SlideShare a Scribd company logo
1 of 83
Download to read offline
Laboratory Method Verification
Ola H. Elgaddar
MD, PhD, MBA, CPHQ, LSSGB,
Lecturer of Chemical Pathology
Alexandria University
Ola.elgaddar@alexu.edu.eg
Quality
Doing the right thing right, from the
first time and every time!
Accreditation
An institution or a program meets
standards of quality set forth by an
accrediting agency
Validation / Verification??
Method validation:(Manufacturer concern)
Establishing the performance of a new
diagnostic tool.
Confirmation, through the provision of
objective evidence, that the requirements for
a specific intended use or application have
been fulfilled’ (doing correct test)......
ISO 9001:2005
Validation / Verification??
Method verification: (Lab / user concern)
A process to determine performance
characteristics before a test system is utilized
for patient testing.
Confirmation, through the provision of
objective evidence, that specified
requirements have been fulfilled’ (doing test
correctly)……ISO 9001:2005
According to Westgard
Ø The inner, hidden, deeper, secret meaning
of method validation is error assessment.
ØHow much error might be present in the
test result within your laboratory ?
ØCould this degree of error affect the
interpretation and possibly patient care ?
If the potential error is large enough to lead to
misinterpretation, then the method is not
acceptable.
Ø Random error, RE, or imprecision is
described as an error that can be either
positive or negative, whose direction and
exact magnitude cannot be predicted, where
the distribution of results when replicate
measurements are made on a single
specimen.
ØUsually, due to error in Pipetting
Ø Systematic error, SE, or inaccuracy is an
error that is always in one direction,
displacing the mean of the distribution from
its original value.
ØIn contrast to random errors, systematic
errors are in one direction and cause all the
test results to be either high or low.
ØEither constant or proportionate
ØUsually, due to error in calibration
Internal Quality Control (IQC) is used, on daily
basis, in the decision to accept or reject
results of patients samples and enables the
lab to describe and monitor the quality of its
work.
-Usually it has 2 levels (Sometimes 3);
representing the “Normal” and the
“Pathological” analyte level.
- Judged according to Westgard Multi-QC
rules
External Quality Control (EQC) =
Proficiency test is used, on monthly or Bi-
weekly (Or others) basis, where labs from all
over the world join the program and send their
used Method / Analyzer.
- A statistical comparison is made and each
lab result is compared to the result of its peer
group in each analyte.
The following items need verification
ØAnalytical Specificity: Interference studies
ØAnalytical Sensitivity: Calibration curve
Detection limit
ØReportable range: Linearity experiment
ØPrecision: Replication study
ØAccuracy: Bias / Recovery study
ØReference Intervals
Analytical Specificity
The ability of an analytical method to detect
“ONLY” the analyte of interest.
Freedom from interference by any element
or compound other than the analyte of
interest
Analytical Specificity
- Analytical Specificity is verified using
interference studies.
- A validated method, known to be free of
the interfering substance is used. A series
of samples containing increased
concentrations of the interfering substance
are analyzed using that method, and the
method under study, then both results are
compared
Some Automated systems have a “HIL” index
Analytical Sensitivity
- The ability of an analytical method to
detect a low concentration of a given
substance in a biological sample. The lower
the detectable concentration, the greater
the analytical sensitivity.
- Detection limits studies
Analytical Sensitivity
OR,
-The ability of an analytical method to
detect (respond to) a change in
concentration of the analyte. The smaller
the detectable change (the change in
concentration that can result in a definite
change in the reported signal), the greater
is the analytical sensitivity.
- Slope of the Calibration Curve
REMEMBER!!
STANDARD
• It is a solution of known concentration.
• Formed by dissolving a known amount of
an analyte in a specific volume of an
aqueous solvent.
• Ex: Dissolving 100 mg glucose in 100 ml
D.W gives a standard of a 100 mg / dl
concentration
CALIBRATOR
• A solution or a device of known
quantitative or qualitative characteristics
(eg: concentration, activity, intensity)
• Used to calibrate or adjust a measurement
procedure.
• Matrix is preserved.
• The calibrator material is reconstituted &
introduced to the analyzer before using a
new lot of reagent, and its concentration is
assigned.
• The calibrator is treated like samples and
the absorbance of the developed color is
determined.
Using the provided concentration, the analyzer
constructs a calibration curve
0
0.1
0.2
0.3
0.4
0.5
0.6
0.7
0.8
0.9
0 50 100 150 200 250
• If the given concentration is 100 mg / dl,
the absorbance of this point is determined.
• By providing the analyzer with the range
where this method is linear, the calibration
curve is extended so that any sample
concentration lying within that linearity
range can be deduced from this curve
Detection Limits
1.Lower Limit of Detection (LLD)= Limit of
the blank (LOB)
2.Biologic Limit of Detection (BLD)= Limit of
Detection (LOD)
3.Functional Sensitivity (FS)= Limit of
Quantification (LOQ)
Detection Limits
1.Lower Limit of Detection (LLD)= Limit
of the blank (LOB)
Most analytical instruments produce a signal even
when a blank (reagent without analyte) is
analyzed. This signal is referred to as the noise
level. The LLD is the analyte concentration that is
required to produce a signal greater than two (or
three) times the standard deviation of the noise
level.
