SlideShare a Scribd company logo
1 of 35
Download to read offline
Risk Management for
Medical Devices
An Overview of ISO 14971 & How To Apply a “Risk
Based Approach” to Your QMS Processes to
Address the Upcoming ISO 13485 Changes

September 2015 Presentation
Background
• Founder & VP QA/RA greenlight.guru
(Quality, Design Control & Risk
Management Software for Medical Devices)

• Founder Creo Quality (Consultancy)

• 17+ Med Device Industry Experience

• @CreoQuality & @greenlightguru

• Email: Jon.Speer@greenlight.guruJon D. Speer
Agenda
• Key risk management terms & definitions 

• Overview of ISO 14971 

• Incorporating risk management throughout the design control process

• Using risk management as a tool during design & development

• Preparing for the coming ISO 13485:201X changes 

• Describe relevance of ISO 14971 and the impending ISO 13485 revision 

• Applying a “risk-based approach” to all your QMS processes

• Q&A
Risk Management Terms &
Definitions
Key Terms & Definitions
• Risk	
  Management	
  - systematic application of management policies, procedures,
and practices to the tasks of analyzing, evaluating, controlling, and monitoring risk

• Risk - combination of the probability of occurrence of harm and the severity of that
harm

• Hazard - potential source of harm

• Hazardous	
  Situation - circumstance in which people, property, or the environment
are exposed to one or more hazard(s)

• Harm - physical injury or damage to the health of people, or damage to property or
the environment

• Severity - measure of the possible consequences of a hazard
Key Terms & Definitions
• Risk	
  Analysis - systematic use of available information to identify hazards and to
estimate the risk

• Risk	
  Estimation - process used to assign values to the probability of occurrence
of harm and the severity of that harm

• Risk	
  Evaluation - process of comparing the estimated risk against given risk
criteria to determine the acceptability of the risk

• Risk	
  Assessment - overall process comprising a risk analysis and a risk evaluation

• Risk	
  Control - process in which decisions are made and measures implemented
by which risks are reduced to, or maintained within, specified levels

• Residual	
  Risk - risk remaining after risk control measures have been taken
ISO 14971 Overview
ISO 14971 Overview
• Risk Management Planning

• Risk Analysis

• Risk Evaluation

• Risk Controls

• Overall Residual Risk Acceptability

• Risk Management Report

• Production & Post-Production Information

• Risk Management File
Risk Management Plan
• Scope of the Risk Management activities. Define the product included. It is
possible to have multiple products described within a single Risk Management
Plan.

• Describe the intended use of the product(s). 

• Identify all Risk Management activities planned throughout the product lifecycle.

• Define roles and responsibilities. Identify the Risk Management team and will be
reviewing and approving risk documentation.
• Criteria for the product’s risk acceptability. (Note, that often times this is likely
to be defined within your Risk Management Procedure.)

• Specify methods to verify Risk Control measures are implemented and reduce
risks.

• Define how post-production information will be captured and fed into Risk
Management activities for the product.
Risk Management Plan
Free Risk Management Plan Template
Free Risk Management Plan
Template + Exclusive Bonus
Offer: 

http://greenlight.guru/risk-
webinar-offer
See More -> http://greenlight.guru/risk-management-software
ISO	
  14971:2007	
  
Figure	
  1	
  -­‐	
  Risk	
  
Management	
  Process
ISO	
  14971:2007	
  
Figure	
  B.1	
  -­‐	
  Overview	
  
of	
  Risk	
  Management	
  
Activities	
  as	
  Applied	
  
to	
  Medical	
  Devices
Risk Assessment = Risk Analysis + Risk
Evaluation
Risk Control
Risk / Benefit Analysis (RBA)
• After you identify Risk Controls and evaluate residual risks, it is still
possible that you will have some risks that are still in the unacceptable
level. In these cases, it might make sense to conduct and document a
risk / benefit analysis (RBA).

• The RBA must be documented and provided objective evidence and
rationale for why the medical benefits outweigh the unacceptable risks.
If you are able to do so, the RBA is a special provision for moving
forward with unacceptable risks.
Overall Residual Risk Evaluation
Risk Management Report
• Summarize all your risk management activities

• Include any risk / benefit analyses 

• Explanation of overall risk acceptability of entire device

• Discuss your plans for evaluating risks in production and post-
production

• Recommend that you have executive management in your
company approve the Risk Management Report.
Production & Post-Production
Information
ISO 14971:2007 vs EN ISO 14971:2012
• Normative standard the same

• 2012 added “Z” annexes

• 2012: reduce risk as far as possible

• 2012: risk controls required for all risks

• 2012: risk / benefit analysis required for all risks
Risk Management 

+ Design Controls
Risk Management + Design Controls
• Understand the importance of Intended Use

• Product Risk Management is a cycle—even during product
development

• Risk Management & Design Controls have the same purpose
Risk Management As a Tool
Risk Management ≠
• The Longer You Wait, The Tougher It Gets – delaying design controls and
risk management documentation actually caused lengthy delays in projects

