1. IMPLANTATION OF PACEMAKER TIPS AND
TRICKS ESSENTIAL TESTING OF
PACEMAKER DURING AND AFTER
IMPLANTATION.
EARLY DIAGNOSIS OF PACEMAKER
RELATED PROBLEMS
implantationofpacemakertipsandtricks,essential
testingofpacemaker,earlydiagnosisofpacemaker
complications
2. INTRODUCTION
Currently available permanent pacemakers contain a
pulse generator and one or more pacing leads.
Early in the era of pacemaker implantation, this
procedure was only performed by the cardiac surgeons
because of the initial mandate for epicardial lead
implantation.
Further advancements in the pacing hardware and
percutaneous venous catheterization simplified the
implantation technique and made it feasible to implant
the transvenous leads.
Simultaneously, further innovations in the pulse
generator and its circuitry extended the utility of the
percutaneous technique even in the very young
patients.
implantationofpacemakertipsand
tricks,essentialtestingofpacemaker,early
diagnosisofpacemakercomplications
3. INCIDENCE
Nearly 250 000 new cardiac pacemakers are
implanted annually in the United States, and an
additional 750 000 are implanted worldwide
implantationofpacemakertipsand
tricks,essentialtestingofpacemaker,early
diagnosisofpacemakercomplications
4. HISTORY OF PACING
Pacing and electrophysiology started with Luigi
Galvani (1737-1798) who studied animal electricity.
Alesandro Giuseppe Anastasio Volta(1745-1827)
developed prototype of battery.
Michael Faraday(1791-1867)pioneered in
electrochemistry,named electrodes,electrolytes,
and ions.
Willem Einthoven (1870-1927) – ECG
Arne Larsson- First pacemakerimplantation by
Dr.Senning and engineer Elmqvist.
First implantable pacemaker was implanted by
Lillehei(1960) .
implantationofpacemakertipsand
tricks,essentialtestingofpacemaker,early
diagnosisofpacemakercomplications
5. Barouh Berkovits in 1964 first demand pacemaker.
Doris Escher -1958-first transvenous pacemaker implant
Seymour Furman first transvenous lead
Wilson Greatbatch in 1970 lithium iodine battery
Dual chamber pacing was pioneered in 1970
Rate responsive pacing 1980s
February 2014 during American Heart Month, Vivek
Reddy, MD, of Mount Sinai Heart at The Mount Sinai
Hospital implanted the United States' first miniature-
sized, leadless cardiac pacemaker directly inside a
patient's heart without surgery.
implantationofpacemakertipsand
tricks,essentialtestingofpacemaker,early
diagnosisofpacemakercomplications
7. The LCP is an entirely self-contained intracardiac device that
includes the pacemaker electronics, lithium battery, and electrodes.
The LCP length is 42 mm with maximum diameter of 5.99 mm.
A distal nonretractable, single-tur (screw-in) steroid-eluting
(dexamethasone sodium phosphate) helix affixes the LCP to the
endocardium.
The maximum depth of penetration of the fixation mechanism in tissue
is 1.3 mm.
implantationofpacemakertipsand
tricks,essentialtestingofpacemaker,early
diagnosisofpacemakercomplications
9. Chest x-ray after LCP implant.
X-ray (posterior-anterior view) of the LCP
position, which was performed the day
after implantation.
implantationofpacemakertipsand
tricks,essentialtestingofpacemaker,early
diagnosisofpacemakercomplications
11. LIMITATIONS
The LCP is only a VVIR pacemaker and is not
appropriate for patients requiring dual-chamber sensing
and pacing
There is a possibility of device dislodgment and
migration into the pulmonary vasculature
LCP has a wider diameter than conventional pacing
leads, which raises the possibility of mechanically
induced proarrhythmia.
The LCP system requires an 18F venous introducer
sheath, there is a possibility of vascular complications
Safety profile within the context of cardiac pacemaker
implantation still requires further study
implantationofpacemakertipsand
tricks,essentialtestingofpacemaker,early
diagnosisofpacemakercomplications
12. Future studies will need to address the
safety/efficacy of alternate-site RV pacing (ie, base,
septum, and outflow tract), especially with regard to
minimizing the potential deleterious effects of
chronic RV apical pacing.
implantationofpacemakertipsand
tricks,essentialtestingofpacemaker,early
diagnosisofpacemakercomplications
13. There are other leadless cardiac pacing systems in
development,
they require 2 components – a subcutaneous
energy transmitter (pulse generator) and a receiver
electrode in the cardiac chamber.
