SlideShare a Scribd company logo
1 of 20
SIZE-EXCLUSION
CHROMATOGRAPHY OF
PERFLUOROSULFONATE
D IONOMERS
Nan Zhao
George Geevarghese
PART – I : INTRO
Nan Zhao
 INTRODUCTIO
N
 The authors want to measure the molar mass
distributions , Polymer conformation and
unperturbed dimensions of the
Polyfluorosulfonated ionomers which are
copolymers of tetrafluorethylene and sulfonic
acid-functionalized (SO3H) perfluorinated vinyl
ethers with SEC(with LS, DRI, DV)
SEC
LS
DRI, DV
 EXPERIMENTAL
 Testing Equipments:
SEC: Agilent 1100 seriesisocratic pump, autosampler and two-wavelength
spectrophotometric detector.
LS: Agilent PD2020 two-angle LS detector
DV: Malvern Model 270 DV detector
DRI: Waters Corporation Model 410 DRI detector
3 Columns: Agilent Olexis 7.5 mm×300 mm columns at 35.0 ◦C were used
with N,N-dimethylformamide (DMF) containing 0.1 M LiNO3.
 Materials: the authors used 3 classes of the Perfluorosulfonated
ionomers.
 The eluent was pre-filtered using 0.22 m Millipore GS filters.
 The flow rate was 1.0 mL/min
 0.2% acetone added to the samples as a flow marker
 The columns were calibrated with 15 PMMA narrow standards from
Agilent with molar masses between 580 and 1,400,000.
 EXPERIMENTAL
 Sample dissolution: He prepares a dilute ionomer
dispersion by using 80/20 n-propanol/water. The
dilute dispersion is then taken in the acid digestion
bomb.
 Then place the bomb in an oven heated from 30 to
230 ◦C at 5 ◦C per min, and then held for 6 h. After
that several hours were required to cool the vessels
down to room temperature.
 Then the autoclaved solutions were diluted 1:1 with
SEC sample solvent to give the SEC sample injection
concentration of ∼0.5 mg/mL.
 RESULTS AND DISCUSSION
 Five strategies the authors used
 SEC optimization
 Ionomer disolution
 increasing SEC detector signals
 (Calculation of molar mass distributions)
 (Polymer conformation and unperturbed dimensions)
 SEC OPTIMIZATION
 Because the sample has two solubility parameters which δ1=9.68
(cal/cm3)1/2 and δ2=16.71 (cal/cm3)1/2, the authors need to find a proper
solvent.
 Previously, Dr. Lousenberg and Hommura et al. used δ values of the
solvent closer to δ1 and closer to δ2 respectively.
 Due to some “unknows reason”, he choose the DMF(δ= 12.1),with as
the solvent.
 Then again, none of the perfluorosulfonatedionomers eluted from the
SEC columns using DMF(pure) he added 0.1M LiNO3 to improve the
elution.
 And because the non-autoclaved samples showed less linearity in the
log M calibration curve, he used autoclaved Nafion.
 SEC OPTIMIZATION
PART II –
4 STRATEGIES + CONCLUSION
George Geevarghese
 Ionomer dissolution
• Martin et al. - autoclave idea!
– 50/50 (v/v) n-propanol/water at 250 deg for 1hr (1wt%
incomer concentration)
– Author repeats (at 260 deg for 1hr & same conc., but fails
– But it worked (almost) at 230 deg for 6hrs & same conc., but
phase separated.
• Author finally used 0.1 M LiNO3 with DMF and added to 80/20
n-propanol/water
– That worked. BUT only for Nafion. C2 and C4 still showed
prepeaks.
• Then again, he realized at low wt % of ionomer (~0.2wt%)
prepeaks were less significant
• BUT, LS observed prepeaks at even small wt %. But, DRI
doesnt (even at higher conc)
• So he found an optimized conc @ 0.1wt%
 Increasing SEC detector signals
• The “optimized” “autoclaved” solution (0.1 wt%) was diluted 1:1
(w/w) with SEC samples
– But, results in large solvent peaks and detector signal
disruptions in the system peak region.
– This also reduced the injection concentrations of polymer ( to
0.5 mg/mL)
• DRI and LS are weak at this conc. because of small specific
refractive index increments (between −0.02 and −0.04
mL/g)
• Viscometry detector signals are weak because of the low
intrinsic viscosities
• So, he increased sample conc. by Solvent Exchange
 to ∼2 mg/mL
 SEC of this sample showed no aggregation(cos no prepeaks in
DRI)
• Can also improve DV signal by inc. the flow rate.
• Changing the solvent to NMP and DMSO will increase the specific
refractive index increment and improve DRI and LS detector signals
accordingly (complicated! high temp!)
1 g of SEC sample solvent is added to 3 g of 0.1 wt % 80/20 n-propanol/water
autoclaved solution. The mixed solvent is then partially evaporated with a nitrogen
stream to approximately 1.5 g. The lower boiling n-propanol and water are
preferentially removed while the sample concentration is increased.
Calculation of molar mass distributions
• Variation in dn/dc across the molecular size distribution caused by
comonomer compositional drift.
• Corrected by:
• Accounting for minor variation in dn/dc across the chromatogram (OR)
• Mark–Houwink calculation method
 Polymer conformation &
unperturbed dimensions
• Conformation plots of three ionomer classes are plotted. All
PFSA ionomers were observed to have lower viscosity at
equivalent molar mass than PMMA
• The unperturbed dimensions was calculated from SEC
viscosity–molar mass data
• It suggested suspicious free rotation of the perfluoroalkane
units (can be attributed to the unusual structure and behavior
of these materials)
Our Conclusions
• Advantages:
– SEC system with triple detection can
• give high accuracy for perfluorosulfonated ionomers.
• Most info can be obtained
– Solvent Exchange process can
• improve all detector signals
• Also reduce the water and n-propanol peaks observed
in the system peak region.
– The autoclaving procedure
• provides molecular solutions without re-aggregation
(better peaks)
Our Conclusions
• Disadvantages:
– Dissolution temp is high compared to other
techniques
– Solvent exchange is somewhat tedious and
requires careful weighing of solutions before and
after evaporation to obtain accurate injected
sample concentrations
– Even though small, Variation in dn/dc causes
error in molar mass and viscosity from its original
value
Reference
• Size-exclusion chromatography of perfluorosulfonated ionomers by T.H.
Moureya, L.A. Slater a, R.C. Galipoa, R.J. Koestnerb :
Journal of Chromatography A, 1218 (2011) 5801– 5809

