Xurography Actuated Valving For Centrifugal Flow Control
David J. Kinahana,b
*, Philip L. Earlya,b
, Abhishek Vembadia,c
, Eoghan MacNamarad
,
Niamh A. Kilcawleya,b
, Thomas Glennond
, Dermot Diamondd
, Dermot Brabazonc
and Jens Ducrée*a,b
Received (in XXX, XXX) Xth XXXXXXXXX 20XX, Accepted Xth XXXXXXXXX 20XX
DOI: 10.1039/b000000x5
We introduce a novel instrument controlled valving scheme for centrifugal platforms which is based upon
xurography. In a first approach, which is akin to previously presented event-triggered flow control, the
valves are composed of a pneumatic chamber sealed by a dissolvable film (DF) and by a pierceable
membrane. Liquid is initially prevented from wetting the DF by the counter pressure of a trapped gas. Via
a channel, this pocket is pneumatically connected to a vent, sealed by the pierceable membrane, located10
on the top surface of the disc. By scouring the top surface of the disc, along a pre-defined track by a
robotic knife-cutter, the trapped gas is released and so the liquid can wet and disintegrate the DF. In order
to automate assay protocols without the need to integrate DFs, we extend this xurography-based flow
control concept by selective venting of chambers subjected to pneumatic over-pressure or vacuum
suction. Unlike most instrument controlled flow-control mechanisms, in this approach to valve actuation15
can occur during disc rotation. To demonstrate the potential of this flow control approach, we designed a
disc architecture to automate the liquid handling as the backbone of a biplex liver assay panel. We
demonstrate valve actuation during rotation, using the robotic arm, using this disc with visualisation via
dyed water. We then demonstrate the biplex liver assay, using calibration reagent, by stopping the disc
and manually piercing the membrane to actuate the same valves.20
1. Introduction
Over the recent decades centrifugal microfluidic systems have
increasingly entered the arena of biomedical point-of-care
diagnostics 1-3
. In these “Lab-on-a-Disc” (LoaD) systems, a typic-
ally disc-shaped cartridge is rotated at defined frequency protocol25
by a robust and low-cost spindle motor. Its simple actuation
scheme renders these LoaD devices independent of external
pressure sources and their pneumatic connectors, thus ensuring
ease-of-use and minimum maintenance.
Centrifugal flow control is predominantly implemented by chan-30
ges in the spin rate4
. For instance, in capillary burst valves, the
centrifugally induced hydrostatic pressure overcomes the counter-
acting capillary pressure imposed by a hydrophobic constriction.
Apart from the system-innate spindle motor, flow control can also
be achieved by additional, instrument-based modules, for exam-35
ple by laser ablating a sacrificial membrane5
or by heating a
phase-change material 6
.
Recently, Kinahan et al. introduced event-triggered flow control7,
8
based on dissolvable-film (DF) membranes9
. These valves are
composed of a dead-end pneumatic chamber sealed by two DFs.40
Upon removal of the so-called control film (CF) through contact
with a secondary liquid, the pneumatic chamber is vented to let
the primary liquid escape by dissolving the so-called the load film
(LF). By cascading such centrifugo-pneumatic DF valves, ‘event-
triggered’ flow control has been implemented, e.g. for automating45
an unprecedented number of individual liquid handling steps in a
single disc at an essentially constant spin rate. However, a major
drawback of this event-triggered valving is its poor flexibility of
timing which depends on the dissolution time of the DFs and time
for liquid to move about the disc. Particularly, these timing issues50
can limit the implementation of biomedical assays requiring
extended incubation.
In order to allow discretionary timing, we introduce here an instr-
ument-controlled scheme which is effectively an actuated centri-
fugo-pneumatic DF valve. In this case the CF is replaced by a55
vent, sealed by a pierceable adhesive tape, which is located on the
top surface of the disc (Fig. 1). A blade, adapted from a
commercial knife cutter, is mounted on a 2-axis robotic arm and
positioned above the disc (Fig. 2). During rotation, the blade is
lowered to perforate the adhesive tape and thus venting the60
pneumatic chamber to atmosphere. Only films located along the
same orbit are actuated (quasi) in parallel, so the order of radial
positions where the knife cutter touches the disc defines the
sequence of valving.
Furthermore, we also demonstrate how similar flow control can65
be achieved by using pneumatic overpressure or vacuum suction
10
. While this technique works without incorporation of DFs, it
comes at the expense of limiting the maximum spin rate.
As a proof-of-concept, we demonstrate the automation of multi-
liquid, multi-step sample preparation and reagent handling70
protocol towards a biplex liver assay panel. Critically for on-disc
blood processing, the valves can be externally actuated during
2 | Journal Name, [year], [vol], 00–00 This journal is © The Royal Society of Chemistry [year]
Figure 1: Operation of the xurography-enabled DF valve. (a) The valve consists of a pneumatic chamber with two outlets closed by a DF tab and PSA. (b)
The gas pocket trapped between the meniscus of the incoming liquid and these outlets initially prevents the wetting of the DF. (c) To open the valve, a
knife blade mounted on a robotic arm (cf. Fig. 2) scours the top of the disc, and thereby piercing the PSA. (d) As the gas pocket decompresses, the DF is
wetted and dissolves to clear the passage of liquid to the outlet.5
rotation, thus ensuring efficient plasma extraction. This highly
controllable valving scheme, where liquid handling steps at any
location on the disc can be controlled through interaction with its
external surface, represents a viable path towards large-scale
integration (LSI) and automation of multiplex bio-assays.10
2. Experimental
2.1 Manufacture of Robotic Arm
The robotic arm was composed of two linear precision rails
(Reliance Precision Ltd, UK). It was mounted in an XZ-axis
configuration on two pillar supports. The pillars were15
manufactured from chemically bonded laser-cut sheets of
poly(methyl-methacrylate) (PMMA). Both axes were powered
using stepper actuators (Firgelli Technologies Inc.). The stepper
motor on the X-axis used an integrated potentiometer to provide
position feedback to custom program (LabVIEW, National In-20
struments) via a data acquisition card (National Instruments).
This stepper motor could also accurately position the knife cutter
assembly, which is mounted on the x-axis, to an accuracy of
0.5mm. Along with the Z-axis rail, the knife-cutter assembly was
composed of a backing plate and a stepper actuator which had a25
stroke of 10 mm without positional feedback. Therefore, the
vertical actuator was connected to the blade holder via a spring-
damped connection. In its retracted state, the blade did not
contact the top surface of the disc. In the extended state, the blade
could scour the top surface of the disc. Thus, tolerances in disc30
thickness and mounting were compensated for by the spring
damping. The knife blade was adapted from a commercial knife-
cutter (Graphtec, Japan).
2.2 Disc Manufacture