These systems use energy delivery sources
ultrasound waves and alternating magnetic fields.
Safety and efficiency these are still under
investigation, and
the potential for interference from external sources
needs further investigation.
implantationofpacemakertipsand
tricks,essentialtestingofpacemaker,early
diagnosisofpacemakercomplications
15. Passive fixation leads distal end contains
extensions liketines,fins,helices and stabilizers
Active fixation leads has distal screw,hook, or helix,
most popular is the extendible –retractable helix.
Leads intended to pace the left ventricle have a
characteristic curve at the distal end and no
conventional right –heart type fixation mechanism.
implantationofpacemakertipsand
tricks,essentialtestingofpacemaker,early
diagnosisofpacemakercomplications
16. Active fixation lead Passive fixation lead
Lead diameter Smaller Larger
Introducer size Smaller Larger
Lead – tissue interface Trauma Atraumatic
Fibrotic ingrowth Slower Faster
Repositioning at
implant
Easy Easy
RV application Less common Very common
Atrial application Very common Very rarely
Proximal manipulation
to secure lead
Yes No
Chronic thresholds Slightly higher Slightly lower
implantationofpacemakertipsand
tricks,essentialtestingofpacemaker,early
diagnosisofpacemakercomplications
17. STEROID ELUTION
Clinical benefits of steroid eluting leads have been
established,but exact mechanism not clear.
When steroid eluting lead is used, the fibrotic
capsule surrounding the lead-tissue interface tends
to be smaller and thinner, but not to the extent that
it would need to be to reduce the threshold by itself.
implantationofpacemakertipsand
tricks,essentialtestingofpacemaker,early
diagnosisofpacemakercomplications
18. ELECTRODE CONFIGURATION
Electrodes are typicallly made from platinum-
irridium, Eigiloy, platinum coated with titanium,
platinum and iridium oxide.
These materials are biologically inert, resist
corrosion, and have excellent conduction
properties.
Unipolar; with one electrode at each end, which as
the cathode and takes the metal outer casing of the
device as the anode.
Bipolar ; with two electrodes at its distal end to form
anode and cathode.
implantationofpacemakertipsand
tricks,essentialtestingofpacemaker,early
diagnosisofpacemakercomplications
19. Bipolar lead Unipolar lead
Pacing artifact Small Large
Pectoral stimulation Almost never Possible
Myopotential
interference
Almost never Possible
Size Larger diameter Smaller diameter
Flexibility Bulkier, stiffer Thinner, more flexible
Reliability record Excellent Near perfect
implantationofpacemakertipsand
tricks,essentialtestingofpacemaker,early
diagnosisofpacemakercomplications
20. Fig.7.9 and 7.10
implantationofpacemakertipsand
tricks,essentialtestingofpacemaker,early
diagnosisofpacemakercomplications
22. PATIENT PREPARATION
informed consent
Routine pre-implant lab tests
patients requiring a pacemaker may be on oral
anticoagulant
standard practice was to discontinue warfarin 48
hours before the procedure, bridge with intravenous
heparin, and then reinitiate warfarin the day of the
procedure or even the night before.
This practice has been associated with higher risk
of hematoma formation compared with that
encountered in unanticoagulated patients (up to
20%)
implantationofpacemakertipsand
tricks,essentialtestingofpacemaker,early
diagnosisofpacemakercomplications
24. Recently, there has been an increasing interest in
performing the pacemaker implantation without
reversal of the anticoagulant.
This practice was associated with lower risk of
pocket bleeding and shorter hospital stay
Pacing Clin Electrophysiol. 2004 Mar;27(3):358-60.Giudici MC1, Paul DL, Bontu
P, Barold SS.
implantationofpacemakertipsand
tricks,essentialtestingofpacemaker,early
diagnosisofpacemakercomplications
25. Antibiotic prophylaxis is a controversial issue, but
most implanters prefer to give oral or intravenous
(IV) antibiotics to decrease the incidence of local or
systemic infections based on limited data available.