• Understanding The Purpose – design, develop, manufacture, and sell
medical devices that are as safe as possible for their intended uses

• Value Of Documentation – FDA inspectors, ISO auditors, and other
regulatory bodies and business associates are likely to review your design
controls and risk management
ISO 13485:201X
ISO 13485:201X – Major Changes
• Regulatory Requirements – not just ISO 13485

• Risk Management – entire QMS should involve risk-based approaches 

• Verification & Validation – plans in place plus supporting documented evidence

• Design Transfer – ensure Design Outputs are suitable for manufacturing before
becoming production specifications

• Purchasing Control – emphasis on purchased product has on safety and
performance

• Supplier Control – emphasis on suppliers ability to meet regulatory
requirements

• Monitoring QMS – during production and post-market
ISO 13485:201X – Some Details
• Terms “risk” & “regulatory” used >200 times (vs. ~50 in ISO
13485:2003)

• Align with industry best practices 

• Better align with regulations; expectation QMS complies with all
applicable regulatory requirements

• Concept of risk-based QMS

• Previously risk more aligned with D&D

• Now risk management methods to all QMS processes

• ISO 9001:2015 not aligned with ISO 13485:201X
ISO 13485:201X – Some Details
• More flexible – non-applications of clauses 6, 7, or 8 allowed
(with justification)

• “Product Quality” replaced with “Product Safety and
Performance”

• Emphasis on traceability during product realization

• Further definition on Design & Development activities (Design
Controls)

• More references to software (as a medical device)
ISO 13485:201X – Some Details
• Requirement to address UDI regulations

• Design & Development Transfer now its own subsection

• Effects of purchasing on safety and performance of medical device

• Ensuring suppliers meet regulatory requirements

• Supplier related risks
ISO 13485:201X – Overall Impact
• If you have ISO 13485:2003 and/or 21 CFR 820 compliant QMS, the
overall impact of the new standard should be relatively minor

• Should help incorporate use of ISO 14971 throughout QMS

• Improved linkage and integration of all clauses

• Expected to be published in 2016; 3 years to adopt (if already
certified)
Risk-Based QMS
Risk-Based QMS
• Greater emphasis on Risk due to ISO 14971 and regulatory bodies

• Manage risks throughout QMS

• Management Review – What is impact of failing to review critical items?

• Training – What are consequences of ineffective training?

• Calibration – What happens if done incorrectly?

• Purchasing – What effect do purchased products have on safety and
performance?

• Supplier Monitoring – Are suppliers able to meet regulatory requirements?

• Manage risks through entire product lifecycle (like ISO 14971)
Want to Work Together?
Free Risk Management Plan
Template + Exclusive Bonus Offer: 

http://greenlight.guru/risk-
webinar-offer
Email: Jon.Speer@greenlight.guru

THANKS!
35

More Related Content

What's hot

ISO 13485:2016 (Medical Devices - Quality Management System) Awareness Training
ISO 13485:2016 (Medical Devices - Quality Management System) Awareness TrainingISO 13485:2016 (Medical Devices - Quality Management System) Awareness Training
ISO 13485:2016 (Medical Devices - Quality Management System) Awareness Training
Operational Excellence Consulting
 

What's hot (20)

medical device regulatory approval in USA
medical device regulatory approval in USAmedical device regulatory approval in USA
medical device regulatory approval in USA
 
GHTF study group 3
GHTF study group 3GHTF study group 3
GHTF study group 3
 
ISO 13485:2016 QMS
ISO 13485:2016  QMSISO 13485:2016  QMS
ISO 13485:2016 QMS
 
ISO 13485: Quality Management System for Medical Device
ISO 13485: Quality Management System for Medical DeviceISO 13485: Quality Management System for Medical Device
ISO 13485: Quality Management System for Medical Device
 
ISO 13485:2016 (Medical Devices - Quality Management System) Awareness Training
ISO 13485:2016 (Medical Devices - Quality Management System) Awareness TrainingISO 13485:2016 (Medical Devices - Quality Management System) Awareness Training
ISO 13485:2016 (Medical Devices - Quality Management System) Awareness Training
 
Iso13485 ppt
Iso13485 pptIso13485 ppt
Iso13485 ppt
 
Process Validation & Verification (V&V) for Medical Devices
Process Validation & Verification (V&V) for Medical DevicesProcess Validation & Verification (V&V) for Medical Devices
Process Validation & Verification (V&V) for Medical Devices
 
NSF certification, Standard for dietary supplement
NSF certification, Standard for dietary supplementNSF certification, Standard for dietary supplement
NSF certification, Standard for dietary supplement
 
STED
STEDSTED
STED
 
Doc in pharma
Doc in pharmaDoc in pharma
Doc in pharma
 
International Medical Device Regulators Forum
International Medical Device Regulators ForumInternational Medical Device Regulators Forum
International Medical Device Regulators Forum
 
IDE Application Process and Best Practices
IDE Application Process and Best PracticesIDE Application Process and Best Practices
IDE Application Process and Best Practices
 
