BASICS OF ECHOCARDIOGRAPHYAND PRINCIPLES OF DOPPLER ECHO Abraha Hailu August 29, 2010
Topic  outline
1. BASICS ultrasound (1923)  vibrations of the same physical nature as sound but with frequencies above the range of human hearing, namely frequencies greater than 20,000 cycles per second. the diagnostic or therapeutic use of ultrasound and esp. a noninvasive technique involving the formation of a two-dimensional image used for the examination and measurement of internal body structures and the detection of bodily abnormalities a diagnostic examination using ultrasound
Medical ultrasound imaging typically uses sound waves at frequencies of 1,000,000 to 20,000,000 Hz (1.0 to 20 MHz). In contrast, the human auditory spectrum (between 20 and 20,000 Hz) Frequency and wavelength are mathematically related to the velocity of the ultrasound beam within the tissue:                            Velocity =  Wavelength (mm)  x  frequency (Hz) The speed with which an acoustic wave moves through a medium is dependent upon the density and resistance of the medium.  ,[object Object],The resolution of a recording, ie, the ability to distinguish two objects that are spatially close together, varies directly with the frequency and inversely with the wavelength High frequency, short wavelength ultrasound can separate objects that are less than 1 mm apart.
Imaging with higher frequency (and lower wavelength) transducers permits enhanced spatial resolution ,[object Object]
As a result, the trade-off for use of higher frequency transducers is reduced tissue penetrationThe trade-off between tissue resolution and penetration guides the choice of transducer frequency for clinical imaging.  ,[object Object],[object Object]
Unlike a reflected beam, scattering results in the US beam being radiated in all directions, with minimal signal returning to the transducerRefraction:  Attenuation: ,[object Object]
The depth of penetration:
30 cm for a 1 MHz transducer,
12 cm for 2.5 MHz transducer, and
6 cm for a 5 MHz transducer
Air has a very high acoustic impedance, resulting in significant signal attenuation when imaging through lung tissue, especially emphysematous lung, or pathologic conditions such as pneumomediastinum or subcutaneous emphysema
In contrast, filling of the pleural space with fluid, generally enhances ultrasound imaging,[object Object]
ULTRASOUND TRANSDUCERS  US transducers use a piezoelectric crystal to generate and receive ultrasound wavesImage formation: is related to the distance of a structure from the transducer,  based upon the time interval between ultrasound transmission and arrival of the reflected signalThe amplitude is proportional to the incident angle and acoustic impedance, and timing is proportional to the distance from the transducer
SECOND HARMONIC IMAGING(improving resolution) An ultrasound wave traveling through tissue becomes distorted, which generates additional sound frequencies that are harmonics of the original or fundamental frequency ,[object Object]
uses broadband transducers that receive double the transmitted frequencyWhen compared to conventional imaging, it reduces variations in ultrasound intensity along endocardial and myocardial surfaces, enhancing these structures  of particular benefit for patients in whom optimal echocardiographic images are technically difficult to obtain ,[object Object],[object Object]
2. IMAGING MODALITIES Two dimensional (2-D) imaging : A 2D image is generated from data obtained mechanically (mechanical transducer) or electronically (phased-array transducer) The signal received undergoes a complex manipulation to form the final image displayed on the monitor including signal amplification, time-gain compensation, filtering, compression and rectification. M-mode: Motion or "M"-mode echocardiography is among the earliest forms of cardiac ultrasound  The very high temporal resolution by M-mode imaging permits: identification of subtle abnormalities such as fluttering of the anterior mitral leaflet due to aortic insufficiency or movement of a vegetation.  dimensional measurements or changes, such as chamber size and endocardial thickening, can be readily appreciated
A.  2-D  ECHOCARDIOGRAPHY
OPTIMIZATION OF 2-D IMAGESTechnical Factors I TRANSDUCER: ,[object Object]
Low-frequency transducers permit good penetration but reduced image resolutionDEPTH: ,[object Object]
The smallest depth that permits display of the region of interest should be employedFOCUS: ,[object Object]
Resolution is greatest in this regionGAIN: ,[object Object],[object Object]
IMAGING PLANES The long-axis plane is the plane perpendicular to the posterior and anterior surfaces of the body and parallel to the long axis of the heart
Sweeping begins at the base of the heart which appears on the rt of the screenThe left atrium, the mitral valve and the right ventricular outflow tract are seen.
Parasternal WindowRight Long-Axis ViewSweeping begins at the right atrium which is on the right of the screen. In this view we can see the right atrium and the right ventricle, with the tricuspid valve in between
The short-axis plane is the plane that runs parallel to the posterior and anterior surfaces of the body and transects the heart from its apex to its base encompassing all four cardiac chambers
                       Parasternal Short-Axis ViewsObtained from the parasternal window by rotating the transducer clockwise by 90ºHence, the image index marker is pointed toward the patient´s right supraclavicularfossa. A series of sweeps transect the heart from the base to the apex by changes in transducer position and angulation. There are three characteristic levels: 1) great vessels, 2) mitral valve, 3) ventricles.
Parasternal Short-Axis ViewGreat Arteries LevelSweeping begins at the left edge of the atrium which appears on the right of the screen
At aortic valve level it demonstrates all three aortic valve leaflets. The pulmonary valve, the right ventricular outflow tract, the right atrium and the left atrium are seen in this view.
Parasternal Short-Axis ViewGreat Arteries LevelIn this view we can see the pulmonary valve (PV) and pulmonary artery (PA) with right (RPA) and left branches (LPA).
Parasternal Short-Axis ViewMitral Valve LevelThe anterior and posterior mitral leaflets are seen as they open in diastole and close in systole
Parasternal Short-Axis ViewMitral Valve Level
The four-chamber plane is the plane that runs parallel to the posterior and anterior surfaces of the body and transects the heart from its apex to its base encompassing all four cardiac chambers
Apical WindowSweeping begins at the left edge of the heart which appears on the right of the screen.
4-Chamber ViewIn this view we can see the left ventricle, the right ventricle, the left atrium and the right atrium. The tricuspid annulus lies slightly (up to 10 mm) closer to the apex than the mitral annulus. The septal and posterior tricuspid leaflets are seen in this view.
Apical Window5-Chamber ViewBy tilting the transducer anteriorly, the aortic root is seen in an oblique long view.
Apical Window2-Chamber ViewSweeping begins at the anterior face of the left ventricle which appears on the right of the screen. From the four-chamber view, the transducer is rotated anti-clockwise by 60º to obtain the two-chamber view of the left ventricle, the mitral valve and the left atrium.
Apical Window2-Chamber ViewThis view shows the LV, the LA and the MV. The LV shows the inferior wall on the left and the anterior wall on the right.
