3. Atomic Absorption and Flame Emission Spectroscopy
Atomic spectroscopy is one of the major tools for qualitative and
quantitative analysis of trace (ppm and ppb) metallic elements in
industrial and environmental laboratories.
About 70 elements have been analyzed by this technique.
In these methods, the solution sample is aspirated into a flame that
is hot enough to break the molecules into their atomic states. The
concentration of the analyte in the flame may be measured by either
its absorption or emission of the radiation.
The absorption mode is known as atomic absorption spectroscopy
(AAS) whereas emission mode as flame emission spectroscopy (FES)
4. Atomic Absorption Spectroscopy, AAS
Atomic Emission Spectroscopy, AES
Principle of Atomic Absorption and Emission Spectrophotometer
Ground state E0
Excited state E1
e
Absorption
Ground state E0
Excited state E1
e
Emission
ee
5. The spectra of gaseous, atomic particles consist of well defined
narrow discrete lines (~.01 nm) arising from electronic
transition of outermost electrons.
Atoms undergo only electronic transition, no vibrational or
rotational transitions since there are no bonds, ; unlike that of
molecular species which give broad bands due to all types of
transitions.
sodium vapor exhibits two sharp absorption peaks at 589.0 and
589.6 nm due to excitation of 3s electron to two 3p states
Atomic Spectra
8. Atomization
It is the conversion of molecules to their component atoms in gaseous
state. It is carried out by introduction of the solution in the flame in the
form of very fine droplets. The precision and accuracy of atomic methods
are dependent upon the atomization step.
In Flame Emission
-Atoms in gaseous state in the flame absorb thermal energy from the
flame itself . Some of the atoms get excited & as they return back to the
ground state they emit radiation having energy equal to that absorbed.
-The emission intensity is proportional to the number of excited atoms,
which is proportional to the total number of atoms in the flame
i.e. the sample concentration. It is of two types:
Flame method
Flameless method
9.
10. Elements detectable by atomic absorption are
highlighted in pink in this periodic table
Elements detected
11. Flame-its functions
One of the features common to both Flame Emission and
Atomic Absorption is the use of a flame to atomize the analyte.
The function of the flame differs in the two techniques:
In AAS, the flame serves to convert the analyte into free
atoms; the conditions should be such that no ionization occurs.
In FES the flame serves to atomize the analyte as well as
provide excitation energy.
12. Flame-its functions
Upper diagram shows FES while
the lower one shows AAS. In FES
the flame also provides the
excitation energy, but in AAS it
provides only for the atomization
.
13. These flames vary in temperature, reducibility and transmission characteristics and
are selected according to the element being analyzed, and properties of the sample.
Flame selection
17. Nebulization: sample solution is introduced through an
orifice into a high velocity gas jet, usually the oxidant,
in either parallel or perpendicular manner
Sample stream is converted into a cloud of droplet in the
aerosol modifier or spray chamber, combined with the
oxidizer/fuel and carried to the burner
18. 1)When aerosol containing
MX enter the flame, the
solvent is evaporated,
leaving particles of dry,
solid MX (desolvation)
2) solid MX is converted to
MX vapor (volatilization)
3) portion of MX molecules
are dissociated to give
free M atoms
(Dissociation)
Steps for atomization
21. 21
Disadvantages of flame atomization
Only 5 – 15 % of the nebulized sample reaches
the flame
A minimum sample volume of 0.5 – 1.0 mL is needed
to give a reliable reading
Samples which are viscous require dilution with a
solvent
23. Limitation of Flame Emission Photometry
1-The number of excited atoms in flame is very small. It is the alkaline
and alkaline earth metals that can be practically determined.
2-It needs perfect control of flame temperature.
3- Interference by other elements is not easy to be eliminated.
4-For heavy and transition metals , the number of absorption
and emission lines is enormous and the spectra are complex.
24. Atomic Absorption Spectroscopy
•In atomic absorption spectroscopy the radiation of
the lines of the analyte are produced by the lamp.
•The presence of the atomic state of the analyte in
the flame attenuates or reduces the intensity of
the
radiation.
•The absorbance is proportional to the
concentration of the analyte, similar to optical
spectroscopy.
25. Atomic Absorption Spectrometer
The flowing fuel and air mixture provides the aspiration action
drawing the solution sample into the flame.
26.
27. ATOMIC ABSORPTION SPECTROSCOPY
• Atoms in the vapor state are subjected to external source of
radiation which produces one line or beam of monochromatic
light with single wavelength.
• This wavelength is a resonance one for the atoms and that
will be absorbed by them.