Medical Devices Regulation (MDR) 2017/745 - Clinical investigations
Medical Devices Regulation (MDR)  2017/745 - Clinical investigationsMedical Devices Regulation (MDR)  2017/745 - Clinical investigations
Medical Devices Regulation (MDR) 2017/745 - Clinical investigations
 
REGULATORY CONSIDERATIONS IN JAPAN.pptx
REGULATORY CONSIDERATIONS IN JAPAN.pptxREGULATORY CONSIDERATIONS IN JAPAN.pptx
REGULATORY CONSIDERATIONS IN JAPAN.pptx
 
China\'s Medical Device Regulatory Process
China\'s Medical Device Regulatory ProcessChina\'s Medical Device Regulatory Process
China\'s Medical Device Regulatory Process
 
GHTF
GHTFGHTF
GHTF
 
ASEAN COUNTRIES.pptx
ASEAN COUNTRIES.pptxASEAN COUNTRIES.pptx
ASEAN COUNTRIES.pptx
 
Understanding IEC 62304
Understanding IEC 62304Understanding IEC 62304
Understanding IEC 62304
 
Iso 13485:2016
Iso 13485:2016Iso 13485:2016
Iso 13485:2016
 
How to Prepare for the New EU Medical Device Regulations (MDR)
How to Prepare for the New EU Medical Device Regulations (MDR)How to Prepare for the New EU Medical Device Regulations (MDR)
How to Prepare for the New EU Medical Device Regulations (MDR)
 

Similar to Risk Management for Medical Devices - ISO 14971 Overview

Presentation OHSAS 18001
Presentation OHSAS  18001Presentation OHSAS  18001
Presentation OHSAS 18001
Gary Wong
 
G12: Implementation to Business Value
G12: Implementation to Business ValueG12: Implementation to Business Value
G12: Implementation to Business Value
HyTrust
 
Developing A Risk Based Information Security Program
Developing A Risk Based Information Security ProgramDeveloping A Risk Based Information Security Program
Developing A Risk Based Information Security Program
Tammy Clark
 

Similar to Risk Management for Medical Devices - ISO 14971 Overview (20)

Topic 1 - Risk Auditing 1-17.pdf
Topic 1 - Risk Auditing 1-17.pdfTopic 1 - Risk Auditing 1-17.pdf
Topic 1 - Risk Auditing 1-17.pdf
 
Risk Based Thinking ISO 9001 Presentation.pdf
Risk Based Thinking ISO 9001 Presentation.pdfRisk Based Thinking ISO 9001 Presentation.pdf
Risk Based Thinking ISO 9001 Presentation.pdf
 
Implementing Risk Based Thinking in HLS OF ISO 9001:2015 - Praneet Surti
Implementing Risk Based Thinking in HLS OF ISO 9001:2015 - Praneet SurtiImplementing Risk Based Thinking in HLS OF ISO 9001:2015 - Praneet Surti
Implementing Risk Based Thinking in HLS OF ISO 9001:2015 - Praneet Surti
 
How Does the New ISO 27001 Impact Your IT Risk Management Processes?
How Does the New ISO 27001 Impact Your IT Risk Management Processes?How Does the New ISO 27001 Impact Your IT Risk Management Processes?
How Does the New ISO 27001 Impact Your IT Risk Management Processes?
 
Neupart Bright Talk - How Does the New ISO 27001 Impact Your IT Risk Manageme...
Neupart Bright Talk - How Does the New ISO 27001 Impact Your IT Risk Manageme...Neupart Bright Talk - How Does the New ISO 27001 Impact Your IT Risk Manageme...
Neupart Bright Talk - How Does the New ISO 27001 Impact Your IT Risk Manageme...
 
How to Prepare Your Organization for the Transition to ISO 13485:2016
How to Prepare Your Organization for the Transition to ISO 13485:2016How to Prepare Your Organization for the Transition to ISO 13485:2016
How to Prepare Your Organization for the Transition to ISO 13485:2016
 
Risk Analysis
Risk AnalysisRisk Analysis
Risk Analysis
 
Presentation OHSAS 18001
Presentation OHSAS  18001Presentation OHSAS  18001
Presentation OHSAS 18001
 
4 Quality System Musts for Medtech Startups to Get Safer Products to Market F...
4 Quality System Musts for Medtech Startups to Get Safer Products to Market F...4 Quality System Musts for Medtech Startups to Get Safer Products to Market F...
4 Quality System Musts for Medtech Startups to Get Safer Products to Market F...
 
