http://www.ppt-to-video.com Presenter     : Dr Unnikrishnan P Coordinator : Dr C MadhusoodhananPillai Moderators  : Dr SheelaRani                         Dr Ravi Vaporizers : 1
1.DO YOU FEEL YOU NEED IT? 2.SOME IMPORTANT             THINGS! 3.CLASSIFICATION 4.FACTORS AFFECTING VAPORIZATION 5.ASTM VAPORIZER STANDARDS 6.FILLING SYSTEMS 7.HAZARDS[lengthy class..is it a hazard?] 8.EVOLUTION
Why we need a vaporizer?  Most volatile agents exist as liquid at room temp & atm pressure Vaporizers convert this liquid form to vapor phase  And add a certain amount of this vapor to the anesthesia circuit... In precisely determined concentrations over a wide range of temperatures, pressures and carrier gas flowrates.
Who is inside the vaporizer? Let me introduce..I am a gas Below a certain temp,I am very flexible ; then I will exist in a form which is interconvertable between the liquid and gaseous states- this gaseous form is known as vapor. But above this temp, I am very rigid ; you cannot change me into a liquid now..yes now I am a perfect gas And I wont tell u the name of that temperature!
SO VAPOR IS THE GASEOUS PHASE OF A SUBSTANCE BELOW ITS CRITICAL TEMPERATURE Love’ [van der Waals forces] promotes the liquid state, whereas ‘hatred’ [ Kinetic energy] promotes vapor phase
Vapor Pressure This vapor exerts a pressure on its surroundings- VAPOR PRESSURE V.P.depends only on the liquid and temperature
Saturated Vapor Pressure Now we close the container;the escaping ‘vapor molecules’ now may collide with the liquid surface and reenter it..an equilibrium forms No of molecules reentering = no of molecules leaving SATURATED VAPOR PRESSUREis the partial pressure of the vapor phase of a substance when at equilibrium with its liquid phase (e.g. in a closed container). Increases rapidly as boiling point approaches.
Boiling Point You go on increasing the temperature The V.P. will increase At a point it equals the atm pressure B.P.OF A LIQUID IS THE TEMPERATURE AT WHICH ITS VAPOR PRESSURE IS EQUAL TO THE ATM PRESSURE Low atm pressure -> Low B.P. High SVP -> Low B.P.[..means I    am very volatile]
PARTIAL PRESSURE A MIXTURE OF GASES IN A CLOSED CONTAINER WILL EXERT A PRESSURE ON THE WALLS OF THE CONTAINER. THE PART OF THE PRESSURE EXERTED BY ANY ONE GAS IN THE MIXTURE IS CALLED ITS PARTIAL PRESSURE DEPENDS ONLY ON THE TEMPERATURE THE HIGHEST PARTIAL PRESSURE THAT CAN BE EXERTED BY A GAS AT A GIVEN TEMPERATURE IS ITS VAPOR PRESSURE. IT IS AN ABSOLUTE VALUE.
VOLUMES PERCENT NO OF UNITS OF VOLUME OF A GAS IN RELATION TO A TOTAL OF 100 UNITS OF VOLUME FOR THE TOTAL GAS MIXTURE.  IT EXPRESSES THE RELATIVE RATIO OF GAS MOLECULES IN A MIXTURE. MOST COMMONLY USED BUT UPTAKE & DEPTH OF ANESTHESIA ARE DIRECTLY RELATED TO PARTIAL PRESSURES  PARTIAL PRESSURE/TOTAL PRESSURE= VOLUMES PERCENT/100
HEAT OF VAPORIZATION THE NUMBER OF CALORIES REQUIRED TO CONVERT 1GM OF LIQUID INTO A VAPOR OR NUMBER OF CALORIES REQUIRED TO CONVERT 1 ML OF LIQUID INTO A VAPOR IN A VAPORIZER, AS THE LIQUID AGENT VAPORIZES,HEAT IS LOST;SO TEMPERATURE DROPS,VAPOR PRESSURE DROPS = LESSER  VOL. AGENT FOR THE CARRIER GAS TO TAKE AWAY↓OUTPUT Oh…tell me some method to supply heat! Else my patient will wake up @#%?
SPECIFIC HEAT THE QUANTITY OF HEAT REQUIRED TO RAISE THE TEMP OF 1gm OF THE SUBSTANCE BY 1⁰C IF ITS LOW, THE TEMPERATURE WILL NOT RISE GRADUALLY WHEN WE SUPPLY HEAT TO A VOLATILE AGENT IN ORDER TO REPLACE THAT LOST DUE TO VAPORIZATION IF ITS HIGH...NICE GUY..VERY OBEDIENT! IS THIS IMPORTANT IN ANY OTHER ASPECT? MATERIAL FROM WHICH VAPORIZER CONSTRUCTED
THERMAL CAPACITY AMOUNT OF HEAT STORED IN THE VAPORISER BODY. IT IS THE PRODUCT OF SPECIFIC HEAT AND MASS.  A VAPORIZER CONSTRUCTED FROM A SUBSTANCE WITH HIGH THERMAL CAPACITY WILL CHANGE TEMP MORE SLOWLY AND SO IS PREFFERED
THERMAL CONDUCTIVITY SPEED WITH WHICH HEAT FLOWS THROUGH A SUBSTANCE; CU > AL > BRASS > STEEL >> GLASS. HIGHER THE THERMAL CONDUCTIVITY, BETTER THE SUBSTANCE CONDUCTS HEAT.
THERMOSTABILIZATION IS ACHIEVED BY CONSTRUCTING VAPORIZERS OF METALS WITH HIGH THERMAL CONDUCTIVITY (COPPER, BRONZE) TO MINIMIZE TEMPERATURE CHANGES WHEN THE VAPORIZER IS IN USE WICKS SHOULD BE IN CONTACT WITH THE METAL PART
GAS LAWS BOYLES LAW:AT CONSTANT TEMPERATURE, THE VOLUME OF A GIVEN MASS OF GAS VARIES INVERSELY WITH THE ABSOLUTE PRESSURE CHARLE’S LAW: AT CONSTANT PRESSURE, THE VOLUME OF A GIVEN MASS OF GAS VARIES DIRECTLY WITH THE ABSOLUTE TEMPERATURE AT CONSTANT VOLUME, THE ABSOLUTE PRESSURE OF A GIVEN MASS OF GAS VARIES DIRECTLY WITH THE ABSOLUTE TEMPERATURE
DALTON’S LAW OF PARTIAL PRESSURES IN A MIXTURE OF GASES THE PRESSURE EXERTED BY EACH GAS IS THE SAME AS THAT WHICH IT WOULD EXERT IF IT ALONE OCCUPIED THE CONTAINER
AVOGADRO’S HYPOTHESIS EQUAL VOLUMES OF GASES AT THE SAME TEMPERATURE AND PRESSURE CONTAIN EQUAL NUMBER OF MOLECULES THIS NO IS 6.023 X 10²³ THIS MUCH NO OF PARTICLES OF ANY GAS AT STP WILL OCCUPY 22.4 Ls
EXAMPLE : CALIBRATION MOLECULAR WT OF ISOFLURANE 184.5 CONTAIN AVOGADRO NO OF PARTICLES; WILL OCCUPY 22.4 Ls @ STP SUPPOSE O₂VAPORIZES 18.45 GM OF ISOFLURANE INTO 224 L =0.1 MOLE = WILL OCCUPY 2.24Ls BUT THIS IS VAPORIZED INTO 224 Ls SO CONCENTRATION OF ISOFLURANE = 2.24/224 = 1%
Q: WHICH LAW IS REPRESENTED HERE?
CLASSIFICATION OF    VAPORIZERS
METHOD OF REGULATING THE OUTPUT CONCENTRATION CONCENTRATION CALIBRATED  MEASURED FLOW
METHOD OF VAPORIZATION FLOW OVER BUBBLE THROUGH INJECTION SIMPLE
TEMPERATURE COMPENSATION THERMOCOMPENSATION SUPPLIED HEAT
SPECIFICITY AGENT SPECIFIC MULTIPLE AGENTS
POSITION VAPORIZER INSIDE CIRCUIT VAPORIZER OUTSIDE CIRCUIT
RESISTANCE PLENUM LOW RESISTANCE/DRAW OVER
		BASIC DESIGN
METHOD OF REGULATING OUTPUT CONCENTRATION
CONCENTRATION CALIBRATED TOTAL FLOW FROM THE MACHINE PASSES THROUGH THE VAPORIZER. THIS IS SPLIT BY A VARIABLE RESISTANCE PROPORTIONATING VALVE INTO TWO: ONE PART[USUALLY MAJOR] FLOWS THROUGH THE BYPASS CHAMBER & THE OTHER [USUALLY SMALL] THROUGH THE VAPORIZING CHAMBER.
CONCENTRATION CALIBRATED AGENT CONCENTRATION IS CONTROLLED BY A DIAL CALIBRATED IN VOLUMES PERCENT VOLUME VAPOURISED - TYPICALLY 200 ML VAPOUR PER ML OF LIQUID ANAESTHETIC ML LIQUID USED/HOUR = 3 X % X FGF
IF WE DIAL A HIGH CONCENTRATION
IF WE DIAL A LOW CONCENTRATION
	SPLITTING RATIO THE RATIO OF THE BYPASS GAS TO THE GAS GOING TO THE VAPORIZING CHAMBER IS CALLED THE SPLITTING RATIO.  SPLITTING RATIO DEPENDS ON:   THE RESISTANCE OF THE TWO PATHWAYS, WHICH INTURN DEPENDS ON THE VARIABLE ORIFICE OF THE INLET/OUTLET. TEMPERATURE OF THE LIQUID /CARRIER GAS. FLOW RATE OF GASES DENSITY OF THE PARTICULAR GAS ENTERING VAP CHAMBER. MOSTLY CARRIER GAS IS O₂
MEASURED FLOW VAPORIZERS THE VAPORIZER HEATS THE ANESTHETIC AGENT TO A TEMPERATURE ABOVE ITS BOILING POINT( SO IT BEHAVES AS GAS ) AND THIS IS THEN METERED INTO THE FRESH GAS FLOW.  A MEASURED FLOW IS SENT BY A SEPARATE OXYGEN FLOW METER TO PASS TO THE VAPORIZER WITH THE OUTPUT BEING AT SVP FOR THE ANESTHETIC AGENT.  IN ORDER TO DILUTE THIS OTHERWISE LETHAL CONCENTRATION, OUTPUT FROM THAT FLOWMETER IS COMBINED WITH GAS PASSING FROM THE MAIN FLOWMETER e.g. COPPER KETTLE, VERNI-TRIOL
MEASURED FLOW VAPORIZERS OPERATOR HAS TO SET THE FLOW TO THE VAPORIZER AND BYPASS WITH SEPARATE FLOWMETERS  THIS MEANS THAT RESPECTIVE FLOWS HAVE TO BE CALCULATED FOR EACH AGENT FOR A GIVEN TEMPERATURE AND VAPOUR OUTPUT.
VAPORIZATION METHODS
FLOW OVER HERE THE CARRIER GAS PASSES OVER THE SURFACE OF A LIQUID. SURFACE AREA OF VAPORISATION CAN BE INCREASED BY USING WICKS.
		FLOW OVER
		FLOW OVER DUE TO CAPILLARY ACTION, THE ANAESTHETIC AGENT RISES INTO THE WICKS. THIS DRAMATICALLY INCREASES THE SURFACE AREA OF ANAESTHETIC AGENT EXPOSED TO THE FRESH GAS AND THEREBY IMPROVES THE EFFICIENCY OF VAPORISATION. CARRIER GAS CAN BE DIRECTED USING BAFFLES OR SPIRAL TRACKS THAT LENGTHEN THE GAS PATHWAYS OVER THE LIQUID.THIS INCREASES THE TIME AND AREA OF CONTACT.
		FLOW OVER
BUBBLE THROUGH CARRIER GAS IS ALLOWED TO BUBBLE THROUGH THE LIQUID.  SMALLER THE BUBBLE ,LARGER WILL BE THE SURFACE AREA.  METHODS TO BREAK THE GAS INTO SMALL BUBBLES ARE AGITATION & SPLASHING. FACTORS INFLUENCING THE VAPORIZER OUTPUT : SIZE OF THE BUBBLE, DEPTH OF LIQUID, BUBBLING SPEED
INJECTION VAPORIZERS A KNOWN AMOUNT OF LIQUID AGENT OR PURE VAPOUR IS INJECTED INTO THE GAS STREAM TO PROVIDE THE DESIRED CONCENTRATION E.G. TEC 6 DESFLURANE VAPORIZER
TEMPERATURE COMPENSATION TO MAINTAIN A CONSTANT O/P FROM THE VAPORIZER, MECHANISMS TO COMPENSATE FOR THE FLUCTUATIONS IN TEMPERATURE ARE TO BE EMPLOYED
TEMPERATURE COMPENSATION
THERMOCOMPENSATION MOST VARIABLE BYPASS VAPORIZERS COMPENSATE FOR CHANGES IN VAPOR PRESSURE BY ALTERING THE SPLITTING RATIO. DONE BY USING A THERMOSENSITIVE ELEMENT [BIMETALLIC STRIP] INCORPORATED IN THE VAPORIZING CHAMBER OR BYPASS CHAMBER. SO, THE SPLITTING OF GAS IS CONTROLLED BY TWO VALVES: (1)THE DIAL WE SET (SPLITTING VALVE) AND (2)THE TEMPERATURE COMPENSATING VALVE
THERMOCOMPENSATION IN A BIMETALLIC STRIP, TWO METALS WITH VERY DIFFERENT COEFFICIENTS OF THERMAL EXPANSION ARE FIXED TOGETHER.
THERMOCOMPENSATION WHEN THE TEMPERATURE OF THE VAPORISING CHAMBER DROPS, THE BIMETALLIC STRIP BENDS AND MOVES AWAY.  THIS REDUCES THE RESISTANCE TO FLOW AND THUS MORE FLOW OCCURS INTO THE VAPORISING CHAMBER.
SUPPLIED HEAT AN ELECTRIC HEATER CAN BE USED TO SUPPLY HEAT TO A VAPORIZER AND MAINTAIN IT AT A CONSTANT TEMPERATURE.E.G.TEC 6
		RESISTANCE
PLENUM [Latin= fullness] VAPORIZERS WITH HIGH RESISTANCE WHICH DEPEND ON COMPRESSED GAS DRIVEN UNDER PRESSURE ARE CALLED PLENUM VAPORIZERS. e.g. BOYLE BOTTLE,COPPER KETTLE
DRAW OVER CARRIER GAS IS DRAWN THROUGH THE VAPORISER EITHER BY THE PATIENT’S OWN RESPIRATORY EFFORTS, OR BY A SELF-INFLATING BAG OR MANUAL BELLOWS DRAWOVER SYSTEMS OPERATE AT LESS THAN, OR AT AMBIENT PRESSURE FLOW THROUGH THE SYSTEM IS INTERMITTENT, VARYING WITH DIFFERENT PHASES OF INSPIRATION, AND CEASING IN EXPIRATION. A ONE-WAY VALVE PREVENTS REVERSE FLOW IN THE CIRCUIT.  E.G. EMO
DRAW OVER HAVE A LOW INTERNAL RESISTANCE TO GAS FLOW  SO THAT THEY MAY BE USED WITHIN THE BREATHING CIRCUIT,  THE GAS FLOW BEING DRIVEN THROUGH THEM BY THE PATIENT'S BREATHING. THEY MAY BE USED IN A NON-REBREATHING DRAW-OVER APPARATUS, OR AS IN-CIRCUIT VAPORIZERS IN A CIRCLE ABSORBER SYSTEM.     ARE USUALLY OF THE SIMPLE BOTTLE OR BOWL TYPE (E.G. THE OHIO 8 BOTTLE, GOLDMAN OR STEPHENS),  BUT MORE ACCURATE, LOW RESISTANCE DRAW-OVER VAPORIZERS, SUCH AS THE PENLON OXFORD MINIATURE VAPORIZER (OMV) AND OHMEDA DRAW-OVER TEC VAPORIZERS, HAVE BEEN MANUFACTURED.