Although there is a distinct lack of either national or
international guidance in this area, meta-analysis of
the randomized trials suggests a benefit from pre-
procedure intravenous antibiotics
implantationofpacemakertipsand
tricks,essentialtestingofpacemaker,early
diagnosisofpacemakercomplications
26. INTRAOPERATIVE MEASUREMENTS
Intrinsic signal from the heart should be measured first.
If pacing threshold is tested before the intrinsic signal is
measured , the patient can become pacemaker
dependent, making it impossible to measure sensing.
Sensing is tested by evaluating the signals that would
make an intracardiac electrogram using a device called
a pacing system analyzer(PSA).
The intracardiac signal for the ventricle must be atleast
5mV and ideally between 6 and10mV in order to be
useful
For the atrium, any signal >2mV Is considered desirable.
If signals are inadequate, adjust the leads by mapping
correct position.
implantationofpacemakertipsand
tricks,essentialtestingofpacemaker,early
diagnosisofpacemakercomplications
27. Pacing thresholds in both chambers should be < 1V
Thresholds depends largely on leads placement.
Long-term performance depends on obtaining good
sensing and capture threshold values.
implantationofpacemakertipsand
tricks,essentialtestingofpacemaker,early
diagnosisofpacemakercomplications
29. Battery :
Lithium iodine battery
High energy density ,
Long shelf life ,
Predictable loss of
battery
BOL (vol) – 2.8v
BOL (res) - <1komhs
2.0 – 2.2V with 20,000-30,000Ω
impedence –battery is nearing
depletion
Electrical Testing Of Pacemaker
implantationofpacemakertipsand
tricks,essentialtestingofpacemaker,early
diagnosisofpacemakercomplications
30. Pacing impedance :
Pacing impedance refers to the opposition to current
flow. Two sources contribute to pacing impedance:
1. Pacing lead
2. Electrode - tissue
Tissue of contact
Electrode tip size
Polarization
Normal lead impedance vary from 250-1200ohms.
Single impedance value may be of little use with out
previous values for comparison.
implantationofpacemakertipsand
tricks,essentialtestingofpacemaker,early
diagnosisofpacemakercomplications
31. 1. Pulse generator output circuit
Capture threshold , Pacing threshold , stimulation threshold
Minimum amount of energy required to constantly cause
depolarization
Volts and pulse duration
Electrical Testing Of Pacemaker
implantationofpacemakertipsand
tricks,essentialtestingofpacemaker,early
diagnosisofpacemakercomplications
32. 2.0 v 1.5 v 1 v
Electrical Testing Of Pacemaker
1. Pulse generator output circuit
implantationofpacemakertipsand
tricks,essentialtestingofpacemaker,early
diagnosisofpacemakercomplications
33. Rheobase ; point at
which the plateau
begins and roughly
establishes the
minimum voltage
requirements to
capture the heart.
Chronaxie; the point at
which twice the
rheobase voltage value
meets the curve
implantationofpacemakertipsand
tricks,essentialtestingofpacemaker,early
diagnosisofpacemakercomplications
34. 1. Pulse generator output circuit
Site At
implantation
Acute Chronic
Atrium <1.5mv 3-5 times
threshold
voltage
Twice the
Threshold
voltage
Ventricle <1mv With PW
0.5ms
With PW of
0.5ms
implantationofpacemakertipsand
tricks,essentialtestingofpacemaker,early
diagnosisofpacemakercomplications
35. 1. Pulse generator output circuit
High Pacemaker Output can cause
Reduce longevity
Diaphragmatic stimulation
Muscle Sti. in Unipolar pacemakers
Patient may “feel” heart beat
Algorithm for checking pacemaker output
threshold every beat and maintaining threshold
just above it - Auto capture.
Electrical Testing Of Pacemaker
implantationofpacemakertipsand
tricks,essentialtestingofpacemaker,early
diagnosisofpacemakercomplications
36. Electrical Testing Of Pacemaker
2. Pulse generator sensing circuit :
Ability of the device to detect intrinsic beat of the heart
Measured - peak to peak magnitude (mv) & slew rate(mv/ms)
implantationofpacemakertipsand
tricks,essentialtestingofpacemaker,early
diagnosisofpacemakercomplications
37. 2. Pulse generator sensing circuit :
Electrical Testing Of Pacemaker
Reduce Lower Rate below intrinsic rate to inhibit pacing
and ensure intrinsic activity
Increase sensitivity setting while observing EGM. The
sensitivity value at which sensing is lost on the EGM is
the sensing threshold.
