AUTOMATION IN HEMATOLOGY
By
Dr.Varughese George
PART 2
I N D E X
PART 1
I. Necessity for Automation.
II. Advantages & Disadvantages of Automation.
III. Types of Automated Hematology Analyzers.
IV. Principles involved in Automation.
V. Pentra ES 60 Haematology Analyzer.
VI. Pentra DF Nexus Haematology Analyzer.
PART 2
I. Histograms.
II. Flags
III. Quality Control
Histograms
 Graphical representation
of numerical data of
different cell populations
in a cell counter.
 x-axis => cell size
 y-axis => no: of cells
 Gives info on
 Average size
 Distribution of size
Discrimination thresholds
 Moving or fixed discriminator separates the
distribution curve for the volume
 WBC Discriminator
 The automated counter sets a LD flunctuating between 30-60 fl & an UD fixed at
300 fl.
 WBC is calculated from particle counts > than this LD
 RBC Discriminator
Has two flexible discriminators LD (25-75 fl) & UD (200-250 fl).
RBC is calculated from particle counts between this LD & UD.
 Platelet Discriminator
Three discriminators – LD (2-6 fl), UD (12-30 fl) & a fixed discriminator (8-12 fl)
Estimation of RBC Count & MCV – RBC Histogram
Normal RBC distribution curve is Gaussian bell shaped curve.
Analyzer counts RBCs as those which ranges from 36-360 fl.
MCV is perpendicular line from peak of the curve to the base.
Peak of curve should fall within the normal MCV range of 80-100 fl.
There are 2 flexible discriminators – LD (25-75 fl) & UD (200-250 fl)
Abnormalities of RBC Histogram
Left shift of the curve in microcytosis.
Right shift of the curve in macrocytosis.
Bimodal peak of the curve in dimorphic population of cells.
Estimation of Hematocrit, MCH &
MCHC
Estimation of RDW
RDW is expressed as a coefficient of variation of RBC size distribution.
RDW- CV is a better indicator of anisocytosis than RDW-SD.
RBC distribution curve ll get wider as RBC vary more in size.
Normal Range – RDW-CV – 11.0-15.0% RDW-SD – 40.0 - 55.0 fL
RDW
Estimation of WBC Count
 Hematology analyzers can generate a
 3-part differential count – lymphocytes, monocytes & granulocytes
based on the principle of electrical impedence
Or
 5-part differential count - lymphocytes, monocytes, neutrophils,
eosinophils & basophils
based on different principles –
 light scatter
 electrical impedence
 electrical conductivity
 peroxidase staining
Estimation of WBC Count – WBC Histogram
 Cells > 35 fl are counted as WBCs in the WBC/Hb chamber.
Cells with volume 35-90 fl  Lymphocytes.
Cells with volume 90-160 fl  Mononuclear cells.
Cells with volume 160-450 fl  Neutrophils.
Abnormalities of WBC Histogram
 Peak to the left of lymphocyte peak – nucleated cells.
 Peak between lymphocytes & monocytes – blast cells, eosinophilia, basophilia,
plasma cells & atypical lymphocytes.
 Peak between monocytes & neutrophils – left shift
2-6 fl 12-30 fl
fixed at
12 fl
Platelet Histogram
PL PU
PLT RBC
100%
20%
 PLT size: 8-12 fL
 PLT detection: between 2 and 30 fL
 Fixed discriminator at 12 fL
Estimation of Platelet Count
Platelets are counted by the electrical impedence method in the RBC aperture.
Particles between 2 fl and less than 20 fl are classified as platelets by the analyzer.
MPV is a measurement of the average size of platelets found in blood Normal MPV is 7-10 fl.
↑ MPV ( > 10 fl) d/t destruction of platelets in circulation.
↓ MPV ( < 7 fl) d/t impaired production of platelets in circulation.
 Plateletocrit (PCT) is volume of circulating platelets in a unit volume of blood. Normal PCT is 0.19-0.36%.
 PCT ↑ in thrombocytosis and ↓ in thrombocytopenia.
PDW is a measure of variation in the size of platelets. Standard PDW ranges from 9 to 14 fL
↑ PDW is observed in megaloblastic anemia, CML & after chemotherapy.
I N D E X
PART 1
I. Necessity for Automation.
II. Advantages & Disadvantages of Automation.
III. Types of Automated Hematology Analyzers.
IV. Principles involved in Automation.
V. Pentra ES 60 Haematology Analyzer.
VI. Pentra DF Nexus Haematology Analyzer.
PART 2
I. Histograms.
II. Flags.
III. Quality Control.
Flagging
 Flags are signals that occur when
an abnormal result is detected
by the automated blood
analyzer.
 Flags are signaled by certain
‘asteriks’ on the report.
 They reduce the False +ve &
False –ve results by mandating
the results of blood smear
examination.
RBC Flags
 Seen when LD > preset
height by 10 %
 Shown by
 RBC Count
 HCT, MCV,MCH,MCHC
 Occurs when there is
 platelet aggregation
 RBC fragments
RL Flag
RBC Flags
 Seen when UD > preset
height by 10%.
 Shown by
 RBC Count
 HCT, MCV,MCH,MCHC
 Occurs when there are
 Cold agglutinins
RU Flag
RBC Flags
 Shown by RDW-SD
 Seen in
 Post Blood tranfusion
 Treated Fe deficiency anemia.
MP- Flag
WBC Flags
 Generated when curve deviates
from the baseline on the LD.
 Various causes for this are –
 Platelet aggregates (clotted
sample, EDTA incompactibility)
 Lyse-resistant RBCs.
 Erythroblasts.
 Cryoagglutinates.
 Giant platelets.
WL Flag
WBC Flags
 Generated when there is
deviation of the curve on the UD
or if it does not end at the
baseline.
 Various causes for this are :-
 Hyperleucocytosis.
WU Flag
WBC Flags - T1 & T2 Flags
Peak between T1-T2 : The middle cell count - Eosinophils, Monocytes, Blasts, promyelocytes,
myelocytes and metamyelocytes.
Peak between LD-T1 : Lymphocytes
Peak between T2-UD : Neutrophils.
T1 & T2 flags appears when discrimination between lymphocytes, middle cells & neutrophils is
not possible which happens in presence of abnormal/higher leucocyte counts as in Chronic
Myeloid Leukemia.
WBC Flags - F1, F2, F3 Flags
 Sometimes, the cell Fractions
may be mixed.
 F1 & F2 or F2 or F3 merge into
each other over large areas.
 F1 (small cell inaccurate) flag :
Acute Lymphocytic Leukemia
 F2 (middle cell inaccurate) flag :
eosinophil, Acute Myeloid
Leukemia, monocytosis, etc
 F3 (large cell inaccurate) flag:
Platelet Flags
 This occurs when the LD
> the preset height by
10%
 Shown by
 Platelet count
 MPV
 P-LCR
 Occurs due to noise.
PL Flag
Platelet Flags
 This occurs when the
UD> the preset height by
> 40%.
 Occurs in
 Hemolytic anemias with
fragmented cells.
 Large platelets.
PU Flag
Platelet Flags
I N D E X
PART 1
I. Necessity for Automation.
II. Advantages & Disadvantages of Automation.
III. Types of Automated Hematology Analyzers.
IV. Principles involved in Automation.
V. Pentra ES 60 Haematology Analyzer.
VI. Pentra DF Nexus Haematology Analyzer.
PART 2
I. Histograms.
II. Flags
III. Quality Control
Quality Control / Quality Assurance/Quality Assessment
 Quality Control => measures that must be included
during each assay run that the test is working properly.
 Quality Assurance => overall program that ensures
that the final results reported by the lab are correct
 Quality Assessment => means to determine the
quality of results generated by the lab
QUALITY CONTROL
 Measure of precision – how well the measurement system reproduces the
same results over time and under varying operating conditions.
 Designed to detect, reduce & correct deficiencies in a laboratory’s internal
analytical processes prior to release of patient’s reports.
 Should approximate the same matrix as patient’s specimens, taking into
account properties such as viscosity, turbidity, composition and color.
 Should be simple to use, with minimal vial to vial comparability.
 Should be stable for long periods of time and available in large enough
quantities for a single batch to last at least 1 year.
Types of Quality Control
Internal
 Continuous evaluation of reliability
of the daily works of the laboratory
with validation of tests.
 Primary tool required is called a
control – a specimen with a
predetermined range of result values,
processed in the same manner as
patient sample.
 If the result of a test on a control
sample is different from its known
value, this indicates a problem in the
equipment or the methods being
used.
External
 Evaluation by an outside agency of
the comparability a laboratory's
testing to a source outside the
laboratory.
 This comparison can be made to
the performance of a peer group of
laboratories or to the performance
of a reference laboratory.
 The analysis of performance is
retrospective.
Accuracy and Precision
 Accuracy is a measure of rightness.
 It refers to closeness to the true value.
 Precision is a measure of exactness.
 It refers to reproducibility of the test.
Control
 A solution that contains the same constituents as those being
analyzed in the patient sample.
 Commercially produced pooled RBCs/sera or stabilized
anticoagulated whole blood.
 Should have same test properties as that of a blood sample.
 At least 1 control specimen should be used for every batch.
 If large specimens, use 1 control for every 20 specimens.
 For most tests, a “normal” control and an “abnormal” control are
analyzed with each patient test or batch of tests.
 The results are compared with the manufacturer’s range of values
and plotted on a Levey-Jennings Chart.
Levey Jennings Graphs
 A graphical method for displaying control
results and evaluating whether a procedure is
in-control or out-of-control
 Control values are plotted versus time
 Lines are drawn from point to point to accent
any trends, shifts, or random excursions
 Consecutive values of control are recorded and
the standard deviation is calculated.
 The mean and ± 2 SD are plotted on a Levey-
Jennings chart.
 As long as the control value are between the ± 2
SD lines on the L-J chart, the control values are
“in control”
 If they are outside the ± 2 SD lines, they are 'out
of control'
Westgard Control Rules
 Proposed by Dr. James Westgard
on lab quality control.
 Set the basis for evaluating
analytical run for medical
laboratories.
 There are 6 basic rules.
12s Rule . Warning rule to trigger careful inspection of the control data
A single control measurement exceeding 2 standard deviations of control limits either
above or below the mean.
No cause for rejecting a run
13s rule
A single control measurement exceeds the mean plus 3s or the mean minus 3s
control limit.
The run must be rejected.
22s rule
2 consecutive control measurements exceed exceeds the +2SD or -2SD
control limit .
The run must be rejected.
R4s rule
1 control measurement exceed the +2SD and the other exceeds the -2SD
control limit .
The run must be rejected.
41s rule
4 consecutive control measurements exceed exceeds the +1SD or -1SD
control limit .
The run must be rejected.
10x rule
10 consecutive quality control results for one level of control are on one
side of the mean
The run must be rejected.
‘Out of Control’
 Stop testing
 Identify and correct the issue.
 Repeat testing on patient samples and controls.
 Don’t report patient results until the issue is sorted out
and the controls indicate proper performance
Errors
 Systematic error is evidenced by a change in the mean of
control values.
 The change in mean maybe
• Gradual - demonstrated as a trend in control values.
• Abrupt – demonstrated as a shift in control values.
 Random error is any deviation away from the expected
result.
Errors
Trend Shift
 Gradual loss of reliability in test
system.
Causes
 Deterioration of instrument light
source.
 Gradual accumulation of debris in
sample/reagent tubing and
electrode surfaces.
 Aging of reagents.
 Gradual deterioration of –
 Control materials.
 Incubation chamber
temperature.
 Light filter integrity.
 Calibration.
 Sudden/dramatic +ve or –ve change in
test system
Causes
 Sudden failure or change in the light
source.
 Change in reagent formulation.
 Major instrument maintenance.
 Sudden change in incubation
temperature.
 Change in room temperature or
humidity.
 Failure in sampling system.
 Failure in reagent dispense system.
 Inaccurate calibration/recalibration.
Calibrators
 Determines the accuracy and precision of the
analyzer using a specifically formulated product
in order to recover each parameter within close
tolerances of known target values and limits.
 Finetunes the hematology analyzer to provide the
most accurate results possible.
 Coefficients of variation and percent difference
recovery must be within their specified limits.
Why to Calibrate?
Calibration is necessary
To ensure readings from an instrument
are consistent with other measurements.
To determine the accuracy of the
instrument readings.
To establish the reliability of the
instrument i.e. that it can be trusted.
When to Calibrate?
At installation
After the replacement of any component that
involves dilution characteristics or the primary
measurements (such as the apertures)
When advised to do so by your service
representative
Automation in Hematology part 2