The discs are manufactured using multilayer polymer fabrication35
methods using methods previously described 7, 11
. The over-
pressure / vacuum-suction valves were manufacture using disc
which were assembled from five layers. Laser-cut PMMA sheets
were used for the upper layer, the reservoir layer and the base
layer. Vents and loading holes were created from voids in the40
upper layer and reservoirs were created from voids in the base
layer. Connecting microchannels were created on the upper and
lower pressure sensitive adhesive (PSA) (Adhesives Research,
Limerick, Ireland) from voids created by a knife-cutter. The
stacking of the two PSA layers permitted microchannels cros-45
sings. The discs for demonstrating xurography actuated DF
valves were manufactured from four PMMA and four PSA layers
as previously described 7
in order to incorporate the DF tabs.
Finally the vent layer of the disc was covered with a transparent
film (Adhesives Research) to seal the disc. Loading holes and50
appropriate vents were then pierced by a scalpel.
2.3 Xurography Actuated DF Valves
The disc implements a biplex liver assay test (see Section 2.4)
starting with plasma extraction. Next, 10-µl volumes of plasma
are metered and then transferred to the periphery of the disc55
where they mix with the assay reagents alkaline phosphatase
(ALP), active direct bilirubin (DB) and passive DB. While the
DB assays involve simpler end-point measurements, the ALP re-
quires the addition of a stop reagent after a defined reaction
interval. Assay automation thus requires five sequentially actu-60
ated valves.
Figure 2: Robotic arm. (a) 3D rendering. (b) The robotic arm mounted on spin stand. The liver assay disc (cf. Fig. 3) is mounted on the experimental rig.
(c) Monochrome image acquired while the disc is spinning at 50 Hz. The track of knife cutter can be seen scratching the PSA layer to vent and open
Valve 1.65
This journal is © The Royal Society of Chemistry [2013] Lab Chip, 2013, [vol], 00–00 | 3
Figure 3: The Lab-on-a-Disc for a biplex liver assay panel measuring Direct Bilirubin (DB) and Alkaline Phosphatase (ALP). (a) The disc displays three se-
parate structures which are operated in parallel. (b-e) Schematic of blood processing on the disc, (b) sample centrifugation, (c) opening of Valve 1 and
metering of plasma, (d) opening of Valve 2, 3 and 4, and transfer of metered sample (10 µl) to mix with reagents. (e) After 3 minutes, the knife cutter
opens Valve 5 to release the reagent stopping the ALP assay. Note that, at all times, the disc is subject to Euler-force based “shake-mode” mixing. (f)5
Image of the disc just prior to opening of Valve 5. This disc is processing 100% Calibrant. (g) Comparison of calibration data acquired on benchtop (n = 3)
and on-disc (n = 1).
The disc was tested on a custom spin stand12
on which the robotic
arm was mounted. During disc rotation, a servo-motor (FESTO,
Germany) periodically generates a trigger pulse.10
This motor was controlled using a CANOpen interface via a
custom LabVIEW (National Instruments, UK) program. The
angular location of the trigger pulse is set using this software
interface, thus permitting visualization of an arbitrary azimuthal
position on the disc.15
As the robotic arm was above the disc, conventional ‘top down’
visualisation resulted in a limited view of the disc. Thus, a mono-
chrome CCD camera (PCO, Germany), synchronised with the
motor and a strobed light source, was mounted at an angle to vi-
sualise both, the movement of liquid as well as the knife cutter.20
The efficacy of the knife cutter was evaluated with (red) dyed
4 | Journal Name, [year], [vol], 00–00 This journal is © The Royal Society of Chemistry [year]
Figure 4: A Lab-on-a-Disc for the biplex liver assay using over-pressure / vacuum suction xurography valves. Chambers with overpressure are
represented by light green and negatively pressurised valves are represented by light red. This disc was tested using dye solutions. Each LUO was
initiated by stopping the disc and manually venting the pneumatic chambers.
water. Initially, the disc (manufactured using DFs) is rotated at5
60 Hz and the valves remained closed (the suction / over-pressure
disc remained sealed up to 20 Hz). Next, the spin rate was
reduced to 5 Hz when the knife cutter was engaged to scour the
top of the disc along the radius of Valve 1. The valve vented
within ~5 s of the blade touching top PSA. Subsequently, the10
blade was retracted and the disc was accelerated up to 30 Hz to
drive the liquid into the metering structures. Next, the disc was
decelerated to 5 Hz and the blade was re-positioned to trigger
Valves 2, 3 and 4 in parallel about 5 s after touchdown of the
blade. The disc was again accelerated to 30 Hz to empty the15
metering structures. This procedure was repeated to trigger Valve
5.
Actuation of the valves was also attempted at higher spin rates
(10 Hz initially), but excessive vibration in the robotic arm as a
consequence of the mechanical drag caused damage to the appa-20
ratus. Yet, the maximum spin rate of xurography actuation might
be elevated through improved mechanical design. Due to this,
implementation of the biplex liver assay on the disc was
implemented by stopping the disc between each Laboratory Unit
Operation4
(LUO) for piercing the upper layer of PSA with a25
scalpel. This method was also used for the overpressure / suction
valves.
2.4 Biplex Liver Assay Panel
To demonstrate timing of flow control with the xurography-actu-
ated flow control scheme, we implemented a multi-step.30
absorbance based biplex liver assay panel comprising of DB
(DB411, Randox, Ireland) and ALP (AP311, Randox, Ireland) as
adapted from that previously described by Nwankire et al. 13
.
While the DB assay completes after 15 minutes, the ALP reaction
must be halted after 3 minutes by addition of ‘stop reagent’35
immediately followed by the absorbance measurement. Along
with processing a blood sample, the assay needs to be calibrated
with known concentrations of DB (0 and 100%) and ALP
(CAL2350, Randox, Ireland). The DB concentration is
determined from two assays, termed ‘active’ and ‘passive’ which40
are read out on a commercial plate reader (TECAN Infinite® 200
PRO).
Thus a disc with nine reservoirs is required to array 0% calibrant,
100% calibrant and the blood sample against active DB, passive
DB and ALP. This is achieved by placing three identical45
structures on a single disc, each of which can prove one sample
against active DB, passive DB and ALP (Fig. 3). Absorbance
data from a disc, used to process 0% calibrant and 100% calibrant
(Fig. 3h).
50
2.5 Xurography Actuated Overpressure / Suction Valves
In an effort to eliminate the integration of the DF, we developed
an alternative method where the liquid is initially retained in the
reservoirs by trapped air. Rather than by the DF, the liquid is
initially either retained by an upstream underpressure or a down-55
stream overpressure. Based on the same xurography-based me-
chanism, pressures are equilibrated with atmosphere by piercing
the top membrane. Figure 4 demonstrates the potential for auto-
mating the same biplex liver assay panel (Section 2.4) with this
alternative approach, thus permitting simpler disc manufacture;60
on the downside, this valving scheme limits the maximum spin