Sensitivity threshold safety is twice the attained valve.
Sensitivity Slew rate
Atrium 1-2mv(0.5mv) > 0.5 v/s
ventricle 2-3mv > 0.75 v/s
implantationofpacemakertipsand
tricks,essentialtestingofpacemaker,early
diagnosisofpacemakercomplications
38. AUTOMATIC OPTIMIZATION OF OTHER PACEMAKER
FUNCTION BASED ON SENSING
These include algorithms to prevent inhibition during
oversensing and loss of pacemaker capture.
Ventricular safety pacing prevents inappropriate pacemaker
inhibition caused by ventricular oversensing of atrial pacing
stimuli.
Safety pacing may be identified on ECGs by noting a shorter
than programmed AV delay, usually 80 to 130 milliseconds.
Noise reversion to fixed-rate asynchronous pacing prevents
pacemaker inhibition during continuous ventricular
oversensing, including that occurring during electromagnetic
interference from sources such as electrocautery.
Automatic assessment of the pacing capture threshold is
performed by closed-loop feedback algorithms that
periodically test capture and adjust the output based on test
results.
implantationofpacemakertipsand
tricks,essentialtestingofpacemaker,early
diagnosisofpacemakercomplications
39. 3. Pulse generator timing circuit :
a. Lower rate limit (LRL)
b. Hysteresis rate
c. Refractory and blanking periods
d. Ventricular safety pacing interval .
e. Upper rate response .
Electrical Testing Of Pacemaker
implantationofpacemakertipsand
tricks,essentialtestingofpacemaker,early
diagnosisofpacemakercomplications
40. Pulse generator timing circuit
Lower rate interval - lowest rate that the pacemaker will
pace .
A paced or non-refractory sensed event restarts the rate
timer at the programmed rate.
implantationofpacemakertipsand
tricks,essentialtestingofpacemaker,early
diagnosisofpacemakercomplications
42. 3. Pulse generator timing circuit :
Hysteresis :
Hysteresis allows the rate to drop below the programmed
pacing LRL.
Advantages of hysteresis :
1. Encourages native rhythm – maintain AV sync in VVI , prolong
battery life
2. Prevent retrograde conduction – avoids pacemaker syndrome
implantationofpacemakertipsand
tricks,essentialtestingofpacemaker,early
diagnosisofpacemakercomplications
43. AV delay (AVI) – pacemaker equivalent of PR interval.
Sensed vs paced AVI – paced AVI is programmed at 125-
200ms , sensed AV interval is programmed at 20-50ms shorter
than paced.
Dynamic AV delay allow pacemaker to respond to exercise
3. Pulse generator timing circuit :
sAVI – 150ms
pAVI – 200ms
implantationofpacemakertipsand
tricks,essentialtestingofpacemaker,early
diagnosisofpacemakercomplications
44. AV delay (AVI)
Longer AVI :
Good AV conduction – maintains AV synchrony , long battery life
Achieved by following methods :Programming longer AVI , managed
ventricular pacing , AV delay hysteresis .
Shorter AVI:
HOCM – RV apical pacing decreases HOCM gradient
CRT – usually 80-120ms , for 100%ventricular pacing and
optimize CO
3. Pulse generator timing circuit :
implantationofpacemakertipsand
tricks,essentialtestingofpacemaker,early
diagnosisofpacemakercomplications
45. Refractory and
blanking periods :
Refractory period
– sensing present
but no action
Blanking period -
sensing absent
and hence no
action
3. Pulse generator timing circuit :
implantationofpacemakertipsand
tricks,essentialtestingofpacemaker,early
diagnosisofpacemakercomplications
46. Blanking periods :3. Pulse generator timing circuit :
Blanking period Time Importance
Atrial blanking
period
50-100ms Non programmable ,
Avoid atrial sensing of its own
paced beat
Post ventricular
atrial blanking
period
220ms Avoid sensing of ventricular beat
Long PVAB decreases detection
of AF,AFL
Ventricular
blanking period
50-100ms Non programmable,
Avoid ventricular sensing of its
paced beat
Post atrial
ventricular
blanking period
28ms if the PAvB period is too long, R
on T - ventricular
tachyarrhythmia.
implantationofpacemakertipsand
tricks,essentialtestingofpacemaker,early
diagnosisofpacemakercomplications
47. Refractory period:
Refractory period Importance
Ventricular refractory period
(VRP)
Prevent sensing of T wave .