FACTORS AFFECTING VAPORIZER OUTPUT 1. FLOW THROUGH THE VAPORIZING CHAMBER: Varying the proportion of gas passing through the vaporizing chamber and bypass chamber 2.SURFACE AREA OF THE LIQUID GAS INTERFACE: Greater the surface area, more will be the vaporization. Bubble through > flow over
FACTORS AFFECTING VAPORIZER OUTPUT 3.TEMPERATURE: As temperature increase, output increase 4.TIME: Output concentration tend to fall over time
FACTORS AFFECTING VAPORIZER OUTPUT 5.GAS FLOW RATE: (A)At high flowrates,the gas leaving vaporization chamber is  less saturated    	              (B)Alters the total flow that passes through the vaporization chamber
FACTORS AFFECTING VAPORIZER OUTPUT 6. CARRIER GAS COMPOSITION:  		(a) changes in viscosity & density may affect the proportion of the total flow passing through the vaporization chamber 	 	(b)N₂O dissolves in the flow, thus altering the effective volume passing through the vaporization chamber
FACTORS AFFECTING VAPORIZER OUTPUT 7.BOILING POINT :HIGHER THE BOILING POINT,LESS WILL BE THE VAPOUR O/P. 8. AMBIENT PRESSURE : SVP IS SOLELY A FUNCTION OF TEMP. SO IF AMBIENT PRESSURE IS REDUCED,THE CONSTANT SVP BECOMES A GREATER PROPORTION OF THE TOTAL PRESSURE OUTPUT INCREASES. AGENTS WITH LOW BOILING POINTS ARE MORE SUSCEPTIBLE TO THE INFLUENCE OF AMBIENT PRESSURE
EFFECT OF LOW ATMOSPHERIC PRESSURE
CONCENTRATION CALIBRATED HIGH RESISTANCE PATHWAY THROUGH THE VAPORISING CHAMBER OFFERS LESS RESISTANCE, UNDER HYPOBARIC CONDITIONS AND SO A SLIGHT INCREASE IN VAPOR O/P OCCURS.
MEASURED FLOW HERE THE DELIVERED PARTIAL PRESSURE & VOLUME PERCENT INCREASES. AMOUNT OF INCREASE DEPENDS ON THE BAROMETRIC PRESSURE & THE VAPOR PRESSURE OF THE AGENT. CLOSER THE VAPOR PRESSURE IS TO BAROMETRIC PRESSURE, GREATER THE EFFECT.
EFFECT OF HIGH ATMOSPHERIC PRESSURE
CONCENTRATION CALIBRATED ATM PRESSURE INCREASESDENSITY OF GAS CHANGESMORE RESISTANCE TO FLOW OF GAS THROUGH THE VAPORIZING CHAMBERDECREASED VAPOUR O/P (PARTIAL PRESSURE & VOLUME PERCENT)  EFFECT ON PARTIAL PRESSURE IS LESS DRAMATIC
MEASURED FLOW LOWER CONCENTRATION IN TERMS OF PP/VOLUME PERCENT
EFFECT OF INTERMITTENT BACK PRESSURE
EFFECT OF INTERMITTENT BACK PRESSURE WHEN ASSISTED OR CONTROLLED VENTILATION  IS USED ,THE POSITIVE PRESSURE GENERATED DURING INSPIRATION IS TRANSMITTED FROM THE BREATHING SYSTEM BACK TO THE MACHINE & SOME WAY MAY BE TRANSMITTED TO THE VAPORIZERS. ALSO SEEN WITH THE USE OF OXYGEN FLUSH 2 EFFECTS ARE SEEN:
PUMPING EFFECT THE  INCREASE IN VAPORIZER OUTPUT CONCENTRATION DUE TO BACK PRESSURE
PUMPING EFFECT When the bag is squeezed ,pressure is transmitted back into both, the "by pass" channel and also to the vaporising chamber.  The fresh gas tries to move forward and gets compressed both in the 'by pass' channel and the vaporising chamber.  However, the vaporising chamber volume is much larger than the 'by pass' channel volume, and thus, more fresh gas gets compressed into it than into the 'by pass' channel.
PUMPING EFFECT  This extra fresh gas that enters the vaporising chamber collects anaesthetic vapor
PUMPING EFFECT when the positive pressure is suddenly released (expiration) the previously compressed gases now suddenly expands in all directions.   Some of the rapidly expanding gas (containing vapor) enter the inlet of the vaporiser and cross over into the 'by pass' channel
PUMPING EFFECT Normally, a vaporiser 'by pass' channel does not have vapor.  So this vapor due the 'pumping effect' is additional.  When this 'by pass' vapor flows across to the exit, it meets the vapor from the vaporising chamber.  The addition of the 'by pass' vapor to the vapor from the vaporising chamber raises the final concentration of anaesthetic delivered.  i.e. The 'pumping effect' increases the delivered concentration of anaesthetic agent.
PUMPING EFFECT SEEN  ESPECIALLY WHEN CARRIER GAS IS LOW AGENT  IN  VAPORIZING CHAMBER IS LOW DIAL SETTING IS LOW PRESSURE    FLUCTUATIONS ARE HIGH & FREQUENT.
PUMPING EFFECT MECHANISM IN MEASURED FLOW: GAS FLOW TO THESE VAPORIZERS BECOME SATURATED WITH VAPOUR & IS JOINED BY GAS FROM OTHER FLOWMETER,WHICH DILUTES ITS CONCENTRATION. WHEN BACK PRESSURE IS APPLIED,THERE IS RETROGRADE FLOW OF GAS SO THAT THE DILUTED GAS MIXTURE IS FORCED BACK INTO THE VAPORIZER.BECAUSE THIS GAS IS NOT SATURATED ,IT WILL THEN PICK UP ANESTHETIC VAPOUR.THE RESULT IS AN INCREASE IN O/P.
MODIFICATIONS TO MINIMISE PUMPING EFFECT KEEP  THE  VAPORIING   CHAMBER  SMALL  OR INCREASING  THE   SIZE   OF  THE   BYPASS CHAMBER .
MODIFICATIONS TO MINIMISE PUMPING EFFECT ADD LONG SPIRAL OR LARGE DIAMETER TUBE TO LEAD TO THE VAPORIZER CHAMBER.THE EXTRA GAS FORCED INTO THIS TUBE  & SUBSEQUENTLY RETURNED TO THE BYPASS DOES NOT REACH THE VAPORIZING CHAMBER
MODIFICATIONS TO MINIMISE PUMPING EFFECT INCREASE RESISTANCE TO GAS FLOW THROUGH THE VAPORIZER.
MODIFICATIONS TO MINIMISE PUMPING EFFECT CHECK VALVE TO PREVENT BACKWARD FLOW OF GAS
MODIFICATIONS TO MINIMISE PUMPING EFFECT EXCLUDE WICKS FROM THE AREA WHERE THE INLET TUBE JOINS THE VAPORIZING CHAMBER. OUTLET TUBE MAY BE MADE LONGER SO THAT UNSATURATED GAS WILL HAVE TO PASS FURTHER BACK BEFORE PICKING UP ANESTHETIC VAPOUR. CONNECTIONS  OF  OXYGEN  FLUSH  VALVE  LINE TO  THE  COMMON  GAS   OUTLET  BE DESIGNED  TO  MINIMISE  PRESSURE FLUCTUATIONS  THAT  MAY  PRODUCE  A PUMPING  EFFECT LIMIT  PRESSURE   TRANSMITTED   TO VAPORIZER   TO  <10KPa   ABOVE   NORMAL WORKING   PRESSURE   ,CONC   NOT  TO INCREASE   >20%
PRESSURIZING EFFECT THE O/P OF SOME VAPORIZER USED IN CONJUNCTION WITH AUTOMATIC VENTILATOR HAS BEEN FOUND TO BE LOWER THAN DURING FREE FLOW TO ATMOSPHERE. MOSTLY SEEN WHEN HIGH FLOW LARGE PRESSURE FLUCTUATIONS LOW DIAL SETTINGS
PRESSURIZING EFFECT No of mol.s of the agent picked by carrier gas ∝ density of vapor mol.s in vaporizing chamber ∝ vapor pressure   Increased pressure in vaporizer chamber compress carrier gas more mol.s/mL No of mols of the anesthetic agent same o/p ↓ed The Changes In Vaporizer O/P Caused By The Pumping Effect Usually Are Greater In Magnitude Than Those Associated With The Pressurizing Effect
1988 ASTM MACHINE STANDARDS FOR VAPORIZERS A VAPORIZER MUST BE CAPABLE OF ACCEPTING A TOTAL GAS FLOW OF 15 lts/min FROM THE ANESTHESIA MACHINE AND IN TURN, DELIVERING A GAS FLOW WITH A  PREDICTABLE CONC OF VAPOUR. THE EFFECTS OF THE CONDITIONS OF USE(VARIATIONS IN TEMP,PRESSURE,BACKPRESSURE,INPUT FLOW RATES)ON VAPORIZER PERFORMANCE MUST BE STATED IN CATALOGS.THE EFFECT OF CARRIER GAS COMPOSITION ON VAPORIZER O/P SHOULD ALSO BE SUPPLIED.
1988 ASTM MACHINE STANDARDS FOR VAPORIZERS THE EXTENT TO WHICH TEMP & INFLOW RATES INFLUENCE THE VAPORIZER CONC MUST BE STATED. VAPORIZER MUST BE DESIGNED SO THAT IT CANNOT BE OVERFILLED WHEN IN NORMAL OPERATING POSITION. A SYSTEM THAT ISOLATES THE VAPORIZER FROM EACH OTHER & PREVENTS GAS FROM PASSING THROUGH THE VAPORIZER CHAMBER OF ONE VAPO & THEN THROUGH THAT OF ANOTHER MUST BE PROVIDED CONTROL MUST BE PROVIDED TO LIMIT THE ESCAPE OF ANESTHETIC VAPOUR FROM THE VAP CHAMBER INTO THE FRESH GAS ,SO  THAT THE DELVERED CONC IS <0.1% WHEN THE VAPORIZER IS TURNED OFF
1988 ASTM MACHINE STANDARDS FOR VAPORIZERS ALL THE VAPORIZER KNOB MUST TURN COUNTER CLOCKWISE TO INCREASE  THE CONC. VAPORIZER MUST BE PROVIDED WITH A LIQUID LEVEL INDICATOR VISIBLE FROM THE FRONT OF THE MACHINE THE VAPORIZER MUST PERMIT MAXIMAL CALIBRATED  FLOWS OF O2&N20 IN AN ON & OFF POSITIONS WITH THE VAPORIZER FILLED TO THE MAXIMUM SAFE INDICATED LEVEL WITHOUT DISCHARGING LIQUID THROUGH ITS OUTLET WHEN IT IS MOUNTED & USED IN ACCORDANCE  WITH THE MANUFACTURERS INSTRUCTIONS
1988 ASTM MACHINE STANDARDS FOR VAPORIZERS VAPORIZER UNSUITABLE FOR USE IN THE BREATHING SYSTEM MUST ‘VE NONINTERCHANGEABLE 23MM FITTINGS.THE 22MM OR 15MM FITTINGS CANNOT BE USED.THE INLET OF THE VAPORIZER MUST BE MALE,OUTLET MUST BE FEMALE & THE DIRECTION OF GAS FLOW MUST BE MARKED WHEN 23MM FITTINGS ARE USED. VAPORIZER  SUITABLE FOR USE IN THE BREATHING SYSTEM MUST ‘VE STANDARD 22MM FITINGS OR SCREW THREADED,WEIGHT BEARING FITTINGS WITH INLET FEMALE & OUTLET MALE.THE INLET &OUTLET PORTS MUST BE MARKED,THE DIRECTION OF GAS FLOW MUST BE INDICATED BY ARROWS & VAPORIZER MUST BE MARKED “FOR USE IN BREATHING SYSTEM”.
1988 ASTM MACHINE STANDARDS FOR VAPORIZERS ALL VAPORIZER LOCATED IN THE FRESH GAS CIRCUIT BE CONCENTRATION CALIBRATED  & THAT THE CONTROL OF VAPOUR CAN BE PROVIDED BY MEANS OF CALIBRATED KNOBS OR DIALS.THE MEASURED FLOW SYSTEMS NOT MENTIONED IN THE STANDARDS ARE THEREFORE NOW CONSIDERED BY MOST TO BE OBSOLETE ,ALTHOUGH MILITARY FIELD ANESTHESIA MACHINE STILL USE SUCH SYSTEM
AGENT SPECIFIC FILLING SYSTEMS
AGENT SPECIFIC FILLING SYSTEMS  A VAPORIZER DESIGNED FOR A SINGLE AGENT BE FITTED WITH A PERMANENTLY ATTACHED AGENT SPECIFIC DEVICE TO PREVENT ACCIDENTAL FILLING WITH A WRONG AGENT.  THEY PREVENT ACCIDENTAL FILLING WITH THE WRONG AGENT REDUCE AIR POLLUTION PREVENT CONTAMINANTS ENTERING VAPORIZATION CHAMBER TYPES 	KEYED FILLING SYSTEM 	SCREW  CAPPED FILLING SYSTEM  	PIN SAFETY SYSTEM
AGENT SPECIFIC FILLING SYSTEMS BOTTLE COLLAR ATTACHED AT THE NECK OF THE BOTTLE. 2 PROJECTIONS,ONE THICKER THAN THE OTHER ARE THERE. THIS MATES WITH THE CORRESPONDING INDENTATIONS ON THE BOTTLE ADAPTOR.
AGENT SPECIFIC FILLING SYSTEMS
AGENT SPECIFIC FILLING SYSTEMS BOTTLE ADAPTOR AT ONE END IS 1) THE BOTTLE CONNECTOR WITH A SCREW THREAD TO MATCH THE THREAD ON THE BOTTLE 2)SKIRT THAT EXTENDS BEYOND THE SCREW THREADS  3)SLOTS THAT MATCH THE PROJECTIONS ON THE BOTTLE COLLAR
AGENT SPECIFIC FILLING SYSTEMS AT THE OTHER END IS 1.THE MALE ADAPTOR THAT FITS INTO THE VAPORIZER FILLER RECEPTACLE. 2.A SHORT LENGTH OF PLASTIC TUBING WITH 2 INNER TUBES CONNECTS THE ENDS. 3.THE TUBE ALLOWS THE BOTTLE TO BE HELD HIGHER OR LOWER THAN THE VAPORIZER.
AGENT SPECIFIC FILLING SYSTEMS MALE ADAPTER HAS A GROOVE ON ONE SIDE TO PREVENT THE PROBE FROM BEING PLACED IN THE INCORRECT VAPORIZER TWO HOLES ON THE OTHER SIDE LARGER FOR AGENT TO LEAVE / ENTER SMALLER FOR AIR
AGENT SPECIFIC FILLING SYSTEMS FILLER RECEPTACLE [FILLER SOCKET/VAPORIZER FILLER UNIT/FILL & DRAIN SYSTEM] MUST PERMIT THE INSERTION OF THE INTENDED BOTTLE . MUST ‘VE A MEANS OF TIGHTENING THE MALE ADAPTER  ,TO FORM A TIGHT SEAL.
