MAPLESON BREATHING SYSTEMS 
DR DEEPAK CHAVAN 
TMH, MUMBAI
Mapleson A (Magill‘s) System 
• The Mapleson A system was designed by Sir Ivan Magill in the 1930 
• The Mapleson A or Magill system is good for spontaneous breathing 
patients, so the fresh gas flow can be lower. 
• However as the APL valve is close to the patient, it is difficult to use.
• It consist of a three-way T-tube connected to the fresh gas outlet a 
reservoir bag and a corrrugated tube. 
• The other end of the reservoir tube is connected to the patient and a 
spring-loaded expiratory valve. 
• Corrugated rubber or plastic tubing: 110-130 cm in length. 
• Reservoir Bag at Machine end. 
• APL valve at the patient end. 
• 
• Tube volume > Tidal volume.
Mapleson A : Functional Analysis 
Inspiration – Gas inhaled from reservoir bag during 
inspiration & RB collapses 
Expiration – During early phase of expiration, the reservoir 
bag is not full 
& hence pressure within the system does not increase. 
Exhaled gas (first portion is dead space gas) passes back 
towards the reservoir bag. 
Towards the later part of expiration , the bag fills further 
& the pressure within the system increases. 
This results in the opening of expiratory valve & venting of 
alveolar gas.
Expiratory pause - Fresh gas washes 
the expired gas out of the reservoir 
tube, filling it with fresh gas for the 
next inspiration. 
Rebreathing of alveolar gas can be 
prevented if the fresh gas flow = 
patient's minute ventilation. 
However, the last gas to be washed 
out of the circuit is dead space gas, 
which consists of warmed and 
humidified fresh gas, and no CO2. If 
some rebreathing of this dead space 
gas is accepted, a flow approximating 
to around 70% of the minute volume 
can be used:
Controlled Ventilation: 
• To facilitate IPPV the expiratory valve has to 
be partly closed. 
• During inspiration the patients gets ventilated 
with FG and part of the FG is vented through 
the valve after sufficient pressure has 
developed to open the valve. 
• During expiration , the FG from the machine 
flows into the reservoir bag and all the 
expired gas ( i.e. dead space and alveolar gas 
flows back into the corrugated tube till the 
system is full.
• During the next inspiration the alveolar gas is pushed 
back into the alveoli followed by the fresh gas. 
• When sufficient pressure is developed, part of the 
expired gas and part of the FG escape through the 
valve. 
• This leads to considerable rebreathing as well as 
excessive wastage of fresh gas . 
• Hence these system are inefficient for controlled 
ventilation.
Advantages: 
• Best among all Maplesons systems for spontaneous ventilation 
• Minimal wastage of gases during spontaneous ventilation 
Disadvantages: 
• Not efficient for controlled ventilation. 
• Wastage of gases & operation theatre pollution. 
• Expiratory valve required –produces slight resistance during expiration. 
• Expiratory valve is heavy(Headbrink valve). 
• Expiratory valve is near patient end inconvenient to use especially in head & neck 
surgeries. 
• Not suitable for paediatric use.
Mapleson A – Lack Modification 
• A co-axial modification of the Mapleson A system. 
• Designed to facilitate scavenging of expired gas & 
make more efficient for controlled ventilation 
• 1.6 m in length 
• FGF through outside tube ( 30mm), exhaled gases 
from inner tube. 
• Inner tube wide in diameter (14 mm) to reduce 
resistance to expiration(1.6 cm H2O). 
• Reservoir bag at machine end 
• APL valve at machine end.
• The Lack circuit is essentially similar in function to the Magill, 
except that the expiratory valve is located at the machine-end of 
the circuit, being connected to the patient adapter by the inner 
coaxial tube.
• Inspiration - The valve closes and the patient inspires fresh gas from 
the outer reservoir tube. 
• Expiratory pause - Fresh gas washes the expired gas out of the 
reservoir tube, filling it with fresh gas for the next inspiration. 
• Expiration - The patient expires into the reservoir tube. Towards the 
end of expiration, the bag fills and positive pressure opens the valve, 
allowing expired gas to escape via the inner exhaust tube
Advantages 
• The location of the valve is more convenient, facilitating intermittent 
positive pressure ventilation and scavenging of expired gas. 
Disadvantages 
• In common with other co-axial systems, if the inner tube becomes 
disconnected or breaks, the entire reservoir tube becomes dead-space. 
• This can be avoided by use of the 'parallel Lack' system, in which the 
inner and outer tubes are replaced by conventional breathing tubing and 
a Y-piece:
Mapleson B System 
• The Mapleson B system features the fresh gas inlet near the 
patient, distal to the expiratory valve. 
• The expiratory valve opens when pressure in the circuit 
increases, and a mixture of alveolar gas and fresh gas is 
discharged. 
• During the next inspiration, a mixture of retained fresh gas 
and alveolar gas is inhaled. 
• Rebreathing is avoided with fresh gas flow rates of greater 
than twice the minute ventilation for both spontaneous and 
controlled ventilation.
Mapleson C system 
• This circuit is also known as Water’s circuit. 
• It is similar in construction to the Mapleson B , but the main 
tube is shorter. 
• A FGF equal to twice the to twice the minute ventilation is 
required to prevent rebreathing. 
• CO2 builds up slowly with this circuit. 
• Mapleson B &C : In order to reduce rebreathing of alveolar 
gas FG entry was shifted to near the patient. 
• This allows a complete mixing of FG and expired gas. 
• The end result is that these system are neither efficient during 
spontaneous nor during controlled ventilation.
Mapleson D System 
• It consists of fresh gas inlet nearer the patient end , a 
corrugated rubber tubing one end which is connected 
with expiratory valve and then reservoir bag. 
• It is mainly used for assisted or controlled ventilation 
• During the controlled ventilation there is little chance 
of rebreathing. 
• The FGF which enters during expiratory pause 
accumulates in the patient end is forced during the 
inflation.
• In spontaneous breathing during inspiration the 
patient will inhale the fresh gas & gas in corrugated 
tube depending on FGF, TV, length of expiratory pause 
& volume of corrugated tube. 
• Rebreathing can be minimized by increasing FGF 2-3 
times the MV. 
• For an adult 15L/min FGF which seems uneconomical 
is required. 
• In some cases 250 ml/kg/min required to prevent 
rebreathing.
Bain circuit 
• In principal it is modification of the system used by 
Macintosh and Pask during Second World War to 
administer anaesthesia. 
• It was introduced by Bain and Spoerel in 1972 
• It is a modification of Mapleson D system. 
• It is a coaxial system in which fresh gas flows through 
a narrow inner tube within outer corrugated tubing 
• It functions like T-piece except that tube supplying FG 
to the patient is located inside the reservoir tube.
Specifications:- 
• Length-1.8 meters. 
• Diameter of outer tube-22mm(transparent,carries 
expiratory gases) 
• Diameter of inner tubing-7 mm(inspiratory) 
• Resistance-Less than0.7 cmH2O 
• Dead space-Outer tube upto expiratory valve( around 
500ml=TV) 
• Flow rates-100-150 ml/kg/min for controlled ventilation. 
Average 300 ml/kg/min for spontaneous ventilation
Bain system (Mapleson D)- Functional Analysis 
Spontaneous respiration: 
