SlideShare a Scribd company logo
1 of 32
Download to read offline
TI00083D/06/EN/13.11
71136721
Technical Information
Proline Prosonic Flow 93P
Ultrasonic flow measuring system
Volume flow measurement of liquids in
chemical or process applications
Applications
The sensors are perfectly suited for the non-contact
measurement of pure or slightly contaminated liquids,
regardless of the pressure or electrical conductivity.
• Suitable for pipe diameters
from DN 15 to 4000 (½ to 160")
• Suitable for fluid temperatures ranging
from –40 to +170 °C (–40 to +338 °F)
• Can be used with all metal and plastic pipes lined or
unlined and with composite pipes
• Ideal solution for applications with chemicals,
solvents, liquid hydrocarbons, acids and alkalis
• Particularly suitable for corrosive, hazardous or
ultrapure fluids thanks to non-contact measurement
concept
• Interface for easy integration into all common
distributed control systems:
– HART
– PROFIBUS DP/PA
– FOUNDATION Fieldbus
• Approval for use in hazardous area Zone 1
(ATEX, FM, CSA) available as standard
Features and benefits
The Prosonic Flow ultrasonic clamp-on system allows
accurate and cost-effective flow measurement from
outside the pipe and without the need to interrupt the
process. The flow measurement is bidirectional and
causes no pressure loss.
• Easy, safe and menu-guided sensor mounting
ensures precise measuring results
• Long-term system integrity thanks to robust sensor
and industrial mounting kit design
• Automatic frequency scan for optimized installation
and maximum measuring performance
• Remote configuration using Endress+Hauser's
FieldCare software
Proline Prosonic Flow 93P
2 Endress+Hauser
Table of contents
Function and system design. . . . . . . . . . . . . . . . . . . . . 3
Measuring principle . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3
Measuring system . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3
Sensor selection and arrangement . . . . . . . . . . . . . . . . . . . . . . . . . 5
Two-channel operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
Input . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
Measured variable . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
Measuring range . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
Operable flow range . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
Input signal . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
Output . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
Output signal . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
Signal on alarm . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
Load . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
Switching output . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
Low flow cutoff . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
Galvanic isolation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
Power supply. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
Electrical connection, measuring unit . . . . . . . . . . . . . . . . . . . . . . 9
Electrical connection, terminal assignment . . . . . . . . . . . . . . . . . 10
Connecting the connecting cable . . . . . . . . . . . . . . . . . . . . . . . . 10
Supply voltage . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11
Cable entry . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11
Connecting cable (sensor/transmitter) . . . . . . . . . . . . . . . . . . . . 12
Power consumption . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12
Power supply failure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12
Potential equalization . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12
Performance characteristics. . . . . . . . . . . . . . . . . . . . 12
Reference operating conditions . . . . . . . . . . . . . . . . . . . . . . . . . . 12
Maximum measured error . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12
Repeatability . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13
Operating conditions: installation . . . . . . . . . . . . . . . 13
Installation instructions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13
Inlet and outlet runs . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14
Operating conditions: environment . . . . . . . . . . . . . . 15
Ambient temperature range . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15
Storage temperature . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15
Degree of protection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15
Shock and vibration resistance . . . . . . . . . . . . . . . . . . . . . . . . . . 15
Electromagnetic compatibility (EMC) . . . . . . . . . . . . . . . . . . . . . 15
Operating conditions: process . . . . . . . . . . . . . . . . . . 16
Medium temperature range . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16
Medium pressure range (nominal pressure) . . . . . . . . . . . . . . . . . 16
Pressure loss . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16
Mechanical construction . . . . . . . . . . . . . . . . . . . . . . 17
Design, dimensions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17
Weight . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 24
Materials . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 24
Human interface . . . . . . . . . . . . . . . . . . . . . . . . . . . . 25
Display elements . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 25
Operating elements . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 25
Language group . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 25
Remote operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 25
Certificates and approvals . . . . . . . . . . . . . . . . . . . . . 25
CE mark . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 25
C-Tick mark . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 25
Ex approval . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 25
PROFIBUS DP/PA certification . . . . . . . . . . . . . . . . . . . . . . . . . 25
FOUNDATION Fieldbus certification . . . . . . . . . . . . . . . . . . . . . 25
Other standards and guidelines . . . . . . . . . . . . . . . . . . . . . . . . . . 26
Ordering information. . . . . . . . . . . . . . . . . . . . . . . . . 26
Accessories . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 27
Device-specific accessories . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 27
Measuring principle-specific accessories . . . . . . . . . . . . . . . . . . . 27
Communication-specific accessories . . . . . . . . . . . . . . . . . . . . . . 29
Service-specific accessories . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 30
Documentation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 31
Registered trademarks. . . . . . . . . . . . . . . . . . . . . . . . 31
Proline Prosonic Flow 93P
Endress+Hauser 3
Function and system design
Measuring principle The measuring system operates on the principle of transit time difference. In this measurement method,
acoustic (ultrasonic) signals are transmitted between two sensors. The signals are sent in both directions,
i.e. the sensor in question works as both a sound transmitter and a sound receiver.
As the propagation velocity of the waves is less when the waves travel against the direction of flow than along
the direction of flow, a transit time difference occurs. This transit time difference is directly proportional to the
flow velocity.
A0005961
Principle of the transit time difference measurement method
Q = v · A
a Sensor
b Sensor
Q Volume flow
v Flow velocity (v & t )
t Transit time difference (t = ta – tb)
A Pipe cross-sectional area
The measuring system calculates the volume flow of the fluid from the measured transit time difference and
the pipe cross-sectional area. In addition to measuring the transit time difference, the system simultaneously
measures the sound velocity of the fluid. This additional measured variable can be used to distinguish different
fluids or as a measure of product quality.
The measuring device can be configured onsite to suit the specific application using Quick Setup menus.
Measuring system The measuring system consists of one transmitter and two sensors.
The transmitter is used both to control the sensors and to prepare, process and evaluate the measuring signals,
and to convert the signals to a desired output variable.
The transmitter is optionally equipped for two-channel operation  ä 5.
The sensors work as sound transmitters and sound receivers. Depending on the application and version, the
sensors can be arranged for measurement via one or two traverses  ä 5.
b
a
Q
ta
tb
Proline Prosonic Flow 93P
4 Endress+Hauser
Transmitter
Sensor
Mounting accessories
The requisite mounting distances must be determined for the sensors. Information on the fluid, the pipe
material used and the exact pipe dimensions is needed to determine these values. The values for the sound
velocity of the following fluids, pipe materials and lining materials are stored in the transmitter:
If you are using fluids, pipe materials or lining materials other than those listed in the table, and you do not
have the corresponding sound velocities for these fluids/materials, you can use the DDU18 and DDU19
sensors to determine the values.
Prosonic Flow 93 wall-mount housing Prosonic Flow 93 field housing
For mounting in non-hazardous zone, Ex Zone 2
A0009629
For mounting in Ex Zone 1
A0009676
Prosonic Flow P Prosonic Flow P
DN 15 to 65 (½ to 2½")
A0011484
DN 50 to 4000 (2 to 160")
A0013475
Fluid Pipe material Lining
• Water
• Sea water
• Distilled water
• Ammonia
• Alcohol
• Benzene
• Bromide
• Ethanol
• Glycol
• Kerosene
• Milk
• Methanol
• Toluene
• Lube oil
• Diesel
• Gasoline
• Carbon steel
• Ductile iron
• Stainless steel
• Alloy C
• PVC
• PE
• LDPE
• HDPE
• GRP
• PVDF
• PA
• PP
• PTFE
• Glass pyrex
• Asbestos cement
• Copper
• Mortar
• Rubber
• Tar Epoxy
DDU18
(sound velocity measurement)
DDU19
(wall thickness measurement)
Nominal diameter range: DN 50 to 3000 (2 to 120")
A0009784
Wall thickness range:
• Steel pipes: 2 to 50 mm (0.08 to 2.0")
• Plastic pipes: 4 to 15 mm (0.16 to 0.60")
(only suitable for PTFE and PE pipes to a certain extent)
A0009673
Proline Prosonic Flow 93P
Endress+Hauser 5
Sensor selection and
arrangement
The sensors can be arranged in two ways:
• Mounting arrangement for measurement via one traverse:
the sensors are located on opposite sides of the pipe.
• Mounting arrangement for measurement via two traverses:
the sensors are located on the same side of the pipe.
A0001108
Sensor mounting arrangement
A Mounting arrangement for measurement via one traverse
B Mounting arrangement for measurement via two traverses
The number of traverses required depends on the sensor type, the nominal diameter and the thickness of the
pipe wall. We recommend the following types of mounting:
Two-channel operation The transmitter is able to operate two independent measuring channels (measuring channel 1 and measuring
channel 2). A pair of sensors is connected per measuring channel. Both measuring channels operate
independently of one another and are supported by the transmitter to an equal extent.
Two-channel operation can be used for the following measurements:
• Two-channel measurement = flow measurement at two separate measuring points
• Two-path measurement = redundant flow measurement at one measuring point
BA
Sensor Type Nominal Diameter Sensor Frequency Sensor ID Type of Mounting 1)
Prosonic Flow P
DN 15 to 65 (½ to 2½") 6 MHz P-CL-6F* 2 traverses 3)
DN 50 to 65 (2 to 2½") 6 MHz (or 2 MHz)
P-CL-6F*
P-CL-2F*
2 (or 1) traverses 2)
DN 80 (3") 2 MHz P-CL-2F* 2 traverses
DN 100 to 300 (4 to 12") 2 MHz (or 1 MHz)
P-CL-2F*
P-CL-1F*
2 traverses
DN 300 to 600 (12 to 24") 1 MHz (or 2 MHz)
P-CL-1F*
P-CL-2F*
2 traverses
DN 650 to 4000 (26 to 160") 1 MHz P-CL-1F* 1 traverse
1)
The installation of clamp-on sensors is principally recommended in the 2 traverse type installation. This type of
installation allows the easiest and most comfortable type of mounting and means that a system can also be mounted
even if the pipe can only be accessed from one side. However, in certain applications a 1 traverse installation may be
preferred. These include:
• Certain plastic pipes with wall thickness > 4 mm (0.16")
• Pipes made of composite materials such as GRP
• Lined pipes
• Applications with fluids with high acoustic damping
2)
If the pipe nominal diameter is small (DN 65 / 2½" and smaller), the sensor spacing with Prosonic Flow P can be too
small for two traverse installation using sensor P-CL-2F*. In this case, the 1 traverse type of installation must be used.
3) 6 MHz sensors for applications where flow velocity 10 m/s (32.8 ft/s).
Proline Prosonic Flow 93P
6 Endress+Hauser
Two-channel measurement
The flow is measured at two separate measuring points in the case of two-channel measurement.
The measured values of the two measuring channels can be processed and displayed differently.
The following measured values can be output for two-channel measurement:
• Individual measured values per measuring channel (output independently of one another)
• The difference between the two measured values
• The sum of the two measured values
The two measuring channels can be configured individually. This makes it possible to independently configure
and select the display, outputs, sensor type and type of installation.
A0001159
Two-channel measurement: example of arranging sensor pairs at two separate measuring points
A Measuring channel 1: mounting the sensor pair for measurement via two traverses
B Measuring channel 2: mounting the sensor pair for measurement via one traverse
Two-path measurement
The flow is measured redundantly at one measuring point in the case of two-path measurement.
The measured values of the two measuring channels can be processed and displayed differently.
The following measured values can be output for two-path measurement:
• Individual measured values per measuring channel (output independently of one another)
• The average of the two measured values
The "Averaging" function generally provides you with a more stable measured value. The function is thus
suitable for measurements under conditions that are not ideal (e.g. short inlet runs).
The two measuring channels can be configured individually. This makes it possible to independently configure
and select the display, outputs, sensor type and type of installation.
It is generally not necessary to individually configure the two measuring channels in the case of two-path
measurement. However, in certain situations individual channel configuration can be used to balance out
application-specific asymmetries.
A0001160
Two-path measurement: examples of arranging sensor pairs at one measuring point
A Measuring channel 1 and measuring channel 2: mounting the sensor pair for measurement via two traverses
B Measuring channel 1 and measuring channel 2: mounting the sensor pair for measurement via one traverse
BA
BA
Proline Prosonic Flow 93P
Endress+Hauser 7
Input
Measured variable Flow velocity (transit time difference proportional to flow velocity)
Measuring range Typically v = 0 to 15 m/s (0 to 50 ft/s)
Operable flow range Over 150 : 1
Input signal Status input (auxiliary input)
U = 3 to 30 V DC, Ri = 5 k, galvanically isolated
Configurable for:
totalizer(s) reset, measured-value suppression, error-message reset
Output
Output signal Current output
• Galvanically isolated
• Active/passive selectable
– Active: 0/4 to 20 mA, RL < 700  (for HART: RL 250 )
– Passive: 4 to 20 mA, max. 30 V DC, Ri 150 
• Time constant selectable (0.01 to 100 s)
• Full scale value adjustable
• Temperature coefficient: typ. 0.005 % o.r./°C (o.r. = of reading)
• Resolution: 0.5 μA
Pulse/frequency output
• Galvanically isolated
• Active/passive selectable
– Active: 24 V DC, 25 mA (max. 250 mA during 20 ms), RL > 100 
– Passive: open collector, 30 V DC, 250 mA
• Time constant selectable (0.05 to 100 s)
• Frequency output
– End frequency: 2 to 10000 Hz (fmax = 12500 Hz)
– End frequency for EEx ia 2 to 5000 Hz
– On/off ratio 1:1, pulse width max. 10 s
• Pulse output
– Pulse value and pulse polarity selectable
– Max. pulse width adjustable (0.05 to 2000 ms)
– As of a frequency of 1 / (2 × pulse width), the on/off ratio is 1:1
PROFIBUS DP interface
• PROFIBUS DP in accordance with EN 50170 Volume 2
• Profile version 3.0
• Data transmission rate: 9.6 kBaud to 12 MBaud
• Automatic data transmission rate recognition
• Signal encoding = NRZ Code
• Function blocks: 8 × Analog Input (AI), 3 × Totalizer
• Output data: volume flow channel 1 or channel 2, sound velocity channel 1 or channel 2,
flow velocity channel 1 or channel 2, average volume flow, average sound velocity, average flow velocity,
volume flow sum, volume flow difference, totalizer 1 to 3
• Input data: positive zero return (ON/OFF), zero point adjustment, measuring mode, totalizer control
• Bus address adjustable via miniature switches or local display (optional) at the measuring device
• Available output combination  ä 9
Proline Prosonic Flow 93P
8 Endress+Hauser
PROFIBUS PA interface
• PROFIBUS PA in accordance with EN 50170 Volume 2, IEC 61158-2 (MBP)
• Galvanically isolated
• Data transmission rate, supported baudrate: 31.25 kBit/s
• Current consumption = 11 mA
• Error current FDE (fault disconnection electronic) = 0 mA
• Signal encoding = Manchester II
• Function blocks: 8 × Analog Input (AI), 3 × Totalizer
• Output data: volume flow channel 1 or channel 2, sound velocity channel 1 or channel 2,
flow velocity channel 1 or channel 2, average volume flow, average sound velocity, average flow velocity,
volume flow sum, volume flow difference, totalizer 1 to 3
• Input data: positive zero return (ON/OFF), operation control, totalizer control, zero point adjustment
control, display value
• Bus address can be set via DIP switch on device
FOUNDATION Fieldbus interface
• FOUNDATION Fieldbus H1, IEC 61158-2
• Galvanically isolated
• Data transmission rate, supported baudrate: 31.25 kBit/s
• Current consumption = 12 mA
• Error current FDE (fault disconnection electronic) = 0 mA
• Signal encoding = Manchester II
• Function blocks: 8 × Analog Input (AI), 1 × Discrete Output, 1 × PID
• Output data: volume flow channel 1 or channel 2, sound velocity channel 1 or channel 2,
flow velocity channel 1 or channel 2, signal strength channel 1 or 2, average volume flow,
average sound velocity, average flow velocity, volume flow sum, difference, volume flow,
totalizer 1 to 3
• Input data: positive zero return (ON/OFF), reset totalizer, zero point adjustment control
• Link master function (LAS) is supported
Signal on alarm • Current output  failsafe mode selectable
• Pulse/frequency output  failsafe mode selectable
• Relay output  "deenergized" in the event of a fault or if the power supply fails
Load See "Output signal"
Switching output Relay output
• NC or NO contact available
Factory setting: relay 1 = NO contact, relay 2 = NC contact
• Max. 30 V / 0.5 A AC; 60 V / 0.1 A DC
• Galvanically isolated
• Configurable for: error messages, flow direction, limit values
Low flow cutoff Switch points for low flow are selectable
Galvanic isolation All circuits for inputs, outputs, and power supply are galvanically isolated from each other.
Proline Prosonic Flow 93P
Endress+Hauser 9
Power supply
Electrical connection,
measuring unit
A0013851
Connecting the transmitter, cable cross-section max. 2.5 mm2
(14 AWG)
A Aluminum field housing
B Wall-mount housing
*) Permanent assignment communication boards
**) Flexible assignment communication boards
a Cover of the connection compartment
b Cable for power supply: 85 to 260 V AC, 20 to 55 V AC, 16 to 62 V DC
Terminal No. 1: L1 for AC, L+ for DC
Terminal No. 1: N for AC, L- for DC
c Ground terminal for protective ground
d Signal cable: see terminal assignment
Fieldbus:
- Terminal 26: DP (B) / PA (+), FF (+) with reverse polarity protection
- Terminal 27: DP (A) / PA (–), FF (–) with reverse polarity protection
e Ground terminal for signal cable / fieldbus cable
f Service connector for connecting service interface FXA 193 (Fieldcheck, FieldCare)
g Signal cable: see terminal assignment
Cable for external termination (only for PROFIBUS DP with permanent assignment communication board)
- Terminal No. 24: +5V
- Terminal No. 25: DGND
HART*
PROFIBUS DP**
27
25
23
21
2
1
26
24
22
20
L1 (L+)
N (L-)
–
–
–
+
+
+
PROFIBUS PA*
FOUNDATION Fieldbus*
PA(–)/FF(–)
27
25
23
21
2
1
26
24
22
20
L1 (L+)
N (L-)
–
–
–
+
+
+
A (RxD/TxD-N)
B (RxD/TxD-P)
PA(+)/FF(+)
d
c
e
b
– 27
– 25
– 23
– 21
+ 26
+ 24
+ 22
+ 20
N (L-) 2
L1 (L+)1
25
23
21
2
1
24
22
20
L1 (L+)
N (L-)
–
–
–
+
+
+
27
26
A (RxD/TxD-N)
B (RxD/TxD-P)
d
c
e
b
d
c
e
b
g
d
c
e
b
g
PROFIBUS DP*
f f
f f
a
a
A B
d
b
g
(d)b d/(g)
Proline Prosonic Flow 93P
10 Endress+Hauser
Electrical connection,
terminal assignment
The inputs and outputs on the communication board can be either permanently assigned or variable, depending
on the version ordered (see table). Replacements for modules which are defective or which have to be replaced
can be ordered as accessories.
Connecting the connecting
cable
Connecting the wall-mount housing
A0011390
Connecting the connecting cable, wall-mount housing
1 Channel 1 upstream
2 Channel 1 downstream
3 Channel 2 upstream (only available when ordered)
4 Channel 2 downstream (only available when ordered)
Order variant Terminal No. (inputs/outputs)
20 (+) / 21 (–) 22 (+) / 23 (–) 24 (+) / 25 (–) 26 (+) / 27 (–)
Fixed communication boards (fixed assignment)
93***-***********A – – Frequency output Current output HART
93***-***********B Relay output 2 Relay output 1 Frequency output Current output HART
93***-***********F – – – PROFIBUS PA, Ex i
93***-***********G – – – FOUNDATION Fieldbus, Ex i
93***-***********H – – – PROFIBUS PA
93***-***********J – – – PROFIBUS DP
93***-***********K – – – FOUNDATION Fieldbus
93***-***********S – – Frequency output, Ex i Current output, Ex i,
passive, HART
93***-***********T – – Frequency output, Ex i Current output, Ex i,
passive, HART
Flexible communication boards
93***-***********C Relay output 2 Relay output 1 Frequency output Current output HART
93***-***********D Status input Relay output Frequency output Current output HART
93***-***********L Status input Relay output 2 Relay output 1 Current output HART
93***-***********M Status input Frequency output Frequency output Current output HART
93***-***********P Current output Frequency output Status input PROFIBUS DP
93***-***********V Relay output 2 Relay output 1 Status input PROFIBUS DP
93***-***********W Relay output Current output Current output Current output HART
93***-***********2 Relay output Current output Frequency output Current output HART
93***-***********4 Current input Relay output Frequency output Current output HART
93***-***********6 Relay output Relay output Current output Current output HART
A
1
A
32 4
Proline Prosonic Flow 93P
Endress+Hauser 11
Connecting the field housing
A0008314
Connecting the connecting cable, field housing
1 Channel 1 upstream
2 Channel 1 downstream
3 Channel 2 upstream (only available when ordered)
4 Channel 2 downstream (only available when ordered)
Supply voltage Transmitter
HART
• 85 to 260 V AC, 45 to 65 Hz
• 20 to 55 V AC, 45 to 65 Hz
• 16 to 62 V DC
PROFIBUS DP/PA
• 9 to 32 V DC
FOUNDATION Fieldbus
• 9 to 32 V DC
Sensor
Powered by the transmitter
Cable entry Power supply and signal cables (inputs/outputs)
• Cable gland M20 × 1.5
– Cable gland for cables 8 to 12 mm (0.31 to 0.47")
– Cable gland for cables 6 to 12 mm (0.24 to 0.47")
• Thread for cable entry ½" NPT, G ½"
Connecting cable (sensor/transmitter)
Cable gland for one connecting cable per cable entry, 1 × 8 mm (1 × 0.31")
• Cable gland M20 × 1.5
• Thread for cable entry ½" NPT, G ½"
Cable gland for two connecting cables per cable entry, 2 × 4 mm (2 × 0.16")
• Cable gland M20 × 1.5
• Thread for cable entry ½" NPT, G ½"
A0008152
Cable gland for two connecting cables per cable entry
A
1
2
3
4
A
Proline Prosonic Flow 93P
12 Endress+Hauser
Connecting cable (sensor/
transmitter)
Only use the connecting cables supplied by Endress+Hauser.
Different versions of the connecting cables are available  ä 27.
• Cable material:
– Prosonic Flow 93P (DN 50 to 4000 / 2 to 160"): PVC (standard) or PTFE (for higher temperatures)
– Prosonic Flow 93P (DN 15 to 65 / ½ to 2½"): TPE-V
• Cable length:
– For use in a non-hazardous zone: 5 to 60 m (16.4 to 196.8 ft)
– For use in a hazardous zone: 5 to 30 m (16.4 to 98.4 ft)
! Note!
To ensure correct measuring results, route the connecting cable well clear of electrical machines and switching
elements.
Power consumption AC: < 18 VA (incl. sensor)
DC: < 10 W (incl. sensor)
Switch-on current
• max. 13.5 A (< 50 ms) at 24 V DC
• max. 3 A (< 5 ms) at 260 V AC
Power supply failure Lasting min. 1 power cycle
HistoROM/T-DAT save measuring system data if the power supply fails
Potential equalization For potential equalization, no special measures are necessary.
Performance characteristics
Reference operating
conditions
• Fluid temperature: +20 to +30 °C
• Ambient temperature: +22 °C ± 2 K
• Warm-up period: 30 minutes
• Sensors and transmitter are grounded
• The measuring sensors are correctly installed
Maximum measured error The measured error depends on a number of factors. A distinction is made between the measured error of the
device (Prosonic Flow 93 = 0.5 % of the measured value) and an additional installation-specific measured error
(typically 1.5 % of the measured value) that is independent of the device.
The installation-specific measured error depends on the installation conditions on site, such as the nominal
diameter, wall thickness, real pipe geometry, fluid, etc.
The sum of the two measured errors is the measured error at the measuring point.
A0011347
Example of the measured error in a pipe with a nominal diameter DN > 200 (8")
a Measured error of the device (0.5 % o.r. ± 3 mm/s)
b Measured error due to installation conditions (typically 1.5 % o.r.)
c Measured error at the measuring point: 0.5 % o.r. ± 3 mm/s + 1.5 % o.r. = 2 % o.r. ± 3 mm/s
0
0.5
2.0
3.5
%
m/s2 4 6 8 10 12
b
c
a
Proline Prosonic Flow 93P
Endress+Hauser 13
Measured error at the measuring point
The measured error at the measuring point is made up of the measured error of the device (0.5 % o.r.) and the
measured error resulting from the installation conditions on site. Given a flow velocity > 0.3 m/s (1 ft/s) and
a Reynolds number > 10000, the following are typical error limits:
Measurement Report
If required, the device can be supplied with a factory measurement report. To certify the performance of the
device, a measurement is performed under reference conditions. Here, the sensors are mounted on a pipe with
a nominal diameter of DN 15 (½"), DN 25 (1"), DN 40 (1½"), DN 50 (2") or DN 100 (4") respectively.
The measurement report guarantees the following error limits of the device
[at a flow velocity > 0.3 m/s (1 ft/s) and a Reynolds number > 10000]:
Repeatability ±0.3 % for flow velocities > 0.3 m/s (1 ft/s)
Operating conditions: installation
Installation instructions Mounting location
Correct flow measurement is possible only if the pipe is full. It is preferable to install the sensors in a riser.
! Note!
Entrained air or gas bubbles in the measuring tube can result in an increase in measuring errors.
For this reason, avoid the following mounting locations:
• Highest point of a pipeline. Risk of air accumulating.
• Directly upstream of a free pipe outlet in a vertical pipe. Risk of partial pipe filling.
A0001103
Nominal diameter Device error limits +
Installation-specific
error limits (typical)