AGENT SPECIFIC FILLING SYSTEMS FILLER RECEPTACLE MUST ‘VE A MEANS OF SEALING THE ADAPTOR WHEN BOTTLE ADAPTOR NOT INSERTED. SINGLE PORT FOR FILLING & DRAINAGE. A VALVE ATTACHED TO A KNOB AT THE TOP CONTROLS THE OPENING INTO THE VAPORIZER A BALL VALVE  IN THE  AIRLINE   OCCLUDES  THE AIR  PORT  AFTER  THE  VAP  IS  FILLED  TO PREVENT   OVERFILLING  &  FLOODING OF AIRLINE  WITH  THE  AGENT.
AGENT SPECIFIC FILLING SYSTEMS FILLING CAP OF BOTTLE REMOVED ADAPTER SCREWED TO THE COLLAR
AGENT SPECIFIC FILLING SYSTEMS FILLING VAPORIZER TURNED OFF  PLUG REMOVED BOTTLE WITH ADAPTER INSERTED SUCH THAT THE GROOVES MATCH; [TUBE BENT SUCH THAT BOTTLE IS BELOW INLET]
AGENT SPECIFIC FILLING SYSTEMS FILLING RETAINING SCREW TIGHTENED FILL VALVE [VENT] OPENED BOTTLE HELD HIGH AIR FROM THE VAPORIZER   DISPLACED BY THE LIQUID   MOVES THROUGH THE OTHER   TUBE AND ENTERS THE AIR SPACE INSIDE THE BOTTLE GENTLE UP AND DOWN MOTION MAY HELP
AGENT SPECIFIC FILLING SYSTEMS DRAINING ADAPTER ATTACHED TO BOTTLEFILLER PLUG REMOVED MALE ADAPTOR INSERTEDRETAINING SCREW TIGHTENEDBOTTLE HELD BELOW THE RECEPTACLE DRAIN VALVE OPENEDDRAINED SCREW LOOSENEDADAPTOR REMOVEDFILLER PLUG REINSERTED
AGENT SPECIFIC FILLING SYSTEMS-PROBLEMS 1.DIFFICULTY IN FILLING 2.MISALIGNMENT OF ADAPTOR IN FILLER RECEPTACLE . 3.ADAPTOR NOT SEALING AT THE BOTTLE END . 4.LEAK IN THE BOTTLE ADAPTOR 5.AIR BUBBLES. 6.LOST BOTTLE ADAPTOR 7.FAILURE OF KEYED SYSTEM 8.LIQUID LEAKS 9.INCOMPLETE EMPTYING
AGENT SPECIFIC FILLING SYSTEMS-PROBLEMS 10.VAPORIZER TIPPING IF FILLER RECEPTACLE ON VAPORIZER EXTENDS BEYOND THE BASEIT CANNOT BE SET UPRIGHT ON A FLAT SURFACETHIS CAN CAUSE TIPPING  PREVENTED BY SETTING THE VAPORIZER RECEPTACLE AT THE EDGE OF THE SURFACE  OR A RING CAN BE FITTED AT THE BASE THAT EXTENDS BELOW THE PROJECTION OF THE FILLER REEPTACLE . 11.POOR DRAINAGE 12.BROKEN INNER TUBE
LOCATION BETWEEN THE FLOWMETERS & COMMON GAS OUTLET BETWEEN THE COMMON GAS OUTLET & BREATHING SYSTEM IN- SYSTEM VAPORIZERS
HAZARDS INCORRECT AGENT LOW OUTPUT OR HIGH OUTPUT MISPLACING DESFLURANE DANGEROUS GAS ALLOWEDTO FLOW THRUGH IT UNTIL NO AGENT DETECTED IN THE OUTFLOW
HAZARDS TIPPING LIQUIDFROM VAPORIZING CHAMBERBYPASS/OUTLETHIGH OUTPUT DRAINED BEFORE MOVING
HAZARDS OVERFILLING IN MAJORITY, THE DESIGN OF THE FILLING PORT AND AGENT SPECIFIC FILLING SYSTEMS PREVENT THIS DURING FILLING DO NOT:   TURN THE DIAL ON/ UNSCREW THE BOTTLE ADAPTOR
HAZARDS REVERSED FLOW INLET MALE & OUTLET FEMALE INCRESED OUTPUT CONCENTRATION DIAL IN WRONG POSITION CONTAMINANTS IN VAPORIZING CHAMBER PHYSICAL DAMAGE OBSTRUCTION TO FRESH GAS FLOW INTERLOCK MALFUNCTION LEAKS
Evolution of vaporizers MORTONS ETHER INHALER 1846,OCT 16
SNOW ETHER INHALER 1847 It incorporates the basic principles of vaporization He knew the importance of ambient temperature on vaporization Had a metal box as bath for water at 50-60⁰F
CLOVER PORTABLE REGULATING ETHER INHALER 1877 By Joseph Clover  [surgeon ->anesthetist] For ether & chloroform Can regulate the amount of vapor inhaled, rapid onset,no valves Metal sphere with ether; sphere rotated around a central tube through which patient inhales Model for Ombredanne’s inhaler Markings not related to output concentration
CLOVER GIVING CHLOROFORM
VERNON HARCOURT CHLOROFORM INHALER 1903 BY AG Vernon Harcourt For chloroform [amount inhaled <2%,hence safe] Simple, accurate, portable Apparatus could be worn around the chloroformist’s neck Danger! Don’t lean to take something from the floor
MORE BURDENS ON THE NECK………
OMBREDANNE ETHER INHALER 1908 Louis Ombredanne [plastic Sxn] Strongly argued for air! Against ether! Criticized Clovers ‘useless’ water chamber  Had a modern bag [made from cow’s caecum]
OMBREDANNE INHALER
SOMNOFORM INHALER 1908 SOMNOFORM MIXTURE: Ethyl chloride+ Methyl chloride+ Ethyl bromide [60:35:5]
OPEN DROP MASKS ESMARCH 1879 SCHIMMELBUSCH 1890  YANKAUER MASK 1910 For agents like ether,chloroform, divinyl ether & ethyl chloride Mask was covered with flannel/gauze onto which,these agents were poured
YANKAUER’S MASK
YANKAUER’S MASK (1)Yankauer’s facemask with mesh and (2)removable spiral wire collar for holding in place, lint or gauze, on to which agents such as ether &chloroform were applied Length 15cm  Width 7.9cm Depth 5.5cm Materials: Copper alloy, chrome plate
SCHIMMELBUSCH MASK
SCHIMMELBUSCH MASK SCHIMMELBUSCH FOLDING FACE MASK with a hinged frame for holding gauze/lint in place Height : 18.3cm Width 11.3cm depth 7.3cm Materials: copper alloy, nickel plate For ether 16 layers of gauze; ether dropped over the whole area
SCHIMMELBUSCH MASK Heat of vaporization high;so require much latent heat Mask cools & water vapour condenses on it and freezes Increased resistance and reduced rate of vaporisation Keep a second mask and fresh gauze in supply There should be a Drop Bottle to drop the agent into the mask
SCHIMMELBUSCH MASK For Chloroform, 12 layers of gauze Liquid agent poured over one half of the mask in order to ensure that air is entrained to avoid high concentration  Cover the patients face with a piece of cloth of 25X20 cm.  In this make a central hole to expose patients nose and mouth. This prevents vapor or liquid from entering eyes. Also reduces the amount of air drawn in
OXFORD VINETHENE INHALER 1940 Used for Divinyl ether
OXY-COLUMBUS TRILENE INHALER 1950 By Hans Hosemann Analgesic properties of Trichloroethylene utilized in labour,dressingchanges,dentistry,ENT Held over nose / mouth Chain passed around patients neck Vaporization by warmth of patients hand Inhaler fell from hand as she become unconcious
DUKES TRILENE INHALER 1952 Draw over Non rebreathing mechanism to prevent CO2 accumulation Trileneupto 0.5% ; conc. Controlled by the patient Wrist strap to prevent falling away Mainly in Rx of pain during child birth
DUKES TRILENE INHALER
DRAGER BAR TRILENE INHALER 1955 Safe self administration Overdosing practically impossible Max 1%
PENTHRANE ANALGIZER 1968 For Methoxyflurane Light weight Disposable  No face mask
SKINNERS WIRE FACE MASK 1862 By Thomas Skinner , an Obstetric Physician For chloroform Wire frame and gauze Can be folded and kept in pocket
JUNKERS CHLOROFORM BOTTLE FIRST BLOW OVER APPARATUS Air blown by hand held bellows through glass bottles containing chloroform Modified later with wire face mask,which is safer
OGSTON INHALER Mask is like that of Schimmelbusch Surmounted by a wire frame Around this a gauze could be erected  Semiopen method, economical use of ether
BOYLES BOTTLE PARTS:(1)VAPORIZING BOTTLE 300 mL (2) METAL TOP INCORPORATING CONTROLS (3)LEVER,PLUNGER WHICH IS CHROME PLATED (4)STOPPER & RETAINING CHAIN
BOYLES BOTTLE Stoper cork should be air tight It is retained by small chain Top of cork insulated to avoid electric spark
BOYLES BOTTLE
BOYLES BOTTLE Copper in U-Tube & hood of plunger prevents decomposition of ether Before operating the ether vaporizer, the filling orifices should be kept open for some seconds and then closed ;otherwise high concentration may be delivered Temperature compensated by copper plate or water bath
BOYLES BOTTLE: ETHER-CHLOROFORM-TRILENE
PINSON ETHER BOMB Avoids carrying of complex machines & heavy cylinders 100% warmed ether vapor
TRILENE POCKET VAPORIZER 1941 Self administration of trilene in war Small, non spillable, easy to use Easily refilled  .3-.4% vapor @ room temperature
OXFORD MINIATURE VAPORIZER[OMV] Advantages Small, portable, simple, less servicing May be drained of one agent & filled with the other Detachable scales for several agents
OXFORD MINIATURE VAPORIZER[OMV] PARTS: ETHER PLACED IN INNER CONTAINER SURROUNDED BY CRYSTALS OF HYDRATED CaCl₂  MIXING CHAMBER CONTROL VALVE  50 Ml  CAPACITY WEIGHT WITH WATER 1060gm
OXFORD MINIATURE VAPORIZER[OMV] hot water jackets causes melting of CaCl₂. It cools and solidifies and relases heat of crystalisation. This heat absorbed by ether vaporisation  mixed with O₂ in the mixing chamber  delivered Prone to tipping Screw cap filling Back pressure compensation by circuit valves
TRISERVICE ANESTHESIA APPARATUS Consists of 2 OMVs + LAERDEL folding manual resuscitator quick change from one to other possible, high concentration of each agent can be obtained All except O₂ cylinder housed in a box Dropped from a parachute in war fields
EPSTEIN MACINTOSH OXFORD VAPORIZER 1952 [EMO]
EPSTEIN   MACINTOSH  OXFORD VAPORIZER 1952 [EMO]
EPSTEIN MACINTOSH OXFORD VAPORIZER 1952 [EMO]
EMO-INTERIOR PARTS VAPORIZATION CHAMBER BYPASS CHAMBER MIXING CHAMBER CONTROL VALVE METAL BELLOWS CONTAINING LIQUID IN VAPORIZATION CHAMBER WICKS AROUND VAPORIZING CHAMBER WALL WATER JACKET OF 1250 ML DISTILLED WATER ALUMINIUM OR STEEL CONTROL VALVE AT THE MOUTH OF THE VAPORIZATION CHAMBER
EMO-INTERIOR PARTS
EMO- HOW IT WORKS CONTROL VALVE CONTROLS AIR FLOW INTO THE CHAMBER METAL BELLOWS IS THE THERMOCOMPENSATOR USED WITH OXFORD INFLATING BELLOWS A MAGNET IS USED TO IMMOBILIZE NON REBREATHING VALVE DURING SPONTANEOUS RESPIRATION IN CONTROLLED VENTILATION, NO MAGNET IS USED APL VALVE NEAR PATIENT END KEPT OPEN FOR EXPIRATION
EMO- HOW IT WORKS TIPPING PREVENTED BY ON-OFF VALVES BACK PRESSURE COMPENSATION PREVENTED BY CIRCUIT VALVES ADVANTAGES: LOW COST, PORTABLE,CAN BE USED IN HIGH ALTITUDES AS IT AUTOMATICALLY ADJUSTS FOR CHANGES IN BAROMETRIC PRESSURES IN TROPICS IT HAS TO BE COOLED BEFORE USE IN CLOD TEMPERATURE, ANTIFREEZE SOLUTION 25% GLYCOL USED IN WATER JACKET AIR IS THE CARRIER
EMO HOW IT WORKS
GOLDMAN VAPORIZER
GOLDMAN VAPORIZER-PARTS SMALL GLASS POT METAL HEAD  INLET & OUTLET CONTROL LEVER ON TOP OF METAL HEAD CAPACITY 30 ML MAX CONCENTRATIO 2±0.5%
GOLDMAN VAPORIZER
Goldman Adjustment is made by means of a ratchet type control valve. The glass jar is screwed on to the body and can be quickly detached for easy filling
ROWBOTHOM VAPORIZER MODIFICATION OF GOLDMAN VAPORIZER
FOREGGER COPPER KETTLE 1952
FOREGGER COPPER KETTLE 1952 COPPER VESSELHIGH HEAT CAPACITY & THERMAL CONDUCTIVITY GAS BREAKS AS IT PASSES THROUGH THE DIFFUSER AND GETS FULLY SATURATED IN ITS PASSAGE THROUGH THE LIQUID HIGH DEGREE OF ACCURACY HALOX VAPORIZER- MODIFICATION
[object Object]
TEMPERATURE COMPENSATION IN BOYLE BOTTLE IS BY ETHYL CHLORIDE
WHEN I SPEAK OF COPPER KETTLE THE FIRST WORD THAT COMES TO MY MIND IS MEASURED FLOW
GOLDMAN IS A FLOW OVER TYPE VAPORIZER

