SlideShare a Scribd company logo
1 of 38
Download to read offline
Siham Abdoun
Msc., PhD
Introduction:


It is a dynamic techniques, in which current

passes through the electrochemical cell,


Coulometry

is

based

on

an

exhaustive

electrolysis of the analyte. By exhaustive we
mean that the analyte is completely oxidized or

reduced at the working electrode or that it reacts
completely with a reagent generated at the

working electrode.

1.

There are two forms of coulometry:
Controlled potential coulometry, in which we
apply a constant potential to the
electrochemical cell, and

2.

Controlled-current coulometry, in which we
pass a constant current through the
electrochemical cell.


During an electrolysis, the total charge, Q, in
coulombs, passing through the electrochemical
cell is proportional to the absolute amount of
analyte by faraday’s law
Q = nFNA………..(1)



where n is the number of electrons per mole of
analyte, F is Faraday’s constant (96487Cmol-1),
and NA is the moles of analyte.


A coulomb is equivalent to an Amber. second;

thus, when passing a constant current, i, the
total charge is Q =ite……………(2)


where te is the electrolysis time. If the current
varies with time, as it does in controlled-

potential coulometry, then the total charge is
.………….(3)


In coulometry, we monitor current as a function
of time and use either equation (2) or equation
(3) to calculate Q. Knowing the total charge, we
then use equation (1) to determine the moles of
analyte.



To obtain an accurate value for NA, all the

current must be used to oxidize or reduce the
analyte. In other words, coulometry requires

100% current efficiency


Current efficiency is the percentage of current

that actually leads to the analyte’s oxidation or
reduction; to ensure 100% current efficiency is
to hold the working electrode at a constant
potential, chosen so that the analyte reacts
completely without simultaneously oxidizing
or reducing an interfering species


As electrolysis progresses the analyte’s
concentration decreases, as does the current.
The resulting current-versus-time profile for
controlled-potential coulometry is shown in the
Figure below



Integrating the area under the curve (equation 3)
from (t = 0 to t = te) gives the total charge


We can use the Nernst equation to estimate the
minimum potential for quantitatively electrolysis



If we define a quantitative electrolysis as one in
which we reduce or oxidized 99.99% of substance

Example;
……..(4)
the required potential can be calculated by;


....................(5)
then the concentration of Cu2+ at te is
…….. (6)


where [Cu2+]0 is the initial concentration of

Cu2+ in the sample. Substituting equation (6)
into equation (5) allows us to calculate the

desired potential.
………(7)
MINIMIZING ELECTROLYSIS TIME


In controlled-potential coulometry, as shown in

the above figure, the current decreases over time.
As a result, the rate of electrolysis becomes

slower and complete electrolysis of the analyte
may require a long time.

Factors affecting the analysis time:


We can approximate the change in current as a

function of time as

……….(8)


where i0 is the current at t = 0 and k is a rate
constant



For complete electrolysis in which we oxidize or
reduce 99.99% of the analyte, the current at the end
of the analysis, te, is



………..(9)

Substituting equation (9) into equation (8) and
solving for te gives the minimum time for complete
electrolysis as


As k is directly proportional to the area of the working
electrode and the rate of stirring, and that is inversely

proportional to the volume of solution so factors affecting
analysis time are;
1.

Volume of electrochemical cell

2.

Electrode surface area

3.

Stirring rate



A quantitative electrolysis typically requires approximately
30–60 min, although shorter or longer times are possible.


INSTRUMENTATION



A three-electrode potentiostat is used to set the

potential in controlled potential coulometry. The
working electrodes is usually one of two types: a
cylindrical Pt electrode manufactured from
platinum-gauze or a Hg pool electrode


Platinum is the working electrode of choice when
we need to apply a positive potential while Hg is the
electrode of choice for an analyte requiring a

negative potential..


The auxiliary electrode, which is often a Pt wire,
is separated by a salt bridge from the analytical
solution. This is necessary to prevent the
electrolysis products generated at the auxiliary
electrode from reacting with the analyte and
interfering in the analysis.



A saturated calomel or Ag/AgCl electrode serves
as the reference electrode.


Methods for determining the total charge;

1.

One method is to monitor the current as a function

of time and determine the area under the curve, as
shown in the above figure.
2.