Atrial refractory period (ARP) AVI (120-200ms) .
Post ventricular atrial refractory
period
Avoid sensing retrograde P
waves (PMT) , far field R
waves .
3. Pulse generator timing circuit :
implantationofpacemakertipsand
tricks,essentialtestingofpacemaker,early
diagnosisofpacemakercomplications
49. Rate responsive pacing refer to ability of pacemaker to increase its
lower rate in response to physiological stimulus
Sinus node dysfunction .
HRR should start with in 10s of exercise , peak at 90 – 120s and
should return to baseline with in 60 – 120s after exercise.
Fastest rate at which pacemaker will pace upper rate response.
If intrinsic atrial rate exceeds URR then wenckebach or 2:1 AVB
Choosing URR : young patients (150b/mt) , old angina (<110b/mt).
Various sensors (activity , minute ventilation , QT)
4. Pulse generator rate responsive pacing:
Electrical Testing Of Pacemaker
implantationofpacemakertipsand
tricks,essentialtestingofpacemaker,early
diagnosisofpacemakercomplications
53. VVI mode it is the basic single-chamber ventricular
pacing mode; it allows pacing to occur when the
ventricular rate slows below the programmed lower rate
limit..
There is no atrial sensing, so AV synchrony is not
preserved. This mode is indicated for patients with
permanent AF.
AAI mode is the corresponding single-chamber atrial
pacing mode.
It is appropriate for patients with sinus node
dysfunction and normal AV conduction.
Because it does not provide ventricular pacing, it should
not be used in patients at risk for AV block
implantationofpacemakertipsand
tricks,essentialtestingofpacemaker,early
diagnosisofpacemakercomplications
54. DDD pacing mode is most commonly used in the patients
whose rhythm is not permanent AF .
In this mode the atrial rate cannot go lower than the
programmed lower rate.
In the setting of AV block, all ventricular events are paced. A
special characteristic of the DDD pacing mode is the ability to
“track” intrinsic atrial activity to maintain AV synchrony.
The DDD mode has an upper rate limit, the maximum rate that
intrinsic atrial activity will be tracked.
The maximum rate is selected to exceed the maximum sinus
rate that the patient is capable of achieving.
The upper rate limit is predominantly of importance to prevent
tracking of rapid atrial activity in spontaneous atrial
arrhythmias such as AF.
implantationofpacemakertipsand
tricks,essentialtestingofpacemaker,early
diagnosisofpacemakercomplications
55. Automatic Mode Switching
Automatic mode switching in the DDD pacing mode
initiates a temporary change in mode to a nontracking
one (usually DDI or DDIR) during paroxysmal atrial
tachyarrhythmias.
This prevents the adverse consequences of rapid
ventricular pacing as a result of tracking nonphysiologic
high atrial rates.
Most mode-switching algorithms use the atrial rate as an
indicator for the onset of an atrial tachyarrhythmia.
When the atrial rhythm again meets the defined criteria
for a physiologic rhythm, the mode switches back to an
atrial tracking mode
implantationofpacemakertipsand
tricks,essentialtestingofpacemaker,early
diagnosisofpacemakercomplications
57. Pacemaker follow up guidelines:
NASPE guidelines
Single or dual pacing
1st visit 6 – 8 week post implant , if symptomatic prior
to this
5th month
From 6th month q 3month
Battery wear present q 1month
implantationofpacemakertipsand
tricks,essentialtestingofpacemaker,early
diagnosisofpacemakercomplications
60. Pocket hematoma :
The risk of haematoma is increased in patients taking
antithrombotic or anticoagulant drugs (Goldstein et al., 1998).
Most small hematomas can be managed conservatively with
cold compress and withdrawal of antiplatelet or antithrombotic
agents.