More Related Content

What's hot

Physics In Anaesthesia
Physics In AnaesthesiaPhysics In Anaesthesia
Physics In AnaesthesiaNARENDRA PATIL
 
Vaporizers
VaporizersVaporizers
VaporizersKIMS
 
Vaporizers dr. anju bhalotra
Vaporizers  dr. anju  bhalotraVaporizers  dr. anju  bhalotra
Vaporizers dr. anju bhalotraisakakinada
 
Low pressure system in anaesthesia machine
Low pressure system in anaesthesia machineLow pressure system in anaesthesia machine
Low pressure system in anaesthesia machineSwadheen Rout
 
O2 cascade flux n odc
O2 cascade flux n odcO2 cascade flux n odc
O2 cascade flux n odcRony Mathew
 
anaesthesia Vaporizers tec1 to tec5
anaesthesia Vaporizers tec1 to tec5anaesthesia Vaporizers tec1 to tec5
anaesthesia Vaporizers tec1 to tec5Preeti Loona
 
Mapleson circuits
Mapleson circuitsMapleson circuits
Mapleson circuitsArun Shetty
 
Circle system low flow anesthesia
Circle system low flow anesthesiaCircle system low flow anesthesia
Circle system low flow anesthesiaDrgeeta Choudhary
 
Intermediate & low pressure system
Intermediate & low pressure systemIntermediate & low pressure system
Intermediate & low pressure systemDeepa Sinha
 
Intro to Hypoxic pulmonary vasoconstriction
Intro to Hypoxic pulmonary vasoconstriction Intro to Hypoxic pulmonary vasoconstriction
Intro to Hypoxic pulmonary vasoconstriction Arun Shetty
 
Anaesthesia Vaporizers
Anaesthesia VaporizersAnaesthesia Vaporizers
Anaesthesia VaporizersRahul Varshney
 
Anaesthesia breathing systems
Anaesthesia breathing systemsAnaesthesia breathing systems
Anaesthesia breathing systemsD Nkar
 
Anaesthesia gas cylinders & pipeline gas supply
Anaesthesia gas cylinders & pipeline gas supplyAnaesthesia gas cylinders & pipeline gas supply
Anaesthesia gas cylinders & pipeline gas supplyUnnikrishnan Prathapadas
 
Airway local blocks
Airway local blocksAirway local blocks
Airway local blocksNisar Arain
 
Low flow Anaesthesia & Gas Monitoring
Low flow Anaesthesia & Gas MonitoringLow flow Anaesthesia & Gas Monitoring
Low flow Anaesthesia & Gas MonitoringKalpesh Shah
 

What's hot (20)

Physics In Anaesthesia
Physics In AnaesthesiaPhysics In Anaesthesia
Physics In Anaesthesia
 
Vaporizers
VaporizersVaporizers
Vaporizers
 
Vaporizers anand ram final
Vaporizers anand ram finalVaporizers anand ram final
Vaporizers anand ram final
 
Vaporizers dr. anju bhalotra
Vaporizers  dr. anju  bhalotraVaporizers  dr. anju  bhalotra
Vaporizers dr. anju bhalotra
 
Anaesthesia machine 2
Anaesthesia machine 2Anaesthesia machine 2
Anaesthesia machine 2
 
Vaporizers
VaporizersVaporizers
Vaporizers
 
Low pressure system in anaesthesia machine
Low pressure system in anaesthesia machineLow pressure system in anaesthesia machine
Low pressure system in anaesthesia machine
 
O2 cascade flux n odc
O2 cascade flux n odcO2 cascade flux n odc
O2 cascade flux n odc
 
anaesthesia Vaporizers tec1 to tec5
anaesthesia Vaporizers tec1 to tec5anaesthesia Vaporizers tec1 to tec5
anaesthesia Vaporizers tec1 to tec5
 
Mapleson circuits
Mapleson circuitsMapleson circuits
Mapleson circuits
 
Vaporizers 2015 spmc
Vaporizers 2015 spmcVaporizers 2015 spmc
Vaporizers 2015 spmc
 
Circle system low flow anesthesia
Circle system low flow anesthesiaCircle system low flow anesthesia
Circle system low flow anesthesia
 
Intermediate & low pressure system
Intermediate & low pressure systemIntermediate & low pressure system
Intermediate & low pressure system
 
Intro to Hypoxic pulmonary vasoconstriction
Intro to Hypoxic pulmonary vasoconstriction Intro to Hypoxic pulmonary vasoconstriction
Intro to Hypoxic pulmonary vasoconstriction
 
Anaesthesia Vaporizers
Anaesthesia VaporizersAnaesthesia Vaporizers
Anaesthesia Vaporizers
 
Anaesthesia breathing systems
Anaesthesia breathing systemsAnaesthesia breathing systems
Anaesthesia breathing systems
 
Anaesthesia gas cylinders & pipeline gas supply
Anaesthesia gas cylinders & pipeline gas supplyAnaesthesia gas cylinders & pipeline gas supply
Anaesthesia gas cylinders & pipeline gas supply
 
Airway local blocks
Airway local blocksAirway local blocks
Airway local blocks
 
Humidification in Anaesthesia
Humidification in AnaesthesiaHumidification in Anaesthesia
Humidification in Anaesthesia
 
Low flow Anaesthesia & Gas Monitoring
Low flow Anaesthesia & Gas MonitoringLow flow Anaesthesia & Gas Monitoring
Low flow Anaesthesia & Gas Monitoring
 

Viewers also liked

Copy of vaporizers
Copy of vaporizersCopy of vaporizers
Copy of vaporizersimran80
 
Anesthesia Ventilator
Anesthesia VentilatorAnesthesia Ventilator
Anesthesia VentilatorVIT
 
Anaesthesia machine 2014 spmc
Anaesthesia machine 2014 spmcAnaesthesia machine 2014 spmc
Anaesthesia machine 2014 spmcDeepak Choudhary
 
Anesthesia Machine
Anesthesia MachineAnesthesia Machine
Anesthesia MachineKhalid
 
Anaesthesia machine
Anaesthesia machineAnaesthesia machine
Anaesthesia machineimran80
 

Viewers also liked (7)

Copy of vaporizers
Copy of vaporizersCopy of vaporizers
Copy of vaporizers
 
vaporizer(dr.s.p)
vaporizer(dr.s.p)vaporizer(dr.s.p)
vaporizer(dr.s.p)
 
Anesthesia Ventilator
Anesthesia VentilatorAnesthesia Ventilator
Anesthesia Ventilator
 
Anesthesia machine
Anesthesia machineAnesthesia machine
Anesthesia machine
 
Anaesthesia machine 2014 spmc
Anaesthesia machine 2014 spmcAnaesthesia machine 2014 spmc
Anaesthesia machine 2014 spmc
 
Anesthesia Machine
Anesthesia MachineAnesthesia Machine
Anesthesia Machine
 
Anaesthesia machine
Anaesthesia machineAnaesthesia machine
Anaesthesia machine
 

Similar to Vaporizer Fundamentals: Classification, Design, and Operation

Similar to Vaporizer Fundamentals: Classification, Design, and Operation (20)

Vapourizer
Vapourizer Vapourizer
Vapourizer
 
Vaporizers
VaporizersVaporizers
Vaporizers
 
Vaporizers in Anaesthesia
Vaporizers in AnaesthesiaVaporizers in Anaesthesia
Vaporizers in Anaesthesia
 
Anaesthesia Vaporizers by Dr shailendra
Anaesthesia Vaporizers by Dr shailendraAnaesthesia Vaporizers by Dr shailendra
Anaesthesia Vaporizers by Dr shailendra
 
Anaesthetic vapourizers
Anaesthetic vapourizersAnaesthetic vapourizers
Anaesthetic vapourizers
 
Modified gas laws
Modified gas lawsModified gas laws
Modified gas laws
 
Distillation
Distillation Distillation
Distillation
 
Basics of anaesthetic vaporisers
Basics of anaesthetic vaporisersBasics of anaesthetic vaporisers
Basics of anaesthetic vaporisers
 
Simple-distillation by SGP.pptx
Simple-distillation by SGP.pptxSimple-distillation by SGP.pptx
Simple-distillation by SGP.pptx
 
Distillation
Distillation Distillation
Distillation
 
Distillation
Distillation Distillation
Distillation
 
Vaporoizer
VaporoizerVaporoizer
Vaporoizer
 
Condenser in Thermal Power Plant
Condenser in Thermal Power PlantCondenser in Thermal Power Plant
Condenser in Thermal Power Plant
 
GAS LAWS AND THEIR APPLICATION.ppt
GAS LAWS AND THEIR APPLICATION.pptGAS LAWS AND THEIR APPLICATION.ppt
GAS LAWS AND THEIR APPLICATION.ppt
 
Behavior of Gases
Behavior of GasesBehavior of Gases
Behavior of Gases
 
Anesthesia workstation
Anesthesia workstationAnesthesia workstation
Anesthesia workstation
 
Agc wp-dissolved gases in transformer insulation fluids
Agc wp-dissolved gases in transformer insulation fluidsAgc wp-dissolved gases in transformer insulation fluids
Agc wp-dissolved gases in transformer insulation fluids
 
Distillation
DistillationDistillation
Distillation
 
Gas laws
Gas lawsGas laws
Gas laws
 
Evaporator
EvaporatorEvaporator
Evaporator
 

More from Unnikrishnan Prathapadas

Awake Craniotomy and the neurosurgeon.pptx
Awake Craniotomy and the neurosurgeon.pptxAwake Craniotomy and the neurosurgeon.pptx
Awake Craniotomy and the neurosurgeon.pptxUnnikrishnan Prathapadas
 
PHYSIOLOGICAL CHANGES IN AGING IN CNS1.pptx
PHYSIOLOGICAL CHANGES IN AGING IN CNS1.pptxPHYSIOLOGICAL CHANGES IN AGING IN CNS1.pptx
PHYSIOLOGICAL CHANGES IN AGING IN CNS1.pptxUnnikrishnan Prathapadas
 
ANAESTHESIA GAS CYLINDERS & PIPELINE GAS SUPPLY (2).pptx
ANAESTHESIA GAS CYLINDERS & PIPELINE GAS SUPPLY (2).pptxANAESTHESIA GAS CYLINDERS & PIPELINE GAS SUPPLY (2).pptx
ANAESTHESIA GAS CYLINDERS & PIPELINE GAS SUPPLY (2).pptxUnnikrishnan Prathapadas
 
An introduction to paediatric anaesthesia for undergraduates
An introduction to paediatric anaesthesia for undergraduatesAn introduction to paediatric anaesthesia for undergraduates
An introduction to paediatric anaesthesia for undergraduatesUnnikrishnan Prathapadas
 

More from Unnikrishnan Prathapadas (20)

NEWER DRUGS IN ANAESTHESIA.pptx
NEWER DRUGS IN ANAESTHESIA.pptxNEWER DRUGS IN ANAESTHESIA.pptx
NEWER DRUGS IN ANAESTHESIA.pptx
 
ARTERIAL BLOOD GAS ANALYSIS.pptx
ARTERIAL BLOOD GAS ANALYSIS.pptxARTERIAL BLOOD GAS ANALYSIS.pptx
ARTERIAL BLOOD GAS ANALYSIS.pptx
 
BS MAC BS CP50 .pptx
BS MAC BS CP50 .pptxBS MAC BS CP50 .pptx
BS MAC BS CP50 .pptx
 
Awake Craniotomy Anaesthesia.pptx
Awake Craniotomy Anaesthesia.pptxAwake Craniotomy Anaesthesia.pptx
Awake Craniotomy Anaesthesia.pptx
 
Awake Craniotomy and the neurosurgeon.pptx
Awake Craniotomy and the neurosurgeon.pptxAwake Craniotomy and the neurosurgeon.pptx
Awake Craniotomy and the neurosurgeon.pptx
 
TIVA IN NEUROANAESTHESIA.pptx
TIVA IN NEUROANAESTHESIA.pptxTIVA IN NEUROANAESTHESIA.pptx
TIVA IN NEUROANAESTHESIA.pptx
 
SAFE & APPROPRIATE TIVA.pptx
SAFE & APPROPRIATE TIVA.pptxSAFE & APPROPRIATE TIVA.pptx
SAFE & APPROPRIATE TIVA.pptx
 
ARTERIAL BLOOD GAS ANALYSIS FINAL.pptx
ARTERIAL BLOOD GAS ANALYSIS FINAL.pptxARTERIAL BLOOD GAS ANALYSIS FINAL.pptx
ARTERIAL BLOOD GAS ANALYSIS FINAL.pptx
 
PHYSIOLOGICAL CHANGES IN AGING IN CNS1.pptx
PHYSIOLOGICAL CHANGES IN AGING IN CNS1.pptxPHYSIOLOGICAL CHANGES IN AGING IN CNS1.pptx
PHYSIOLOGICAL CHANGES IN AGING IN CNS1.pptx
 
OBSTETRIC HAEMORRHAGE.pptx
OBSTETRIC HAEMORRHAGE.pptxOBSTETRIC HAEMORRHAGE.pptx
OBSTETRIC HAEMORRHAGE.pptx
 
ANAESTHESIA GAS CYLINDERS & PIPELINE GAS SUPPLY (2).pptx
ANAESTHESIA GAS CYLINDERS & PIPELINE GAS SUPPLY (2).pptxANAESTHESIA GAS CYLINDERS & PIPELINE GAS SUPPLY (2).pptx
ANAESTHESIA GAS CYLINDERS & PIPELINE GAS SUPPLY (2).pptx
 
Sodium Imbalances in NEURO ICU.pptx
Sodium Imbalances in NEURO ICU.pptxSodium Imbalances in NEURO ICU.pptx
Sodium Imbalances in NEURO ICU.pptx
 
Delirium in Intensive Care Unit
Delirium in Intensive Care UnitDelirium in Intensive Care Unit
Delirium in Intensive Care Unit
 