Modern instruments use electronic integration to
monitor charge as a function of time. The total

charge at the end of the electrolysis is read directly
from a digital readout
3. Gravimetric technique


If the product of controlled-potential coulometry forms

a deposit on the working electrode, then we can use the
change in the electrode’s mass as the analytical signal.

For example, if we apply a potential that reduces Cu2+
to Cu at a Pt working electrode, the difference in the
electrode’s mass before and after electrolysis is a direct
measurement of the amount of copper in the sample.


we call this electrogravimetry


Is a technique use a constant current in place of
a constant potential, which produces the

current-versus-time profile shown in the figure
below


Controlled-current coulometry has two advantages
over controlled-potential coulometry.



First, the analysis time is shorter because the
current does not decrease over time. A typical

analysis time for controlled-current coulometry is
less than 10 min, compared to approximately 30–

60 min for controlled-potential coulometry.


Second the total charge is simply the product of
current and time

1.

Problems with this technique;
We need a method for determining when the
analyte's electrolysis is complete.

2.

The current efficiency may be <100%, due to
secondary reactions
Maintaining current efficiency


The change in the working electrode’s potential
may result in a current efficiency <100%, as
shown in the following reaction
………(10)


At the beginning, the potential of the working electrode
remains nearly constant at a level near its initial value. As

the concentration of Fe2+ decreases, the working
electrode’s potential shifts toward more positive values

until the oxidation of H2O begins.
……..(11)


Because a portion of the total current comes from the
oxidation of H2O, the current efficiency for the analysis
is less than 100% and we cannot use eq.( 1) to determine
the amount of Fe2+ in the sample.


Although we cannot prevent the potential from drifting
until another species undergoes oxidation, we can

maintain a 100% current efficiency if the product of that
oxidation reacts both rapidly and quantitatively with the

remaining Fe2+.


To accomplish this we can add an excess of Ce3+ to the
analytical solution, when the potential of the working
electrode shifts to a more positive potential Ce3+
eventually begins to oxidize as shown;



…….(12)


The Ce4+ that forms at the working electrode rapidly
mixes with the solution where it reacts with any

available Fe2+ maintaining a current efficiency of 100%.
…… ……(13)


Combining reaction (12) and reaction (13) shows that
the net reaction is the oxidation of Fe2+ to Fe3+
…………..(14)



A species, such as Ce3+, which is used to maintain
100% current efficiency, is called a mediator.


Endpoint determination;



As shown from the above example, when the oxidation
of Fe2+ is complete current continues to flow from the
oxidation of Ce3+, and, eventually, the oxidation of
H2O. So there is no signal that indicate there is no

more Fe2+ in solution.


In this example it is convenient to treat a controlled-

current coulometric analysis as a reaction between the
analyte, Fe2+, and the mediator, Ce3+, as shown by

reaction 13.


This reaction is identical to a redox titration;
thus, we can use the end points for a redox
titration to signal the end of a controlled-current
coulometric analysis. For example, ferroin
provides a useful visual endpoint for the Ce3+
mediated coulometric analysis for Fe2+,
changing color from red to blue when the
electrolysis of Fe2+ is complete.


INSTRUMENTATION



Controlled-current coulometry normally is carried out

using a two-electrode galvanostat, consisting of a
working electrode and a counter electrode.


The working electrode—often a simple Pt electrode—
is also called the generator electrode since it is where

the mediator reacts to generate the species that reacts
with the analyte.


There are two other crucial needs for controlledcurrent coulometry instrument:

1.

An accurate clock for measuring the electrolysis
time, te, and

2.


A switch for starting and stopping the electrolysis
The switch must control both the current and the

clock, so that we can make an accurate determination
of the electrolysis time.


Quantitative Applications



Coulometry is used for the quantitative analysis of both

inorganic and organic analytes.
1. Controlled-potential coulometry


The majority of controlled-potential coulometric
analyses involve the determination of inorganic cations
and anions, including trace metals and halides ions.



It is also used for the quantitative analysis of organic
compounds, example is the reduction of trichloroacetate
to dichloroacetate, and of dichloroacetate to
monochloroacetate.
2. Controlled-current coulometry (coulometric titrations)


The use of a mediator makes a coulometric

titration a more adaptable analytical technique
than controlled-potential coulometry


For an analyte that is not easily oxidized or
reduced, we can complete a coulometric titration
by coupling a mediator’s oxidation or reduction to
an acid–base, precipitation, or complexation
reaction involving the analyte.