More Related Content

What's hot

Automation in clinical hematology
Automation in clinical hematologyAutomation in clinical hematology
Automation in clinical hematologyTitto Rahim
 
special and routine stains in haematology 1
special and routine stains in haematology 1special and routine stains in haematology 1
special and routine stains in haematology 1Dr.SHAHID Raza
 
Raju automation Msc MLT PATHOLOGY
Raju automation Msc MLT PATHOLOGYRaju automation Msc MLT PATHOLOGY
Raju automation Msc MLT PATHOLOGYrajusehrawat
 
CBC Histogram DR NARMADA PRASAD TIWARI
 CBC Histogram DR NARMADA PRASAD TIWARI CBC Histogram DR NARMADA PRASAD TIWARI
CBC Histogram DR NARMADA PRASAD TIWARINarmada Tiwari
 
Automation in cytology.
Automation in cytology.Automation in cytology.
Automation in cytology.Manan Shah
 
Automation in urine analysis
Automation in urine analysisAutomation in urine analysis
Automation in urine analysisVarun Singh
 
Demonstration of le cells
Demonstration of le cellsDemonstration of le cells
Demonstration of le cellsSHRUTHI VASAN
 
Gel card technology ppt nc
Gel card technology ppt ncGel card technology ppt nc
Gel card technology ppt ncNainshi Bhatt
 
Interpretation of histograms
Interpretation of histogramsInterpretation of histograms
Interpretation of histogramsPankaj Gupta
 
Quality Control in Blood Bank
Quality Control in Blood BankQuality Control in Blood Bank
Quality Control in Blood BankNashwa Elsayed
 
PRINCIPLES OF OPERATIONOF THE AUTOMATED COAGULATION ANALZER.pptx
PRINCIPLES OF OPERATIONOF THE AUTOMATED COAGULATION ANALZER.pptxPRINCIPLES OF OPERATIONOF THE AUTOMATED COAGULATION ANALZER.pptx
PRINCIPLES OF OPERATIONOF THE AUTOMATED COAGULATION ANALZER.pptxEmmanuelAdoku
 
CYTOLOGY OF CSF
CYTOLOGY OF CSFCYTOLOGY OF CSF
CYTOLOGY OF CSFMusa Khan
 
Cbc coulter principle
Cbc  coulter principleCbc  coulter principle
Cbc coulter principlejadcaesar
 
Automated cell counters: principle and types
Automated cell counters: principle and typesAutomated cell counters: principle and types
Automated cell counters: principle and typesSivaranjini N
 

What's hot (20)

Automation in clinical hematology
Automation in clinical hematologyAutomation in clinical hematology
Automation in clinical hematology
 
Automated cell counters
Automated cell countersAutomated cell counters
Automated cell counters
 
special and routine stains in haematology 1
special and routine stains in haematology 1special and routine stains in haematology 1
special and routine stains in haematology 1
 
Raju automation Msc MLT PATHOLOGY
Raju automation Msc MLT PATHOLOGYRaju automation Msc MLT PATHOLOGY
Raju automation Msc MLT PATHOLOGY
 
Laboratory hematology
Laboratory hematologyLaboratory hematology
Laboratory hematology
 
CBC Histogram DR NARMADA PRASAD TIWARI
 CBC Histogram DR NARMADA PRASAD TIWARI CBC Histogram DR NARMADA PRASAD TIWARI
CBC Histogram DR NARMADA PRASAD TIWARI
 
Reticulocyte count
Reticulocyte countReticulocyte count
Reticulocyte count
 
Automation in cytology.
Automation in cytology.Automation in cytology.
Automation in cytology.
 