Error limits at the measuring
point (typical)
DN 15 (½") ±0.5 % o.r. ± 5 mm/s + ±2.5 % o.r.  ±3 % o.r. ± 5 mm/s
DN 25 to 200 (1 to 8") ±0.5 % o.r. ± 7.5 mm/s + ±1.5 % o.r.  ±2 % o.r. ± 7.5 mm/s
> DN 200 (8") ±0.5 % o.r. ± 3 mm/s + ±1.5 % o.r.  ±2 % o.r. ± 3 mm/s
o.r. = of reading
Nominal diameter Guaranteed error limits of the device
DN 15 (½"), DN 25 (1"), DN 40 (1½"), DN 50 (2") ±0.5 % o.r. ± 5 mm/s
DN 100 (4") ±0.5 % o.r. ± 7.5 mm/s
o.r. = of reading
Proline Prosonic Flow 93P
14 Endress+Hauser
Orientation
Vertical
Recommended orientation with upward direction of flow (View A). With this orientation, entrained solids will
sink and gases will rise away from the sensor when the fluid is stagnant. The piping can be completely drained
and protected against solids buildup.
Horizontal
In the recommended installation range in a horizontal installation position (View B), gas and air collections at
the pipe cover and problematic deposits at the bottom of the pipe have a smaller influence on measurement.
A0001105
A Recommended orientation with upward direction of flow
B Recommended installation range with horizontal orientation
C Recommended installation range max. 120°
Inlet and outlet runs If possible, install the sensor well clear of fittings such as valves, T-pieces, elbows, etc. Compliance with the
following inlet and outlet runs is required in order to ensure measuring accuracy.
A0013459
Inlet and outlet runs (top view)
1 Valve (2/3 open)
2 Pump
3 Two pipe bends in different directions
A
B
C C
³ 15 x DN
³ 20 x DN
³ 20 x DN
³ 15 x DN
³ 3 x DN
1
2
3
Proline Prosonic Flow 93P
Endress+Hauser 15
Operating conditions: environment
Ambient temperature range Transmitter
• Standard: –20 to +60 °C (–4 to +140 °F)
• Optional: –40 to +60 °C (–40 to +140 °F)
Prosonic Flow P sensor
Prosonic Flow P (DN 15 to 65 / ½ to 2½")
• Standard: –40 to +100 °C (–40 to +212 °F)
• Optional: –40 to +150 °C (–40 to +302 °F)
Prosonic Flow P (DN 50 to 4000 / 2 to 160")
• Standard: –40 to +80 °C (–40 to +176 °F)
• Optional: 0 to +170 °C (+32 to +338 °F)
DDU18 sensor (accessories: sound velocity measurement)
• Standard: –40 to +80 °C (–40 to +176 °F)
• Optional: 0 to +170 °C (+32 to +338 °F)
DDU19 sensor (accessories: wall thickness measurement)
–20 to +60 °C (–4 to +140 °F)
Connecting cable (sensor/transmitter)
Prosonic Flow 93P (DN 15 to 65 / ½ to 2½"):
• Standard (TPE-V): –40 to +80 °C (–40 to +176 °F)
Prosonic Flow 93P (DN 50 to 4000 / 2 to 160"):
• Standard (PVC): –20 to +70 °C (–4 to +158 °F)
• Optional (PTFE): –40 to +170 °C (–40 to +338 °F)
! Note!
• It is permitted to insulate the sensors mounted on the pipe.
• Mount the transmitter in a shady location and avoid direct sunlight, particularly in warm climatic regions.
Storage temperature The storage temperature corresponds to the ambient temperature range.
Degree of protection Transmitter
IP 67 (NEMA 4X)
Sensor
IP 68 (NEMA 6P)
DDU18 sensor (accessories: sound velocity measurement)
IP 68 (NEMA 6P)
DDU19 sensor (accessories: wall thickness measurement)
IP 67 (NEMA 4X)
Shock and vibration resistance According to IEC 68-2-6
Electromagnetic compatibility
(EMC)
Electromagnetic compatibility (EMC requirements) according to IEC/EN 61326 "Emission to class A
requirements" and NAMUR Recommendation NE 21 and NE 43.
Proline Prosonic Flow 93P
16 Endress+Hauser
Operating conditions: process
Medium temperature range Prosonic Flow P sensor
Prosonic Flow P (DN 15 to 65)
• Standard: –40 to +100 °C (–40 to +212 °F)
• Optional: –40 to +150 °C (–40 to +302 °F)
Prosonic Flow P (DN 50 to 4000)
• Standard: –40 to +80 °C (–40 to +176 °F)
• Optional: 0 to +170 °C (+32 to +338 °F)
DDU18 sensor (accessories: sound velocity measurement)
• Standard: –40 to +80 °C (–40 to +176 °F)
• Optional: 0 to +170 °C (+32 to +338 °F)
DDU19 sensor (accessories: wall thickness measurement)
–20 to +60 °C (–4 to +140 °F)
Medium pressure range
(nominal pressure)
No pressure limitation, however perfect measurement requires that the static fluid pressure is higher than vapor
pressure.
Pressure loss There is no pressure loss.
Proline Prosonic Flow 93P
Endress+Hauser 17
Mechanical construction
Design, dimensions Transmitter wall-mount housing
A0001150
Dimensions (SI units)
Dimensions (US units)
A B C D E F G H J K
215 250 90.5 159.5 135 90 45 > 50 81 53
L M N O P Q R S T 1)
95 53 102 81.5 11.5 192 8 × M5 20 2 ×  6.5
1) Securing screw for wall mounting: M6 (screw head max. 10.5 mm)
All dimensions in [mm]
A B C D E F G H J K
8.46 9.84 3.56 6.27 5.31 3.54 1.77 > 1.97 3.18 2.08
L M N O P Q R S T 1)
3.74 2.08 4.01 3.20 0.45 7.55 8 × M5 0.79 2 ×  0.26
1)
Securing screw for wall mounting: M6 (screw head max. 0.41")
All dimensions in [inch]
Esc
E- +
DC
B
A
F
E
G
KJ
Q
NL
O
R
H J
M
P P
SS
T
Proline Prosonic Flow 93P
18 Endress+Hauser
There is a separate mounting kit for the wall-mounted housing. It can be ordered from Endress+Hauser as an
accessory ( ä 27). The following installation variants are possible:
• Panel-mounted installation
• Pipe mounting
Panel-mounted installation
A0001131
Pipe mounting
A0001132
245 (9.65)
~110 (~4.33)
210 (8.27)
+0.5 (+0.019)
–0.5 (–0.019)
+0.5 (+0.019)
–0.5 (–0.019)
mm (inch)
Ø 20…70
(Ø 0.79…2.75)
~ ~ 6.1)155 (
mm (inch)
Proline Prosonic Flow 93P
Endress+Hauser 19
Transmitter Field housing
There is a separate mounting kit for the field housing. It can be ordered from Endress+Hauser as an accessory
( ä 27). The following installation variants are possible:
• Wall mounting
• Pipe mounting
Wall mounting
A0006999
Dimensions in SI units
Dimensions in US units
A A* B B* C D E F G H J K L M
265 242 240 217 206 186 178
8.6
(M8)
100 130 100 144 170 355
* Blind version (without local display)
All dimensions in [mm]
A A* B B* C D E F G H J K L M
10.4 9.53 9.45 8.54 8.11 7.32 7.01
8.6
(M8)
3.94 5.12 3.94 5.67 6.69 13.9
* Blind version (without local display)
All dimensions in [inch]
Nicht unter Spannungöffnen
Keepcovertightwhile
circuitsarealive
Nepasouvrirl’appareil soustension
Keepcovertightwhile
circuitsarealive
Nicht-eigensichere
Stromkreise durch
IP40-Abdeckung geschützt
Non-intrinsically safe
circuits Ip40 protected
Boucles de courant
sans sécurité intrinsèque
protégées par Ip40
B*
B
A*
A
F
G
D
C
H
J
K
L
M
E
Esc
E+–
Proline Prosonic Flow 93P
20 Endress+Hauser
Pipe mounting
A0005157
a = Pipe mounting (separate mounting set, see accessories  ä 27)
Nicht-eigensichere
Stromkreise durch
IP40-Abdeckung geschützt
Non-intrinsically safe
circuits Ip40 protected
Boucles de courant
sans sécurité intrinsèque
protégées par Ip40
a
240 (9.45)
mm (inch)
Nicht unter Spannungöffnen
Keepcovertightwhile
circuitsarealive
Nepasouvrirl’appareil soustension
Keepcovertightwhile
circuitsarealive
Esc
E+–
Proline Prosonic Flow 93P
Endress+Hauser 21
Prosonic Flow P sensor (DN 15 to 65 / ½ to 2½")
A0011502
Dimensions in SI units
Dimensions in US units
A B C D E F
72 331 39 28 233 450
All dimensions in [mm]
A B C D E F
2.83 13.03 1.54 1.10 9.17 17.72
All dimensions in [inch]
B
A C
EF
D
Proline Prosonic Flow 93P
22 Endress+Hauser
Prosonic Flow P sensor (DN 50 to 4000 / 2 to 160")
A0001154
Mounting arrangement for measurement via two traverses
Dimensions in SI units
Dimensions in US units
A
C
I
G
DB J
E
F
H
A B C D E F G H
56 62 145 111  58 max. 872 min. 0.5 439 to 790
I J
Depends on the measuring point conditions (pipe, fluid, etc.).
Dimension "I" can be determined:
• Via transmitter programming Quick Setup (or FieldCare)
• Online (Applicator)
Pipe outer diameter
All dimensions in [mm]
A B C D E F G H
2.20 2.44 5.71 4.37  2.28 max. 34.3 min. 0.2 16.3 to 31.1
I J
Depends on the measuring point conditions (pipe, fluid, etc.).
Dimension "I" can be determined:
• Via transmitter programming Quick Setup (or FieldCare)
• Online (Applicator)
Pipe outer diameter
All dimensions in [inch]
Proline Prosonic Flow 93P
Endress+Hauser 23
A0001155
Mounting arrangement for measurement via one traverse
Dimensions in SI units
Dimensions in US units
E
DH
A
B
G
C
F
A B C D E F
56 62 145 111  58 max. 872
G H
Depends on the measuring point conditions (pipe, fluid, etc.).
Dimension "G" can be determined:
• Via transmitter programming Quick Setup (or FieldCare)
• Online (Applicator)
Pipe outer diameter
All dimensions in [mm]
A B C D E F
2.20 2.44 5.71 4.37  2.28 max. 34.3
G H
Depends on the measuring point conditions (pipe, fluid, etc.).
Dimension "G" can be determined:
• Via transmitter programming Quick Setup (or FieldCare)
• Online (Applicator)
Pipe outer diameter
All dimensions in [inch]
Proline Prosonic Flow 93P
24 Endress+Hauser
Weight Transmitter
• Wall-mount housing: 6.0 kg (13.2 lbs)
• Field housing: 6.7 kg (14.8 lbs)
Sensor Prosonic Flow P
• DN 15 to 65 (½ to 2½") (incl. mounting material): 1.2 kg (2.65 lbs)
• DN 50 to 4000 (2 to 160") (incl. mounting material): 2.8 kg (6.2 lbs)
Sensor (accessories)
• Prosonic Flow DDU18 (incl. mounting material): 2.4 kg (5.3 lbs)
• Prosonic Flow DDU19 (incl. mounting material): 1.5 kg (3.3 lbs)
! Note!
Weight information without packaging material.
Materials Transmitter
• Wall-mounted housing: powder-coated die-cast aluminum
• Field housing: powder-coated die-cast aluminum
Sensor Prosonic Flow P
DN 15 to 65 (½ to 2½"); DN 50 to 4000 (2 to 160")
• Sensor holder: stainless steel 1.4308/CF-8
• Sensor housing: stainless steel 1.4301/304
• Strapping bands/bracket: stainless steel 1.4301/304
• Sensor contact surfaces: chemically stable plastic
Sensor (accessories)
Prosonic Flow DDU18; Prosonic Flow DDU19
• Sensor holder: stainless steel 1.4308/CF-8
• Sensor housing: stainless steel 1.4301/304
• Strapping bands/bracket: stainless steel 1.4301/304
• Sensor contact surfaces: chemically stable plastic
Connecting cable (sensor/transmitter)
Prosonic Flow 93P (DN 15 to 65 / ½ to 2½")
• TPE-V connecting cable
– Cable sheath: TPE-V
– Cable connector: stainless steel 1.40301
Prosonic Flow 93P (DN 50 to 4000 / 2 to 160")
• PVC connecting cable
– Cable sheath: PVC
– Cable connector: nickeled brass 2.0401
• PTFE connecting cable
– Cable sheath: PTFE
– Cable connector: stainless steel 1.4301
Proline Prosonic Flow 93P
Endress+Hauser 25
Human interface
Display elements • Liquid crystal display: illuminated, four lines each with 16 characters
• Custom configuration for presenting different measured values and status variables
• 3 totalizers
Operating elements • Local operation with three optical keys
• Application specific Quick Setup menus for straightforward commissioning
Language group Language groups available for operation in different countries:
• Western Europe and America (WEA):
English, German, Spanish, Italian, French, Dutch and Portuguese
• Eastern Europe/Scandinavia (EES):
English, Russian, Polish, Norwegian, Finnish, Swedish and Czech
• South and Eastern Asia (SEA):
English, Japanese, Indonesian
• China (CN):
English, Chinese
You can change the language group via the FieldCare operating program.
Remote operation Operation via HART, PROFIBUS DP/PA, FOUNDATION Fieldbus
Certificates and approvals
CE mark The measuring system is in conformity with the statutory requirements of the EC Directives.
Endress+Hauser confirms successful testing of the device by affixing to it the CE mark.
C-Tick mark The measuring system is in conformity with the EMC requirements of the "Australian Communications and
Media Authority" (ACMA).
Ex approval Information on the currently available Ex-rated versions (ATEX, IECEx, FM, CSA, NEPSI, etc.) can be supplied
by your Endress+Hauser Sales Center on request. All information relevant to explosion protection is available
in separate documents that you can order as necessary.
PROFIBUS DP/PA
certification
The flowmeter has successfully passed all the test procedures carried out and is certified and registered by the
PNO (PROFIBUS/DP User Organization). The device thus meets all the requirements of the following
specifications:
• Certified to PROFIBUS DP/PA, profile version 3.0 (device certification number: on request)
• The device can also be operated with certified devices of other manufacturers (interoperability)
FOUNDATION Fieldbus
certification
The flowmeter has successfully passed all the test procedures carried out and is certified and registered by the
Fieldbus Foundation. The device thus meets all the requirements of the following specifications:
• Certified to FOUNDATION Fieldbus Specification
• The device meets all the specifications of the FOUNDATION Fieldbus H1.
• Interoperability Test Kit (ITK), revision status 5.01 (device certification number: on request)
• The device can also be operated with certified devices of other manufacturers
• Physical Layer Conformance Test of the Fieldbus Foundation
Proline Prosonic Flow 93P
26 Endress+Hauser
Other standards and
guidelines
• EN 60529
Degrees of protection provided by enclosures (IP code)
• EN 61010-1
Safety requirements for electrical equipment for measurement, control and laboratory use
• IEC/EN 61326
"Emission in accordance with Class A requirements".
Electromagnetic compatibility (EMC requirements).
• ANSI/ISA-S82.01
Safety Standard for Electrical and Electronic Test, Measuring, Controlling and Related Equipment -
General Requirements. Pollution Degree 2, Installation Category II.
• CAN/CSA-C22.2 No. 1010.1-92
Safety Requirements for Electrical Equipment for Measurement and Control and Laboratory Use.
Pollution degree 2, Installation Category II
• NAMUR NE 21
Electromagnetic compatibility (EMC) of industrial process and laboratory control equipment.
• NAMUR NE 43
Standardization of the signal level for the breakdown information of digital transmitters with analog output
signal.
• NAMUR NE 53
Software of field devices and signal-processing devices with digital electronics.
Ordering information
The Endress+Hauser sales and service organization can provide detailed ordering information and information
on the order codes on request.
Proline Prosonic Flow 93P
Endress+Hauser 27
Accessories
Various accessories, which can be ordered separately from Endress+Hauser, are available for the transmitter
and the sensor. The Endress+Hauser service organization can provide detailed information on the order codes
on request.
Device-specific accessories
Measuring principle-specific
accessories
Accessory Description Order code
Wall-mount housing,
transmitter
Prosonic Flow 93
Transmitter for replacement or for stock. Use the order code to
define the following specifications:
• Approvals
• Degree of protection/version
• Cable entry
• Display / power supply / operation
• Software
• Outputs / inputs
Single-channel version:
93XXX - XX1XX********
Two-channel version:
93XXX - XX2XX********
Conversion kit,
inputs/outputs
Conversion kit with appropriate plug-in point modules for
converting the current input/output configuration to a new
version.
DK9UI - **
Sensor P
(DN 15 to 65 / ½ to 2½")
Clamp-on version
DN 15 to 65 (½ to 2½")
• –40 to +100 °C (–40 to 212 °F)
• –40 to +150 °C (–40 to 302 °F)
DK9PS - 1*
DK9PS - 2*
Sensor P
(DN 50 to 4000 / 2 to 160")
Clamp-on version
DN 50 to 300 (2 to 12")
• –40 to +80 °C (–40 to 176 °F)
• 0 to +170 °C (+32 to 338 °F)
DK9PS - B*
DK9PS - F*
DN 100 to 4000 (4 to 160")
• –40 to +80 °C (–40 to 176 °F)
• 0 to +170 °C (+32 to 338 °F)
DK9PS - A*
DK9PS - E*
Sensor DDU18 Sensor for sound velocity measurement
• –40 to +80 °C (–40 to 176 °F)
• 0 to +170 °C (+32 to 338 °F)
50091703
50091704
Sensor DDU19 Sensor for wall thickness measurement. 50091713
Accessory Description Order code
Mounting kit for
aluminum field housing
Mounting kit for wall-mount housing.
Suitable for:
• Wall mounting
• Pipe mounting
• Panel mounting
DK9WM - A
Mounting kit for field housing Mounting kit for aluminum field housing:
Suitable for pipe mounting (¾ to 3")
DK9WM - B
Sensor holder set • Prosonic Flow P (DN 15 to 65 / ½ to 2½"):
Sensor holder, clamp-on version
DK9SH - 1
• Prosonic Flow P (DN 50 to 4000 / 2 to 160")
– Sensor holder, fixed retaining nut, clamp-on version
– Sensor holder, removable retaining nut, clamp-on version
DK9SH - A
DK9SH - B
Proline Prosonic Flow 93P
28 Endress+Hauser
Clamp-on installation set Sensor fastening for Prosonic Flow P
(DN 15 to 65 / ½ to 2½")
• U-shaped screw DN 15 to 32 (½ to 1¼")
• Strapping bands DN 40 to 65 (1½ to 2½")
DK9IC - 1*
DK9IC - 2*
Sensor fastening for Prosonic Flow P
(DN 50 to 4000 / 2 to 160")
• Without sensor fastening
• Strapping bands DN 50 to 200 (2 to 8")
• Strapping bands DN 200 to 600 (8 to 24")
• Strapping bands DN 600 to 2000 (24 to 80")
• Strapping bands DN 2000 to 4000 (80 to 160")
DK9IC - A*
DK9IC - B*
DK9IC - C*
DK9IC - D*
DK9IC - E*
• Without mounting tools
• Spacing ruler DN 50 to 200 (2 to 8")
• Spacing ruler DN 200 to 600 (8 to 24")
• Mounting rail DN 50 to 200 (2 to 8")
• Mounting rail DN 200 to 600 (8 to 24")
• Fastening, 1 traverse DN 50 to 4000 (2 to 160")
DK9IC - *1
DK9IC - *2
DK9IC - *3
DK9IC - *4
DK9IC - *5
DK9IC - *6
Conduit adapter for
connecting cable
Prosonic Flow P (DN 15 to 65 / ½ to 2½")
• Conduit adapter incl. cable entry M20 × 1.5
• Conduit adapter incl. cable entry ½" NPT
• Conduit adapter incl. cable entry G ½"
DK9CB - BA1
DK9CB - BA2
DK9CB - BA3
Prosonic Flow P (DN 50 to 4000 / 2 to 160")
• Conduit adapter incl. cable entry M20 × 1.5
• Conduit adapter incl. cable entry ½" NPT
• Conduit adapter incl. cable entry G ½"
DK9CB - BB1
DK9CB - BB2
DK9CB - BB3
Connecting cable Prosonic Flow P (DN 15 to 65 / ½ to 2½")
5 m sensor cable, TPE-V, –20 to +70 °C (–4 to +158 °F)
10 m sensor cable, TPE-V, –20 to +70 °C (–4 to +158 °F)
15 m sensor cable, TPE-V, –20 to +70 °C (–4 to +158 °F)
30 m sensor cable, TPE-V, –20 to +70 °C (–4 to +158 °F)
DK9SS - BAA
DK9SS - BAB
DK9SS - BAC
DK9SS - BAD
Prosonic Flow P (DN 50 to 4000 / 2 to 160")
5 m sensor cable, PVC, –20 to +70 °C (–4 to +158 °F)
10 m sensor cable, PVC, –20 to +70 °C (–4 to +158 °F)
15 m sensor cable, PVC, –20 to +70 °C (–4 to +158 °F)
30 m sensor cable, PVC, –20 to +70 °C (–4 to +158 °F)
DK9SS - BBA
DK9SS - BBB
DK9SS - BBC
DK9SS - BBD
5 m sensor cable, PTFE, –40 to +170 °C (–40 to +158 °F)
10 m sensor cable, PTFE, –40 to +170 °C (–40 to +158 °F)
15 m sensor cable, PTFE, –40 to +170 °C (–40 to +158 °F)
30 m sensor cable, PTFE, –40 to +170 °C (–40 to +158 °F)
DK9SS - BBE
DK9SS - BBF
DK9SS - BBG
DK9SS - BBH
Acoustic coupling fluid • Coupling fluid 0 to 170 °C (+32 to 338 °F),
high temperature, standard
• Adhesive coupling fluid –40 to +80 °C (–40 to +176 °F)
• Water-soluble coupling fluid –20 to +80 °C (–4 to +176 °F)
• Coupling fluid DDU 19, –20 to +60 °C (–4 to +140 °F)
• Coupling fluid –40 to +80 °C (–40 to +176 °F), standard,
type MBG2000
DK9CM - 2
DK9CM - 3
DK9CM - 4
DK9CM - 6
DK9CM - 7
Accessory Description Order code
Proline Prosonic Flow 93P
Endress+Hauser 29
Communication-specific
accessories
Accessory Description Order code
HART handheld terminal
DXR375
Handheld terminal for remote configuration and for obtaining
measured values via the HART current output (4 to 20 mA)
and FOUNDATION Fieldbus.
Contact your Endress+Hauser representative for more
information.
DXR375 - *******
Fieldgate FXA320 Gateway for remote interrogation of HART sensors and
actuators via Web browser:
• 2-channel analog input (4 to 20 mA)
• 4 binary inputs with event counter function and frequency
measurement
• Communication via modem, Ethernet or GSM
• Visualization via Internet/Intranet in the Web browser and/
or WAP cellular phone
• Limit value monitoring with alarm signaling via e-mail or
SMS
• Synchronized time stamping of all measured values
FXA320 - *****
Fieldgate FXA520 Gateway for remote interrogation of HART sensors and
actuators via Web browser:
• Web server for remote monitoring of up to 30 measuring
points
• Intrinsically safe version [EEx ia]IIC for applications in
hazardous areas
• Communication via modem, Ethernet or GSM
• Visualization via Internet/Intranet in the Web browser and/
or WAP cellular phone
• Limit value monitoring with alarm signaling via e-mail or
SMS
• Synchronized time stamping of all measured values
• Remote diagnosis and remote configuration of
connected HART devices
FXA520 - ****
Proline Prosonic Flow 93P
30 Endress+Hauser
Service-specific accessories Accessory Description Order code
Applicator Software for selecting and planning flowmeters.
The Applicator can be downloaded from the Internet or
ordered on CD-ROM for installation on a local PC.
Contact your Endress+Hauser representative for more
information.
DXA80 - *
Fieldcheck Tester/simulator for testing flowmeters in the field.
When used in conjunction with the "FieldCare" software
package, test results can be imported into a database, printed
out and used for official certification.
Contact your Endress+Hauser representative for more
information.
50098801
FieldCare FieldCare is Endress+Hauser's FDT-based plant asset
management tool. It can configure all intelligent field units in
your system and helps you manage them.
By using the status information, it is also a simple but effective
way of checking their status and condition.
See the product page on
the Endress+Hauser
website:
www.endress.com
FXA193 Service interface from the measuring device to the PC for
operation via FieldCare.
FXA193 – *
Communication cable Communication cable for connecting the Prosonic Flow 93P
transmitter to the FXA193 service interface.
DK9ZT – A
Proline Prosonic Flow 93P
Endress+Hauser 31
Documentation
• Flow measurement (FA005D)
• Operating Instructions for Prosonic Flow 93 (BA070D and BA071D)
• Operating Instructions for Prosonic Flow 93 PROFIBUS DP/PA (BA076D and BA077D)
• Operating Instructions for Prosonic Flow 93 FOUNDATION Fieldbus (BA078D and BA079D)
• Supplementary documentation on Ex-ratings: ATEX, FM, CSA, IECEx, NEPSI
Registered trademarks
HART®
Registered trademark of HART Communication Foundation, Austin, USA
PROFIBUS®
Registered trademark of the PROFIBUS User Organization, Karlsruhe, Germany
FOUNDATION™ Fieldbus
Registered trademark of the Fieldbus FOUNDATION, Austin, USA
HistoROM™, T-DAT™, F-CHIP®, FieldCare®, Fieldcheck®
Registered or registration-pending trademarks of Endress+Hauser Flowtec AG, Reinach, CH
Instruments International
Endress+Hauser
Instruments International AG
Kaegenstrasse 2
4153 Reinach
Switzerland
Tel.+41 61 715 81 00
Fax+41 61 715 25 00
www.endress.com
info@ii.endress.com
TI00083D/06/EN/13.11
71136721
FM+SGML6.0 ProMoDo

