SlideShare a Scribd company logo
1 of 71
ADVANCES IN
CHROMATOGRAPHIC
TECHNIQUES
MEENAL AGGARWAL
L-2011-BS-72-IM
1
HISTORY
o Tswett used chromatography to separate
plant pigments (1906)
o colorful separation of plant pigments was
done using a column of calcium
carbonate(chalk)
o the new technique was called
chromatography because the result of the
analysis was written in color.(Chroma
means color and graphein means to write)
Mikhail Tswett,
Russian Botanist,
1872-1919
2
CHROMATOGRAPHY
Technique used to separate and identify the
components of a mixture
PRINCIPLE: Works by allowing the molecules
present in the mixture to distribute themselves
between a stationary and a mobile medium.
Molecules that spend most of their time in the
mobile phase are carried along faster.
3
Components:
mobile phase: a solvent that flows through the supporting
medium
stationary phase: a layer or coating on the supporting
medium that interacts with the analytes
supporting medium: a solid surface on which the stationary
phase is bound or coated
4
BASIC TERMS….
 Adsorbtion: Interaction of solute molecules with the
surface of the stationary phase
 Eluent: The mobile phase
 Elution: Motion of the mobile phase through the
stationary phase
 Elution time: The time taken for a solute to pass
through the system. A solute with a short elution time
travels through the stationary phase rapidly, i.e. it elutes
fast
 Stationary phase: The part of the chromatography
system that is fixed in place
5
 Normal phase: “Unmodified” stationary phase where
POLAR solutes interact strongly and run slowly
 Reverse phase: “Modified” stationary phase where
POLAR solutes run fast i.e. reverse order
 Resolution: Degree of separation of different solutes.
In principle, resolution can be improved by using a
longer stationary phase, finer stationary phase or
slower elution.
 Rf value: distance travelled by solute
distance travelled by solvent
 Rf = retardation factor. The Rf value has to be
between 0 and 1, and it is typically reported to two
decimal places.
6
CLASSIFICATION
On the basis of
interaction of
solute to the
stationary phase
On the basis of
chromatographic
bed shape
Techniques by
physical state of
mobile phase 7
On the basis of
interaction of
solute to the
stationary phase
Adsorption
chromatography
Partition
chromatography
Ion exchange
chromatography
Molecular
exclusion
chromatography
8
On the basis of
chromatographic bed shape
Column
chromatography
TLC
Paper
chromatography
9
Techniques by
physical state of
mobile phase
Gas
chromatography
Liquid
chromatography
Affinity
chromatography
Supercritical
fluid
chromatography 10
ADSORPTION CHROMATOGRAPHY
 Principle of separation:
utilizes a mobile liquid or
gaseous phase that is
adsorbed onto the surface
of a stationary solid phase
 Stationary phase:
adsorbent filled in a tube
(column)
 Mobile phase: various
solvents (eluents)
11
 One of the oldest type
 solute molecules bond directly to the surface of the
stationary phase
 Stationary phases may contain a variety of
adsorption sites differing in the tenacity with which
they bind the molecules and in their relative
abundance
12
PARTITION
CHROMATOGRAPHY
PRINCIPLE
partiton of component of sample between
sample and liquid/gas stationary phase retard
some components of sample more as compared
to others. This gives the basis of separation
13
 Based on thin film formed
on the surface of a solid
support by a liquid
stationary phase
 Solute equilibrates
between mobile phase
and stationary liquid
phase
14
ION EXCHANGE CHROMATOGRAPHY
 Resin( stationary phase)
used to covalently attach
anions or cations onto it
 Solute ions of the
opposite charge in the
mobile liquid phase are
attracted to the resin by
electrostatic forces
 Ion exchange mechanism
separates analytes based
on their respective
charges
15
MECHANISM
 To optimize binding of all charged molecules, the mobile
phase is generally a low to medium conductivity (i.e., low to
medium salt concentration) solution
 adsorption of the molecules to the solid support is driven by
the ionic interaction between the oppositely charged ionic
groups in the sample molecule and in the functional ligand on
the support
 The strength of the interaction is determined by the number
and location of the charges on the molecule and on the
functional group
 By increasing the salt concentration the molecules with the
weakest ionic interactions start to elute from the column first
 Molecules that have a stronger ionic interaction require a
higher salt concentration and elute later in the gradient
 The binding capacities of ion exchange resins are generally
quite high
16
MOLECULAR EXCLUSION
CHROMATOGRAPHY
 Also known as gel
permeation or gel filtration
chromatography
 Lacks attractive
interaction between solute
and stationary phase
17
 Liquid or gaseous phase passes a porous gel which
separates the molecule according to its size
 The pores are normally small and exclude the
larger solute molecules, but allows the smaller
molecules to enter the gel, causing them to flow
through a larger volume. This causes the larger
molecules to pass through column at faster rate than smaller
ones
 It is generally low resolution technique and thus it is
often reserved for the final “polishing” step of a
purification
18
COLUMN CHROMATOGRAPHY
Stationary phase is
held in a narrow
tube through which
the mobile phase
is forced under
pressure or under
the effect of gravity
19
Factors affecting solutes separation in CC
( Factors affecting column efficiency)
EffectFactor
Decrease of size improves separation (but very small
particles need high pressure).
Particle size of solid stationary
phase (or of support)
Efficiency increases as ratio length / width increases.Column dimensions
Non uniform packing results in irregular movement
of solutes through column & less uniform zone
formation, (i.e. band broadning or tailing).
Uniformity of packing
Increase in column temperature results in speed of
elution but does not improve separation (tailing).
Column temperature
Solvents should be of low viscosity (to give efficient
resolution) & h igh volatility (to get rapid recovery of
the substances).
Eluting solvent
Uniform & low flow rate gives better resolution.Solvent flow rate
Discontinuous flow disturbs resolutionContinuity of flow
Deactivation of adsorbent decreases separation.Condition of adsorbent
Substances of high concentration move slowly.Concentration of solutes 20
ProcedureTechnique
Addition of solvent mixture of fixed composition
during the whole process.
Isocratic elution
Continuous or linear elution: in which there is
continuous change in the composition of the
mobile phase over a period of time (e.g. polarity,
pH or ionic strength).
Gradient elution
Step wise or fractional elution: in which the
change is not continuous i.e. a sudden change in
the composition of the mobile phase is followed
by a period where the mobile phase is held
constant.
Elution techniques
21
PAPER CHROMATOGRAPHY
uses filter paper strips as carrier or
inert support.
The factor governing separation of
mixtures of solutes on filter
paper is the partition between
two immiscible phases.
One is usually water adsorbed on
cellulose fibres in the paper
(stationary phase).
The second is the organic solvent
flows past the sample on the
paper (stationary phase).
22
PAPER CHROMATOGRAPHY
 a small dot or line of sample solution is placed onto a
strip of chromatography paper
 The paper is placed in a jar containing a shallow layer of
solvent and sealed
 As the solvent rises through the paper, it meets the
sample mixture which starts to travel up the paper with
the solvent.
 This paper is made of cellulose, a polar substance,
 and the compounds within the mixture travel farther if
they
are non-polar
More polar substances bond with the cellulose paper
more quickly, and therefore do not travel as far 23
THIN LAYER CHROMATOGRAPHY (TLC)
involves a stationary phase
of a thin layer of
adsorbent like silica gel,
alumina, or cellulose on a
flat, inert substrate
24
25
HPTLC
 Sophisticated form of thin layer chromatography. It
involves the same theoretical principle of thin layer
chromatography.
 It is also known as planar chromatography or Flat-bed
chromatography.
 Traditional Thin Layer Chromatography & its modern
instrumental quantitative analysis version HPTLC are
very popular for many reasons such as
 visual chromatogram
 simplicity
 multiple sample handling
 low running and maintenance costs, disposable layer
etc.
26
PRINCIPLE
o HPTLC have similar approach and employ the same
physical principles of TLC (adsorption chromatography)
i.e. the principle of separation is adsorption.
o Solvent flows through because of capillary action
o Components move according to their affinities towards the
adsorbent Component with more affinity towards the
stationary phase travels slower
o Component with lesser affinity towards the stationary
phase travels faster
Thus the components are separated on a chromatographic
plate 27
STEPS INVOLVING IN HPTLC
Sample Preparation
Selection of
chromatography layer
Pre-washing
Pre-conditioning
Application of
sample
Chromatography development
Detection of spots
Scanning & documentation
28
Pre washing of pre coated plates
main purpose of the pre-washing is to remove
impurities which include water vapours and other
volatile substances from the atmosphere when they
get exposed in the lab environment.
Silica gel 60F is most widely used sorbent
major disadvantage of this sorbent is that it contain
iron as impurity
This iron is removed by using Methanol : water in the
ratio of 9:1.This is the major advantage of the step of
pre-washing.
29
ACTIVATION OF PLATES
 Freshly opened box of HPTLC plates doesn’t need
activation.
 Plates exposed to high humidity or kept in hand for
long time require activation.
 Plates are placed in oven at 110o-120oc for 30 min
prior to the sample application.
30
PRE-CONDITIONING
• Also called Chamber Saturation
•Un- saturated chamber causes high Rf values
Sample application
• Usual concentration range is 0.1-1µg / µl
Above this causes poor separation
• sample and standard application from
syringe on TLC plates as bands
• Band wise application - better separation
31
POST CHROMATOGRAPHY STEPS
Detection
Photo
documentation
Densitometry
measurements
32
DETECTION
UV CABINET
Detection under UV light is first choice - non
destructive
- Spots of fluorescent compounds can be seen at 254
nm (short wave length) or at 366 nm (long wave
length)
- Spots of non fluorescent compounds can be seen -
fluorescent stationary phase is used - silica gel GF
33
DENSITOMETRY MEASUREMENTS
34
 Measures visible, UV absorbance or
Fluorescence.
 Convert the spot/band into chromatogram
consisting of peaks
Instrumentation of HPTLC consists of following:
Lamp selector
Entrance lens slit
Monochromator entry slit
Grating
Mirror
Slit aperture disc
Mirror
Beam splitter
Reference photo multiplier
Measuring photo multiplier
Photo diode for transmission measurements.
35