NON OPERATING ROOM ANAESTHESIA
NON OPERATING ROOM ANAESTHESIANON OPERATING ROOM ANAESTHESIA
NON OPERATING ROOM ANAESTHESIA
 
DVT AND PE
DVT AND PEDVT AND PE
DVT AND PE
 
Introduction to anesthesia
Introduction to anesthesiaIntroduction to anesthesia
Introduction to anesthesia
 
An introduction to paediatric anaesthesia for undergraduates
An introduction to paediatric anaesthesia for undergraduatesAn introduction to paediatric anaesthesia for undergraduates
An introduction to paediatric anaesthesia for undergraduates
 
TRIGEMINAL NEURALGIA
TRIGEMINAL NEURALGIATRIGEMINAL NEURALGIA
TRIGEMINAL NEURALGIA
 
Intensive fcare for spinal cord injury
Intensive fcare for spinal cord injuryIntensive fcare for spinal cord injury
Intensive fcare for spinal cord injury
 
Arise,awake doctors
Arise,awake doctorsArise,awake doctors
Arise,awake doctors
 

Recently uploaded

epilepsy and status epilepticus for undergraduate.pptx
epilepsy and status epilepticus  for undergraduate.pptxepilepsy and status epilepticus  for undergraduate.pptx
epilepsy and status epilepticus for undergraduate.pptxMohamed Rizk Khodair
 
Case Report Peripartum Cardiomyopathy.pptx
Case Report Peripartum Cardiomyopathy.pptxCase Report Peripartum Cardiomyopathy.pptx
Case Report Peripartum Cardiomyopathy.pptxNiranjan Chavan
 
Radiation Dosimetry Parameters and Isodose Curves.pptx
Radiation Dosimetry Parameters and Isodose Curves.pptxRadiation Dosimetry Parameters and Isodose Curves.pptx
Radiation Dosimetry Parameters and Isodose Curves.pptxDr. Dheeraj Kumar
 
97111 47426 Call Girls In Delhi MUNIRKAA
97111 47426 Call Girls In Delhi MUNIRKAA97111 47426 Call Girls In Delhi MUNIRKAA
97111 47426 Call Girls In Delhi MUNIRKAAjennyeacort
 
Introduction to Sports Injuries by- Dr. Anjali Rai
Introduction to Sports Injuries by- Dr. Anjali RaiIntroduction to Sports Injuries by- Dr. Anjali Rai
Introduction to Sports Injuries by- Dr. Anjali RaiGoogle
 
Presentació "Real-Life VR Integration for Mild Cognitive Impairment Rehabilit...
Presentació "Real-Life VR Integration for Mild Cognitive Impairment Rehabilit...Presentació "Real-Life VR Integration for Mild Cognitive Impairment Rehabilit...
Presentació "Real-Life VR Integration for Mild Cognitive Impairment Rehabilit...Badalona Serveis Assistencials
 
VarSeq 2.6.0: Advancing Pharmacogenomics and Genomic Analysis
VarSeq 2.6.0: Advancing Pharmacogenomics and Genomic AnalysisVarSeq 2.6.0: Advancing Pharmacogenomics and Genomic Analysis
VarSeq 2.6.0: Advancing Pharmacogenomics and Genomic AnalysisGolden Helix
 
Measurement of Radiation and Dosimetric Procedure.pptx
Measurement of Radiation and Dosimetric Procedure.pptxMeasurement of Radiation and Dosimetric Procedure.pptx
Measurement of Radiation and Dosimetric Procedure.pptxDr. Dheeraj Kumar
 
call girls in Connaught Place DELHI 🔝 >༒9540349809 🔝 genuine Escort Service ...
call girls in Connaught Place  DELHI 🔝 >༒9540349809 🔝 genuine Escort Service ...call girls in Connaught Place  DELHI 🔝 >༒9540349809 🔝 genuine Escort Service ...
call girls in Connaught Place DELHI 🔝 >༒9540349809 🔝 genuine Escort Service ...saminamagar
 
Primary headache and facial pain. (2024)
Primary headache and facial pain. (2024)Primary headache and facial pain. (2024)
Primary headache and facial pain. (2024)Mohamed Rizk Khodair
 
Wessex Health Partners Wessex Integrated Care, Population Health, Research & ...
Wessex Health Partners Wessex Integrated Care, Population Health, Research & ...Wessex Health Partners Wessex Integrated Care, Population Health, Research & ...
Wessex Health Partners Wessex Integrated Care, Population Health, Research & ...Wessex Health Partners
 
Glomerular Filtration and determinants of glomerular filtration .pptx
Glomerular Filtration and  determinants of glomerular filtration .pptxGlomerular Filtration and  determinants of glomerular filtration .pptx
Glomerular Filtration and determinants of glomerular filtration .pptxDr.Nusrat Tariq
 
Statistical modeling in pharmaceutical research and development.
Statistical modeling in pharmaceutical research and development.Statistical modeling in pharmaceutical research and development.
Statistical modeling in pharmaceutical research and development.ANJALI
 
Music Therapy's Impact in Palliative Care| IAPCON2024| Dr. Tara Rajendran
Music Therapy's Impact in Palliative Care| IAPCON2024| Dr. Tara RajendranMusic Therapy's Impact in Palliative Care| IAPCON2024| Dr. Tara Rajendran
Music Therapy's Impact in Palliative Care| IAPCON2024| Dr. Tara RajendranTara Rajendran
 
SWD (Short wave diathermy)- Physiotherapy.ppt
SWD (Short wave diathermy)- Physiotherapy.pptSWD (Short wave diathermy)- Physiotherapy.ppt
SWD (Short wave diathermy)- Physiotherapy.pptMumux Mirani
 
Glomerular Filtration rate and its determinants.pptx
Glomerular Filtration rate and its determinants.pptxGlomerular Filtration rate and its determinants.pptx
Glomerular Filtration rate and its determinants.pptxDr.Nusrat Tariq
 
Apiculture Chapter 1. Introduction 2.ppt
Apiculture Chapter 1. Introduction 2.pptApiculture Chapter 1. Introduction 2.ppt
Apiculture Chapter 1. Introduction 2.pptkedirjemalharun
 
Biomechanics- Shoulder Joint!!!!!!!!!!!!
Biomechanics- Shoulder Joint!!!!!!!!!!!!Biomechanics- Shoulder Joint!!!!!!!!!!!!
Biomechanics- Shoulder Joint!!!!!!!!!!!!ibtesaam huma
 
call girls in aerocity DELHI 🔝 >༒9540349809 🔝 genuine Escort Service 🔝✔️✔️
call girls in aerocity DELHI 🔝 >༒9540349809 🔝 genuine Escort Service 🔝✔️✔️call girls in aerocity DELHI 🔝 >༒9540349809 🔝 genuine Escort Service 🔝✔️✔️
call girls in aerocity DELHI 🔝 >༒9540349809 🔝 genuine Escort Service 🔝✔️✔️saminamagar
 
world health day presentation ppt download
world health day presentation ppt downloadworld health day presentation ppt download
world health day presentation ppt downloadAnkitKumar311566
 

Recently uploaded (20)

epilepsy and status epilepticus for undergraduate.pptx
epilepsy and status epilepticus  for undergraduate.pptxepilepsy and status epilepticus  for undergraduate.pptx
epilepsy and status epilepticus for undergraduate.pptx
 
Case Report Peripartum Cardiomyopathy.pptx
Case Report Peripartum Cardiomyopathy.pptxCase Report Peripartum Cardiomyopathy.pptx
Case Report Peripartum Cardiomyopathy.pptx
 
Radiation Dosimetry Parameters and Isodose Curves.pptx
Radiation Dosimetry Parameters and Isodose Curves.pptxRadiation Dosimetry Parameters and Isodose Curves.pptx
Radiation Dosimetry Parameters and Isodose Curves.pptx
 
97111 47426 Call Girls In Delhi MUNIRKAA
97111 47426 Call Girls In Delhi MUNIRKAA97111 47426 Call Girls In Delhi MUNIRKAA
97111 47426 Call Girls In Delhi MUNIRKAA
 
Introduction to Sports Injuries by- Dr. Anjali Rai
Introduction to Sports Injuries by- Dr. Anjali RaiIntroduction to Sports Injuries by- Dr. Anjali Rai
Introduction to Sports Injuries by- Dr. Anjali Rai
 
Presentació "Real-Life VR Integration for Mild Cognitive Impairment Rehabilit...
Presentació "Real-Life VR Integration for Mild Cognitive Impairment Rehabilit...Presentació "Real-Life VR Integration for Mild Cognitive Impairment Rehabilit...
Presentació "Real-Life VR Integration for Mild Cognitive Impairment Rehabilit...
 
VarSeq 2.6.0: Advancing Pharmacogenomics and Genomic Analysis
VarSeq 2.6.0: Advancing Pharmacogenomics and Genomic AnalysisVarSeq 2.6.0: Advancing Pharmacogenomics and Genomic Analysis
VarSeq 2.6.0: Advancing Pharmacogenomics and Genomic Analysis
 
Measurement of Radiation and Dosimetric Procedure.pptx
Measurement of Radiation and Dosimetric Procedure.pptxMeasurement of Radiation and Dosimetric Procedure.pptx
Measurement of Radiation and Dosimetric Procedure.pptx
 
call girls in Connaught Place DELHI 🔝 >༒9540349809 🔝 genuine Escort Service ...
call girls in Connaught Place  DELHI 🔝 >༒9540349809 🔝 genuine Escort Service ...call girls in Connaught Place  DELHI 🔝 >༒9540349809 🔝 genuine Escort Service ...
call girls in Connaught Place DELHI 🔝 >༒9540349809 🔝 genuine Escort Service ...
 
Primary headache and facial pain. (2024)
Primary headache and facial pain. (2024)Primary headache and facial pain. (2024)
Primary headache and facial pain. (2024)
 
Wessex Health Partners Wessex Integrated Care, Population Health, Research & ...
Wessex Health Partners Wessex Integrated Care, Population Health, Research & ...Wessex Health Partners Wessex Integrated Care, Population Health, Research & ...
Wessex Health Partners Wessex Integrated Care, Population Health, Research & ...
 
Glomerular Filtration and determinants of glomerular filtration .pptx
Glomerular Filtration and  determinants of glomerular filtration .pptxGlomerular Filtration and  determinants of glomerular filtration .pptx
Glomerular Filtration and determinants of glomerular filtration .pptx
 
Statistical modeling in pharmaceutical research and development.
Statistical modeling in pharmaceutical research and development.Statistical modeling in pharmaceutical research and development.
Statistical modeling in pharmaceutical research and development.
 
Music Therapy's Impact in Palliative Care| IAPCON2024| Dr. Tara Rajendran
Music Therapy's Impact in Palliative Care| IAPCON2024| Dr. Tara RajendranMusic Therapy's Impact in Palliative Care| IAPCON2024| Dr. Tara Rajendran
Music Therapy's Impact in Palliative Care| IAPCON2024| Dr. Tara Rajendran
 
SWD (Short wave diathermy)- Physiotherapy.ppt
SWD (Short wave diathermy)- Physiotherapy.pptSWD (Short wave diathermy)- Physiotherapy.ppt
SWD (Short wave diathermy)- Physiotherapy.ppt
 
Glomerular Filtration rate and its determinants.pptx
Glomerular Filtration rate and its determinants.pptxGlomerular Filtration rate and its determinants.pptx
Glomerular Filtration rate and its determinants.pptx
 
Apiculture Chapter 1. Introduction 2.ppt
Apiculture Chapter 1. Introduction 2.pptApiculture Chapter 1. Introduction 2.ppt
Apiculture Chapter 1. Introduction 2.ppt
 
Biomechanics- Shoulder Joint!!!!!!!!!!!!
Biomechanics- Shoulder Joint!!!!!!!!!!!!Biomechanics- Shoulder Joint!!!!!!!!!!!!
Biomechanics- Shoulder Joint!!!!!!!!!!!!
 
call girls in aerocity DELHI 🔝 >༒9540349809 🔝 genuine Escort Service 🔝✔️✔️
call girls in aerocity DELHI 🔝 >༒9540349809 🔝 genuine Escort Service 🔝✔️✔️call girls in aerocity DELHI 🔝 >༒9540349809 🔝 genuine Escort Service 🔝✔️✔️
call girls in aerocity DELHI 🔝 >༒9540349809 🔝 genuine Escort Service 🔝✔️✔️
 
world health day presentation ppt download
world health day presentation ppt downloadworld health day presentation ppt download
world health day presentation ppt download
 