Automation in urine analysis
Automation in urine analysisAutomation in urine analysis
Automation in urine analysis
 
Demonstration of le cells
Demonstration of le cellsDemonstration of le cells
Demonstration of le cells
 
Hematology auto analyzer
Hematology auto analyzerHematology auto analyzer
Hematology auto analyzer
 
Gel card technology ppt nc
Gel card technology ppt ncGel card technology ppt nc
Gel card technology ppt nc
 
Interpretation of histograms
Interpretation of histogramsInterpretation of histograms
Interpretation of histograms
 
Quality Control in Blood Bank
Quality Control in Blood BankQuality Control in Blood Bank
Quality Control in Blood Bank
 
PRINCIPLES OF OPERATIONOF THE AUTOMATED COAGULATION ANALZER.pptx
PRINCIPLES OF OPERATIONOF THE AUTOMATED COAGULATION ANALZER.pptxPRINCIPLES OF OPERATIONOF THE AUTOMATED COAGULATION ANALZER.pptx
PRINCIPLES OF OPERATIONOF THE AUTOMATED COAGULATION ANALZER.pptx
 
CYTOLOGY OF CSF
CYTOLOGY OF CSFCYTOLOGY OF CSF
CYTOLOGY OF CSF
 
Thalhgbopathy
ThalhgbopathyThalhgbopathy
Thalhgbopathy
 
Cbc coulter principle
Cbc  coulter principleCbc  coulter principle
Cbc coulter principle
 
morphology of red blood cells
morphology of red blood cellsmorphology of red blood cells
morphology of red blood cells
 
Automated cell counters: principle and types
Automated cell counters: principle and typesAutomated cell counters: principle and types
Automated cell counters: principle and types
 

Viewers also liked

Automated cell counters
Automated  cell countersAutomated  cell counters
Automated cell countersRicha Sharma
 
Medical Economics of Automatic Hematology Analysis
Medical Economics of Automatic Hematology AnalysisMedical Economics of Automatic Hematology Analysis
Medical Economics of Automatic Hematology Analysismitch81
 
Soft tissue lecture
Soft tissue lectureSoft tissue lecture
Soft tissue lectureGhie Santos
 
Medical revalidation for histopathologists
Medical revalidation for histopathologistsMedical revalidation for histopathologists
Medical revalidation for histopathologistsDr. Varughese George
 
good practices in the clinical laboratory
good practices in the clinical laboratorygood practices in the clinical laboratory
good practices in the clinical laboratoryGhie Santos
 
Lower female gen tract lecture
Lower female gen tract lectureLower female gen tract lecture
Lower female gen tract lectureGhie Santos
 
Sudden unexpected death in infancy
Sudden unexpected death in infancySudden unexpected death in infancy
Sudden unexpected death in infancyDr. Varughese George
 
5. amyloidosis dr. sinhasan, mdzah
5. amyloidosis dr. sinhasan, mdzah5. amyloidosis dr. sinhasan, mdzah
5. amyloidosis dr. sinhasan, mdzahkciapm
 
1. pediatric tumors dr. sinhasan, mdzah
1. pediatric tumors  dr. sinhasan, mdzah1. pediatric tumors  dr. sinhasan, mdzah
1. pediatric tumors dr. sinhasan, mdzahkciapm
 
Use of blood components in clinical practice - Part 2
Use of blood components in clinical practice - Part 2Use of blood components in clinical practice - Part 2
Use of blood components in clinical practice - Part 2Dr. Varughese George
 
KCIAPM SLIDE SEMINAR 2016: RV METROPOLIS. CASE FILES
KCIAPM SLIDE SEMINAR 2016: RV METROPOLIS. CASE FILESKCIAPM SLIDE SEMINAR 2016: RV METROPOLIS. CASE FILES
KCIAPM SLIDE SEMINAR 2016: RV METROPOLIS. CASE FILESkciapm
 

Viewers also liked (20)

Laboratory diagnosis of anemia
Laboratory diagnosis of anemiaLaboratory diagnosis of anemia
Laboratory diagnosis of anemia
 
Differential leukocyte count
Differential leukocyte countDifferential leukocyte count
Differential leukocyte count
 
Laboratory Diagnosis of Jaundice
Laboratory Diagnosis of JaundiceLaboratory Diagnosis of Jaundice
Laboratory Diagnosis of Jaundice
 
Hemoglobin estimation
Hemoglobin estimationHemoglobin estimation
Hemoglobin estimation
 
STOOL EXAMINATION
STOOL EXAMINATIONSTOOL EXAMINATION
STOOL EXAMINATION
 
Automated cell counters
Automated  cell countersAutomated  cell counters
Automated cell counters
 
Fecalysis
FecalysisFecalysis
Fecalysis
 
Medical Economics of Automatic Hematology Analysis
Medical Economics of Automatic Hematology AnalysisMedical Economics of Automatic Hematology Analysis
Medical Economics of Automatic Hematology Analysis
 
Soft tissue lecture
Soft tissue lectureSoft tissue lecture
Soft tissue lecture
 
Medical revalidation for histopathologists
Medical revalidation for histopathologistsMedical revalidation for histopathologists
Medical revalidation for histopathologists
 
good practices in the clinical laboratory
good practices in the clinical laboratorygood practices in the clinical laboratory
good practices in the clinical laboratory
 
Lower female gen tract lecture
Lower female gen tract lectureLower female gen tract lecture
Lower female gen tract lecture
 
Sudden unexpected death in infancy
Sudden unexpected death in infancySudden unexpected death in infancy
Sudden unexpected death in infancy
 
5. amyloidosis dr. sinhasan, mdzah
5. amyloidosis dr. sinhasan, mdzah5. amyloidosis dr. sinhasan, mdzah
5. amyloidosis dr. sinhasan, mdzah
 
Haematology Analyzer
Haematology Analyzer  Haematology Analyzer
Haematology Analyzer
 
Thyroid neoplasms
Thyroid neoplasmsThyroid neoplasms
Thyroid neoplasms
 
1. pediatric tumors dr. sinhasan, mdzah
1. pediatric tumors  dr. sinhasan, mdzah1. pediatric tumors  dr. sinhasan, mdzah
1. pediatric tumors dr. sinhasan, mdzah
 
Hematoxylin and Eosin Staining
Hematoxylin and Eosin StainingHematoxylin and Eosin Staining
Hematoxylin and Eosin Staining
 
Use of blood components in clinical practice - Part 2
Use of blood components in clinical practice - Part 2Use of blood components in clinical practice - Part 2
Use of blood components in clinical practice - Part 2
 
KCIAPM SLIDE SEMINAR 2016: RV METROPOLIS. CASE FILES
KCIAPM SLIDE SEMINAR 2016: RV METROPOLIS. CASE FILESKCIAPM SLIDE SEMINAR 2016: RV METROPOLIS. CASE FILES
KCIAPM SLIDE SEMINAR 2016: RV METROPOLIS. CASE FILES
 

Similar to Automation in Hematology part 2

Histogram Interpretation for result analysis
Histogram Interpretation for result analysisHistogram Interpretation for result analysis
Histogram Interpretation for result analysissiddhidivekar2
 