• The breathing system should 
be filled with FG before 
connecting to the patient. 
• When the patient takes an 
inspiration, the FG from the 
machine , the reservoir bag and 
the corrugated tube flow to the 
patient.
• During the expiration there is a continuous FGF into the system at 
the patients end .The expired gas gets continuously mixed with the 
FG as it flows back into corrugated tubing and the reservoir bag 
• Once the system is full the excess gas is vented to the atmosphere 
through the valve situated at the end of the corrugated tube near 
the reservoir bag. 
• During the expiratory pause the FG continues to flow and fill the 
proximal portion of the corrugated tube while mix gas is vented 
through valve. 
• During the next inspiration , the patient breathes FG as well as 
mixed gas from the corrugated tube. 
• It is calculated and clinically prove that FGF should at least 1.5- 2 
times the patient’s MV
• Controlled ventilation : 
• To facilitate intermittent positive 
pressure ventilation, the expiratory 
valve has to be partly closed so that 
it opens only after sufficient 
pressure has developed in the 
system. 
• When the system is filled with fresh 
gas, the patient gets ventilated with 
the FGF from the machine, 
corrugated tubing and the reservoir 
bag.
• During expiration expired gas continuously gets mixed 
with FG that is flowing into the system at the patient 
end. 
• During the expiratory pause the FG continues to enter 
the system and pushes the mixed gas towards the 
reservoir. 
• When next inspiration is initiated , the patient gets 
ventilated with the gas in the corrugated tube i.e.a 
mixture of FG, alveolar gas and dead space gas. 
• As the pressure in the system increases,APL valve open 
and the contents of reservoir bag are discharged into 
the atmosphere.
ADVANTAGES OF BAIN'S SYSTEM 
• Light weight. 
• Minimal drag on ETT as compared to Magill's circuit. 
• Low resistance. 
• As the outer tube is transparent, it is easy to detect any kinking 
or disconnection of the inner fresh gas flow tube. 
• It can be used both during spontaneous and controlled 
ventilation and change over is easier. 
• It is useful where patient is not accessible as in MRI suites. 
• Exhaled gases do not accumulate near surgical field, so risk of 
flash fires is abolished. 
• Easy for scavenging of gases as scavenging valve is at machine 
end of the circuit. 
• Easy to connect to ventilator. 
• There is some warming of the inspired fresh gas by the exhaled 
gas present in outer tubing.
DISADVANTAGES OF BAIN'S SYSTEM 
• Due to multiple connections in the circuit there is a risk of disconnections. 
• Wrong assembling of the parts can lead to malfunction of the circuit. 
• Theatre pollution occurs due to high fresh gas flow. However, it can be prevented 
by using scavenging system. 
• Increases the cost due to high fresh gas flows. 
• There can be kinking of the fresh gas supply inner tube blocking the fresh gas 
supply leading to hypoxia 
• There can be crack in the inner tube causing leakage 
• Case report available about the defect in metal head so that fresh gas and exhaled 
gas mix and entire limb becomes dead space 
• It cannot be used in paediatric patients with weight less than 20 kg.
Checking the circuit 
•Mapleson D system is checked for leaks by occluding the patient end, closing the APL 
valve and pressurizing the system. The APL valve is then opened. The bag should deflate 
easily if the valve is working properly. Outer tube integrity should also be checked by 
following the simplest innovative method. Wet the hands with spirit. Blow air through the 
tube. Wipe the tube with wet hands. Leak will produce chillness in the hands. 
•For checking integrity of inner tube of Bains system, a test is performed by setting a low 
flow on the oxygen flowmeter and occluding the inner tube with a finger or barrel of a 
small syringe at the patient end while observing the flow meter indicator. If the inner tube 
is intact and correctly connected, the indicator will fall. 
•Pethicks test - To check the integrity of the inner tube, activate the oxygen flush and 
observe the bag. Due to venturi effect the high flow from the inner tube at the patient end 
will create a negative pressure in the outer exhalation tubing and this will suck gas from 
the bag and bag will deflate. If the inner tube is not intact, this maneuver will cause the 
bag to inflate slightly.
Mapleson E and F 
• Valveless breathing system used for children upto 30 
kg. 
• Suitable for spontaneous and controlled ventilation 
• Components:- 
• - T shaped tubing with 3 ports. 
• -FGF delivered to one port 
• -2 nd port goes to patient & 3rd to reservoir tube.
Mapleson E OR Ayre’s T- PIECE 
• Introduced by Phillips Arye in 1937. 
• Belongs to Mapleson E. 
• Available as meatllic / plastic. 
• Length – 2 inches. 
• Parts – inlet, outlet, side tube. 
• Inlet size-10 mm, outlet size-10mm metallic 
&15 mm plastic
Advantages 
• Simple to use , Light weight . 
• No dead space , no resistance. 
• For pediatric patients less than 20 kgs. 
• Expiratory limb is attached to the outlet of T piece.- 
• It should accommodate air space equal to 1/3 rd of 
TV. 
• If too short – air dilution in spont. Breathing patients 
& patient become light. 
• 1 inch of expiratory tube can accommodate 2-3 ml of 
gas. 
• Fresh Gas Flows – 2- 3 times MV
Dis advantages: 
• High flow rates are required. 
• Loss of heat & humidity. 
• Risk of accidental occlusion of expiratory limb- risk of increased 
airway pressure & barotrauma to lungs.
Mapleson F OR Jackson Rees Circuit 
• It is a modification of Mapleson E by Jackson Rees and is known as 
Jackson Rees modification. 
• It has a 500 ml bag attached to the expiratory limb. 
• This bag helps in respiratory monitoring or assisting the respiration. 
• It also helps in venting out excess gases. 
• The bag has a hole in the tail of the bag that is occluded by using a 
finger to provide pressure. 
• The bags with valve are also available. 
• It is used in neonates, infants, and paediatric patients less than 20 kg 
in weight or less than 5 years of age.
• Technique of use 
For spontaneous respiration: 
The relief mechanism of the bag is left fully open. 
• For controlled respiration: 
The hole in the bag can be occluded by the user during inspiration 
and ventilation is done by squeezing the bag.
• Functional analysis 
• It also functions like Mapleson D system. 
• The flows required to prevent rebreathing are 2.5-3.0 times minute volume 
during spontaneous ventilation and 1.5 to 2 times the minute volume during 
controlled ventilation. 
• In this system, during expiration fresh gas and exhaled gas will collect and mix in 
the bag. 
• The next inspiration results in patient inhaling fresh gas both direct from inlet 
and from expiratory part of the circuit as in Mapleson E.
• During expiratory pause the expired gases are replaced by fresh gas in the 
expiratory limb. 
• Observation of bag movements helps in assessing respiration during 
spontaneous breathing. 
• It also allows controlled ventilation by squeezing the bag. 
• Heat and moisture exchanger should not be used with Mapleson E and F 
during spontaneous respiration as it increases resistance. 
• So most of the fresh gas will enter expiratory limb leading to wastage of 
fresh gases and delaying induction by inhalation agents.