Subcostal Window4-Chamber ViewSweeping starts at the apex which appears on the right of the screen. A view of all four chambers shows the right ventricular free wall, the midsection of the interventricular septum and the posterolateral left ventricular wall. In this view, the interatrial septum is perpendicular to the direction of the ultrasound beam
Subcostal WindowThe inferior vena cava as it enters the right atrium and the central hepatic vein are seen.
Suprasternal Notch WindowSweeping begins at the descending aorta which appears on the right of the screen. The long-axis view shows the ascending aorta, the arch, the proximal descending aorta and the origins of the right brachiocephalic and left common carotid and subclavian arteries.
Suprasternal Notch Window
B.  M-MODE ECHOCARDIOGRAPHY a single line of sight is included, the repetition frequency of the pulse transmission is very high and sampling rates of around 1800 cycles/sec are usedContinuously-moving structures may be identified more accurately when motion versus time, as well as depth, is displayed clearly on the M-mode recording
AORTIC VALVE AND LEFT ATRIUM
AORTIC VALVE AND LEFT ATRIUM The aortic root is moving anteriorly in systole and posteriorly in diastole. The left atrium is posterior to the aortic root.  The aortic leaflet coaptation point is seen as a thin line in diastole.  AO is measured in end-diastole and LA in end-systole.
In early diastole, the leaflets separate widely, with the maximum early diastolic motion of the anterior leaflet termed the E point. The leaflets move together in mid-diastole and then separate again with atrial systole, the A point. Closure at the end of diastole is termed C point.
M-mode recording perpendicular to the long axis of and through the centre of the left ventricle at papillary muscle level provides standard measurements of systolic and diastolic thickness and chamber dimensions:
Normal Values Left ventricular end-diastole:  37 - 57 mm (23 -31 mm/m²) Left ventricular end-systole:  21 - 40 mm (14 -21 mm/m²) Interventricular septum:  7 - 11 mm Posterior wall:  7 - 11 mm
3. DOPPLER ECHOCARDIOGRAPHY BASIC PRINCIPLES: utilizes ultrasound to record blood flow within the cardiovascular system (While M-mode and 2D echo create ultrasonic images of the heart) is based upon the changes in frequency of the backscatter signal from small moving structures, ie, red blood cells, intercepted by the ultrasound beam
A moving target will backscatter an ultrasound beam to the transducer so that the frequency observed when the target is moving toward the transducer is higher and the frequency observed when the target is moving away from the transducer is lower than the original transmitter frequency ,[object Object]
This difference in frequency between the transmitted frequency (F[t]) and received frequency (F[r]) is the Doppler shift:
  Doppler shift (F[d]) = F[r] - F[t],[object Object]
  With a stationary target (panel A): the carrier frequency [f(t)] from the transmitting transducer strikes the target and is reflected back to the receiving transducer at the reflected frequency [f(r)], which is unaltered
  with a target moving toward the transducer (panel B):  An increase in f(r) is seen
  with a target moving away from the transducer (panel C): f(r) is reduced
In all cases, the extent to which f(t) is increased or reduced is proportional to the velocity of the target,[object Object]
Blood flow velocity (V) is related to the Doppler shift by the speed of sound in blood (C) and ø (the intercept angle between the ultrasound beam and the direction of blood flow) ,[object Object]
  F[d] = 2 x F[t] x [(Vx cos ø)] ÷ CThis equation can be solved for V, by substituting (F[r] - F[t]) for F[d]: ,[object Object],[object Object]
For ø of 90º (perpendicular to blood flow), cosine ø = 0 and the Doppler shift is 0
For ø up to 20º, cos ø results in a minimal (<10 percent) change in the Doppler shift
For ø of 60º, cosine ø = 0.50 The value of ø is particularly important for accurate assessment of high velocity jets, which occur in aortic stenosis or pulmonary artery hypertension It is generally assumed that ø is 0º and cos ø is therefore 1 ,[object Object]
When the beam does not lie parallel, it is possible to introduce a correction into the calculation of flow velocity by measuring the cosine of the angle of interrogation and introducing this value into the Doppler equation,[object Object]
By convention, time is displayed on the x axis and frequency shift on the y axis
Shifts toward the transducer are represented as "positive" deflections from the "zero" baseline, and shifts away from the transducer are displayed as "negative" deflections,[object Object],[object Object]
Right Ventricular InflowCan be recorded from an apical approach. The pattern is similar to mitral flow.
Left Ventricular OutflowAn apical window is used with a pulsed Doppler sample volume positioned on the left ventricular side of the aortic valve. Note the narrow band of the systolic velocities.
Right Ventricular OutflowFrom a parasternal short-axis view, the sample volume is located in the right ventricular outflow tract. The Doppler shape is similar to the left ventricular outflow curve
Suprahepatic Vein FlowFrom the longitudinal subcostal view, the sample volume is located at the main suprahepatic vein. There are two positive waves, (A) and (V), and two negative ones, (X) and (Y).
DOPPLER MODALITIES  Doppler methods used for cardiac evaluation : continuous wave doppler Pulsed  wave doppler color flow doppler
CONTINUOUS WAVE DOPPLER employs two dedicated ultrasound crystals, one for continuous transmission and a second for continuous reception ,[object Object],Limitations of this technique:  ,[object Object]
Thus, there may be overlap in certain settings, such as:
stenoses in series (eg, left ventricular outflow tract gradient and aortic stenosis) or
flows that are in close proximity/alignment (eg, AS and MR),[object Object]
An ideal Doppler profile is one with a smooth "outer" contour, well-defined edge and maximum velocity, and abrupt onset and termination
PULSED DOPPLER  permits sampling of blood flow velocities from a specific region  ,[object Object],is always performed with 2D guidance to determine the sample volume position Particularly useful for assessing the relatively low velocity flows associated with: transmitral or transtricuspid blood flow,  pulmonary venous flow,  left atrial appendage flow, or  for confirming the location of eccentric jets of aortic insufficiency or mitral regurgitation
[object Object],[object Object]
Because PWD repeatedly samples the returning signal, there is a maximum limit to the frequency shift that can be measured unambiguouslyThe maximum detectable frequency shift (the Nyquist limit) is one-half the PRFIf velocity exceeds the Nyquist limit, signal aliasing is seen with the signal cut off at the edge of the display and the top of the waveform appearing in the reverse partHigh-PRF increases the number of sample volumes
COLOR FLOW IMAGING ,[object Object]
with flow toward the transducer displayed in orange/red
flow away from the transducer displayed as blue,[object Object]
apical four chamber view with color flow Doppler during diastole This color signal is used to position a pulsed wave Doppler sample volume so that quantitatable signals of flow can be obtained from the pulmonary veins and from the mitral leaflet tips