For example, if we use H2O as a mediator, we can
generate H3O+ at the anode and generate OH– at the

cathode.


Coulometric acid–base titrations have been used for the
analysis of strong and weak acids and bases, in both

aqueous and non-aqueous matrices.


In comparison to a conventional titration, a coulometric

titration has many important advantages.
1.

The first advantage is that electrochemically generating
a titrant allows us to use an unstable reagent.
Although we cannot easily prepare and store a solution
of a highly reactive reagent, such as Ag2+ or Mn3+, we

can generate them electrochemically and use them in a
coulometric titration.


Second, because it is relatively easy to measure a small
quantity of charge, we can use a coulometric titration

to determine an analyte whose concentration is too
small for a conventional titration.


Third, standardizations using primary standards are not
necessary,


The coulometric method of analysis is accurate,
precise , sensitive, selective and not expensive but
Controlled-potential coulometry is a relatively time
consuming analysis, with a typical analysis requiring

30–60 min. Coulometric titrations, on the other hand,
require only a few minutes,


Another important example of redox titrimetry is the

determination of water in substance such as
pharmaceutical preparation and raw material . The
titrant for this analysis is known as the Karl Fischer
reagent and consists of a mixture of iodine, sulfur
dioxide, pyridine, and methanol.


Because the concentration of pyridine is sufficiently
large, I2 and SO2 react with pyridine (py) to form the

complexes py•I2 and py•SO2. When added to a
sample containing water, I2 is reduced to I– and SO2

is oxidized to SO3.


Methanol is included to prevent the further reaction of
py•SO3 with water. The titration’s end point is
signaled when the solution changes from the product’s
yellow color to the brown color of the Karl Fischer

reagent.
Thanks

More Related Content

What's hot

What's hot (20)

Electroanalytical Methods of analysis
Electroanalytical Methods of analysisElectroanalytical Methods of analysis
Electroanalytical Methods of analysis
 
Cyclic voltammetry
Cyclic voltammetryCyclic voltammetry
Cyclic voltammetry
 
Amperometry
AmperometryAmperometry
Amperometry
 
Stripping voltammetry
Stripping voltammetryStripping voltammetry
Stripping voltammetry
 
AUGER & ESCA Spectroscopy( Mass Spectroscopy )
AUGER & ESCA Spectroscopy( Mass Spectroscopy )AUGER & ESCA Spectroscopy( Mass Spectroscopy )
AUGER & ESCA Spectroscopy( Mass Spectroscopy )
 
Dc,pulse,ac and square wave polarographic techniques new
Dc,pulse,ac and square wave polarographic techniques newDc,pulse,ac and square wave polarographic techniques new
Dc,pulse,ac and square wave polarographic techniques new
 
Coulometry and Electrogravimetry
Coulometry and ElectrogravimetryCoulometry and Electrogravimetry
Coulometry and Electrogravimetry
 
AMPEROMETRY
AMPEROMETRYAMPEROMETRY
AMPEROMETRY
 
Basics of Voltammetry and Potentiometry
Basics of Voltammetry and Potentiometry Basics of Voltammetry and Potentiometry
Basics of Voltammetry and Potentiometry
 
Cyclic voltammetry
Cyclic voltammetryCyclic voltammetry
Cyclic voltammetry
 
Electrogravimetry
ElectrogravimetryElectrogravimetry
Electrogravimetry
 
Voltammetry and Polarography
Voltammetry and PolarographyVoltammetry and Polarography
Voltammetry and Polarography
 
High Frequency Titrations
High Frequency TitrationsHigh Frequency Titrations
High Frequency Titrations
 
Voltammetry vipul
Voltammetry vipulVoltammetry vipul
Voltammetry vipul
 
Cyclic voltammetry
Cyclic voltammetryCyclic voltammetry
Cyclic voltammetry
 
Actinometry-sobia.pptx
Actinometry-sobia.pptxActinometry-sobia.pptx
Actinometry-sobia.pptx
 
Electrogravimetry
ElectrogravimetryElectrogravimetry
Electrogravimetry
 
Cyclic Voltammetry Application
Cyclic Voltammetry  Application Cyclic Voltammetry  Application
Cyclic Voltammetry Application
 