AAPS 2015_W3081_Biomarker Screening Poster_Russell
AAPS 2015_W3081_Biomarker Screening Poster_RussellAAPS 2015_W3081_Biomarker Screening Poster_Russell
AAPS 2015_W3081_Biomarker Screening Poster_RussellLawrence Hwang
 
Hematology analyzer channels-modes- handling errors
Hematology analyzer  channels-modes- handling errorsHematology analyzer  channels-modes- handling errors
Hematology analyzer channels-modes- handling errorsDrAbdulrazzaqAlagbar
 
Interpretation of cbc
Interpretation of cbcInterpretation of cbc
Interpretation of cbcAlaa Abozied
 
Automation in haematology
Automation in haematologyAutomation in haematology
Automation in haematologyraj kumar
 
Paper id 212014108
Paper id 212014108Paper id 212014108
Paper id 212014108IJRAT
 
9637a2ed-9b43-4645-aba0-950b8bc23296.pdf
9637a2ed-9b43-4645-aba0-950b8bc23296.pdf9637a2ed-9b43-4645-aba0-950b8bc23296.pdf
9637a2ed-9b43-4645-aba0-950b8bc23296.pdfkarenflores225
 
Xtra online platelet_distribution_curves
Xtra online platelet_distribution_curvesXtra online platelet_distribution_curves
Xtra online platelet_distribution_curvesAlaa Fakhri
 
Blood cells counters types - linearity
Blood cells counters  types - linearityBlood cells counters  types - linearity
Blood cells counters types - linearityDrAbdulrazzaqAlagbar
 
AUTOMATION IN HAEMATOLOGY.pptx
AUTOMATION IN HAEMATOLOGY.pptxAUTOMATION IN HAEMATOLOGY.pptx
AUTOMATION IN HAEMATOLOGY.pptxFraishu
 
L25 b met_complete_blood_count
L25 b met_complete_blood_countL25 b met_complete_blood_count
L25 b met_complete_blood_countShakeel Mujahid
 
Histosch cw ieng
Histosch cw iengHistosch cw ieng
Histosch cw iengbabu3151
 
SCREENING TUBEs_ICCS_2016-AR
SCREENING TUBEs_ICCS_2016-ARSCREENING TUBEs_ICCS_2016-AR
SCREENING TUBEs_ICCS_2016-ARAmr Rajab
 

Similar to Automation in Hematology part 2 (20)

Histogram Interpretation for result analysis
Histogram Interpretation for result analysisHistogram Interpretation for result analysis
Histogram Interpretation for result analysis
 
Lab test 1.pptx
Lab test 1.pptxLab test 1.pptx
Lab test 1.pptx
 
AAPS 2015_W3081_Biomarker Screening Poster_Russell
AAPS 2015_W3081_Biomarker Screening Poster_RussellAAPS 2015_W3081_Biomarker Screening Poster_Russell
AAPS 2015_W3081_Biomarker Screening Poster_Russell
 
Abnormalities in leukocyte number.ppt
Abnormalities in leukocyte number.pptAbnormalities in leukocyte number.ppt
Abnormalities in leukocyte number.ppt
 
Hematology analyzer channels-modes- handling errors
Hematology analyzer  channels-modes- handling errorsHematology analyzer  channels-modes- handling errors
Hematology analyzer channels-modes- handling errors
 
CellCounter_DAE_V0.0.pptx
CellCounter_DAE_V0.0.pptxCellCounter_DAE_V0.0.pptx
CellCounter_DAE_V0.0.pptx
 
Interpretation of cbc
Interpretation of cbcInterpretation of cbc
Interpretation of cbc
 
Hematology
HematologyHematology
Hematology
 
Automation in haematology
Automation in haematologyAutomation in haematology
Automation in haematology
 
Heamatology test and techniques
Heamatology test and techniquesHeamatology test and techniques
Heamatology test and techniques
 
Heamatology test and techniques
Heamatology test and techniquesHeamatology test and techniques
Heamatology test and techniques
 
Paper id 212014108
Paper id 212014108Paper id 212014108
Paper id 212014108
 
9637a2ed-9b43-4645-aba0-950b8bc23296.pdf
9637a2ed-9b43-4645-aba0-950b8bc23296.pdf9637a2ed-9b43-4645-aba0-950b8bc23296.pdf
9637a2ed-9b43-4645-aba0-950b8bc23296.pdf
 
cbc histogram.pdf
cbc histogram.pdfcbc histogram.pdf
cbc histogram.pdf
 
Xtra online platelet_distribution_curves
Xtra online platelet_distribution_curvesXtra online platelet_distribution_curves
Xtra online platelet_distribution_curves
 
Blood cells counters types - linearity
Blood cells counters  types - linearityBlood cells counters  types - linearity
Blood cells counters types - linearity
 
AUTOMATION IN HAEMATOLOGY.pptx
AUTOMATION IN HAEMATOLOGY.pptxAUTOMATION IN HAEMATOLOGY.pptx
AUTOMATION IN HAEMATOLOGY.pptx
 
L25 b met_complete_blood_count
L25 b met_complete_blood_countL25 b met_complete_blood_count
L25 b met_complete_blood_count
 
Histosch cw ieng
Histosch cw iengHistosch cw ieng
Histosch cw ieng
 
SCREENING TUBEs_ICCS_2016-AR
SCREENING TUBEs_ICCS_2016-ARSCREENING TUBEs_ICCS_2016-AR
SCREENING TUBEs_ICCS_2016-AR
 

More from Dr. Varughese George (20)

Benign White blood cell (WBC) Disorders
Benign White blood cell (WBC) DisordersBenign White blood cell (WBC) Disorders
Benign White blood cell (WBC) Disorders
 
Urine analysis Class I
Urine analysis   Class IUrine analysis   Class I
Urine analysis Class I
 
Urine analysis Class II
Urine analysis   Class IIUrine analysis   Class II
Urine analysis Class II
 
Stool Examination
Stool ExaminationStool Examination
Stool Examination
 
Stool Examination Abridged What A Medical Graduate Should Know
Stool Examination Abridged   What A Medical Graduate Should KnowStool Examination Abridged   What A Medical Graduate Should Know
Stool Examination Abridged What A Medical Graduate Should Know
 
Muscle biopsy
Muscle biopsyMuscle biopsy
Muscle biopsy
 
Cpc meeting 12032020
Cpc meeting 12032020Cpc meeting 12032020
Cpc meeting 12032020
 
Cpc meeting 17022020
Cpc meeting 17022020Cpc meeting 17022020
Cpc meeting 17022020
 
Urine analysis
Urine analysisUrine analysis
Urine analysis
 
MOSCMC CPC 28102019 - Follicular Lymphoma
MOSCMC CPC 28102019 - Follicular LymphomaMOSCMC CPC 28102019 - Follicular Lymphoma
MOSCMC CPC 28102019 - Follicular Lymphoma
 
Infectious diseases
Infectious diseasesInfectious diseases
Infectious diseases
 
Diseases of the kidney
Diseases of the kidneyDiseases of the kidney
Diseases of the kidney
 
Aneurysms
AneurysmsAneurysms
Aneurysms
 
Tumors of the breast
Tumors of the breastTumors of the breast
Tumors of the breast
 
Tumor board Ca stomach
Tumor board Ca stomachTumor board Ca stomach
Tumor board Ca stomach
 
Tumor board soft tissue sarcoma
Tumor board  soft tissue sarcomaTumor board  soft tissue sarcoma
Tumor board soft tissue sarcoma
 
Approach to infertility
Approach to infertilityApproach to infertility
Approach to infertility
 
Tumor board melanoma 28082018
Tumor board melanoma 28082018Tumor board melanoma 28082018
Tumor board melanoma 28082018
 
Tumor board ca stomach 28082018
Tumor board ca stomach 28082018Tumor board ca stomach 28082018
Tumor board ca stomach 28082018
 