More Related Content

What's hot

Proline proline prowirl 72 f 72w_73f_73w-endress+hauser datasheet-vortex flow...
Proline proline prowirl 72 f 72w_73f_73w-endress+hauser datasheet-vortex flow...Proline proline prowirl 72 f 72w_73f_73w-endress+hauser datasheet-vortex flow...
Proline proline prowirl 72 f 72w_73f_73w-endress+hauser datasheet-vortex flow...ENVIMART
 
Liquistation csf48 endress+hauser datasheet-automatic stationary sampler
Liquistation csf48 endress+hauser datasheet-automatic stationary samplerLiquistation csf48 endress+hauser datasheet-automatic stationary sampler
Liquistation csf48 endress+hauser datasheet-automatic stationary samplerENVIMART
 
Proline promag 50 l endress+hauser datasheet-electromagnetic flowmeter
Proline promag 50 l endress+hauser datasheet-electromagnetic flowmeterProline promag 50 l endress+hauser datasheet-electromagnetic flowmeter
Proline promag 50 l endress+hauser datasheet-electromagnetic flowmeterENVIMART
 
Proline promass 80 f 83f-endress+hauser datasheet-coriolis mass flowmeter
Proline promass 80 f 83f-endress+hauser datasheet-coriolis mass flowmeterProline promass 80 f 83f-endress+hauser datasheet-coriolis mass flowmeter
Proline promass 80 f 83f-endress+hauser datasheet-coriolis mass flowmeterENVIMART
 
Proline promag 50 w endress+hauser datasheet-electromagnetic flowmeter
Proline promag 50 w endress+hauser datasheet-electromagnetic flowmeterProline promag 50 w endress+hauser datasheet-electromagnetic flowmeter
Proline promag 50 w endress+hauser datasheet-electromagnetic flowmeterENVIMART
 
Proline t mass b 150-endress+hauser datasheet-thermal mass flow measuring system
Proline t mass b 150-endress+hauser datasheet-thermal mass flow measuring systemProline t mass b 150-endress+hauser datasheet-thermal mass flow measuring system
Proline t mass b 150-endress+hauser datasheet-thermal mass flow measuring systemENVIMART
 
Prosonic m fmu40 41 42_43_44-endress+hauser datasheet-ultrasonic level measur...
Prosonic m fmu40 41 42_43_44-endress+hauser datasheet-ultrasonic level measur...Prosonic m fmu40 41 42_43_44-endress+hauser datasheet-ultrasonic level measur...
Prosonic m fmu40 41 42_43_44-endress+hauser datasheet-ultrasonic level measur...ENVIMART
 
Proline promag w 400 endress+hauser datasheet-electromagnetic flowmeter
Proline promag w 400 endress+hauser datasheet-electromagnetic flowmeterProline promag w 400 endress+hauser datasheet-electromagnetic flowmeter
Proline promag w 400 endress+hauser datasheet-electromagnetic flowmeterENVIMART
 
Proline promag 55 s endress+hauser datasheet-electromagnetic flowmeter
Proline promag 55 s endress+hauser datasheet-electromagnetic flowmeterProline promag 55 s endress+hauser datasheet-electromagnetic flowmeter
Proline promag 55 s endress+hauser datasheet-electromagnetic flowmeterENVIMART
 
Proline promag w 800 endress+hauser datasheet-electromagnetic flowmeter
Proline promag w 800 endress+hauser datasheet-electromagnetic flowmeterProline promag w 800 endress+hauser datasheet-electromagnetic flowmeter
Proline promag w 800 endress+hauser datasheet-electromagnetic flowmeterENVIMART
 
Liquiphant m density densitycontroller fml621 endress+hauser datasheet
Liquiphant m density densitycontroller fml621 endress+hauser datasheetLiquiphant m density densitycontroller fml621 endress+hauser datasheet
Liquiphant m density densitycontroller fml621 endress+hauser datasheetENVIMART
 
Proline promag 50 h endress+hauser datasheet-electromagnetic flowmeter
Proline promag 50 h endress+hauser datasheet-electromagnetic flowmeterProline promag 50 h endress+hauser datasheet-electromagnetic flowmeter
Proline promag 50 h endress+hauser datasheet-electromagnetic flowmeterENVIMART
 
Proline promag 10 w endress+hauser datasheet-electromagnetic flowmeter
Proline promag 10 w endress+hauser datasheet-electromagnetic flowmeterProline promag 10 w endress+hauser datasheet-electromagnetic flowmeter
Proline promag 10 w endress+hauser datasheet-electromagnetic flowmeterENVIMART
 
Liquipoint t ftw31 32 endress+hauser datasheet-level limit switch for multipl...
Liquipoint t ftw31 32 endress+hauser datasheet-level limit switch for multipl...Liquipoint t ftw31 32 endress+hauser datasheet-level limit switch for multipl...
Liquipoint t ftw31 32 endress+hauser datasheet-level limit switch for multipl...ENVIMART
 

What's hot (20)

Proline proline prowirl 72 f 72w_73f_73w-endress+hauser datasheet-vortex flow...
Proline proline prowirl 72 f 72w_73f_73w-endress+hauser datasheet-vortex flow...Proline proline prowirl 72 f 72w_73f_73w-endress+hauser datasheet-vortex flow...
Proline proline prowirl 72 f 72w_73f_73w-endress+hauser datasheet-vortex flow...
 
Fmu90 ti
Fmu90 tiFmu90 ti
Fmu90 ti
 
Liquistation csf48 endress+hauser datasheet-automatic stationary sampler
Liquistation csf48 endress+hauser datasheet-automatic stationary samplerLiquistation csf48 endress+hauser datasheet-automatic stationary sampler
Liquistation csf48 endress+hauser datasheet-automatic stationary sampler
 
Proline promag 50 l endress+hauser datasheet-electromagnetic flowmeter
Proline promag 50 l endress+hauser datasheet-electromagnetic flowmeterProline promag 50 l endress+hauser datasheet-electromagnetic flowmeter
Proline promag 50 l endress+hauser datasheet-electromagnetic flowmeter
 
Proline promass 80 f 83f-endress+hauser datasheet-coriolis mass flowmeter
Proline promass 80 f 83f-endress+hauser datasheet-coriolis mass flowmeterProline promass 80 f 83f-endress+hauser datasheet-coriolis mass flowmeter
Proline promass 80 f 83f-endress+hauser datasheet-coriolis mass flowmeter
 
Proline promag 50 w endress+hauser datasheet-electromagnetic flowmeter
Proline promag 50 w endress+hauser datasheet-electromagnetic flowmeterProline promag 50 w endress+hauser datasheet-electromagnetic flowmeter
Proline promag 50 w endress+hauser datasheet-electromagnetic flowmeter
 
Proline t mass b 150-endress+hauser datasheet-thermal mass flow measuring system
Proline t mass b 150-endress+hauser datasheet-thermal mass flow measuring systemProline t mass b 150-endress+hauser datasheet-thermal mass flow measuring system
Proline t mass b 150-endress+hauser datasheet-thermal mass flow measuring system
 
Prosonic m fmu40 41 42_43_44-endress+hauser datasheet-ultrasonic level measur...
Prosonic m fmu40 41 42_43_44-endress+hauser datasheet-ultrasonic level measur...Prosonic m fmu40 41 42_43_44-endress+hauser datasheet-ultrasonic level measur...
Prosonic m fmu40 41 42_43_44-endress+hauser datasheet-ultrasonic level measur...
 
Proline promag w 400 endress+hauser datasheet-electromagnetic flowmeter
Proline promag w 400 endress+hauser datasheet-electromagnetic flowmeterProline promag w 400 endress+hauser datasheet-electromagnetic flowmeter
Proline promag w 400 endress+hauser datasheet-electromagnetic flowmeter
 
Proline promag 55 s endress+hauser datasheet-electromagnetic flowmeter
Proline promag 55 s endress+hauser datasheet-electromagnetic flowmeterProline promag 55 s endress+hauser datasheet-electromagnetic flowmeter
Proline promag 55 s endress+hauser datasheet-electromagnetic flowmeter
 
Proline promag w 800 endress+hauser datasheet-electromagnetic flowmeter
Proline promag w 800 endress+hauser datasheet-electromagnetic flowmeterProline promag w 800 endress+hauser datasheet-electromagnetic flowmeter
Proline promag w 800 endress+hauser datasheet-electromagnetic flowmeter
 
Liquiphant m density densitycontroller fml621 endress+hauser datasheet
Liquiphant m density densitycontroller fml621 endress+hauser datasheetLiquiphant m density densitycontroller fml621 endress+hauser datasheet
Liquiphant m density densitycontroller fml621 endress+hauser datasheet
 
Electromagnetic flowmeter - Proline Promag D 400
Electromagnetic flowmeter - Proline Promag D 400Electromagnetic flowmeter - Proline Promag D 400
Electromagnetic flowmeter - Proline Promag D 400
 
Proline Promag 10L-Electromagnetic Flowmeter
Proline Promag 10L-Electromagnetic FlowmeterProline Promag 10L-Electromagnetic Flowmeter
Proline Promag 10L-Electromagnetic Flowmeter
 
Electromagnetic flowmeter - Proline promag 10D
Electromagnetic flowmeter - Proline promag 10DElectromagnetic flowmeter - Proline promag 10D
Electromagnetic flowmeter - Proline promag 10D
 
Proline promag 50 h endress+hauser datasheet-electromagnetic flowmeter
Proline promag 50 h endress+hauser datasheet-electromagnetic flowmeterProline promag 50 h endress+hauser datasheet-electromagnetic flowmeter
Proline promag 50 h endress+hauser datasheet-electromagnetic flowmeter
 
Proline Promag L 800-Electromagnetic Flowmeter
Proline Promag L 800-Electromagnetic FlowmeterProline Promag L 800-Electromagnetic Flowmeter
Proline Promag L 800-Electromagnetic Flowmeter
 
Proline promag 10 w endress+hauser datasheet-electromagnetic flowmeter
Proline promag 10 w endress+hauser datasheet-electromagnetic flowmeterProline promag 10 w endress+hauser datasheet-electromagnetic flowmeter
Proline promag 10 w endress+hauser datasheet-electromagnetic flowmeter
 
Liquipoint t ftw31 32 endress+hauser datasheet-level limit switch for multipl...
Liquipoint t ftw31 32 endress+hauser datasheet-level limit switch for multipl...Liquipoint t ftw31 32 endress+hauser datasheet-level limit switch for multipl...
Liquipoint t ftw31 32 endress+hauser datasheet-level limit switch for multipl...
 
Electromagnetic flowmeter - Proline Promag 55H
Electromagnetic flowmeter - Proline Promag 55HElectromagnetic flowmeter - Proline Promag 55H
Electromagnetic flowmeter - Proline Promag 55H
 

Viewers also liked (6)

Flow indicator for bulk solids-Solimotion FTR20
Flow indicator for bulk solids-Solimotion FTR20Flow indicator for bulk solids-Solimotion FTR20
Flow indicator for bulk solids-Solimotion FTR20
 
Level limit switch-Soliswitch FTE30
Level limit switch-Soliswitch FTE30Level limit switch-Soliswitch FTE30
Level limit switch-Soliswitch FTE30
 
Microwave barrier-Soliwave M FQR50 / FDR50
Microwave barrier-Soliwave M FQR50 / FDR50Microwave barrier-Soliwave M FQR50 / FDR50
Microwave barrier-Soliwave M FQR50 / FDR50
 
Microwave barrier- Soliwave FQR56/FDR56
Microwave barrier- Soliwave FQR56/FDR56Microwave barrier- Soliwave FQR56/FDR56
Microwave barrier- Soliwave FQR56/FDR56
 
Liquiphant M Density and Density Computer FML621
Liquiphant M Density and Density Computer FML621Liquiphant M Density and Density Computer FML621
Liquiphant M Density and Density Computer FML621
 
Level Limit Switch-Soliswitch FTE31
Level Limit Switch-Soliswitch FTE31Level Limit Switch-Soliswitch FTE31
Level Limit Switch-Soliswitch FTE31
 

Similar to Ultrasonic flowmeter-Chemical or process applications

Proline prosonic flow 91 w endress+hauser datasheet-ultrasonic flowmeter
Proline prosonic flow 91 w endress+hauser datasheet-ultrasonic flowmeterProline prosonic flow 91 w endress+hauser datasheet-ultrasonic flowmeter
Proline prosonic flow 91 w endress+hauser datasheet-ultrasonic flowmeterENVIMART
 
Proline promag 10 p endress+hauser datasheet-electromagnetic flowmeter
Proline promag 10 p endress+hauser datasheet-electromagnetic flowmeterProline promag 10 p endress+hauser datasheet-electromagnetic flowmeter
Proline promag 10 p endress+hauser datasheet-electromagnetic flowmeterENVIMART
 
Proline promag 10 l endress+hauser datasheet-electromagnetic flowmeter
Proline promag 10 l endress+hauser datasheet-electromagnetic flowmeterProline promag 10 l endress+hauser datasheet-electromagnetic flowmeter
Proline promag 10 l endress+hauser datasheet-electromagnetic flowmeterENVIMART
 
Cerabar t pmc131 pmp131 pmp135-endress+hauser datasheet-process pressure meas...
Cerabar t pmc131 pmp131 pmp135-endress+hauser datasheet-process pressure meas...Cerabar t pmc131 pmp131 pmp135-endress+hauser datasheet-process pressure meas...
Cerabar t pmc131 pmp131 pmp135-endress+hauser datasheet-process pressure meas...ENVIMART
 
Proline promag d 400 endress+hauser datasheet-electromagnetic flowmeter
Proline promag d 400 endress+hauser datasheet-electromagnetic flowmeterProline promag d 400 endress+hauser datasheet-electromagnetic flowmeter
Proline promag d 400 endress+hauser datasheet-electromagnetic flowmeterENVIMART
 
Prosonic s fmu90 endress+hauser datasheet-transmitter in housing
Prosonic s fmu90 endress+hauser datasheet-transmitter in housingProsonic s fmu90 endress+hauser datasheet-transmitter in housing
Prosonic s fmu90 endress+hauser datasheet-transmitter in housingENVIMART
 

Similar to Ultrasonic flowmeter-Chemical or process applications (19)

Proline prosonic flow 91 w endress+hauser datasheet-ultrasonic flowmeter
Proline prosonic flow 91 w endress+hauser datasheet-ultrasonic flowmeterProline prosonic flow 91 w endress+hauser datasheet-ultrasonic flowmeter
Proline prosonic flow 91 w endress+hauser datasheet-ultrasonic flowmeter
 
Electromagnetic flowmeter - Proline Promag 10H
Electromagnetic flowmeter - Proline Promag 10HElectromagnetic flowmeter - Proline Promag 10H
Electromagnetic flowmeter - Proline Promag 10H
 
Ga dat cam bien
Ga dat cam bien Ga dat cam bien
Ga dat cam bien
 
Proline Promag 50L-Electromagnetic Flowmeter
Proline Promag 50L-Electromagnetic FlowmeterProline Promag 50L-Electromagnetic Flowmeter
Proline Promag 50L-Electromagnetic Flowmeter
 
Proline promag 10 p endress+hauser datasheet-electromagnetic flowmeter
Proline promag 10 p endress+hauser datasheet-electromagnetic flowmeterProline promag 10 p endress+hauser datasheet-electromagnetic flowmeter
Proline promag 10 p endress+hauser datasheet-electromagnetic flowmeter
 
Proline Promag 10P-Electromagnetic Flowmeter
Proline Promag 10P-Electromagnetic FlowmeterProline Promag 10P-Electromagnetic Flowmeter
Proline Promag 10P-Electromagnetic Flowmeter
 
Flowmeter Promag 10W-Electromagnetic-Water and Waste Water
Flowmeter Promag 10W-Electromagnetic-Water and Waste WaterFlowmeter Promag 10W-Electromagnetic-Water and Waste Water
Flowmeter Promag 10W-Electromagnetic-Water and Waste Water
 
Flowmeter Promag 55S-Electromagnetic-Endress Hauser
Flowmeter Promag 55S-Electromagnetic-Endress HauserFlowmeter Promag 55S-Electromagnetic-Endress Hauser
Flowmeter Promag 55S-Electromagnetic-Endress Hauser
 
Dosimag. Electromagnetic Flow Measuring System
Dosimag. Electromagnetic Flow Measuring SystemDosimag. Electromagnetic Flow Measuring System
Dosimag. Electromagnetic Flow Measuring System
 
Electromagnetic flowmeter - Proline Promag 50H, 53H
Electromagnetic flowmeter - Proline Promag 50H, 53HElectromagnetic flowmeter - Proline Promag 50H, 53H
Electromagnetic flowmeter - Proline Promag 50H, 53H
 