Occasionally, large hematomas that compromise the suture
line or skin integrity may have to be surgically evacuated.
Needle aspiration increases risk of infection and should not
be done.
Pacemaker complications
implantationofpacemakertipsand
tricks,essentialtestingofpacemaker,early
diagnosisofpacemakercomplications
61. PACEMAKER COMPLICATIONS
Pocket hematoma :
In patients requiring oral anticoagulants (warfarin), to take INR
of about 2.0 at the time of implantation is safe (Belott &
Reynolds, 2000).
Unfractionated heparin or low-molecular-weight heparin are
always discontinued prior to device implant and ideally avoided
for a minimum of 24 hours post implantation.
Administration of anticoagulants can be resumed within 48-72
h after implantation if there is no evidence of substantial
hematoma formation.
implantationofpacemakertipsand
tricks,essentialtestingofpacemaker,early
diagnosisofpacemakercomplications
62. Device-related infections :
The reported incidence of pacemaker-related infection ranges
from 0.5% to 6% in early series
The use of prophylactic antibiotics and pocket irrigation with
antibiotic solutions has decreased the rate of acute infections
following pacemaker implantations to <1 to 2 percent in most
series
The mortality of persistent infection when infected leads are not
removed can be as high as 66%.
DM, malignancy, operator inexperience, advanced age,
corticosteroid use, anticoagulation, recent device manipulation,
CRF, and bacteremia from a distant focus of infection.
Pacemaker complications
implantationofpacemakertipsand
tricks,essentialtestingofpacemaker,early
diagnosisofpacemakercomplications
63. Device related infection :
Device infection is defined as either:
(a) deep infection - infection involving the generator pocket
and/or the intravenous portion of the leads, with bacteremia,
requiring device extraction or
(b) superficial infection - characterized by local
inflammation, involving the skin but not the generator pocket,
and treated with oral antibiotics.
Pacemaker complications
implantationofpacemakertipsand
tricks,essentialtestingofpacemaker,early
diagnosisofpacemakercomplications
64. Pacemaker complications
Device related infection :
2007;49;1851-1859 J. Am. Coll. Cardiol.
implantationofpacemakertipsand
tricks,essentialtestingofpacemaker,early
diagnosisofpacemakercomplications
65. Device related infection :
2007;49;1851-1859 J. Am. Coll. C
implantationofpacemakertipsand
tricks,essentialtestingofpacemaker,early
diagnosisofpacemakercomplications
66. Pacemaker complications
Device related infection :
2007;49;1851-1859 J. Am. Coll. C
implantationofpacemakertipsand
tricks,essentialtestingofpacemaker,early
diagnosisofpacemakercomplications
67. Wound pain :
Infection , Pacemaker implanted too superficially ,
Pacemaker implanted too laterally , Pacemaker allergy .
Skin erosion :
Incidence has been estimated around 0.8% .Old age ,
infection.
Surgical revision of pocket and reimplantation .
Allergic reactions :
Always rule out infection before coming to diagnosis of allergy
Pacemaker complications
implantationofpacemakertipsand
tricks,essentialtestingofpacemaker,early
diagnosisofpacemakercomplications
68. Lead dislodgement:
Relatively common – 5-10%
of patients(ICD database
2001)
Atrial more common than
ventricular(2-3% vs. 1%)
Micro dislodgement , macro
dislodgement
Increased pacing threshold ,
failure to pace and sense
Active fixation (decreases risk)
Pacemaker complications
implantationofpacemakertipsand
tricks,essentialtestingofpacemaker,early
diagnosisofpacemakercomplications
69. Pneumothorax , :
Uncommon complication – 1.6-
2.6%
During or 48 hrs after procedure
Inadvent puncture and laceration
of subclavian vein , artery or lung
Related to operator experience
and underlying anatomy
Avoided by
1. Venogram – flouroscpic puncture
2. Axillary venous access (Martin
etal’96)
implantationofpacemakertipsand
tricks,essentialtestingofpacemaker,early
diagnosisofpacemakercomplications
70. Cardiac Perforation :
Uncommon but potentially serious complication - lower
than 1%.