Zero field splitting
Zero field splittingZero field splitting
Zero field splitting
 
Polarography
PolarographyPolarography
Polarography
 

Viewers also liked

Electrochemical method of analysis
Electrochemical method of analysisElectrochemical method of analysis
Electrochemical method of analysis
Siham Abdallaha
 
Advances in Ion Selective Electrodes(ISE)
Advances in Ion Selective Electrodes(ISE) Advances in Ion Selective Electrodes(ISE)
Advances in Ion Selective Electrodes(ISE)
Nur Fatihah
 

Viewers also liked (20)

Electrochemical method of analysis
Electrochemical method of analysisElectrochemical method of analysis
Electrochemical method of analysis
 
Electrogravimetry and coulometry
Electrogravimetry and coulometryElectrogravimetry and coulometry
Electrogravimetry and coulometry
 
Polarography
PolarographyPolarography
Polarography
 
Electrogravimetry and coulometry
Electrogravimetry and coulometryElectrogravimetry and coulometry
Electrogravimetry and coulometry
 
Potentiometry
PotentiometryPotentiometry
Potentiometry
 
Voltammetry
VoltammetryVoltammetry
Voltammetry
 
Cyclic Voltammetry
Cyclic VoltammetryCyclic Voltammetry
Cyclic Voltammetry
 
Karl fischer
Karl fischerKarl fischer
Karl fischer
 
1 Potentiometry
1  Potentiometry1  Potentiometry
1 Potentiometry
 
Unit 6 Water Content Determination and Moisture analysis
Unit 6   Water Content Determination and Moisture analysisUnit 6   Water Content Determination and Moisture analysis
Unit 6 Water Content Determination and Moisture analysis
 
potentiometry
potentiometrypotentiometry
potentiometry
 
Potentiometry
PotentiometryPotentiometry
Potentiometry
 
9 polarography jntu pharmacy
9 polarography jntu pharmacy9 polarography jntu pharmacy
9 polarography jntu pharmacy
 
Analytical chemistry
Analytical chemistryAnalytical chemistry
Analytical chemistry
 
Conducto ppt
Conducto pptConducto ppt
Conducto ppt
 
potentiometry
potentiometrypotentiometry
potentiometry
 
Types of titrations
Types of titrationsTypes of titrations
Types of titrations
 
Mass spectrometry
Mass spectrometryMass spectrometry
Mass spectrometry
 
MASS SPECTROSCOPY
MASS SPECTROSCOPYMASS SPECTROSCOPY
MASS SPECTROSCOPY
 
Advances in Ion Selective Electrodes(ISE)
Advances in Ion Selective Electrodes(ISE) Advances in Ion Selective Electrodes(ISE)
Advances in Ion Selective Electrodes(ISE)
 

Similar to Coulometric method of analysis

Voltametry- Pharmaceutical Analysis
Voltametry- Pharmaceutical AnalysisVoltametry- Pharmaceutical Analysis
Voltametry- Pharmaceutical Analysis
Sanchit Dhankhar
 
Electrochemistry, electrophoresis, ise
Electrochemistry, electrophoresis, iseElectrochemistry, electrophoresis, ise
Electrochemistry, electrophoresis, ise
Angelica Nhoj Gemora
 
Copper Design BEUKES
Copper Design BEUKESCopper Design BEUKES
Copper Design BEUKES
Nick Beukes
 

Similar to Coulometric method of analysis (20)

potentiometry
potentiometrypotentiometry
potentiometry
 
CBSE Class 12 Chemistry Chapter 3 (Electrochemistry) | Homi Institute
CBSE Class 12 Chemistry Chapter 3 (Electrochemistry) | Homi InstituteCBSE Class 12 Chemistry Chapter 3 (Electrochemistry) | Homi Institute
CBSE Class 12 Chemistry Chapter 3 (Electrochemistry) | Homi Institute
 
Voltametry- Pharmaceutical Analysis
Voltametry- Pharmaceutical AnalysisVoltametry- Pharmaceutical Analysis
Voltametry- Pharmaceutical Analysis
 
Potentiometry1 for mpharm ist sem notes
Potentiometry1 for mpharm ist sem notes Potentiometry1 for mpharm ist sem notes
Potentiometry1 for mpharm ist sem notes
 