rate of the valves as, at a sufficiently high spin-rate, the suction
back-pressure can be overcome 10
. This can therefore limit the
usefulness of these valves in applications, such as leukocyte
extraction from blood 14
, where high centrifugal forces are65
advantageous.
Similarly, for the over pressure valve, the ‘burst frequency’ of the
unvented chamber, where the liquid will overcome the over
pressure and enter the collection chamber, is dependent on ratio
of the volume of an intermediate chamber (ΔV) to the combined70
volume of the intermediate chamber and the collection chamber
(V). This ratio similarly determines the ‘design’ burst frequency
of centrifugo-pneumatic DF valves 9
and the ‘failure’ frequency
of event-triggered DF valves 7
. In the non DF case, the
intermediate chamber must be relatively large to reflect the size75
of the collection chamber. However, the presence of these large
intermediate chambers effectively increases the dead volume of
liquid trapped within the valves.
3. Conclusions & Outlook
This novel, instrument-actuated flow-control mechanism renders80
the timing of assay protocols independent from both, the spin rate
of the disc and the dissolution time of the DF to permit freely
This journal is © The Royal Society of Chemistry [2013] Lab Chip, 2013, [vol], 00–00 | 5
Figure 5: Concept of the proposed next-generation, xurography driven
platform. This human-powered and thus fully autonomous system will
employ a hand-crank to power the spindle. Pushing down spring-loaded
cutter arms will initiate each LUO. The proposed system might also be5
equipped with a mechanical tachometer to monitor the spin rate.
Alternatively a mechanical brake system might be incorporated to
throttle the (maximum) speed. Particularly applicable for non-
quantitative assays delivering ‘Yes-No’ answers which can be read out
without instrumentation, this fully mechanically operated platform will10
be well-suitable for deployment to extremely resource-poor settings.
programmable timing of automated assay protocols on the Lab-
on-a-Disc platform, even “on the fly”. These enhanced capabili-
ties of the novel valving scheme are demonstrated by processing
a complex, multi-parameter liver assay including a calibrant as15
well as stopping reagents. In this case, the valves were actuated
by stopping the disc and manually piercing the sealing
membrane; this approach was chosen due to the limited
mechanical reliability of the robotic arm. However, it is clear that
with more robust design this valving technology has great20
potential for implementation of automated assays.
It should be noted that, during this study, care was taken during
disc design to place the actuation membranes on different radii
from loading and venting holes. This prevents possible damage to
the cutting blade during rotation. Similarly, it is important to25
ensure that scouring the upper surface of the disc does not impact
on optical clarity should on disc measurement be required.
However, by connecting the actuation membranes to the valves
by long pneumatic channels, these valves provide great flexibility
to locate the actuation membranes (mostly) arbitrary radial30
locations. One potential way to limit these issues might be to
locate the pierceable membranes on the underside of the disc.
This xurography-based flow control scheme has potential to be
particularly powerful for applications in low-resource settings15, 16
including diagnostics associated with emerging issues in35
developing countries17
. While our work shown here uses either a
2-axis robotic arm to actuate the valves, or we stop the disc and
manually pierce the membrane, in the future we propose to
simplify the platform where a number of low-cost, spring loaded
knife-blades are mounted at different disc radii. These blades40
could be actuated by low-cost solenoid actuators. Mounting one
actuator / blade per LUO would reduce the cost and complexity
of this platform. In this approach the disc only requires to be
rotated using a low-cost servo-motor without position or speed
control. Indeed, for making the system fully autonomous, the45
system could even operate in a power-free manual mode by
replacing the motor with a hand-crank while actuating the valves
through applying finger pressure to (appropriately shielded)
spring-loaded blades (Fig. 5). Such an approach will also enable
greater miniaturisation which can make the system more50
appealing for point-of-use applications.
Acknowledgements
This work was supported by the European Union under Grant No,
FP7-KBBE-2013-7-613908-DECATHLON and Science
Foundation Ireland under Grant Nos. 10/CE/B1821, 12/IA/156755
and 12/RC/2289-INSIGHT.
Footnotes
* To whom correspondance should be addressed: david.kinahan@dcu.ie;
jens.ducree@dcu.ie
a
Biomedical Diagnostics Institute, Dublin City University, Glasnevin,60
Dublin 9, Ireland
b
School of Physical Sciences, Dublin City University, Glasnevin, Dublin
9, Ireland
c
School of Mechanical and Manufacturing Engineering, Dublin City
University, Glasnevin, Dublin 9, Ireland65
d
The Insight Centre for Data Analytics, National Centre for Sensor
Research, Dublin City University, Glasnevin, Dublin 9, Ireland
References
1. M. Madou, J. Zoval, G. Jia, H. Kido, J. Kim and N. Kim,
Annu. Rev. Biomed. Eng., 2006, 8, 601-628.70
2. J. Ducrée, S. Haeberle, S. Lutz, S. Pausch, F. V. Stetten and R.
Zengerle, Journal of Micromechanics and Microengineering,
2007, 17, S103-S115.
3. R. Gorkin, J. Park, J. Siegrist, M. Amasia, B. S. Lee, J.-M.
Park, J. Kim, H. Kim, M. Madou and Y.-K. Cho, Lab on a75
Chip, 2010, 10, 1758-1773.
4. O. Strohmeier, M. Keller, F. Schwemmer, S. Zehnle, D. Mark,
F. von Stetten, R. Zengerle and N. Paust, Chemical Society
Reviews, 2015, 44, 6187-6229.
5. J. L. Garcia-Cordero, D. Kurzbuch, F. Benito-Lopez, D.80
Diamond, L. P. Lee and A. J. Ricco, Lab on a Chip, 2010, 10,
2680-2687.
6. Y.-k. Cho, J.-g. Lee, J.-m. Park, B.-s. Lee, Y. Lee and C. Ko,
Lab on a Chip, 2007, 7, 565-565.
7. D. J. Kinahan, S. M. Kearney, N. Dimov, M. T. Glynn and J.85
Ducrée, Lab on a Chip, 2014, 14, 2249-2258.
8. D. J. Kinahan, S. M. Kearney, O. P. Faneuil, M. T. Glynn, N.
Dimov and J. Ducrée, RSC Advances, 2015, 5, 1818-1826.
9. R. Gorkin III, C. E. Nwankire, J. Gaughran, X. Zhang, G. G.
Donohoe, M. Rook, R. O'Kennedy and J. Ducrée, Lab on a90
Chip, 2012, 12, 2894-2902.
10. W. Al-Faqheri, F. Ibrahim, T. H. G. Thio, J. Moebius, K.
Joseph, H. Arof and M. Madou, PloS one, 2013, 8, e58523.
11. D. A. Bartholomeusz, R. W. Boutté and J. D. Andrade,
Microelectromechanical Systems, Journal of, 2005, 14, 1364-95
1374.
12. M. Grumann, T. Brenner, C. Beer, R. Zengerle and J. Ducrée,
Review of Scientific Instruments, 2005, 76, 1-6.
13. C. E. Nwankire, M. Czugala, R. Burger, K. J. Fraser, T.
Glennon, B. E. Onwuliri, I. E. Nduaguibe, D. Diamond and J.100
Ducrée, Biosensors and Bioelectronics, 2014, 56, 352-358.
14. D. J. Kinahan, S. M. Kearney, N. A. Kilcawley, P. L. Early,
M. T. Glynn and J. Ducrée, PloS one, 2016, 11, e0155545.
15. M. T. Glynn, D. J. Kinahan and J. Ducrée, Lab on a Chip,
2014, 14, 2844-2851.105
16. M. T. Glynn, D. J. Kinahan and J. Ducrée, Lab on a Chip,
2013, 13, 2731-2748.
17. B. M. Popkin, L. S. Adair and S. W. Ng, Nutrition reviews,
2012, 70, 3-21.
110