Acute (<5 days) ,
subacute(5d-1month) ,
chronic (>1month)
Increasing stimulation threshold , RBBB pattern for RV
pacing, intercostal muscle or diaphragmatic contraction,
friction rub, and pericarditis, pericardial effusion, or
cardiac tamponade.
Pacemaker complications
implantationofpacemakertipsand
tricks,essentialtestingofpacemaker,early
diagnosisofpacemakercomplications
71. CARDIAC
PERFORATION
CXR ,
ECHO ,
CT
Management;
Lead withdrawal and
repositioning.
Surgical back up
implantationofpacemakertipsand
tricks,essentialtestingofpacemaker,early
diagnosisofpacemakercomplications
72. Extracardiac stimulation
The diaphragm or pectoral or intercostal muscles
Diaphragmatic stimulation - direct stimulation of the diaphragm
(left) or stimulation of the phrenic nerve (right).
Early postimplantation period , dislodgment of the pacing lead.
MC in patients with LV coronary vein branch lead placement
for CRT
Output pacing importance (testing and treatment)
Pectoral stimulation - incorrect orientation of the pacemaker or
a current leak from a lead insulation failure or exposed
connector.
Pacemaker complications
implantationofpacemakertipsand
tricks,essentialtestingofpacemaker,early
diagnosisofpacemakercomplications
73. Venous thrombosis :
Venous thrombosis occurs in 30% to 50% of patients and only
1-3% of patients become symptomatic.
Manifestations vary from usually asymptomatic, acute
symptomatic thrombosis, and even SVCS .
Early or late after pacemaker implantation.
Predictors of severe stenosis are multiple pacemaker leads,
previous pacing , double coils , hormone therapy .
Asymptomatic (no treatment) , symptomatic (anticoagulants –
endovascular stents – surgical correction).
Pacemaker complications
implantationofpacemakertipsand
tricks,essentialtestingofpacemaker,early
diagnosisofpacemakercomplications
74. Twiddler syndrome:
Pacemaker complications
Obese women with loose, fatty subcutaneous tissue
Small size of the implanted generator with a large pocket
Twisting of pulse generator in long axis
Lead dislodgement and lead fracture
Failure to capture
implantationofpacemakertipsand
tricks,essentialtestingofpacemaker,early
diagnosisofpacemakercomplications
75. The prevelance of this
syndrome is 0.07% (Gungor et
al., 2009)
Rotated along the transverse
axis it is referred by us as the
reel syndrome.
Pocket should be revised.
Avoid by
Limit the pocket size,
Suture the device to the fascia
The patients not to manipulate
their device pocket
Twiddler syndrome:
Pacemaker complications
implantationofpacemakertipsand
tricks,essentialtestingofpacemaker,early
diagnosisofpacemakercomplications
76. Failure to capture
Failure to output
Sensing abnormalities(under and over
sensing)
Specific mode complications
1. Pacemaker related tachycardia
2. Pacemaker syndrome
Pacemaker malfunction
implantationofpacemakertipsand
tricks,essentialtestingofpacemaker,early
diagnosisofpacemakercomplications
77. Failure to capture:
Pacing artifact present but no evoked potential .
Causes
1. Lead dislodgement or perforation
2. Lead maturation(inflammation/fibrosis)(exit block)
3. Battery depletion
4. Circuit failure(coil fracture , insulation defect)
5. Capture management algorithm failure
6. Inappropriate programming
7. Pseudo malfunction
8. Functional non capture
9. Metabolic , drugs , cardiomyopathies
Pacemaker malfunction
implantationofpacemakertipsand
tricks,essentialtestingofpacemaker,early
diagnosisofpacemakercomplications
78. Electrocardiographic tracing from a patient with a DDDR pacemaker. All ventricular
pacing artifacts but one failed to result in ventricular depolarization—
that is, failure to capture
Failure to capture:
Pacemaker malfunction
implantationofpacemakertipsand
tricks,essentialtestingofpacemaker,early
diagnosisofpacemakercomplications
79. Failure to capture:
Pacing threshold
Normal
Increased
Battery depletion
Functional non capture
Impedance
Normal
Dislodgement
Exit block
Decreased
Insulation
failure/break
Increased
Lead fracture
Loose screw
implantationofpacemakertipsand
tricks,essentialtestingofpacemaker,early
diagnosisofpacemakercomplications
80. Failure to output:
Absence of pacing stimuli and hence no capture .