More Related Content

What's hot

Supraglottic airway device
Supraglottic airway deviceSupraglottic airway device
Supraglottic airway deviceDebojyoti Dutta
 
Breathing circuit's
Breathing circuit'sBreathing circuit's
Breathing circuit'sImran Sheikh
 
Breathing System
Breathing SystemBreathing System
Breathing SystemKhalid
 
Intermediate & low pressure system
Intermediate & low pressure systemIntermediate & low pressure system
Intermediate & low pressure systemDeepa Sinha
 
Breathing circuits
Breathing circuitsBreathing circuits
Breathing circuitsgramanathan
 
Bougie, trachlite , laryngeal tube , combitube , i gel ,truview
Bougie, trachlite , laryngeal tube , combitube , i gel ,truviewBougie, trachlite , laryngeal tube , combitube , i gel ,truview
Bougie, trachlite , laryngeal tube , combitube , i gel ,truviewDhritiman Chakrabarti
 
Double Lumen Endobronchial Tubes ppt
Double Lumen Endobronchial Tubes pptDouble Lumen Endobronchial Tubes ppt
Double Lumen Endobronchial Tubes pptImran Sheikh
 
anaesthesia Breathing circuits and its classification and functional analysis
anaesthesia Breathing circuits and its classification and functional analysisanaesthesia Breathing circuits and its classification and functional analysis
anaesthesia Breathing circuits and its classification and functional analysisprateek gupta
 
Anaesthesia breathing systems
Anaesthesia breathing systemsAnaesthesia breathing systems
Anaesthesia breathing systemsD Nkar
 
Neuromuscular Monitoring
Neuromuscular MonitoringNeuromuscular Monitoring
Neuromuscular MonitoringMohtasib Madaoo
 
Anaesthesia Workstation checklist and safety features
Anaesthesia Workstation checklist and safety features Anaesthesia Workstation checklist and safety features
Anaesthesia Workstation checklist and safety features ZIKRULLAH MALLICK
 
Low pressure system in anaesthesia machine
Low pressure system in anaesthesia machineLow pressure system in anaesthesia machine
Low pressure system in anaesthesia machineSwadheen Rout
 
Gas laws in anaesthesia
Gas laws in anaesthesiaGas laws in anaesthesia
Gas laws in anaesthesiaDavis Kurian
 

What's hot (20)

Supraglottic airway device
Supraglottic airway deviceSupraglottic airway device
Supraglottic airway device
 
Breathing circuit's
Breathing circuit'sBreathing circuit's
Breathing circuit's
 
Breathing System
Breathing SystemBreathing System
Breathing System
 
Intermediate & low pressure system
Intermediate & low pressure systemIntermediate & low pressure system
Intermediate & low pressure system
 
Breathing circuits
Breathing circuitsBreathing circuits
Breathing circuits
 
Bougie, trachlite , laryngeal tube , combitube , i gel ,truview
Bougie, trachlite , laryngeal tube , combitube , i gel ,truviewBougie, trachlite , laryngeal tube , combitube , i gel ,truview
Bougie, trachlite , laryngeal tube , combitube , i gel ,truview
 
Double Lumen Endobronchial Tubes ppt
Double Lumen Endobronchial Tubes pptDouble Lumen Endobronchial Tubes ppt
Double Lumen Endobronchial Tubes ppt
 
The canister (the absorber)
The canister (the absorber)The canister (the absorber)
The canister (the absorber)
 
anaesthesia Breathing circuits and its classification and functional analysis
anaesthesia Breathing circuits and its classification and functional analysisanaesthesia Breathing circuits and its classification and functional analysis
anaesthesia Breathing circuits and its classification and functional analysis
 
Anaesthesia breathing systems
Anaesthesia breathing systemsAnaesthesia breathing systems
Anaesthesia breathing systems
 
Laryngeal mask-airway
Laryngeal mask-airwayLaryngeal mask-airway
Laryngeal mask-airway
 
Anaesthesia machine 2
Anaesthesia machine 2Anaesthesia machine 2
Anaesthesia machine 2
 
Breathing circuits
Breathing circuitsBreathing circuits
Breathing circuits
 
Spinal & epidural needle
Spinal & epidural needleSpinal & epidural needle
Spinal & epidural needle
 
Neuromuscular Monitoring
Neuromuscular MonitoringNeuromuscular Monitoring
Neuromuscular Monitoring
 
Anaesthesia machine 1
Anaesthesia machine 1Anaesthesia machine 1
Anaesthesia machine 1
 
Anaesthesia Workstation checklist and safety features
Anaesthesia Workstation checklist and safety features Anaesthesia Workstation checklist and safety features
Anaesthesia Workstation checklist and safety features
 
Low pressure system in anaesthesia machine
Low pressure system in anaesthesia machineLow pressure system in anaesthesia machine
Low pressure system in anaesthesia machine
 
Gas laws in anaesthesia
Gas laws in anaesthesiaGas laws in anaesthesia
Gas laws in anaesthesia
 
Capnography
CapnographyCapnography
Capnography
 

Viewers also liked

Mapleson breathing systems
Mapleson breathing systemsMapleson breathing systems
Mapleson breathing systemsgaganbrar18
 
Breathing systems open circuit- shoeib
Breathing systems  open circuit- shoeibBreathing systems  open circuit- shoeib
Breathing systems open circuit- shoeibhavalprit
 
Anesthesia breathing systems
Anesthesia breathing systemsAnesthesia breathing systems
Anesthesia breathing systemsKaren Haddock
 
OPOIDS by Dr. Nadeem Korai
OPOIDS by Dr. Nadeem KoraiOPOIDS by Dr. Nadeem Korai
OPOIDS by Dr. Nadeem KoraiNadeemkorai
 
Difficult airway management for nursing staff
Difficult airway management for nursing staffDifficult airway management for nursing staff
Difficult airway management for nursing staffrashidmkhan
 
Opioids & Their Use in Anaesthesia
Opioids & Their Use in Anaesthesia Opioids & Their Use in Anaesthesia
Opioids & Their Use in Anaesthesia Zareer Tafadar
 
Face masks, laryngeal tube, airways yuvaraj
Face masks, laryngeal tube, airways  yuvarajFace masks, laryngeal tube, airways  yuvaraj
Face masks, laryngeal tube, airways yuvarajhavalprit
 
Advanced cardiac life support(acls)
Advanced cardiac life support(acls)Advanced cardiac life support(acls)
Advanced cardiac life support(acls)omar143
 
OPIOID AGONISTS AND ANTAGONISTS
OPIOID AGONISTS AND ANTAGONISTSOPIOID AGONISTS AND ANTAGONISTS
OPIOID AGONISTS AND ANTAGONISTSjamal53
 
Muscle Relaxants
Muscle RelaxantsMuscle Relaxants
Muscle RelaxantsKhalid
 
Opioid analgesics
Opioid analgesicsOpioid analgesics
Opioid analgesicsfarhan_aq91
 
Basic Of Mechanical Ventilation
Basic Of Mechanical VentilationBasic Of Mechanical Ventilation
Basic Of Mechanical VentilationDang Thanh Tuan
 