More Related Content

What's hot

What's hot (20)

Stress echocardiography
Stress echocardiographyStress echocardiography
Stress echocardiography
 
TEE VIEWS
TEE VIEWSTEE VIEWS
TEE VIEWS
 
Electrophysiology study basics
Electrophysiology study basicsElectrophysiology study basics
Electrophysiology study basics
 
Echo assessment of RV function
Echo assessment of RV functionEcho assessment of RV function
Echo assessment of RV function
 
Asd echo assessment
Asd echo assessmentAsd echo assessment
Asd echo assessment
 
2D ECHO Basics
2D ECHO Basics2D ECHO Basics
2D ECHO Basics
 
Contrast Echocardiography
Contrast EchocardiographyContrast Echocardiography
Contrast Echocardiography
 
Transesophageal echocardiography
Transesophageal echocardiographyTransesophageal echocardiography
Transesophageal echocardiography
 
Left ventricular angiogram (1)
Left ventricular angiogram (1)Left ventricular angiogram (1)
Left ventricular angiogram (1)
 
Echo Mitral Stenosis
Echo Mitral StenosisEcho Mitral Stenosis
Echo Mitral Stenosis
 
M-mode echo-1-1.pptx
M-mode echo-1-1.pptxM-mode echo-1-1.pptx
M-mode echo-1-1.pptx
 
Intravascular Ultrasound (IVUS)
Intravascular Ultrasound (IVUS)Intravascular Ultrasound (IVUS)
Intravascular Ultrasound (IVUS)
 
Diastolic Dysfunction 2016
Diastolic Dysfunction 2016Diastolic Dysfunction 2016
Diastolic Dysfunction 2016
 
Echo assessment of cardiomyopathy pdf
Echo assessment of cardiomyopathy pdfEcho assessment of cardiomyopathy pdf
Echo assessment of cardiomyopathy pdf
 
Echocardiography of CHD in Adults
Echocardiography of CHD in AdultsEchocardiography of CHD in Adults
Echocardiography of CHD in Adults
 
How to echo... tricuspid regurgitation.ppt
How to echo... tricuspid regurgitation.pptHow to echo... tricuspid regurgitation.ppt
How to echo... tricuspid regurgitation.ppt
 
Echo views
Echo viewsEcho views
Echo views
 
Lv systolic function
Lv systolic functionLv systolic function
Lv systolic function
 
Coronary angiography
Coronary angiographyCoronary angiography
Coronary angiography
 
Echo assessment of mitral regurgitation
Echo assessment of mitral regurgitationEcho assessment of mitral regurgitation
Echo assessment of mitral regurgitation
 

Viewers also liked (12)

Introduction to echocardiography
Introduction to echocardiography Introduction to echocardiography
Introduction to echocardiography
 
Echocardiogram Basics
Echocardiogram BasicsEchocardiogram Basics
Echocardiogram Basics
 
Echocardiogram - Basics
Echocardiogram - BasicsEchocardiogram - Basics
Echocardiogram - Basics
 