Vaporizer Fundamentals: Classification, Design, and Operation

  • 1. http://www.ppt-to-video.com Presenter : Dr Unnikrishnan P Coordinator : Dr C MadhusoodhananPillai Moderators : Dr SheelaRani Dr Ravi Vaporizers : 1
  • 2. 1.DO YOU FEEL YOU NEED IT? 2.SOME IMPORTANT THINGS! 3.CLASSIFICATION 4.FACTORS AFFECTING VAPORIZATION 5.ASTM VAPORIZER STANDARDS 6.FILLING SYSTEMS 7.HAZARDS[lengthy class..is it a hazard?] 8.EVOLUTION
  • 3. Why we need a vaporizer? Most volatile agents exist as liquid at room temp & atm pressure Vaporizers convert this liquid form to vapor phase And add a certain amount of this vapor to the anesthesia circuit... In precisely determined concentrations over a wide range of temperatures, pressures and carrier gas flowrates.
  • 4. Who is inside the vaporizer? Let me introduce..I am a gas Below a certain temp,I am very flexible ; then I will exist in a form which is interconvertable between the liquid and gaseous states- this gaseous form is known as vapor. But above this temp, I am very rigid ; you cannot change me into a liquid now..yes now I am a perfect gas And I wont tell u the name of that temperature!
  • 5. SO VAPOR IS THE GASEOUS PHASE OF A SUBSTANCE BELOW ITS CRITICAL TEMPERATURE Love’ [van der Waals forces] promotes the liquid state, whereas ‘hatred’ [ Kinetic energy] promotes vapor phase
  • 6. Vapor Pressure This vapor exerts a pressure on its surroundings- VAPOR PRESSURE V.P.depends only on the liquid and temperature
  • 7.
  • 8. Saturated Vapor Pressure Now we close the container;the escaping ‘vapor molecules’ now may collide with the liquid surface and reenter it..an equilibrium forms No of molecules reentering = no of molecules leaving SATURATED VAPOR PRESSUREis the partial pressure of the vapor phase of a substance when at equilibrium with its liquid phase (e.g. in a closed container). Increases rapidly as boiling point approaches.
  • 9.
  • 10. Boiling Point You go on increasing the temperature The V.P. will increase At a point it equals the atm pressure B.P.OF A LIQUID IS THE TEMPERATURE AT WHICH ITS VAPOR PRESSURE IS EQUAL TO THE ATM PRESSURE Low atm pressure -> Low B.P. High SVP -> Low B.P.[..means I am very volatile]
  • 11. PARTIAL PRESSURE A MIXTURE OF GASES IN A CLOSED CONTAINER WILL EXERT A PRESSURE ON THE WALLS OF THE CONTAINER. THE PART OF THE PRESSURE EXERTED BY ANY ONE GAS IN THE MIXTURE IS CALLED ITS PARTIAL PRESSURE DEPENDS ONLY ON THE TEMPERATURE THE HIGHEST PARTIAL PRESSURE THAT CAN BE EXERTED BY A GAS AT A GIVEN TEMPERATURE IS ITS VAPOR PRESSURE. IT IS AN ABSOLUTE VALUE.
  • 12. VOLUMES PERCENT NO OF UNITS OF VOLUME OF A GAS IN RELATION TO A TOTAL OF 100 UNITS OF VOLUME FOR THE TOTAL GAS MIXTURE. IT EXPRESSES THE RELATIVE RATIO OF GAS MOLECULES IN A MIXTURE. MOST COMMONLY USED BUT UPTAKE & DEPTH OF ANESTHESIA ARE DIRECTLY RELATED TO PARTIAL PRESSURES PARTIAL PRESSURE/TOTAL PRESSURE= VOLUMES PERCENT/100
  • 13. HEAT OF VAPORIZATION THE NUMBER OF CALORIES REQUIRED TO CONVERT 1GM OF LIQUID INTO A VAPOR OR NUMBER OF CALORIES REQUIRED TO CONVERT 1 ML OF LIQUID INTO A VAPOR IN A VAPORIZER, AS THE LIQUID AGENT VAPORIZES,HEAT IS LOST;SO TEMPERATURE DROPS,VAPOR PRESSURE DROPS = LESSER VOL. AGENT FOR THE CARRIER GAS TO TAKE AWAY↓OUTPUT Oh…tell me some method to supply heat! Else my patient will wake up @#%?
  • 14. SPECIFIC HEAT THE QUANTITY OF HEAT REQUIRED TO RAISE THE TEMP OF 1gm OF THE SUBSTANCE BY 1⁰C IF ITS LOW, THE TEMPERATURE WILL NOT RISE GRADUALLY WHEN WE SUPPLY HEAT TO A VOLATILE AGENT IN ORDER TO REPLACE THAT LOST DUE TO VAPORIZATION IF ITS HIGH...NICE GUY..VERY OBEDIENT! IS THIS IMPORTANT IN ANY OTHER ASPECT? MATERIAL FROM WHICH VAPORIZER CONSTRUCTED
  • 15. THERMAL CAPACITY AMOUNT OF HEAT STORED IN THE VAPORISER BODY. IT IS THE PRODUCT OF SPECIFIC HEAT AND MASS. A VAPORIZER CONSTRUCTED FROM A SUBSTANCE WITH HIGH THERMAL CAPACITY WILL CHANGE TEMP MORE SLOWLY AND SO IS PREFFERED
  • 16. THERMAL CONDUCTIVITY SPEED WITH WHICH HEAT FLOWS THROUGH A SUBSTANCE; CU > AL > BRASS > STEEL >> GLASS. HIGHER THE THERMAL CONDUCTIVITY, BETTER THE SUBSTANCE CONDUCTS HEAT.
  • 17. THERMOSTABILIZATION IS ACHIEVED BY CONSTRUCTING VAPORIZERS OF METALS WITH HIGH THERMAL CONDUCTIVITY (COPPER, BRONZE) TO MINIMIZE TEMPERATURE CHANGES WHEN THE VAPORIZER IS IN USE WICKS SHOULD BE IN CONTACT WITH THE METAL PART
  • 18. GAS LAWS BOYLES LAW:AT CONSTANT TEMPERATURE, THE VOLUME OF A GIVEN MASS OF GAS VARIES INVERSELY WITH THE ABSOLUTE PRESSURE CHARLE’S LAW: AT CONSTANT PRESSURE, THE VOLUME OF A GIVEN MASS OF GAS VARIES DIRECTLY WITH THE ABSOLUTE TEMPERATURE AT CONSTANT VOLUME, THE ABSOLUTE PRESSURE OF A GIVEN MASS OF GAS VARIES DIRECTLY WITH THE ABSOLUTE TEMPERATURE
  • 19. DALTON’S LAW OF PARTIAL PRESSURES IN A MIXTURE OF GASES THE PRESSURE EXERTED BY EACH GAS IS THE SAME AS THAT WHICH IT WOULD EXERT IF IT ALONE OCCUPIED THE CONTAINER
  • 20. AVOGADRO’S HYPOTHESIS EQUAL VOLUMES OF GASES AT THE SAME TEMPERATURE AND PRESSURE CONTAIN EQUAL NUMBER OF MOLECULES THIS NO IS 6.023 X 10²³ THIS MUCH NO OF PARTICLES OF ANY GAS AT STP WILL OCCUPY 22.4 Ls
  • 21. EXAMPLE : CALIBRATION MOLECULAR WT OF ISOFLURANE 184.5 CONTAIN AVOGADRO NO OF PARTICLES; WILL OCCUPY 22.4 Ls @ STP SUPPOSE O₂VAPORIZES 18.45 GM OF ISOFLURANE INTO 224 L =0.1 MOLE = WILL OCCUPY 2.24Ls BUT THIS IS VAPORIZED INTO 224 Ls SO CONCENTRATION OF ISOFLURANE = 2.24/224 = 1%
  • 22. Q: WHICH LAW IS REPRESENTED HERE?
  • 23. CLASSIFICATION OF VAPORIZERS
  • 24. METHOD OF REGULATING THE OUTPUT CONCENTRATION CONCENTRATION CALIBRATED MEASURED FLOW
  • 25. METHOD OF VAPORIZATION FLOW OVER BUBBLE THROUGH INJECTION SIMPLE
  • 27. SPECIFICITY AGENT SPECIFIC MULTIPLE AGENTS
  • 28. POSITION VAPORIZER INSIDE CIRCUIT VAPORIZER OUTSIDE CIRCUIT
  • 29. RESISTANCE PLENUM LOW RESISTANCE/DRAW OVER
  • 31. METHOD OF REGULATING OUTPUT CONCENTRATION
  • 32. CONCENTRATION CALIBRATED TOTAL FLOW FROM THE MACHINE PASSES THROUGH THE VAPORIZER. THIS IS SPLIT BY A VARIABLE RESISTANCE PROPORTIONATING VALVE INTO TWO: ONE PART[USUALLY MAJOR] FLOWS THROUGH THE BYPASS CHAMBER & THE OTHER [USUALLY SMALL] THROUGH THE VAPORIZING CHAMBER.
  • 33. CONCENTRATION CALIBRATED AGENT CONCENTRATION IS CONTROLLED BY A DIAL CALIBRATED IN VOLUMES PERCENT VOLUME VAPOURISED - TYPICALLY 200 ML VAPOUR PER ML OF LIQUID ANAESTHETIC ML LIQUID USED/HOUR = 3 X % X FGF
  • 34. IF WE DIAL A HIGH CONCENTRATION
  • 35. IF WE DIAL A LOW CONCENTRATION
  • 36. SPLITTING RATIO THE RATIO OF THE BYPASS GAS TO THE GAS GOING TO THE VAPORIZING CHAMBER IS CALLED THE SPLITTING RATIO. SPLITTING RATIO DEPENDS ON: THE RESISTANCE OF THE TWO PATHWAYS, WHICH INTURN DEPENDS ON THE VARIABLE ORIFICE OF THE INLET/OUTLET. TEMPERATURE OF THE LIQUID /CARRIER GAS. FLOW RATE OF GASES DENSITY OF THE PARTICULAR GAS ENTERING VAP CHAMBER. MOSTLY CARRIER GAS IS O₂
  • 37. MEASURED FLOW VAPORIZERS THE VAPORIZER HEATS THE ANESTHETIC AGENT TO A TEMPERATURE ABOVE ITS BOILING POINT( SO IT BEHAVES AS GAS ) AND THIS IS THEN METERED INTO THE FRESH GAS FLOW. A MEASURED FLOW IS SENT BY A SEPARATE OXYGEN FLOW METER TO PASS TO THE VAPORIZER WITH THE OUTPUT BEING AT SVP FOR THE ANESTHETIC AGENT. IN ORDER TO DILUTE THIS OTHERWISE LETHAL CONCENTRATION, OUTPUT FROM THAT FLOWMETER IS COMBINED WITH GAS PASSING FROM THE MAIN FLOWMETER e.g. COPPER KETTLE, VERNI-TRIOL
  • 38. MEASURED FLOW VAPORIZERS OPERATOR HAS TO SET THE FLOW TO THE VAPORIZER AND BYPASS WITH SEPARATE FLOWMETERS THIS MEANS THAT RESPECTIVE FLOWS HAVE TO BE CALCULATED FOR EACH AGENT FOR A GIVEN TEMPERATURE AND VAPOUR OUTPUT.
  • 40. FLOW OVER HERE THE CARRIER GAS PASSES OVER THE SURFACE OF A LIQUID. SURFACE AREA OF VAPORISATION CAN BE INCREASED BY USING WICKS.
  • 42. FLOW OVER DUE TO CAPILLARY ACTION, THE ANAESTHETIC AGENT RISES INTO THE WICKS. THIS DRAMATICALLY INCREASES THE SURFACE AREA OF ANAESTHETIC AGENT EXPOSED TO THE FRESH GAS AND THEREBY IMPROVES THE EFFICIENCY OF VAPORISATION. CARRIER GAS CAN BE DIRECTED USING BAFFLES OR SPIRAL TRACKS THAT LENGTHEN THE GAS PATHWAYS OVER THE LIQUID.THIS INCREASES THE TIME AND AREA OF CONTACT.
  • 44. BUBBLE THROUGH CARRIER GAS IS ALLOWED TO BUBBLE THROUGH THE LIQUID. SMALLER THE BUBBLE ,LARGER WILL BE THE SURFACE AREA. METHODS TO BREAK THE GAS INTO SMALL BUBBLES ARE AGITATION & SPLASHING. FACTORS INFLUENCING THE VAPORIZER OUTPUT : SIZE OF THE BUBBLE, DEPTH OF LIQUID, BUBBLING SPEED
  • 45. INJECTION VAPORIZERS A KNOWN AMOUNT OF LIQUID AGENT OR PURE VAPOUR IS INJECTED INTO THE GAS STREAM TO PROVIDE THE DESIRED CONCENTRATION E.G. TEC 6 DESFLURANE VAPORIZER
  • 46. TEMPERATURE COMPENSATION TO MAINTAIN A CONSTANT O/P FROM THE VAPORIZER, MECHANISMS TO COMPENSATE FOR THE FLUCTUATIONS IN TEMPERATURE ARE TO BE EMPLOYED
  • 48. THERMOCOMPENSATION MOST VARIABLE BYPASS VAPORIZERS COMPENSATE FOR CHANGES IN VAPOR PRESSURE BY ALTERING THE SPLITTING RATIO. DONE BY USING A THERMOSENSITIVE ELEMENT [BIMETALLIC STRIP] INCORPORATED IN THE VAPORIZING CHAMBER OR BYPASS CHAMBER. SO, THE SPLITTING OF GAS IS CONTROLLED BY TWO VALVES: (1)THE DIAL WE SET (SPLITTING VALVE) AND (2)THE TEMPERATURE COMPENSATING VALVE
  • 49. THERMOCOMPENSATION IN A BIMETALLIC STRIP, TWO METALS WITH VERY DIFFERENT COEFFICIENTS OF THERMAL EXPANSION ARE FIXED TOGETHER.
  • 50. THERMOCOMPENSATION WHEN THE TEMPERATURE OF THE VAPORISING CHAMBER DROPS, THE BIMETALLIC STRIP BENDS AND MOVES AWAY.  THIS REDUCES THE RESISTANCE TO FLOW AND THUS MORE FLOW OCCURS INTO THE VAPORISING CHAMBER.
  • 51. SUPPLIED HEAT AN ELECTRIC HEATER CAN BE USED TO SUPPLY HEAT TO A VAPORIZER AND MAINTAIN IT AT A CONSTANT TEMPERATURE.E.G.TEC 6
  • 53. PLENUM [Latin= fullness] VAPORIZERS WITH HIGH RESISTANCE WHICH DEPEND ON COMPRESSED GAS DRIVEN UNDER PRESSURE ARE CALLED PLENUM VAPORIZERS. e.g. BOYLE BOTTLE,COPPER KETTLE
  • 54. DRAW OVER CARRIER GAS IS DRAWN THROUGH THE VAPORISER EITHER BY THE PATIENT’S OWN RESPIRATORY EFFORTS, OR BY A SELF-INFLATING BAG OR MANUAL BELLOWS DRAWOVER SYSTEMS OPERATE AT LESS THAN, OR AT AMBIENT PRESSURE FLOW THROUGH THE SYSTEM IS INTERMITTENT, VARYING WITH DIFFERENT PHASES OF INSPIRATION, AND CEASING IN EXPIRATION. A ONE-WAY VALVE PREVENTS REVERSE FLOW IN THE CIRCUIT. E.G. EMO
  • 55. DRAW OVER HAVE A LOW INTERNAL RESISTANCE TO GAS FLOW SO THAT THEY MAY BE USED WITHIN THE BREATHING CIRCUIT, THE GAS FLOW BEING DRIVEN THROUGH THEM BY THE PATIENT'S BREATHING. THEY MAY BE USED IN A NON-REBREATHING DRAW-OVER APPARATUS, OR AS IN-CIRCUIT VAPORIZERS IN A CIRCLE ABSORBER SYSTEM.     ARE USUALLY OF THE SIMPLE BOTTLE OR BOWL TYPE (E.G. THE OHIO 8 BOTTLE, GOLDMAN OR STEPHENS), BUT MORE ACCURATE, LOW RESISTANCE DRAW-OVER VAPORIZERS, SUCH AS THE PENLON OXFORD MINIATURE VAPORIZER (OMV) AND OHMEDA DRAW-OVER TEC VAPORIZERS, HAVE BEEN MANUFACTURED.
