3. INTRODUCTIO
N
The authors want to measure the molar mass
distributions , Polymer conformation and
unperturbed dimensions of the
Polyfluorosulfonated ionomers which are
copolymers of tetrafluorethylene and sulfonic
acid-functionalized (SO3H) perfluorinated vinyl
ethers with SEC(with LS, DRI, DV)
SEC
LS
DRI, DV
4. EXPERIMENTAL
Testing Equipments:
SEC: Agilent 1100 seriesisocratic pump, autosampler and two-wavelength
spectrophotometric detector.
LS: Agilent PD2020 two-angle LS detector
DV: Malvern Model 270 DV detector
DRI: Waters Corporation Model 410 DRI detector
3 Columns: Agilent Olexis 7.5 mm×300 mm columns at 35.0 ◦C were used
with N,N-dimethylformamide (DMF) containing 0.1 M LiNO3.
Materials: the authors used 3 classes of the Perfluorosulfonated
ionomers.
The eluent was pre-filtered using 0.22 m Millipore GS filters.
The flow rate was 1.0 mL/min
0.2% acetone added to the samples as a flow marker
The columns were calibrated with 15 PMMA narrow standards from
Agilent with molar masses between 580 and 1,400,000.
5. EXPERIMENTAL
Sample dissolution: He prepares a dilute ionomer
dispersion by using 80/20 n-propanol/water. The
dilute dispersion is then taken in the acid digestion
bomb.
Then place the bomb in an oven heated from 30 to
230 ◦C at 5 ◦C per min, and then held for 6 h. After
that several hours were required to cool the vessels
down to room temperature.
Then the autoclaved solutions were diluted 1:1 with
SEC sample solvent to give the SEC sample injection
concentration of ∼0.5 mg/mL.
6. RESULTS AND DISCUSSION
Five strategies the authors used
SEC optimization
Ionomer disolution
increasing SEC detector signals
(Calculation of molar mass distributions)
(Polymer conformation and unperturbed dimensions)
7. SEC OPTIMIZATION
Because the sample has two solubility parameters which δ1=9.68
(cal/cm3)1/2 and δ2=16.71 (cal/cm3)1/2, the authors need to find a proper
solvent.
Previously, Dr. Lousenberg and Hommura et al. used δ values of the
solvent closer to δ1 and closer to δ2 respectively.
Due to some “unknows reason”, he choose the DMF(δ= 12.1),with as
the solvent.
Then again, none of the perfluorosulfonatedionomers eluted from the
SEC columns using DMF(pure) he added 0.1M LiNO3 to improve the
elution.
8. And because the non-autoclaved samples showed less linearity in the
log M calibration curve, he used autoclaved Nafion.
SEC OPTIMIZATION
9. PART II –
4 STRATEGIES + CONCLUSION
George Geevarghese
10. Ionomer dissolution
• Martin et al. - autoclave idea!
– 50/50 (v/v) n-propanol/water at 250 deg for 1hr (1wt%
incomer concentration)
– Author repeats (at 260 deg for 1hr & same conc., but fails
– But it worked (almost) at 230 deg for 6hrs & same conc., but
phase separated.
• Author finally used 0.1 M LiNO3 with DMF and added to 80/20
n-propanol/water
– That worked. BUT only for Nafion. C2 and C4 still showed
prepeaks.
11. • Then again, he realized at low wt % of ionomer (~0.2wt%)
prepeaks were less significant
• BUT, LS observed prepeaks at even small wt %. But, DRI
doesnt (even at higher conc)
• So he found an optimized conc @ 0.1wt%
12. Increasing SEC detector signals
• The “optimized” “autoclaved” solution (0.1 wt%) was diluted 1:1
(w/w) with SEC samples
– But, results in large solvent peaks and detector signal
disruptions in the system peak region.
– This also reduced the injection concentrations of polymer ( to
0.5 mg/mL)
• DRI and LS are weak at this conc. because of small specific
refractive index increments (between −0.02 and −0.04
mL/g)
• Viscometry detector signals are weak because of the low
intrinsic viscosities
13. • So, he increased sample conc. by Solvent Exchange
to ∼2 mg/mL
SEC of this sample showed no aggregation(cos no prepeaks in
DRI)
• Can also improve DV signal by inc. the flow rate.
• Changing the solvent to NMP and DMSO will increase the specific
refractive index increment and improve DRI and LS detector signals
accordingly (complicated! high temp!)
1 g of SEC sample solvent is added to 3 g of 0.1 wt % 80/20 n-propanol/water
autoclaved solution. The mixed solvent is then partially evaporated with a nitrogen
stream to approximately 1.5 g. The lower boiling n-propanol and water are
preferentially removed while the sample concentration is increased.
15. • Variation in dn/dc across the molecular size distribution caused by
comonomer compositional drift.
• Corrected by:
• Accounting for minor variation in dn/dc across the chromatogram (OR)
• Mark–Houwink calculation method
16. Polymer conformation &
unperturbed dimensions
• Conformation plots of three ionomer classes are plotted. All
PFSA ionomers were observed to have lower viscosity at
equivalent molar mass than PMMA
17. • The unperturbed dimensions was calculated from SEC
viscosity–molar mass data
• It suggested suspicious free rotation of the perfluoroalkane
units (can be attributed to the unusual structure and behavior
of these materials)
18. Our Conclusions
• Advantages:
– SEC system with triple detection can
• give high accuracy for perfluorosulfonated ionomers.
• Most info can be obtained
– Solvent Exchange process can
• improve all detector signals
• Also reduce the water and n-propanol peaks observed
in the system peak region.
– The autoclaving procedure
• provides molecular solutions without re-aggregation
(better peaks)
19. Our Conclusions
• Disadvantages:
– Dissolution temp is high compared to other
techniques
– Solvent exchange is somewhat tedious and
requires careful weighing of solutions before and
after evaporation to obtain accurate injected
sample concentrations
– Even though small, Variation in dn/dc causes
error in molar mass and viscosity from its original
value
20. Reference
• Size-exclusion chromatography of perfluorosulfonated ionomers by T.H.
Moureya, L.A. Slater a, R.C. Galipoa, R.J. Koestnerb :
Journal of Chromatography A, 1218 (2011) 5801– 5809