Ventricular Septal Defects - A Review
Ventricular Septal Defects - A ReviewVentricular Septal Defects - A Review
Ventricular Septal Defects - A Review
 
Ion channelopathy
Ion channelopathyIon channelopathy
Ion channelopathy
 
Acs anticoagulation
Acs anticoagulationAcs anticoagulation
Acs anticoagulation
 
Left Atrial Thrombus - A Review
Left Atrial Thrombus - A ReviewLeft Atrial Thrombus - A Review
Left Atrial Thrombus - A Review
 
Svt evaluation
Svt evaluationSvt evaluation
Svt evaluation
 
Tetrology of Fallot (TOF) - A Review
Tetrology of Fallot (TOF) - A ReviewTetrology of Fallot (TOF) - A Review
Tetrology of Fallot (TOF) - A Review
 
Heart sounds s1
Heart sounds s1Heart sounds s1
Heart sounds s1
 
PCI vs OMT vs CABG in Stable CAD
PCI vs OMT vs CABG in Stable CADPCI vs OMT vs CABG in Stable CAD
PCI vs OMT vs CABG in Stable CAD
 
Approach to pituitary tumours
Approach to pituitary tumoursApproach to pituitary tumours
Approach to pituitary tumours
 

Similar to Basics of echo & principles of doppler echocardiography

Basics of Chest Sonography and Anatomy of Chest Wall
Basics of Chest Sonography and Anatomy of Chest WallBasics of Chest Sonography and Anatomy of Chest Wall
Basics of Chest Sonography and Anatomy of Chest Wall
Gamal Agmy
 
basicechocardiography-170913182344 (2).pdf
basicechocardiography-170913182344 (2).pdfbasicechocardiography-170913182344 (2).pdf
basicechocardiography-170913182344 (2).pdf
littlealphonsa
 
basicsofultrasound-170807183128 (1).pdf
basicsofultrasound-170807183128 (1).pdfbasicsofultrasound-170807183128 (1).pdf
basicsofultrasound-170807183128 (1).pdf
littlealphonsa
 

Similar to Basics of echo & principles of doppler echocardiography (20)

Basics of Chest Sonography and Anatomy of Chest Wall
Basics of Chest Sonography and Anatomy of Chest WallBasics of Chest Sonography and Anatomy of Chest Wall
Basics of Chest Sonography and Anatomy of Chest Wall
 
Echo tee and tte
Echo tee and tteEcho tee and tte
Echo tee and tte
 
Usg applications in anaesthesia dr gs
Usg applications in anaesthesia dr gsUsg applications in anaesthesia dr gs
Usg applications in anaesthesia dr gs
 
PPT 1 Introduction to ultrasound.pptx
PPT 1 Introduction to ultrasound.pptxPPT 1 Introduction to ultrasound.pptx
PPT 1 Introduction to ultrasound.pptx
 
USG THORAX [Autosaved].pptx
USG THORAX [Autosaved].pptxUSG THORAX [Autosaved].pptx
USG THORAX [Autosaved].pptx
 
basic of US 2022.ppt
basic of US 2022.pptbasic of US 2022.ppt
basic of US 2022.ppt
 
basicechocardiography-170913182344 (2).pdf
basicechocardiography-170913182344 (2).pdfbasicechocardiography-170913182344 (2).pdf
basicechocardiography-170913182344 (2).pdf
 
Paper
PaperPaper
Paper
 
ultra sound.pptx
ultra sound.pptxultra sound.pptx
ultra sound.pptx
 
basicsofultrasound-170807183128 (1).pdf
basicsofultrasound-170807183128 (1).pdfbasicsofultrasound-170807183128 (1).pdf
basicsofultrasound-170807183128 (1).pdf
 
Ultrasound Physics Made easy - By Dr Chandni Wadhwani
Ultrasound Physics Made easy - By Dr Chandni WadhwaniUltrasound Physics Made easy - By Dr Chandni Wadhwani
Ultrasound Physics Made easy - By Dr Chandni Wadhwani
 
basicsofultrasound-170807183128 (1).pdf
basicsofultrasound-170807183128 (1).pdfbasicsofultrasound-170807183128 (1).pdf
basicsofultrasound-170807183128 (1).pdf
 
Ultrasound & MRI method of examination
Ultrasound & MRI method of examinationUltrasound & MRI method of examination
Ultrasound & MRI method of examination
 
ultrasonography.pptx
ultrasonography.pptxultrasonography.pptx
ultrasonography.pptx
 
chestultrasou.ppt
chestultrasou.pptchestultrasou.ppt
chestultrasou.ppt
 
TRANS ESOPHAGEAL ECHOCARDIOGRAPHY
TRANS ESOPHAGEAL ECHOCARDIOGRAPHYTRANS ESOPHAGEAL ECHOCARDIOGRAPHY
TRANS ESOPHAGEAL ECHOCARDIOGRAPHY
 
Pocket Guide to Chest X-Rays - sample chapter
Pocket Guide to Chest X-Rays - sample chapterPocket Guide to Chest X-Rays - sample chapter
Pocket Guide to Chest X-Rays - sample chapter
 
Principles and technology of two dimensional echocardiography (2)
Principles and technology of two dimensional echocardiography (2)Principles and technology of two dimensional echocardiography (2)
Principles and technology of two dimensional echocardiography (2)
 
