set of critical results
  • 19. 19 Qualitatively … Sources of Error Testing Tip Perform an audit of your lab, equipment ,and operating procedures to see which of these variables could be causing variability in your results!
  • 20. 20 Quantitatively … GR&R •Once you have completed a lab audit you can measure the quality of the results using a GR&R study •GR&R = Gage Repeatability & Reliability •Non-destructive testing (typically) •Can be performed destructively, but BE CAREFUL! •Used to qualify and/or validate equipment •“Quantifies” total system error based on three components •Part variation (PV) •Equipment variation (EV) •Appraiser (operator) variation (AV) •GR&R threshold values (by AIAG) •R&R < 10% = good (can identify process changes) •10% < R&R < 30% = okay (can identify part variability) •R&R > 30% = bad (can’t identify part variability)
  • 21. 21 Regulatory Compliance Small investments to protect you for the future 3
  • 22. 22 Traceability & Data Security •Instron® has been partnered with Xybion Corporation, experts in FDA compliance, for more than 10 years •Together, we created ComplianceBuilder™ •Instron Bluehill® 3, WaveMatrix™, and Partner™ Software with ComplianceBuilder allows for 21 CFR § 11 Compliance All methods & data are secured, with user sign-in and traceability of user, changes & all revisions of data and method files can be restored through the ComplianceBuilder file management system
  • 23. 23 How does ComplianceBuilder™ Help? •Allows use of electronic records; without an electronic file management system – you need to store paper files in a secured location (paper cost, space cost) •ComplianceBuilder monitors data / methods used for Instron® systems and can produce a simple report to prove traceability for audit (required by FDA) •Relevant Files: Sample, Method, Raw Data & Report Files •Relevant Actions: Additions, Modifications & Deletions •If any change is made to a relevant Instron file, ComplianceBuilder will capture it in a secure database, ready to be retrieved for an FDA audit •Can use e-signature feature to sign-off on changes and modifications When a change is made to an Instron sample or method file, the user can log in with a unique ID and password, and also explain the reason for the change; all changes (who, when, what) are captured
  • 24. 24 Equipment Validation and Documentation •21 CFR 820.72: Inspection, measuring, and testing equipment •This requires IQ/OQ IQ – Installation Qualification •Installation Conditions •Operation & Safety features •Provided system manuals and documentation •Environmental conditions •Appropriate services & utilities OQ – Operational Qualification •Validation of correct results •Transducer validation •Speed and displacement validation •Software functionality check As part of our standard service offerings, Instron® can provide all IQ/OQ documentation needed for IQ/OQ – saving you time and ensuring full compliance with standards.
  • 25. 25 THANK YOU FOR YOUR TIME! Please contact your local Instron® Sales Representative with any questions.