Auditing Information Security Management System Using ISO 27001 2013
Auditing Information Security Management System Using ISO 27001 2013Auditing Information Security Management System Using ISO 27001 2013
Auditing Information Security Management System Using ISO 27001 2013
 
Quality assurance and management, software engineering
Quality assurance and management, software engineeringQuality assurance and management, software engineering
Quality assurance and management, software engineering
 
ISO-45001 : 2018 _Awareness Training .pdf
ISO-45001 : 2018  _Awareness Training .pdfISO-45001 : 2018  _Awareness Training .pdf
ISO-45001 : 2018 _Awareness Training .pdf
 
Water Safety Plans with Odoo
Water Safety Plans with OdooWater Safety Plans with Odoo
Water Safety Plans with Odoo
 
Unleashing agile testing under medical regulations
Unleashing agile testing under medical regulationsUnleashing agile testing under medical regulations
Unleashing agile testing under medical regulations
 
(5) integrated management system (ims)
(5) integrated management system (ims)(5) integrated management system (ims)
(5) integrated management system (ims)
 
Quality management system(qrm)
Quality management system(qrm)Quality management system(qrm)
Quality management system(qrm)
 
Iso 22000 & 9000
Iso 22000 & 9000Iso 22000 & 9000
Iso 22000 & 9000
 
PECB Webinar: QMS Risk Assessment
PECB Webinar: QMS Risk AssessmentPECB Webinar: QMS Risk Assessment
PECB Webinar: QMS Risk Assessment
 
G12: Implementation to Business Value
G12: Implementation to Business ValueG12: Implementation to Business Value
G12: Implementation to Business Value
 
Developing A Risk Based Information Security Program
Developing A Risk Based Information Security ProgramDeveloping A Risk Based Information Security Program
Developing A Risk Based Information Security Program
 

More from Greenlight Guru

More from Greenlight Guru (20)

The Compliance Concern: Challenges and Roadblocks for Ensuring Compliance in ...
The Compliance Concern: Challenges and Roadblocks for Ensuring Compliance in ...The Compliance Concern: Challenges and Roadblocks for Ensuring Compliance in ...
The Compliance Concern: Challenges and Roadblocks for Ensuring Compliance in ...
 
How Electronic Data Capture Is Transforming the MedTech Industry
How Electronic Data Capture Is Transforming the MedTech IndustryHow Electronic Data Capture Is Transforming the MedTech Industry
How Electronic Data Capture Is Transforming the MedTech Industry
 
The ROI of Shifting Mindset From Compliance to Quality
The ROI of Shifting Mindset From Compliance to QualityThe ROI of Shifting Mindset From Compliance to Quality
The ROI of Shifting Mindset From Compliance to Quality
 
Leveraging Modern Software Technologies: MedTech’s Best Kept Secret
Leveraging Modern Software Technologies: MedTech’s Best Kept SecretLeveraging Modern Software Technologies: MedTech’s Best Kept Secret
Leveraging Modern Software Technologies: MedTech’s Best Kept Secret
 
Behind the Stats: Expectations vs. Reality of Bringing a Device to Market
Behind the Stats: Expectations vs. Reality of Bringing a Device to MarketBehind the Stats: Expectations vs. Reality of Bringing a Device to Market
Behind the Stats: Expectations vs. Reality of Bringing a Device to Market
 
Common Misconceptions on Medical Device Risk & Design Controls
Common Misconceptions on Medical Device Risk & Design ControlsCommon Misconceptions on Medical Device Risk & Design Controls
Common Misconceptions on Medical Device Risk & Design Controls
 
QMSR Harmonization: The Future of FDA's Quality Management System Regulation
QMSR Harmonization: The Future of FDA's Quality Management System RegulationQMSR Harmonization: The Future of FDA's Quality Management System Regulation
QMSR Harmonization: The Future of FDA's Quality Management System Regulation
 
QMS from a Regulatory Perspective
QMS from a Regulatory PerspectiveQMS from a Regulatory Perspective
QMS from a Regulatory Perspective
 
Modernizing your QMS to keep up with the Modern Age of Requirements
Modernizing your QMS to keep up with the Modern Age of RequirementsModernizing your QMS to keep up with the Modern Age of Requirements
Modernizing your QMS to keep up with the Modern Age of Requirements
 
eMDR Program and Process
eMDR Program and ProcesseMDR Program and Process
eMDR Program and Process
 
Moving up to the State of the Art in Risk Management
Moving up to the State of the Art in Risk ManagementMoving up to the State of the Art in Risk Management
Moving up to the State of the Art in Risk Management
 
ISO 13485: What's Next?
ISO 13485: What's Next?ISO 13485: What's Next?
ISO 13485: What's Next?
 
Computer Software Assurance (CSA): Understanding the FDA’s New Draft Guidance
Computer Software Assurance (CSA): Understanding the FDA’s New Draft GuidanceComputer Software Assurance (CSA): Understanding the FDA’s New Draft Guidance
Computer Software Assurance (CSA): Understanding the FDA’s New Draft Guidance
 
Making Headway Despite the Turbulence: Regulatory Requirements, Quality, and ...
Making Headway Despite the Turbulence: Regulatory Requirements, Quality, and ...Making Headway Despite the Turbulence: Regulatory Requirements, Quality, and ...
Making Headway Despite the Turbulence: Regulatory Requirements, Quality, and ...
 