Ultrasonic Level Measurement-Prosonic T FMU30
Ultrasonic Level Measurement-Prosonic T FMU30Ultrasonic Level Measurement-Prosonic T FMU30
Ultrasonic Level Measurement-Prosonic T FMU30
 
Proline promag 10 l endress+hauser datasheet-electromagnetic flowmeter
Proline promag 10 l endress+hauser datasheet-electromagnetic flowmeterProline promag 10 l endress+hauser datasheet-electromagnetic flowmeter
Proline promag 10 l endress+hauser datasheet-electromagnetic flowmeter
 
Vortex flowmeter for gas, steam and liquids
Vortex flowmeter for gas, steam and liquidsVortex flowmeter for gas, steam and liquids
Vortex flowmeter for gas, steam and liquids
 
Cerabar t pmc131 pmp131 pmp135-endress+hauser datasheet-process pressure meas...
Cerabar t pmc131 pmp131 pmp135-endress+hauser datasheet-process pressure meas...Cerabar t pmc131 pmp131 pmp135-endress+hauser datasheet-process pressure meas...
Cerabar t pmc131 pmp131 pmp135-endress+hauser datasheet-process pressure meas...
 
pH.ORP Transmitter-Analog sensors-Digital sensors
pH.ORP Transmitter-Analog sensors-Digital sensorspH.ORP Transmitter-Analog sensors-Digital sensors
pH.ORP Transmitter-Analog sensors-Digital sensors
 
Radiometric measuring technology-Gammapilot FTG20
Radiometric measuring technology-Gammapilot FTG20Radiometric measuring technology-Gammapilot FTG20
Radiometric measuring technology-Gammapilot FTG20
 
Proline promag d 400 endress+hauser datasheet-electromagnetic flowmeter
Proline promag d 400 endress+hauser datasheet-electromagnetic flowmeterProline promag d 400 endress+hauser datasheet-electromagnetic flowmeter
Proline promag d 400 endress+hauser datasheet-electromagnetic flowmeter
 
Proline Promag P200-Electromagnetic Flowmeter
Proline Promag P200-Electromagnetic FlowmeterProline Promag P200-Electromagnetic Flowmeter
Proline Promag P200-Electromagnetic Flowmeter
 
Prosonic s fmu90 endress+hauser datasheet-transmitter in housing
Prosonic s fmu90 endress+hauser datasheet-transmitter in housingProsonic s fmu90 endress+hauser datasheet-transmitter in housing
Prosonic s fmu90 endress+hauser datasheet-transmitter in housing
 

More from Edress Hauser: Flow Meter, Level, Pressure, Temperature

More from Edress Hauser: Flow Meter, Level, Pressure, Temperature (20)

Level measurement with Prothermo NMT532
Level measurement with Prothermo NMT532 Level measurement with Prothermo NMT532
Level measurement with Prothermo NMT532
 
High accuracy liquid level measurement-Proservo NMS5
High accuracy liquid level measurement-Proservo NMS5High accuracy liquid level measurement-Proservo NMS5
High accuracy liquid level measurement-Proservo NMS5
 
Conductive Limit Detection-Rod Probe 11263
Conductive Limit Detection-Rod Probe 11263Conductive Limit Detection-Rod Probe 11263
Conductive Limit Detection-Rod Probe 11263
 
FTW360 compact pump protector
FTW360 compact pump protectorFTW360 compact pump protector
FTW360 compact pump protector
 
Conductive-Point level switch for multiple point
Conductive-Point level switch for multiple pointConductive-Point level switch for multiple point
Conductive-Point level switch for multiple point
 
Level limit switch compact design-Liquiphant T FTL20H
Level limit switch compact design-Liquiphant T FTL20HLevel limit switch compact design-Liquiphant T FTL20H
Level limit switch compact design-Liquiphant T FTL20H
 
Vibronic-Point level switch for all kinds of liquids
Vibronic-Point level switch for all kinds of liquidsVibronic-Point level switch for all kinds of liquids
Vibronic-Point level switch for all kinds of liquids
 
Vibronic-Point level switch-Liquiphant M FTL51C
Vibronic-Point level switch-Liquiphant M FTL51CVibronic-Point level switch-Liquiphant M FTL51C
Vibronic-Point level switch-Liquiphant M FTL51C
 
Level limit switch for liquids-Liquiphant T FTL20
Level limit switch for liquids-Liquiphant T FTL20Level limit switch for liquids-Liquiphant T FTL20
Level limit switch for liquids-Liquiphant T FTL20
 
Differential pressure measurement-Level transmiter
Differential pressure measurement-Level transmiterDifferential pressure measurement-Level transmiter
Differential pressure measurement-Level transmiter
 
Hydrostatic Level Measurement-Deltapilot S DB50/50L/51/52/53
Hydrostatic Level Measurement-Deltapilot S DB50/50L/51/52/53Hydrostatic Level Measurement-Deltapilot S DB50/50L/51/52/53
Hydrostatic Level Measurement-Deltapilot S DB50/50L/51/52/53
 
Hydrostatic level measurement-Deltapilot M FMB50/51/52/53
Hydrostatic level measurement-Deltapilot M FMB50/51/52/53Hydrostatic level measurement-Deltapilot M FMB50/51/52/53
Hydrostatic level measurement-Deltapilot M FMB50/51/52/53
 
Capacitance level measurement-Liquicap M FMI51, FMI52
Capacitance level measurement-Liquicap M FMI51, FMI52Capacitance level measurement-Liquicap M FMI51, FMI52
Capacitance level measurement-Liquicap M FMI51, FMI52
 
Ultrasonic Level Measurement-Prosonic M FMU40/41/42/43/44
Ultrasonic Level Measurement-Prosonic M FMU40/41/42/43/44Ultrasonic Level Measurement-Prosonic M FMU40/41/42/43/44
Ultrasonic Level Measurement-Prosonic M FMU40/41/42/43/44
 
Electromechanical Level Measuring System-Silopilot M FMM50
Electromechanical Level Measuring System-Silopilot M FMM50Electromechanical Level Measuring System-Silopilot M FMM50
Electromechanical Level Measuring System-Silopilot M FMM50
 
Electromechanical Level System-Silopilot T FMM20
Electromechanical Level System-Silopilot T FMM20Electromechanical Level System-Silopilot T FMM20
Electromechanical Level System-Silopilot T FMM20
 
Guided Level Radar-Levelflex M FMP40
Guided Level Radar-Levelflex M FMP40Guided Level Radar-Levelflex M FMP40
Guided Level Radar-Levelflex M FMP40
 
Level Radar-Tank Side Monitor NRF590
Level Radar-Tank Side Monitor NRF590Level Radar-Tank Side Monitor NRF590
Level Radar-Tank Side Monitor NRF590
 
Level radar measurement in liquids-Micropilot FMR50
Level radar measurement in liquids-Micropilot FMR50Level radar measurement in liquids-Micropilot FMR50
Level radar measurement in liquids-Micropilot FMR50
 
Automatic stationary sampler-Liquistation CSF34
Automatic stationary sampler-Liquistation CSF34Automatic stationary sampler-Liquistation CSF34
Automatic stationary sampler-Liquistation CSF34
 

Recently uploaded

Merck Moving Beyond Passwords: FIDO Paris Seminar.pptx
Merck Moving Beyond Passwords: FIDO Paris Seminar.pptxMerck Moving Beyond Passwords: FIDO Paris Seminar.pptx
Merck Moving Beyond Passwords: FIDO Paris Seminar.pptxLoriGlavin3
 
How to Effectively Monitor SD-WAN and SASE Environments with ThousandEyes
How to Effectively Monitor SD-WAN and SASE Environments with ThousandEyesHow to Effectively Monitor SD-WAN and SASE Environments with ThousandEyes
How to Effectively Monitor SD-WAN and SASE Environments with ThousandEyesThousandEyes
 
Modern Roaming for Notes and Nomad – Cheaper Faster Better Stronger
Modern Roaming for Notes and Nomad – Cheaper Faster Better StrongerModern Roaming for Notes and Nomad – Cheaper Faster Better Stronger
Modern Roaming for Notes and Nomad – Cheaper Faster Better Strongerpanagenda
 
Top 10 Hubspot Development Companies in 2024
Top 10 Hubspot Development Companies in 2024Top 10 Hubspot Development Companies in 2024
Top 10 Hubspot Development Companies in 2024TopCSSGallery
 
The Fit for Passkeys for Employee and Consumer Sign-ins: FIDO Paris Seminar.pptx
The Fit for Passkeys for Employee and Consumer Sign-ins: FIDO Paris Seminar.pptxThe Fit for Passkeys for Employee and Consumer Sign-ins: FIDO Paris Seminar.pptx
The Fit for Passkeys for Employee and Consumer Sign-ins: FIDO Paris Seminar.pptxLoriGlavin3
 
How to write a Business Continuity Plan
How to write a Business Continuity PlanHow to write a Business Continuity Plan
How to write a Business Continuity PlanDatabarracks
 
Transcript: New from BookNet Canada for 2024: Loan Stars - Tech Forum 2024
Transcript: New from BookNet Canada for 2024: Loan Stars - Tech Forum 2024Transcript: New from BookNet Canada for 2024: Loan Stars - Tech Forum 2024
Transcript: New from BookNet Canada for 2024: Loan Stars - Tech Forum 2024BookNet Canada
 
Decarbonising Buildings: Making a net-zero built environment a reality
Decarbonising Buildings: Making a net-zero built environment a realityDecarbonising Buildings: Making a net-zero built environment a reality
Decarbonising Buildings: Making a net-zero built environment a realityIES VE
 
So einfach geht modernes Roaming fuer Notes und Nomad.pdf
So einfach geht modernes Roaming fuer Notes und Nomad.pdfSo einfach geht modernes Roaming fuer Notes und Nomad.pdf
So einfach geht modernes Roaming fuer Notes und Nomad.pdfpanagenda
 
Genislab builds better products and faster go-to-market with Lean project man...
Genislab builds better products and faster go-to-market with Lean project man...Genislab builds better products and faster go-to-market with Lean project man...
Genislab builds better products and faster go-to-market with Lean project man...Farhan Tariq
 
Data governance with Unity Catalog Presentation
Data governance with Unity Catalog PresentationData governance with Unity Catalog Presentation
Data governance with Unity Catalog PresentationKnoldus Inc.
 
Connecting the Dots for Information Discovery.pdf
Connecting the Dots for Information Discovery.pdfConnecting the Dots for Information Discovery.pdf
Connecting the Dots for Information Discovery.pdfNeo4j
 
Microsoft 365 Copilot: How to boost your productivity with AI – Part one: Ado...
Microsoft 365 Copilot: How to boost your productivity with AI – Part one: Ado...Microsoft 365 Copilot: How to boost your productivity with AI – Part one: Ado...
Microsoft 365 Copilot: How to boost your productivity with AI – Part one: Ado...Nikki Chapple
 
Use of FIDO in the Payments and Identity Landscape: FIDO Paris Seminar.pptx
Use of FIDO in the Payments and Identity Landscape: FIDO Paris Seminar.pptxUse of FIDO in the Payments and Identity Landscape: FIDO Paris Seminar.pptx
Use of FIDO in the Payments and Identity Landscape: FIDO Paris Seminar.pptxLoriGlavin3
 
Unleashing Real-time Insights with ClickHouse_ Navigating the Landscape in 20...
Unleashing Real-time Insights with ClickHouse_ Navigating the Landscape in 20...Unleashing Real-time Insights with ClickHouse_ Navigating the Landscape in 20...
Unleashing Real-time Insights with ClickHouse_ Navigating the Landscape in 20...Alkin Tezuysal
 
Abdul Kader Baba- Managing Cybersecurity Risks and Compliance Requirements i...
Abdul Kader Baba- Managing Cybersecurity Risks  and Compliance Requirements i...Abdul Kader Baba- Managing Cybersecurity Risks  and Compliance Requirements i...
Abdul Kader Baba- Managing Cybersecurity Risks and Compliance Requirements i...itnewsafrica
 
Generative Artificial Intelligence: How generative AI works.pdf
Generative Artificial Intelligence: How generative AI works.pdfGenerative Artificial Intelligence: How generative AI works.pdf
Generative Artificial Intelligence: How generative AI works.pdfIngrid Airi González
 
Testing tools and AI - ideas what to try with some tool examples
Testing tools and AI - ideas what to try with some tool examplesTesting tools and AI - ideas what to try with some tool examples
Testing tools and AI - ideas what to try with some tool examplesKari Kakkonen
 
Design pattern talk by Kaya Weers - 2024 (v2)
Design pattern talk by Kaya Weers - 2024 (v2)Design pattern talk by Kaya Weers - 2024 (v2)
Design pattern talk by Kaya Weers - 2024 (v2)Kaya Weers
 
QCon London: Mastering long-running processes in modern architectures
QCon London: Mastering long-running processes in modern architecturesQCon London: Mastering long-running processes in modern architectures
QCon London: Mastering long-running processes in modern architecturesBernd Ruecker
 

Recently uploaded (20)

Merck Moving Beyond Passwords: FIDO Paris Seminar.pptx
Merck Moving Beyond Passwords: FIDO Paris Seminar.pptxMerck Moving Beyond Passwords: FIDO Paris Seminar.pptx
Merck Moving Beyond Passwords: FIDO Paris Seminar.pptx
 
How to Effectively Monitor SD-WAN and SASE Environments with ThousandEyes
How to Effectively Monitor SD-WAN and SASE Environments with ThousandEyesHow to Effectively Monitor SD-WAN and SASE Environments with ThousandEyes
How to Effectively Monitor SD-WAN and SASE Environments with ThousandEyes
 
Modern Roaming for Notes and Nomad – Cheaper Faster Better Stronger
Modern Roaming for Notes and Nomad – Cheaper Faster Better StrongerModern Roaming for Notes and Nomad – Cheaper Faster Better Stronger
Modern Roaming for Notes and Nomad – Cheaper Faster Better Stronger
 
Top 10 Hubspot Development Companies in 2024
Top 10 Hubspot Development Companies in 2024Top 10 Hubspot Development Companies in 2024
Top 10 Hubspot Development Companies in 2024
 
The Fit for Passkeys for Employee and Consumer Sign-ins: FIDO Paris Seminar.pptx
The Fit for Passkeys for Employee and Consumer Sign-ins: FIDO Paris Seminar.pptxThe Fit for Passkeys for Employee and Consumer Sign-ins: FIDO Paris Seminar.pptx
The Fit for Passkeys for Employee and Consumer Sign-ins: FIDO Paris Seminar.pptx
 
How to write a Business Continuity Plan
How to write a Business Continuity PlanHow to write a Business Continuity Plan
How to write a Business Continuity Plan
 
Transcript: New from BookNet Canada for 2024: Loan Stars - Tech Forum 2024
Transcript: New from BookNet Canada for 2024: Loan Stars - Tech Forum 2024Transcript: New from BookNet Canada for 2024: Loan Stars - Tech Forum 2024
Transcript: New from BookNet Canada for 2024: Loan Stars - Tech Forum 2024
 
Decarbonising Buildings: Making a net-zero built environment a reality
Decarbonising Buildings: Making a net-zero built environment a realityDecarbonising Buildings: Making a net-zero built environment a reality
Decarbonising Buildings: Making a net-zero built environment a reality
 
So einfach geht modernes Roaming fuer Notes und Nomad.pdf
So einfach geht modernes Roaming fuer Notes und Nomad.pdfSo einfach geht modernes Roaming fuer Notes und Nomad.pdf
So einfach geht modernes Roaming fuer Notes und Nomad.pdf
 
Genislab builds better products and faster go-to-market with Lean project man...
Genislab builds better products and faster go-to-market with Lean project man...Genislab builds better products and faster go-to-market with Lean project man...
Genislab builds better products and faster go-to-market with Lean project man...
 
Data governance with Unity Catalog Presentation
Data governance with Unity Catalog PresentationData governance with Unity Catalog Presentation
Data governance with Unity Catalog Presentation
 
Connecting the Dots for Information Discovery.pdf
Connecting the Dots for Information Discovery.pdfConnecting the Dots for Information Discovery.pdf
Connecting the Dots for Information Discovery.pdf
 
Microsoft 365 Copilot: How to boost your productivity with AI – Part one: Ado...
Microsoft 365 Copilot: How to boost your productivity with AI – Part one: Ado...Microsoft 365 Copilot: How to boost your productivity with AI – Part one: Ado...
Microsoft 365 Copilot: How to boost your productivity with AI – Part one: Ado...
 
Use of FIDO in the Payments and Identity Landscape: FIDO Paris Seminar.pptx
Use of FIDO in the Payments and Identity Landscape: FIDO Paris Seminar.pptxUse of FIDO in the Payments and Identity Landscape: FIDO Paris Seminar.pptx
Use of FIDO in the Payments and Identity Landscape: FIDO Paris Seminar.pptx
 
Unleashing Real-time Insights with ClickHouse_ Navigating the Landscape in 20...
Unleashing Real-time Insights with ClickHouse_ Navigating the Landscape in 20...Unleashing Real-time Insights with ClickHouse_ Navigating the Landscape in 20...
Unleashing Real-time Insights with ClickHouse_ Navigating the Landscape in 20...
 