INSTRUMENT FOR ATOMIC ABSORPTION
1- Source of radiation 2- Chopper
3- Atomizer
4-Monochromator 5- Detector
6- Read out meter
29. Source of radiation - hollow cathode lamp
• It is a tube with a front quartz window
• Contains an anode of tungsten and a cylindrical cathode , the material
of which is the same element as that of the sample to be determined.
•Passage of a dc voltage through the lamp produces the specific lines of
those elements.
• The glass tube is filled with neon or argon at a pressure of 1 to 5 torr.
.
32. Hollow cathode lamp
Variety of hollow cathode
lamps are available
commercially
Some cathode are made up
of a mixture of several
metals permitting analysis
of many elements
33. a)
• in order for Beer’s law, to be followed, band-width
of the source must be narrow relative to the width of
an absorption peak (0.002-0.005nm)
•as a result nonlinear calibration curves are inevitable
when continuous radiation source are used
• problem is solved by using the line sources with
bandwidths narrower than absorption peaks
Radiation source
Why cannot we use continuous radiation source?
34. Radiation source
The bandwidth of the
absorption line is much
broader than the
bandwidth of the spectral
lines of the hollow cathode
lamp.
35. Hollow cathode lamp
Spectral output of the
multi-element steel
hollow cathode lamp.
Note the extremely
sharp spectral lines.
36. The chopper
• The function of the chopper is to chop the light leaving the source so
that when the incident beam hits the chopper at the solid surface,
the beam will be blocked and detector will only read the emitted
signal from the flame. As the chopper rotates and the beam emerges
to the detector, the detector signal will be the sum of the
transmitted signal plus that emitted from the flame. The signal
processor will be able to subtract the first signal from the second
one, thus excluding the signal from emission in flames. This can be
represented by the following equations:
• Signal 1 (Blocked Beam) = Pe
• Signal 2 (Transmitted Beam) = P + Pe
• Overall Difference Signal = (P + Pe) - Pe= P (Corrected Signal)
• This correction method for background emission in flames is called
source modulation
37. Interferences with FES and AAS
There are 3 types of interferences associated with both FES
and AAS:
Spectral Interferences arise whenever the line of interest
cannot be easily resolved from another element or from a
molecular band present in the background.
Generally, the bandpass of even very
goodmonochromators 10 times greater than the width of
the lines of the HCL. In some cases, amplitude modulation
of the HCL helps to detect these interferences.
38. Background absorption is caused by the large number of
species present in the flame (metal oxide, OH radical, H2,
fragments of solvent mol.etc)
Incident radiation is absorbed by these species as well as by
the analyte atoms
Correction:
“blank” solution (solution devoid of analyte) can be used to
correct the measurement from sample solution (in
practice it is difficult to prepare exact blank)
Spectral interferences
39. 39
Spectral interferences
----- spectral overlap
Cu 324.754 nm, Eu 324.753 nm
Al 308.215 nm , V 308.211nm,
Al 309.27 nm
Avoid the interference by observing the
aluminum line at 309.27 nm
40. 40
Spectral interferences
----- light scatter
• Metal oxide particles with diameters greater than the
wavelength of light; eg CaOH interferes with Ba analysis.
• When sample contains organic species or when organic
solvents are used to dissolve the sample, incomplete
combustion of the organic matrix leaves carbonaceous
particles that are capable of scattering light
41. 41
Spectral interferences
• The interference can be avoided by
variation in analytical variables, such as
flame temperature and fuel to oxidant
ratio. The solution to many of these
problems is a hotter flame, such as the use
of acetylene/nitrous oxide.
• Standard addition method
42. 42
Chemical interferences
----- Formation of compounds of low
volatility
Increase in flame temperature
Use of releasing agents- La 3+,
Sr+3
)
Ca 2+
, PO4
3-
Use of protective agents (EDTA)
43. Ionization Interference:
• atoms with low ionization potential become
ionized reducing the population of both the ground
state and excited state free atoms
• by adding an excess of easily ionized element (viz.
K, Cs or Sr), ionization in the sample and calibration
solution can be suppressed
• more easily ionized atoms produces a large
concentration of electrons in the vapor and, by mass
action, suppresses the ionization of analyte atoms
44. Standard Addition Method
Mg concentration
after filled up
X X+0.1 X+0.2 X+0.3
100 ml
Solvent
No.1 No.2 No.4No.3
10 ml Unknown sample
10 ml 10 ml 10 ml 10 ml
1.0 ppm X Standard solution (ppm : mg/1000ml)
20 ml
30 ml
10 ml