Dermatitis herpetiformis
Dermatitis herpetiformisDermatitis herpetiformis
Dermatitis herpetiformis
 

Recently uploaded

Top Quality Call Girl Service Kalyanpur 6378878445 Available Call Girls Any Time
Top Quality Call Girl Service Kalyanpur 6378878445 Available Call Girls Any TimeTop Quality Call Girl Service Kalyanpur 6378878445 Available Call Girls Any Time
Top Quality Call Girl Service Kalyanpur 6378878445 Available Call Girls Any TimeCall Girls Delhi
 
Call Girls Gwalior Just Call 8617370543 Top Class Call Girl Service Available
Call Girls Gwalior Just Call 8617370543 Top Class Call Girl Service AvailableCall Girls Gwalior Just Call 8617370543 Top Class Call Girl Service Available
Call Girls Gwalior Just Call 8617370543 Top Class Call Girl Service AvailableDipal Arora
 
Call Girls Bareilly Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Bareilly Just Call 9907093804 Top Class Call Girl Service AvailableCall Girls Bareilly Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Bareilly Just Call 9907093804 Top Class Call Girl Service AvailableDipal Arora
 
Call Girls Nagpur Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Nagpur Just Call 9907093804 Top Class Call Girl Service AvailableCall Girls Nagpur Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Nagpur Just Call 9907093804 Top Class Call Girl Service AvailableDipal Arora
 
Russian Call Girls in Jaipur Riya WhatsApp ❤8445551418 VIP Call Girls Jaipur
Russian Call Girls in Jaipur Riya WhatsApp ❤8445551418 VIP Call Girls JaipurRussian Call Girls in Jaipur Riya WhatsApp ❤8445551418 VIP Call Girls Jaipur
Russian Call Girls in Jaipur Riya WhatsApp ❤8445551418 VIP Call Girls Jaipurparulsinha
 
Manyata Tech Park ( Call Girls ) Bangalore ✔ 6297143586 ✔ Hot Model With Sexy...
Manyata Tech Park ( Call Girls ) Bangalore ✔ 6297143586 ✔ Hot Model With Sexy...Manyata Tech Park ( Call Girls ) Bangalore ✔ 6297143586 ✔ Hot Model With Sexy...
Manyata Tech Park ( Call Girls ) Bangalore ✔ 6297143586 ✔ Hot Model With Sexy...vidya singh
 
Best Rate (Guwahati ) Call Girls Guwahati ⟟ 8617370543 ⟟ High Class Call Girl...
Best Rate (Guwahati ) Call Girls Guwahati ⟟ 8617370543 ⟟ High Class Call Girl...Best Rate (Guwahati ) Call Girls Guwahati ⟟ 8617370543 ⟟ High Class Call Girl...
Best Rate (Guwahati ) Call Girls Guwahati ⟟ 8617370543 ⟟ High Class Call Girl...Dipal Arora
 
Top Rated Bangalore Call Girls Richmond Circle ⟟ 8250192130 ⟟ Call Me For Gen...
Top Rated Bangalore Call Girls Richmond Circle ⟟ 8250192130 ⟟ Call Me For Gen...Top Rated Bangalore Call Girls Richmond Circle ⟟ 8250192130 ⟟ Call Me For Gen...
Top Rated Bangalore Call Girls Richmond Circle ⟟ 8250192130 ⟟ Call Me For Gen...narwatsonia7
 
Best Rate (Hyderabad) Call Girls Jahanuma ⟟ 8250192130 ⟟ High Class Call Girl...
Best Rate (Hyderabad) Call Girls Jahanuma ⟟ 8250192130 ⟟ High Class Call Girl...Best Rate (Hyderabad) Call Girls Jahanuma ⟟ 8250192130 ⟟ High Class Call Girl...
Best Rate (Hyderabad) Call Girls Jahanuma ⟟ 8250192130 ⟟ High Class Call Girl...astropune
 
Call Girls Kochi Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Kochi Just Call 9907093804 Top Class Call Girl Service AvailableCall Girls Kochi Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Kochi Just Call 9907093804 Top Class Call Girl Service AvailableDipal Arora
 
All Time Service Available Call Girls Marine Drive 📳 9820252231 For 18+ VIP C...
All Time Service Available Call Girls Marine Drive 📳 9820252231 For 18+ VIP C...All Time Service Available Call Girls Marine Drive 📳 9820252231 For 18+ VIP C...
All Time Service Available Call Girls Marine Drive 📳 9820252231 For 18+ VIP C...Arohi Goyal
 
Lucknow Call girls - 8800925952 - 24x7 service with hotel room
Lucknow Call girls - 8800925952 - 24x7 service with hotel roomLucknow Call girls - 8800925952 - 24x7 service with hotel room
Lucknow Call girls - 8800925952 - 24x7 service with hotel roomdiscovermytutordmt
 
Call Girls Service Jaipur Grishma WhatsApp ❤8445551418 VIP Call Girls Jaipur
Call Girls Service Jaipur Grishma WhatsApp ❤8445551418 VIP Call Girls JaipurCall Girls Service Jaipur Grishma WhatsApp ❤8445551418 VIP Call Girls Jaipur
Call Girls Service Jaipur Grishma WhatsApp ❤8445551418 VIP Call Girls Jaipurparulsinha
 
Call Girls Aurangabad Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Aurangabad Just Call 9907093804 Top Class Call Girl Service AvailableCall Girls Aurangabad Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Aurangabad Just Call 9907093804 Top Class Call Girl Service AvailableDipal Arora
 
Top Rated Bangalore Call Girls Mg Road ⟟ 8250192130 ⟟ Call Me For Genuine Sex...
Top Rated Bangalore Call Girls Mg Road ⟟ 8250192130 ⟟ Call Me For Genuine Sex...Top Rated Bangalore Call Girls Mg Road ⟟ 8250192130 ⟟ Call Me For Genuine Sex...
Top Rated Bangalore Call Girls Mg Road ⟟ 8250192130 ⟟ Call Me For Genuine Sex...narwatsonia7
 
Call Girls Visakhapatnam Just Call 9907093804 Top Class Call Girl Service Ava...
Call Girls Visakhapatnam Just Call 9907093804 Top Class Call Girl Service Ava...Call Girls Visakhapatnam Just Call 9907093804 Top Class Call Girl Service Ava...
Call Girls Visakhapatnam Just Call 9907093804 Top Class Call Girl Service Ava...Dipal Arora
 
Call Girls Cuttack Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Cuttack Just Call 9907093804 Top Class Call Girl Service AvailableCall Girls Cuttack Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Cuttack Just Call 9907093804 Top Class Call Girl Service AvailableDipal Arora
 
Call Girls Siliguri Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Siliguri Just Call 9907093804 Top Class Call Girl Service AvailableCall Girls Siliguri Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Siliguri Just Call 9907093804 Top Class Call Girl Service AvailableDipal Arora
 
Call Girls Coimbatore Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Coimbatore Just Call 9907093804 Top Class Call Girl Service AvailableCall Girls Coimbatore Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Coimbatore Just Call 9907093804 Top Class Call Girl Service AvailableDipal Arora
 
College Call Girls in Haridwar 9667172968 Short 4000 Night 10000 Best call gi...
College Call Girls in Haridwar 9667172968 Short 4000 Night 10000 Best call gi...College Call Girls in Haridwar 9667172968 Short 4000 Night 10000 Best call gi...
College Call Girls in Haridwar 9667172968 Short 4000 Night 10000 Best call gi...perfect solution
 

Recently uploaded (20)

Top Quality Call Girl Service Kalyanpur 6378878445 Available Call Girls Any Time
Top Quality Call Girl Service Kalyanpur 6378878445 Available Call Girls Any TimeTop Quality Call Girl Service Kalyanpur 6378878445 Available Call Girls Any Time
Top Quality Call Girl Service Kalyanpur 6378878445 Available Call Girls Any Time
 