Abdul Kader Baba- Managing Cybersecurity Risks and Compliance Requirements i...
Abdul Kader Baba- Managing Cybersecurity Risks  and Compliance Requirements i...Abdul Kader Baba- Managing Cybersecurity Risks  and Compliance Requirements i...
Abdul Kader Baba- Managing Cybersecurity Risks and Compliance Requirements i...
 
Generative Artificial Intelligence: How generative AI works.pdf
Generative Artificial Intelligence: How generative AI works.pdfGenerative Artificial Intelligence: How generative AI works.pdf
Generative Artificial Intelligence: How generative AI works.pdf
 
Testing tools and AI - ideas what to try with some tool examples
Testing tools and AI - ideas what to try with some tool examplesTesting tools and AI - ideas what to try with some tool examples
Testing tools and AI - ideas what to try with some tool examples
 
Design pattern talk by Kaya Weers - 2024 (v2)
Design pattern talk by Kaya Weers - 2024 (v2)Design pattern talk by Kaya Weers - 2024 (v2)
Design pattern talk by Kaya Weers - 2024 (v2)
 
QCon London: Mastering long-running processes in modern architectures
QCon London: Mastering long-running processes in modern architecturesQCon London: Mastering long-running processes in modern architectures
QCon London: Mastering long-running processes in modern architectures
 