Reducing Friction Between Companies and Regulatory Bodies
Reducing Friction Between Companies and Regulatory BodiesReducing Friction Between Companies and Regulatory Bodies
Reducing Friction Between Companies and Regulatory Bodies
 
Latin American Regulations - What you Don't Know
Latin American Regulations - What you Don't KnowLatin American Regulations - What you Don't Know
Latin American Regulations - What you Don't Know
 
Insights on the MedTech Regulatory and Clinical Environment in Israel
Insights on the MedTech Regulatory and Clinical Environment in IsraelInsights on the MedTech Regulatory and Clinical Environment in Israel
Insights on the MedTech Regulatory and Clinical Environment in Israel
 
The Global Guide to Human Factors and Usability Engineering Regulations
The Global Guide to Human Factors and Usability Engineering RegulationsThe Global Guide to Human Factors and Usability Engineering Regulations
The Global Guide to Human Factors and Usability Engineering Regulations
 
MDSAP Certification: Success and Failures
MDSAP Certification: Success and FailuresMDSAP Certification: Success and Failures
MDSAP Certification: Success and Failures
 
QMSR Harmonization - The Good the Bad and the Ugly
QMSR Harmonization - The Good the Bad and the UglyQMSR Harmonization - The Good the Bad and the Ugly
QMSR Harmonization - The Good the Bad and the Ugly
 

Recently uploaded

Call Girls In Majnu Ka Tilla 959961~3876 Shot 2000 Night 8000
Call Girls In Majnu Ka Tilla 959961~3876 Shot 2000 Night 8000Call Girls In Majnu Ka Tilla 959961~3876 Shot 2000 Night 8000
Call Girls In Majnu Ka Tilla 959961~3876 Shot 2000 Night 8000
dlhescort
 
Quick Doctor In Kuwait +2773`7758`557 Kuwait Doha Qatar Dubai Abu Dhabi Sharj...
Quick Doctor In Kuwait +2773`7758`557 Kuwait Doha Qatar Dubai Abu Dhabi Sharj...Quick Doctor In Kuwait +2773`7758`557 Kuwait Doha Qatar Dubai Abu Dhabi Sharj...
Quick Doctor In Kuwait +2773`7758`557 Kuwait Doha Qatar Dubai Abu Dhabi Sharj...
daisycvs
 
Call Girls In Noida 959961⊹3876 Independent Escort Service Noida
Call Girls In Noida 959961⊹3876 Independent Escort Service NoidaCall Girls In Noida 959961⊹3876 Independent Escort Service Noida
Call Girls In Noida 959961⊹3876 Independent Escort Service Noida
dlhescort
 
Call Now ☎️🔝 9332606886🔝 Call Girls ❤ Service In Bhilwara Female Escorts Serv...
Call Now ☎️🔝 9332606886🔝 Call Girls ❤ Service In Bhilwara Female Escorts Serv...Call Now ☎️🔝 9332606886🔝 Call Girls ❤ Service In Bhilwara Female Escorts Serv...
Call Now ☎️🔝 9332606886🔝 Call Girls ❤ Service In Bhilwara Female Escorts Serv...
Anamikakaur10
 
Call Girls Hebbal Just Call 👗 7737669865 👗 Top Class Call Girl Service Bangalore
Call Girls Hebbal Just Call 👗 7737669865 👗 Top Class Call Girl Service BangaloreCall Girls Hebbal Just Call 👗 7737669865 👗 Top Class Call Girl Service Bangalore
Call Girls Hebbal Just Call 👗 7737669865 👗 Top Class Call Girl Service Bangalore
amitlee9823
 
FULL ENJOY Call Girls In Majnu Ka Tilla, Delhi Contact Us 8377877756
FULL ENJOY Call Girls In Majnu Ka Tilla, Delhi Contact Us 8377877756FULL ENJOY Call Girls In Majnu Ka Tilla, Delhi Contact Us 8377877756
FULL ENJOY Call Girls In Majnu Ka Tilla, Delhi Contact Us 8377877756
dollysharma2066
 
Russian Call Girls In Rajiv Chowk Gurgaon ❤️8448577510 ⊹Best Escorts Service ...
Russian Call Girls In Rajiv Chowk Gurgaon ❤️8448577510 ⊹Best Escorts Service ...Russian Call Girls In Rajiv Chowk Gurgaon ❤️8448577510 ⊹Best Escorts Service ...
Russian Call Girls In Rajiv Chowk Gurgaon ❤️8448577510 ⊹Best Escorts Service ...
lizamodels9
 
Russian Call Girls In Gurgaon ❤️8448577510 ⊹Best Escorts Service In 24/7 Delh...
Russian Call Girls In Gurgaon ❤️8448577510 ⊹Best Escorts Service In 24/7 Delh...Russian Call Girls In Gurgaon ❤️8448577510 ⊹Best Escorts Service In 24/7 Delh...
Russian Call Girls In Gurgaon ❤️8448577510 ⊹Best Escorts Service In 24/7 Delh...
lizamodels9
 