Basic Mechanical Ventilation
Basic Mechanical VentilationBasic Mechanical Ventilation
Basic Mechanical VentilationAndrew Ferguson
 
Emergency Airway Management 2014: Mark P. Brady PA-C
Emergency Airway Management 2014: Mark P. Brady PA-CEmergency Airway Management 2014: Mark P. Brady PA-C
Emergency Airway Management 2014: Mark P. Brady PA-Cmarkpbrady
 

Viewers also liked (20)

Mapleson breathing systems
Mapleson breathing systemsMapleson breathing systems
Mapleson breathing systems
 
Breathing systems open circuit- shoeib
Breathing systems  open circuit- shoeibBreathing systems  open circuit- shoeib
Breathing systems open circuit- shoeib
 
Anesthesia breathing systems
Anesthesia breathing systemsAnesthesia breathing systems
Anesthesia breathing systems
 
OPOIDS by Dr. Nadeem Korai
OPOIDS by Dr. Nadeem KoraiOPOIDS by Dr. Nadeem Korai
OPOIDS by Dr. Nadeem Korai
 
Difficult airway management for nursing staff
Difficult airway management for nursing staffDifficult airway management for nursing staff
Difficult airway management for nursing staff
 
Breathing systems (2)
Breathing systems (2)Breathing systems (2)
Breathing systems (2)
 
Basic ventilatory modes
Basic ventilatory modesBasic ventilatory modes
Basic ventilatory modes
 
Physics and its laws in anaesthesia
Physics and its laws in anaesthesiaPhysics and its laws in anaesthesia
Physics and its laws in anaesthesia
 
Opioids & Their Use in Anaesthesia
Opioids & Their Use in Anaesthesia Opioids & Their Use in Anaesthesia
Opioids & Their Use in Anaesthesia
 
Face masks, laryngeal tube, airways yuvaraj
Face masks, laryngeal tube, airways  yuvarajFace masks, laryngeal tube, airways  yuvaraj
Face masks, laryngeal tube, airways yuvaraj
 
Advanced cardiac life support(acls)
Advanced cardiac life support(acls)Advanced cardiac life support(acls)
Advanced cardiac life support(acls)
 
OPIOID AGONISTS AND ANTAGONISTS
OPIOID AGONISTS AND ANTAGONISTSOPIOID AGONISTS AND ANTAGONISTS
OPIOID AGONISTS AND ANTAGONISTS
 
Muscle Relaxants
Muscle RelaxantsMuscle Relaxants
Muscle Relaxants
 
Opioid analgesics
Opioid analgesicsOpioid analgesics
Opioid analgesics
 
Basic Of Mechanical Ventilation
Basic Of Mechanical VentilationBasic Of Mechanical Ventilation
Basic Of Mechanical Ventilation
 
Skeletal muscle relaxants - drdhriti
Skeletal muscle relaxants - drdhritiSkeletal muscle relaxants - drdhriti
Skeletal muscle relaxants - drdhriti
 
OPIOID ANALGESICS
OPIOID ANALGESICSOPIOID ANALGESICS
OPIOID ANALGESICS
 
Basic Mechanical Ventilation
Basic Mechanical VentilationBasic Mechanical Ventilation
Basic Mechanical Ventilation
 
Emergency Airway Management 2014: Mark P. Brady PA-C
Emergency Airway Management 2014: Mark P. Brady PA-CEmergency Airway Management 2014: Mark P. Brady PA-C
Emergency Airway Management 2014: Mark P. Brady PA-C
 
Opioids
OpioidsOpioids
Opioids
 

Similar to Mapleson breathing systems

BREATHING_CIRCUITS POTC 2022 march.pptx
BREATHING_CIRCUITS POTC 2022 march.pptxBREATHING_CIRCUITS POTC 2022 march.pptx
BREATHING_CIRCUITS POTC 2022 march.pptxsanjotNinave2
 
Bain’s circuit (Mapelson D) by Zikrullah mallick
Bain’s circuit (Mapelson D) by Zikrullah mallickBain’s circuit (Mapelson D) by Zikrullah mallick
Bain’s circuit (Mapelson D) by Zikrullah mallickZIKRULLAH MALLICK
 
Breathing systems - Mapleson Classification
Breathing systems - Mapleson ClassificationBreathing systems - Mapleson Classification
Breathing systems - Mapleson ClassificationMr.Harshad Khade
 
bains circuit copy copy.pptx
bains circuit copy copy.pptxbains circuit copy copy.pptx
bains circuit copy copy.pptxTushar Mankar
 
Anaesthesia breathing systems
Anaesthesia breathing systemsAnaesthesia breathing systems
Anaesthesia breathing systemsZIKRULLAH MALLICK
 
BREATHING CICUITS.pptx
BREATHING CICUITS.pptxBREATHING CICUITS.pptx
BREATHING CICUITS.pptxmohamedrifky10
 
3. breathing system bsc
3. breathing system bsc3. breathing system bsc
3. breathing system bscCHERUDUGASE
 
Presentation.power point presentation for
Presentation.power point presentation forPresentation.power point presentation for
Presentation.power point presentation forPranavTrehan2
 
Chapter 3 anesthetic_equipment_and_breathing_systems_october_20121
Chapter 3 anesthetic_equipment_and_breathing_systems_october_20121Chapter 3 anesthetic_equipment_and_breathing_systems_october_20121
Chapter 3 anesthetic_equipment_and_breathing_systems_october_20121SamuelDaksa
 
lowflowsystem-200601004249.pptx
lowflowsystem-200601004249.pptxlowflowsystem-200601004249.pptx
lowflowsystem-200601004249.pptxJaseerAk1
 
ventilator mode.pdf
ventilator mode.pdfventilator mode.pdf
ventilator mode.pdfsyedumair76
 

Similar to Mapleson breathing systems (20)

BREATHING SYSTEMS 1.pptx
BREATHING SYSTEMS 1.pptxBREATHING SYSTEMS 1.pptx
BREATHING SYSTEMS 1.pptx
 
Breathing systems (1)
Breathing systems (1)Breathing systems (1)
Breathing systems (1)
 
Breathing Circuits.pptx
Breathing Circuits.pptxBreathing Circuits.pptx
Breathing Circuits.pptx
 
Breathing Circuits.pptx
Breathing Circuits.pptxBreathing Circuits.pptx
Breathing Circuits.pptx
 
BREATHING_CIRCUITS POTC 2022 march.pptx
BREATHING_CIRCUITS POTC 2022 march.pptxBREATHING_CIRCUITS POTC 2022 march.pptx
BREATHING_CIRCUITS POTC 2022 march.pptx
 
Final circuits
Final circuitsFinal circuits
Final circuits
 
#Reservoir bag
#Reservoir bag#Reservoir bag
#Reservoir bag
 
Bain’s circuit (Mapelson D) by Zikrullah mallick
Bain’s circuit (Mapelson D) by Zikrullah mallickBain’s circuit (Mapelson D) by Zikrullah mallick
Bain’s circuit (Mapelson D) by Zikrullah mallick
 