More Related Content

What's hot

Ion Exchange Chromatography ppt
Ion Exchange Chromatography pptIon Exchange Chromatography ppt
Ion Exchange Chromatography pptAlexa Jacob
 
Lcms gcms and its applications
Lcms gcms and its applicationsLcms gcms and its applications
Lcms gcms and its applicationsNihal Calicut
 
Ion pair chromatography final
Ion pair chromatography finalIon pair chromatography final
Ion pair chromatography finalsnehal dhobale
 
High performance liquid chromatography
High performance liquid chromatographyHigh performance liquid chromatography
High performance liquid chromatographyMehwish Nawaz
 
Ion pair , reversed pair liquid chromatography
Ion   pair , reversed pair liquid chromatographyIon   pair , reversed pair liquid chromatography
Ion pair , reversed pair liquid chromatographyjain university
 
Ion Exchange Chromatography, ppt
Ion Exchange Chromatography, pptIon Exchange Chromatography, ppt
Ion Exchange Chromatography, pptshaisejacob
 
Affinity chromatography
Affinity chromatographyAffinity chromatography
Affinity chromatographySagar Savale
 
Lecture 04 size exclusion chromatography
Lecture 04 size exclusion chromatographyLecture 04 size exclusion chromatography
Lecture 04 size exclusion chromatographyDeepak Sharma
 