More Related Content

What's hot

Internship Report
Internship ReportInternship Report
Internship ReportMalik Ehsan
 
Well control intro presentation
Well control intro presentationWell control intro presentation
Well control intro presentationamrhaggag
 
Drilling fluids
Drilling fluidsDrilling fluids
Drilling fluidsAAPG SU SC
 
Basic Well Control
Basic Well ControlBasic Well Control
Basic Well ControlM.T.H Group
 
Introduction to manage pressure drilling
Introduction to manage pressure drillingIntroduction to manage pressure drilling
Introduction to manage pressure drillingMubarik Rao
 
Well control and blowout preventer
Well control  and blowout preventerWell control  and blowout preventer
Well control and blowout preventerShamim Akram
 
Well completion workshop april-2018
Well completion workshop april-2018Well completion workshop april-2018
Well completion workshop april-2018mohamed atwa
 

What's hot (12)

Internship Report
Internship ReportInternship Report
Internship Report
 
Well control intro presentation
Well control intro presentationWell control intro presentation
Well control intro presentation
 
Skin
SkinSkin
Skin
 
AADE-10-DF-HO-05
AADE-10-DF-HO-05AADE-10-DF-HO-05
AADE-10-DF-HO-05
 
Dk well control
Dk well controlDk well control
Dk well control
 
Drilling fluids
Drilling fluidsDrilling fluids
Drilling fluids
 
Basic Well Control
Basic Well ControlBasic Well Control
Basic Well Control
 
Introduction to manage pressure drilling
Introduction to manage pressure drillingIntroduction to manage pressure drilling
Introduction to manage pressure drilling
 
Spe 121557-ms
Spe 121557-msSpe 121557-ms
Spe 121557-ms
 
Well control and blowout preventer
Well control  and blowout preventerWell control  and blowout preventer
Well control and blowout preventer
 
Well control system
Well control systemWell control system
Well control system
 
Well completion workshop april-2018
Well completion workshop april-2018Well completion workshop april-2018
Well completion workshop april-2018
 

Viewers also liked

Logo Rijk van Nijmegen 2025
Logo Rijk van Nijmegen 2025Logo Rijk van Nijmegen 2025
Logo Rijk van Nijmegen 2025Rogier Kersten
 
Primary health center by Sachin dwivedi
Primary health center by Sachin dwivediPrimary health center by Sachin dwivedi
Primary health center by Sachin dwivediSachin Dwivedi
 
Gram Games | 2Tons Acceleration Program
Gram Games | 2Tons Acceleration ProgramGram Games | 2Tons Acceleration Program
Gram Games | 2Tons Acceleration ProgramGram Games
 
Colliers International LD CV
Colliers International LD CVColliers International LD CV
Colliers International LD CVLuke Dixon
 
The Female Social Network Case Study - Fresh Produce Social Media
The Female Social Network Case Study - Fresh Produce Social MediaThe Female Social Network Case Study - Fresh Produce Social Media
The Female Social Network Case Study - Fresh Produce Social MediaSteven Bradley
 
механізми підтримки генеруючих потужностей в єс
механізми підтримки генеруючих потужностей в єсмеханізми підтримки генеруючих потужностей в єс
механізми підтримки генеруючих потужностей в єсYuliya Nosulko
 
Miller Greg Portfolio
Miller Greg PortfolioMiller Greg Portfolio
Miller Greg PortfolioShakea Lue
 
Gram Games | GDG DevFest 15 "Releasing Your Unity Game"
Gram Games | GDG DevFest 15 "Releasing Your Unity Game"Gram Games | GDG DevFest 15 "Releasing Your Unity Game"
Gram Games | GDG DevFest 15 "Releasing Your Unity Game"Gram Games
 
Presentation to the City of Cape Town for relaunch of Main Rd and Muizenberg
Presentation to the City of Cape Town for relaunch of Main Rd and MuizenbergPresentation to the City of Cape Town for relaunch of Main Rd and Muizenberg
Presentation to the City of Cape Town for relaunch of Main Rd and MuizenbergPaul Kahanovitz
 
Gamification for Utilities - Energy 2014
Gamification for Utilities - Energy 2014Gamification for Utilities - Energy 2014
Gamification for Utilities - Energy 2014Steven Bradley
 

Viewers also liked (18)

Juan
JuanJuan
Juan
 
Logo Rijk van Nijmegen 2025
Logo Rijk van Nijmegen 2025Logo Rijk van Nijmegen 2025
Logo Rijk van Nijmegen 2025
 
Primary health center by Sachin dwivedi
Primary health center by Sachin dwivediPrimary health center by Sachin dwivedi
Primary health center by Sachin dwivedi
 
nagarkot 2
nagarkot 2nagarkot 2
nagarkot 2
 
Gram Games | 2Tons Acceleration Program
Gram Games | 2Tons Acceleration ProgramGram Games | 2Tons Acceleration Program
Gram Games | 2Tons Acceleration Program
 
Cafe_power system
Cafe_power systemCafe_power system
Cafe_power system
 
Colliers International LD CV
Colliers International LD CVColliers International LD CV
Colliers International LD CV
 
Forever 21
Forever 21Forever 21
Forever 21
 
What is news?
What is news?What is news?
What is news?
 
The Female Social Network Case Study - Fresh Produce Social Media
The Female Social Network Case Study - Fresh Produce Social MediaThe Female Social Network Case Study - Fresh Produce Social Media
The Female Social Network Case Study - Fresh Produce Social Media
 
механізми підтримки генеруючих потужностей в єс
механізми підтримки генеруючих потужностей в єсмеханізми підтримки генеруючих потужностей в єс
механізми підтримки генеруючих потужностей в єс
 
Forget Me Not
Forget Me NotForget Me Not
Forget Me Not
 
Miller Greg Portfolio
Miller Greg PortfolioMiller Greg Portfolio
Miller Greg Portfolio
 
Gram Games | GDG DevFest 15 "Releasing Your Unity Game"
Gram Games | GDG DevFest 15 "Releasing Your Unity Game"Gram Games | GDG DevFest 15 "Releasing Your Unity Game"
Gram Games | GDG DevFest 15 "Releasing Your Unity Game"
 
методичні рекомендації
методичні рекомендаціїметодичні рекомендації
методичні рекомендації
 
Presentation to the City of Cape Town for relaunch of Main Rd and Muizenberg
Presentation to the City of Cape Town for relaunch of Main Rd and MuizenbergPresentation to the City of Cape Town for relaunch of Main Rd and Muizenberg
Presentation to the City of Cape Town for relaunch of Main Rd and Muizenberg
 
2017 01-18 water seminar lecture
2017 01-18 water seminar lecture2017 01-18 water seminar lecture
2017 01-18 water seminar lecture
 