DIFFERENCES BETWEEN TLC AND
HPTLC
36
DISPLACEMENT CHROMATOGRAPHY
 The basic principle of displacement chromatography is,
“A molecule with a high affinity for the chromatography
matrix (the displacer) will compete effectively for binding
sites, and thus displace all molecules with lesser
affinities”
 Displacement chromatography has advantages over
elution chromatography
 components are resolved into consecutive zones of pure
substances rather than “peaks”.
 Because the process takes advantage of the
nonlinearity of the isotherms, a larger column feed can
be separated on a given column with the purified
components recovered at significantly higher
concentrations 37
GAS CHROMATOGRAPHY (GC)
 Also called as Gas liquid chromatography (GLC)
 based on a partition equilibrium of analyte between
a solid stationary phase (often a liquid silicone-
based material) and a mobile gas (most often
Helium).
 The stationary phase is adhered to the inside of a
small-diameter glass tube (a capillary column) or a
solid matrix inside a larger metal tube (a packed
column).
38
 high temperatures used in GC make it unsuitable for
high molecular weight biopolymers or proteins (heat
will denature them)
 Well suited for use in the petrochemical, environmental
monitoring, and industrial chemical fields. It is also
used extensively in chemistry research.
 Here the mobile phase is an unreactive gas ( e.g.
Nitrogen) flowing through a tube.
 And the stationary phase is an involatile liquid held on
particles of a solid support.
39
40
In the animation below the red molecules are more soluble in the
liquid (or less volatile) than are the green molecules.
41
INSTRUMENTATION:
o Mobile Phase: He, Ar, N2, H2
o Flow regulators & flow meters:
o Injection Port: Rubber septum barrier (usually
maintained at a higher temperature than the boiling point
of the least volatile component in the sample mixture)
o Column: (fused silica with a thin coating of stationary
phase on the inner surface)
o Oven: Thermostat controlled forced air oven
o Detector:
o Data System: recorders & integrators
42
COLUMNS:
Columns in GC are two types
1) packed column
2) capillary column
Packed column
(capillary column)
Open tubular column
43
Packed column:
Glass or metals
2-3 m long, 2-4 mm I'd.
Densely packed with packing
materials or solid support coated
with thin layer of stationary liquid
phase
Diatomaceous earth
Size: 60-80 mesh (250-170 m)
or 80-100 mesh (170-149 m)
Better resolution – efficient mass
transfer between gas and SP
Tubing – fused silica, glass, copper,
stainless steel
Open tubular column
44
Liquid phase
Solid support coated
with liquid phase
Porous Adsorbent
Porous Layer Open
Tubular
(PLOT)
Wall-coated Open
Tubular
(WCOT)
Support-coated Open
Tubular
(SCOT)
Types of open tubular column:
45
GC DETECTORS
 TCD (thermal conductivity detector)
 FID (Flame ionization Detector)
 MSD (Mass Selective Detector)
 ECD (Electron Capture Detector)
 NPD (Nitrogen Phosphorus Detector)
 FTIRD (Fourier transformation infrared detector)
 AED (Atomic emission detector)
46
GC-MS
 Gas chromatography-Mass spectroscopy (GC-MS) is
one of the so-called hyphenated analytical technique
 As the name implies, it is actually two techniques that
are combined to form a single method of analyzing
mixtures of chemicals
 Gas chromatography is a technique capable of
separating, detecting and partially characterizing the
organic compounds particularly when present in small
quantity.
 Mass spectroscopy provides some definite structural
information from in small quantity.
47
 Combination of GC-MS provides extremely powerful
tool because it permits direct and effectively
continuous correlation of chromatographic and mass
spectroscopic properties
 The separation and identification of the components
of complex natural and synthetic mixture are
achieved more quickly than any other technique with
less sample
48
PRINCIPLE OF GC-MS
 The sample solution is injected into the GC inlet
where it is vaporized and swept onto a
chromatographic column by the carrier gas (usually
helium).
 The sample flows through the column and the
compounds comprising the mixture of interest are
separated by virtue of their relative interaction with the
coating of the column (stationary phase) and the
carrier gas (mobile phase).
 The latter part of the column passes through a heated
transfer line and ends at the entrance to ion source
where compounds eluting from the column are
converted to ions.
49
GC
Computer
MS
GC-MS
50
GC-MS INSTRUMENT
The insides of the GC-MS, with the column of the gas
chromatograph in the oven on the right. 51
LIQUID CHROMATOGRAPHY
 separation technique in which the mobile phase is a
liquid.
 can be carried out either in a column or a plane.
 Present day it utilizes very small packing particles and a
relatively high pressure is referred as high performance
liquid chromatography (HPLC).
 In the HPLC technique, the sample is forced through a
column that is packed with irregularly or spherically
shaped particles or a porous monolithic layer (stationary
phase) by a liquid (mobile phase) at high pressure.
52
SUPERCRITICAL
FLUIDCHROMATOGRAPHY
 Supercritical fluid chromatography is a separation technique in
which the mobile phase is a fluid above and relatively close to
its critical temperature and pressure
53
AFFINITY CHROMATOGRAPHY
Based on selective non-covalent interaction between an analyte
and specific molecules
Stationary Phase
 Highly specific and selective
 Molecular recognition groups (affinity ligands) covalently
attached to agarose or cellulose beads
 Affinity ligand binds selectively and reversibly the analyte
 Mechanism of retention: interaction with highly specialized
molecular recognition systems which are attached to the
stationary phase
Mobile Phase
 Two distinct roles
– Support the strong binding of the analyte molecule to the
ligand
– Weaken and eliminate the analyte-ligand interaction
54
HPLC
 one of the most powerful tools in analytical chemistry
 has the ability to separate, identify, and quantitate the
compounds that are present in any sample that can be
dissolved in a liquid
 Similar to gas chromatography but uses a liquid mobile
phase
 The stationary phase is usually an inert solid or a liquid
held on the inert solid
 Mobile phase travels through the column forcibly with
the aid of the high pressure pump
 Solutes of the sample separated on column and eluted
with mobile phase
 The technique is applicable to thermally fragile samples,
e.g. amino acids, proteins, nucleic acids, hydrocarbons,
antibiotics, steroids, drugs, inorganic and may organic
substances.
55
56
.
COLUMNS
 Stainless steel, or heavily walled glass tubing of 10-
30 cm, capable of withstanding 6000 psi, 4 to 10
mm inside dia meter, 0.5 μm particles provide 40-
60000 plates/meter.
 1 - 4.6 mm dia, 3.5 mm particles, 3-5 cm length
columns offer 1, 00,000 Plates/meter. Such
columns have low solvent consumption and speed
up the separations
57
Filter and separate irreversibly bonding
compounds.
The particles have similar composition to that of
the analytical column but slightly larger particle
size to minimize pressure drop.
Column thermostats must be capable of
controlling ± 0.10 C. This is accomplished by
using column heaters or water jackets
GUARD COLUMN:
58
DETECTION METHODS
􀂄 UV – Ultraviolet light--- most popular
– Lamp
– Grating/Lens - Wave length 190-350 nm
– FlowCell
– PhotoDiode - Differential Light Output
􀂄 RI – Refractive Index
– Universal analyte detector
– Solvent must remain the same throughout separation
– VERY temperature sensitive
– Sometimes difficult to stabilize baseline
􀂄 FD – Fluorescence-greater sensitivity, not so popular
– Excitation wavelength generates fluorescence emission at a higher wavelength
– Analytes must have fluorophore group---not very common
– Very sensitive and selective
􀂄 MS – Mass Spectrometry
– Mass to charge ratio (m/z)
– Allows specific compound ID
59
NORMAL-PHASE HPLC
 Adsorption of analytes on the polar, weakly acidic
surface of silica gel
 Stationary Phase.: Silica (pH 2-8), Alumina (pH 2 -
12), Bonded Diol and NH2
 Mobile Phase: Non-polar solvents (Hexane, CHCl3)
 Applications: Non-polar and semi-polar samples;
hexane soluble; positional isomers
 Polar solutes elute later than non-polar lyophilic
ones
60
REVERSED-PHASE HPLC
 Partition of analytes between mobile phase and
stagnant phase inside the pore space + adsorption
on the surface of bonded phase
 Stationary Phase: Hydrophobic surfaces of
moieties bonded on silica (C18, C8, C5, Phenyl,
CN)
 Mobile phase: Methanol or Acetonitrile and Water
 Applications: ~80% of all separations done on RP
HPLC
 organic molecules are separated based on their
degree of hydrophobicity
61
UPLC
 In 2004, further advances in instrumentation and
column technology were made to achieve very
significant increase in:
 RESOLUTION
 SPEED
 SENSITIVITY
 Increase separation EFFICIENCY
 Columns with smaller particles [<1.7um]
 Mobile phase delivery is done at >15,000psi
62
CONTRASTING HPLC AND UPLC
 UPLC gives faster results with better resolution
 UPLC uses less of valuable solvents like
acetonitrile which lowers cost
 The reduction of solvent use is more
environmentally friendly
63
64
INSTRUMENTATION
 Autosampler – Reduced Cycle Time and niglisible
carryove
 Flow cell
 Van guar column – for protecting column
performance
 Tubings – Special for UPLC, reduced volume,
optimized connection
 Mobile Phase Filters – To prevent system damage
 column
 Detectors
65
SCHEMATIC PRESENTATION OF UPLC
Gradient
Controller
Pump
•
Column
Detector
Injector
Mobile Phases
66
TUBING AND FITTING
 Tubing and fittings are carefully designed to
provide leak-free connections, with low dead
volume, and minimal band spreading. For best
performance, these components must be clean,
and designed to work together.
67
MOBILE PHASE FILTERS
 Good practice to always filter your
solvents
 As it prevent system damage
 Filters should be changed periodically
depending on usage and mobile phase
68
DETECTORS
 Evaporative Light Scattering (ELS) Detector
 Fluorescence (FLR) Detector
 Photodiode Array (PDA) Detector
 Tunable UV (TUV) Detector
 Single Quadrapole Detector (SQD)
69
ADVANTAGES OF UPLC
 Decreases run time and increases sensitivity
 Provides the selectivity, sensitivity, and dynamic range of LC analysis
 Maintaining resolution performance.
 Expands scope of Multiresidue Methods
 UPLC’s fast resolving power quickly quantifies related and unrelated
compounds
 Faster analysis through the use of a novel separation material of very
fine particle size
 Operation cost is reduced
 Less solvent consumption
 Reduces process cycle times, so that more product can be produced
with existing resources
 Delivers real-time analysis in step with manufacturing processes
 Assures end-product quality, including final release testing
70
Disadvantages OF UPLC
Due to increased pressure requires more
maintenance and reduces the life of the
columns of this type.
THANK YOU
71