Chromatographic Methods of Analysis ( Gel Chromatography Method )
Chromatographic Methods of Analysis ( Gel Chromatography Method )Chromatographic Methods of Analysis ( Gel Chromatography Method )
Chromatographic Methods of Analysis ( Gel Chromatography Method )FLI
 
Liquid liquid extraction and flocculation
Liquid liquid extraction and flocculationLiquid liquid extraction and flocculation
Liquid liquid extraction and flocculationvikash_94
 
Ion pair chromatography for pharmacy students
Ion pair chromatography for pharmacy studentsIon pair chromatography for pharmacy students
Ion pair chromatography for pharmacy studentsabhishek rai
 
High performance liquid chromatography
High performance liquid chromatographyHigh performance liquid chromatography
High performance liquid chromatographyDr Alok Bharti
 
Dart ion source- mass spectrometry
Dart ion source- mass spectrometryDart ion source- mass spectrometry
Dart ion source- mass spectrometrykeerthana151
 
Gel permeation chromatography
Gel permeation chromatographyGel permeation chromatography
Gel permeation chromatographyShradha Basu
 

What's hot (20)

Ion Exchange Chromatography ppt
Ion Exchange Chromatography pptIon Exchange Chromatography ppt
Ion Exchange Chromatography ppt
 
Lcms gcms and its applications
Lcms gcms and its applicationsLcms gcms and its applications
Lcms gcms and its applications
 
Ion pair chromatography final
Ion pair chromatography finalIon pair chromatography final
Ion pair chromatography final
 
Column chromatography
Column chromatographyColumn chromatography
Column chromatography
 
High performance liquid chromatography
High performance liquid chromatographyHigh performance liquid chromatography
High performance liquid chromatography
 
Ion pair , reversed pair liquid chromatography
Ion   pair , reversed pair liquid chromatographyIon   pair , reversed pair liquid chromatography
Ion pair , reversed pair liquid chromatography
 
Mass spectroscopy
Mass spectroscopyMass spectroscopy
Mass spectroscopy
 
Ion Exchange Chromatography, ppt
Ion Exchange Chromatography, pptIon Exchange Chromatography, ppt
Ion Exchange Chromatography, ppt
 
Affinity chromatography
Affinity chromatographyAffinity chromatography
Affinity chromatography
 
Ion exchange chromatography
Ion exchange chromatographyIon exchange chromatography
Ion exchange chromatography
 
Lecture 04 size exclusion chromatography
Lecture 04 size exclusion chromatographyLecture 04 size exclusion chromatography
Lecture 04 size exclusion chromatography
 
Chromatographic Methods of Analysis ( Gel Chromatography Method )
Chromatographic Methods of Analysis ( Gel Chromatography Method )Chromatographic Methods of Analysis ( Gel Chromatography Method )
Chromatographic Methods of Analysis ( Gel Chromatography Method )
 
Hplc
HplcHplc
Hplc
 
Liquid liquid extraction and flocculation
Liquid liquid extraction and flocculationLiquid liquid extraction and flocculation
Liquid liquid extraction and flocculation
 
Ion pair chromatography for pharmacy students
Ion pair chromatography for pharmacy studentsIon pair chromatography for pharmacy students
Ion pair chromatography for pharmacy students
 
High performance liquid chromatography
High performance liquid chromatographyHigh performance liquid chromatography
High performance liquid chromatography
 
Dart ion source- mass spectrometry
Dart ion source- mass spectrometryDart ion source- mass spectrometry
Dart ion source- mass spectrometry
 
Capillary electrophoresis
Capillary electrophoresisCapillary electrophoresis
Capillary electrophoresis
 
Isoelectric focusing
Isoelectric focusingIsoelectric focusing
Isoelectric focusing
 
Gel permeation chromatography
Gel permeation chromatographyGel permeation chromatography
Gel permeation chromatography
 