Physics
PhysicsPhysics
Physics
 
ultrasound JP.pptx
ultrasound JP.pptxultrasound JP.pptx
ultrasound JP.pptx
 

Recently uploaded

Dehradun Call Girls Service {8854095900} ❤️VVIP ROCKY Call Girl in Dehradun U...
Dehradun Call Girls Service {8854095900} ❤️VVIP ROCKY Call Girl in Dehradun U...Dehradun Call Girls Service {8854095900} ❤️VVIP ROCKY Call Girl in Dehradun U...
Dehradun Call Girls Service {8854095900} ❤️VVIP ROCKY Call Girl in Dehradun U...
Sheetaleventcompany
 
Call Girl in Chennai | Whatsapp No 📞 7427069034 📞 VIP Escorts Service Availab...
Call Girl in Chennai | Whatsapp No 📞 7427069034 📞 VIP Escorts Service Availab...Call Girl in Chennai | Whatsapp No 📞 7427069034 📞 VIP Escorts Service Availab...
Call Girl in Chennai | Whatsapp No 📞 7427069034 📞 VIP Escorts Service Availab...
amritaverma53
 
👉 Chennai Sexy Aunty’s WhatsApp Number 👉📞 7427069034 👉📞 Just📲 Call Ruhi Colle...
👉 Chennai Sexy Aunty’s WhatsApp Number 👉📞 7427069034 👉📞 Just📲 Call Ruhi Colle...👉 Chennai Sexy Aunty’s WhatsApp Number 👉📞 7427069034 👉📞 Just📲 Call Ruhi Colle...
👉 Chennai Sexy Aunty’s WhatsApp Number 👉📞 7427069034 👉📞 Just📲 Call Ruhi Colle...
rajnisinghkjn
 

Recently uploaded (20)

Call 8250092165 Patna Call Girls ₹4.5k Cash Payment With Room Delivery
Call 8250092165 Patna Call Girls ₹4.5k Cash Payment With Room DeliveryCall 8250092165 Patna Call Girls ₹4.5k Cash Payment With Room Delivery
Call 8250092165 Patna Call Girls ₹4.5k Cash Payment With Room Delivery
 
Chandigarh Call Girls Service ❤️🍑 9809698092 👄🫦Independent Escort Service Cha...
Chandigarh Call Girls Service ❤️🍑 9809698092 👄🫦Independent Escort Service Cha...Chandigarh Call Girls Service ❤️🍑 9809698092 👄🫦Independent Escort Service Cha...
Chandigarh Call Girls Service ❤️🍑 9809698092 👄🫦Independent Escort Service Cha...
 
Kolkata Call Girls Naktala 💯Call Us 🔝 8005736733 🔝 💃 Top Class Call Girl Se...
Kolkata Call Girls Naktala  💯Call Us 🔝 8005736733 🔝 💃  Top Class Call Girl Se...Kolkata Call Girls Naktala  💯Call Us 🔝 8005736733 🔝 💃  Top Class Call Girl Se...
Kolkata Call Girls Naktala 💯Call Us 🔝 8005736733 🔝 💃 Top Class Call Girl Se...
 
Bhawanipatna Call Girls 📞9332606886 Call Girls in Bhawanipatna Escorts servic...
Bhawanipatna Call Girls 📞9332606886 Call Girls in Bhawanipatna Escorts servic...Bhawanipatna Call Girls 📞9332606886 Call Girls in Bhawanipatna Escorts servic...
Bhawanipatna Call Girls 📞9332606886 Call Girls in Bhawanipatna Escorts servic...
 
tongue disease lecture Dr Assadawy legacy
tongue disease lecture Dr Assadawy legacytongue disease lecture Dr Assadawy legacy
tongue disease lecture Dr Assadawy legacy
 
Kolkata Call Girls Shobhabazar 💯Call Us 🔝 8005736733 🔝 💃 Top Class Call Gir...
Kolkata Call Girls Shobhabazar  💯Call Us 🔝 8005736733 🔝 💃  Top Class Call Gir...Kolkata Call Girls Shobhabazar  💯Call Us 🔝 8005736733 🔝 💃  Top Class Call Gir...
Kolkata Call Girls Shobhabazar 💯Call Us 🔝 8005736733 🔝 💃 Top Class Call Gir...
 
Dehradun Call Girls Service {8854095900} ❤️VVIP ROCKY Call Girl in Dehradun U...
Dehradun Call Girls Service {8854095900} ❤️VVIP ROCKY Call Girl in Dehradun U...Dehradun Call Girls Service {8854095900} ❤️VVIP ROCKY Call Girl in Dehradun U...
Dehradun Call Girls Service {8854095900} ❤️VVIP ROCKY Call Girl in Dehradun U...
 
ANATOMY AND PHYSIOLOGY OF REPRODUCTIVE SYSTEM.pptx
ANATOMY AND PHYSIOLOGY OF REPRODUCTIVE SYSTEM.pptxANATOMY AND PHYSIOLOGY OF REPRODUCTIVE SYSTEM.pptx
ANATOMY AND PHYSIOLOGY OF REPRODUCTIVE SYSTEM.pptx
 
Call Girls Mussoorie Just Call 8854095900 Top Class Call Girl Service Available
Call Girls Mussoorie Just Call 8854095900 Top Class Call Girl Service AvailableCall Girls Mussoorie Just Call 8854095900 Top Class Call Girl Service Available
Call Girls Mussoorie Just Call 8854095900 Top Class Call Girl Service Available
 
Race Course Road } Book Call Girls in Bangalore | Whatsapp No 6378878445 VIP ...
Race Course Road } Book Call Girls in Bangalore | Whatsapp No 6378878445 VIP ...Race Course Road } Book Call Girls in Bangalore | Whatsapp No 6378878445 VIP ...
Race Course Road } Book Call Girls in Bangalore | Whatsapp No 6378878445 VIP ...
 