Breathing systems - Mapleson Classification
Breathing systems - Mapleson ClassificationBreathing systems - Mapleson Classification
Breathing systems - Mapleson Classification
 
bains circuit copy copy.pptx
bains circuit copy copy.pptxbains circuit copy copy.pptx
bains circuit copy copy.pptx
 
Anaesthesia breathing systems
Anaesthesia breathing systemsAnaesthesia breathing systems
Anaesthesia breathing systems
 
Breathing systems
Breathing systemsBreathing systems
Breathing systems
 
BREATHING CICUITS.pptx
BREATHING CICUITS.pptxBREATHING CICUITS.pptx
BREATHING CICUITS.pptx
 
3. breathing system bsc
3. breathing system bsc3. breathing system bsc
3. breathing system bsc
 
Presentation.power point presentation for
Presentation.power point presentation forPresentation.power point presentation for
Presentation.power point presentation for
 
breathin system anesthesia.pptx
breathin system anesthesia.pptxbreathin system anesthesia.pptx
breathin system anesthesia.pptx
 
Chapter 3 anesthetic_equipment_and_breathing_systems_october_20121
Chapter 3 anesthetic_equipment_and_breathing_systems_october_20121Chapter 3 anesthetic_equipment_and_breathing_systems_october_20121
Chapter 3 anesthetic_equipment_and_breathing_systems_october_20121
 
lowflowsystem-200601004249.pptx
lowflowsystem-200601004249.pptxlowflowsystem-200601004249.pptx
lowflowsystem-200601004249.pptx
 
ventilator mode.pdf
ventilator mode.pdfventilator mode.pdf
ventilator mode.pdf
 
Ventilator mode
Ventilator modeVentilator mode
Ventilator mode
 

Recently uploaded

Call Girls LB Nagar 7001305949 all area service COD available Any Time
Call Girls LB Nagar 7001305949 all area service COD available Any TimeCall Girls LB Nagar 7001305949 all area service COD available Any Time
Call Girls LB Nagar 7001305949 all area service COD available Any Timedelhimodelshub1
 
Models Call Girls Electronic City | 7001305949 At Low Cost Cash Payment Booking
Models Call Girls Electronic City | 7001305949 At Low Cost Cash Payment BookingModels Call Girls Electronic City | 7001305949 At Low Cost Cash Payment Booking
Models Call Girls Electronic City | 7001305949 At Low Cost Cash Payment Bookingnarwatsonia7
 
Basics of Anatomy- Language of Anatomy.pptx
Basics of Anatomy- Language of Anatomy.pptxBasics of Anatomy- Language of Anatomy.pptx
Basics of Anatomy- Language of Anatomy.pptxAyushGupta813444
 
Russian Call Girls in Hyderabad Ishita 9907093804 Independent Escort Service ...
Russian Call Girls in Hyderabad Ishita 9907093804 Independent Escort Service ...Russian Call Girls in Hyderabad Ishita 9907093804 Independent Escort Service ...
Russian Call Girls in Hyderabad Ishita 9907093804 Independent Escort Service ...delhimodelshub1
 
2025 Inpatient Prospective Payment System (IPPS) Proposed Rule
2025 Inpatient Prospective Payment System (IPPS) Proposed Rule2025 Inpatient Prospective Payment System (IPPS) Proposed Rule
2025 Inpatient Prospective Payment System (IPPS) Proposed RuleShelby Lewis
 
Kukatpally Call Girls Services 9907093804 High Class Babes Here Call Now
Kukatpally Call Girls Services 9907093804 High Class Babes Here Call NowKukatpally Call Girls Services 9907093804 High Class Babes Here Call Now
Kukatpally Call Girls Services 9907093804 High Class Babes Here Call NowHyderabad Call Girls Services
 
Leading transformational change: inner and outer skills
Leading transformational change: inner and outer skillsLeading transformational change: inner and outer skills
Leading transformational change: inner and outer skillsHelenBevan4
 
Experience learning - lessons from 25 years of ATACC - Mark Forrest and Halde...
Experience learning - lessons from 25 years of ATACC - Mark Forrest and Halde...Experience learning - lessons from 25 years of ATACC - Mark Forrest and Halde...
Experience learning - lessons from 25 years of ATACC - Mark Forrest and Halde...scanFOAM
 
Call Girl Service ITPL - [ Cash on Delivery ] Contact 7001305949 Escorts Service
Call Girl Service ITPL - [ Cash on Delivery ] Contact 7001305949 Escorts ServiceCall Girl Service ITPL - [ Cash on Delivery ] Contact 7001305949 Escorts Service
Call Girl Service ITPL - [ Cash on Delivery ] Contact 7001305949 Escorts Servicenarwatsonia7
 
Russian Call Girls Hyderabad Saloni 9907093804 Independent Escort Service Hyd...
Russian Call Girls Hyderabad Saloni 9907093804 Independent Escort Service Hyd...Russian Call Girls Hyderabad Saloni 9907093804 Independent Escort Service Hyd...
Russian Call Girls Hyderabad Saloni 9907093804 Independent Escort Service Hyd...delhimodelshub1
 
Russian Call Girls in Raipur 9873940964 Book Hot And Sexy Girls
Russian Call Girls in Raipur 9873940964 Book Hot And Sexy GirlsRussian Call Girls in Raipur 9873940964 Book Hot And Sexy Girls
Russian Call Girls in Raipur 9873940964 Book Hot And Sexy Girlsddev2574
 
VIP Call Girls Hyderabad Megha 9907093804 Independent Escort Service Hyderabad
VIP Call Girls Hyderabad Megha 9907093804 Independent Escort Service HyderabadVIP Call Girls Hyderabad Megha 9907093804 Independent Escort Service Hyderabad
VIP Call Girls Hyderabad Megha 9907093804 Independent Escort Service Hyderabaddelhimodelshub1
 
Gurgaon Sector 90 Call Girls ( 9873940964 ) Book Hot And Sexy Girls In A Few ...
Gurgaon Sector 90 Call Girls ( 9873940964 ) Book Hot And Sexy Girls In A Few ...Gurgaon Sector 90 Call Girls ( 9873940964 ) Book Hot And Sexy Girls In A Few ...
Gurgaon Sector 90 Call Girls ( 9873940964 ) Book Hot And Sexy Girls In A Few ...ggsonu500
 
Call Girl Chandigarh Mallika ❤️🍑 9907093804 👄🫦 Independent Escort Service Cha...
Call Girl Chandigarh Mallika ❤️🍑 9907093804 👄🫦 Independent Escort Service Cha...Call Girl Chandigarh Mallika ❤️🍑 9907093804 👄🫦 Independent Escort Service Cha...
Call Girl Chandigarh Mallika ❤️🍑 9907093804 👄🫦 Independent Escort Service Cha...High Profile Call Girls Chandigarh Aarushi
 
Hi,Fi Call Girl In Marathahalli - 7001305949 with real photos and phone numbers
Hi,Fi Call Girl In Marathahalli - 7001305949 with real photos and phone numbersHi,Fi Call Girl In Marathahalli - 7001305949 with real photos and phone numbers
Hi,Fi Call Girl In Marathahalli - 7001305949 with real photos and phone numbersnarwatsonia7
 