Similar to Size-exclusion chromatography of perfluorosulfonated ionomers

Trace metal analysis in sediments using aas and asv techniques
Trace metal analysis in sediments using aas and asv techniquesTrace metal analysis in sediments using aas and asv techniques
Trace metal analysis in sediments using aas and asv techniquesAwad Albalwi
 
HPLC (HIGH PERFOMANCE LIQUID CHROMATOGRAPHY)
HPLC (HIGH PERFOMANCE LIQUID CHROMATOGRAPHY)HPLC (HIGH PERFOMANCE LIQUID CHROMATOGRAPHY)
HPLC (HIGH PERFOMANCE LIQUID CHROMATOGRAPHY)RAJA K
 
Atomic absorption spectroscopy- AAS
Atomic absorption spectroscopy- AASAtomic absorption spectroscopy- AAS
Atomic absorption spectroscopy- AASJustin K George
 
Ultra Performance Liquid Chromatography
Ultra Performance Liquid ChromatographyUltra Performance Liquid Chromatography
Ultra Performance Liquid ChromatographyMahewash Sana Pathan
 
High performance liquid chromatography
High performance liquid chromatographyHigh performance liquid chromatography
High performance liquid chromatographyAshfaq Ahmad
 
Monitoring Cerium Dioxide and Zinc Oxide Nanoparticles through Drinking Water...
Monitoring Cerium Dioxide and Zinc Oxide Nanoparticles through Drinking Water...Monitoring Cerium Dioxide and Zinc Oxide Nanoparticles through Drinking Water...
Monitoring Cerium Dioxide and Zinc Oxide Nanoparticles through Drinking Water...PerkinElmer, Inc.
 
High Performance Liquid Chromatography- Dr. A. Amsavel
High Performance Liquid Chromatography- Dr. A. AmsavelHigh Performance Liquid Chromatography- Dr. A. Amsavel
High Performance Liquid Chromatography- Dr. A. AmsavelDr. Amsavel A
 
phyto chemical finger printing in phytochemistry
phyto chemical finger printing in phytochemistryphyto chemical finger printing in phytochemistry
phyto chemical finger printing in phytochemistryssuser22bccd
 
Detection Method for Low Level of Potent Toxins
Detection Method for Low Level of Potent ToxinsDetection Method for Low Level of Potent Toxins
Detection Method for Low Level of Potent ToxinsShreyas Patel
 
CE8512- WATER & WASTE WATER ANALYSIS LAB MANUVAL
CE8512- WATER & WASTE WATER ANALYSIS LAB MANUVALCE8512- WATER & WASTE WATER ANALYSIS LAB MANUVAL
CE8512- WATER & WASTE WATER ANALYSIS LAB MANUVALLokesh Kalliz
 
do-and-bod-analysis.ppt
do-and-bod-analysis.pptdo-and-bod-analysis.ppt
do-and-bod-analysis.pptRodrigo664210
 
wrd-ot-do-and-bod-analysis_445265_7.ppt
wrd-ot-do-and-bod-analysis_445265_7.pptwrd-ot-do-and-bod-analysis_445265_7.ppt
wrd-ot-do-and-bod-analysis_445265_7.pptMyrnaFiel
 
do-and-bod-analysis.ppt
do-and-bod-analysis.pptdo-and-bod-analysis.ppt
do-and-bod-analysis.pptshubham871246
 
Uv vis spectroscopy practical.
Uv vis spectroscopy practical.Uv vis spectroscopy practical.
Uv vis spectroscopy practical.Salum Mkata
 

Similar to Size-exclusion chromatography of perfluorosulfonated ionomers (20)

Trace metal analysis in sediments using aas and asv techniques
Trace metal analysis in sediments using aas and asv techniquesTrace metal analysis in sediments using aas and asv techniques
Trace metal analysis in sediments using aas and asv techniques
 
Analytical Chem Presentation
Analytical Chem PresentationAnalytical Chem Presentation
Analytical Chem Presentation
 