Gamification for Utilities - Energy 2014
Gamification for Utilities - Energy 2014Gamification for Utilities - Energy 2014
Gamification for Utilities - Energy 2014
 

Similar to knife cutter

addae-mensah_loc_c002349c
addae-mensah_loc_c002349caddae-mensah_loc_c002349c
addae-mensah_loc_c002349cMatthew Rendely
 
Spe yp monthly session hydraulic fracturing technology - april 2021
Spe yp monthly session   hydraulic fracturing technology - april 2021Spe yp monthly session   hydraulic fracturing technology - april 2021
Spe yp monthly session hydraulic fracturing technology - april 2021mohamed atwa
 
Создан первый в мире мягкий автономный робот
Создан первый в мире мягкий автономный роботСоздан первый в мире мягкий автономный робот
Создан первый в мире мягкий автономный роботAnatol Alizar
 
The effect of solids on the behaviour of the downcomer of a jameson cell
The effect of solids on the behaviour of the downcomer of a jameson cellThe effect of solids on the behaviour of the downcomer of a jameson cell
The effect of solids on the behaviour of the downcomer of a jameson celleSAT Journals
 
Controlled Formation of Low-Volume Liquid Pillars between Plates with Lattic...
Controlled Formation of Low-Volume Liquid Pillars  between Plates with Lattic...Controlled Formation of Low-Volume Liquid Pillars  between Plates with Lattic...
Controlled Formation of Low-Volume Liquid Pillars between Plates with Lattic...Keiji Takamoto
 
Setzer_Arch-Fairing_Experiment
Setzer_Arch-Fairing_ExperimentSetzer_Arch-Fairing_Experiment
Setzer_Arch-Fairing_ExperimentThomas Setzer
 
Consolidated Drained (CD) Triaxial Test.pdf
Consolidated Drained (CD) Triaxial Test.pdfConsolidated Drained (CD) Triaxial Test.pdf
Consolidated Drained (CD) Triaxial Test.pdfA K
 
Pilot plant testing for hydrocyclone design
Pilot plant testing for hydrocyclone  design Pilot plant testing for hydrocyclone  design
Pilot plant testing for hydrocyclone design reyhane mazahernasab
 
Training materials for students final r1-uf-ro
Training materials for students final r1-uf-roTraining materials for students final r1-uf-ro
Training materials for students final r1-uf-rosembcorpwatertechprize
 
Fast, reconfigurable switch for chemical sensors
Fast, reconfigurable switch for chemical sensorsFast, reconfigurable switch for chemical sensors
Fast, reconfigurable switch for chemical sensorsIowa State University
 
Slug Test Procedures
Slug Test ProceduresSlug Test Procedures
Slug Test Procedurestpttk
 
A Novel Technique to Solve Mathematical Model of Pressure Swing Adsorption Sy...
A Novel Technique to Solve Mathematical Model of Pressure Swing Adsorption Sy...A Novel Technique to Solve Mathematical Model of Pressure Swing Adsorption Sy...
A Novel Technique to Solve Mathematical Model of Pressure Swing Adsorption Sy...ijsrd.com
 
Paper presenting practical explanation of kinetic energy separation technolog...
Paper presenting practical explanation of kinetic energy separation technolog...Paper presenting practical explanation of kinetic energy separation technolog...
Paper presenting practical explanation of kinetic energy separation technolog...Carroll Cobb
 
Part 1 introduction to triaxial testing
Part 1 introduction to triaxial testingPart 1 introduction to triaxial testing
Part 1 introduction to triaxial testingJohn Tran
 

Similar to knife cutter (20)

addae-mensah_loc_c002349c
addae-mensah_loc_c002349caddae-mensah_loc_c002349c
addae-mensah_loc_c002349c
 
Spe 164580-ms
Spe 164580-msSpe 164580-ms
Spe 164580-ms
 
Us8296901
Us8296901Us8296901
Us8296901
 
US8733090
US8733090US8733090
US8733090
 
VPOD
VPODVPOD
VPOD
 
Spe yp monthly session hydraulic fracturing technology - april 2021
Spe yp monthly session   hydraulic fracturing technology - april 2021Spe yp monthly session   hydraulic fracturing technology - april 2021
Spe yp monthly session hydraulic fracturing technology - april 2021
 
2 delugevalve a
2 delugevalve a2 delugevalve a
2 delugevalve a
 
Создан первый в мире мягкий автономный робот
Создан первый в мире мягкий автономный роботСоздан первый в мире мягкий автономный робот
Создан первый в мире мягкий автономный робот
 
01 hydro ripple+clarifier
01 hydro ripple+clarifier01 hydro ripple+clarifier
01 hydro ripple+clarifier
 
The effect of solids on the behaviour of the downcomer of a jameson cell
The effect of solids on the behaviour of the downcomer of a jameson cellThe effect of solids on the behaviour of the downcomer of a jameson cell
The effect of solids on the behaviour of the downcomer of a jameson cell
 
Controlled Formation of Low-Volume Liquid Pillars between Plates with Lattic...
Controlled Formation of Low-Volume Liquid Pillars  between Plates with Lattic...Controlled Formation of Low-Volume Liquid Pillars  between Plates with Lattic...
Controlled Formation of Low-Volume Liquid Pillars between Plates with Lattic...
 
Setzer_Arch-Fairing_Experiment
Setzer_Arch-Fairing_ExperimentSetzer_Arch-Fairing_Experiment
Setzer_Arch-Fairing_Experiment
 
Consolidated Drained (CD) Triaxial Test.pdf
Consolidated Drained (CD) Triaxial Test.pdfConsolidated Drained (CD) Triaxial Test.pdf
Consolidated Drained (CD) Triaxial Test.pdf
 
Pilot plant testing for hydrocyclone design
Pilot plant testing for hydrocyclone  design Pilot plant testing for hydrocyclone  design
Pilot plant testing for hydrocyclone design
 
Training materials for students final r1-uf-ro
Training materials for students final r1-uf-roTraining materials for students final r1-uf-ro
Training materials for students final r1-uf-ro
 
Fast, reconfigurable switch for chemical sensors
Fast, reconfigurable switch for chemical sensorsFast, reconfigurable switch for chemical sensors
Fast, reconfigurable switch for chemical sensors
 
Slug Test Procedures
Slug Test ProceduresSlug Test Procedures
Slug Test Procedures
 
A Novel Technique to Solve Mathematical Model of Pressure Swing Adsorption Sy...
A Novel Technique to Solve Mathematical Model of Pressure Swing Adsorption Sy...A Novel Technique to Solve Mathematical Model of Pressure Swing Adsorption Sy...
A Novel Technique to Solve Mathematical Model of Pressure Swing Adsorption Sy...
 
Paper presenting practical explanation of kinetic energy separation technolog...
Paper presenting practical explanation of kinetic energy separation technolog...Paper presenting practical explanation of kinetic energy separation technolog...
Paper presenting practical explanation of kinetic energy separation technolog...
 