Call Girls Electronic City Just Call 👗 7737669865 👗 Top Class Call Girl Servi...
Call Girls Electronic City Just Call 👗 7737669865 👗 Top Class Call Girl Servi...Call Girls Electronic City Just Call 👗 7737669865 👗 Top Class Call Girl Servi...
Call Girls Electronic City Just Call 👗 7737669865 👗 Top Class Call Girl Servi...
amitlee9823
 

Recently uploaded (20)

Call Girls In Majnu Ka Tilla 959961~3876 Shot 2000 Night 8000
Call Girls In Majnu Ka Tilla 959961~3876 Shot 2000 Night 8000Call Girls In Majnu Ka Tilla 959961~3876 Shot 2000 Night 8000
Call Girls In Majnu Ka Tilla 959961~3876 Shot 2000 Night 8000
 
Quick Doctor In Kuwait +2773`7758`557 Kuwait Doha Qatar Dubai Abu Dhabi Sharj...
Quick Doctor In Kuwait +2773`7758`557 Kuwait Doha Qatar Dubai Abu Dhabi Sharj...Quick Doctor In Kuwait +2773`7758`557 Kuwait Doha Qatar Dubai Abu Dhabi Sharj...
Quick Doctor In Kuwait +2773`7758`557 Kuwait Doha Qatar Dubai Abu Dhabi Sharj...
 
Call Girls In Noida 959961⊹3876 Independent Escort Service Noida
Call Girls In Noida 959961⊹3876 Independent Escort Service NoidaCall Girls In Noida 959961⊹3876 Independent Escort Service Noida
Call Girls In Noida 959961⊹3876 Independent Escort Service Noida
 
Call Now ☎️🔝 9332606886🔝 Call Girls ❤ Service In Bhilwara Female Escorts Serv...
Call Now ☎️🔝 9332606886🔝 Call Girls ❤ Service In Bhilwara Female Escorts Serv...Call Now ☎️🔝 9332606886🔝 Call Girls ❤ Service In Bhilwara Female Escorts Serv...
Call Now ☎️🔝 9332606886🔝 Call Girls ❤ Service In Bhilwara Female Escorts Serv...
 
Uneak White's Personal Brand Exploration Presentation
Uneak White's Personal Brand Exploration PresentationUneak White's Personal Brand Exploration Presentation
Uneak White's Personal Brand Exploration Presentation
 
Call Girls Hebbal Just Call 👗 7737669865 👗 Top Class Call Girl Service Bangalore
Call Girls Hebbal Just Call 👗 7737669865 👗 Top Class Call Girl Service BangaloreCall Girls Hebbal Just Call 👗 7737669865 👗 Top Class Call Girl Service Bangalore
Call Girls Hebbal Just Call 👗 7737669865 👗 Top Class Call Girl Service Bangalore
 
Organizational Transformation Lead with Culture
Organizational Transformation Lead with CultureOrganizational Transformation Lead with Culture
Organizational Transformation Lead with Culture
 
Phases of Negotiation .pptx
 Phases of Negotiation .pptx Phases of Negotiation .pptx
Phases of Negotiation .pptx
 
The Path to Product Excellence: Avoiding Common Pitfalls and Enhancing Commun...
The Path to Product Excellence: Avoiding Common Pitfalls and Enhancing Commun...The Path to Product Excellence: Avoiding Common Pitfalls and Enhancing Commun...
The Path to Product Excellence: Avoiding Common Pitfalls and Enhancing Commun...
 
BAGALUR CALL GIRL IN 98274*61493 ❤CALL GIRLS IN ESCORT SERVICE❤CALL GIRL
BAGALUR CALL GIRL IN 98274*61493 ❤CALL GIRLS IN ESCORT SERVICE❤CALL GIRLBAGALUR CALL GIRL IN 98274*61493 ❤CALL GIRLS IN ESCORT SERVICE❤CALL GIRL
BAGALUR CALL GIRL IN 98274*61493 ❤CALL GIRLS IN ESCORT SERVICE❤CALL GIRL
 
Value Proposition canvas- Customer needs and pains
Value Proposition canvas- Customer needs and painsValue Proposition canvas- Customer needs and pains
Value Proposition canvas- Customer needs and pains
 
FULL ENJOY Call Girls In Majnu Ka Tilla, Delhi Contact Us 8377877756
FULL ENJOY Call Girls In Majnu Ka Tilla, Delhi Contact Us 8377877756FULL ENJOY Call Girls In Majnu Ka Tilla, Delhi Contact Us 8377877756
FULL ENJOY Call Girls In Majnu Ka Tilla, Delhi Contact Us 8377877756
 
Russian Call Girls In Rajiv Chowk Gurgaon ❤️8448577510 ⊹Best Escorts Service ...
Russian Call Girls In Rajiv Chowk Gurgaon ❤️8448577510 ⊹Best Escorts Service ...Russian Call Girls In Rajiv Chowk Gurgaon ❤️8448577510 ⊹Best Escorts Service ...
Russian Call Girls In Rajiv Chowk Gurgaon ❤️8448577510 ⊹Best Escorts Service ...
 