More Related Content

What's hot

Chromatography and its types
Chromatography and its typesChromatography and its types
Chromatography and its typesnadeem akhter
 
(Gel Filtration Chromatography)GFC
(Gel Filtration Chromatography)GFC(Gel Filtration Chromatography)GFC
(Gel Filtration Chromatography)GFCAthira athira
 
High performance liquid chromatography (HPLC)
High performance liquid chromatography (HPLC)High performance liquid chromatography (HPLC)
High performance liquid chromatography (HPLC)Htet Wai Moe
 
HPLC - High Performance Liquid Chromatography
HPLC - High Performance Liquid ChromatographyHPLC - High Performance Liquid Chromatography
HPLC - High Performance Liquid ChromatographyDivya Basuti
 
Principles and application of chromatography
Principles and application of chromatographyPrinciples and application of chromatography
Principles and application of chromatographysuniu
 
Affinity chromatography ppt
Affinity chromatography pptAffinity chromatography ppt
Affinity chromatography pptpoojakamble1609
 
Applications of thin layer chromatography ppt by ann
Applications of thin layer chromatography ppt by annApplications of thin layer chromatography ppt by ann
Applications of thin layer chromatography ppt by annAnjali Rarichan
 
Column chromatography
Column chromatographyColumn chromatography
Column chromatographyRakesh Guptha
 
Capillary Electrophoresis
Capillary ElectrophoresisCapillary Electrophoresis
Capillary ElectrophoresisSantoshi10
 
TLC, thin layer chromatography
TLC, thin layer chromatographyTLC, thin layer chromatography
TLC, thin layer chromatographyshaisejacob
 
Electrophoresis, principle and types
Electrophoresis, principle and typesElectrophoresis, principle and types
Electrophoresis, principle and typesKAUSHAL SAHU
 

What's hot (20)

HPLC
HPLCHPLC
HPLC
 
Chromatography and its types
Chromatography and its typesChromatography and its types
Chromatography and its types
 
(Gel Filtration Chromatography)GFC
(Gel Filtration Chromatography)GFC(Gel Filtration Chromatography)GFC
(Gel Filtration Chromatography)GFC
 
Gas chromatography
Gas chromatographyGas chromatography
Gas chromatography
 
High performance liquid chromatography (HPLC)
High performance liquid chromatography (HPLC)High performance liquid chromatography (HPLC)
High performance liquid chromatography (HPLC)
 
HPLC - High Performance Liquid Chromatography
HPLC - High Performance Liquid ChromatographyHPLC - High Performance Liquid Chromatography
HPLC - High Performance Liquid Chromatography
 
Principles and application of chromatography
Principles and application of chromatographyPrinciples and application of chromatography
Principles and application of chromatography
 
Affinity chromatography ppt
Affinity chromatography pptAffinity chromatography ppt
Affinity chromatography ppt
 
Mass spectroscopy
Mass spectroscopyMass spectroscopy
Mass spectroscopy
 
Paper chromatography
Paper chromatographyPaper chromatography
Paper chromatography
 