Gurgaon iffco chowk 🔝 Call Girls Service 🔝 ( 8264348440 ) unlimited hard sex ...
Gurgaon iffco chowk 🔝 Call Girls Service 🔝 ( 8264348440 ) unlimited hard sex ...Gurgaon iffco chowk 🔝 Call Girls Service 🔝 ( 8264348440 ) unlimited hard sex ...
Gurgaon iffco chowk 🔝 Call Girls Service 🔝 ( 8264348440 ) unlimited hard sex ...soniya singh
 

Recently uploaded (20)

Call Girl Guwahati Aashi 👉 7001305949 👈 🔝 Independent Escort Service Guwahati
Call Girl Guwahati Aashi 👉 7001305949 👈 🔝 Independent Escort Service GuwahatiCall Girl Guwahati Aashi 👉 7001305949 👈 🔝 Independent Escort Service Guwahati
Call Girl Guwahati Aashi 👉 7001305949 👈 🔝 Independent Escort Service Guwahati
 
Call Girls LB Nagar 7001305949 all area service COD available Any Time
Call Girls LB Nagar 7001305949 all area service COD available Any TimeCall Girls LB Nagar 7001305949 all area service COD available Any Time
Call Girls LB Nagar 7001305949 all area service COD available Any Time
 
Models Call Girls Electronic City | 7001305949 At Low Cost Cash Payment Booking
Models Call Girls Electronic City | 7001305949 At Low Cost Cash Payment BookingModels Call Girls Electronic City | 7001305949 At Low Cost Cash Payment Booking
Models Call Girls Electronic City | 7001305949 At Low Cost Cash Payment Booking
 
Basics of Anatomy- Language of Anatomy.pptx
Basics of Anatomy- Language of Anatomy.pptxBasics of Anatomy- Language of Anatomy.pptx
Basics of Anatomy- Language of Anatomy.pptx
 
Russian Call Girls in Hyderabad Ishita 9907093804 Independent Escort Service ...
Russian Call Girls in Hyderabad Ishita 9907093804 Independent Escort Service ...Russian Call Girls in Hyderabad Ishita 9907093804 Independent Escort Service ...
Russian Call Girls in Hyderabad Ishita 9907093804 Independent Escort Service ...
 
2025 Inpatient Prospective Payment System (IPPS) Proposed Rule
2025 Inpatient Prospective Payment System (IPPS) Proposed Rule2025 Inpatient Prospective Payment System (IPPS) Proposed Rule
2025 Inpatient Prospective Payment System (IPPS) Proposed Rule
 
Kukatpally Call Girls Services 9907093804 High Class Babes Here Call Now
Kukatpally Call Girls Services 9907093804 High Class Babes Here Call NowKukatpally Call Girls Services 9907093804 High Class Babes Here Call Now
Kukatpally Call Girls Services 9907093804 High Class Babes Here Call Now
 
Leading transformational change: inner and outer skills
Leading transformational change: inner and outer skillsLeading transformational change: inner and outer skills
Leading transformational change: inner and outer skills
 
Experience learning - lessons from 25 years of ATACC - Mark Forrest and Halde...
Experience learning - lessons from 25 years of ATACC - Mark Forrest and Halde...Experience learning - lessons from 25 years of ATACC - Mark Forrest and Halde...
Experience learning - lessons from 25 years of ATACC - Mark Forrest and Halde...
 
Call Girl Dehradun Aashi 🔝 7001305949 🔝 💃 Independent Escort Service Dehradun
Call Girl Dehradun Aashi 🔝 7001305949 🔝 💃 Independent Escort Service DehradunCall Girl Dehradun Aashi 🔝 7001305949 🔝 💃 Independent Escort Service Dehradun
Call Girl Dehradun Aashi 🔝 7001305949 🔝 💃 Independent Escort Service Dehradun
 
Call Girl Service ITPL - [ Cash on Delivery ] Contact 7001305949 Escorts Service
Call Girl Service ITPL - [ Cash on Delivery ] Contact 7001305949 Escorts ServiceCall Girl Service ITPL - [ Cash on Delivery ] Contact 7001305949 Escorts Service
Call Girl Service ITPL - [ Cash on Delivery ] Contact 7001305949 Escorts Service
 
Russian Call Girls Hyderabad Saloni 9907093804 Independent Escort Service Hyd...
Russian Call Girls Hyderabad Saloni 9907093804 Independent Escort Service Hyd...Russian Call Girls Hyderabad Saloni 9907093804 Independent Escort Service Hyd...
Russian Call Girls Hyderabad Saloni 9907093804 Independent Escort Service Hyd...
 
Russian Call Girls in Raipur 9873940964 Book Hot And Sexy Girls
Russian Call Girls in Raipur 9873940964 Book Hot And Sexy GirlsRussian Call Girls in Raipur 9873940964 Book Hot And Sexy Girls
Russian Call Girls in Raipur 9873940964 Book Hot And Sexy Girls
 
VIP Call Girls Hyderabad Megha 9907093804 Independent Escort Service Hyderabad
VIP Call Girls Hyderabad Megha 9907093804 Independent Escort Service HyderabadVIP Call Girls Hyderabad Megha 9907093804 Independent Escort Service Hyderabad
VIP Call Girls Hyderabad Megha 9907093804 Independent Escort Service Hyderabad
 
Call Girls Guwahati Aaradhya 👉 7001305949👈 🎶 Independent Escort Service Guwahati
Call Girls Guwahati Aaradhya 👉 7001305949👈 🎶 Independent Escort Service GuwahatiCall Girls Guwahati Aaradhya 👉 7001305949👈 🎶 Independent Escort Service Guwahati
Call Girls Guwahati Aaradhya 👉 7001305949👈 🎶 Independent Escort Service Guwahati
 
Gurgaon Sector 90 Call Girls ( 9873940964 ) Book Hot And Sexy Girls In A Few ...
Gurgaon Sector 90 Call Girls ( 9873940964 ) Book Hot And Sexy Girls In A Few ...Gurgaon Sector 90 Call Girls ( 9873940964 ) Book Hot And Sexy Girls In A Few ...
Gurgaon Sector 90 Call Girls ( 9873940964 ) Book Hot And Sexy Girls In A Few ...
 
Call Girl Chandigarh Mallika ❤️🍑 9907093804 👄🫦 Independent Escort Service Cha...
Call Girl Chandigarh Mallika ❤️🍑 9907093804 👄🫦 Independent Escort Service Cha...Call Girl Chandigarh Mallika ❤️🍑 9907093804 👄🫦 Independent Escort Service Cha...
Call Girl Chandigarh Mallika ❤️🍑 9907093804 👄🫦 Independent Escort Service Cha...
 
College Call Girls Dehradun Kavya 🔝 7001305949 🔝 📍 Independent Escort Service...
College Call Girls Dehradun Kavya 🔝 7001305949 🔝 📍 Independent Escort Service...College Call Girls Dehradun Kavya 🔝 7001305949 🔝 📍 Independent Escort Service...
College Call Girls Dehradun Kavya 🔝 7001305949 🔝 📍 Independent Escort Service...
 