Electrochemistry, electrophoresis, ise
Electrochemistry, electrophoresis, iseElectrochemistry, electrophoresis, ise
Electrochemistry, electrophoresis, ise
 
Electrochemistry
ElectrochemistryElectrochemistry
Electrochemistry
 
Basic concepts in electrochemistry
Basic concepts in electrochemistryBasic concepts in electrochemistry
Basic concepts in electrochemistry
 
Modelling and simulation approach.pdf
Modelling and simulation approach.pdfModelling and simulation approach.pdf
Modelling and simulation approach.pdf
 
Electrochemical methods: Environmental Analysis
Electrochemical methods: Environmental Analysis Electrochemical methods: Environmental Analysis
Electrochemical methods: Environmental Analysis
 
Potentiometry
PotentiometryPotentiometry
Potentiometry
 
electrogravimetry-211216084524.pdf
electrogravimetry-211216084524.pdfelectrogravimetry-211216084524.pdf
electrogravimetry-211216084524.pdf
 
421-821-chapter-25.ppt
421-821-chapter-25.ppt421-821-chapter-25.ppt
421-821-chapter-25.ppt
 
Electrodes and potentiometry
Electrodes and potentiometryElectrodes and potentiometry
Electrodes and potentiometry
 
Knocking Door of Cyclic Voltammetry - cv of CV by Monalin Mishra
Knocking Door of Cyclic Voltammetry - cv of CV by Monalin MishraKnocking Door of Cyclic Voltammetry - cv of CV by Monalin Mishra
Knocking Door of Cyclic Voltammetry - cv of CV by Monalin Mishra
 
Electrochemistry
ElectrochemistryElectrochemistry
Electrochemistry
 
Fuel cell thermodynamics (2)
Fuel cell thermodynamics (2)Fuel cell thermodynamics (2)
Fuel cell thermodynamics (2)
 
Cyclic Voltammetry: Principle, Instrumentation & Applications
Cyclic Voltammetry: Principle, Instrumentation & ApplicationsCyclic Voltammetry: Principle, Instrumentation & Applications
Cyclic Voltammetry: Principle, Instrumentation & Applications
 
Copper Design BEUKES
Copper Design BEUKESCopper Design BEUKES
Copper Design BEUKES
 