Nmr spectroscopy
Nmr spectroscopyNmr spectroscopy
Nmr spectroscopy
 
Applications of thin layer chromatography ppt by ann
Applications of thin layer chromatography ppt by annApplications of thin layer chromatography ppt by ann
Applications of thin layer chromatography ppt by ann
 
Ion exchange chromatography
Ion  exchange chromatographyIon  exchange chromatography
Ion exchange chromatography
 
Column chromatography
Column chromatographyColumn chromatography
Column chromatography
 
Capillary Electrophoresis
Capillary ElectrophoresisCapillary Electrophoresis
Capillary Electrophoresis
 
TLC, thin layer chromatography
TLC, thin layer chromatographyTLC, thin layer chromatography
TLC, thin layer chromatography
 
Paper chromatography
Paper chromatographyPaper chromatography
Paper chromatography
 
Thin layer chromatography
Thin layer chromatographyThin layer chromatography
Thin layer chromatography
 
Liquid chromatography
Liquid chromatographyLiquid chromatography
Liquid chromatography
 
Electrophoresis, principle and types
Electrophoresis, principle and typesElectrophoresis, principle and types
Electrophoresis, principle and types
 

Viewers also liked

Attracting Curiosity - EAInnovates2015
Attracting Curiosity - EAInnovates2015Attracting Curiosity - EAInnovates2015
Attracting Curiosity - EAInnovates2015James Tiffin Jr.
 
Magnets in orthodontics.1
Magnets in orthodontics.1Magnets in orthodontics.1
Magnets in orthodontics.1shafeeq rahman
 
Provoking Original Ideas from Student Makers
Provoking Original Ideas from Student Makers Provoking Original Ideas from Student Makers
Provoking Original Ideas from Student Makers James Tiffin Jr.
 
Presentation on principle of paper chromatography and Rf Value
Presentation on principle of paper chromatography and Rf ValuePresentation on principle of paper chromatography and Rf Value
Presentation on principle of paper chromatography and Rf ValueManoj Kumar Tekuri
 
Types of chromatographic methods
Types of chromatographic methodsTypes of chromatographic methods
Types of chromatographic methodssumit prajapati
 
Paper Chromatography
Paper ChromatographyPaper Chromatography
Paper ChromatographyDavid Genis
 
Thin layer chromatography
Thin layer chromatographyThin layer chromatography
Thin layer chromatographyGuruprasad Rao
 
Thin layer chromatography (tlc)
Thin layer chromatography (tlc)Thin layer chromatography (tlc)
Thin layer chromatography (tlc)Dr.M.Prasad Naidu
 
Classification of chromatography
Classification of chromatographyClassification of chromatography
Classification of chromatographykhadeeja ikram01
 
Paper chromatography(final)
Paper chromatography(final)Paper chromatography(final)
Paper chromatography(final)Myquie de Guzman
 
Paper Chromatography PPT (new)
Paper Chromatography PPT (new)Paper Chromatography PPT (new)
Paper Chromatography PPT (new)shaisejacob
 

Viewers also liked (18)

Chromatography
ChromatographyChromatography
Chromatography
 
Paper chromatography (2)
Paper chromatography (2)Paper chromatography (2)
Paper chromatography (2)
 
Attracting Curiosity - EAInnovates2015
Attracting Curiosity - EAInnovates2015Attracting Curiosity - EAInnovates2015
Attracting Curiosity - EAInnovates2015
 
Magnets in orthodontics.1
Magnets in orthodontics.1Magnets in orthodontics.1
Magnets in orthodontics.1
 
Provoking Original Ideas from Student Makers
Provoking Original Ideas from Student Makers Provoking Original Ideas from Student Makers
Provoking Original Ideas from Student Makers
 
Presentation on principle of paper chromatography and Rf Value
Presentation on principle of paper chromatography and Rf ValuePresentation on principle of paper chromatography and Rf Value
Presentation on principle of paper chromatography and Rf Value
 
Magnets
MagnetsMagnets
Magnets
 
Paper chromatography
Paper chromatographyPaper chromatography
Paper chromatography
 
Magnets
MagnetsMagnets
Magnets
 
Types of chromatographic methods
Types of chromatographic methodsTypes of chromatographic methods
Types of chromatographic methods
 
Paper chromatography experiment
Paper chromatography experimentPaper chromatography experiment
Paper chromatography experiment
 
Paper Chromatography
Paper ChromatographyPaper Chromatography
Paper Chromatography
 
TLC
TLCTLC
TLC
 
Thin layer chromatography
Thin layer chromatographyThin layer chromatography
Thin layer chromatography
 
Thin layer chromatography (tlc)
Thin layer chromatography (tlc)Thin layer chromatography (tlc)
Thin layer chromatography (tlc)
 
Classification of chromatography
Classification of chromatographyClassification of chromatography
Classification of chromatography
 
Paper chromatography(final)
Paper chromatography(final)Paper chromatography(final)
Paper chromatography(final)
 
Paper Chromatography PPT (new)
Paper Chromatography PPT (new)Paper Chromatography PPT (new)
Paper Chromatography PPT (new)
 

Similar to chromatographic techniques

column chromatography and ultra performance liquid chromatography
column chromatography and ultra performance liquid chromatographycolumn chromatography and ultra performance liquid chromatography
column chromatography and ultra performance liquid chromatographyHenisha Patel
 
Chromatography-PPT.pptx
Chromatography-PPT.pptxChromatography-PPT.pptx
Chromatography-PPT.pptxNagen87
 
chromotography-140516031516-phpapp02.pdf
chromotography-140516031516-phpapp02.pdfchromotography-140516031516-phpapp02.pdf
chromotography-140516031516-phpapp02.pdfPratyushNahak
 
DSE-2, ANALYTICAL METHODS -Ch-II.pptx
DSE-2, ANALYTICAL METHODS -Ch-II.pptxDSE-2, ANALYTICAL METHODS -Ch-II.pptx
DSE-2, ANALYTICAL METHODS -Ch-II.pptxMathabhanga College
 
Chromatography-PPT.pptx
Chromatography-PPT.pptxChromatography-PPT.pptx
Chromatography-PPT.pptxAmul29
 
Classification of Chromatography
Classification of ChromatographyClassification of Chromatography
Classification of Chromatographykhadeeja ikram01
 
Chromatography introduction
Chromatography introductionChromatography introduction
Chromatography introductionDr. Samia
 
Classification of chromatography
Classification of chromatographyClassification of chromatography
Classification of chromatographyJagdish Jat
 
Partition chromatography 3
Partition chromatography 3Partition chromatography 3
Partition chromatography 3MrSyedAmmar
 
Introduction to chromatography and its applications 2
Introduction to chromatography and its applications 2Introduction to chromatography and its applications 2
Introduction to chromatography and its applications 2Kalsoom Mohammed
 
chromatography and its application.ppt
chromatography and its application.pptchromatography and its application.ppt
chromatography and its application.pptDenekeAlemayehu
 
CHM260 - Separations Methods
CHM260 - Separations MethodsCHM260 - Separations Methods
CHM260 - Separations MethodsAlia Najiha
 

Similar to chromatographic techniques (20)

column chromatography and ultra performance liquid chromatography
column chromatography and ultra performance liquid chromatographycolumn chromatography and ultra performance liquid chromatography
column chromatography and ultra performance liquid chromatography
 
Chromatography-PPT.pptx
Chromatography-PPT.pptxChromatography-PPT.pptx
Chromatography-PPT.pptx
 
Chromatography-PPT.pptx
Chromatography-PPT.pptxChromatography-PPT.pptx
Chromatography-PPT.pptx
 