(RIYA)🎄Airhostess Call Girl Jaipur Call Now 8445551418 Premium Collection Of ...
(RIYA)🎄Airhostess Call Girl Jaipur Call Now 8445551418 Premium Collection Of ...(RIYA)🎄Airhostess Call Girl Jaipur Call Now 8445551418 Premium Collection Of ...
(RIYA)🎄Airhostess Call Girl Jaipur Call Now 8445551418 Premium Collection Of ...
 
Call Girls in Lucknow Just Call 👉👉 8875999948 Top Class Call Girl Service Ava...
Call Girls in Lucknow Just Call 👉👉 8875999948 Top Class Call Girl Service Ava...Call Girls in Lucknow Just Call 👉👉 8875999948 Top Class Call Girl Service Ava...
Call Girls in Lucknow Just Call 👉👉 8875999948 Top Class Call Girl Service Ava...
 
Low Cost Call Girls Bangalore {9179660964} ❤️VVIP NISHA Call Girls in Bangalo...
Low Cost Call Girls Bangalore {9179660964} ❤️VVIP NISHA Call Girls in Bangalo...Low Cost Call Girls Bangalore {9179660964} ❤️VVIP NISHA Call Girls in Bangalo...
Low Cost Call Girls Bangalore {9179660964} ❤️VVIP NISHA Call Girls in Bangalo...
 
Cardiac Output, Venous Return, and Their Regulation
Cardiac Output, Venous Return, and Their RegulationCardiac Output, Venous Return, and Their Regulation
Cardiac Output, Venous Return, and Their Regulation
 
Call Girl in Chennai | Whatsapp No 📞 7427069034 📞 VIP Escorts Service Availab...
Call Girl in Chennai | Whatsapp No 📞 7427069034 📞 VIP Escorts Service Availab...Call Girl in Chennai | Whatsapp No 📞 7427069034 📞 VIP Escorts Service Availab...
Call Girl in Chennai | Whatsapp No 📞 7427069034 📞 VIP Escorts Service Availab...
 
Call Girls Shahdol Just Call 8250077686 Top Class Call Girl Service Available
Call Girls Shahdol Just Call 8250077686 Top Class Call Girl Service AvailableCall Girls Shahdol Just Call 8250077686 Top Class Call Girl Service Available
Call Girls Shahdol Just Call 8250077686 Top Class Call Girl Service Available
 
👉 Chennai Sexy Aunty’s WhatsApp Number 👉📞 7427069034 👉📞 Just📲 Call Ruhi Colle...
👉 Chennai Sexy Aunty’s WhatsApp Number 👉📞 7427069034 👉📞 Just📲 Call Ruhi Colle...👉 Chennai Sexy Aunty’s WhatsApp Number 👉📞 7427069034 👉📞 Just📲 Call Ruhi Colle...
👉 Chennai Sexy Aunty’s WhatsApp Number 👉📞 7427069034 👉📞 Just📲 Call Ruhi Colle...
 
Gastric Cancer: Сlinical Implementation of Artificial Intelligence, Synergeti...
Gastric Cancer: Сlinical Implementation of Artificial Intelligence, Synergeti...Gastric Cancer: Сlinical Implementation of Artificial Intelligence, Synergeti...
Gastric Cancer: Сlinical Implementation of Artificial Intelligence, Synergeti...
 
Ahmedabad Call Girls Book Now 9630942363 Top Class Ahmedabad Escort Service A...
Ahmedabad Call Girls Book Now 9630942363 Top Class Ahmedabad Escort Service A...Ahmedabad Call Girls Book Now 9630942363 Top Class Ahmedabad Escort Service A...
Ahmedabad Call Girls Book Now 9630942363 Top Class Ahmedabad Escort Service A...
 
Call Girls Rishikesh Just Call 9667172968 Top Class Call Girl Service Available
Call Girls Rishikesh Just Call 9667172968 Top Class Call Girl Service AvailableCall Girls Rishikesh Just Call 9667172968 Top Class Call Girl Service Available
Call Girls Rishikesh Just Call 9667172968 Top Class Call Girl Service Available
 