SEO Case Study: How I Increased SEO Traffic & Ranking by 50-60% in 6 Months
SEO Case Study: How I Increased SEO Traffic & Ranking by 50-60%  in 6 MonthsSEO Case Study: How I Increased SEO Traffic & Ranking by 50-60%  in 6 Months
SEO Case Study: How I Increased SEO Traffic & Ranking by 50-60% in 6 Months
 
Cheap Rate Call Girls In Noida Sector 62 Metro 959961乂3876
Cheap Rate Call Girls In Noida Sector 62 Metro 959961乂3876Cheap Rate Call Girls In Noida Sector 62 Metro 959961乂3876
Cheap Rate Call Girls In Noida Sector 62 Metro 959961乂3876
 
Falcon's Invoice Discounting: Your Path to Prosperity
Falcon's Invoice Discounting: Your Path to ProsperityFalcon's Invoice Discounting: Your Path to Prosperity
Falcon's Invoice Discounting: Your Path to Prosperity
 
Russian Call Girls In Gurgaon ❤️8448577510 ⊹Best Escorts Service In 24/7 Delh...
Russian Call Girls In Gurgaon ❤️8448577510 ⊹Best Escorts Service In 24/7 Delh...Russian Call Girls In Gurgaon ❤️8448577510 ⊹Best Escorts Service In 24/7 Delh...
Russian Call Girls In Gurgaon ❤️8448577510 ⊹Best Escorts Service In 24/7 Delh...
 
Famous Olympic Siblings from the 21st Century
Famous Olympic Siblings from the 21st CenturyFamous Olympic Siblings from the 21st Century
Famous Olympic Siblings from the 21st Century
 
Call Girls Ludhiana Just Call 98765-12871 Top Class Call Girl Service Available
Call Girls Ludhiana Just Call 98765-12871 Top Class Call Girl Service AvailableCall Girls Ludhiana Just Call 98765-12871 Top Class Call Girl Service Available
Call Girls Ludhiana Just Call 98765-12871 Top Class Call Girl Service Available
 
Call Girls Electronic City Just Call 👗 7737669865 👗 Top Class Call Girl Servi...
Call Girls Electronic City Just Call 👗 7737669865 👗 Top Class Call Girl Servi...Call Girls Electronic City Just Call 👗 7737669865 👗 Top Class Call Girl Servi...
Call Girls Electronic City Just Call 👗 7737669865 👗 Top Class Call Girl Servi...
 

