SlideShare a Scribd company logo
1 of 90
Calibration of
CMMs
Using E&R Test and
Laser Interferometer
By: Hassan Habib
Things we are going to learn
• About Coordinate Measuring Machines (CMMs)
• Measurements
• What is a CMM?
• How does a CMM work?
• Calibration
• What is calibration?
• Calibration standards for CMM
• Calibration procedure for CMM
• E&R Test
• Introduction
• Definition of errors & Material Standard
• Choice of Artifacts
• Preliminary Setup
• Measurements for error in length
• Calculation of results
• Probing test error
Things we are going to learn
• Laser Interferometery
• Introduction
• Principle of Laser Interferometers
• Widely used interferometers in the market
• Components of Laser interferometer
• Preparing the machine for calibration
• Definition of geometrical deviations to be measured
• Setting up the laser
• Set up the measurements/optics
• Collect the data and analyze according to international
standards
• Compensation of results
• Some Post Checks
• Beauty is in the perfection of creation. Humans have been
creating since their presence on earth
• However, accuracy of these measurements has evolved
over time and so has the beauty they have created
• The earliest examples of accuracy can be found in the
construction of Great Pyramids (The difference between
height of two opposite corners at its base is 13 mm.)
About Coordinate Measuring Machines (CMM)
Measurements
About Coordinate Measuring Machines (CMM)
Measurements
Evolution of Measuring Instruments
Cubit
Micrometer
Gauge
Blocks
Length
Comparator
Gauges and
Dial
Indicators
Visual
Inspection
machines
Universal
Measuring
Machines
CMM
About Coordinate Measuring Machines (CMM)
What is a CMM?
• Modern machines used for very accurate and precise
measurements
• CMM works on the principle of Coordinate Measuring i.e.
measurement based on collection of data points taken in a
Cartesian Coordinate System
• “The primary function of a CMM is to measure the actual
shape of a work piece, compare it against the desired
shape, and evaluate the metrological information such as
size, form, location, and orientation.” [Ref: Read ‘Notes’]
About Coordinate Measuring Machines (CMM)
What is a CMM?
• CMM can measure complex geometrical tolerances and
deviations on manufactured parts
• The accuracy of these machines today are closer to 1µm
• CMMs are capable of measuring point to point coordinates
and they can also analyze continuous data points using
advanced touch probes (SP600 and PH20)
• CMMs are extensively used in Aerospace and automotive
sectors where increased accuracy of measurement is
required
Features
About Coordinate Measuring Machines (CMM)
How does CMM work?
About Coordinate Measuring Machines (CMM)
How does CMM work?
Types of CMMs
Calibration
What is calibration?
• It is the process of verifying and adjusting the accuracy
of Measuring & Monitoring Equipment (MME) and
machines by comparing them with standards of known
accuracy.
• The adjustment of instruments is performed by
compensating the errors into the instrument.
• OEM (Original Equipment Manufacturer) of the
instrument usually defines calibration interval,
environment of usage and tolerance limits within which
the instrument will conform to its performance standard.
• It is performed in regular intervals so as to ensure that
the instrument is reliable.
Calibration
What is calibration?
• Calibration provides the confidence that their accuracy
is as per the given specifications of OEM.
• It ensures the repeatability of the measurements taken
by the equipment.
• The uncertainty is kept at minimum level further
building the confidence of measurements.
• It is performed in regular intervals so as to ensure that
the instrument is reliable.
• Calibrating instruments through certified bodies
increases the confidence level of customers for your
organization.
Importance of Calibration
Calibration
Calibration of CMMs
• The calibration of CMM benefits as per the stated
benefits of calibration.
• For a CMM there can be number of sources of error that
will remain undetected thereby nullifying usability.
• The calibration of CMM is performed according to the
guidelines provided in ISO 10360-2.
• National Physics Laboratory (NPL) has also provided
standard procedures for verification of performance
level of CMMs by detailing guidelines. These
guidelines are provided to perform E&R test on the
machine using organizational standards used for
calibration.
Calibration
Calibration of CMMs
Calibration Standards for CMMs
Calibration
Calibration of CMMs
• In the scope of this presentation we are going to study
two standards of calibration of CMMs:
• E&R test for verification of length measurements
• Verification and Compensation of geometric errors
with Laser Interferometry
Calibration
Calibration Procedure for CMMs
Choice of
Artifacts
Preliminary
Setup
Length
Measurement
Calculation of
results
Probing Test
Procedure of verification for length measurement through E&R test
Calibration
Calibration Procedure for CMMs
Procedure of verification for geometric deviations through Laser Interferometer
E&R Test
Introduction
• The tests help in demonstrating traceability to national
standards and estimating the accuracy of measurements
made with three dimensional CMMs for maintaining
confidence and reliability in the measurements.
• The tests also indicate health of the machine that is
necessary to perform maintenance.
E&R Test
Introduction
• Wear of components - the guide ways, the scales, the
probe system and the qualification sphere;
• Environment in which the CMM operates - the ambient
temperature, temperature gradients, humidity and
vibration;
• The probing strategy used – the magnitude and direction
of the probe force, the type of probe stylus used and the
measuring speed of the probe; and
• Characteristics of the workpiece – elasticity, surface
roughness, hardness and the mass of the component
Possible sources of error in CMM
E&R Test
Definition of Errors & material standard
• ISO 10360 strongly recommended that the material
standard should be either a step gauge, end bar or a
series of gauge blocks conforming to ISO 3650
• The material standard of size used for the tests must be
calibrated.
• The uncertainty of calibration must be taken into
consideration and the calibrations must be traceable to
the relevant national standard.
E&R Test
• It is the term that specifies the length measuring
accuracy of their CMM. EMPE,L is defined as the extreme
value of the error of indication of a CMM for size
measurement, permitted by specifications, regulations
etc.
• It is measured in one of the following ways:
• a) EL,MPE = ± minimum of (A + L/K) and B
• b) EL, MPE = ± (A + L/K)
• c) EL, MPE = ± B
where
A is a positive constant, expressed in micrometres and supplied by the
manufacturer;
K is a dimensionless positive constant supplied by the manufacturer;
L is the measured size, in millimetres; and
B is the maximum permissible error
a)
b)
c)
Definition of Errors & material standard
E&R Test
• The E&R test involves two types of measurement errors.
• Volumetric length measuring error E
• It applies to all measurements of distances, diameters, and positional
tolerances.
• Volumetric Probing Error P
• It applies to all Form measurements of straightness, flatness, Cylindricity,
roundness and free form tolerances.
Definition of Errors & material standard
E&R Test
Choice of artifacts
• The choice of artifacts depend on the recommendations
from the manufacturer and or the size of you CMM.
Various types of artifacts are available for verification
and re-verification, and interim check tests, some of
them are:
• Step Gauge
• Length bar
• Ball Plate
• Hole Plate
• Purpose made test piece
E&R Test
Definition of Errors & material standard
Comparison between different artifacts
E&R Test
Preliminary Setup
• Below are some of the pre-requisites for performing the
E&R test:
• CMM must be operated in accordance with the
procedure stated in the instruction manual
including machine start up, probe qualification and
probe configuration.
• Manufacturer supplied test sphere must be used.
• Limits for permissible environmental conditions,
such as temperature conditions, air humidity and
vibration that influence the measurements are
usually specified by the manufacturer
• Cleaning of stylus tip;
• Thermal stability of the probing system
• Weight of stylus system and/or probing system;
and location, type, number of thermal sensors
E&R Test
• For the E test a set of 5 length gauges is measured three
times in 7 spatial positions.
• Total number of measurements:
• 3 x 5 x 7 = 105
• 100 % of results must be in the specified limits
• The seven spatial positions are:
• Along x-axis
• Along y-axis
• Along z-axis
• Along s-partial 1 (Diagonal in XY)
• Along s-partial 2 (Diagonal in YZ)
• Along s-partial 3 (Diagonal in XZ)
• Along s-partial 4 (Diagonal in XYZ)
Measurements for error in length
E Test Sample positions
E&R Test
Measurements for error in length
E Test Measuring Lines for 7 spatial locations
E&R Test
Calculation of results
• For each of the 105 measurements the error of length
measurement, EL is calculated.
• Its value is the absolute value of the difference between
the indicated value of the relevant test length and the
true value of the material standard.
• Sample results are shown:
E&R Test
Calculation of results
E&R Test
Calculation of results
E&R Test
Calculation of results
Graphical representation of the results
E&R Test
Calculation of results
• From the results it can be seen that some of the thirty-
five test lengths have values of the error of length
measurement
• These values will have to be measured again ten times
each at the relevant configuration
Interpretation of the results
E&R Test
Probing Test Error
• This test of the CMM probing system is used to
establish whether the CMM is capable of measuring
within the manufacturer‘s stated value of PFTU, MPE by
determining the range of values of the radial distance r
when measuring a reference sphere.
Where,
P: associated with the probing system
F: apparent Form error
T: contact probing (that is to say Tactile)
U: single (that is to say Unique)
• It is advisable to carry out this test before an acceptance
or re-verification test.
Introduction
E&R Test
Probing Test Error
• The sphere supplied by the manufacturer for probe
qualifying purposes (reference sphere) should not be
used for the probing error test.
Introduction
E&R Test
Probing Test Error
• The probing error is a positive constant, the value of
which is supplied by the CMM manufacturer.
• The test sphere should be between 10 mm and 50 mm
diameter.
• The test sphere should be mounted rigidly to overcome
errors due to bending of the mounting stem.
• Twenty-five points are measured and recorded. It is a
requirement that the points are approximately evenly
distributed over at least a hemisphere of the test sphere.
• Their position is at the discretion of the user
Procedure
E&R Test
Probing Test Error
• one point on the pole (defined by the direction of the
stylus shaft) of the test sphere;
• four points (equally spaced) 22.5° below the pole;
• eight points (equally spaced) 45° below the pole and
rotated 22.5° relative to the previous group;
• four points (equally spaced) 67.5° below the pole and
rotated 22.5° relative to the previous group; and
• eight points (equally spaced) 90° below the pole (i.e., on
the equator) and rotated 22.5° relative to the previous
group.
Measurements
Measurement Pattern
E&R Test
Probing Test Error
Results
E&R Test
Probing Test Error
Graphical representation of results
E&R Test
Probing Test Error
Interpretation of Results
• If the range rmax - rmin of the twenty-five radial distances
(PFTU) is no greater than the manufacturer‘s stated
value of PFTU, MPE when taking into account the
measurement uncertainty, then the performance of the
probing system is verified
Laser Interferomter
Introduction
• As stated before laser interferometer is a higher standard
of measurement that is used for the calibration of CMM
for its geometrical deviations.
• The values from this calibration are also used as
compensations for the deviations in the machine
controller.
• Once these compensations are provided to the controller
all machine errors are compensated and the machine
returns to the factory provided performance standard.
• Laser interferometry has slowly evolved into easy to use
equipment that can help to perform various tasks.
• This type of calibrations all started with Michelson’s