Chromotography
ChromotographyChromotography
Chromotography
 
chromotography-140516031516-phpapp02.pdf
chromotography-140516031516-phpapp02.pdfchromotography-140516031516-phpapp02.pdf
chromotography-140516031516-phpapp02.pdf
 
chromatography
chromatographychromatography
chromatography
 
Chromatography
ChromatographyChromatography
Chromatography
 
DSE-2, ANALYTICAL METHODS -Ch-II.pptx
DSE-2, ANALYTICAL METHODS -Ch-II.pptxDSE-2, ANALYTICAL METHODS -Ch-II.pptx
DSE-2, ANALYTICAL METHODS -Ch-II.pptx
 
Chromatography-PPT.pptx
Chromatography-PPT.pptxChromatography-PPT.pptx
Chromatography-PPT.pptx
 
Classification of Chromatography
Classification of ChromatographyClassification of Chromatography
Classification of Chromatography
 
Chromatography introduction
Chromatography introductionChromatography introduction
Chromatography introduction
 
Classification of chromatography
Classification of chromatographyClassification of chromatography
Classification of chromatography
 
Chromatography.pptx
Chromatography.pptxChromatography.pptx
Chromatography.pptx
 
Chromatograph yfinal
Chromatograph yfinalChromatograph yfinal
Chromatograph yfinal
 
Partition chromatography 3
Partition chromatography 3Partition chromatography 3
Partition chromatography 3
 
Introduction to chromatography and its applications 2
Introduction to chromatography and its applications 2Introduction to chromatography and its applications 2
Introduction to chromatography and its applications 2
 
chromatography and its application.ppt
chromatography and its application.pptchromatography and its application.ppt
chromatography and its application.ppt
 
HPLC
HPLCHPLC
HPLC
 
CHM260 - Separations Methods
CHM260 - Separations MethodsCHM260 - Separations Methods
CHM260 - Separations Methods
 
Chromatography
ChromatographyChromatography
Chromatography
 

Recently uploaded

SKILL OF INTRODUCING THE LESSON MICRO SKILLS.pptx
SKILL OF INTRODUCING THE LESSON MICRO SKILLS.pptxSKILL OF INTRODUCING THE LESSON MICRO SKILLS.pptx
SKILL OF INTRODUCING THE LESSON MICRO SKILLS.pptxAmanpreet Kaur
 
Making communications land - Are they received and understood as intended? we...
Making communications land - Are they received and understood as intended? we...Making communications land - Are they received and understood as intended? we...
Making communications land - Are they received and understood as intended? we...Association for Project Management
 
Micro-Scholarship, What it is, How can it help me.pdf
Micro-Scholarship, What it is, How can it help me.pdfMicro-Scholarship, What it is, How can it help me.pdf
Micro-Scholarship, What it is, How can it help me.pdfPoh-Sun Goh
 
Application orientated numerical on hev.ppt
Application orientated numerical on hev.pptApplication orientated numerical on hev.ppt
Application orientated numerical on hev.pptRamjanShidvankar
 
Interdisciplinary_Insights_Data_Collection_Methods.pptx
Interdisciplinary_Insights_Data_Collection_Methods.pptxInterdisciplinary_Insights_Data_Collection_Methods.pptx
Interdisciplinary_Insights_Data_Collection_Methods.pptxPooja Bhuva
 
How to Create and Manage Wizard in Odoo 17
How to Create and Manage Wizard in Odoo 17How to Create and Manage Wizard in Odoo 17
How to Create and Manage Wizard in Odoo 17Celine George
 
How to Give a Domain for a Field in Odoo 17
How to Give a Domain for a Field in Odoo 17How to Give a Domain for a Field in Odoo 17
How to Give a Domain for a Field in Odoo 17Celine George
 
Sensory_Experience_and_Emotional_Resonance_in_Gabriel_Okaras_The_Piano_and_Th...
Sensory_Experience_and_Emotional_Resonance_in_Gabriel_Okaras_The_Piano_and_Th...Sensory_Experience_and_Emotional_Resonance_in_Gabriel_Okaras_The_Piano_and_Th...
Sensory_Experience_and_Emotional_Resonance_in_Gabriel_Okaras_The_Piano_and_Th...Pooja Bhuva
 
Food safety_Challenges food safety laboratories_.pdf
Food safety_Challenges food safety laboratories_.pdfFood safety_Challenges food safety laboratories_.pdf
Food safety_Challenges food safety laboratories_.pdfSherif Taha
 
Wellbeing inclusion and digital dystopias.pptx
Wellbeing inclusion and digital dystopias.pptxWellbeing inclusion and digital dystopias.pptx
Wellbeing inclusion and digital dystopias.pptxJisc
 
Towards a code of practice for AI in AT.pptx
Towards a code of practice for AI in AT.pptxTowards a code of practice for AI in AT.pptx
Towards a code of practice for AI in AT.pptxJisc
 
How to Manage Global Discount in Odoo 17 POS
How to Manage Global Discount in Odoo 17 POSHow to Manage Global Discount in Odoo 17 POS
How to Manage Global Discount in Odoo 17 POSCeline George
 
Unit 3 Emotional Intelligence and Spiritual Intelligence.pdf
Unit 3 Emotional Intelligence and Spiritual Intelligence.pdfUnit 3 Emotional Intelligence and Spiritual Intelligence.pdf
Unit 3 Emotional Intelligence and Spiritual Intelligence.pdfDr Vijay Vishwakarma
 
This PowerPoint helps students to consider the concept of infinity.
This PowerPoint helps students to consider the concept of infinity.This PowerPoint helps students to consider the concept of infinity.
This PowerPoint helps students to consider the concept of infinity.christianmathematics
 
REMIFENTANIL: An Ultra short acting opioid.pptx
REMIFENTANIL: An Ultra short acting opioid.pptxREMIFENTANIL: An Ultra short acting opioid.pptx
REMIFENTANIL: An Ultra short acting opioid.pptxDr. Ravikiran H M Gowda
 
Google Gemini An AI Revolution in Education.pptx
Google Gemini An AI Revolution in Education.pptxGoogle Gemini An AI Revolution in Education.pptx
Google Gemini An AI Revolution in Education.pptxDr. Sarita Anand
 
FSB Advising Checklist - Orientation 2024
FSB Advising Checklist - Orientation 2024FSB Advising Checklist - Orientation 2024
FSB Advising Checklist - Orientation 2024Elizabeth Walsh
 
ICT Role in 21st Century Education & its Challenges.pptx
ICT Role in 21st Century Education & its Challenges.pptxICT Role in 21st Century Education & its Challenges.pptx
ICT Role in 21st Century Education & its Challenges.pptxAreebaZafar22
 
Salient Features of India constitution especially power and functions
Salient Features of India constitution especially power and functionsSalient Features of India constitution especially power and functions
Salient Features of India constitution especially power and functionsKarakKing
 
2024-NATIONAL-LEARNING-CAMP-AND-OTHER.pptx
2024-NATIONAL-LEARNING-CAMP-AND-OTHER.pptx2024-NATIONAL-LEARNING-CAMP-AND-OTHER.pptx
2024-NATIONAL-LEARNING-CAMP-AND-OTHER.pptxMaritesTamaniVerdade
 

Recently uploaded (20)

SKILL OF INTRODUCING THE LESSON MICRO SKILLS.pptx
SKILL OF INTRODUCING THE LESSON MICRO SKILLS.pptxSKILL OF INTRODUCING THE LESSON MICRO SKILLS.pptx
SKILL OF INTRODUCING THE LESSON MICRO SKILLS.pptx
 
Making communications land - Are they received and understood as intended? we...
Making communications land - Are they received and understood as intended? we...Making communications land - Are they received and understood as intended? we...
Making communications land - Are they received and understood as intended? we...
 