Call Girls Gwalior Just Call 8617370543 Top Class Call Girl Service Available
Call Girls Gwalior Just Call 8617370543 Top Class Call Girl Service AvailableCall Girls Gwalior Just Call 8617370543 Top Class Call Girl Service Available
Call Girls Gwalior Just Call 8617370543 Top Class Call Girl Service Available
 
Call Girls Bareilly Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Bareilly Just Call 9907093804 Top Class Call Girl Service AvailableCall Girls Bareilly Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Bareilly Just Call 9907093804 Top Class Call Girl Service Available
 
Call Girls Nagpur Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Nagpur Just Call 9907093804 Top Class Call Girl Service AvailableCall Girls Nagpur Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Nagpur Just Call 9907093804 Top Class Call Girl Service Available
 
Russian Call Girls in Jaipur Riya WhatsApp ❤8445551418 VIP Call Girls Jaipur
Russian Call Girls in Jaipur Riya WhatsApp ❤8445551418 VIP Call Girls JaipurRussian Call Girls in Jaipur Riya WhatsApp ❤8445551418 VIP Call Girls Jaipur
Russian Call Girls in Jaipur Riya WhatsApp ❤8445551418 VIP Call Girls Jaipur
 
Manyata Tech Park ( Call Girls ) Bangalore ✔ 6297143586 ✔ Hot Model With Sexy...
Manyata Tech Park ( Call Girls ) Bangalore ✔ 6297143586 ✔ Hot Model With Sexy...Manyata Tech Park ( Call Girls ) Bangalore ✔ 6297143586 ✔ Hot Model With Sexy...
Manyata Tech Park ( Call Girls ) Bangalore ✔ 6297143586 ✔ Hot Model With Sexy...
 
Best Rate (Guwahati ) Call Girls Guwahati ⟟ 8617370543 ⟟ High Class Call Girl...
Best Rate (Guwahati ) Call Girls Guwahati ⟟ 8617370543 ⟟ High Class Call Girl...Best Rate (Guwahati ) Call Girls Guwahati ⟟ 8617370543 ⟟ High Class Call Girl...
Best Rate (Guwahati ) Call Girls Guwahati ⟟ 8617370543 ⟟ High Class Call Girl...
 
Top Rated Bangalore Call Girls Richmond Circle ⟟ 8250192130 ⟟ Call Me For Gen...
Top Rated Bangalore Call Girls Richmond Circle ⟟ 8250192130 ⟟ Call Me For Gen...Top Rated Bangalore Call Girls Richmond Circle ⟟ 8250192130 ⟟ Call Me For Gen...
Top Rated Bangalore Call Girls Richmond Circle ⟟ 8250192130 ⟟ Call Me For Gen...
 
Best Rate (Hyderabad) Call Girls Jahanuma ⟟ 8250192130 ⟟ High Class Call Girl...
Best Rate (Hyderabad) Call Girls Jahanuma ⟟ 8250192130 ⟟ High Class Call Girl...Best Rate (Hyderabad) Call Girls Jahanuma ⟟ 8250192130 ⟟ High Class Call Girl...
Best Rate (Hyderabad) Call Girls Jahanuma ⟟ 8250192130 ⟟ High Class Call Girl...
 
Call Girls Kochi Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Kochi Just Call 9907093804 Top Class Call Girl Service AvailableCall Girls Kochi Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Kochi Just Call 9907093804 Top Class Call Girl Service Available
 
All Time Service Available Call Girls Marine Drive 📳 9820252231 For 18+ VIP C...
All Time Service Available Call Girls Marine Drive 📳 9820252231 For 18+ VIP C...All Time Service Available Call Girls Marine Drive 📳 9820252231 For 18+ VIP C...
All Time Service Available Call Girls Marine Drive 📳 9820252231 For 18+ VIP C...
 
Lucknow Call girls - 8800925952 - 24x7 service with hotel room
Lucknow Call girls - 8800925952 - 24x7 service with hotel roomLucknow Call girls - 8800925952 - 24x7 service with hotel room
Lucknow Call girls - 8800925952 - 24x7 service with hotel room
 
Call Girls Service Jaipur Grishma WhatsApp ❤8445551418 VIP Call Girls Jaipur
Call Girls Service Jaipur Grishma WhatsApp ❤8445551418 VIP Call Girls JaipurCall Girls Service Jaipur Grishma WhatsApp ❤8445551418 VIP Call Girls Jaipur
Call Girls Service Jaipur Grishma WhatsApp ❤8445551418 VIP Call Girls Jaipur
 
Call Girls Aurangabad Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Aurangabad Just Call 9907093804 Top Class Call Girl Service AvailableCall Girls Aurangabad Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Aurangabad Just Call 9907093804 Top Class Call Girl Service Available
 
Top Rated Bangalore Call Girls Mg Road ⟟ 8250192130 ⟟ Call Me For Genuine Sex...
Top Rated Bangalore Call Girls Mg Road ⟟ 8250192130 ⟟ Call Me For Genuine Sex...Top Rated Bangalore Call Girls Mg Road ⟟ 8250192130 ⟟ Call Me For Genuine Sex...
Top Rated Bangalore Call Girls Mg Road ⟟ 8250192130 ⟟ Call Me For Genuine Sex...
 
Call Girls Visakhapatnam Just Call 9907093804 Top Class Call Girl Service Ava...
Call Girls Visakhapatnam Just Call 9907093804 Top Class Call Girl Service Ava...Call Girls Visakhapatnam Just Call 9907093804 Top Class Call Girl Service Ava...
Call Girls Visakhapatnam Just Call 9907093804 Top Class Call Girl Service Ava...
 
Call Girls Cuttack Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Cuttack Just Call 9907093804 Top Class Call Girl Service AvailableCall Girls Cuttack Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Cuttack Just Call 9907093804 Top Class Call Girl Service Available
 
Call Girls Siliguri Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Siliguri Just Call 9907093804 Top Class Call Girl Service AvailableCall Girls Siliguri Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Siliguri Just Call 9907093804 Top Class Call Girl Service Available
 
Call Girls Coimbatore Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Coimbatore Just Call 9907093804 Top Class Call Girl Service AvailableCall Girls Coimbatore Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Coimbatore Just Call 9907093804 Top Class Call Girl Service Available
 
College Call Girls in Haridwar 9667172968 Short 4000 Night 10000 Best call gi...
College Call Girls in Haridwar 9667172968 Short 4000 Night 10000 Best call gi...College Call Girls in Haridwar 9667172968 Short 4000 Night 10000 Best call gi...
College Call Girls in Haridwar 9667172968 Short 4000 Night 10000 Best call gi...
 