Ultrasonic flowmeter-Chemical or process applications

  • 1. TI00083D/06/EN/13.11 71136721 Technical Information Proline Prosonic Flow 93P Ultrasonic flow measuring system Volume flow measurement of liquids in chemical or process applications Applications The sensors are perfectly suited for the non-contact measurement of pure or slightly contaminated liquids, regardless of the pressure or electrical conductivity. • Suitable for pipe diameters from DN 15 to 4000 (½ to 160") • Suitable for fluid temperatures ranging from –40 to +170 °C (–40 to +338 °F) • Can be used with all metal and plastic pipes lined or unlined and with composite pipes • Ideal solution for applications with chemicals, solvents, liquid hydrocarbons, acids and alkalis • Particularly suitable for corrosive, hazardous or ultrapure fluids thanks to non-contact measurement concept • Interface for easy integration into all common distributed control systems: – HART – PROFIBUS DP/PA – FOUNDATION Fieldbus • Approval for use in hazardous area Zone 1 (ATEX, FM, CSA) available as standard Features and benefits The Prosonic Flow ultrasonic clamp-on system allows accurate and cost-effective flow measurement from outside the pipe and without the need to interrupt the process. The flow measurement is bidirectional and causes no pressure loss. • Easy, safe and menu-guided sensor mounting ensures precise measuring results • Long-term system integrity thanks to robust sensor and industrial mounting kit design • Automatic frequency scan for optimized installation and maximum measuring performance • Remote configuration using Endress+Hauser's FieldCare software
  • 2. Proline Prosonic Flow 93P 2 Endress+Hauser Table of contents Function and system design. . . . . . . . . . . . . . . . . . . . . 3 Measuring principle . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3 Measuring system . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3 Sensor selection and arrangement . . . . . . . . . . . . . . . . . . . . . . . . . 5 Two-channel operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5 Input . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7 Measured variable . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7 Measuring range . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7 Operable flow range . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7 Input signal . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7 Output . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7 Output signal . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7 Signal on alarm . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8 Load . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8 Switching output . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8 Low flow cutoff . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8 Galvanic isolation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8 Power supply. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9 Electrical connection, measuring unit . . . . . . . . . . . . . . . . . . . . . . 9 Electrical connection, terminal assignment . . . . . . . . . . . . . . . . . 10 Connecting the connecting cable . . . . . . . . . . . . . . . . . . . . . . . . 10 Supply voltage . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11 Cable entry . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11 Connecting cable (sensor/transmitter) . . . . . . . . . . . . . . . . . . . . 12 Power consumption . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12 Power supply failure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12 Potential equalization . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12 Performance characteristics. . . . . . . . . . . . . . . . . . . . 12 Reference operating conditions . . . . . . . . . . . . . . . . . . . . . . . . . . 12 Maximum measured error . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12 Repeatability . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13 Operating conditions: installation . . . . . . . . . . . . . . . 13 Installation instructions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13 Inlet and outlet runs . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14 Operating conditions: environment . . . . . . . . . . . . . . 15 Ambient temperature range . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15 Storage temperature . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15 Degree of protection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15 Shock and vibration resistance . . . . . . . . . . . . . . . . . . . . . . . . . . 15 Electromagnetic compatibility (EMC) . . . . . . . . . . . . . . . . . . . . . 15 Operating conditions: process . . . . . . . . . . . . . . . . . . 16 Medium temperature range . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16 Medium pressure range (nominal pressure) . . . . . . . . . . . . . . . . . 16 Pressure loss . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16 Mechanical construction . . . . . . . . . . . . . . . . . . . . . . 17 Design, dimensions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17 Weight . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 24 Materials . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 24 Human interface . . . . . . . . . . . . . . . . . . . . . . . . . . . . 25 Display elements . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 25 Operating elements . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 25 Language group . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 25 Remote operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 25 Certificates and approvals . . . . . . . . . . . . . . . . . . . . . 25 CE mark . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 25 C-Tick mark . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 25 Ex approval . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 25 PROFIBUS DP/PA certification . . . . . . . . . . . . . . . . . . . . . . . . . 25 FOUNDATION Fieldbus certification . . . . . . . . . . . . . . . . . . . . . 25 Other standards and guidelines . . . . . . . . . . . . . . . . . . . . . . . . . . 26 Ordering information. . . . . . . . . . . . . . . . . . . . . . . . . 26 Accessories . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 27 Device-specific accessories . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 27 Measuring principle-specific accessories . . . . . . . . . . . . . . . . . . . 27 Communication-specific accessories . . . . . . . . . . . . . . . . . . . . . . 29 Service-specific accessories . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 30 Documentation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 31 Registered trademarks. . . . . . . . . . . . . . . . . . . . . . . . 31
  • 3. Proline Prosonic Flow 93P Endress+Hauser 3 Function and system design Measuring principle The measuring system operates on the principle of transit time difference. In this measurement method, acoustic (ultrasonic) signals are transmitted between two sensors. The signals are sent in both directions, i.e. the sensor in question works as both a sound transmitter and a sound receiver. As the propagation velocity of the waves is less when the waves travel against the direction of flow than along the direction of flow, a transit time difference occurs. This transit time difference is directly proportional to the flow velocity. A0005961 Principle of the transit time difference measurement method Q = v · A a Sensor b Sensor Q Volume flow v Flow velocity (v & t ) t Transit time difference (t = ta – tb) A Pipe cross-sectional area The measuring system calculates the volume flow of the fluid from the measured transit time difference and the pipe cross-sectional area. In addition to measuring the transit time difference, the system simultaneously measures the sound velocity of the fluid. This additional measured variable can be used to distinguish different fluids or as a measure of product quality. The measuring device can be configured onsite to suit the specific application using Quick Setup menus. Measuring system The measuring system consists of one transmitter and two sensors. The transmitter is used both to control the sensors and to prepare, process and evaluate the measuring signals, and to convert the signals to a desired output variable. The transmitter is optionally equipped for two-channel operation  ä 5. The sensors work as sound transmitters and sound receivers. Depending on the application and version, the sensors can be arranged for measurement via one or two traverses  ä 5. b a Q ta tb
  • 4. Proline Prosonic Flow 93P 4 Endress+Hauser Transmitter Sensor Mounting accessories The requisite mounting distances must be determined for the sensors. Information on the fluid, the pipe material used and the exact pipe dimensions is needed to determine these values. The values for the sound velocity of the following fluids, pipe materials and lining materials are stored in the transmitter: If you are using fluids, pipe materials or lining materials other than those listed in the table, and you do not have the corresponding sound velocities for these fluids/materials, you can use the DDU18 and DDU19 sensors to determine the values. Prosonic Flow 93 wall-mount housing Prosonic Flow 93 field housing For mounting in non-hazardous zone, Ex Zone 2 A0009629 For mounting in Ex Zone 1 A0009676 Prosonic Flow P Prosonic Flow P DN 15 to 65 (½ to 2½") A0011484 DN 50 to 4000 (2 to 160") A0013475 Fluid Pipe material Lining • Water • Sea water • Distilled water • Ammonia • Alcohol • Benzene • Bromide • Ethanol • Glycol • Kerosene • Milk • Methanol • Toluene • Lube oil • Diesel • Gasoline • Carbon steel • Ductile iron • Stainless steel • Alloy C • PVC • PE • LDPE • HDPE • GRP • PVDF • PA • PP • PTFE • Glass pyrex • Asbestos cement • Copper • Mortar • Rubber • Tar Epoxy DDU18 (sound velocity measurement) DDU19 (wall thickness measurement) Nominal diameter range: DN 50 to 3000 (2 to 120") A0009784 Wall thickness range: • Steel pipes: 2 to 50 mm (0.08 to 2.0") • Plastic pipes: 4 to 15 mm (0.16 to 0.60") (only suitable for PTFE and PE pipes to a certain extent) A0009673
  • 5. Proline Prosonic Flow 93P Endress+Hauser 5 Sensor selection and arrangement The sensors can be arranged in two ways: • Mounting arrangement for measurement via one traverse: the sensors are located on opposite sides of the pipe. • Mounting arrangement for measurement via two traverses: the sensors are located on the same side of the pipe. A0001108 Sensor mounting arrangement A Mounting arrangement for measurement via one traverse B Mounting arrangement for measurement via two traverses The number of traverses required depends on the sensor type, the nominal diameter and the thickness of the pipe wall. We recommend the following types of mounting: Two-channel operation The transmitter is able to operate two independent measuring channels (measuring channel 1 and measuring channel 2). A pair of sensors is connected per measuring channel. Both measuring channels operate independently of one another and are supported by the transmitter to an equal extent. Two-channel operation can be used for the following measurements: • Two-channel measurement = flow measurement at two separate measuring points • Two-path measurement = redundant flow measurement at one measuring point BA Sensor Type Nominal Diameter Sensor Frequency Sensor ID Type of Mounting 1) Prosonic Flow P DN 15 to 65 (½ to 2½") 6 MHz P-CL-6F* 2 traverses 3) DN 50 to 65 (2 to 2½") 6 MHz (or 2 MHz) P-CL-6F* P-CL-2F* 2 (or 1) traverses 2) DN 80 (3") 2 MHz P-CL-2F* 2 traverses DN 100 to 300 (4 to 12") 2 MHz (or 1 MHz) P-CL-2F* P-CL-1F* 2 traverses DN 300 to 600 (12 to 24") 1 MHz (or 2 MHz) P-CL-1F* P-CL-2F* 2 traverses DN 650 to 4000 (26 to 160") 1 MHz P-CL-1F* 1 traverse 1) The installation of clamp-on sensors is principally recommended in the 2 traverse type installation. This type of installation allows the easiest and most comfortable type of mounting and means that a system can also be mounted even if the pipe can only be accessed from one side. However, in certain applications a 1 traverse installation may be preferred. These include: • Certain plastic pipes with wall thickness > 4 mm (0.16") • Pipes made of composite materials such as GRP • Lined pipes • Applications with fluids with high acoustic damping 2) If the pipe nominal diameter is small (DN 65 / 2½" and smaller), the sensor spacing with Prosonic Flow P can be too small for two traverse installation using sensor P-CL-2F*. In this case, the 1 traverse type of installation must be used. 3) 6 MHz sensors for applications where flow velocity 10 m/s (32.8 ft/s).
  • 6. Proline Prosonic Flow 93P 6 Endress+Hauser Two-channel measurement The flow is measured at two separate measuring points in the case of two-channel measurement. The measured values of the two measuring channels can be processed and displayed differently. The following measured values can be output for two-channel measurement: • Individual measured values per measuring channel (output independently of one another) • The difference between the two measured values • The sum of the two measured values The two measuring channels can be configured individually. This makes it possible to independently configure and select the display, outputs, sensor type and type of installation. A0001159 Two-channel measurement: example of arranging sensor pairs at two separate measuring points A Measuring channel 1: mounting the sensor pair for measurement via two traverses B Measuring channel 2: mounting the sensor pair for measurement via one traverse Two-path measurement The flow is measured redundantly at one measuring point in the case of two-path measurement. The measured values of the two measuring channels can be processed and displayed differently. The following measured values can be output for two-path measurement: • Individual measured values per measuring channel (output independently of one another) • The average of the two measured values The "Averaging" function generally provides you with a more stable measured value. The function is thus suitable for measurements under conditions that are not ideal (e.g. short inlet runs). The two measuring channels can be configured individually. This makes it possible to independently configure and select the display, outputs, sensor type and type of installation. It is generally not necessary to individually configure the two measuring channels in the case of two-path measurement. However, in certain situations individual channel configuration can be used to balance out application-specific asymmetries. A0001160 Two-path measurement: examples of arranging sensor pairs at one measuring point A Measuring channel 1 and measuring channel 2: mounting the sensor pair for measurement via two traverses B Measuring channel 1 and measuring channel 2: mounting the sensor pair for measurement via one traverse BA BA
  • 7. Proline Prosonic Flow 93P Endress+Hauser 7 Input Measured variable Flow velocity (transit time difference proportional to flow velocity) Measuring range Typically v = 0 to 15 m/s (0 to 50 ft/s) Operable flow range Over 150 : 1 Input signal Status input (auxiliary input) U = 3 to 30 V DC, Ri = 5 k, galvanically isolated Configurable for: totalizer(s) reset, measured-value suppression, error-message reset Output Output signal Current output • Galvanically isolated • Active/passive selectable – Active: 0/4 to 20 mA, RL < 700  (for HART: RL 250 ) – Passive: 4 to 20 mA, max. 30 V DC, Ri 150  • Time constant selectable (0.01 to 100 s) • Full scale value adjustable • Temperature coefficient: typ. 0.005 % o.r./°C (o.r. = of reading) • Resolution: 0.5 μA Pulse/frequency output • Galvanically isolated • Active/passive selectable – Active: 24 V DC, 25 mA (max. 250 mA during 20 ms), RL > 100  – Passive: open collector, 30 V DC, 250 mA • Time constant selectable (0.05 to 100 s) • Frequency output – End frequency: 2 to 10000 Hz (fmax = 12500 Hz) – End frequency for EEx ia 2 to 5000 Hz – On/off ratio 1:1, pulse width max. 10 s • Pulse output – Pulse value and pulse polarity selectable – Max. pulse width adjustable (0.05 to 2000 ms) – As of a frequency of 1 / (2 × pulse width), the on/off ratio is 1:1 PROFIBUS DP interface • PROFIBUS DP in accordance with EN 50170 Volume 2 • Profile version 3.0 • Data transmission rate: 9.6 kBaud to 12 MBaud • Automatic data transmission rate recognition • Signal encoding = NRZ Code • Function blocks: 8 × Analog Input (AI), 3 × Totalizer • Output data: volume flow channel 1 or channel 2, sound velocity channel 1 or channel 2, flow velocity channel 1 or channel 2, average volume flow, average sound velocity, average flow velocity, volume flow sum, volume flow difference, totalizer 1 to 3 • Input data: positive zero return (ON/OFF), zero point adjustment, measuring mode, totalizer control • Bus address adjustable via miniature switches or local display (optional) at the measuring device • Available output combination  ä 9