Interferometer
Laser Interferomter
Principle
• The working principle of laser interferometers today
used, work on the principle of the Michelson’s
Interferometer.
• The Michelson interferometer is common configuration
for optical interferometry and was invented by Albert
Abraham Michelson.
• Albert Michelson and Edward Morley performed their
famous Michelson-Morley experiment in 1887.
Edward Morley
Laser Interferomter
Principle
• Using a beam splitter, a light source is split into two
arms.
• Each of those is reflected back toward the beam splitter
which then combines their amplitudes
interferometrically.
• The resulting interference pattern that is not directed
back toward the source is typically directed to some type
of photoelectric detector or camera.
• Depending on the interferometer's particular
application, the two paths may be of different lengths or
include optical materials or components under test.
Interference Patterns from an
interferometer
Michelson’s Interferometer
Laser Interferomter
Principle
Michelson’s Interferometer
Laser Interferomter
Principle
• M is partially reflective, so part of the light is
transmitted through to point B while some is reflected in
the direction of A.
• Both beams recombine at point C' to produce an
interference pattern incident on the detector at point E
(or on the retina of a person's eye).
• If there is a slight angle between the two returning
beams, for instance, then an imaging detector will
record a sinusoidal fringe pattern.
• If there is perfect spatial alignment between the
returning beams, then there will not be any such pattern
but rather a constant intensity over the beam dependent
on the differential path length.
Michelson’s Interferometer
Laser Interferomter
Principle
Modern Laser Interferometer
Example of laser measurement
Laser Interferomter
Widely used interferometers in the market
5529A
XL 80
Laser Interferomter
Components of laser interferometer
• Laser Head
• Environment compensation unit
• Material temperature sensors
• Power Supply
• Air sensors
• Tripod stand with stage
• Laptop with necessary software
• Optics for different measurements
Major Components
Laser Interferomter
Preparing the machine for calibration
Preliminary setup
Check the air
filters
•Including
machine filters
and air dryer
filters
Check air tubing
for replacement
•Wet air would
probably
require
replacement of
tubing
Check air
bearings
•They must
have specified
gap of air
cushion
Remove rubber
pads from base
•Removal
makes
foundation stiff
Balance the
machine bed
•Use
inclinometers
Note:
– You can use the master square to balance the z-axis
– Balance the machine on three nodes and remove any redundant
rests from the base
– There should be no turbulence in air. Turbulence will cause laser
error
– The environment must be controlled as much as possible.
Laser Interferomter
Definition of geometrical deviations to be measured
Measurements needed for error mapping
• Error mapping is done to calculate the 21 geometric
deviations that can occur in an articulating machine
including CMM and machining centers.
• First these are measured and then compensated using the
appropriate software to the controller of the machine.
• The deviations are majorly of these types:
• Linearity (3)
• Straightness (6)
• Rotation (9)
• Squareness (3)
Laser Interferomter
Definition of geometrical deviations to be measured
Measurements needed for error mapping
Laser Interferomter
Definition of geometrical deviations to be measured
Geometrical compensation parameters
• Usually the geometric deviations are termed as
compensation parameters and are designated values
such as:
Rxx = Linear straightness in x-axis
Ryy = Linear straightness in y-axis
Rxy = Horizontal straightness of x-axis
Rxz = Vertical straightness of x-axis
Dxy = Yaw of x-axis
Dxz = Pitch of x-axis
.
.
.
.
etc.
Laser Interferomter
Definition of geometrical deviations to be measured
Geometrical compensation parameters
• For the complete error map these measurements are to
be compensated in the controller of the machine.
• Software such as Geocomp are used to upload the
valued of these deviations into the controller.
• Two of these measurements can be taken from
inclinometer. These are the straightness of x-axis and y-
axis.
• Rest of the measurements are taken from a laser
interferometer
• The linear straightness of the z-axis can be measured by
using two Dzy measurements.
Laser Interferomter
Setting up the laser
Generic Procedure
• Here we will outline a generic procedure to perform the
linear measurement on ML10 laser interferometer of
Renishaw. It will give us an overview of the procedure.
It is similar to that of Agilent laser interferometer.
Laser Interferomter
Setting up the laser
Generic Procedure
• First step is to setup the stage on the tripod stand.
Secure and tight it on the stand.
• Position the Laser head on the tripod stand with stage.
• Arrange the laser interferometer according to the
measurement you are about to take.
• Switch on the laser. You have to wait for a specified
time for the laser to get stabilized.
Laser Interferomter
Setting up the laser
Generic Procedure
• Align and fix the optics for the measurement in
our case it will be the linear interferometer that
includes linear reflector and linear beam splitter
as well as clamp blocks to fix it on the machine
head.
• Now you have to mount the interferometer on the
machine bed and reflector has to be attached on
machine spindle. Side surfaces of linear beam
splitter and reflector have to be exactly parallel.
Note: In our case the linear interferometer stays stationary while the
splitter moves along the axis.
Laser Interferomter
Setting up the laser
Generic Procedure
• The shutter of the laser head can be rotated. Rotate it
so that the laser leaving the head has a reduced beam
diameter.
• Now adjust the tripod so as the spirit level of the tripod
is at the central position.
• Bring closer the reflector to the laser head and observe
a white spot target on the front. Now move the
machine in the x-axis until the beam hits the target.
Shutter at closed position with
no laser emitted
Laser Interferomter
Setting up the laser
Generic Procedure
• Now remove white target and check if the beam hits the
center of the laser head target on its shutter. If it doesn’t,
keep on adjusting the position of the machine until it hits
the target at the center.
• Now adjust reflector and splitter as close as possible and
align them together.
• Make sure that the faces are parallel with the one another
and with the machine axis.
Laser Interferomter
Setting up the laser
Generic Procedure
• Now use a target at the input aperture with white spot
at the top and translate machine axis vertically and
horizontally so that the beam hits the target.
• Now take away the target and check if the returned
beam from the interferometer hits the center of the
shutter. If it does not repeat the motion of the machine
until it does hit the center.
Laser Interferomter
Set up the measurements/optics
Introduction
• We have elaborated about the 21 measurements needed
for complete calibration of CMM.
• Majorly these measurements are divided into three
groups:
– Linearity
– Straightness
– Rotation and
– Squareness
• Now we will elaborate the different configurations of
the optics that are used to take these measurements
Laser Interferomter
Set up the measurements/optics
Linearity
• Linear measurements are made at multiple points
along a machine’s travel path to measure linear
displacement and velocity.
• It is checked to improve positioning accuracy along an
axis for any machine that requires positioning
accuracy and velocity control.
Laser Interferomter
Set up the measurements/optics
Optics required
• To make linear measurements following optics are
required:
– Beam Splitter
– Linear Reflectors
– Targets
– Height adjustment fixtures
Renishaw XL 80
Agilent 5529A
Laser Interferomter
Set up the measurements/optics
Principle
Laser Interferomter
Set up the measurements/optics
Configuration on machine
Laser Interferomter
Set up the measurements/optics
Straightness
• Straightness measurements evaluate the unwanted side
to-side or up-and-down motion of a machine tool’s
travel in a specified direction.
Laser Interferomter
Set up the measurements/optics
Optics required
• To make straightness measurements following optics
are required:
– Straightness reflector
– Straightness interferometer
– Targets
– Height adjustment fixtures
Renishaw XL 80
Agilent 5529A
Laser Interferomter
Set up the measurements/optics
Principle
Laser Interferomter
Set up the measurements/optics
Configuration on machine
Laser Interferomter
Set up the measurements/optics
Configuration on machine
Laser Interferomter
Set up the measurements/optics
Rotation
• Angular measurements are made at multiple points
along a machine’s travel path to test for rotation about
an axis perpendicular to the axis of motion (roll, pitch
and yaw)
• Unwanted angular motion in a machine tool causes
positioning errors that reduce the overall accuracy of
your machine.
Laser Interferomter
Set up the measurements/optics
Optics required
• To make straightness measurements following optics
are required:
– Angular reflector
– Angular interferometer
– Targets
– Height adjustment fixtures
Renishaw XL 80
Agilent 5529A
Laser Interferomter
Set up the measurements/optics
Principle
Laser Interferomter
Set up the measurements/optics
Configuration on machine
Laser Interferomter
Set up the measurements/optics
Configuration on machine
Laser Interferomter
Set up the measurements/optics
Squareness
• Angular measurements are made at multiple points
along a machine’s travel path to test for rotation about
an axis perpendicular to the axis of motion (roll, pitch
and yaw)
• Unwanted angular motion in a machine tool causes
positioning errors that reduce the overall accuracy of
your machine.
Laser Interferomter
Set up the measurements/optics
Optics required
• To make squareness measurements following optics
are required:
– Optical Square
– Brackets
– Targets
– Height adjustment fixtures
Renishaw XL 80
Agilent 5529A
Laser Interferomter
Set up the measurements/optics
Principle
Laser Interferomter
Set up the measurements/optics
Configuration on the machine
Laser Interferomter
Collect the data and analyze according to
international standards
Parameters to be entered
• After the required measurement optics have been setup
and is ready to be measured. Usually the following
information is entered in the software to begin taking
measurements:
– Start position for the measurement
– End position (Length of the axis of your machine)
– Interval (It can be 10, 50, 100)
– No. of points
– No. of cycles (For bidirectional checks)
Laser Interferomter
Collect the data and analyze according to
international standards
Sample screen for entering parameters
Laser Interferomter
Collect the data and analyze according to
international standards
Sample for measurement screen
Laser Interferomter
Collect the data and analyze according to
international standards
Analyzing data
• After the readings have been taken the software shows
statistical chart that represents the points that are out of
tolerance.
• This graphical representation is according to the
standard you have chosen. You choose the standard
that your OEM has specified for the calibration of the
machine.
Laser Interferomter
Collect the data and analyze according to
international standards
Sample for analyzing data
Laser Interferomter
Compensation of results
• After the data has been analyzed according to a
standard, software provides the compensation table for
the measurement you have taken.
• This information is then fed into the specific software
of the machine used for this purpose. E.g Geocomp is
used for compensation for Global Image CMM.
Laser Interferomter
Compensation of results
Example of compensation table
Laser Interferomter
Some post checks
Post Checks
• After all the measurements have been compensated,
machine is operated in dry run mode. That checks
healthy operation.
• E&R test elaborated before is also performed to check
that all the measurements of the machine are in the
tolerance defined by the macnufacturer
Calibration of CMMs
References
• Coordinate Measuring Machines and Systems by
Robert J. Hocken & Paulo H. Pereira
• Leitz 10360-2 guide
• NPL Guidelines – Good Practice Guide No. 42
• Agilent 5529A Manual
• Renishaw user guide for XL-80
• Wikipedia
Calibration of CMMs