Causes
1. Pseudo malfunction - hysteresis , PMT termination ,
sleep rate
2. Over sensing - EMI ; T P R over sensing ;
Myopotential/diaphragmatic ; Cross talk ; Make break
signals
3. Open circuit - lead fracture , loose screw , air in the
pocket , incompatible lead .
4. Battery depletion
5. Recording artifact.
Pacemaker malfunction
implantationofpacemakertipsand
tricks,essentialtestingofpacemaker,early
diagnosisofpacemakercomplications
81. Failure to output:
Pacemaker malfunction
VVIR pacemaker This patient had a pacemaker programmed to
a unipolar sensing configuration. The sensing of myopotentials
led to symptomatic pauses, and reprogramming the pacemaker
to a bipolar sensing configuration prevented subsequent
myopotential over sensing.
implantationofpacemakertipsand
tricks,essentialtestingofpacemaker,early
diagnosisofpacemakercomplications
82. Application of magnet
Failure to output:
Eliminates pauses
Pauses persistent
Impedance
Normal
Decreased
Insulation
failure/break
Increased
Lead fracture
Loose screw
Battery
depletion
Over sensing
Pseudo malfunction
implantationofpacemakertipsand
tricks,essentialtestingofpacemaker,early
diagnosisofpacemakercomplications
83. Battery depletion :
Elective replacement indicators (ERI)
1. Low voltage(2.1-2.4)
2. Low pacing rate on magnet application
3. Elevated battery impedance
4. Increased pulse width duration
5. Restricted programmability
6. Change to simpler pacing mode
End of life (EOL)
1. Low voltage(≤2.1vol)
Pacemaker malfunction
implantationofpacemakertipsand
tricks,essentialtestingofpacemaker,early
diagnosisofpacemakercomplications
84. Pacemaker undersensing :
Pacing artifact present but no sensing(sensed beat
doesn’t reset cycle)
Causes are
1. Defect in signal production – scar /fibrosis , BBB , ectopic ,
cardioversion , defibrillation , metabolic.
2. Defect in signal transmission – lead fracture/dislodgement,
insulation failure , partial open circuit.
3. Defect in pacemaker – battery depletion , sensing circuit
abnormalities , committed DVI.
Pacemaker malfunction
implantationofpacemakertipsand
tricks,essentialtestingofpacemaker,early
diagnosisofpacemakercomplications
86. Pacemaker over
sensing :
Present as failure to pace
Causes
1. EMI
2. T , P , R over sensing .
3. Cross talk
4. Myopotential (unipolar)
5. Make break signals
Pacemaker malfunction
Cross talk :
High atrial output
High ventricular sensitivity
Low VBP
Ventricular sensing of
paced atrial impulse
Pts with Poor AV conduction –
Ventricular Asystole
implantationofpacemakertipsand
tricks,essentialtestingofpacemaker,early
diagnosisofpacemakercomplications
87. Electromagnetic interference :
Pacemaker malfunction
Source Pacer
damage
Inhibition Rate
increase
Asynchronou
s noise
Uni/
bipola
r
Cardioversion/
Defibrillation
Y N N N U/B
Anti theft devices /
Weapon detector
N Y N N U
Phone
(cell/cordless)
N Y Y Y U/B
Ablation Y Y Y N U/B
Diathermy/
lithotripsy
Y Y Y Y U/B
FM radio
TV transmitter
N Y N Y U
MRI/PET Y Y(N) Y(N) Y(N) U/B
implantationofpacemakertipsand
tricks,essentialtestingofpacemaker,early
diagnosisofpacemakercomplications
88. Pacemaker syndrome :
Seen in 20% of PPI (5%
severe symptomatic)
VVI/DDD/AAI
Pulsations in neck , fatigue
, cough ,chest fullness ,
headache , chocking
sensation , PND, confusion
, syncope , pulmonary
edema.
Rx : VVI – program
hysteresis , or change to
DDD ; DDD –atrial lead
reprogrammed or changed
Pacemaker malfunction
implantationofpacemakertipsand
tricks,essentialtestingofpacemaker,early
diagnosisofpacemakercomplications