Part 1 introduction to triaxial testing
Part 1 introduction to triaxial testingPart 1 introduction to triaxial testing
Part 1 introduction to triaxial testing
 

knife cutter

  • 1. Lab on a Chip Cite this: DOI: 10.1039/c0xx00000x www.rsc.org/loc Dynamic Article Links ► TECHNICAL INNOVATION This journal is © The Royal Society of Chemistry [year] [journal], [year], [vol], 00–00 | 1 Xurography Actuated Valving For Centrifugal Flow Control David J. Kinahana,b *, Philip L. Earlya,b , Abhishek Vembadia,c , Eoghan MacNamarad , Niamh A. Kilcawleya,b , Thomas Glennond , Dermot Diamondd , Dermot Brabazonc and Jens Ducrée*a,b Received (in XXX, XXX) Xth XXXXXXXXX 20XX, Accepted Xth XXXXXXXXX 20XX DOI: 10.1039/b000000x5 We introduce a novel instrument controlled valving scheme for centrifugal platforms which is based upon xurography. In a first approach, which is akin to previously presented event-triggered flow control, the valves are composed of a pneumatic chamber sealed by a dissolvable film (DF) and by a pierceable membrane. Liquid is initially prevented from wetting the DF by the counter pressure of a trapped gas. Via a channel, this pocket is pneumatically connected to a vent, sealed by the pierceable membrane, located10 on the top surface of the disc. By scouring the top surface of the disc, along a pre-defined track by a robotic knife-cutter, the trapped gas is released and so the liquid can wet and disintegrate the DF. In order to automate assay protocols without the need to integrate DFs, we extend this xurography-based flow control concept by selective venting of chambers subjected to pneumatic over-pressure or vacuum suction. Unlike most instrument controlled flow-control mechanisms, in this approach to valve actuation15 can occur during disc rotation. To demonstrate the potential of this flow control approach, we designed a disc architecture to automate the liquid handling as the backbone of a biplex liver assay panel. We demonstrate valve actuation during rotation, using the robotic arm, using this disc with visualisation via dyed water. We then demonstrate the biplex liver assay, using calibration reagent, by stopping the disc and manually piercing the membrane to actuate the same valves.20 1. Introduction Over the recent decades centrifugal microfluidic systems have increasingly entered the arena of biomedical point-of-care diagnostics 1-3 . In these “Lab-on-a-Disc” (LoaD) systems, a typic- ally disc-shaped cartridge is rotated at defined frequency protocol25 by a robust and low-cost spindle motor. Its simple actuation scheme renders these LoaD devices independent of external pressure sources and their pneumatic connectors, thus ensuring ease-of-use and minimum maintenance. Centrifugal flow control is predominantly implemented by chan-30 ges in the spin rate4 . For instance, in capillary burst valves, the centrifugally induced hydrostatic pressure overcomes the counter- acting capillary pressure imposed by a hydrophobic constriction. Apart from the system-innate spindle motor, flow control can also be achieved by additional, instrument-based modules, for exam-35 ple by laser ablating a sacrificial membrane5 or by heating a phase-change material 6 . Recently, Kinahan et al. introduced event-triggered flow control7, 8 based on dissolvable-film (DF) membranes9 . These valves are composed of a dead-end pneumatic chamber sealed by two DFs.40 Upon removal of the so-called control film (CF) through contact with a secondary liquid, the pneumatic chamber is vented to let the primary liquid escape by dissolving the so-called the load film (LF). By cascading such centrifugo-pneumatic DF valves, ‘event- triggered’ flow control has been implemented, e.g. for automating45 an unprecedented number of individual liquid handling steps in a single disc at an essentially constant spin rate. However, a major drawback of this event-triggered valving is its poor flexibility of timing which depends on the dissolution time of the DFs and time for liquid to move about the disc. Particularly, these timing issues50 can limit the implementation of biomedical assays requiring extended incubation. In order to allow discretionary timing, we introduce here an instr- ument-controlled scheme which is effectively an actuated centri- fugo-pneumatic DF valve. In this case the CF is replaced by a55 vent, sealed by a pierceable adhesive tape, which is located on the top surface of the disc (Fig. 1). A blade, adapted from a commercial knife cutter, is mounted on a 2-axis robotic arm and positioned above the disc (Fig. 2). During rotation, the blade is lowered to perforate the adhesive tape and thus venting the60 pneumatic chamber to atmosphere. Only films located along the same orbit are actuated (quasi) in parallel, so the order of radial positions where the knife cutter touches the disc defines the sequence of valving. Furthermore, we also demonstrate how similar flow control can65 be achieved by using pneumatic overpressure or vacuum suction 10 . While this technique works without incorporation of DFs, it comes at the expense of limiting the maximum spin rate. As a proof-of-concept, we demonstrate the automation of multi- liquid, multi-step sample preparation and reagent handling70 protocol towards a biplex liver assay panel. Critically for on-disc blood processing, the valves can be externally actuated during
  • 2. 2 | Journal Name, [year], [vol], 00–00 This journal is © The Royal Society of Chemistry [year] Figure 1: Operation of the xurography-enabled DF valve. (a) The valve consists of a pneumatic chamber with two outlets closed by a DF tab and PSA. (b) The gas pocket trapped between the meniscus of the incoming liquid and these outlets initially prevents the wetting of the DF. (c) To open the valve, a knife blade mounted on a robotic arm (cf. Fig. 2) scours the top of the disc, and thereby piercing the PSA. (d) As the gas pocket decompresses, the DF is wetted and dissolves to clear the passage of liquid to the outlet.5 rotation, thus ensuring efficient plasma extraction. This highly controllable valving scheme, where liquid handling steps at any location on the disc can be controlled through interaction with its external surface, represents a viable path towards large-scale integration (LSI) and automation of multiplex bio-assays.10 2. Experimental 2.1 Manufacture of Robotic Arm The robotic arm was composed of two linear precision rails (Reliance Precision Ltd, UK). It was mounted in an XZ-axis configuration on two pillar supports. The pillars were15 manufactured from chemically bonded laser-cut sheets of poly(methyl-methacrylate) (PMMA). Both axes were powered using stepper actuators (Firgelli Technologies Inc.). The stepper motor on the X-axis used an integrated potentiometer to provide position feedback to custom program (LabVIEW, National In-20 struments) via a data acquisition card (National Instruments). This stepper motor could also accurately position the knife cutter assembly, which is mounted on the x-axis, to an accuracy of 0.5mm. Along with the Z-axis rail, the knife-cutter assembly was composed of a backing plate and a stepper actuator which had a25 stroke of 10 mm without positional feedback. Therefore, the vertical actuator was connected to the blade holder via a spring- damped connection. In its retracted state, the blade did not contact the top surface of the disc. In the extended state, the blade could scour the top surface of the disc. Thus, tolerances in disc30 thickness and mounting were compensated for by the spring damping. The knife blade was adapted from a commercial knife- cutter (Graphtec, Japan). 