More Related Content

What's hot

gauges, Metrology, Types of gauges
gauges, Metrology, Types of gaugesgauges, Metrology, Types of gauges
gauges, Metrology, Types of gaugesRaju D
 
Coordinate Measuring Machine(CMM)
Coordinate Measuring Machine(CMM)Coordinate Measuring Machine(CMM)
Coordinate Measuring Machine(CMM)Tushar Makvana
 
Introduction to Mechanical Measurements and Metrology
Introduction to Mechanical Measurements and Metrology Introduction to Mechanical Measurements and Metrology
Introduction to Mechanical Measurements and Metrology taruian
 
6 CHAPTER SIX - COMPUTER AIDED INSPECTION.ppt
6 CHAPTER SIX - COMPUTER AIDED INSPECTION.ppt6 CHAPTER SIX - COMPUTER AIDED INSPECTION.ppt
6 CHAPTER SIX - COMPUTER AIDED INSPECTION.pptfahamkhan7
 
Metrology and measurements(Coordinate Measuring Machine)
Metrology and measurements(Coordinate Measuring Machine)Metrology and measurements(Coordinate Measuring Machine)
Metrology and measurements(Coordinate Measuring Machine)Vignesh Ramanathan
 
Metrology & measurement
Metrology & measurement Metrology & measurement
Metrology & measurement Nouman Ali
 
Computer aided quality control
Computer aided quality controlComputer aided quality control
Computer aided quality controlAhmad Bajwa
 
LINEAR and ANGULAR Measurement
LINEAR and ANGULAR MeasurementLINEAR and ANGULAR Measurement
LINEAR and ANGULAR MeasurementDhruv Parekh
 
Auto collimator angle dekkor
Auto collimator angle dekkorAuto collimator angle dekkor
Auto collimator angle dekkorNaman Dave
 
Introduction to Mechanical Measurement
Introduction to Mechanical MeasurementIntroduction to Mechanical Measurement
Introduction to Mechanical MeasurementNaman Dave
 
Antifriction guideways, feed drives, spindles
Antifriction guideways, feed drives, spindlesAntifriction guideways, feed drives, spindles
Antifriction guideways, feed drives, spindlesvedamsh95
 

What's hot (20)

Form measurement
Form measurementForm measurement
Form measurement
 
gauges, Metrology, Types of gauges
gauges, Metrology, Types of gaugesgauges, Metrology, Types of gauges
gauges, Metrology, Types of gauges
 
Coordinate Measuring Machine(CMM)
Coordinate Measuring Machine(CMM)Coordinate Measuring Machine(CMM)
Coordinate Measuring Machine(CMM)
 
FORM MEASUREMENTS
FORM MEASUREMENTSFORM MEASUREMENTS
FORM MEASUREMENTS
 
Screw thread measurements and Gear measurement
Screw thread measurements and Gear measurementScrew thread measurements and Gear measurement
Screw thread measurements and Gear measurement
 
Introduction to Mechanical Measurements and Metrology
Introduction to Mechanical Measurements and Metrology Introduction to Mechanical Measurements and Metrology
Introduction to Mechanical Measurements and Metrology
 
6 CHAPTER SIX - COMPUTER AIDED INSPECTION.ppt
6 CHAPTER SIX - COMPUTER AIDED INSPECTION.ppt6 CHAPTER SIX - COMPUTER AIDED INSPECTION.ppt
6 CHAPTER SIX - COMPUTER AIDED INSPECTION.ppt
 
interchangeability
interchangeabilityinterchangeability
interchangeability
 
Metrology and measurements(Coordinate Measuring Machine)
Metrology and measurements(Coordinate Measuring Machine)Metrology and measurements(Coordinate Measuring Machine)
Metrology and measurements(Coordinate Measuring Machine)
 
UNIT II LINEAR AND ANGULAR MEASUREMENTS
UNIT II LINEAR AND ANGULAR MEASUREMENTSUNIT II LINEAR AND ANGULAR MEASUREMENTS
UNIT II LINEAR AND ANGULAR MEASUREMENTS
 
Metrology & measurement
Metrology & measurement Metrology & measurement
Metrology & measurement
 
Unit 3
Unit 3Unit 3
Unit 3
 
Coordinate measuring machine
Coordinate measuring machineCoordinate measuring machine
Coordinate measuring machine
 
Computer aided quality control
Computer aided quality controlComputer aided quality control
Computer aided quality control
 
Cnc machine centre
Cnc machine centreCnc machine centre
Cnc machine centre
 
LINEAR and ANGULAR Measurement
LINEAR and ANGULAR MeasurementLINEAR and ANGULAR Measurement
LINEAR and ANGULAR Measurement
 
Gear and Screw measurement
Gear and Screw measurementGear and Screw measurement
Gear and Screw measurement
 
Auto collimator angle dekkor
Auto collimator angle dekkorAuto collimator angle dekkor
Auto collimator angle dekkor
 
Introduction to Mechanical Measurement
Introduction to Mechanical MeasurementIntroduction to Mechanical Measurement
Introduction to Mechanical Measurement
 
Antifriction guideways, feed drives, spindles
Antifriction guideways, feed drives, spindlesAntifriction guideways, feed drives, spindles
Antifriction guideways, feed drives, spindles
 