Automation in Hematology part 2

  • 3. I N D E X PART 1 I. Necessity for Automation. II. Advantages & Disadvantages of Automation. III. Types of Automated Hematology Analyzers. IV. Principles involved in Automation. V. Pentra ES 60 Haematology Analyzer. VI. Pentra DF Nexus Haematology Analyzer. PART 2 I. Histograms. II. Flags III. Quality Control
  • 4. Histograms  Graphical representation of numerical data of different cell populations in a cell counter.  x-axis => cell size  y-axis => no: of cells  Gives info on  Average size  Distribution of size
  • 5. Discrimination thresholds  Moving or fixed discriminator separates the distribution curve for the volume  WBC Discriminator  The automated counter sets a LD flunctuating between 30-60 fl & an UD fixed at 300 fl.  WBC is calculated from particle counts > than this LD  RBC Discriminator Has two flexible discriminators LD (25-75 fl) & UD (200-250 fl). RBC is calculated from particle counts between this LD & UD.  Platelet Discriminator Three discriminators – LD (2-6 fl), UD (12-30 fl) & a fixed discriminator (8-12 fl)
  • 6. Estimation of RBC Count & MCV – RBC Histogram Normal RBC distribution curve is Gaussian bell shaped curve. Analyzer counts RBCs as those which ranges from 36-360 fl. MCV is perpendicular line from peak of the curve to the base. Peak of curve should fall within the normal MCV range of 80-100 fl. There are 2 flexible discriminators – LD (25-75 fl) & UD (200-250 fl)
  • 7. Abnormalities of RBC Histogram Left shift of the curve in microcytosis. Right shift of the curve in macrocytosis. Bimodal peak of the curve in dimorphic population of cells.
  • 9. Estimation of RDW RDW is expressed as a coefficient of variation of RBC size distribution. RDW- CV is a better indicator of anisocytosis than RDW-SD. RBC distribution curve ll get wider as RBC vary more in size. Normal Range – RDW-CV – 11.0-15.0% RDW-SD – 40.0 - 55.0 fL
  • 10. RDW
  • 11. Estimation of WBC Count  Hematology analyzers can generate a  3-part differential count – lymphocytes, monocytes & granulocytes based on the principle of electrical impedence Or  5-part differential count - lymphocytes, monocytes, neutrophils, eosinophils & basophils based on different principles –  light scatter  electrical impedence  electrical conductivity  peroxidase staining
  • 12. Estimation of WBC Count – WBC Histogram  Cells > 35 fl are counted as WBCs in the WBC/Hb chamber. Cells with volume 35-90 fl  Lymphocytes. Cells with volume 90-160 fl  Mononuclear cells. Cells with volume 160-450 fl  Neutrophils.
  • 13. Abnormalities of WBC Histogram  Peak to the left of lymphocyte peak – nucleated cells.  Peak between lymphocytes & monocytes – blast cells, eosinophilia, basophilia, plasma cells & atypical lymphocytes.  Peak between monocytes & neutrophils – left shift
  • 14. 2-6 fl 12-30 fl fixed at 12 fl Platelet Histogram PL PU PLT RBC 100% 20%  PLT size: 8-12 fL  PLT detection: between 2 and 30 fL  Fixed discriminator at 12 fL
  • 15. Estimation of Platelet Count Platelets are counted by the electrical impedence method in the RBC aperture. Particles between 2 fl and less than 20 fl are classified as platelets by the analyzer. MPV is a measurement of the average size of platelets found in blood Normal MPV is 7-10 fl. ↑ MPV ( > 10 fl) d/t destruction of platelets in circulation. ↓ MPV ( < 7 fl) d/t impaired production of platelets in circulation.  Plateletocrit (PCT) is volume of circulating platelets in a unit volume of blood. Normal PCT is 0.19-0.36%.  PCT ↑ in thrombocytosis and ↓ in thrombocytopenia. PDW is a measure of variation in the size of platelets. Standard PDW ranges from 9 to 14 fL ↑ PDW is observed in megaloblastic anemia, CML & after chemotherapy.
  • 16. I N D E X PART 1 I. Necessity for Automation. II. Advantages & Disadvantages of Automation. III. Types of Automated Hematology Analyzers. IV. Principles involved in Automation. V. Pentra ES 60 Haematology Analyzer. VI. Pentra DF Nexus Haematology Analyzer. PART 2 I. Histograms. II. Flags. III. Quality Control.
  • 17. Flagging  Flags are signals that occur when an abnormal result is detected by the automated blood analyzer.  Flags are signaled by certain ‘asteriks’ on the report.  They reduce the False +ve & False –ve results by mandating the results of blood smear examination.
  • 18. RBC Flags  Seen when LD > preset height by 10 %  Shown by  RBC Count  HCT, MCV,MCH,MCHC  Occurs when there is  platelet aggregation  RBC fragments RL Flag
  • 19. RBC Flags  Seen when UD > preset height by 10%.  Shown by  RBC Count  HCT, MCV,MCH,MCHC  Occurs when there are  Cold agglutinins RU Flag
  • 20. RBC Flags  Shown by RDW-SD  Seen in  Post Blood tranfusion  Treated Fe deficiency anemia. MP- Flag
  • 21. WBC Flags  Generated when curve deviates from the baseline on the LD.  Various causes for this are –  Platelet aggregates (clotted sample, EDTA incompactibility)  Lyse-resistant RBCs.  Erythroblasts.  Cryoagglutinates.  Giant platelets. WL Flag
  • 22. WBC Flags  Generated when there is deviation of the curve on the UD or if it does not end at the baseline.  Various causes for this are :-  Hyperleucocytosis. WU Flag
  • 23. WBC Flags - T1 & T2 Flags Peak between T1-T2 : The middle cell count - Eosinophils, Monocytes, Blasts, promyelocytes, myelocytes and metamyelocytes. Peak between LD-T1 : Lymphocytes Peak between T2-UD : Neutrophils. T1 & T2 flags appears when discrimination between lymphocytes, middle cells & neutrophils is not possible which happens in presence of abnormal/higher leucocyte counts as in Chronic Myeloid Leukemia.
  • 24. WBC Flags - F1, F2, F3 Flags  Sometimes, the cell Fractions may be mixed.  F1 & F2 or F2 or F3 merge into each other over large areas.  F1 (small cell inaccurate) flag : Acute Lymphocytic Leukemia  F2 (middle cell inaccurate) flag : eosinophil, Acute Myeloid Leukemia, monocytosis, etc  F3 (large cell inaccurate) flag:
  • 25. Platelet Flags  This occurs when the LD > the preset height by 10%  Shown by  Platelet count  MPV  P-LCR  Occurs due to noise. PL Flag
  • 26. Platelet Flags  This occurs when the UD> the preset height by > 40%.  Occurs in  Hemolytic anemias with fragmented cells.  Large platelets. PU Flag
  • 28.
  • 29. I N D E X PART 1 I. Necessity for Automation. II. Advantages & Disadvantages of Automation. III. Types of Automated Hematology Analyzers. IV. Principles involved in Automation. V. Pentra ES 60 Haematology Analyzer. VI. Pentra DF Nexus Haematology Analyzer. PART 2 I. Histograms. II. Flags III. Quality Control
  • 30. Quality Control / Quality Assurance/Quality Assessment  Quality Control => measures that must be included during each assay run that the test is working properly.  Quality Assurance => overall program that ensures that the final results reported by the lab are correct  Quality Assessment => means to determine the quality of results generated by the lab
  • 31. QUALITY CONTROL  Measure of precision – how well the measurement system reproduces the same results over time and under varying operating conditions.  Designed to detect, reduce & correct deficiencies in a laboratory’s internal analytical processes prior to release of patient’s reports.  Should approximate the same matrix as patient’s specimens, taking into account properties such as viscosity, turbidity, composition and color.  Should be simple to use, with minimal vial to vial comparability.  Should be stable for long periods of time and available in large enough quantities for a single batch to last at least 1 year.
  • 32. Types of Quality Control Internal  Continuous evaluation of reliability of the daily works of the laboratory with validation of tests.  Primary tool required is called a control – a specimen with a predetermined range of result values, processed in the same manner as patient sample.  If the result of a test on a control sample is different from its known value, this indicates a problem in the equipment or the methods being used. External  Evaluation by an outside agency of the comparability a laboratory's testing to a source outside the laboratory.  This comparison can be made to the performance of a peer group of laboratories or to the performance of a reference laboratory.  The analysis of performance is retrospective.
  • 33. Accuracy and Precision  Accuracy is a measure of rightness.  It refers to closeness to the true value.  Precision is a measure of exactness.  It refers to reproducibility of the test.
  • 34. Control  A solution that contains the same constituents as those being analyzed in the patient sample.  Commercially produced pooled RBCs/sera or stabilized anticoagulated whole blood.  Should have same test properties as that of a blood sample.  At least 1 control specimen should be used for every batch.  If large specimens, use 1 control for every 20 specimens.  For most tests, a “normal” control and an “abnormal” control are analyzed with each patient test or batch of tests.  The results are compared with the manufacturer’s range of values and plotted on a Levey-Jennings Chart.
  • 35. Levey Jennings Graphs  A graphical method for displaying control results and evaluating whether a procedure is in-control or out-of-control  Control values are plotted versus time  Lines are drawn from point to point to accent any trends, shifts, or random excursions  Consecutive values of control are recorded and the standard deviation is calculated.  The mean and ± 2 SD are plotted on a Levey- Jennings chart.  As long as the control value are between the ± 2 SD lines on the L-J chart, the control values are “in control”  If they are outside the ± 2 SD lines, they are 'out of control'
  • 36. Westgard Control Rules  Proposed by Dr. James Westgard on lab quality control.  Set the basis for evaluating analytical run for medical laboratories.  There are 6 basic rules.
  • 37. 12s Rule . Warning rule to trigger careful inspection of the control data A single control measurement exceeding 2 standard deviations of control limits either above or below the mean. No cause for rejecting a run
  • 38. 13s rule A single control measurement exceeds the mean plus 3s or the mean minus 3s control limit. The run must be rejected.
  • 39. 22s rule 2 consecutive control measurements exceed exceeds the +2SD or -2SD control limit . The run must be rejected.
  • 40. R4s rule 1 control measurement exceed the +2SD and the other exceeds the -2SD control limit . The run must be rejected.
  • 41. 41s rule 4 consecutive control measurements exceed exceeds the +1SD or -1SD control limit . The run must be rejected.
  • 42. 10x rule 10 consecutive quality control results for one level of control are on one side of the mean The run must be rejected.
  • 43. ‘Out of Control’  Stop testing  Identify and correct the issue.  Repeat testing on patient samples and controls.  Don’t report patient results until the issue is sorted out and the controls indicate proper performance
  • 44. Errors  Systematic error is evidenced by a change in the mean of control values.  The change in mean maybe • Gradual - demonstrated as a trend in control values. • Abrupt – demonstrated as a shift in control values.  Random error is any deviation away from the expected result.
  • 45. Errors Trend Shift  Gradual loss of reliability in test system. Causes  Deterioration of instrument light source.  Gradual accumulation of debris in sample/reagent tubing and electrode surfaces.  Aging of reagents.  Gradual deterioration of –  Control materials.  Incubation chamber temperature.  Light filter integrity.  Calibration.  Sudden/dramatic +ve or –ve change in test system Causes  Sudden failure or change in the light source.  Change in reagent formulation.  Major instrument maintenance.  Sudden change in incubation temperature.  Change in room temperature or humidity.  Failure in sampling system.  Failure in reagent dispense system.  Inaccurate calibration/recalibration.
  • 46. Calibrators  Determines the accuracy and precision of the analyzer using a specifically formulated product in order to recover each parameter within close tolerances of known target values and limits.  Finetunes the hematology analyzer to provide the most accurate results possible.  Coefficients of variation and percent difference recovery must be within their specified limits.
  • 47. Why to Calibrate? Calibration is necessary To ensure readings from an instrument are consistent with other measurements. To determine the accuracy of the instrument readings. To establish the reliability of the instrument i.e. that it can be trusted.
  • 48. When to Calibrate? At installation After the replacement of any component that involves dilution characteristics or the primary measurements (such as the apertures) When advised to do so by your service representative