  • 56. FACTORS AFFECTING VAPORIZER OUTPUT 1. FLOW THROUGH THE VAPORIZING CHAMBER: Varying the proportion of gas passing through the vaporizing chamber and bypass chamber 2.SURFACE AREA OF THE LIQUID GAS INTERFACE: Greater the surface area, more will be the vaporization. Bubble through > flow over
  • 57. FACTORS AFFECTING VAPORIZER OUTPUT 3.TEMPERATURE: As temperature increase, output increase 4.TIME: Output concentration tend to fall over time
  • 58. FACTORS AFFECTING VAPORIZER OUTPUT 5.GAS FLOW RATE: (A)At high flowrates,the gas leaving vaporization chamber is less saturated (B)Alters the total flow that passes through the vaporization chamber
  • 59. FACTORS AFFECTING VAPORIZER OUTPUT 6. CARRIER GAS COMPOSITION: (a) changes in viscosity & density may affect the proportion of the total flow passing through the vaporization chamber (b)N₂O dissolves in the flow, thus altering the effective volume passing through the vaporization chamber
  • 60. FACTORS AFFECTING VAPORIZER OUTPUT 7.BOILING POINT :HIGHER THE BOILING POINT,LESS WILL BE THE VAPOUR O/P. 8. AMBIENT PRESSURE : SVP IS SOLELY A FUNCTION OF TEMP. SO IF AMBIENT PRESSURE IS REDUCED,THE CONSTANT SVP BECOMES A GREATER PROPORTION OF THE TOTAL PRESSURE OUTPUT INCREASES. AGENTS WITH LOW BOILING POINTS ARE MORE SUSCEPTIBLE TO THE INFLUENCE OF AMBIENT PRESSURE
  • 61. EFFECT OF LOW ATMOSPHERIC PRESSURE
  • 62. CONCENTRATION CALIBRATED HIGH RESISTANCE PATHWAY THROUGH THE VAPORISING CHAMBER OFFERS LESS RESISTANCE, UNDER HYPOBARIC CONDITIONS AND SO A SLIGHT INCREASE IN VAPOR O/P OCCURS.
  • 63. MEASURED FLOW HERE THE DELIVERED PARTIAL PRESSURE & VOLUME PERCENT INCREASES. AMOUNT OF INCREASE DEPENDS ON THE BAROMETRIC PRESSURE & THE VAPOR PRESSURE OF THE AGENT. CLOSER THE VAPOR PRESSURE IS TO BAROMETRIC PRESSURE, GREATER THE EFFECT.
  • 64. EFFECT OF HIGH ATMOSPHERIC PRESSURE
  • 65. CONCENTRATION CALIBRATED ATM PRESSURE INCREASESDENSITY OF GAS CHANGESMORE RESISTANCE TO FLOW OF GAS THROUGH THE VAPORIZING CHAMBERDECREASED VAPOUR O/P (PARTIAL PRESSURE & VOLUME PERCENT) EFFECT ON PARTIAL PRESSURE IS LESS DRAMATIC
  • 66. MEASURED FLOW LOWER CONCENTRATION IN TERMS OF PP/VOLUME PERCENT
  • 67. EFFECT OF INTERMITTENT BACK PRESSURE
  • 68. EFFECT OF INTERMITTENT BACK PRESSURE WHEN ASSISTED OR CONTROLLED VENTILATION IS USED ,THE POSITIVE PRESSURE GENERATED DURING INSPIRATION IS TRANSMITTED FROM THE BREATHING SYSTEM BACK TO THE MACHINE & SOME WAY MAY BE TRANSMITTED TO THE VAPORIZERS. ALSO SEEN WITH THE USE OF OXYGEN FLUSH 2 EFFECTS ARE SEEN:
  • 69. PUMPING EFFECT THE INCREASE IN VAPORIZER OUTPUT CONCENTRATION DUE TO BACK PRESSURE
  • 70. PUMPING EFFECT When the bag is squeezed ,pressure is transmitted back into both, the "by pass" channel and also to the vaporising chamber.  The fresh gas tries to move forward and gets compressed both in the 'by pass' channel and the vaporising chamber.  However, the vaporising chamber volume is much larger than the 'by pass' channel volume, and thus, more fresh gas gets compressed into it than into the 'by pass' channel.
  • 71. PUMPING EFFECT This extra fresh gas that enters the vaporising chamber collects anaesthetic vapor
  • 72. PUMPING EFFECT when the positive pressure is suddenly released (expiration) the previously compressed gases now suddenly expands in all directions.  Some of the rapidly expanding gas (containing vapor) enter the inlet of the vaporiser and cross over into the 'by pass' channel
  • 73. PUMPING EFFECT Normally, a vaporiser 'by pass' channel does not have vapor.  So this vapor due the 'pumping effect' is additional.  When this 'by pass' vapor flows across to the exit, it meets the vapor from the vaporising chamber.  The addition of the 'by pass' vapor to the vapor from the vaporising chamber raises the final concentration of anaesthetic delivered.  i.e. The 'pumping effect' increases the delivered concentration of anaesthetic agent.
  • 74. PUMPING EFFECT SEEN ESPECIALLY WHEN CARRIER GAS IS LOW AGENT IN VAPORIZING CHAMBER IS LOW DIAL SETTING IS LOW PRESSURE FLUCTUATIONS ARE HIGH & FREQUENT.
  • 75. PUMPING EFFECT MECHANISM IN MEASURED FLOW: GAS FLOW TO THESE VAPORIZERS BECOME SATURATED WITH VAPOUR & IS JOINED BY GAS FROM OTHER FLOWMETER,WHICH DILUTES ITS CONCENTRATION. WHEN BACK PRESSURE IS APPLIED,THERE IS RETROGRADE FLOW OF GAS SO THAT THE DILUTED GAS MIXTURE IS FORCED BACK INTO THE VAPORIZER.BECAUSE THIS GAS IS NOT SATURATED ,IT WILL THEN PICK UP ANESTHETIC VAPOUR.THE RESULT IS AN INCREASE IN O/P.
  • 76. MODIFICATIONS TO MINIMISE PUMPING EFFECT KEEP THE VAPORIING CHAMBER SMALL OR INCREASING THE SIZE OF THE BYPASS CHAMBER .
  • 77. MODIFICATIONS TO MINIMISE PUMPING EFFECT ADD LONG SPIRAL OR LARGE DIAMETER TUBE TO LEAD TO THE VAPORIZER CHAMBER.THE EXTRA GAS FORCED INTO THIS TUBE & SUBSEQUENTLY RETURNED TO THE BYPASS DOES NOT REACH THE VAPORIZING CHAMBER
  • 78. MODIFICATIONS TO MINIMISE PUMPING EFFECT INCREASE RESISTANCE TO GAS FLOW THROUGH THE VAPORIZER.
  • 79. MODIFICATIONS TO MINIMISE PUMPING EFFECT CHECK VALVE TO PREVENT BACKWARD FLOW OF GAS
  • 80. MODIFICATIONS TO MINIMISE PUMPING EFFECT EXCLUDE WICKS FROM THE AREA WHERE THE INLET TUBE JOINS THE VAPORIZING CHAMBER. OUTLET TUBE MAY BE MADE LONGER SO THAT UNSATURATED GAS WILL HAVE TO PASS FURTHER BACK BEFORE PICKING UP ANESTHETIC VAPOUR. CONNECTIONS OF OXYGEN FLUSH VALVE LINE TO THE COMMON GAS OUTLET BE DESIGNED TO MINIMISE PRESSURE FLUCTUATIONS THAT MAY PRODUCE A PUMPING EFFECT LIMIT PRESSURE TRANSMITTED TO VAPORIZER TO <10KPa ABOVE NORMAL WORKING PRESSURE ,CONC NOT TO INCREASE >20%
  • 81. PRESSURIZING EFFECT THE O/P OF SOME VAPORIZER USED IN CONJUNCTION WITH AUTOMATIC VENTILATOR HAS BEEN FOUND TO BE LOWER THAN DURING FREE FLOW TO ATMOSPHERE. MOSTLY SEEN WHEN HIGH FLOW LARGE PRESSURE FLUCTUATIONS LOW DIAL SETTINGS
  • 82.
  • 83. PRESSURIZING EFFECT No of mol.s of the agent picked by carrier gas ∝ density of vapor mol.s in vaporizing chamber ∝ vapor pressure Increased pressure in vaporizer chamber compress carrier gas more mol.s/mL No of mols of the anesthetic agent same o/p ↓ed The Changes In Vaporizer O/P Caused By The Pumping Effect Usually Are Greater In Magnitude Than Those Associated With The Pressurizing Effect
  • 84. 1988 ASTM MACHINE STANDARDS FOR VAPORIZERS A VAPORIZER MUST BE CAPABLE OF ACCEPTING A TOTAL GAS FLOW OF 15 lts/min FROM THE ANESTHESIA MACHINE AND IN TURN, DELIVERING A GAS FLOW WITH A PREDICTABLE CONC OF VAPOUR. THE EFFECTS OF THE CONDITIONS OF USE(VARIATIONS IN TEMP,PRESSURE,BACKPRESSURE,INPUT FLOW RATES)ON VAPORIZER PERFORMANCE MUST BE STATED IN CATALOGS.THE EFFECT OF CARRIER GAS COMPOSITION ON VAPORIZER O/P SHOULD ALSO BE SUPPLIED.
  • 85. 1988 ASTM MACHINE STANDARDS FOR VAPORIZERS THE EXTENT TO WHICH TEMP & INFLOW RATES INFLUENCE THE VAPORIZER CONC MUST BE STATED. VAPORIZER MUST BE DESIGNED SO THAT IT CANNOT BE OVERFILLED WHEN IN NORMAL OPERATING POSITION. A SYSTEM THAT ISOLATES THE VAPORIZER FROM EACH OTHER & PREVENTS GAS FROM PASSING THROUGH THE VAPORIZER CHAMBER OF ONE VAPO & THEN THROUGH THAT OF ANOTHER MUST BE PROVIDED CONTROL MUST BE PROVIDED TO LIMIT THE ESCAPE OF ANESTHETIC VAPOUR FROM THE VAP CHAMBER INTO THE FRESH GAS ,SO THAT THE DELVERED CONC IS <0.1% WHEN THE VAPORIZER IS TURNED OFF
  • 86. 1988 ASTM MACHINE STANDARDS FOR VAPORIZERS ALL THE VAPORIZER KNOB MUST TURN COUNTER CLOCKWISE TO INCREASE THE CONC. VAPORIZER MUST BE PROVIDED WITH A LIQUID LEVEL INDICATOR VISIBLE FROM THE FRONT OF THE MACHINE THE VAPORIZER MUST PERMIT MAXIMAL CALIBRATED FLOWS OF O2&N20 IN AN ON & OFF POSITIONS WITH THE VAPORIZER FILLED TO THE MAXIMUM SAFE INDICATED LEVEL WITHOUT DISCHARGING LIQUID THROUGH ITS OUTLET WHEN IT IS MOUNTED & USED IN ACCORDANCE WITH THE MANUFACTURERS INSTRUCTIONS
  • 87. 1988 ASTM MACHINE STANDARDS FOR VAPORIZERS VAPORIZER UNSUITABLE FOR USE IN THE BREATHING SYSTEM MUST ‘VE NONINTERCHANGEABLE 23MM FITTINGS.THE 22MM OR 15MM FITTINGS CANNOT BE USED.THE INLET OF THE VAPORIZER MUST BE MALE,OUTLET MUST BE FEMALE & THE DIRECTION OF GAS FLOW MUST BE MARKED WHEN 23MM FITTINGS ARE USED. VAPORIZER SUITABLE FOR USE IN THE BREATHING SYSTEM MUST ‘VE STANDARD 22MM FITINGS OR SCREW THREADED,WEIGHT BEARING FITTINGS WITH INLET FEMALE & OUTLET MALE.THE INLET &OUTLET PORTS MUST BE MARKED,THE DIRECTION OF GAS FLOW MUST BE INDICATED BY ARROWS & VAPORIZER MUST BE MARKED “FOR USE IN BREATHING SYSTEM”.
  • 88. 1988 ASTM MACHINE STANDARDS FOR VAPORIZERS ALL VAPORIZER LOCATED IN THE FRESH GAS CIRCUIT BE CONCENTRATION CALIBRATED & THAT THE CONTROL OF VAPOUR CAN BE PROVIDED BY MEANS OF CALIBRATED KNOBS OR DIALS.THE MEASURED FLOW SYSTEMS NOT MENTIONED IN THE STANDARDS ARE THEREFORE NOW CONSIDERED BY MOST TO BE OBSOLETE ,ALTHOUGH MILITARY FIELD ANESTHESIA MACHINE STILL USE SUCH SYSTEM
  • 90. AGENT SPECIFIC FILLING SYSTEMS A VAPORIZER DESIGNED FOR A SINGLE AGENT BE FITTED WITH A PERMANENTLY ATTACHED AGENT SPECIFIC DEVICE TO PREVENT ACCIDENTAL FILLING WITH A WRONG AGENT. THEY PREVENT ACCIDENTAL FILLING WITH THE WRONG AGENT REDUCE AIR POLLUTION PREVENT CONTAMINANTS ENTERING VAPORIZATION CHAMBER TYPES KEYED FILLING SYSTEM SCREW CAPPED FILLING SYSTEM PIN SAFETY SYSTEM
  • 91. AGENT SPECIFIC FILLING SYSTEMS BOTTLE COLLAR ATTACHED AT THE NECK OF THE BOTTLE. 2 PROJECTIONS,ONE THICKER THAN THE OTHER ARE THERE. THIS MATES WITH THE CORRESPONDING INDENTATIONS ON THE BOTTLE ADAPTOR.
  • 93. AGENT SPECIFIC FILLING SYSTEMS BOTTLE ADAPTOR AT ONE END IS 1) THE BOTTLE CONNECTOR WITH A SCREW THREAD TO MATCH THE THREAD ON THE BOTTLE 2)SKIRT THAT EXTENDS BEYOND THE SCREW THREADS 3)SLOTS THAT MATCH THE PROJECTIONS ON THE BOTTLE COLLAR
  • 94. AGENT SPECIFIC FILLING SYSTEMS AT THE OTHER END IS 1.THE MALE ADAPTOR THAT FITS INTO THE VAPORIZER FILLER RECEPTACLE. 2.A SHORT LENGTH OF PLASTIC TUBING WITH 2 INNER TUBES CONNECTS THE ENDS. 3.THE TUBE ALLOWS THE BOTTLE TO BE HELD HIGHER OR LOWER THAN THE VAPORIZER.
  • 95. AGENT SPECIFIC FILLING SYSTEMS MALE ADAPTER HAS A GROOVE ON ONE SIDE TO PREVENT THE PROBE FROM BEING PLACED IN THE INCORRECT VAPORIZER TWO HOLES ON THE OTHER SIDE LARGER FOR AGENT TO LEAVE / ENTER SMALLER FOR AIR
  • 96. AGENT SPECIFIC FILLING SYSTEMS FILLER RECEPTACLE [FILLER SOCKET/VAPORIZER FILLER UNIT/FILL & DRAIN SYSTEM] MUST PERMIT THE INSERTION OF THE INTENDED BOTTLE . MUST ‘VE A MEANS OF TIGHTENING THE MALE ADAPTER ,TO FORM A TIGHT SEAL.
  • 97. AGENT SPECIFIC FILLING SYSTEMS FILLER RECEPTACLE MUST ‘VE A MEANS OF SEALING THE ADAPTOR WHEN BOTTLE ADAPTOR NOT INSERTED. SINGLE PORT FOR FILLING & DRAINAGE. A VALVE ATTACHED TO A KNOB AT THE TOP CONTROLS THE OPENING INTO THE VAPORIZER A BALL VALVE IN THE AIRLINE OCCLUDES THE AIR PORT AFTER THE VAP IS FILLED TO PREVENT OVERFILLING & FLOODING OF AIRLINE WITH THE AGENT.