HPLC (HIGH PERFOMANCE LIQUID CHROMATOGRAPHY)
HPLC (HIGH PERFOMANCE LIQUID CHROMATOGRAPHY)HPLC (HIGH PERFOMANCE LIQUID CHROMATOGRAPHY)
HPLC (HIGH PERFOMANCE LIQUID CHROMATOGRAPHY)
 
Atomic absorption spectroscopy- AAS
Atomic absorption spectroscopy- AASAtomic absorption spectroscopy- AAS
Atomic absorption spectroscopy- AAS
 
Ultra Performance Liquid Chromatography
Ultra Performance Liquid ChromatographyUltra Performance Liquid Chromatography
Ultra Performance Liquid Chromatography
 
High performance liquid chromatography
High performance liquid chromatographyHigh performance liquid chromatography
High performance liquid chromatography
 
Monitoring Cerium Dioxide and Zinc Oxide Nanoparticles through Drinking Water...
Monitoring Cerium Dioxide and Zinc Oxide Nanoparticles through Drinking Water...Monitoring Cerium Dioxide and Zinc Oxide Nanoparticles through Drinking Water...
Monitoring Cerium Dioxide and Zinc Oxide Nanoparticles through Drinking Water...
 
High Performance Liquid Chromatography- Dr. A. Amsavel
High Performance Liquid Chromatography- Dr. A. AmsavelHigh Performance Liquid Chromatography- Dr. A. Amsavel
High Performance Liquid Chromatography- Dr. A. Amsavel
 
phyto chemical finger printing in phytochemistry
phyto chemical finger printing in phytochemistryphyto chemical finger printing in phytochemistry
phyto chemical finger printing in phytochemistry
 
Detection Method for Low Level of Potent Toxins
Detection Method for Low Level of Potent ToxinsDetection Method for Low Level of Potent Toxins
Detection Method for Low Level of Potent Toxins
 
CE8512- WATER & WASTE WATER ANALYSIS LAB MANUVAL
CE8512- WATER & WASTE WATER ANALYSIS LAB MANUVALCE8512- WATER & WASTE WATER ANALYSIS LAB MANUVAL
CE8512- WATER & WASTE WATER ANALYSIS LAB MANUVAL
 
Rahman,1983
Rahman,1983Rahman,1983
Rahman,1983
 
do-and-bod-analysis.ppt
do-and-bod-analysis.pptdo-and-bod-analysis.ppt
do-and-bod-analysis.ppt
 
wrd-ot-do-and-bod-analysis_445265_7.ppt
wrd-ot-do-and-bod-analysis_445265_7.pptwrd-ot-do-and-bod-analysis_445265_7.ppt
wrd-ot-do-and-bod-analysis_445265_7.ppt
 
do-and-bod-analysis.ppt
do-and-bod-analysis.pptdo-and-bod-analysis.ppt
do-and-bod-analysis.ppt
 
Expanding Your High Performance Liquid Chromatography and Ultra High Performa...
Expanding Your High Performance Liquid Chromatography and Ultra High Performa...Expanding Your High Performance Liquid Chromatography and Ultra High Performa...
Expanding Your High Performance Liquid Chromatography and Ultra High Performa...
 
Expanding Your High Performance Liquid Chromatography and Ultra High Performa...
Expanding Your High Performance Liquid Chromatography and Ultra High Performa...Expanding Your High Performance Liquid Chromatography and Ultra High Performa...
Expanding Your High Performance Liquid Chromatography and Ultra High Performa...
 
Uv vis spectroscopy practical.
Uv vis spectroscopy practical.Uv vis spectroscopy practical.
Uv vis spectroscopy practical.
 