Micro-Scholarship, What it is, How can it help me.pdf
Micro-Scholarship, What it is, How can it help me.pdfMicro-Scholarship, What it is, How can it help me.pdf
Micro-Scholarship, What it is, How can it help me.pdf
 
Application orientated numerical on hev.ppt
Application orientated numerical on hev.pptApplication orientated numerical on hev.ppt
Application orientated numerical on hev.ppt
 
Interdisciplinary_Insights_Data_Collection_Methods.pptx
Interdisciplinary_Insights_Data_Collection_Methods.pptxInterdisciplinary_Insights_Data_Collection_Methods.pptx
Interdisciplinary_Insights_Data_Collection_Methods.pptx
 
How to Create and Manage Wizard in Odoo 17
How to Create and Manage Wizard in Odoo 17How to Create and Manage Wizard in Odoo 17
How to Create and Manage Wizard in Odoo 17
 
How to Give a Domain for a Field in Odoo 17
How to Give a Domain for a Field in Odoo 17How to Give a Domain for a Field in Odoo 17
How to Give a Domain for a Field in Odoo 17
 
Sensory_Experience_and_Emotional_Resonance_in_Gabriel_Okaras_The_Piano_and_Th...
Sensory_Experience_and_Emotional_Resonance_in_Gabriel_Okaras_The_Piano_and_Th...Sensory_Experience_and_Emotional_Resonance_in_Gabriel_Okaras_The_Piano_and_Th...
Sensory_Experience_and_Emotional_Resonance_in_Gabriel_Okaras_The_Piano_and_Th...
 
Food safety_Challenges food safety laboratories_.pdf
Food safety_Challenges food safety laboratories_.pdfFood safety_Challenges food safety laboratories_.pdf
Food safety_Challenges food safety laboratories_.pdf
 
Wellbeing inclusion and digital dystopias.pptx
Wellbeing inclusion and digital dystopias.pptxWellbeing inclusion and digital dystopias.pptx
Wellbeing inclusion and digital dystopias.pptx
 
Towards a code of practice for AI in AT.pptx
Towards a code of practice for AI in AT.pptxTowards a code of practice for AI in AT.pptx
Towards a code of practice for AI in AT.pptx
 
How to Manage Global Discount in Odoo 17 POS
How to Manage Global Discount in Odoo 17 POSHow to Manage Global Discount in Odoo 17 POS
How to Manage Global Discount in Odoo 17 POS
 
Unit 3 Emotional Intelligence and Spiritual Intelligence.pdf
Unit 3 Emotional Intelligence and Spiritual Intelligence.pdfUnit 3 Emotional Intelligence and Spiritual Intelligence.pdf
Unit 3 Emotional Intelligence and Spiritual Intelligence.pdf
 
This PowerPoint helps students to consider the concept of infinity.
This PowerPoint helps students to consider the concept of infinity.This PowerPoint helps students to consider the concept of infinity.
This PowerPoint helps students to consider the concept of infinity.
 
REMIFENTANIL: An Ultra short acting opioid.pptx
REMIFENTANIL: An Ultra short acting opioid.pptxREMIFENTANIL: An Ultra short acting opioid.pptx
REMIFENTANIL: An Ultra short acting opioid.pptx
 
Google Gemini An AI Revolution in Education.pptx
Google Gemini An AI Revolution in Education.pptxGoogle Gemini An AI Revolution in Education.pptx
Google Gemini An AI Revolution in Education.pptx
 
FSB Advising Checklist - Orientation 2024
FSB Advising Checklist - Orientation 2024FSB Advising Checklist - Orientation 2024
FSB Advising Checklist - Orientation 2024
 
ICT Role in 21st Century Education & its Challenges.pptx
ICT Role in 21st Century Education & its Challenges.pptxICT Role in 21st Century Education & its Challenges.pptx
ICT Role in 21st Century Education & its Challenges.pptx
 
Salient Features of India constitution especially power and functions
Salient Features of India constitution especially power and functionsSalient Features of India constitution especially power and functions
Salient Features of India constitution especially power and functions
 
2024-NATIONAL-LEARNING-CAMP-AND-OTHER.pptx
2024-NATIONAL-LEARNING-CAMP-AND-OTHER.pptx2024-NATIONAL-LEARNING-CAMP-AND-OTHER.pptx
2024-NATIONAL-LEARNING-CAMP-AND-OTHER.pptx
 