2.2 Disc Manufacture The discs are manufactured using multilayer polymer fabrication35 methods using methods previously described 7, 11 . The over- pressure / vacuum-suction valves were manufacture using disc which were assembled from five layers. Laser-cut PMMA sheets were used for the upper layer, the reservoir layer and the base layer. Vents and loading holes were created from voids in the40 upper layer and reservoirs were created from voids in the base layer. Connecting microchannels were created on the upper and lower pressure sensitive adhesive (PSA) (Adhesives Research, Limerick, Ireland) from voids created by a knife-cutter. The stacking of the two PSA layers permitted microchannels cros-45 sings. The discs for demonstrating xurography actuated DF valves were manufactured from four PMMA and four PSA layers as previously described 7 in order to incorporate the DF tabs. Finally the vent layer of the disc was covered with a transparent film (Adhesives Research) to seal the disc. Loading holes and50 appropriate vents were then pierced by a scalpel. 2.3 Xurography Actuated DF Valves The disc implements a biplex liver assay test (see Section 2.4) starting with plasma extraction. Next, 10-µl volumes of plasma are metered and then transferred to the periphery of the disc55 where they mix with the assay reagents alkaline phosphatase (ALP), active direct bilirubin (DB) and passive DB. While the DB assays involve simpler end-point measurements, the ALP re- quires the addition of a stop reagent after a defined reaction interval. Assay automation thus requires five sequentially actu-60 ated valves. Figure 2: Robotic arm. (a) 3D rendering. (b) The robotic arm mounted on spin stand. The liver assay disc (cf. Fig. 3) is mounted on the experimental rig. (c) Monochrome image acquired while the disc is spinning at 50 Hz. The track of knife cutter can be seen scratching the PSA layer to vent and open Valve 1.65
  • 3. This journal is © The Royal Society of Chemistry [2013] Lab Chip, 2013, [vol], 00–00 | 3 Figure 3: The Lab-on-a-Disc for a biplex liver assay panel measuring Direct Bilirubin (DB) and Alkaline Phosphatase (ALP). (a) The disc displays three se- parate structures which are operated in parallel. (b-e) Schematic of blood processing on the disc, (b) sample centrifugation, (c) opening of Valve 1 and metering of plasma, (d) opening of Valve 2, 3 and 4, and transfer of metered sample (10 µl) to mix with reagents. (e) After 3 minutes, the knife cutter opens Valve 5 to release the reagent stopping the ALP assay. Note that, at all times, the disc is subject to Euler-force based “shake-mode” mixing. (f)5 Image of the disc just prior to opening of Valve 5. This disc is processing 100% Calibrant. (g) Comparison of calibration data acquired on benchtop (n = 3) and on-disc (n = 1). The disc was tested on a custom spin stand12 on which the robotic arm was mounted. During disc rotation, a servo-motor (FESTO, Germany) periodically generates a trigger pulse.10 This motor was controlled using a CANOpen interface via a custom LabVIEW (National Instruments, UK) program. The angular location of the trigger pulse is set using this software interface, thus permitting visualization of an arbitrary azimuthal position on the disc.15 As the robotic arm was above the disc, conventional ‘top down’ visualisation resulted in a limited view of the disc. Thus, a mono- chrome CCD camera (PCO, Germany), synchronised with the motor and a strobed light source, was mounted at an angle to vi- sualise both, the movement of liquid as well as the knife cutter.20 The efficacy of the knife cutter was evaluated with (red) dyed
  • 4. 4 | Journal Name, [year], [vol], 00–00 This journal is © The Royal Society of Chemistry [year] Figure 4: A Lab-on-a-Disc for the biplex liver assay using over-pressure / vacuum suction xurography valves. Chambers with overpressure are represented by light green and negatively pressurised valves are represented by light red. This disc was tested using dye solutions. Each LUO was initiated by stopping the disc and manually venting the pneumatic chambers. water. Initially, the disc (manufactured using DFs) is rotated at5 60 Hz and the valves remained closed (the suction / over-pressure disc remained sealed up to 20 Hz). Next, the spin rate was reduced to 5 Hz when the knife cutter was engaged to scour the top of the disc along the radius of Valve 1. The valve vented within ~5 s of the blade touching top PSA. Subsequently, the10 blade was retracted and the disc was accelerated up to 30 Hz to drive the liquid into the metering structures. Next, the disc was decelerated to 5 Hz and the blade was re-positioned to trigger Valves 2, 3 and 4 in parallel about 5 s after touchdown of the blade. The disc was again accelerated to 30 Hz to empty the15 metering structures. This procedure was repeated to trigger Valve 5. Actuation of the valves was also attempted at higher spin rates (10 Hz initially), but excessive vibration in the robotic arm as a consequence of the mechanical drag caused damage to the appa-20 ratus. Yet, the maximum spin rate of xurography actuation might be elevated through improved mechanical design. Due to this, implementation of the biplex liver assay on the disc was implemented by stopping the disc between each Laboratory Unit Operation4 (LUO) for piercing the upper layer of PSA with a25 scalpel. This method was also used for the overpressure / suction valves. 2.4 Biplex Liver Assay Panel To demonstrate timing of flow control with the xurography-actu- ated flow control scheme, we implemented a multi-step.30 absorbance based biplex liver assay panel comprising of DB (DB411, Randox, Ireland) and ALP (AP311, Randox, Ireland) as adapted from that previously described by Nwankire et al. 13 . While the DB assay completes after 15 minutes, the ALP reaction must be halted after 3 minutes by addition of ‘stop reagent’35 immediately followed by the absorbance measurement. Along with processing a blood sample, the assay needs to be calibrated with known concentrations of DB (0 and 100%) and ALP (CAL2350, Randox, Ireland). The DB concentration is determined from two assays, termed ‘active’ and ‘passive’ which40 are read out on a commercial plate reader (TECAN Infinite® 200 PRO). Thus a disc with nine reservoirs is required to array 0% calibrant, 100% calibrant and the blood sample against active DB, passive DB and ALP. This is achieved by placing three identical45 structures on a single disc, each of which can prove one sample against active DB, passive DB and ALP (Fig. 3). Absorbance data from a disc, used to process 0% calibrant and 100% calibrant (Fig. 3h). 