Similar to Calibration of Coordinate Measuring Machines (CMM)

Uncertainty of Coordinate Measuring Machines
Uncertainty of Coordinate Measuring MachinesUncertainty of Coordinate Measuring Machines
Uncertainty of Coordinate Measuring MachinesHassan Habib
 
Advances in metrology i
Advances in metrology  iAdvances in metrology  i
Advances in metrology iravi8929
 
RAPID PRODUCT DEVELOPMENT seminar
RAPID PRODUCT DEVELOPMENT seminarRAPID PRODUCT DEVELOPMENT seminar
RAPID PRODUCT DEVELOPMENT seminarVijay Mohangekar
 
Metrology and Quality Engineering.pptx
Metrology and Quality Engineering.pptxMetrology and Quality Engineering.pptx
Metrology and Quality Engineering.pptxMandarPadhye5
 
co_ordinate_measuring_machine.pdf
co_ordinate_measuring_machine.pdfco_ordinate_measuring_machine.pdf
co_ordinate_measuring_machine.pdfPradeepKumarb24
 
Module 1 mmm 17 e46b
Module 1 mmm 17 e46bModule 1 mmm 17 e46b
Module 1 mmm 17 e46bD N ROOPA
 
PDM IMPLIMENTATION
PDM IMPLIMENTATIONPDM IMPLIMENTATION
PDM IMPLIMENTATIONUttam Misra
 
Andrew Littlejohn Die Cast Co-op
Andrew Littlejohn Die Cast Co-opAndrew Littlejohn Die Cast Co-op
Andrew Littlejohn Die Cast Co-opAndrew Littlejohn
 
metrolgy and measurements unit 1 introduction
metrolgy and measurements unit 1 introductionmetrolgy and measurements unit 1 introduction
metrolgy and measurements unit 1 introductionntamilselvaneec
 
Unit 1 Basics of Metrology
Unit 1 Basics of MetrologyUnit 1 Basics of Metrology
Unit 1 Basics of Metrologydavid blessley
 
APPLICATION OF MECHATRONICS IN CMM
APPLICATION OF MECHATRONICS IN CMMAPPLICATION OF MECHATRONICS IN CMM
APPLICATION OF MECHATRONICS IN CMMSajid Sheikh
 
Ch-4: Measurement systems and basic concepts of measurement methods
Ch-4: Measurement systems and basic concepts of measurement methodsCh-4: Measurement systems and basic concepts of measurement methods
Ch-4: Measurement systems and basic concepts of measurement methodsSuraj Shukla
 
Calipers presentation mukesh
Calipers presentation mukeshCalipers presentation mukesh
Calipers presentation mukeshmukeshjeswani
 

Similar to Calibration of Coordinate Measuring Machines (CMM) (20)

Uncertainty of Coordinate Measuring Machines
Uncertainty of Coordinate Measuring MachinesUncertainty of Coordinate Measuring Machines
Uncertainty of Coordinate Measuring Machines
 
Advances in metrology i
Advances in metrology  iAdvances in metrology  i
Advances in metrology i
 
RAPID PRODUCT DEVELOPMENT seminar
RAPID PRODUCT DEVELOPMENT seminarRAPID PRODUCT DEVELOPMENT seminar
RAPID PRODUCT DEVELOPMENT seminar
 
Metrology and Quality Engineering.pptx
Metrology and Quality Engineering.pptxMetrology and Quality Engineering.pptx
Metrology and Quality Engineering.pptx
 
co_ordinate_measuring_machine.pdf
co_ordinate_measuring_machine.pdfco_ordinate_measuring_machine.pdf
co_ordinate_measuring_machine.pdf
 
Module 1 mmm 17 e46b
Module 1 mmm 17 e46bModule 1 mmm 17 e46b
Module 1 mmm 17 e46b
 
Meter Testing 101.pptx
Meter Testing 101.pptxMeter Testing 101.pptx
Meter Testing 101.pptx
 
Introduction to metrology
Introduction to metrologyIntroduction to metrology
Introduction to metrology
 
Measurement standards
 Measurement standards Measurement standards
Measurement standards
 
PDM IMPLIMENTATION
PDM IMPLIMENTATIONPDM IMPLIMENTATION
PDM IMPLIMENTATION
 
Andrew Littlejohn Die Cast Co-op
Andrew Littlejohn Die Cast Co-opAndrew Littlejohn Die Cast Co-op
Andrew Littlejohn Die Cast Co-op
 
Meter Testing 101
Meter Testing 101Meter Testing 101
Meter Testing 101
 
ICP QC protocol
ICP  QC  protocolICP  QC  protocol
ICP QC protocol
 
Customer Request Field Meter Testing Programs
Customer Request Field Meter Testing ProgramsCustomer Request Field Meter Testing Programs
Customer Request Field Meter Testing Programs
 
metrolgy and measurements unit 1 introduction
metrolgy and measurements unit 1 introductionmetrolgy and measurements unit 1 introduction
metrolgy and measurements unit 1 introduction
 
Unit 1 Basics of Metrology
Unit 1 Basics of MetrologyUnit 1 Basics of Metrology
Unit 1 Basics of Metrology
 
Meter Testing 101
Meter Testing 101Meter Testing 101
Meter Testing 101
 
APPLICATION OF MECHATRONICS IN CMM
APPLICATION OF MECHATRONICS IN CMMAPPLICATION OF MECHATRONICS IN CMM
APPLICATION OF MECHATRONICS IN CMM
 
Ch-4: Measurement systems and basic concepts of measurement methods
Ch-4: Measurement systems and basic concepts of measurement methodsCh-4: Measurement systems and basic concepts of measurement methods
Ch-4: Measurement systems and basic concepts of measurement methods
 
Calipers presentation mukesh
Calipers presentation mukeshCalipers presentation mukesh
Calipers presentation mukesh
 

More from Hassan Habib

Basics of Dimensional Metrology
Basics of Dimensional MetrologyBasics of Dimensional Metrology
Basics of Dimensional MetrologyHassan Habib
 
Gears Terminologies, Standards, Design, and Inspections
Gears Terminologies, Standards, Design, and InspectionsGears Terminologies, Standards, Design, and Inspections
Gears Terminologies, Standards, Design, and InspectionsHassan Habib
 
Acceptance sampling
Acceptance samplingAcceptance sampling
Acceptance samplingHassan Habib
 
Surface finishing processes - Electroplating
Surface finishing processes - ElectroplatingSurface finishing processes - Electroplating
Surface finishing processes - ElectroplatingHassan Habib
 
Datums and Datum Targets
Datums and Datum TargetsDatums and Datum Targets
Datums and Datum TargetsHassan Habib
 
Geometrical Dimensioning & Tolerancing
Geometrical Dimensioning & TolerancingGeometrical Dimensioning & Tolerancing
Geometrical Dimensioning & TolerancingHassan Habib
 
Geometrical tolerances
Geometrical tolerancesGeometrical tolerances
Geometrical tolerancesHassan Habib
 
Impementing cellular manufacturing
Impementing cellular manufacturingImpementing cellular manufacturing
Impementing cellular manufacturingHassan Habib
 
Simple analysis on ansys
Simple analysis on ansysSimple analysis on ansys
Simple analysis on ansysHassan Habib
 
Terrorism, Eradication, Reasons, Effects
Terrorism, Eradication, Reasons, EffectsTerrorism, Eradication, Reasons, Effects
Terrorism, Eradication, Reasons, EffectsHassan Habib
 

More from Hassan Habib (14)

Basics of Dimensional Metrology
Basics of Dimensional MetrologyBasics of Dimensional Metrology
Basics of Dimensional Metrology
 
Asme y14.5.1
Asme y14.5.1Asme y14.5.1
Asme y14.5.1
 
Gears Terminologies, Standards, Design, and Inspections
Gears Terminologies, Standards, Design, and InspectionsGears Terminologies, Standards, Design, and Inspections
Gears Terminologies, Standards, Design, and Inspections
 
Acceptance sampling
Acceptance samplingAcceptance sampling
Acceptance sampling
 
Surface finishing processes - Electroplating
Surface finishing processes - ElectroplatingSurface finishing processes - Electroplating
Surface finishing processes - Electroplating
 
Datums and Datum Targets
Datums and Datum TargetsDatums and Datum Targets
Datums and Datum Targets
 
Geometrical Dimensioning & Tolerancing
Geometrical Dimensioning & TolerancingGeometrical Dimensioning & Tolerancing
Geometrical Dimensioning & Tolerancing
 
Child labour1
Child labour1Child labour1
Child labour1
 
Risk management
Risk managementRisk management
Risk management
 
Geometrical tolerances
Geometrical tolerancesGeometrical tolerances
Geometrical tolerances
 
Impementing cellular manufacturing
Impementing cellular manufacturingImpementing cellular manufacturing
Impementing cellular manufacturing
 
Simple analysis on ansys
Simple analysis on ansysSimple analysis on ansys
Simple analysis on ansys
 