chromatographic techniques

  • 2. HISTORY o Tswett used chromatography to separate plant pigments (1906) o colorful separation of plant pigments was done using a column of calcium carbonate(chalk) o the new technique was called chromatography because the result of the analysis was written in color.(Chroma means color and graphein means to write) Mikhail Tswett, Russian Botanist, 1872-1919 2
  • 3. CHROMATOGRAPHY Technique used to separate and identify the components of a mixture PRINCIPLE: Works by allowing the molecules present in the mixture to distribute themselves between a stationary and a mobile medium. Molecules that spend most of their time in the mobile phase are carried along faster. 3
  • 4. Components: mobile phase: a solvent that flows through the supporting medium stationary phase: a layer or coating on the supporting medium that interacts with the analytes supporting medium: a solid surface on which the stationary phase is bound or coated 4
  • 5. BASIC TERMS….  Adsorbtion: Interaction of solute molecules with the surface of the stationary phase  Eluent: The mobile phase  Elution: Motion of the mobile phase through the stationary phase  Elution time: The time taken for a solute to pass through the system. A solute with a short elution time travels through the stationary phase rapidly, i.e. it elutes fast  Stationary phase: The part of the chromatography system that is fixed in place 5
  • 6.  Normal phase: “Unmodified” stationary phase where POLAR solutes interact strongly and run slowly  Reverse phase: “Modified” stationary phase where POLAR solutes run fast i.e. reverse order  Resolution: Degree of separation of different solutes. In principle, resolution can be improved by using a longer stationary phase, finer stationary phase or slower elution.  Rf value: distance travelled by solute distance travelled by solvent  Rf = retardation factor. The Rf value has to be between 0 and 1, and it is typically reported to two decimal places. 6
  • 7. CLASSIFICATION On the basis of interaction of solute to the stationary phase On the basis of chromatographic bed shape Techniques by physical state of mobile phase 7
  • 8. On the basis of interaction of solute to the stationary phase Adsorption chromatography Partition chromatography Ion exchange chromatography Molecular exclusion chromatography 8
  • 9. On the basis of chromatographic bed shape Column chromatography TLC Paper chromatography 9
  • 10. Techniques by physical state of mobile phase Gas chromatography Liquid chromatography Affinity chromatography Supercritical fluid chromatography 10
  • 11. ADSORPTION CHROMATOGRAPHY  Principle of separation: utilizes a mobile liquid or gaseous phase that is adsorbed onto the surface of a stationary solid phase  Stationary phase: adsorbent filled in a tube (column)  Mobile phase: various solvents (eluents) 11
  • 12.  One of the oldest type  solute molecules bond directly to the surface of the stationary phase  Stationary phases may contain a variety of adsorption sites differing in the tenacity with which they bind the molecules and in their relative abundance 12
  • 13. PARTITION CHROMATOGRAPHY PRINCIPLE partiton of component of sample between sample and liquid/gas stationary phase retard some components of sample more as compared to others. This gives the basis of separation 13
  • 14.  Based on thin film formed on the surface of a solid support by a liquid stationary phase  Solute equilibrates between mobile phase and stationary liquid phase 14
  • 15. ION EXCHANGE CHROMATOGRAPHY  Resin( stationary phase) used to covalently attach anions or cations onto it  Solute ions of the opposite charge in the mobile liquid phase are attracted to the resin by electrostatic forces  Ion exchange mechanism separates analytes based on their respective charges 15
  • 16. MECHANISM  To optimize binding of all charged molecules, the mobile phase is generally a low to medium conductivity (i.e., low to medium salt concentration) solution  adsorption of the molecules to the solid support is driven by the ionic interaction between the oppositely charged ionic groups in the sample molecule and in the functional ligand on the support  The strength of the interaction is determined by the number and location of the charges on the molecule and on the functional group  By increasing the salt concentration the molecules with the weakest ionic interactions start to elute from the column first  Molecules that have a stronger ionic interaction require a higher salt concentration and elute later in the gradient  The binding capacities of ion exchange resins are generally quite high 16
  • 17. MOLECULAR EXCLUSION CHROMATOGRAPHY  Also known as gel permeation or gel filtration chromatography  Lacks attractive interaction between solute and stationary phase 17
  • 18.  Liquid or gaseous phase passes a porous gel which separates the molecule according to its size  The pores are normally small and exclude the larger solute molecules, but allows the smaller molecules to enter the gel, causing them to flow through a larger volume. This causes the larger molecules to pass through column at faster rate than smaller ones  It is generally low resolution technique and thus it is often reserved for the final “polishing” step of a purification 18
  • 19. COLUMN CHROMATOGRAPHY Stationary phase is held in a narrow tube through which the mobile phase is forced under pressure or under the effect of gravity 19
  • 20. Factors affecting solutes separation in CC ( Factors affecting column efficiency) EffectFactor Decrease of size improves separation (but very small particles need high pressure). Particle size of solid stationary phase (or of support) Efficiency increases as ratio length / width increases.Column dimensions Non uniform packing results in irregular movement of solutes through column & less uniform zone formation, (i.e. band broadning or tailing). Uniformity of packing Increase in column temperature results in speed of elution but does not improve separation (tailing). Column temperature Solvents should be of low viscosity (to give efficient resolution) & h igh volatility (to get rapid recovery of the substances). Eluting solvent Uniform & low flow rate gives better resolution.Solvent flow rate Discontinuous flow disturbs resolutionContinuity of flow Deactivation of adsorbent decreases separation.Condition of adsorbent Substances of high concentration move slowly.Concentration of solutes 20
  • 21. ProcedureTechnique Addition of solvent mixture of fixed composition during the whole process. Isocratic elution Continuous or linear elution: in which there is continuous change in the composition of the mobile phase over a period of time (e.g. polarity, pH or ionic strength). Gradient elution Step wise or fractional elution: in which the change is not continuous i.e. a sudden change in the composition of the mobile phase is followed by a period where the mobile phase is held constant. Elution techniques 21
  • 22. PAPER CHROMATOGRAPHY uses filter paper strips as carrier or inert support. The factor governing separation of mixtures of solutes on filter paper is the partition between two immiscible phases. One is usually water adsorbed on cellulose fibres in the paper (stationary phase). The second is the organic solvent flows past the sample on the paper (stationary phase). 22
  • 23. PAPER CHROMATOGRAPHY  a small dot or line of sample solution is placed onto a strip of chromatography paper  The paper is placed in a jar containing a shallow layer of solvent and sealed  As the solvent rises through the paper, it meets the sample mixture which starts to travel up the paper with the solvent.  This paper is made of cellulose, a polar substance,  and the compounds within the mixture travel farther if they are non-polar More polar substances bond with the cellulose paper more quickly, and therefore do not travel as far 23
  • 24. THIN LAYER CHROMATOGRAPHY (TLC) involves a stationary phase of a thin layer of adsorbent like silica gel, alumina, or cellulose on a flat, inert substrate 24
  • 25. 25
  • 26. HPTLC  Sophisticated form of thin layer chromatography. It involves the same theoretical principle of thin layer chromatography.  It is also known as planar chromatography or Flat-bed chromatography.  Traditional Thin Layer Chromatography & its modern instrumental quantitative analysis version HPTLC are very popular for many reasons such as  visual chromatogram  simplicity  multiple sample handling  low running and maintenance costs, disposable layer etc. 26
  • 27. PRINCIPLE o HPTLC have similar approach and employ the same physical principles of TLC (adsorption chromatography) i.e. the principle of separation is adsorption. o Solvent flows through because of capillary action o Components move according to their affinities towards the adsorbent Component with more affinity towards the stationary phase travels slower o Component with lesser affinity towards the stationary phase travels faster Thus the components are separated on a chromatographic plate 27
  • 28. STEPS INVOLVING IN HPTLC Sample Preparation Selection of chromatography layer Pre-washing Pre-conditioning Application of sample Chromatography development Detection of spots Scanning & documentation 28
  • 29. Pre washing of pre coated plates main purpose of the pre-washing is to remove impurities which include water vapours and other volatile substances from the atmosphere when they get exposed in the lab environment. Silica gel 60F is most widely used sorbent major disadvantage of this sorbent is that it contain iron as impurity This iron is removed by using Methanol : water in the ratio of 9:1.This is the major advantage of the step of pre-washing. 29
  • 30. ACTIVATION OF PLATES  Freshly opened box of HPTLC plates doesn’t need activation.  Plates exposed to high humidity or kept in hand for long time require activation.  Plates are placed in oven at 110o-120oc for 30 min prior to the sample application. 30
  • 31. PRE-CONDITIONING • Also called Chamber Saturation •Un- saturated chamber causes high Rf values Sample application • Usual concentration range is 0.1-1µg / µl Above this causes poor separation • sample and standard application from syringe on TLC plates as bands • Band wise application - better separation 31
  • 33. DETECTION UV CABINET Detection under UV light is first choice - non destructive - Spots of fluorescent compounds can be seen at 254 nm (short wave length) or at 366 nm (long wave length) - Spots of non fluorescent compounds can be seen - fluorescent stationary phase is used - silica gel GF 33
  • 34. DENSITOMETRY MEASUREMENTS 34  Measures visible, UV absorbance or Fluorescence.  Convert the spot/band into chromatogram consisting of peaks
  • 35. Instrumentation of HPTLC consists of following: Lamp selector Entrance lens slit Monochromator entry slit Grating Mirror Slit aperture disc Mirror Beam splitter Reference photo multiplier Measuring photo multiplier Photo diode for transmission measurements. 35
  • 36. DIFFERENCES BETWEEN TLC AND HPTLC 36