Basics of echo & principles of doppler echocardiography

  • 1. BASICS OF ECHOCARDIOGRAPHYAND PRINCIPLES OF DOPPLER ECHO Abraha Hailu August 29, 2010
  • 3. 1. BASICS ultrasound (1923) vibrations of the same physical nature as sound but with frequencies above the range of human hearing, namely frequencies greater than 20,000 cycles per second. the diagnostic or therapeutic use of ultrasound and esp. a noninvasive technique involving the formation of a two-dimensional image used for the examination and measurement of internal body structures and the detection of bodily abnormalities a diagnostic examination using ultrasound
  • 4.
  • 5.
  • 6.
  • 7.
  • 8. The depth of penetration:
  • 9. 30 cm for a 1 MHz transducer,
  • 10. 12 cm for 2.5 MHz transducer, and
  • 11. 6 cm for a 5 MHz transducer
  • 12. Air has a very high acoustic impedance, resulting in significant signal attenuation when imaging through lung tissue, especially emphysematous lung, or pathologic conditions such as pneumomediastinum or subcutaneous emphysema
  • 13.
  • 14. ULTRASOUND TRANSDUCERS US transducers use a piezoelectric crystal to generate and receive ultrasound wavesImage formation: is related to the distance of a structure from the transducer, based upon the time interval between ultrasound transmission and arrival of the reflected signalThe amplitude is proportional to the incident angle and acoustic impedance, and timing is proportional to the distance from the transducer
  • 15.
  • 16.
  • 17. 2. IMAGING MODALITIES Two dimensional (2-D) imaging : A 2D image is generated from data obtained mechanically (mechanical transducer) or electronically (phased-array transducer) The signal received undergoes a complex manipulation to form the final image displayed on the monitor including signal amplification, time-gain compensation, filtering, compression and rectification. M-mode: Motion or "M"-mode echocardiography is among the earliest forms of cardiac ultrasound The very high temporal resolution by M-mode imaging permits: identification of subtle abnormalities such as fluttering of the anterior mitral leaflet due to aortic insufficiency or movement of a vegetation. dimensional measurements or changes, such as chamber size and endocardial thickening, can be readily appreciated
  • 18. A. 2-D ECHOCARDIOGRAPHY
  • 19.
  • 20.
  • 21.
  • 22.
  • 23. IMAGING PLANES The long-axis plane is the plane perpendicular to the posterior and anterior surfaces of the body and parallel to the long axis of the heart
  • 24. Sweeping begins at the base of the heart which appears on the rt of the screenThe left atrium, the mitral valve and the right ventricular outflow tract are seen.
  • 25.
  • 26. Parasternal WindowRight Long-Axis ViewSweeping begins at the right atrium which is on the right of the screen. In this view we can see the right atrium and the right ventricle, with the tricuspid valve in between
  • 27.
  • 28. The short-axis plane is the plane that runs parallel to the posterior and anterior surfaces of the body and transects the heart from its apex to its base encompassing all four cardiac chambers
  • 29. Parasternal Short-Axis ViewsObtained from the parasternal window by rotating the transducer clockwise by 90ºHence, the image index marker is pointed toward the patient´s right supraclavicularfossa. A series of sweeps transect the heart from the base to the apex by changes in transducer position and angulation. There are three characteristic levels: 1) great vessels, 2) mitral valve, 3) ventricles.
  • 30. Parasternal Short-Axis ViewGreat Arteries LevelSweeping begins at the left edge of the atrium which appears on the right of the screen
  • 31. At aortic valve level it demonstrates all three aortic valve leaflets. The pulmonary valve, the right ventricular outflow tract, the right atrium and the left atrium are seen in this view.
  • 32. Parasternal Short-Axis ViewGreat Arteries LevelIn this view we can see the pulmonary valve (PV) and pulmonary artery (PA) with right (RPA) and left branches (LPA).
  • 33. Parasternal Short-Axis ViewMitral Valve LevelThe anterior and posterior mitral leaflets are seen as they open in diastole and close in systole
  • 35. The four-chamber plane is the plane that runs parallel to the posterior and anterior surfaces of the body and transects the heart from its apex to its base encompassing all four cardiac chambers
  • 36. Apical WindowSweeping begins at the left edge of the heart which appears on the right of the screen.
  • 37. 4-Chamber ViewIn this view we can see the left ventricle, the right ventricle, the left atrium and the right atrium. The tricuspid annulus lies slightly (up to 10 mm) closer to the apex than the mitral annulus. The septal and posterior tricuspid leaflets are seen in this view.
  • 38. Apical Window5-Chamber ViewBy tilting the transducer anteriorly, the aortic root is seen in an oblique long view.
  • 39. Apical Window2-Chamber ViewSweeping begins at the anterior face of the left ventricle which appears on the right of the screen. From the four-chamber view, the transducer is rotated anti-clockwise by 60º to obtain the two-chamber view of the left ventricle, the mitral valve and the left atrium.
  • 40. Apical Window2-Chamber ViewThis view shows the LV, the LA and the MV. The LV shows the inferior wall on the left and the anterior wall on the right.
  • 41. Subcostal Window4-Chamber ViewSweeping starts at the apex which appears on the right of the screen. A view of all four chambers shows the right ventricular free wall, the midsection of the interventricular septum and the posterolateral left ventricular wall. In this view, the interatrial septum is perpendicular to the direction of the ultrasound beam
  • 42. Subcostal WindowThe inferior vena cava as it enters the right atrium and the central hepatic vein are seen.
  • 43. Suprasternal Notch WindowSweeping begins at the descending aorta which appears on the right of the screen. The long-axis view shows the ascending aorta, the arch, the proximal descending aorta and the origins of the right brachiocephalic and left common carotid and subclavian arteries.