Terrorism, Eradication, Reasons, Effects
Terrorism, Eradication, Reasons, EffectsTerrorism, Eradication, Reasons, Effects
Terrorism, Eradication, Reasons, Effects
 
Six Sigma
Six SigmaSix Sigma
Six Sigma
 

Recently uploaded

multiple access in wireless communication
multiple access in wireless communicationmultiple access in wireless communication
multiple access in wireless communicationpanditadesh123
 
Cost estimation approach: FP to COCOMO scenario based question
Cost estimation approach: FP to COCOMO scenario based questionCost estimation approach: FP to COCOMO scenario based question
Cost estimation approach: FP to COCOMO scenario based questionSneha Padhiar
 
"Exploring the Essential Functions and Design Considerations of Spillways in ...
"Exploring the Essential Functions and Design Considerations of Spillways in ..."Exploring the Essential Functions and Design Considerations of Spillways in ...
"Exploring the Essential Functions and Design Considerations of Spillways in ...Erbil Polytechnic University
 
CME 397 - SURFACE ENGINEERING - UNIT 1 FULL NOTES
CME 397 - SURFACE ENGINEERING - UNIT 1 FULL NOTESCME 397 - SURFACE ENGINEERING - UNIT 1 FULL NOTES
CME 397 - SURFACE ENGINEERING - UNIT 1 FULL NOTESkarthi keyan
 
CS 3251 Programming in c all unit notes pdf
CS 3251 Programming in c all unit notes pdfCS 3251 Programming in c all unit notes pdf
CS 3251 Programming in c all unit notes pdfBalamuruganV28
 
2022 AWS DNA Hackathon 장애 대응 솔루션 jarvis.
2022 AWS DNA Hackathon 장애 대응 솔루션 jarvis.2022 AWS DNA Hackathon 장애 대응 솔루션 jarvis.
2022 AWS DNA Hackathon 장애 대응 솔루션 jarvis.elesangwon
 
Novel 3D-Printed Soft Linear and Bending Actuators
Novel 3D-Printed Soft Linear and Bending ActuatorsNovel 3D-Printed Soft Linear and Bending Actuators
Novel 3D-Printed Soft Linear and Bending ActuatorsResearcher Researcher
 
Prach: A Feature-Rich Platform Empowering the Autism Community
Prach: A Feature-Rich Platform Empowering the Autism CommunityPrach: A Feature-Rich Platform Empowering the Autism Community
Prach: A Feature-Rich Platform Empowering the Autism Communityprachaibot
 
FUNCTIONAL AND NON FUNCTIONAL REQUIREMENT
FUNCTIONAL AND NON FUNCTIONAL REQUIREMENTFUNCTIONAL AND NON FUNCTIONAL REQUIREMENT
FUNCTIONAL AND NON FUNCTIONAL REQUIREMENTSneha Padhiar
 
Industrial Applications of Centrifugal Compressors
Industrial Applications of Centrifugal CompressorsIndustrial Applications of Centrifugal Compressors
Industrial Applications of Centrifugal CompressorsAlirezaBagherian3
 
Immutable Image-Based Operating Systems - EW2024.pdf
Immutable Image-Based Operating Systems - EW2024.pdfImmutable Image-Based Operating Systems - EW2024.pdf
Immutable Image-Based Operating Systems - EW2024.pdfDrew Moseley
 
11. Properties of Liquid Fuels in Energy Engineering.pdf
11. Properties of Liquid Fuels in Energy Engineering.pdf11. Properties of Liquid Fuels in Energy Engineering.pdf
11. Properties of Liquid Fuels in Energy Engineering.pdfHafizMudaserAhmad
 
Gravity concentration_MI20612MI_________
Gravity concentration_MI20612MI_________Gravity concentration_MI20612MI_________
Gravity concentration_MI20612MI_________Romil Mishra
 
Levelling - Rise and fall - Height of instrument method
Levelling - Rise and fall - Height of instrument methodLevelling - Rise and fall - Height of instrument method
Levelling - Rise and fall - Height of instrument methodManicka Mamallan Andavar
 
Energy Awareness training ppt for manufacturing process.pptx
Energy Awareness training ppt for manufacturing process.pptxEnergy Awareness training ppt for manufacturing process.pptx
Energy Awareness training ppt for manufacturing process.pptxsiddharthjain2303
 
Engineering Drawing section of solid
Engineering Drawing     section of solidEngineering Drawing     section of solid
Engineering Drawing section of solidnamansinghjarodiya
 
KCD Costa Rica 2024 - Nephio para parvulitos
KCD Costa Rica 2024 - Nephio para parvulitosKCD Costa Rica 2024 - Nephio para parvulitos
KCD Costa Rica 2024 - Nephio para parvulitosVictor Morales
 
Module-1-(Building Acoustics) Noise Control (Unit-3). pdf
Module-1-(Building Acoustics) Noise Control (Unit-3). pdfModule-1-(Building Acoustics) Noise Control (Unit-3). pdf
Module-1-(Building Acoustics) Noise Control (Unit-3). pdfManish Kumar
 
Mine Environment II Lab_MI10448MI__________.pptx
Mine Environment II Lab_MI10448MI__________.pptxMine Environment II Lab_MI10448MI__________.pptx
Mine Environment II Lab_MI10448MI__________.pptxRomil Mishra
 

Recently uploaded (20)

multiple access in wireless communication
multiple access in wireless communicationmultiple access in wireless communication
multiple access in wireless communication
 
Cost estimation approach: FP to COCOMO scenario based question
Cost estimation approach: FP to COCOMO scenario based questionCost estimation approach: FP to COCOMO scenario based question
Cost estimation approach: FP to COCOMO scenario based question
 
"Exploring the Essential Functions and Design Considerations of Spillways in ...
"Exploring the Essential Functions and Design Considerations of Spillways in ..."Exploring the Essential Functions and Design Considerations of Spillways in ...
"Exploring the Essential Functions and Design Considerations of Spillways in ...
 
CME 397 - SURFACE ENGINEERING - UNIT 1 FULL NOTES
CME 397 - SURFACE ENGINEERING - UNIT 1 FULL NOTESCME 397 - SURFACE ENGINEERING - UNIT 1 FULL NOTES
CME 397 - SURFACE ENGINEERING - UNIT 1 FULL NOTES
 
CS 3251 Programming in c all unit notes pdf
CS 3251 Programming in c all unit notes pdfCS 3251 Programming in c all unit notes pdf
CS 3251 Programming in c all unit notes pdf
 
2022 AWS DNA Hackathon 장애 대응 솔루션 jarvis.
2022 AWS DNA Hackathon 장애 대응 솔루션 jarvis.2022 AWS DNA Hackathon 장애 대응 솔루션 jarvis.
2022 AWS DNA Hackathon 장애 대응 솔루션 jarvis.
 
Novel 3D-Printed Soft Linear and Bending Actuators
Novel 3D-Printed Soft Linear and Bending ActuatorsNovel 3D-Printed Soft Linear and Bending Actuators
Novel 3D-Printed Soft Linear and Bending Actuators
 
Prach: A Feature-Rich Platform Empowering the Autism Community
Prach: A Feature-Rich Platform Empowering the Autism CommunityPrach: A Feature-Rich Platform Empowering the Autism Community
Prach: A Feature-Rich Platform Empowering the Autism Community
 
FUNCTIONAL AND NON FUNCTIONAL REQUIREMENT
FUNCTIONAL AND NON FUNCTIONAL REQUIREMENTFUNCTIONAL AND NON FUNCTIONAL REQUIREMENT
FUNCTIONAL AND NON FUNCTIONAL REQUIREMENT
 
Industrial Applications of Centrifugal Compressors
Industrial Applications of Centrifugal CompressorsIndustrial Applications of Centrifugal Compressors
Industrial Applications of Centrifugal Compressors
 
Designing pile caps according to ACI 318-19.pptx
Designing pile caps according to ACI 318-19.pptxDesigning pile caps according to ACI 318-19.pptx
Designing pile caps according to ACI 318-19.pptx
 
Immutable Image-Based Operating Systems - EW2024.pdf
Immutable Image-Based Operating Systems - EW2024.pdfImmutable Image-Based Operating Systems - EW2024.pdf
Immutable Image-Based Operating Systems - EW2024.pdf
 
11. Properties of Liquid Fuels in Energy Engineering.pdf
11. Properties of Liquid Fuels in Energy Engineering.pdf11. Properties of Liquid Fuels in Energy Engineering.pdf
11. Properties of Liquid Fuels in Energy Engineering.pdf
 
Gravity concentration_MI20612MI_________
Gravity concentration_MI20612MI_________Gravity concentration_MI20612MI_________
Gravity concentration_MI20612MI_________
 
Levelling - Rise and fall - Height of instrument method
Levelling - Rise and fall - Height of instrument methodLevelling - Rise and fall - Height of instrument method
Levelling - Rise and fall - Height of instrument method
 