Adobe Scan 01-Sep-2022.pdf
Adobe Scan 01-Sep-2022.pdfAdobe Scan 01-Sep-2022.pdf
Adobe Scan 01-Sep-2022.pdf
 
UPLC
UPLCUPLC
UPLC
 

More from Geevarghese George

Natural Rubber - Sources, Coagulation & Processing of Coagulate, Structure & ...
Natural Rubber - Sources, Coagulation & Processing of Coagulate, Structure & ...Natural Rubber - Sources, Coagulation & Processing of Coagulate, Structure & ...
Natural Rubber - Sources, Coagulation & Processing of Coagulate, Structure & ...Geevarghese George
 
Embrittlement of Polymers in Space
Embrittlement of Polymers in SpaceEmbrittlement of Polymers in Space
Embrittlement of Polymers in SpaceGeevarghese George
 
"Mathematics in day to day life"
"Mathematics in day to day life""Mathematics in day to day life"
"Mathematics in day to day life"Geevarghese George
 

More from Geevarghese George (6)

Epoxy mechanical properties
Epoxy mechanical propertiesEpoxy mechanical properties
Epoxy mechanical properties
 
Market Research
Market ResearchMarket Research
Market Research
 
Natural Rubber - Sources, Coagulation & Processing of Coagulate, Structure & ...
Natural Rubber - Sources, Coagulation & Processing of Coagulate, Structure & ...Natural Rubber - Sources, Coagulation & Processing of Coagulate, Structure & ...
Natural Rubber - Sources, Coagulation & Processing of Coagulate, Structure & ...
 
Embrittlement of Polymers in Space
Embrittlement of Polymers in SpaceEmbrittlement of Polymers in Space
Embrittlement of Polymers in Space
 
"Mathematics in day to day life"
"Mathematics in day to day life""Mathematics in day to day life"
"Mathematics in day to day life"
 
Electromechanical Relays
Electromechanical RelaysElectromechanical Relays
Electromechanical Relays
 