IEEE 2004
IEEE 2004IEEE 2004
IEEE 2004
 
Polarography
PolarographyPolarography
Polarography
 

Coulometric method of analysis

  • 2. Introduction:  It is a dynamic techniques, in which current passes through the electrochemical cell,  Coulometry is based on an exhaustive electrolysis of the analyte. By exhaustive we mean that the analyte is completely oxidized or reduced at the working electrode or that it reacts completely with a reagent generated at the working electrode.
  • 3.  1. There are two forms of coulometry: Controlled potential coulometry, in which we apply a constant potential to the electrochemical cell, and 2. Controlled-current coulometry, in which we pass a constant current through the electrochemical cell.
  • 4.  During an electrolysis, the total charge, Q, in coulombs, passing through the electrochemical cell is proportional to the absolute amount of analyte by faraday’s law Q = nFNA………..(1)  where n is the number of electrons per mole of analyte, F is Faraday’s constant (96487Cmol-1), and NA is the moles of analyte.
  • 5.  A coulomb is equivalent to an Amber. second; thus, when passing a constant current, i, the total charge is Q =ite……………(2)  where te is the electrolysis time. If the current varies with time, as it does in controlled- potential coulometry, then the total charge is .………….(3)
  • 6.  In coulometry, we monitor current as a function of time and use either equation (2) or equation (3) to calculate Q. Knowing the total charge, we then use equation (1) to determine the moles of analyte.  To obtain an accurate value for NA, all the current must be used to oxidize or reduce the analyte. In other words, coulometry requires 100% current efficiency
  • 7.  Current efficiency is the percentage of current that actually leads to the analyte’s oxidation or reduction; to ensure 100% current efficiency is to hold the working electrode at a constant potential, chosen so that the analyte reacts completely without simultaneously oxidizing or reducing an interfering species
  • 8.  As electrolysis progresses the analyte’s concentration decreases, as does the current. The resulting current-versus-time profile for controlled-potential coulometry is shown in the Figure below  Integrating the area under the curve (equation 3) from (t = 0 to t = te) gives the total charge
  • 9.
  • 10.  We can use the Nernst equation to estimate the minimum potential for quantitatively electrolysis  If we define a quantitative electrolysis as one in which we reduce or oxidized 99.99% of substance Example; ……..(4) the required potential can be calculated by;  ....................(5) then the concentration of Cu2+ at te is …….. (6)
  • 11.  where [Cu2+]0 is the initial concentration of Cu2+ in the sample. Substituting equation (6) into equation (5) allows us to calculate the desired potential. ………(7)
  • 12. MINIMIZING ELECTROLYSIS TIME  In controlled-potential coulometry, as shown in the above figure, the current decreases over time. As a result, the rate of electrolysis becomes slower and complete electrolysis of the analyte may require a long time. Factors affecting the analysis time:  We can approximate the change in current as a function of time as ……….(8)
  • 13.  where i0 is the current at t = 0 and k is a rate constant  For complete electrolysis in which we oxidize or reduce 99.99% of the analyte, the current at the end of the analysis, te, is  ………..(9) Substituting equation (9) into equation (8) and solving for te gives the minimum time for complete electrolysis as
  • 14.  As k is directly proportional to the area of the working electrode and the rate of stirring, and that is inversely proportional to the volume of solution so factors affecting analysis time are; 1. Volume of electrochemical cell 2. Electrode surface area 3. Stirring rate  A quantitative electrolysis typically requires approximately 30–60 min, although shorter or longer times are possible.
  • 15.  INSTRUMENTATION  A three-electrode potentiostat is used to set the potential in controlled potential coulometry. The working electrodes is usually one of two types: a cylindrical Pt electrode manufactured from platinum-gauze or a Hg pool electrode  Platinum is the working electrode of choice when we need to apply a positive potential while Hg is the electrode of choice for an analyte requiring a negative potential..
  • 16.  The auxiliary electrode, which is often a Pt wire, is separated by a salt bridge from the analytical solution. This is necessary to prevent the electrolysis products generated at the auxiliary electrode from reacting with the analyte and interfering in the analysis.  A saturated calomel or Ag/AgCl electrode serves as the reference electrode.
  • 17.  Methods for determining the total charge; 1. One method is to monitor the current as a function of time and determine the area under the curve, as shown in the above figure. 2. Modern instruments use electronic integration to monitor charge as a function of time. The total charge at the end of the electrolysis is read directly from a digital readout
  • 18. 3. Gravimetric technique  If the product of controlled-potential coulometry forms a deposit on the working electrode, then we can use the change in the electrode’s mass as the analytical signal. For example, if we apply a potential that reduces Cu2+ to Cu at a Pt working electrode, the difference in the electrode’s mass before and after electrolysis is a direct measurement of the amount of copper in the sample.  we call this electrogravimetry
  • 19.  Is a technique use a constant current in place of a constant potential, which produces the current-versus-time profile shown in the figure below
  • 20.  Controlled-current coulometry has two advantages over controlled-potential coulometry.  First, the analysis time is shorter because the current does not decrease over time. A typical analysis time for controlled-current coulometry is less than 10 min, compared to approximately 30– 60 min for controlled-potential coulometry.  Second the total charge is simply the product of current and time