Energy Awareness training ppt for manufacturing process.pptx
Energy Awareness training ppt for manufacturing process.pptxEnergy Awareness training ppt for manufacturing process.pptx
Energy Awareness training ppt for manufacturing process.pptx
 
Engineering Drawing section of solid
Engineering Drawing     section of solidEngineering Drawing     section of solid
Engineering Drawing section of solid
 
KCD Costa Rica 2024 - Nephio para parvulitos
KCD Costa Rica 2024 - Nephio para parvulitosKCD Costa Rica 2024 - Nephio para parvulitos
KCD Costa Rica 2024 - Nephio para parvulitos
 
Module-1-(Building Acoustics) Noise Control (Unit-3). pdf
Module-1-(Building Acoustics) Noise Control (Unit-3). pdfModule-1-(Building Acoustics) Noise Control (Unit-3). pdf
Module-1-(Building Acoustics) Noise Control (Unit-3). pdf
 
Mine Environment II Lab_MI10448MI__________.pptx
Mine Environment II Lab_MI10448MI__________.pptxMine Environment II Lab_MI10448MI__________.pptx
Mine Environment II Lab_MI10448MI__________.pptx
 

Calibration of Coordinate Measuring Machines (CMM)

  • 1. Calibration of CMMs Using E&R Test and Laser Interferometer By: Hassan Habib
  • 2. Things we are going to learn • About Coordinate Measuring Machines (CMMs) • Measurements • What is a CMM? • How does a CMM work? • Calibration • What is calibration? • Calibration standards for CMM • Calibration procedure for CMM • E&R Test • Introduction • Definition of errors & Material Standard • Choice of Artifacts • Preliminary Setup • Measurements for error in length • Calculation of results • Probing test error
  • 3. Things we are going to learn • Laser Interferometery • Introduction • Principle of Laser Interferometers • Widely used interferometers in the market • Components of Laser interferometer • Preparing the machine for calibration • Definition of geometrical deviations to be measured • Setting up the laser • Set up the measurements/optics • Collect the data and analyze according to international standards • Compensation of results • Some Post Checks
  • 4.
  • 5. • Beauty is in the perfection of creation. Humans have been creating since their presence on earth • However, accuracy of these measurements has evolved over time and so has the beauty they have created • The earliest examples of accuracy can be found in the construction of Great Pyramids (The difference between height of two opposite corners at its base is 13 mm.) About Coordinate Measuring Machines (CMM) Measurements
  • 6. About Coordinate Measuring Machines (CMM) Measurements Evolution of Measuring Instruments Cubit Micrometer Gauge Blocks Length Comparator Gauges and Dial Indicators Visual Inspection machines Universal Measuring Machines CMM
  • 7. About Coordinate Measuring Machines (CMM) What is a CMM? • Modern machines used for very accurate and precise measurements • CMM works on the principle of Coordinate Measuring i.e. measurement based on collection of data points taken in a Cartesian Coordinate System • “The primary function of a CMM is to measure the actual shape of a work piece, compare it against the desired shape, and evaluate the metrological information such as size, form, location, and orientation.” [Ref: Read ‘Notes’]
  • 8. About Coordinate Measuring Machines (CMM) What is a CMM? • CMM can measure complex geometrical tolerances and deviations on manufactured parts • The accuracy of these machines today are closer to 1µm • CMMs are capable of measuring point to point coordinates and they can also analyze continuous data points using advanced touch probes (SP600 and PH20) • CMMs are extensively used in Aerospace and automotive sectors where increased accuracy of measurement is required Features
  • 9. About Coordinate Measuring Machines (CMM) How does CMM work?
  • 10. About Coordinate Measuring Machines (CMM) How does CMM work? Types of CMMs
  • 11.
  • 12. Calibration What is calibration? • It is the process of verifying and adjusting the accuracy of Measuring & Monitoring Equipment (MME) and machines by comparing them with standards of known accuracy. • The adjustment of instruments is performed by compensating the errors into the instrument. • OEM (Original Equipment Manufacturer) of the instrument usually defines calibration interval, environment of usage and tolerance limits within which the instrument will conform to its performance standard. • It is performed in regular intervals so as to ensure that the instrument is reliable.
  • 13. Calibration What is calibration? • Calibration provides the confidence that their accuracy is as per the given specifications of OEM. • It ensures the repeatability of the measurements taken by the equipment. • The uncertainty is kept at minimum level further building the confidence of measurements. • It is performed in regular intervals so as to ensure that the instrument is reliable. • Calibrating instruments through certified bodies increases the confidence level of customers for your organization. Importance of Calibration
  • 14. Calibration Calibration of CMMs • The calibration of CMM benefits as per the stated benefits of calibration. • For a CMM there can be number of sources of error that will remain undetected thereby nullifying usability. • The calibration of CMM is performed according to the guidelines provided in ISO 10360-2. • National Physics Laboratory (NPL) has also provided standard procedures for verification of performance level of CMMs by detailing guidelines. These guidelines are provided to perform E&R test on the machine using organizational standards used for calibration.
  • 16. Calibration Calibration of CMMs • In the scope of this presentation we are going to study two standards of calibration of CMMs: • E&R test for verification of length measurements • Verification and Compensation of geometric errors with Laser Interferometry
  • 17. Calibration Calibration Procedure for CMMs Choice of Artifacts Preliminary Setup Length Measurement Calculation of results Probing Test Procedure of verification for length measurement through E&R test
  • 18. Calibration Calibration Procedure for CMMs Procedure of verification for geometric deviations through Laser Interferometer
  • 19.
  • 20. E&R Test Introduction • The tests help in demonstrating traceability to national standards and estimating the accuracy of measurements made with three dimensional CMMs for maintaining confidence and reliability in the measurements. • The tests also indicate health of the machine that is necessary to perform maintenance.
  • 21. E&R Test Introduction • Wear of components - the guide ways, the scales, the probe system and the qualification sphere; • Environment in which the CMM operates - the ambient temperature, temperature gradients, humidity and vibration; • The probing strategy used – the magnitude and direction of the probe force, the type of probe stylus used and the measuring speed of the probe; and • Characteristics of the workpiece – elasticity, surface roughness, hardness and the mass of the component Possible sources of error in CMM
  • 22. E&R Test Definition of Errors & material standard • ISO 10360 strongly recommended that the material standard should be either a step gauge, end bar or a series of gauge blocks conforming to ISO 3650 • The material standard of size used for the tests must be calibrated. • The uncertainty of calibration must be taken into consideration and the calibrations must be traceable to the relevant national standard.
  • 23. E&R Test • It is the term that specifies the length measuring accuracy of their CMM. EMPE,L is defined as the extreme value of the error of indication of a CMM for size measurement, permitted by specifications, regulations etc. • It is measured in one of the following ways: • a) EL,MPE = ± minimum of (A + L/K) and B • b) EL, MPE = ± (A + L/K) • c) EL, MPE = ± B where A is a positive constant, expressed in micrometres and supplied by the manufacturer; K is a dimensionless positive constant supplied by the manufacturer; L is the measured size, in millimetres; and B is the maximum permissible error a) b) c) Definition of Errors & material standard
  • 24. E&R Test • The E&R test involves two types of measurement errors. • Volumetric length measuring error E • It applies to all measurements of distances, diameters, and positional tolerances. • Volumetric Probing Error P • It applies to all Form measurements of straightness, flatness, Cylindricity, roundness and free form tolerances. Definition of Errors & material standard
  • 25. E&R Test Choice of artifacts • The choice of artifacts depend on the recommendations from the manufacturer and or the size of you CMM. Various types of artifacts are available for verification and re-verification, and interim check tests, some of them are: • Step Gauge • Length bar • Ball Plate • Hole Plate • Purpose made test piece
  • 26. E&R Test Definition of Errors & material standard Comparison between different artifacts
  • 27. E&R Test Preliminary Setup • Below are some of the pre-requisites for performing the E&R test: • CMM must be operated in accordance with the procedure stated in the instruction manual including machine start up, probe qualification and probe configuration. • Manufacturer supplied test sphere must be used. • Limits for permissible environmental conditions, such as temperature conditions, air humidity and vibration that influence the measurements are usually specified by the manufacturer • Cleaning of stylus tip; • Thermal stability of the probing system • Weight of stylus system and/or probing system; and location, type, number of thermal sensors
  • 28. E&R Test • For the E test a set of 5 length gauges is measured three times in 7 spatial positions. • Total number of measurements: • 3 x 5 x 7 = 105 • 100 % of results must be in the specified limits • The seven spatial positions are: • Along x-axis • Along y-axis • Along z-axis • Along s-partial 1 (Diagonal in XY) • Along s-partial 2 (Diagonal in YZ) • Along s-partial 3 (Diagonal in XZ) • Along s-partial 4 (Diagonal in XYZ) Measurements for error in length E Test Sample positions
  • 29. E&R Test Measurements for error in length E Test Measuring Lines for 7 spatial locations
  • 30. E&R Test Calculation of results • For each of the 105 measurements the error of length measurement, EL is calculated. • Its value is the absolute value of the difference between the indicated value of the relevant test length and the true value of the material standard. • Sample results are shown:
  • 33. E&R Test Calculation of results Graphical representation of the results
  • 34. E&R Test Calculation of results • From the results it can be seen that some of the thirty- five test lengths have values of the error of length measurement • These values will have to be measured again ten times each at the relevant configuration Interpretation of the results
  • 35. E&R Test Probing Test Error • This test of the CMM probing system is used to establish whether the CMM is capable of measuring within the manufacturer‘s stated value of PFTU, MPE by determining the range of values of the radial distance r when measuring a reference sphere. Where, P: associated with the probing system F: apparent Form error T: contact probing (that is to say Tactile) U: single (that is to say Unique) • It is advisable to carry out this test before an acceptance or re-verification test. Introduction
  • 36. E&R Test Probing Test Error • The sphere supplied by the manufacturer for probe qualifying purposes (reference sphere) should not be used for the probing error test. Introduction
  • 37. E&R Test Probing Test Error • The probing error is a positive constant, the value of which is supplied by the CMM manufacturer. • The test sphere should be between 10 mm and 50 mm diameter. • The test sphere should be mounted rigidly to overcome errors due to bending of the mounting stem. • Twenty-five points are measured and recorded. It is a requirement that the points are approximately evenly distributed over at least a hemisphere of the test sphere. • Their position is at the discretion of the user Procedure
  • 38. E&R Test Probing Test Error • one point on the pole (defined by the direction of the stylus shaft) of the test sphere; • four points (equally spaced) 22.5° below the pole; • eight points (equally spaced) 45° below the pole and rotated 22.5° relative to the previous group; • four points (equally spaced) 67.5° below the pole and rotated 22.5° relative to the previous group; and • eight points (equally spaced) 90° below the pole (i.e., on the equator) and rotated 22.5° relative to the previous group. Measurements Measurement Pattern
  • 39. E&R Test Probing Test Error Results
  • 40. E&R Test Probing Test Error Graphical representation of results
  • 41. E&R Test Probing Test Error Interpretation of Results • If the range rmax - rmin of the twenty-five radial distances (PFTU) is no greater than the manufacturer‘s stated value of PFTU, MPE when taking into account the measurement uncertainty, then the performance of the probing system is verified
  • 42.