Size-exclusion chromatography of perfluorosulfonated ionomers

  • 2. PART – I : INTRO Nan Zhao
  • 3.  INTRODUCTIO N  The authors want to measure the molar mass distributions , Polymer conformation and unperturbed dimensions of the Polyfluorosulfonated ionomers which are copolymers of tetrafluorethylene and sulfonic acid-functionalized (SO3H) perfluorinated vinyl ethers with SEC(with LS, DRI, DV) SEC LS DRI, DV
  • 4.  EXPERIMENTAL  Testing Equipments: SEC: Agilent 1100 seriesisocratic pump, autosampler and two-wavelength spectrophotometric detector. LS: Agilent PD2020 two-angle LS detector DV: Malvern Model 270 DV detector DRI: Waters Corporation Model 410 DRI detector 3 Columns: Agilent Olexis 7.5 mm×300 mm columns at 35.0 ◦C were used with N,N-dimethylformamide (DMF) containing 0.1 M LiNO3.  Materials: the authors used 3 classes of the Perfluorosulfonated ionomers.  The eluent was pre-filtered using 0.22 m Millipore GS filters.  The flow rate was 1.0 mL/min  0.2% acetone added to the samples as a flow marker  The columns were calibrated with 15 PMMA narrow standards from Agilent with molar masses between 580 and 1,400,000.
  • 5.  EXPERIMENTAL  Sample dissolution: He prepares a dilute ionomer dispersion by using 80/20 n-propanol/water. The dilute dispersion is then taken in the acid digestion bomb.  Then place the bomb in an oven heated from 30 to 230 ◦C at 5 ◦C per min, and then held for 6 h. After that several hours were required to cool the vessels down to room temperature.  Then the autoclaved solutions were diluted 1:1 with SEC sample solvent to give the SEC sample injection concentration of ∼0.5 mg/mL.
  • 6.  RESULTS AND DISCUSSION  Five strategies the authors used  SEC optimization  Ionomer disolution  increasing SEC detector signals  (Calculation of molar mass distributions)  (Polymer conformation and unperturbed dimensions)
  • 7.  SEC OPTIMIZATION  Because the sample has two solubility parameters which δ1=9.68 (cal/cm3)1/2 and δ2=16.71 (cal/cm3)1/2, the authors need to find a proper solvent.  Previously, Dr. Lousenberg and Hommura et al. used δ values of the solvent closer to δ1 and closer to δ2 respectively.  Due to some “unknows reason”, he choose the DMF(δ= 12.1),with as the solvent.  Then again, none of the perfluorosulfonatedionomers eluted from the SEC columns using DMF(pure) he added 0.1M LiNO3 to improve the elution.
  • 8.  And because the non-autoclaved samples showed less linearity in the log M calibration curve, he used autoclaved Nafion.  SEC OPTIMIZATION
  • 9. PART II – 4 STRATEGIES + CONCLUSION George Geevarghese
  • 10.  Ionomer dissolution • Martin et al. - autoclave idea! – 50/50 (v/v) n-propanol/water at 250 deg for 1hr (1wt% incomer concentration) – Author repeats (at 260 deg for 1hr & same conc., but fails – But it worked (almost) at 230 deg for 6hrs & same conc., but phase separated. • Author finally used 0.1 M LiNO3 with DMF and added to 80/20 n-propanol/water – That worked. BUT only for Nafion. C2 and C4 still showed prepeaks.
  • 11. • Then again, he realized at low wt % of ionomer (~0.2wt%) prepeaks were less significant • BUT, LS observed prepeaks at even small wt %. But, DRI doesnt (even at higher conc) • So he found an optimized conc @ 0.1wt%
  • 12.  Increasing SEC detector signals • The “optimized” “autoclaved” solution (0.1 wt%) was diluted 1:1 (w/w) with SEC samples – But, results in large solvent peaks and detector signal disruptions in the system peak region. – This also reduced the injection concentrations of polymer ( to 0.5 mg/mL) • DRI and LS are weak at this conc. because of small specific refractive index increments (between −0.02 and −0.04 mL/g) • Viscometry detector signals are weak because of the low intrinsic viscosities
  • 13. • So, he increased sample conc. by Solvent Exchange  to ∼2 mg/mL  SEC of this sample showed no aggregation(cos no prepeaks in DRI) • Can also improve DV signal by inc. the flow rate. • Changing the solvent to NMP and DMSO will increase the specific refractive index increment and improve DRI and LS detector signals accordingly (complicated! high temp!) 1 g of SEC sample solvent is added to 3 g of 0.1 wt % 80/20 n-propanol/water autoclaved solution. The mixed solvent is then partially evaporated with a nitrogen stream to approximately 1.5 g. The lower boiling n-propanol and water are preferentially removed while the sample concentration is increased.
  • 14. Calculation of molar mass distributions
  • 15. • Variation in dn/dc across the molecular size distribution caused by comonomer compositional drift. • Corrected by: • Accounting for minor variation in dn/dc across the chromatogram (OR) • Mark–Houwink calculation method
  • 16.  Polymer conformation & unperturbed dimensions • Conformation plots of three ionomer classes are plotted. All PFSA ionomers were observed to have lower viscosity at equivalent molar mass than PMMA
  • 17. • The unperturbed dimensions was calculated from SEC viscosity–molar mass data • It suggested suspicious free rotation of the perfluoroalkane units (can be attributed to the unusual structure and behavior of these materials)
  • 18. Our Conclusions • Advantages: – SEC system with triple detection can • give high accuracy for perfluorosulfonated ionomers. • Most info can be obtained – Solvent Exchange process can • improve all detector signals • Also reduce the water and n-propanol peaks observed in the system peak region. – The autoclaving procedure • provides molecular solutions without re-aggregation (better peaks)
  • 19. Our Conclusions • Disadvantages: – Dissolution temp is high compared to other techniques – Solvent exchange is somewhat tedious and requires careful weighing of solutions before and after evaporation to obtain accurate injected sample concentrations – Even though small, Variation in dn/dc causes error in molar mass and viscosity from its original value
  • 20. Reference • Size-exclusion chromatography of perfluorosulfonated ionomers by T.H. Moureya, L.A. Slater a, R.C. Galipoa, R.J. Koestnerb : Journal of Chromatography A, 1218 (2011) 5801– 5809