Risk Management for Medical Devices - ISO 14971 Overview

  • 1. Risk Management for Medical Devices An Overview of ISO 14971 & How To Apply a “Risk Based Approach” to Your QMS Processes to Address the Upcoming ISO 13485 Changes September 2015 Presentation
  • 2. Background • Founder & VP QA/RA greenlight.guru (Quality, Design Control & Risk Management Software for Medical Devices) • Founder Creo Quality (Consultancy) • 17+ Med Device Industry Experience • @CreoQuality & @greenlightguru • Email: Jon.Speer@greenlight.guruJon D. Speer
  • 3. Agenda • Key risk management terms & definitions  • Overview of ISO 14971  • Incorporating risk management throughout the design control process • Using risk management as a tool during design & development • Preparing for the coming ISO 13485:201X changes  • Describe relevance of ISO 14971 and the impending ISO 13485 revision  • Applying a “risk-based approach” to all your QMS processes • Q&A
  • 4. Risk Management Terms & Definitions
  • 5. Key Terms & Definitions • Risk  Management  - systematic application of management policies, procedures, and practices to the tasks of analyzing, evaluating, controlling, and monitoring risk • Risk - combination of the probability of occurrence of harm and the severity of that harm • Hazard - potential source of harm • Hazardous  Situation - circumstance in which people, property, or the environment are exposed to one or more hazard(s) • Harm - physical injury or damage to the health of people, or damage to property or the environment • Severity - measure of the possible consequences of a hazard
  • 6. Key Terms & Definitions • Risk  Analysis - systematic use of available information to identify hazards and to estimate the risk • Risk  Estimation - process used to assign values to the probability of occurrence of harm and the severity of that harm • Risk  Evaluation - process of comparing the estimated risk against given risk criteria to determine the acceptability of the risk • Risk  Assessment - overall process comprising a risk analysis and a risk evaluation • Risk  Control - process in which decisions are made and measures implemented by which risks are reduced to, or maintained within, specified levels • Residual  Risk - risk remaining after risk control measures have been taken
  • 8. ISO 14971 Overview • Risk Management Planning • Risk Analysis • Risk Evaluation • Risk Controls • Overall Residual Risk Acceptability • Risk Management Report • Production & Post-Production Information • Risk Management File
  • 9. Risk Management Plan • Scope of the Risk Management activities. Define the product included. It is possible to have multiple products described within a single Risk Management Plan. • Describe the intended use of the product(s). • Identify all Risk Management activities planned throughout the product lifecycle. • Define roles and responsibilities. Identify the Risk Management team and will be reviewing and approving risk documentation.
  • 10. • Criteria for the product’s risk acceptability. (Note, that often times this is likely to be defined within your Risk Management Procedure.) • Specify methods to verify Risk Control measures are implemented and reduce risks. • Define how post-production information will be captured and fed into Risk Management activities for the product. Risk Management Plan
  • 11. Free Risk Management Plan Template Free Risk Management Plan Template + Exclusive Bonus Offer: http://greenlight.guru/risk- webinar-offer
  • 12. See More -> http://greenlight.guru/risk-management-software
  • 13. ISO  14971:2007   Figure  1  -­‐  Risk   Management  Process
  • 14. ISO  14971:2007   Figure  B.1  -­‐  Overview   of  Risk  Management   Activities  as  Applied   to  Medical  Devices
  • 15. Risk Assessment = Risk Analysis + Risk Evaluation
  • 17. Risk / Benefit Analysis (RBA) • After you identify Risk Controls and evaluate residual risks, it is still possible that you will have some risks that are still in the unacceptable level. In these cases, it might make sense to conduct and document a risk / benefit analysis (RBA). • The RBA must be documented and provided objective evidence and rationale for why the medical benefits outweigh the unacceptable risks. If you are able to do so, the RBA is a special provision for moving forward with unacceptable risks.
  • 18. Overall Residual Risk Evaluation
  • 19. Risk Management Report • Summarize all your risk management activities • Include any risk / benefit analyses • Explanation of overall risk acceptability of entire device • Discuss your plans for evaluating risks in production and post- production • Recommend that you have executive management in your company approve the Risk Management Report.
  • 21. ISO 14971:2007 vs EN ISO 14971:2012 • Normative standard the same • 2012 added “Z” annexes • 2012: reduce risk as far as possible • 2012: risk controls required for all risks • 2012: risk / benefit analysis required for all risks
  • 22. Risk Management 
 + Design Controls
  • 23.
  • 24. Risk Management + Design Controls • Understand the importance of Intended Use • Product Risk Management is a cycle—even during product development • Risk Management & Design Controls have the same purpose
  • 26. Risk Management ≠ • The Longer You Wait, The Tougher It Gets – delaying design controls and risk management documentation actually caused lengthy delays in projects • Understanding The Purpose – design, develop, manufacture, and sell medical devices that are as safe as possible for their intended uses • Value Of Documentation – FDA inspectors, ISO auditors, and other regulatory bodies and business associates are likely to review your design controls and risk management
  • 28. ISO 13485:201X – Major Changes • Regulatory Requirements – not just ISO 13485 • Risk Management – entire QMS should involve risk-based approaches • Verification & Validation – plans in place plus supporting documented evidence • Design Transfer – ensure Design Outputs are suitable for manufacturing before becoming production specifications • Purchasing Control – emphasis on purchased product has on safety and performance • Supplier Control – emphasis on suppliers ability to meet regulatory requirements • Monitoring QMS – during production and post-market
  • 29. ISO 13485:201X – Some Details • Terms “risk” & “regulatory” used >200 times (vs. ~50 in ISO 13485:2003) • Align with industry best practices • Better align with regulations; expectation QMS complies with all applicable regulatory requirements • Concept of risk-based QMS • Previously risk more aligned with D&D • Now risk management methods to all QMS processes • ISO 9001:2015 not aligned with ISO 13485:201X
  • 30. ISO 13485:201X – Some Details • More flexible – non-applications of clauses 6, 7, or 8 allowed (with justification) • “Product Quality” replaced with “Product Safety and Performance” • Emphasis on traceability during product realization • Further definition on Design & Development activities (Design Controls) • More references to software (as a medical device)
  • 31. ISO 13485:201X – Some Details • Requirement to address UDI regulations • Design & Development Transfer now its own subsection • Effects of purchasing on safety and performance of medical device • Ensuring suppliers meet regulatory requirements • Supplier related risks
  • 32. ISO 13485:201X – Overall Impact • If you have ISO 13485:2003 and/or 21 CFR 820 compliant QMS, the overall impact of the new standard should be relatively minor • Should help incorporate use of ISO 14971 throughout QMS • Improved linkage and integration of all clauses • Expected to be published in 2016; 3 years to adopt (if already certified)
  • 34. Risk-Based QMS • Greater emphasis on Risk due to ISO 14971 and regulatory bodies • Manage risks throughout QMS • Management Review – What is impact of failing to review critical items? • Training – What are consequences of ineffective training? • Calibration – What happens if done incorrectly? • Purchasing – What effect do purchased products have on safety and performance? • Supplier Monitoring – Are suppliers able to meet regulatory requirements? • Manage risks through entire product lifecycle (like ISO 14971)
  • 35. Want to Work Together? Free Risk Management Plan Template + Exclusive Bonus Offer: http://greenlight.guru/risk- webinar-offer Email: Jon.Speer@greenlight.guru THANKS! 35