Hi,Fi Call Girl In Marathahalli - 7001305949 with real photos and phone numbers
Hi,Fi Call Girl In Marathahalli - 7001305949 with real photos and phone numbersHi,Fi Call Girl In Marathahalli - 7001305949 with real photos and phone numbers
Hi,Fi Call Girl In Marathahalli - 7001305949 with real photos and phone numbers
 
Gurgaon iffco chowk 🔝 Call Girls Service 🔝 ( 8264348440 ) unlimited hard sex ...
Gurgaon iffco chowk 🔝 Call Girls Service 🔝 ( 8264348440 ) unlimited hard sex ...Gurgaon iffco chowk 🔝 Call Girls Service 🔝 ( 8264348440 ) unlimited hard sex ...
Gurgaon iffco chowk 🔝 Call Girls Service 🔝 ( 8264348440 ) unlimited hard sex ...
 

Mapleson breathing systems

  • 1. MAPLESON BREATHING SYSTEMS DR DEEPAK CHAVAN TMH, MUMBAI
  • 2. Mapleson A (Magill‘s) System • The Mapleson A system was designed by Sir Ivan Magill in the 1930 • The Mapleson A or Magill system is good for spontaneous breathing patients, so the fresh gas flow can be lower. • However as the APL valve is close to the patient, it is difficult to use.
  • 3. • It consist of a three-way T-tube connected to the fresh gas outlet a reservoir bag and a corrrugated tube. • The other end of the reservoir tube is connected to the patient and a spring-loaded expiratory valve. • Corrugated rubber or plastic tubing: 110-130 cm in length. • Reservoir Bag at Machine end. • APL valve at the patient end. • • Tube volume > Tidal volume.
  • 4. Mapleson A : Functional Analysis Inspiration – Gas inhaled from reservoir bag during inspiration & RB collapses Expiration – During early phase of expiration, the reservoir bag is not full & hence pressure within the system does not increase. Exhaled gas (first portion is dead space gas) passes back towards the reservoir bag. Towards the later part of expiration , the bag fills further & the pressure within the system increases. This results in the opening of expiratory valve & venting of alveolar gas.
  • 5. Expiratory pause - Fresh gas washes the expired gas out of the reservoir tube, filling it with fresh gas for the next inspiration. Rebreathing of alveolar gas can be prevented if the fresh gas flow = patient's minute ventilation. However, the last gas to be washed out of the circuit is dead space gas, which consists of warmed and humidified fresh gas, and no CO2. If some rebreathing of this dead space gas is accepted, a flow approximating to around 70% of the minute volume can be used:
  • 6. Controlled Ventilation: • To facilitate IPPV the expiratory valve has to be partly closed. • During inspiration the patients gets ventilated with FG and part of the FG is vented through the valve after sufficient pressure has developed to open the valve. • During expiration , the FG from the machine flows into the reservoir bag and all the expired gas ( i.e. dead space and alveolar gas flows back into the corrugated tube till the system is full.
  • 7. • During the next inspiration the alveolar gas is pushed back into the alveoli followed by the fresh gas. • When sufficient pressure is developed, part of the expired gas and part of the FG escape through the valve. • This leads to considerable rebreathing as well as excessive wastage of fresh gas . • Hence these system are inefficient for controlled ventilation.
  • 8. Advantages: • Best among all Maplesons systems for spontaneous ventilation • Minimal wastage of gases during spontaneous ventilation Disadvantages: • Not efficient for controlled ventilation. • Wastage of gases & operation theatre pollution. • Expiratory valve required –produces slight resistance during expiration. • Expiratory valve is heavy(Headbrink valve). • Expiratory valve is near patient end inconvenient to use especially in head & neck surgeries. • Not suitable for paediatric use.
  • 9. Mapleson A – Lack Modification • A co-axial modification of the Mapleson A system. • Designed to facilitate scavenging of expired gas & make more efficient for controlled ventilation • 1.6 m in length • FGF through outside tube ( 30mm), exhaled gases from inner tube. • Inner tube wide in diameter (14 mm) to reduce resistance to expiration(1.6 cm H2O). • Reservoir bag at machine end • APL valve at machine end.
  • 10. • The Lack circuit is essentially similar in function to the Magill, except that the expiratory valve is located at the machine-end of the circuit, being connected to the patient adapter by the inner coaxial tube.
  • 11. • Inspiration - The valve closes and the patient inspires fresh gas from the outer reservoir tube. • Expiratory pause - Fresh gas washes the expired gas out of the reservoir tube, filling it with fresh gas for the next inspiration. • Expiration - The patient expires into the reservoir tube. Towards the end of expiration, the bag fills and positive pressure opens the valve, allowing expired gas to escape via the inner exhaust tube
  • 12. Advantages • The location of the valve is more convenient, facilitating intermittent positive pressure ventilation and scavenging of expired gas. Disadvantages • In common with other co-axial systems, if the inner tube becomes disconnected or breaks, the entire reservoir tube becomes dead-space. • This can be avoided by use of the 'parallel Lack' system, in which the inner and outer tubes are replaced by conventional breathing tubing and a Y-piece:
  • 13. Mapleson B System • The Mapleson B system features the fresh gas inlet near the patient, distal to the expiratory valve. • The expiratory valve opens when pressure in the circuit increases, and a mixture of alveolar gas and fresh gas is discharged. • During the next inspiration, a mixture of retained fresh gas and alveolar gas is inhaled. • Rebreathing is avoided with fresh gas flow rates of greater than twice the minute ventilation for both spontaneous and controlled ventilation.
  • 14. Mapleson C system • This circuit is also known as Water’s circuit. • It is similar in construction to the Mapleson B , but the main tube is shorter. • A FGF equal to twice the to twice the minute ventilation is required to prevent rebreathing. • CO2 builds up slowly with this circuit. • Mapleson B &C : In order to reduce rebreathing of alveolar gas FG entry was shifted to near the patient. • This allows a complete mixing of FG and expired gas. • The end result is that these system are neither efficient during spontaneous nor during controlled ventilation.
  • 15. Mapleson D System • It consists of fresh gas inlet nearer the patient end , a corrugated rubber tubing one end which is connected with expiratory valve and then reservoir bag. • It is mainly used for assisted or controlled ventilation • During the controlled ventilation there is little chance of rebreathing. • The FGF which enters during expiratory pause accumulates in the patient end is forced during the inflation.
  • 16. • In spontaneous breathing during inspiration the patient will inhale the fresh gas & gas in corrugated tube depending on FGF, TV, length of expiratory pause & volume of corrugated tube. • Rebreathing can be minimized by increasing FGF 2-3 times the MV. • For an adult 15L/min FGF which seems uneconomical is required. • In some cases 250 ml/kg/min required to prevent rebreathing.
  • 17. Bain circuit • In principal it is modification of the system used by Macintosh and Pask during Second World War to administer anaesthesia. • It was introduced by Bain and Spoerel in 1972 • It is a modification of Mapleson D system. • It is a coaxial system in which fresh gas flows through a narrow inner tube within outer corrugated tubing • It functions like T-piece except that tube supplying FG to the patient is located inside the reservoir tube.