  • 8. Proline Prosonic Flow 93P 8 Endress+Hauser PROFIBUS PA interface • PROFIBUS PA in accordance with EN 50170 Volume 2, IEC 61158-2 (MBP) • Galvanically isolated • Data transmission rate, supported baudrate: 31.25 kBit/s • Current consumption = 11 mA • Error current FDE (fault disconnection electronic) = 0 mA • Signal encoding = Manchester II • Function blocks: 8 × Analog Input (AI), 3 × Totalizer • Output data: volume flow channel 1 or channel 2, sound velocity channel 1 or channel 2, flow velocity channel 1 or channel 2, average volume flow, average sound velocity, average flow velocity, volume flow sum, volume flow difference, totalizer 1 to 3 • Input data: positive zero return (ON/OFF), operation control, totalizer control, zero point adjustment control, display value • Bus address can be set via DIP switch on device FOUNDATION Fieldbus interface • FOUNDATION Fieldbus H1, IEC 61158-2 • Galvanically isolated • Data transmission rate, supported baudrate: 31.25 kBit/s • Current consumption = 12 mA • Error current FDE (fault disconnection electronic) = 0 mA • Signal encoding = Manchester II • Function blocks: 8 × Analog Input (AI), 1 × Discrete Output, 1 × PID • Output data: volume flow channel 1 or channel 2, sound velocity channel 1 or channel 2, flow velocity channel 1 or channel 2, signal strength channel 1 or 2, average volume flow, average sound velocity, average flow velocity, volume flow sum, difference, volume flow, totalizer 1 to 3 • Input data: positive zero return (ON/OFF), reset totalizer, zero point adjustment control • Link master function (LAS) is supported Signal on alarm • Current output  failsafe mode selectable • Pulse/frequency output  failsafe mode selectable • Relay output  "deenergized" in the event of a fault or if the power supply fails Load See "Output signal" Switching output Relay output • NC or NO contact available Factory setting: relay 1 = NO contact, relay 2 = NC contact • Max. 30 V / 0.5 A AC; 60 V / 0.1 A DC • Galvanically isolated • Configurable for: error messages, flow direction, limit values Low flow cutoff Switch points for low flow are selectable Galvanic isolation All circuits for inputs, outputs, and power supply are galvanically isolated from each other.
  • 9. Proline Prosonic Flow 93P Endress+Hauser 9 Power supply Electrical connection, measuring unit A0013851 Connecting the transmitter, cable cross-section max. 2.5 mm2 (14 AWG) A Aluminum field housing B Wall-mount housing *) Permanent assignment communication boards **) Flexible assignment communication boards a Cover of the connection compartment b Cable for power supply: 85 to 260 V AC, 20 to 55 V AC, 16 to 62 V DC Terminal No. 1: L1 for AC, L+ for DC Terminal No. 1: N for AC, L- for DC c Ground terminal for protective ground d Signal cable: see terminal assignment Fieldbus: - Terminal 26: DP (B) / PA (+), FF (+) with reverse polarity protection - Terminal 27: DP (A) / PA (–), FF (–) with reverse polarity protection e Ground terminal for signal cable / fieldbus cable f Service connector for connecting service interface FXA 193 (Fieldcheck, FieldCare) g Signal cable: see terminal assignment Cable for external termination (only for PROFIBUS DP with permanent assignment communication board) - Terminal No. 24: +5V - Terminal No. 25: DGND HART* PROFIBUS DP** 27 25 23 21 2 1 26 24 22 20 L1 (L+) N (L-) – – – + + + PROFIBUS PA* FOUNDATION Fieldbus* PA(–)/FF(–) 27 25 23 21 2 1 26 24 22 20 L1 (L+) N (L-) – – – + + + A (RxD/TxD-N) B (RxD/TxD-P) PA(+)/FF(+) d c e b – 27 – 25 – 23 – 21 + 26 + 24 + 22 + 20 N (L-) 2 L1 (L+)1 25 23 21 2 1 24 22 20 L1 (L+) N (L-) – – – + + + 27 26 A (RxD/TxD-N) B (RxD/TxD-P) d c e b d c e b g d c e b g PROFIBUS DP* f f f f a a A B d b g (d)b d/(g)
  • 10. Proline Prosonic Flow 93P 10 Endress+Hauser Electrical connection, terminal assignment The inputs and outputs on the communication board can be either permanently assigned or variable, depending on the version ordered (see table). Replacements for modules which are defective or which have to be replaced can be ordered as accessories. Connecting the connecting cable Connecting the wall-mount housing A0011390 Connecting the connecting cable, wall-mount housing 1 Channel 1 upstream 2 Channel 1 downstream 3 Channel 2 upstream (only available when ordered) 4 Channel 2 downstream (only available when ordered) Order variant Terminal No. (inputs/outputs) 20 (+) / 21 (–) 22 (+) / 23 (–) 24 (+) / 25 (–) 26 (+) / 27 (–) Fixed communication boards (fixed assignment) 93***-***********A – – Frequency output Current output HART 93***-***********B Relay output 2 Relay output 1 Frequency output Current output HART 93***-***********F – – – PROFIBUS PA, Ex i 93***-***********G – – – FOUNDATION Fieldbus, Ex i 93***-***********H – – – PROFIBUS PA 93***-***********J – – – PROFIBUS DP 93***-***********K – – – FOUNDATION Fieldbus 93***-***********S – – Frequency output, Ex i Current output, Ex i, passive, HART 93***-***********T – – Frequency output, Ex i Current output, Ex i, passive, HART Flexible communication boards 93***-***********C Relay output 2 Relay output 1 Frequency output Current output HART 93***-***********D Status input Relay output Frequency output Current output HART 93***-***********L Status input Relay output 2 Relay output 1 Current output HART 93***-***********M Status input Frequency output Frequency output Current output HART 93***-***********P Current output Frequency output Status input PROFIBUS DP 93***-***********V Relay output 2 Relay output 1 Status input PROFIBUS DP 93***-***********W Relay output Current output Current output Current output HART 93***-***********2 Relay output Current output Frequency output Current output HART 93***-***********4 Current input Relay output Frequency output Current output HART 93***-***********6 Relay output Relay output Current output Current output HART A 1 A 32 4
  • 11. Proline Prosonic Flow 93P Endress+Hauser 11 Connecting the field housing A0008314 Connecting the connecting cable, field housing 1 Channel 1 upstream 2 Channel 1 downstream 3 Channel 2 upstream (only available when ordered) 4 Channel 2 downstream (only available when ordered) Supply voltage Transmitter HART • 85 to 260 V AC, 45 to 65 Hz • 20 to 55 V AC, 45 to 65 Hz • 16 to 62 V DC PROFIBUS DP/PA • 9 to 32 V DC FOUNDATION Fieldbus • 9 to 32 V DC Sensor Powered by the transmitter Cable entry Power supply and signal cables (inputs/outputs) • Cable gland M20 × 1.5 – Cable gland for cables 8 to 12 mm (0.31 to 0.47") – Cable gland for cables 6 to 12 mm (0.24 to 0.47") • Thread for cable entry ½" NPT, G ½" Connecting cable (sensor/transmitter) Cable gland for one connecting cable per cable entry, 1 × 8 mm (1 × 0.31") • Cable gland M20 × 1.5 • Thread for cable entry ½" NPT, G ½" Cable gland for two connecting cables per cable entry, 2 × 4 mm (2 × 0.16") • Cable gland M20 × 1.5 • Thread for cable entry ½" NPT, G ½" A0008152 Cable gland for two connecting cables per cable entry A 1 2 3 4 A
  • 12. Proline Prosonic Flow 93P 12 Endress+Hauser Connecting cable (sensor/ transmitter) Only use the connecting cables supplied by Endress+Hauser. Different versions of the connecting cables are available  ä 27. • Cable material: – Prosonic Flow 93P (DN 50 to 4000 / 2 to 160"): PVC (standard) or PTFE (for higher temperatures) – Prosonic Flow 93P (DN 15 to 65 / ½ to 2½"): TPE-V • Cable length: – For use in a non-hazardous zone: 5 to 60 m (16.4 to 196.8 ft) – For use in a hazardous zone: 5 to 30 m (16.4 to 98.4 ft) ! Note! To ensure correct measuring results, route the connecting cable well clear of electrical machines and switching elements. Power consumption AC: < 18 VA (incl. sensor) DC: < 10 W (incl. sensor) Switch-on current • max. 13.5 A (< 50 ms) at 24 V DC • max. 3 A (< 5 ms) at 260 V AC Power supply failure Lasting min. 1 power cycle HistoROM/T-DAT save measuring system data if the power supply fails Potential equalization For potential equalization, no special measures are necessary. Performance characteristics Reference operating conditions • Fluid temperature: +20 to +30 °C • Ambient temperature: +22 °C ± 2 K • Warm-up period: 30 minutes • Sensors and transmitter are grounded • The measuring sensors are correctly installed Maximum measured error The measured error depends on a number of factors. A distinction is made between the measured error of the device (Prosonic Flow 93 = 0.5 % of the measured value) and an additional installation-specific measured error (typically 1.5 % of the measured value) that is independent of the device. The installation-specific measured error depends on the installation conditions on site, such as the nominal diameter, wall thickness, real pipe geometry, fluid, etc. The sum of the two measured errors is the measured error at the measuring point. A0011347 Example of the measured error in a pipe with a nominal diameter DN > 200 (8") a Measured error of the device (0.5 % o.r. ± 3 mm/s) b Measured error due to installation conditions (typically 1.5 % o.r.) c Measured error at the measuring point: 0.5 % o.r. ± 3 mm/s + 1.5 % o.r. = 2 % o.r. ± 3 mm/s 0 0.5 2.0 3.5 % m/s2 4 6 8 10 12 b c a
  • 13. Proline Prosonic Flow 93P Endress+Hauser 13 Measured error at the measuring point The measured error at the measuring point is made up of the measured error of the device (0.5 % o.r.) and the measured error resulting from the installation conditions on site. Given a flow velocity > 0.3 m/s (1 ft/s) and a Reynolds number > 10000, the following are typical error limits: Measurement Report If required, the device can be supplied with a factory measurement report. To certify the performance of the device, a measurement is performed under reference conditions. Here, the sensors are mounted on a pipe with a nominal diameter of DN 15 (½"), DN 25 (1"), DN 40 (1½"), DN 50 (2") or DN 100 (4") respectively. The measurement report guarantees the following error limits of the device [at a flow velocity > 0.3 m/s (1 ft/s) and a Reynolds number > 10000]: Repeatability ±0.3 % for flow velocities > 0.3 m/s (1 ft/s) Operating conditions: installation Installation instructions Mounting location Correct flow measurement is possible only if the pipe is full. It is preferable to install the sensors in a riser. ! Note! Entrained air or gas bubbles in the measuring tube can result in an increase in measuring errors. For this reason, avoid the following mounting locations: • Highest point of a pipeline. Risk of air accumulating. • Directly upstream of a free pipe outlet in a vertical pipe. Risk of partial pipe filling. A0001103 Nominal diameter Device error limits + Installation-specific error limits (typical)  Error limits at the measuring point (typical) DN 15 (½") ±0.5 % o.r. ± 5 mm/s + ±2.5 % o.r.  ±3 % o.r. ± 5 mm/s DN 25 to 200 (1 to 8") ±0.5 % o.r. ± 7.5 mm/s + ±1.5 % o.r.  ±2 % o.r. ± 7.5 mm/s > DN 200 (8") ±0.5 % o.r. ± 3 mm/s + ±1.5 % o.r.  ±2 % o.r. ± 3 mm/s o.r. = of reading Nominal diameter Guaranteed error limits of the device DN 15 (½"), DN 25 (1"), DN 40 (1½"), DN 50 (2") ±0.5 % o.r. ± 5 mm/s DN 100 (4") ±0.5 % o.r. ± 7.5 mm/s o.r. = of reading
  • 14. Proline Prosonic Flow 93P 14 Endress+Hauser Orientation Vertical Recommended orientation with upward direction of flow (View A). With this orientation, entrained solids will sink and gases will rise away from the sensor when the fluid is stagnant. The piping can be completely drained and protected against solids buildup. Horizontal In the recommended installation range in a horizontal installation position (View B), gas and air collections at the pipe cover and problematic deposits at the bottom of the pipe have a smaller influence on measurement. A0001105 A Recommended orientation with upward direction of flow B Recommended installation range with horizontal orientation C Recommended installation range max. 120° Inlet and outlet runs If possible, install the sensor well clear of fittings such as valves, T-pieces, elbows, etc. Compliance with the following inlet and outlet runs is required in order to ensure measuring accuracy. A0013459 Inlet and outlet runs (top view) 1 Valve (2/3 open) 2 Pump 3 Two pipe bends in different directions A B C C ³ 15 x DN ³ 20 x DN ³ 20 x DN ³ 15 x DN ³ 3 x DN 1 2 3
  • 15. Proline Prosonic Flow 93P Endress+Hauser 15 Operating conditions: environment Ambient temperature range Transmitter • Standard: –20 to +60 °C (–4 to +140 °F) • Optional: –40 to +60 °C (–40 to +140 °F) Prosonic Flow P sensor Prosonic Flow P (DN 15 to 65 / ½ to 2½") • Standard: –40 to +100 °C (–40 to +212 °F) • Optional: –40 to +150 °C (–40 to +302 °F) Prosonic Flow P (DN 50 to 4000 / 2 to 160") • Standard: –40 to +80 °C (–40 to +176 °F) • Optional: 0 to +170 °C (+32 to +338 °F) DDU18 sensor (accessories: sound velocity measurement) • Standard: –40 to +80 °C (–40 to +176 °F) • Optional: 0 to +170 °C (+32 to +338 °F) DDU19 sensor (accessories: wall thickness measurement) –20 to +60 °C (–4 to +140 °F) Connecting cable (sensor/transmitter) Prosonic Flow 93P (DN 15 to 65 / ½ to 2½"): • Standard (TPE-V): –40 to +80 °C (–40 to +176 °F) Prosonic Flow 93P (DN 50 to 4000 / 2 to 160"): • Standard (PVC): –20 to +70 °C (–4 to +158 °F) • Optional (PTFE): –40 to +170 °C (–40 to +338 °F) ! Note! • It is permitted to insulate the sensors mounted on the pipe. • Mount the transmitter in a shady location and avoid direct sunlight, particularly in warm climatic regions. Storage temperature The storage temperature corresponds to the ambient temperature range. Degree of protection Transmitter IP 67 (NEMA 4X) Sensor IP 68 (NEMA 6P) DDU18 sensor (accessories: sound velocity measurement) IP 68 (NEMA 6P) DDU19 sensor (accessories: wall thickness measurement) IP 67 (NEMA 4X) Shock and vibration resistance According to IEC 68-2-6 Electromagnetic compatibility (EMC) Electromagnetic compatibility (EMC requirements) according to IEC/EN 61326 "Emission to class A requirements" and NAMUR Recommendation NE 21 and NE 43.
  • 16. Proline Prosonic Flow 93P 16 Endress+Hauser Operating conditions: process Medium temperature range Prosonic Flow P sensor Prosonic Flow P (DN 15 to 65) • Standard: –40 to +100 °C (–40 to +212 °F) • Optional: –40 to +150 °C (–40 to +302 °F) Prosonic Flow P (DN 50 to 4000) • Standard: –40 to +80 °C (–40 to +176 °F) • Optional: 0 to +170 °C (+32 to +338 °F) DDU18 sensor (accessories: sound velocity measurement) • Standard: –40 to +80 °C (–40 to +176 °F) • Optional: 0 to +170 °C (+32 to +338 °F) DDU19 sensor (accessories: wall thickness measurement) –20 to +60 °C (–4 to +140 °F) Medium pressure range (nominal pressure) No pressure limitation, however perfect measurement requires that the static fluid pressure is higher than vapor pressure. Pressure loss There is no pressure loss.
  • 17. Proline Prosonic Flow 93P Endress+Hauser 17 Mechanical construction Design, dimensions Transmitter wall-mount housing A0001150 Dimensions (SI units) Dimensions (US units) A B C D E F G H J K 215 250 90.5 159.5 135 90 45 > 50 81 53 L M N O P Q R S T 1) 95 53 102 81.5 11.5 192 8 × M5 20 2 ×  6.5 1) Securing screw for wall mounting: M6 (screw head max. 10.5 mm) All dimensions in [mm] A B C D E F G H J K 8.46 9.84 3.56 6.27 5.31 3.54 1.77 > 1.97 3.18 2.08 L M N O P Q R S T 1) 3.74 2.08 4.01 3.20 0.45 7.55 8 × M5 0.79 2 ×  0.26 1) Securing screw for wall mounting: M6 (screw head max. 0.41") All dimensions in [inch] Esc E- + DC B A F E G KJ Q NL O R H J M P P SS T
  • 18. Proline Prosonic Flow 93P 18 Endress+Hauser There is a separate mounting kit for the wall-mounted housing. It can be ordered from Endress+Hauser as an accessory ( ä 27). The following installation variants are possible: • Panel-mounted installation • Pipe mounting Panel-mounted installation A0001131 Pipe mounting A0001132 245 (9.65) ~110 (~4.33) 210 (8.27) +0.5 (+0.019) –0.5 (–0.019) +0.5 (+0.019) –0.5 (–0.019) mm (inch) Ø 20…70 (Ø 0.79…2.75) ~ ~ 6.1)155 ( mm (inch)