  • 37. DISPLACEMENT CHROMATOGRAPHY  The basic principle of displacement chromatography is, “A molecule with a high affinity for the chromatography matrix (the displacer) will compete effectively for binding sites, and thus displace all molecules with lesser affinities”  Displacement chromatography has advantages over elution chromatography  components are resolved into consecutive zones of pure substances rather than “peaks”.  Because the process takes advantage of the nonlinearity of the isotherms, a larger column feed can be separated on a given column with the purified components recovered at significantly higher concentrations 37
  • 38. GAS CHROMATOGRAPHY (GC)  Also called as Gas liquid chromatography (GLC)  based on a partition equilibrium of analyte between a solid stationary phase (often a liquid silicone- based material) and a mobile gas (most often Helium).  The stationary phase is adhered to the inside of a small-diameter glass tube (a capillary column) or a solid matrix inside a larger metal tube (a packed column). 38
  • 39.  high temperatures used in GC make it unsuitable for high molecular weight biopolymers or proteins (heat will denature them)  Well suited for use in the petrochemical, environmental monitoring, and industrial chemical fields. It is also used extensively in chemistry research.  Here the mobile phase is an unreactive gas ( e.g. Nitrogen) flowing through a tube.  And the stationary phase is an involatile liquid held on particles of a solid support. 39
  • 40. 40
  • 41. In the animation below the red molecules are more soluble in the liquid (or less volatile) than are the green molecules. 41
  • 42. INSTRUMENTATION: o Mobile Phase: He, Ar, N2, H2 o Flow regulators & flow meters: o Injection Port: Rubber septum barrier (usually maintained at a higher temperature than the boiling point of the least volatile component in the sample mixture) o Column: (fused silica with a thin coating of stationary phase on the inner surface) o Oven: Thermostat controlled forced air oven o Detector: o Data System: recorders & integrators 42
  • 43. COLUMNS: Columns in GC are two types 1) packed column 2) capillary column Packed column (capillary column) Open tubular column 43
  • 44. Packed column: Glass or metals 2-3 m long, 2-4 mm I'd. Densely packed with packing materials or solid support coated with thin layer of stationary liquid phase Diatomaceous earth Size: 60-80 mesh (250-170 m) or 80-100 mesh (170-149 m) Better resolution – efficient mass transfer between gas and SP Tubing – fused silica, glass, copper, stainless steel Open tubular column 44
  • 45. Liquid phase Solid support coated with liquid phase Porous Adsorbent Porous Layer Open Tubular (PLOT) Wall-coated Open Tubular (WCOT) Support-coated Open Tubular (SCOT) Types of open tubular column: 45
  • 46. GC DETECTORS  TCD (thermal conductivity detector)  FID (Flame ionization Detector)  MSD (Mass Selective Detector)  ECD (Electron Capture Detector)  NPD (Nitrogen Phosphorus Detector)  FTIRD (Fourier transformation infrared detector)  AED (Atomic emission detector) 46
  • 47. GC-MS  Gas chromatography-Mass spectroscopy (GC-MS) is one of the so-called hyphenated analytical technique  As the name implies, it is actually two techniques that are combined to form a single method of analyzing mixtures of chemicals  Gas chromatography is a technique capable of separating, detecting and partially characterizing the organic compounds particularly when present in small quantity.  Mass spectroscopy provides some definite structural information from in small quantity. 47
  • 48.  Combination of GC-MS provides extremely powerful tool because it permits direct and effectively continuous correlation of chromatographic and mass spectroscopic properties  The separation and identification of the components of complex natural and synthetic mixture are achieved more quickly than any other technique with less sample 48
  • 49. PRINCIPLE OF GC-MS  The sample solution is injected into the GC inlet where it is vaporized and swept onto a chromatographic column by the carrier gas (usually helium).  The sample flows through the column and the compounds comprising the mixture of interest are separated by virtue of their relative interaction with the coating of the column (stationary phase) and the carrier gas (mobile phase).  The latter part of the column passes through a heated transfer line and ends at the entrance to ion source where compounds eluting from the column are converted to ions. 49
  • 51. GC-MS INSTRUMENT The insides of the GC-MS, with the column of the gas chromatograph in the oven on the right. 51
  • 52. LIQUID CHROMATOGRAPHY  separation technique in which the mobile phase is a liquid.  can be carried out either in a column or a plane.  Present day it utilizes very small packing particles and a relatively high pressure is referred as high performance liquid chromatography (HPLC).  In the HPLC technique, the sample is forced through a column that is packed with irregularly or spherically shaped particles or a porous monolithic layer (stationary phase) by a liquid (mobile phase) at high pressure. 52
  • 53. SUPERCRITICAL FLUIDCHROMATOGRAPHY  Supercritical fluid chromatography is a separation technique in which the mobile phase is a fluid above and relatively close to its critical temperature and pressure 53
  • 54. AFFINITY CHROMATOGRAPHY Based on selective non-covalent interaction between an analyte and specific molecules Stationary Phase  Highly specific and selective  Molecular recognition groups (affinity ligands) covalently attached to agarose or cellulose beads  Affinity ligand binds selectively and reversibly the analyte  Mechanism of retention: interaction with highly specialized molecular recognition systems which are attached to the stationary phase Mobile Phase  Two distinct roles – Support the strong binding of the analyte molecule to the ligand – Weaken and eliminate the analyte-ligand interaction 54
  • 55. HPLC  one of the most powerful tools in analytical chemistry  has the ability to separate, identify, and quantitate the compounds that are present in any sample that can be dissolved in a liquid  Similar to gas chromatography but uses a liquid mobile phase  The stationary phase is usually an inert solid or a liquid held on the inert solid  Mobile phase travels through the column forcibly with the aid of the high pressure pump  Solutes of the sample separated on column and eluted with mobile phase  The technique is applicable to thermally fragile samples, e.g. amino acids, proteins, nucleic acids, hydrocarbons, antibiotics, steroids, drugs, inorganic and may organic substances. 55
  • 56. 56
  • 57. . COLUMNS  Stainless steel, or heavily walled glass tubing of 10- 30 cm, capable of withstanding 6000 psi, 4 to 10 mm inside dia meter, 0.5 μm particles provide 40- 60000 plates/meter.  1 - 4.6 mm dia, 3.5 mm particles, 3-5 cm length columns offer 1, 00,000 Plates/meter. Such columns have low solvent consumption and speed up the separations 57
  • 58. Filter and separate irreversibly bonding compounds. The particles have similar composition to that of the analytical column but slightly larger particle size to minimize pressure drop. Column thermostats must be capable of controlling ± 0.10 C. This is accomplished by using column heaters or water jackets GUARD COLUMN: 58
  • 59. DETECTION METHODS 􀂄 UV – Ultraviolet light--- most popular – Lamp – Grating/Lens - Wave length 190-350 nm – FlowCell – PhotoDiode - Differential Light Output 􀂄 RI – Refractive Index – Universal analyte detector – Solvent must remain the same throughout separation – VERY temperature sensitive – Sometimes difficult to stabilize baseline 􀂄 FD – Fluorescence-greater sensitivity, not so popular – Excitation wavelength generates fluorescence emission at a higher wavelength – Analytes must have fluorophore group---not very common – Very sensitive and selective 􀂄 MS – Mass Spectrometry – Mass to charge ratio (m/z) – Allows specific compound ID 59
  • 60. NORMAL-PHASE HPLC  Adsorption of analytes on the polar, weakly acidic surface of silica gel  Stationary Phase.: Silica (pH 2-8), Alumina (pH 2 - 12), Bonded Diol and NH2  Mobile Phase: Non-polar solvents (Hexane, CHCl3)  Applications: Non-polar and semi-polar samples; hexane soluble; positional isomers  Polar solutes elute later than non-polar lyophilic ones 60
  • 61. REVERSED-PHASE HPLC  Partition of analytes between mobile phase and stagnant phase inside the pore space + adsorption on the surface of bonded phase  Stationary Phase: Hydrophobic surfaces of moieties bonded on silica (C18, C8, C5, Phenyl, CN)  Mobile phase: Methanol or Acetonitrile and Water  Applications: ~80% of all separations done on RP HPLC  organic molecules are separated based on their degree of hydrophobicity 61
  • 62. UPLC  In 2004, further advances in instrumentation and column technology were made to achieve very significant increase in:  RESOLUTION  SPEED  SENSITIVITY  Increase separation EFFICIENCY  Columns with smaller particles [<1.7um]  Mobile phase delivery is done at >15,000psi 62
  • 63. CONTRASTING HPLC AND UPLC  UPLC gives faster results with better resolution  UPLC uses less of valuable solvents like acetonitrile which lowers cost  The reduction of solvent use is more environmentally friendly 63
  • 64. 64 INSTRUMENTATION  Autosampler – Reduced Cycle Time and niglisible carryove  Flow cell  Van guar column – for protecting column performance  Tubings – Special for UPLC, reduced volume, optimized connection  Mobile Phase Filters – To prevent system damage  column  Detectors
  • 65. 65 SCHEMATIC PRESENTATION OF UPLC Gradient Controller Pump • Column Detector Injector Mobile Phases
  • 66. 66 TUBING AND FITTING  Tubing and fittings are carefully designed to provide leak-free connections, with low dead volume, and minimal band spreading. For best performance, these components must be clean, and designed to work together.
  • 67. 67 MOBILE PHASE FILTERS  Good practice to always filter your solvents  As it prevent system damage  Filters should be changed periodically depending on usage and mobile phase
  • 68. 68 DETECTORS  Evaporative Light Scattering (ELS) Detector  Fluorescence (FLR) Detector  Photodiode Array (PDA) Detector  Tunable UV (TUV) Detector  Single Quadrapole Detector (SQD)
  • 69. 69 ADVANTAGES OF UPLC  Decreases run time and increases sensitivity  Provides the selectivity, sensitivity, and dynamic range of LC analysis  Maintaining resolution performance.  Expands scope of Multiresidue Methods  UPLC’s fast resolving power quickly quantifies related and unrelated compounds  Faster analysis through the use of a novel separation material of very fine particle size  Operation cost is reduced  Less solvent consumption  Reduces process cycle times, so that more product can be produced with existing resources  Delivers real-time analysis in step with manufacturing processes  Assures end-product quality, including final release testing
  • 70. 70 Disadvantages OF UPLC Due to increased pressure requires more maintenance and reduces the life of the columns of this type.