  • 18. Specifications:- • Length-1.8 meters. • Diameter of outer tube-22mm(transparent,carries expiratory gases) • Diameter of inner tubing-7 mm(inspiratory) • Resistance-Less than0.7 cmH2O • Dead space-Outer tube upto expiratory valve( around 500ml=TV) • Flow rates-100-150 ml/kg/min for controlled ventilation. Average 300 ml/kg/min for spontaneous ventilation
  • 19. Bain system (Mapleson D)- Functional Analysis Spontaneous respiration: • The breathing system should be filled with FG before connecting to the patient. • When the patient takes an inspiration, the FG from the machine , the reservoir bag and the corrugated tube flow to the patient.
  • 20. • During the expiration there is a continuous FGF into the system at the patients end .The expired gas gets continuously mixed with the FG as it flows back into corrugated tubing and the reservoir bag • Once the system is full the excess gas is vented to the atmosphere through the valve situated at the end of the corrugated tube near the reservoir bag. • During the expiratory pause the FG continues to flow and fill the proximal portion of the corrugated tube while mix gas is vented through valve. • During the next inspiration , the patient breathes FG as well as mixed gas from the corrugated tube. • It is calculated and clinically prove that FGF should at least 1.5- 2 times the patient’s MV
  • 21. • Controlled ventilation : • To facilitate intermittent positive pressure ventilation, the expiratory valve has to be partly closed so that it opens only after sufficient pressure has developed in the system. • When the system is filled with fresh gas, the patient gets ventilated with the FGF from the machine, corrugated tubing and the reservoir bag.
  • 22. • During expiration expired gas continuously gets mixed with FG that is flowing into the system at the patient end. • During the expiratory pause the FG continues to enter the system and pushes the mixed gas towards the reservoir. • When next inspiration is initiated , the patient gets ventilated with the gas in the corrugated tube i.e.a mixture of FG, alveolar gas and dead space gas. • As the pressure in the system increases,APL valve open and the contents of reservoir bag are discharged into the atmosphere.
  • 23. ADVANTAGES OF BAIN'S SYSTEM • Light weight. • Minimal drag on ETT as compared to Magill's circuit. • Low resistance. • As the outer tube is transparent, it is easy to detect any kinking or disconnection of the inner fresh gas flow tube. • It can be used both during spontaneous and controlled ventilation and change over is easier. • It is useful where patient is not accessible as in MRI suites. • Exhaled gases do not accumulate near surgical field, so risk of flash fires is abolished. • Easy for scavenging of gases as scavenging valve is at machine end of the circuit. • Easy to connect to ventilator. • There is some warming of the inspired fresh gas by the exhaled gas present in outer tubing.
  • 24. DISADVANTAGES OF BAIN'S SYSTEM • Due to multiple connections in the circuit there is a risk of disconnections. • Wrong assembling of the parts can lead to malfunction of the circuit. • Theatre pollution occurs due to high fresh gas flow. However, it can be prevented by using scavenging system. • Increases the cost due to high fresh gas flows. • There can be kinking of the fresh gas supply inner tube blocking the fresh gas supply leading to hypoxia • There can be crack in the inner tube causing leakage • Case report available about the defect in metal head so that fresh gas and exhaled gas mix and entire limb becomes dead space • It cannot be used in paediatric patients with weight less than 20 kg.
  • 25. Checking the circuit •Mapleson D system is checked for leaks by occluding the patient end, closing the APL valve and pressurizing the system. The APL valve is then opened. The bag should deflate easily if the valve is working properly. Outer tube integrity should also be checked by following the simplest innovative method. Wet the hands with spirit. Blow air through the tube. Wipe the tube with wet hands. Leak will produce chillness in the hands. •For checking integrity of inner tube of Bains system, a test is performed by setting a low flow on the oxygen flowmeter and occluding the inner tube with a finger or barrel of a small syringe at the patient end while observing the flow meter indicator. If the inner tube is intact and correctly connected, the indicator will fall. •Pethicks test - To check the integrity of the inner tube, activate the oxygen flush and observe the bag. Due to venturi effect the high flow from the inner tube at the patient end will create a negative pressure in the outer exhalation tubing and this will suck gas from the bag and bag will deflate. If the inner tube is not intact, this maneuver will cause the bag to inflate slightly.
  • 26. Mapleson E and F • Valveless breathing system used for children upto 30 kg. • Suitable for spontaneous and controlled ventilation • Components:- • - T shaped tubing with 3 ports. • -FGF delivered to one port • -2 nd port goes to patient & 3rd to reservoir tube.
  • 27. Mapleson E OR Ayre’s T- PIECE • Introduced by Phillips Arye in 1937. • Belongs to Mapleson E. • Available as meatllic / plastic. • Length – 2 inches. • Parts – inlet, outlet, side tube. • Inlet size-10 mm, outlet size-10mm metallic &15 mm plastic
  • 28. Advantages • Simple to use , Light weight . • No dead space , no resistance. • For pediatric patients less than 20 kgs. • Expiratory limb is attached to the outlet of T piece.- • It should accommodate air space equal to 1/3 rd of TV. • If too short – air dilution in spont. Breathing patients & patient become light. • 1 inch of expiratory tube can accommodate 2-3 ml of gas. • Fresh Gas Flows – 2- 3 times MV
  • 29. Dis advantages: • High flow rates are required. • Loss of heat & humidity. • Risk of accidental occlusion of expiratory limb- risk of increased airway pressure & barotrauma to lungs.
  • 30. Mapleson F OR Jackson Rees Circuit • It is a modification of Mapleson E by Jackson Rees and is known as Jackson Rees modification. • It has a 500 ml bag attached to the expiratory limb. • This bag helps in respiratory monitoring or assisting the respiration. • It also helps in venting out excess gases. • The bag has a hole in the tail of the bag that is occluded by using a finger to provide pressure. • The bags with valve are also available. • It is used in neonates, infants, and paediatric patients less than 20 kg in weight or less than 5 years of age.
  • 31. • Technique of use For spontaneous respiration: The relief mechanism of the bag is left fully open. • For controlled respiration: The hole in the bag can be occluded by the user during inspiration and ventilation is done by squeezing the bag.
  • 32. • Functional analysis • It also functions like Mapleson D system. • The flows required to prevent rebreathing are 2.5-3.0 times minute volume during spontaneous ventilation and 1.5 to 2 times the minute volume during controlled ventilation. • In this system, during expiration fresh gas and exhaled gas will collect and mix in the bag. • The next inspiration results in patient inhaling fresh gas both direct from inlet and from expiratory part of the circuit as in Mapleson E.
  • 33. • During expiratory pause the expired gases are replaced by fresh gas in the expiratory limb. • Observation of bag movements helps in assessing respiration during spontaneous breathing. • It also allows controlled ventilation by squeezing the bag. • Heat and moisture exchanger should not be used with Mapleson E and F during spontaneous respiration as it increases resistance. • So most of the fresh gas will enter expiratory limb leading to wastage of fresh gases and delaying induction by inhalation agents.