  • 43. Laser Interferomter Introduction • As stated before laser interferometer is a higher standard of measurement that is used for the calibration of CMM for its geometrical deviations. • The values from this calibration are also used as compensations for the deviations in the machine controller. • Once these compensations are provided to the controller all machine errors are compensated and the machine returns to the factory provided performance standard. • Laser interferometry has slowly evolved into easy to use equipment that can help to perform various tasks. • This type of calibrations all started with Michelson’s Interferometer
  • 44. Laser Interferomter Principle • The working principle of laser interferometers today used, work on the principle of the Michelson’s Interferometer. • The Michelson interferometer is common configuration for optical interferometry and was invented by Albert Abraham Michelson. • Albert Michelson and Edward Morley performed their famous Michelson-Morley experiment in 1887. Edward Morley
  • 45. Laser Interferomter Principle • Using a beam splitter, a light source is split into two arms. • Each of those is reflected back toward the beam splitter which then combines their amplitudes interferometrically. • The resulting interference pattern that is not directed back toward the source is typically directed to some type of photoelectric detector or camera. • Depending on the interferometer's particular application, the two paths may be of different lengths or include optical materials or components under test. Interference Patterns from an interferometer Michelson’s Interferometer
  • 47. Laser Interferomter Principle • M is partially reflective, so part of the light is transmitted through to point B while some is reflected in the direction of A. • Both beams recombine at point C' to produce an interference pattern incident on the detector at point E (or on the retina of a person's eye). • If there is a slight angle between the two returning beams, for instance, then an imaging detector will record a sinusoidal fringe pattern. • If there is perfect spatial alignment between the returning beams, then there will not be any such pattern but rather a constant intensity over the beam dependent on the differential path length. Michelson’s Interferometer
  • 48. Laser Interferomter Principle Modern Laser Interferometer Example of laser measurement
  • 49. Laser Interferomter Widely used interferometers in the market 5529A XL 80
  • 50. Laser Interferomter Components of laser interferometer • Laser Head • Environment compensation unit • Material temperature sensors • Power Supply • Air sensors • Tripod stand with stage • Laptop with necessary software • Optics for different measurements Major Components
  • 51. Laser Interferomter Preparing the machine for calibration Preliminary setup Check the air filters •Including machine filters and air dryer filters Check air tubing for replacement •Wet air would probably require replacement of tubing Check air bearings •They must have specified gap of air cushion Remove rubber pads from base •Removal makes foundation stiff Balance the machine bed •Use inclinometers Note: – You can use the master square to balance the z-axis – Balance the machine on three nodes and remove any redundant rests from the base – There should be no turbulence in air. Turbulence will cause laser error – The environment must be controlled as much as possible.
  • 52. Laser Interferomter Definition of geometrical deviations to be measured Measurements needed for error mapping • Error mapping is done to calculate the 21 geometric deviations that can occur in an articulating machine including CMM and machining centers. • First these are measured and then compensated using the appropriate software to the controller of the machine. • The deviations are majorly of these types: • Linearity (3) • Straightness (6) • Rotation (9) • Squareness (3)
  • 53. Laser Interferomter Definition of geometrical deviations to be measured Measurements needed for error mapping
  • 54. Laser Interferomter Definition of geometrical deviations to be measured Geometrical compensation parameters • Usually the geometric deviations are termed as compensation parameters and are designated values such as: Rxx = Linear straightness in x-axis Ryy = Linear straightness in y-axis Rxy = Horizontal straightness of x-axis Rxz = Vertical straightness of x-axis Dxy = Yaw of x-axis Dxz = Pitch of x-axis . . . . etc.
  • 55. Laser Interferomter Definition of geometrical deviations to be measured Geometrical compensation parameters • For the complete error map these measurements are to be compensated in the controller of the machine. • Software such as Geocomp are used to upload the valued of these deviations into the controller. • Two of these measurements can be taken from inclinometer. These are the straightness of x-axis and y- axis. • Rest of the measurements are taken from a laser interferometer • The linear straightness of the z-axis can be measured by using two Dzy measurements.
  • 56. Laser Interferomter Setting up the laser Generic Procedure • Here we will outline a generic procedure to perform the linear measurement on ML10 laser interferometer of Renishaw. It will give us an overview of the procedure. It is similar to that of Agilent laser interferometer.
  • 57. Laser Interferomter Setting up the laser Generic Procedure • First step is to setup the stage on the tripod stand. Secure and tight it on the stand. • Position the Laser head on the tripod stand with stage. • Arrange the laser interferometer according to the measurement you are about to take. • Switch on the laser. You have to wait for a specified time for the laser to get stabilized.
  • 58. Laser Interferomter Setting up the laser Generic Procedure • Align and fix the optics for the measurement in our case it will be the linear interferometer that includes linear reflector and linear beam splitter as well as clamp blocks to fix it on the machine head. • Now you have to mount the interferometer on the machine bed and reflector has to be attached on machine spindle. Side surfaces of linear beam splitter and reflector have to be exactly parallel. Note: In our case the linear interferometer stays stationary while the splitter moves along the axis.
  • 59. Laser Interferomter Setting up the laser Generic Procedure • The shutter of the laser head can be rotated. Rotate it so that the laser leaving the head has a reduced beam diameter. • Now adjust the tripod so as the spirit level of the tripod is at the central position. • Bring closer the reflector to the laser head and observe a white spot target on the front. Now move the machine in the x-axis until the beam hits the target. Shutter at closed position with no laser emitted
  • 60. Laser Interferomter Setting up the laser Generic Procedure • Now remove white target and check if the beam hits the center of the laser head target on its shutter. If it doesn’t, keep on adjusting the position of the machine until it hits the target at the center. • Now adjust reflector and splitter as close as possible and align them together. • Make sure that the faces are parallel with the one another and with the machine axis.
  • 61. Laser Interferomter Setting up the laser Generic Procedure • Now use a target at the input aperture with white spot at the top and translate machine axis vertically and horizontally so that the beam hits the target. • Now take away the target and check if the returned beam from the interferometer hits the center of the shutter. If it does not repeat the motion of the machine until it does hit the center.
  • 62. Laser Interferomter Set up the measurements/optics Introduction • We have elaborated about the 21 measurements needed for complete calibration of CMM. • Majorly these measurements are divided into three groups: – Linearity – Straightness – Rotation and – Squareness • Now we will elaborate the different configurations of the optics that are used to take these measurements
  • 63. Laser Interferomter Set up the measurements/optics Linearity • Linear measurements are made at multiple points along a machine’s travel path to measure linear displacement and velocity. • It is checked to improve positioning accuracy along an axis for any machine that requires positioning accuracy and velocity control.
  • 64. Laser Interferomter Set up the measurements/optics Optics required • To make linear measurements following optics are required: – Beam Splitter – Linear Reflectors – Targets – Height adjustment fixtures Renishaw XL 80 Agilent 5529A
  • 65. Laser Interferomter Set up the measurements/optics Principle
  • 66. Laser Interferomter Set up the measurements/optics Configuration on machine
  • 67. Laser Interferomter Set up the measurements/optics Straightness • Straightness measurements evaluate the unwanted side to-side or up-and-down motion of a machine tool’s travel in a specified direction.
  • 68. Laser Interferomter Set up the measurements/optics Optics required • To make straightness measurements following optics are required: – Straightness reflector – Straightness interferometer – Targets – Height adjustment fixtures Renishaw XL 80 Agilent 5529A
  • 69. Laser Interferomter Set up the measurements/optics Principle
  • 70. Laser Interferomter Set up the measurements/optics Configuration on machine
  • 71. Laser Interferomter Set up the measurements/optics Configuration on machine
  • 72. Laser Interferomter Set up the measurements/optics Rotation • Angular measurements are made at multiple points along a machine’s travel path to test for rotation about an axis perpendicular to the axis of motion (roll, pitch and yaw) • Unwanted angular motion in a machine tool causes positioning errors that reduce the overall accuracy of your machine.
  • 73. Laser Interferomter Set up the measurements/optics Optics required • To make straightness measurements following optics are required: – Angular reflector – Angular interferometer – Targets – Height adjustment fixtures Renishaw XL 80 Agilent 5529A
  • 74. Laser Interferomter Set up the measurements/optics Principle
  • 75. Laser Interferomter Set up the measurements/optics Configuration on machine
  • 76. Laser Interferomter Set up the measurements/optics Configuration on machine
  • 77. Laser Interferomter Set up the measurements/optics Squareness • Angular measurements are made at multiple points along a machine’s travel path to test for rotation about an axis perpendicular to the axis of motion (roll, pitch and yaw) • Unwanted angular motion in a machine tool causes positioning errors that reduce the overall accuracy of your machine.
  • 78. Laser Interferomter Set up the measurements/optics Optics required • To make squareness measurements following optics are required: – Optical Square – Brackets – Targets – Height adjustment fixtures Renishaw XL 80 Agilent 5529A
  • 79. Laser Interferomter Set up the measurements/optics Principle
  • 80. Laser Interferomter Set up the measurements/optics Configuration on the machine
  • 81. Laser Interferomter Collect the data and analyze according to international standards Parameters to be entered • After the required measurement optics have been setup and is ready to be measured. Usually the following information is entered in the software to begin taking measurements: – Start position for the measurement – End position (Length of the axis of your machine) – Interval (It can be 10, 50, 100) – No. of points – No. of cycles (For bidirectional checks)
  • 82. Laser Interferomter Collect the data and analyze according to international standards Sample screen for entering parameters
  • 83. Laser Interferomter Collect the data and analyze according to international standards Sample for measurement screen
  • 84. Laser Interferomter Collect the data and analyze according to international standards Analyzing data • After the readings have been taken the software shows statistical chart that represents the points that are out of tolerance. • This graphical representation is according to the standard you have chosen. You choose the standard that your OEM has specified for the calibration of the machine.
  • 85. Laser Interferomter Collect the data and analyze according to international standards Sample for analyzing data
  • 86. Laser Interferomter Compensation of results • After the data has been analyzed according to a standard, software provides the compensation table for the measurement you have taken. • This information is then fed into the specific software of the machine used for this purpose. E.g Geocomp is used for compensation for Global Image CMM.
  • 87. Laser Interferomter Compensation of results Example of compensation table
  • 88. Laser Interferomter Some post checks Post Checks • After all the measurements have been compensated, machine is operated in dry run mode. That checks healthy operation. • E&R test elaborated before is also performed to check that all the measurements of the machine are in the tolerance defined by the macnufacturer
  • 89. Calibration of CMMs References • Coordinate Measuring Machines and Systems by Robert J. Hocken & Paulo H. Pereira • Leitz 10360-2 guide • NPL Guidelines – Good Practice Guide No. 42 • Agilent 5529A Manual • Renishaw user guide for XL-80 • Wikipedia