  • 19. Proline Prosonic Flow 93P Endress+Hauser 19 Transmitter Field housing There is a separate mounting kit for the field housing. It can be ordered from Endress+Hauser as an accessory ( ä 27). The following installation variants are possible: • Wall mounting • Pipe mounting Wall mounting A0006999 Dimensions in SI units Dimensions in US units A A* B B* C D E F G H J K L M 265 242 240 217 206 186 178 8.6 (M8) 100 130 100 144 170 355 * Blind version (without local display) All dimensions in [mm] A A* B B* C D E F G H J K L M 10.4 9.53 9.45 8.54 8.11 7.32 7.01 8.6 (M8) 3.94 5.12 3.94 5.67 6.69 13.9 * Blind version (without local display) All dimensions in [inch] Nicht unter Spannungöffnen Keepcovertightwhile circuitsarealive Nepasouvrirl’appareil soustension Keepcovertightwhile circuitsarealive Nicht-eigensichere Stromkreise durch IP40-Abdeckung geschützt Non-intrinsically safe circuits Ip40 protected Boucles de courant sans sécurité intrinsèque protégées par Ip40 B* B A* A F G D C H J K L M E Esc E+–
  • 20. Proline Prosonic Flow 93P 20 Endress+Hauser Pipe mounting A0005157 a = Pipe mounting (separate mounting set, see accessories  ä 27) Nicht-eigensichere Stromkreise durch IP40-Abdeckung geschützt Non-intrinsically safe circuits Ip40 protected Boucles de courant sans sécurité intrinsèque protégées par Ip40 a 240 (9.45) mm (inch) Nicht unter Spannungöffnen Keepcovertightwhile circuitsarealive Nepasouvrirl’appareil soustension Keepcovertightwhile circuitsarealive Esc E+–
  • 21. Proline Prosonic Flow 93P Endress+Hauser 21 Prosonic Flow P sensor (DN 15 to 65 / ½ to 2½") A0011502 Dimensions in SI units Dimensions in US units A B C D E F 72 331 39 28 233 450 All dimensions in [mm] A B C D E F 2.83 13.03 1.54 1.10 9.17 17.72 All dimensions in [inch] B A C EF D
  • 22. Proline Prosonic Flow 93P 22 Endress+Hauser Prosonic Flow P sensor (DN 50 to 4000 / 2 to 160") A0001154 Mounting arrangement for measurement via two traverses Dimensions in SI units Dimensions in US units A C I G DB J E F H A B C D E F G H 56 62 145 111  58 max. 872 min. 0.5 439 to 790 I J Depends on the measuring point conditions (pipe, fluid, etc.). Dimension "I" can be determined: • Via transmitter programming Quick Setup (or FieldCare) • Online (Applicator) Pipe outer diameter All dimensions in [mm] A B C D E F G H 2.20 2.44 5.71 4.37  2.28 max. 34.3 min. 0.2 16.3 to 31.1 I J Depends on the measuring point conditions (pipe, fluid, etc.). Dimension "I" can be determined: • Via transmitter programming Quick Setup (or FieldCare) • Online (Applicator) Pipe outer diameter All dimensions in [inch]
  • 23. Proline Prosonic Flow 93P Endress+Hauser 23 A0001155 Mounting arrangement for measurement via one traverse Dimensions in SI units Dimensions in US units E DH A B G C F A B C D E F 56 62 145 111  58 max. 872 G H Depends on the measuring point conditions (pipe, fluid, etc.). Dimension "G" can be determined: • Via transmitter programming Quick Setup (or FieldCare) • Online (Applicator) Pipe outer diameter All dimensions in [mm] A B C D E F 2.20 2.44 5.71 4.37  2.28 max. 34.3 G H Depends on the measuring point conditions (pipe, fluid, etc.). Dimension "G" can be determined: • Via transmitter programming Quick Setup (or FieldCare) • Online (Applicator) Pipe outer diameter All dimensions in [inch]
  • 24. Proline Prosonic Flow 93P 24 Endress+Hauser Weight Transmitter • Wall-mount housing: 6.0 kg (13.2 lbs) • Field housing: 6.7 kg (14.8 lbs) Sensor Prosonic Flow P • DN 15 to 65 (½ to 2½") (incl. mounting material): 1.2 kg (2.65 lbs) • DN 50 to 4000 (2 to 160") (incl. mounting material): 2.8 kg (6.2 lbs) Sensor (accessories) • Prosonic Flow DDU18 (incl. mounting material): 2.4 kg (5.3 lbs) • Prosonic Flow DDU19 (incl. mounting material): 1.5 kg (3.3 lbs) ! Note! Weight information without packaging material. Materials Transmitter • Wall-mounted housing: powder-coated die-cast aluminum • Field housing: powder-coated die-cast aluminum Sensor Prosonic Flow P DN 15 to 65 (½ to 2½"); DN 50 to 4000 (2 to 160") • Sensor holder: stainless steel 1.4308/CF-8 • Sensor housing: stainless steel 1.4301/304 • Strapping bands/bracket: stainless steel 1.4301/304 • Sensor contact surfaces: chemically stable plastic Sensor (accessories) Prosonic Flow DDU18; Prosonic Flow DDU19 • Sensor holder: stainless steel 1.4308/CF-8 • Sensor housing: stainless steel 1.4301/304 • Strapping bands/bracket: stainless steel 1.4301/304 • Sensor contact surfaces: chemically stable plastic Connecting cable (sensor/transmitter) Prosonic Flow 93P (DN 15 to 65 / ½ to 2½") • TPE-V connecting cable – Cable sheath: TPE-V – Cable connector: stainless steel 1.40301 Prosonic Flow 93P (DN 50 to 4000 / 2 to 160") • PVC connecting cable – Cable sheath: PVC – Cable connector: nickeled brass 2.0401 • PTFE connecting cable – Cable sheath: PTFE – Cable connector: stainless steel 1.4301
  • 25. Proline Prosonic Flow 93P Endress+Hauser 25 Human interface Display elements • Liquid crystal display: illuminated, four lines each with 16 characters • Custom configuration for presenting different measured values and status variables • 3 totalizers Operating elements • Local operation with three optical keys • Application specific Quick Setup menus for straightforward commissioning Language group Language groups available for operation in different countries: • Western Europe and America (WEA): English, German, Spanish, Italian, French, Dutch and Portuguese • Eastern Europe/Scandinavia (EES): English, Russian, Polish, Norwegian, Finnish, Swedish and Czech • South and Eastern Asia (SEA): English, Japanese, Indonesian • China (CN): English, Chinese You can change the language group via the FieldCare operating program. Remote operation Operation via HART, PROFIBUS DP/PA, FOUNDATION Fieldbus Certificates and approvals CE mark The measuring system is in conformity with the statutory requirements of the EC Directives. Endress+Hauser confirms successful testing of the device by affixing to it the CE mark. C-Tick mark The measuring system is in conformity with the EMC requirements of the "Australian Communications and Media Authority" (ACMA). Ex approval Information on the currently available Ex-rated versions (ATEX, IECEx, FM, CSA, NEPSI, etc.) can be supplied by your Endress+Hauser Sales Center on request. All information relevant to explosion protection is available in separate documents that you can order as necessary. PROFIBUS DP/PA certification The flowmeter has successfully passed all the test procedures carried out and is certified and registered by the PNO (PROFIBUS/DP User Organization). The device thus meets all the requirements of the following specifications: • Certified to PROFIBUS DP/PA, profile version 3.0 (device certification number: on request) • The device can also be operated with certified devices of other manufacturers (interoperability) FOUNDATION Fieldbus certification The flowmeter has successfully passed all the test procedures carried out and is certified and registered by the Fieldbus Foundation. The device thus meets all the requirements of the following specifications: • Certified to FOUNDATION Fieldbus Specification • The device meets all the specifications of the FOUNDATION Fieldbus H1. • Interoperability Test Kit (ITK), revision status 5.01 (device certification number: on request) • The device can also be operated with certified devices of other manufacturers • Physical Layer Conformance Test of the Fieldbus Foundation
  • 26. Proline Prosonic Flow 93P 26 Endress+Hauser Other standards and guidelines • EN 60529 Degrees of protection provided by enclosures (IP code) • EN 61010-1 Safety requirements for electrical equipment for measurement, control and laboratory use • IEC/EN 61326 "Emission in accordance with Class A requirements". Electromagnetic compatibility (EMC requirements). • ANSI/ISA-S82.01 Safety Standard for Electrical and Electronic Test, Measuring, Controlling and Related Equipment - General Requirements. Pollution Degree 2, Installation Category II. • CAN/CSA-C22.2 No. 1010.1-92 Safety Requirements for Electrical Equipment for Measurement and Control and Laboratory Use. Pollution degree 2, Installation Category II • NAMUR NE 21 Electromagnetic compatibility (EMC) of industrial process and laboratory control equipment. • NAMUR NE 43 Standardization of the signal level for the breakdown information of digital transmitters with analog output signal. • NAMUR NE 53 Software of field devices and signal-processing devices with digital electronics. Ordering information The Endress+Hauser sales and service organization can provide detailed ordering information and information on the order codes on request.
  • 27. Proline Prosonic Flow 93P Endress+Hauser 27 Accessories Various accessories, which can be ordered separately from Endress+Hauser, are available for the transmitter and the sensor. The Endress+Hauser service organization can provide detailed information on the order codes on request. Device-specific accessories Measuring principle-specific accessories Accessory Description Order code Wall-mount housing, transmitter Prosonic Flow 93 Transmitter for replacement or for stock. Use the order code to define the following specifications: • Approvals • Degree of protection/version • Cable entry • Display / power supply / operation • Software • Outputs / inputs Single-channel version: 93XXX - XX1XX******** Two-channel version: 93XXX - XX2XX******** Conversion kit, inputs/outputs Conversion kit with appropriate plug-in point modules for converting the current input/output configuration to a new version. DK9UI - ** Sensor P (DN 15 to 65 / ½ to 2½") Clamp-on version DN 15 to 65 (½ to 2½") • –40 to +100 °C (–40 to 212 °F) • –40 to +150 °C (–40 to 302 °F) DK9PS - 1* DK9PS - 2* Sensor P (DN 50 to 4000 / 2 to 160") Clamp-on version DN 50 to 300 (2 to 12") • –40 to +80 °C (–40 to 176 °F) • 0 to +170 °C (+32 to 338 °F) DK9PS - B* DK9PS - F* DN 100 to 4000 (4 to 160") • –40 to +80 °C (–40 to 176 °F) • 0 to +170 °C (+32 to 338 °F) DK9PS - A* DK9PS - E* Sensor DDU18 Sensor for sound velocity measurement • –40 to +80 °C (–40 to 176 °F) • 0 to +170 °C (+32 to 338 °F) 50091703 50091704 Sensor DDU19 Sensor for wall thickness measurement. 50091713 Accessory Description Order code Mounting kit for aluminum field housing Mounting kit for wall-mount housing. Suitable for: • Wall mounting • Pipe mounting • Panel mounting DK9WM - A Mounting kit for field housing Mounting kit for aluminum field housing: Suitable for pipe mounting (¾ to 3") DK9WM - B Sensor holder set • Prosonic Flow P (DN 15 to 65 / ½ to 2½"): Sensor holder, clamp-on version DK9SH - 1 • Prosonic Flow P (DN 50 to 4000 / 2 to 160") – Sensor holder, fixed retaining nut, clamp-on version – Sensor holder, removable retaining nut, clamp-on version DK9SH - A DK9SH - B
  • 28. Proline Prosonic Flow 93P 28 Endress+Hauser Clamp-on installation set Sensor fastening for Prosonic Flow P (DN 15 to 65 / ½ to 2½") • U-shaped screw DN 15 to 32 (½ to 1¼") • Strapping bands DN 40 to 65 (1½ to 2½") DK9IC - 1* DK9IC - 2* Sensor fastening for Prosonic Flow P (DN 50 to 4000 / 2 to 160") • Without sensor fastening • Strapping bands DN 50 to 200 (2 to 8") • Strapping bands DN 200 to 600 (8 to 24") • Strapping bands DN 600 to 2000 (24 to 80") • Strapping bands DN 2000 to 4000 (80 to 160") DK9IC - A* DK9IC - B* DK9IC - C* DK9IC - D* DK9IC - E* • Without mounting tools • Spacing ruler DN 50 to 200 (2 to 8") • Spacing ruler DN 200 to 600 (8 to 24") • Mounting rail DN 50 to 200 (2 to 8") • Mounting rail DN 200 to 600 (8 to 24") • Fastening, 1 traverse DN 50 to 4000 (2 to 160") DK9IC - *1 DK9IC - *2 DK9IC - *3 DK9IC - *4 DK9IC - *5 DK9IC - *6 Conduit adapter for connecting cable Prosonic Flow P (DN 15 to 65 / ½ to 2½") • Conduit adapter incl. cable entry M20 × 1.5 • Conduit adapter incl. cable entry ½" NPT • Conduit adapter incl. cable entry G ½" DK9CB - BA1 DK9CB - BA2 DK9CB - BA3 Prosonic Flow P (DN 50 to 4000 / 2 to 160") • Conduit adapter incl. cable entry M20 × 1.5 • Conduit adapter incl. cable entry ½" NPT • Conduit adapter incl. cable entry G ½" DK9CB - BB1 DK9CB - BB2 DK9CB - BB3 Connecting cable Prosonic Flow P (DN 15 to 65 / ½ to 2½") 5 m sensor cable, TPE-V, –20 to +70 °C (–4 to +158 °F) 10 m sensor cable, TPE-V, –20 to +70 °C (–4 to +158 °F) 15 m sensor cable, TPE-V, –20 to +70 °C (–4 to +158 °F) 30 m sensor cable, TPE-V, –20 to +70 °C (–4 to +158 °F) DK9SS - BAA DK9SS - BAB DK9SS - BAC DK9SS - BAD Prosonic Flow P (DN 50 to 4000 / 2 to 160") 5 m sensor cable, PVC, –20 to +70 °C (–4 to +158 °F) 10 m sensor cable, PVC, –20 to +70 °C (–4 to +158 °F) 15 m sensor cable, PVC, –20 to +70 °C (–4 to +158 °F) 30 m sensor cable, PVC, –20 to +70 °C (–4 to +158 °F) DK9SS - BBA DK9SS - BBB DK9SS - BBC DK9SS - BBD 5 m sensor cable, PTFE, –40 to +170 °C (–40 to +158 °F) 10 m sensor cable, PTFE, –40 to +170 °C (–40 to +158 °F) 15 m sensor cable, PTFE, –40 to +170 °C (–40 to +158 °F) 30 m sensor cable, PTFE, –40 to +170 °C (–40 to +158 °F) DK9SS - BBE DK9SS - BBF DK9SS - BBG DK9SS - BBH Acoustic coupling fluid • Coupling fluid 0 to 170 °C (+32 to 338 °F), high temperature, standard • Adhesive coupling fluid –40 to +80 °C (–40 to +176 °F) • Water-soluble coupling fluid –20 to +80 °C (–4 to +176 °F) • Coupling fluid DDU 19, –20 to +60 °C (–4 to +140 °F) • Coupling fluid –40 to +80 °C (–40 to +176 °F), standard, type MBG2000 DK9CM - 2 DK9CM - 3 DK9CM - 4 DK9CM - 6 DK9CM - 7 Accessory Description Order code
  • 29. Proline Prosonic Flow 93P Endress+Hauser 29 Communication-specific accessories Accessory Description Order code HART handheld terminal DXR375 Handheld terminal for remote configuration and for obtaining measured values via the HART current output (4 to 20 mA) and FOUNDATION Fieldbus. Contact your Endress+Hauser representative for more information. DXR375 - ******* Fieldgate FXA320 Gateway for remote interrogation of HART sensors and actuators via Web browser: • 2-channel analog input (4 to 20 mA) • 4 binary inputs with event counter function and frequency measurement • Communication via modem, Ethernet or GSM • Visualization via Internet/Intranet in the Web browser and/ or WAP cellular phone • Limit value monitoring with alarm signaling via e-mail or SMS • Synchronized time stamping of all measured values FXA320 - ***** Fieldgate FXA520 Gateway for remote interrogation of HART sensors and actuators via Web browser: • Web server for remote monitoring of up to 30 measuring points • Intrinsically safe version [EEx ia]IIC for applications in hazardous areas • Communication via modem, Ethernet or GSM • Visualization via Internet/Intranet in the Web browser and/ or WAP cellular phone • Limit value monitoring with alarm signaling via e-mail or SMS • Synchronized time stamping of all measured values • Remote diagnosis and remote configuration of connected HART devices FXA520 - ****
  • 30. Proline Prosonic Flow 93P 30 Endress+Hauser Service-specific accessories Accessory Description Order code Applicator Software for selecting and planning flowmeters. The Applicator can be downloaded from the Internet or ordered on CD-ROM for installation on a local PC. Contact your Endress+Hauser representative for more information. DXA80 - * Fieldcheck Tester/simulator for testing flowmeters in the field. When used in conjunction with the "FieldCare" software package, test results can be imported into a database, printed out and used for official certification. Contact your Endress+Hauser representative for more information. 50098801 FieldCare FieldCare is Endress+Hauser's FDT-based plant asset management tool. It can configure all intelligent field units in your system and helps you manage them. By using the status information, it is also a simple but effective way of checking their status and condition. See the product page on the Endress+Hauser website: www.endress.com FXA193 Service interface from the measuring device to the PC for operation via FieldCare. FXA193 – * Communication cable Communication cable for connecting the Prosonic Flow 93P transmitter to the FXA193 service interface. DK9ZT – A
  • 31. Proline Prosonic Flow 93P Endress+Hauser 31 Documentation • Flow measurement (FA005D) • Operating Instructions for Prosonic Flow 93 (BA070D and BA071D) • Operating Instructions for Prosonic Flow 93 PROFIBUS DP/PA (BA076D and BA077D) • Operating Instructions for Prosonic Flow 93 FOUNDATION Fieldbus (BA078D and BA079D) • Supplementary documentation on Ex-ratings: ATEX, FM, CSA, IECEx, NEPSI Registered trademarks HART® Registered trademark of HART Communication Foundation, Austin, USA PROFIBUS® Registered trademark of the PROFIBUS User Organization, Karlsruhe, Germany FOUNDATION™ Fieldbus Registered trademark of the Fieldbus FOUNDATION, Austin, USA HistoROM™, T-DAT™, F-CHIP®, FieldCare®, Fieldcheck® Registered or registration-pending trademarks of Endress+Hauser Flowtec AG, Reinach, CH
  • 32. Instruments International Endress+Hauser Instruments International AG Kaegenstrasse 2 4153 Reinach Switzerland Tel.+41 61 715 81 00 Fax+41 61 715 25 00 www.endress.com info@ii.endress.com TI00083D/06/EN/13.11 71136721 FM+SGML6.0 ProMoDo