  • 21.  1. Problems with this technique; We need a method for determining when the analyte's electrolysis is complete. 2. The current efficiency may be <100%, due to secondary reactions
  • 22. Maintaining current efficiency  The change in the working electrode’s potential may result in a current efficiency <100%, as shown in the following reaction ………(10)
  • 23.  At the beginning, the potential of the working electrode remains nearly constant at a level near its initial value. As the concentration of Fe2+ decreases, the working electrode’s potential shifts toward more positive values until the oxidation of H2O begins. ……..(11)  Because a portion of the total current comes from the oxidation of H2O, the current efficiency for the analysis is less than 100% and we cannot use eq.( 1) to determine the amount of Fe2+ in the sample.
  • 24.  Although we cannot prevent the potential from drifting until another species undergoes oxidation, we can maintain a 100% current efficiency if the product of that oxidation reacts both rapidly and quantitatively with the remaining Fe2+.  To accomplish this we can add an excess of Ce3+ to the analytical solution, when the potential of the working electrode shifts to a more positive potential Ce3+ eventually begins to oxidize as shown;  …….(12)
  • 25.  The Ce4+ that forms at the working electrode rapidly mixes with the solution where it reacts with any available Fe2+ maintaining a current efficiency of 100%. …… ……(13)  Combining reaction (12) and reaction (13) shows that the net reaction is the oxidation of Fe2+ to Fe3+ …………..(14)  A species, such as Ce3+, which is used to maintain 100% current efficiency, is called a mediator.
  • 26.  Endpoint determination;  As shown from the above example, when the oxidation of Fe2+ is complete current continues to flow from the oxidation of Ce3+, and, eventually, the oxidation of H2O. So there is no signal that indicate there is no more Fe2+ in solution.  In this example it is convenient to treat a controlled- current coulometric analysis as a reaction between the analyte, Fe2+, and the mediator, Ce3+, as shown by reaction 13.
  • 27.  This reaction is identical to a redox titration; thus, we can use the end points for a redox titration to signal the end of a controlled-current coulometric analysis. For example, ferroin provides a useful visual endpoint for the Ce3+ mediated coulometric analysis for Fe2+, changing color from red to blue when the electrolysis of Fe2+ is complete.
  • 28.  INSTRUMENTATION  Controlled-current coulometry normally is carried out using a two-electrode galvanostat, consisting of a working electrode and a counter electrode.  The working electrode—often a simple Pt electrode— is also called the generator electrode since it is where the mediator reacts to generate the species that reacts with the analyte.
  • 29.  There are two other crucial needs for controlledcurrent coulometry instrument: 1. An accurate clock for measuring the electrolysis time, te, and 2.  A switch for starting and stopping the electrolysis The switch must control both the current and the clock, so that we can make an accurate determination of the electrolysis time.
  • 30.  Quantitative Applications  Coulometry is used for the quantitative analysis of both inorganic and organic analytes. 1. Controlled-potential coulometry  The majority of controlled-potential coulometric analyses involve the determination of inorganic cations and anions, including trace metals and halides ions.  It is also used for the quantitative analysis of organic compounds, example is the reduction of trichloroacetate to dichloroacetate, and of dichloroacetate to monochloroacetate.
  • 31. 2. Controlled-current coulometry (coulometric titrations)  The use of a mediator makes a coulometric titration a more adaptable analytical technique than controlled-potential coulometry  For an analyte that is not easily oxidized or reduced, we can complete a coulometric titration by coupling a mediator’s oxidation or reduction to an acid–base, precipitation, or complexation reaction involving the analyte.
  • 32.  For example, if we use H2O as a mediator, we can generate H3O+ at the anode and generate OH– at the cathode.  Coulometric acid–base titrations have been used for the analysis of strong and weak acids and bases, in both aqueous and non-aqueous matrices.  In comparison to a conventional titration, a coulometric titration has many important advantages. 1. The first advantage is that electrochemically generating a titrant allows us to use an unstable reagent.
  • 33. Although we cannot easily prepare and store a solution of a highly reactive reagent, such as Ag2+ or Mn3+, we can generate them electrochemically and use them in a coulometric titration.  Second, because it is relatively easy to measure a small quantity of charge, we can use a coulometric titration to determine an analyte whose concentration is too small for a conventional titration.  Third, standardizations using primary standards are not necessary,
  • 34.  The coulometric method of analysis is accurate, precise , sensitive, selective and not expensive but Controlled-potential coulometry is a relatively time consuming analysis, with a typical analysis requiring 30–60 min. Coulometric titrations, on the other hand, require only a few minutes,
  • 35.  Another important example of redox titrimetry is the determination of water in substance such as pharmaceutical preparation and raw material . The titrant for this analysis is known as the Karl Fischer reagent and consists of a mixture of iodine, sulfur dioxide, pyridine, and methanol.
  • 36.  Because the concentration of pyridine is sufficiently large, I2 and SO2 react with pyridine (py) to form the complexes py•I2 and py•SO2. When added to a sample containing water, I2 is reduced to I– and SO2 is oxidized to SO3.
  • 37.  Methanol is included to prevent the further reaction of py•SO3 with water. The titration’s end point is signaled when the solution changes from the product’s yellow color to the brown color of the Karl Fischer reagent.