Editor's Notes

  1. http://www.directindustry.com/prod/mora-aeh-metrology/gantry-coordinate-measuring-machine-cmm-compact-30899-141755.html
  2. http://www.directindustry.com/prod/mora-aeh-metrology/gantry-coordinate-measuring-machine-cmm-compact-30899-141755.html
  3. Pictures taken from: - Coordinate Measuring Machines and Systems by Robert J. Hocken & Paulo H. Pereira - http://blog.coghillcartooning.com/5367/great-pyramid-giza-cartoon-sketch/
  4. Pictures taken from: Coordinate Measuring Machines and Systems by Robert J. Hocken & Paulo H. Pereira http://upload.wikimedia.org/wikipedia/commons/b/bc/Coud%C3%A9e-turin.jpg http://www.directindustry.com/prod/brown-sharpe/high-precision-coordinate-measuring-machine-cmm-7157-788233.html
  5. Pictures taken from: http://www.directindustry.com/prod/brown-sharpe/high-precision-coordinate-measuring-machine-cmm-7157-788233.html Coordinate Measuring Machines and Systems by Robert J. Hocken & Paulo H. Pereira
  6. Pictures taken from: http://www.directindustry.com/prod/brown-sharpe/high-precision-coordinate-measuring-machine-cmm-7157-788233.html Coordinate Measuring Machines and Systems by Robert J. Hocken & Paulo H. Pereira
  7. Pictures taken from: Coordinate Measuring Machines and Systems by Robert J. Hocken & Paulo H. Pereira
  8. Pictures taken from: Coordinate Measuring Machines and Systems by Robert J. Hocken & Paulo H. Pereira
  9. Pictures taken from: http://www.sdm-measuring.it/wp-content/uploads/2013/05/SDM-21coor-.jpg
  10. Pictures taken from: - http://www.leedsmicro.com/blog/2015/02/09/iso-17025-how-accreditation-impacts-forensic-laboratories/
  11. Pictures taken from: iso.org NPL guidelines
  12. Picture taken from: - Coordinate Measuring Machines and Systems by Robert J. Hocken & Paulo H. Pereira
  13. Picture taken from: www.usedcmm.com
  14. Picture taken from: - Leitz 10360-2 guide
  15. Pictures taken from: - NPL Guidelines.
  16. Picture courtesy: http://www.sundamachinetools.com/up_files/metal-machining-service/laser-interferometer.jpg
  17. http://en.wikipedia.org/wiki/Michelson_interferometer
  18. http://en.wikipedia.org/wiki/Michelson_interferometer http://laser.physics.sunysb.edu/~kevin/journal/media/fringes.jpg
  19. http://en.wikipedia.org/wiki/Michelson_interferometer http://laser.physics.sunysb.edu/~kevin/journal/media/fringes.jpg
  20. http://en.wikipedia.org/wiki/Michelson_interferometer
  21. Renishaw’s Online instructions for XL Laser System
  22. http://www.fcpablog.com/storage/Agilent-Technologies-logo.jpg?__SQUARESPACE_CACHEVERSION=1412056472387
  23. http://www.fcpablog.com/storage/Agilent-Technologies-logo.jpg?__SQUARESPACE_CACHEVERSION=1412056472387
  24. Picture: Agilent 5529A manual
  25. Picture: Agilent 5529A manual
  26. Picture: Agilent 5529A manual
  27. Picture: http://www.laser-messtechnik.com/messmittel/images/ML10.jpg
  28. Picture: Renishaw laser calibrator system guide
  29. Picture: Renishaw laser calibrator system guide
  30. Picture: Renishaw laser calibrator system guide
  31. Picture: Renishaw laser calibrator system guide
  32. Picture: Renishaw laser calibrator system guide
  33. Picture: http://www.laser-messtechnik.com/messmittel/images/ML10.jpg
  34. Picture: Agilent 5529A manual
  35. Picture: Renishaw laser calibrator system guide Agilent 5529A manual
  36. Picture: Renishaw laser calibrator system guide
  37. Picture: Agilent 5529A manual
  38. Picture: Agilent 5529A manual
  39. Picture: Renishaw laser calibrator system guide Agilent 5529A manual
  40. Picture: Renishaw laser calibrator system guide
  41. Picture: Agilent 5529A manual
  42. Picture: Agilent 5529A manual
  43. Picture: Agilent 5529A manual
  44. Picture: Renishaw laser calibrator system guide Agilent 5529A manual
  45. Picture: Renishaw laser calibrator system guide
  46. Picture: Agilent 5529A manual
  47. Picture: Agilent 5529A manual
  48. Picture: Agilent 5529A manual
  49. Picture: Renishaw laser calibrator system guide Agilent 5529A manual
  50. Picture: Renishaw laser calibrator system guide
  51. Picture: Renishaw laser calibrator system guide
  52. Picture: Renishaw laser calibrator system guide
  53. Picture: Agilent 5529A manual
  54. Picture: Agilent 5529A manual
  55. Picture: Agilent 5529A manual