  • 98. AGENT SPECIFIC FILLING SYSTEMS FILLING CAP OF BOTTLE REMOVED ADAPTER SCREWED TO THE COLLAR
  • 99. AGENT SPECIFIC FILLING SYSTEMS FILLING VAPORIZER TURNED OFF PLUG REMOVED BOTTLE WITH ADAPTER INSERTED SUCH THAT THE GROOVES MATCH; [TUBE BENT SUCH THAT BOTTLE IS BELOW INLET]
  • 100. AGENT SPECIFIC FILLING SYSTEMS FILLING RETAINING SCREW TIGHTENED FILL VALVE [VENT] OPENED BOTTLE HELD HIGH AIR FROM THE VAPORIZER DISPLACED BY THE LIQUID MOVES THROUGH THE OTHER TUBE AND ENTERS THE AIR SPACE INSIDE THE BOTTLE GENTLE UP AND DOWN MOTION MAY HELP
  • 101.
  • 102. AGENT SPECIFIC FILLING SYSTEMS DRAINING ADAPTER ATTACHED TO BOTTLEFILLER PLUG REMOVED MALE ADAPTOR INSERTEDRETAINING SCREW TIGHTENEDBOTTLE HELD BELOW THE RECEPTACLE DRAIN VALVE OPENEDDRAINED SCREW LOOSENEDADAPTOR REMOVEDFILLER PLUG REINSERTED
  • 103. AGENT SPECIFIC FILLING SYSTEMS-PROBLEMS 1.DIFFICULTY IN FILLING 2.MISALIGNMENT OF ADAPTOR IN FILLER RECEPTACLE . 3.ADAPTOR NOT SEALING AT THE BOTTLE END . 4.LEAK IN THE BOTTLE ADAPTOR 5.AIR BUBBLES. 6.LOST BOTTLE ADAPTOR 7.FAILURE OF KEYED SYSTEM 8.LIQUID LEAKS 9.INCOMPLETE EMPTYING
  • 104. AGENT SPECIFIC FILLING SYSTEMS-PROBLEMS 10.VAPORIZER TIPPING IF FILLER RECEPTACLE ON VAPORIZER EXTENDS BEYOND THE BASEIT CANNOT BE SET UPRIGHT ON A FLAT SURFACETHIS CAN CAUSE TIPPING PREVENTED BY SETTING THE VAPORIZER RECEPTACLE AT THE EDGE OF THE SURFACE OR A RING CAN BE FITTED AT THE BASE THAT EXTENDS BELOW THE PROJECTION OF THE FILLER REEPTACLE . 11.POOR DRAINAGE 12.BROKEN INNER TUBE
  • 105. LOCATION BETWEEN THE FLOWMETERS & COMMON GAS OUTLET BETWEEN THE COMMON GAS OUTLET & BREATHING SYSTEM IN- SYSTEM VAPORIZERS
  • 106. HAZARDS INCORRECT AGENT LOW OUTPUT OR HIGH OUTPUT MISPLACING DESFLURANE DANGEROUS GAS ALLOWEDTO FLOW THRUGH IT UNTIL NO AGENT DETECTED IN THE OUTFLOW
  • 107. HAZARDS TIPPING LIQUIDFROM VAPORIZING CHAMBERBYPASS/OUTLETHIGH OUTPUT DRAINED BEFORE MOVING
  • 108. HAZARDS OVERFILLING IN MAJORITY, THE DESIGN OF THE FILLING PORT AND AGENT SPECIFIC FILLING SYSTEMS PREVENT THIS DURING FILLING DO NOT: TURN THE DIAL ON/ UNSCREW THE BOTTLE ADAPTOR
  • 109. HAZARDS REVERSED FLOW INLET MALE & OUTLET FEMALE INCRESED OUTPUT CONCENTRATION DIAL IN WRONG POSITION CONTAMINANTS IN VAPORIZING CHAMBER PHYSICAL DAMAGE OBSTRUCTION TO FRESH GAS FLOW INTERLOCK MALFUNCTION LEAKS
  • 110. Evolution of vaporizers MORTONS ETHER INHALER 1846,OCT 16
  • 111. SNOW ETHER INHALER 1847 It incorporates the basic principles of vaporization He knew the importance of ambient temperature on vaporization Had a metal box as bath for water at 50-60⁰F
  • 112. CLOVER PORTABLE REGULATING ETHER INHALER 1877 By Joseph Clover [surgeon ->anesthetist] For ether & chloroform Can regulate the amount of vapor inhaled, rapid onset,no valves Metal sphere with ether; sphere rotated around a central tube through which patient inhales Model for Ombredanne’s inhaler Markings not related to output concentration
  • 114. VERNON HARCOURT CHLOROFORM INHALER 1903 BY AG Vernon Harcourt For chloroform [amount inhaled <2%,hence safe] Simple, accurate, portable Apparatus could be worn around the chloroformist’s neck Danger! Don’t lean to take something from the floor
  • 115. MORE BURDENS ON THE NECK………
  • 116. OMBREDANNE ETHER INHALER 1908 Louis Ombredanne [plastic Sxn] Strongly argued for air! Against ether! Criticized Clovers ‘useless’ water chamber Had a modern bag [made from cow’s caecum]
  • 118. SOMNOFORM INHALER 1908 SOMNOFORM MIXTURE: Ethyl chloride+ Methyl chloride+ Ethyl bromide [60:35:5]
  • 119. OPEN DROP MASKS ESMARCH 1879 SCHIMMELBUSCH 1890 YANKAUER MASK 1910 For agents like ether,chloroform, divinyl ether & ethyl chloride Mask was covered with flannel/gauze onto which,these agents were poured
  • 121. YANKAUER’S MASK (1)Yankauer’s facemask with mesh and (2)removable spiral wire collar for holding in place, lint or gauze, on to which agents such as ether &chloroform were applied Length 15cm Width 7.9cm Depth 5.5cm Materials: Copper alloy, chrome plate
  • 123. SCHIMMELBUSCH MASK SCHIMMELBUSCH FOLDING FACE MASK with a hinged frame for holding gauze/lint in place Height : 18.3cm Width 11.3cm depth 7.3cm Materials: copper alloy, nickel plate For ether 16 layers of gauze; ether dropped over the whole area
  • 124. SCHIMMELBUSCH MASK Heat of vaporization high;so require much latent heat Mask cools & water vapour condenses on it and freezes Increased resistance and reduced rate of vaporisation Keep a second mask and fresh gauze in supply There should be a Drop Bottle to drop the agent into the mask
  • 125. SCHIMMELBUSCH MASK For Chloroform, 12 layers of gauze Liquid agent poured over one half of the mask in order to ensure that air is entrained to avoid high concentration Cover the patients face with a piece of cloth of 25X20 cm. In this make a central hole to expose patients nose and mouth. This prevents vapor or liquid from entering eyes. Also reduces the amount of air drawn in
  • 126. OXFORD VINETHENE INHALER 1940 Used for Divinyl ether
  • 127. OXY-COLUMBUS TRILENE INHALER 1950 By Hans Hosemann Analgesic properties of Trichloroethylene utilized in labour,dressingchanges,dentistry,ENT Held over nose / mouth Chain passed around patients neck Vaporization by warmth of patients hand Inhaler fell from hand as she become unconcious
  • 128. DUKES TRILENE INHALER 1952 Draw over Non rebreathing mechanism to prevent CO2 accumulation Trileneupto 0.5% ; conc. Controlled by the patient Wrist strap to prevent falling away Mainly in Rx of pain during child birth
  • 130. DRAGER BAR TRILENE INHALER 1955 Safe self administration Overdosing practically impossible Max 1%
  • 131. PENTHRANE ANALGIZER 1968 For Methoxyflurane Light weight Disposable No face mask
  • 132. SKINNERS WIRE FACE MASK 1862 By Thomas Skinner , an Obstetric Physician For chloroform Wire frame and gauze Can be folded and kept in pocket
  • 133. JUNKERS CHLOROFORM BOTTLE FIRST BLOW OVER APPARATUS Air blown by hand held bellows through glass bottles containing chloroform Modified later with wire face mask,which is safer
  • 134. OGSTON INHALER Mask is like that of Schimmelbusch Surmounted by a wire frame Around this a gauze could be erected Semiopen method, economical use of ether
  • 135. BOYLES BOTTLE PARTS:(1)VAPORIZING BOTTLE 300 mL (2) METAL TOP INCORPORATING CONTROLS (3)LEVER,PLUNGER WHICH IS CHROME PLATED (4)STOPPER & RETAINING CHAIN
  • 136. BOYLES BOTTLE Stoper cork should be air tight It is retained by small chain Top of cork insulated to avoid electric spark
  • 138. BOYLES BOTTLE Copper in U-Tube & hood of plunger prevents decomposition of ether Before operating the ether vaporizer, the filling orifices should be kept open for some seconds and then closed ;otherwise high concentration may be delivered Temperature compensated by copper plate or water bath
  • 140. PINSON ETHER BOMB Avoids carrying of complex machines & heavy cylinders 100% warmed ether vapor
  • 141. TRILENE POCKET VAPORIZER 1941 Self administration of trilene in war Small, non spillable, easy to use Easily refilled .3-.4% vapor @ room temperature
  • 142. OXFORD MINIATURE VAPORIZER[OMV] Advantages Small, portable, simple, less servicing May be drained of one agent & filled with the other Detachable scales for several agents
  • 143. OXFORD MINIATURE VAPORIZER[OMV] PARTS: ETHER PLACED IN INNER CONTAINER SURROUNDED BY CRYSTALS OF HYDRATED CaCl₂ MIXING CHAMBER CONTROL VALVE 50 Ml CAPACITY WEIGHT WITH WATER 1060gm
  • 144. OXFORD MINIATURE VAPORIZER[OMV] hot water jackets causes melting of CaCl₂. It cools and solidifies and relases heat of crystalisation. This heat absorbed by ether vaporisation  mixed with O₂ in the mixing chamber  delivered Prone to tipping Screw cap filling Back pressure compensation by circuit valves
  • 145. TRISERVICE ANESTHESIA APPARATUS Consists of 2 OMVs + LAERDEL folding manual resuscitator quick change from one to other possible, high concentration of each agent can be obtained All except O₂ cylinder housed in a box Dropped from a parachute in war fields
  • 146. EPSTEIN MACINTOSH OXFORD VAPORIZER 1952 [EMO]
  • 147. EPSTEIN MACINTOSH OXFORD VAPORIZER 1952 [EMO]
  • 148. EPSTEIN MACINTOSH OXFORD VAPORIZER 1952 [EMO]
  • 149. EMO-INTERIOR PARTS VAPORIZATION CHAMBER BYPASS CHAMBER MIXING CHAMBER CONTROL VALVE METAL BELLOWS CONTAINING LIQUID IN VAPORIZATION CHAMBER WICKS AROUND VAPORIZING CHAMBER WALL WATER JACKET OF 1250 ML DISTILLED WATER ALUMINIUM OR STEEL CONTROL VALVE AT THE MOUTH OF THE VAPORIZATION CHAMBER
  • 151. EMO- HOW IT WORKS CONTROL VALVE CONTROLS AIR FLOW INTO THE CHAMBER METAL BELLOWS IS THE THERMOCOMPENSATOR USED WITH OXFORD INFLATING BELLOWS A MAGNET IS USED TO IMMOBILIZE NON REBREATHING VALVE DURING SPONTANEOUS RESPIRATION IN CONTROLLED VENTILATION, NO MAGNET IS USED APL VALVE NEAR PATIENT END KEPT OPEN FOR EXPIRATION
  • 152. EMO- HOW IT WORKS TIPPING PREVENTED BY ON-OFF VALVES BACK PRESSURE COMPENSATION PREVENTED BY CIRCUIT VALVES ADVANTAGES: LOW COST, PORTABLE,CAN BE USED IN HIGH ALTITUDES AS IT AUTOMATICALLY ADJUSTS FOR CHANGES IN BAROMETRIC PRESSURES IN TROPICS IT HAS TO BE COOLED BEFORE USE IN CLOD TEMPERATURE, ANTIFREEZE SOLUTION 25% GLYCOL USED IN WATER JACKET AIR IS THE CARRIER
  • 153. EMO HOW IT WORKS
  • 155. GOLDMAN VAPORIZER-PARTS SMALL GLASS POT METAL HEAD INLET & OUTLET CONTROL LEVER ON TOP OF METAL HEAD CAPACITY 30 ML MAX CONCENTRATIO 2±0.5%
  • 157. Goldman Adjustment is made by means of a ratchet type control valve. The glass jar is screwed on to the body and can be quickly detached for easy filling
  • 158. ROWBOTHOM VAPORIZER MODIFICATION OF GOLDMAN VAPORIZER
  • 160. FOREGGER COPPER KETTLE 1952 COPPER VESSELHIGH HEAT CAPACITY & THERMAL CONDUCTIVITY GAS BREAKS AS IT PASSES THROUGH THE DIFFUSER AND GETS FULLY SATURATED IN ITS PASSAGE THROUGH THE LIQUID HIGH DEGREE OF ACCURACY HALOX VAPORIZER- MODIFICATION
  • 161.
  • 162. TEMPERATURE COMPENSATION IN BOYLE BOTTLE IS BY ETHYL CHLORIDE
  • 163. WHEN I SPEAK OF COPPER KETTLE THE FIRST WORD THAT COMES TO MY MIND IS MEASURED FLOW
  • 164. GOLDMAN IS A FLOW OVER TYPE VAPORIZER
  • 165. GOLDMAN OMV EMO BYLES BOTTLE; ALL ARE DRAW OVER TYPE
  • 166. TEMPERATURE STABILISATION AND COMPENSATION ARE THE SAME
  • 167. NO TEMPERATURE COMPENSATION IN GOLDMAN BOYLE BOTTLE
  • 168. COPPER KETTLE: TEMPERATURE COMPENSATION IS MANUAL
  • 169. BIMETALLIC STRIP IN BYPASS CHAMBER IN ALLADIN CASSETTE
  • 170. GOLDMAN IS AN AGENT SPECIFIC VAPORIZER
  • 171.