Editor's Notes

  1. Hemoglobin (gm/dl) × 3 = PCV Red cell count (million/cmm) × 9 = PCV REFERENCE RANGES • Adult males: 40-50% • Adult females (nonpregnant): 38-45% • Adult females (pregnant): 36-42% • Children 6 to 12 years: 37-46% • Children 6 months to 6 years: 36-42% • Infants 2 to 6 months: 32-42% • Newborns: 44-60% CRITICAL VALUES • Packed cell volume: < 20% or > 60% In a RBC histogram, cell numbers are plotted on Y-axis cell volume are on X-axis. The analyzer counts cells between 36 fl and 360 fl as red cells. Although leukocytes are present and counted along with red cells in the diluting fluid, their number is not statistically significant. Only if leukocyte count is markedly elevated (>50,000/μl), histogram and the red cell count will be affected. Area of the peak between 60 fl and 125 fl is used for calculation of mean cell volume and red cell distribution width. Abnormalities in red cell histogram include: (1) Left shift of the curve in microcytosis, (2) Right shift of the curve in macrocytosis, and (3) Bimodal peak of the curve in double (dimorphic) population of red cells.
  2. REFERENCE RANGES • Mean cell volume: 80-100 fl • Mean cell hemoglobin: 27-32 pg • Mean cell hemoglobin concentration: 30-35 g/dl • Red cell distribution width: 9.0-14.5
  3. Red Blood Cell Distribution Width (RDW): Definition and Calculation The red cell distribution width (RDW) is a measurement derived from the red blood cell distribution curves generated on automated hematology analyzers and is an indicator of variation in red blood cell (RBC) size within a blood sample. The RDW is used along with the indices (MCV, MCH, MCHC) to describe a population of RBCs. The RDW measures the deviation of the RBC width, not the actual width or size of individual cells. The two RDW measurements currently in use are the red cell distribution width - coefficient of variation (RDW-CV) and the red cell distribution width - standard deviation (RDW-SD). The RDW-CV is a calculation based on both the width of the distribution curve and the mean cell size. It is calculated by dividing the standard deviation of the mean cell size by the MCV of the red cells and multiplying by 100 to convert to a percentage. A normal range for the RDW-CV is approximately 11.0 - 15.0%. Because it is a calculation, the RDW-CV is dependent not only on the width of the distribution curve but also the MCV of the red cell population and may not always reflect the actual variation in red cell size. Be aware that: A homogenous population of red cells with a narrow distribution curve and low MCV may have an elevated RDW-CV A heterogeneous population of red cells with a broad distribution curve and a high MCV may have a normal RDW-CV. The RDW-SD is an actual measurement of the width of the red cell distribution curve in femtoliters (fL). The width of the distribution curve is measured at the point that is 20% above the baseline. Since the RDW-SD is an actual measurement, it is not influenced by the MCV and more accurately reflects the red cell size variance. The normal RDW-SD range for adults is 40.0 - 55.0 fL.
  4. MAHA – MicroAngiopathic Hemolytic Anemia
  5. ImmNE1 – Band forms ImmNE2 – Immature neutrophils
  6. Platelets are difficult to count because of it small size, marked variation in size, tendency to aggregation and overlapping of size with microcytic red cells, cellular fragments and other debris. A low MPV indicates average size of platelets is small; older platelets are generally smaller than younger ones and generally a low MPV may mean that a condition is affecting the production of platelets by the bone marrow. A high MPV indicates a high no: of larger, younger platelets in the blood: this may be d/t the bone marrow producing and releasing platelets rapidly into circulation.
  7. MP – Multiple peaks
  8.  The aim of quality control is simply to ensure that the results generated by the test are correct. However, quality assurance is concerned with much more: that the right test is carried out on the right specimen, and that the right result and right interpretation is delivered to the right person at the right time”  Quality assessment (also known as proficiency testing) is a challenge to the effectiveness of the QA and QC programs
  9. Accuracy is a measure of rightness. Precision is a measure of exactness. Precision vs. Accuracy
  10. The manufacturer has analyzed each lot of serum for a variety of test components and the expected range of assay values for each component is provided to the laboratory when shipped.
  11. Whenever a patient’s test or a batch of tests are performed and the control(s) is “in control”, the values obtained for the patient test(s) are determined to be “acceptable” and can be released to the doctor as accurate. Whenever a patient’s test or a batch of tests are performed and the control(s) is “out of control”, the values obtained for the patient test(s) are determined to be “not acceptable” and CANNNOT be released to the doctor as accurate until the problem is identified and resolved.