50 2.5 Xurography Actuated Overpressure / Suction Valves In an effort to eliminate the integration of the DF, we developed an alternative method where the liquid is initially retained in the reservoirs by trapped air. Rather than by the DF, the liquid is initially either retained by an upstream underpressure or a down-55 stream overpressure. Based on the same xurography-based me- chanism, pressures are equilibrated with atmosphere by piercing the top membrane. Figure 4 demonstrates the potential for auto- mating the same biplex liver assay panel (Section 2.4) with this alternative approach, thus permitting simpler disc manufacture;60 on the downside, this valving scheme limits the maximum spin rate of the valves as, at a sufficiently high spin-rate, the suction back-pressure can be overcome 10 . This can therefore limit the usefulness of these valves in applications, such as leukocyte extraction from blood 14 , where high centrifugal forces are65 advantageous. Similarly, for the over pressure valve, the ‘burst frequency’ of the unvented chamber, where the liquid will overcome the over pressure and enter the collection chamber, is dependent on ratio of the volume of an intermediate chamber (ΔV) to the combined70 volume of the intermediate chamber and the collection chamber (V). This ratio similarly determines the ‘design’ burst frequency of centrifugo-pneumatic DF valves 9 and the ‘failure’ frequency of event-triggered DF valves 7 . In the non DF case, the intermediate chamber must be relatively large to reflect the size75 of the collection chamber. However, the presence of these large intermediate chambers effectively increases the dead volume of liquid trapped within the valves. 3. Conclusions & Outlook This novel, instrument-actuated flow-control mechanism renders80 the timing of assay protocols independent from both, the spin rate of the disc and the dissolution time of the DF to permit freely
  • 5. This journal is © The Royal Society of Chemistry [2013] Lab Chip, 2013, [vol], 00–00 | 5 Figure 5: Concept of the proposed next-generation, xurography driven platform. This human-powered and thus fully autonomous system will employ a hand-crank to power the spindle. Pushing down spring-loaded cutter arms will initiate each LUO. The proposed system might also be5 equipped with a mechanical tachometer to monitor the spin rate. Alternatively a mechanical brake system might be incorporated to throttle the (maximum) speed. Particularly applicable for non- quantitative assays delivering ‘Yes-No’ answers which can be read out without instrumentation, this fully mechanically operated platform will10 be well-suitable for deployment to extremely resource-poor settings. programmable timing of automated assay protocols on the Lab- on-a-Disc platform, even “on the fly”. These enhanced capabili- ties of the novel valving scheme are demonstrated by processing a complex, multi-parameter liver assay including a calibrant as15 well as stopping reagents. In this case, the valves were actuated by stopping the disc and manually piercing the sealing membrane; this approach was chosen due to the limited mechanical reliability of the robotic arm. However, it is clear that with more robust design this valving technology has great20 potential for implementation of automated assays. It should be noted that, during this study, care was taken during disc design to place the actuation membranes on different radii from loading and venting holes. This prevents possible damage to the cutting blade during rotation. Similarly, it is important to25 ensure that scouring the upper surface of the disc does not impact on optical clarity should on disc measurement be required. However, by connecting the actuation membranes to the valves by long pneumatic channels, these valves provide great flexibility to locate the actuation membranes (mostly) arbitrary radial30 locations. One potential way to limit these issues might be to locate the pierceable membranes on the underside of the disc. This xurography-based flow control scheme has potential to be particularly powerful for applications in low-resource settings15, 16 including diagnostics associated with emerging issues in35 developing countries17 . While our work shown here uses either a 2-axis robotic arm to actuate the valves, or we stop the disc and manually pierce the membrane, in the future we propose to simplify the platform where a number of low-cost, spring loaded knife-blades are mounted at different disc radii. These blades40 could be actuated by low-cost solenoid actuators. Mounting one actuator / blade per LUO would reduce the cost and complexity of this platform. In this approach the disc only requires to be rotated using a low-cost servo-motor without position or speed control. Indeed, for making the system fully autonomous, the45 system could even operate in a power-free manual mode by replacing the motor with a hand-crank while actuating the valves through applying finger pressure to (appropriately shielded) spring-loaded blades (Fig. 5). Such an approach will also enable greater miniaturisation which can make the system more50 appealing for point-of-use applications. Acknowledgements This work was supported by the European Union under Grant No, FP7-KBBE-2013-7-613908-DECATHLON and Science Foundation Ireland under Grant Nos. 10/CE/B1821, 12/IA/156755 and 12/RC/2289-INSIGHT. Footnotes * To whom correspondance should be addressed: david.kinahan@dcu.ie; jens.ducree@dcu.ie a Biomedical Diagnostics Institute, Dublin City University, Glasnevin,60 Dublin 9, Ireland b School of Physical Sciences, Dublin City University, Glasnevin, Dublin 9, Ireland c School of Mechanical and Manufacturing Engineering, Dublin City University, Glasnevin, Dublin 9, Ireland65 d The Insight Centre for Data Analytics, National Centre for Sensor Research, Dublin City University, Glasnevin, Dublin 9, Ireland References 1. M. Madou, J. Zoval, G. Jia, H. Kido, J. Kim and N. Kim, Annu. Rev. Biomed. Eng., 2006, 8, 601-628.70 2. J. Ducrée, S. Haeberle, S. Lutz, S. Pausch, F. V. Stetten and R. Zengerle, Journal of Micromechanics and Microengineering, 2007, 17, S103-S115. 3. R. Gorkin, J. Park, J. Siegrist, M. Amasia, B. S. Lee, J.-M. Park, J. Kim, H. Kim, M. Madou and Y.-K. Cho, Lab on a75 Chip, 2010, 10, 1758-1773. 4. O. Strohmeier, M. Keller, F. Schwemmer, S. Zehnle, D. Mark, F. von Stetten, R. Zengerle and N. Paust, Chemical Society Reviews, 2015, 44, 6187-6229. 5. J. L. Garcia-Cordero, D. Kurzbuch, F. Benito-Lopez, D.80 Diamond, L. P. Lee and A. J. Ricco, Lab on a Chip, 2010, 10, 2680-2687. 6. Y.-k. Cho, J.-g. Lee, J.-m. Park, B.-s. Lee, Y. Lee and C. Ko, Lab on a Chip, 2007, 7, 565-565. 7. D. J. Kinahan, S. M. Kearney, N. Dimov, M. T. Glynn and J.85 Ducrée, Lab on a Chip, 2014, 14, 2249-2258. 8. D. J. Kinahan, S. M. Kearney, O. P. Faneuil, M. T. Glynn, N. Dimov and J. Ducrée, RSC Advances, 2015, 5, 1818-1826. 9. R. Gorkin III, C. E. Nwankire, J. Gaughran, X. Zhang, G. G. Donohoe, M. Rook, R. O'Kennedy and J. Ducrée, Lab on a90 Chip, 2012, 12, 2894-2902. 10. W. Al-Faqheri, F. Ibrahim, T. H. G. Thio, J. Moebius, K. Joseph, H. Arof and M. Madou, PloS one, 2013, 8, e58523. 11. D. A. Bartholomeusz, R. W. Boutté and J. D. Andrade, Microelectromechanical Systems, Journal of, 2005, 14, 1364-95 1374. 12. M. Grumann, T. Brenner, C. Beer, R. Zengerle and J. Ducrée, Review of Scientific Instruments, 2005, 76, 1-6. 13. C. E. Nwankire, M. Czugala, R. Burger, K. J. Fraser, T. Glennon, B. E. Onwuliri, I. E. Nduaguibe, D. Diamond and J.100 Ducrée, Biosensors and Bioelectronics, 2014, 56, 352-358. 14. D. J. Kinahan, S. M. Kearney, N. A. Kilcawley, P. L. Early, M. T. Glynn and J. Ducrée, PloS one, 2016, 11, e0155545. 15. M. T. Glynn, D. J. Kinahan and J. Ducrée, Lab on a Chip, 2014, 14, 2844-2851.105 16. M. T. Glynn, D. J. Kinahan and J. Ducrée, Lab on a Chip, 2013, 13, 2731-2748. 17. B. M. Popkin, L. S. Adair and S. W. Ng, Nutrition reviews, 2012, 70, 3-21. 110