  • 45. B. M-MODE ECHOCARDIOGRAPHY a single line of sight is included, the repetition frequency of the pulse transmission is very high and sampling rates of around 1800 cycles/sec are usedContinuously-moving structures may be identified more accurately when motion versus time, as well as depth, is displayed clearly on the M-mode recording
  • 46.
  • 47. AORTIC VALVE AND LEFT ATRIUM
  • 48. AORTIC VALVE AND LEFT ATRIUM The aortic root is moving anteriorly in systole and posteriorly in diastole. The left atrium is posterior to the aortic root. The aortic leaflet coaptation point is seen as a thin line in diastole. AO is measured in end-diastole and LA in end-systole.
  • 49.
  • 50. In early diastole, the leaflets separate widely, with the maximum early diastolic motion of the anterior leaflet termed the E point. The leaflets move together in mid-diastole and then separate again with atrial systole, the A point. Closure at the end of diastole is termed C point.
  • 51. M-mode recording perpendicular to the long axis of and through the centre of the left ventricle at papillary muscle level provides standard measurements of systolic and diastolic thickness and chamber dimensions:
  • 52. Normal Values Left ventricular end-diastole: 37 - 57 mm (23 -31 mm/m²) Left ventricular end-systole: 21 - 40 mm (14 -21 mm/m²) Interventricular septum: 7 - 11 mm Posterior wall: 7 - 11 mm
  • 53. 3. DOPPLER ECHOCARDIOGRAPHY BASIC PRINCIPLES: utilizes ultrasound to record blood flow within the cardiovascular system (While M-mode and 2D echo create ultrasonic images of the heart) is based upon the changes in frequency of the backscatter signal from small moving structures, ie, red blood cells, intercepted by the ultrasound beam
  • 54.
  • 55. This difference in frequency between the transmitted frequency (F[t]) and received frequency (F[r]) is the Doppler shift:
  • 56.
  • 57. With a stationary target (panel A): the carrier frequency [f(t)] from the transmitting transducer strikes the target and is reflected back to the receiving transducer at the reflected frequency [f(r)], which is unaltered
  • 58. with a target moving toward the transducer (panel B): An increase in f(r) is seen
  • 59. with a target moving away from the transducer (panel C): f(r) is reduced
  • 60.
  • 61.
  • 62.
  • 63. For ø of 90º (perpendicular to blood flow), cosine ø = 0 and the Doppler shift is 0
  • 64. For ø up to 20º, cos ø results in a minimal (<10 percent) change in the Doppler shift
  • 65.
  • 66.
  • 67. By convention, time is displayed on the x axis and frequency shift on the y axis
  • 68.
  • 69.
  • 70.
  • 71.
  • 72.
  • 73. Right Ventricular InflowCan be recorded from an apical approach. The pattern is similar to mitral flow.
  • 74. Left Ventricular OutflowAn apical window is used with a pulsed Doppler sample volume positioned on the left ventricular side of the aortic valve. Note the narrow band of the systolic velocities.
  • 75. Right Ventricular OutflowFrom a parasternal short-axis view, the sample volume is located in the right ventricular outflow tract. The Doppler shape is similar to the left ventricular outflow curve
  • 76. Suprahepatic Vein FlowFrom the longitudinal subcostal view, the sample volume is located at the main suprahepatic vein. There are two positive waves, (A) and (V), and two negative ones, (X) and (Y).
  • 77. DOPPLER MODALITIES  Doppler methods used for cardiac evaluation : continuous wave doppler Pulsed wave doppler color flow doppler
  • 78.
  • 79. Thus, there may be overlap in certain settings, such as:
  • 80. stenoses in series (eg, left ventricular outflow tract gradient and aortic stenosis) or
  • 81.
  • 82.
  • 83. An ideal Doppler profile is one with a smooth "outer" contour, well-defined edge and maximum velocity, and abrupt onset and termination
  • 84.
  • 85.
  • 86.
  • 87.
  • 88. Because PWD repeatedly samples the returning signal, there is a maximum limit to the frequency shift that can be measured unambiguouslyThe maximum detectable frequency shift (the Nyquist limit) is one-half the PRFIf velocity exceeds the Nyquist limit, signal aliasing is seen with the signal cut off at the edge of the display and the top of the waveform appearing in the reverse partHigh-PRF increases the number of sample volumes
  • 89.
  • 90. with flow toward the transducer displayed in orange/red
  • 91.
  • 92. apical four chamber view with color flow Doppler during diastole This color signal is used to position a pulsed wave Doppler sample volume so that quantitatable signals of flow can be obtained from the pulmonary veins and from the mitral leaflet tips
  • 93.
  • 94. The long axis parasternal view with superimposed color flow Doppler mapping of the left ventricular inflow and ouflow tracts obtained during diastole
  • 95. the long axis parasternal view of the left ventricular outflow tract during systole; a normal color flow signal (red-orange) is seen in the left ventricular outflow tract. The occasional blue patches in the signal represent aliasing and suggest that the signal is at or exceeds the Nyquist limit.
  • 96. Short axis view through the base of the heart with CFdoppler imaging short axis view recorded from the base of the heart: Color flow Doppler imaging during systole (panel A), demonstrates normal systolic flow from the right ventricular outflow tract (RVOT) to the main pulmonary artery (MPA). The flow signal is red in the proximal RVOT as it travels towards the transducer. As it moves at right angles to the interrogating beam the signal briefly disappears. When the flow turns away from the transducer and exits into the MPA it is coded blue. short axis view through the base of the heart during diastole: shows diastolic flow signal (red) from pulmonic valve (PV) denoting trivial pulmonary regurgitation (PR) which is found in 90 percent of normals.
  • 97.
  • 98.
  • 99. Five chamber view from a 2-D echo shows a moderate amount of AR and left ventricular enlargement
  • 100. The parasternal long axis view with color Doppler demonstrates severe AR associated with marked dilatation of the aortic root
  • 101.
  • 102. velocity in the stenotic jet (V2)
  • 103.
  • 104.
  • 105.
  • 106. REFERENCES CLINICAL ECHOCARDIOGRAPHY; HOSPITAL UNIVERSITARI VALL D´ HEBRON BARCELONA, Arturo Evangelista and HerminioGarcía del CastilloCo-authors: Teresa González-Alujas, Gustavo Avegliano, ZamiraGómez-Bosch; February 2004 UpToDate 17.3
  • 107. THANKS FOR YOUR ATTENTION