Detection Limits
1.Lower Limit of Detection (LLD)= Limit
of the blank (LOB)
Limit of Detection (LLD) is estimated as, the mean
of the blank sample plus 2 or 3 times the SD
obtained on the blank sample:
LLD = mean(blk) + Zs(blk)
where the Z-value is usually 2 or 3
Detection Limits
2.Biologic Limit of Detection (BLD)= Limit
of Detection (LOD)
-It is the limit of blank, after the addition of the
analyte of interest (Its lowest concentration)
- BLD is estimated as the LLD plus 2 or 3 times the
standard deviation obtained from the "spiked"
sample
BLD = LLD+ Zs(spk)
where the Z-value is usually 2 or 3
- Results between BLD & LLD should be reported
without quantitation
Detection Limits
3.Functional Sensitivity (FS)= Limit of
Quantification (LOQ)
-The lowest concentration of target compounds
that can be quantified confidently, that meets some
pre-specified targets of imprecision, commonly
CV=20%
- Several spiked concentrations must be studied to
determine the precision profile at the low
concentration range and to select the lowest
concentration at which a 20% CV is obtained.
To Sum Up!
1.LoB is the highest apparent analyte
concentration expected to be found when
replicates of a blank sample containing no
analyte are tested.
2.LoD is the lowest analyte concentration likely to
be reliably distinguished from the LoB and at
which detection is feasible.
3.LoQ is the lowest concentration at which the
analyte can not only be reliably detected, but at
which some predefined goals are met. The LoQ
may be equivalent to the LoD or it could be at
a much higher concentration.
Reportable Range
=
Analytical Measurement Range
It is the range of numeric results a method
can produce without any special specimen
pre-treatment, such as dilution.
Ø It should be verified, for the manufacturer’s
claim, using the linearity experiment.
Ø It is performed using either calibrators,
proficiency samples, or samples
Ø Serial dilutions will be made covering the
whole analytical measurement range, and
reaching as close as possible to the claimed
values of the manufacturer.
ØEach dilution is to be processed in duplicate
to remove the element of imprecision.
Ø The observed measures (on the X-axis) are
plotted against the expected measures (on
the Y-axis), and a line point to point graph is
constructed for each analyte.
ØThe line is judged visually for its linearity
and according to each experiment, the
analytical measurement range of each
method is verified, where any patient result
obtained in the future, outside the verified
range, cannot be released without further
processing (Dilution or concentration).
Dilution Expected conc Observed conc
D1 8.27 8.27
D2 4.56 4.7
D4 2.5 2.6
D6 1.5 1.52
D7 1 1.07
D8 0.5 0.68
Correlation: r 0.999886844
Analytical range: fT4 : 0.1 - 12 ng / dl
0
1
2
3
4
5
6
7
8
9
0 2 4 6 8 10
Observed measure
Expectedmeasure
Precision
Closeness of agreement between quantity
values obtained by replicate measurements
of a quantity, under specified conditions.
Precision
ØPrecision should be assessed using
quality control material (A minimum of 2
levels), or pooled serum (Of minimum two
concentration levels)
ØEach level of the QC material / pooled
serum is measured 5 times per day (Within
run), for 5 days (in between runs)
Precision
ØThe measures obtained from this precision
study are to be collected, and the mean, SD,
and CV are calculated for each parameter,
for the used QC levels / pooled serum.
ØThe obtained CVs, are statistically
compared to the manufacturer’s claim using
ANOVA test of significance, to determine if
there is a significant different between the
obtained CVs (and their verification
intervals), and the manufacturer’s claim at a
certain CI; usually 95%
Accuracy
Closeness of the agreement between the
result of a measurement and a true value of
the measurand.
Trueness
Closeness of the agreement between the
replicates of result of a measurement, and
a true value of the measurand.
ØThe difference between the mean of
replicates of a measurement, and its true
value is the BIAS.
Ø The CLSI calculation of bias is based on
the results of 7 – 11 PT samples; each is
measured in duplicate, and then compared
to the true value (Peer’s mean) using
student T-test.
A concept!!!
A concept!!!
A concept!!!
Why is Z-score used in PT results??
Bias is verified for the tested method when
there is no significant difference between:
q Mean Z-score of PT results (7 – 11) is not
significantly different from Zero
q Mean Z-score of PT results (7 – 11) is not
significantly different from peer’s mean
Reference intervals
ØRemember that we are just “Verifying” the
reference intervals stated by the
manufacturer or published in the literature,
and “transferring” them to the lab using the
method under study.
Ø “Establishment” of reference intervals is
another issue.
Reference intervals
ØAcceptability of the transfer shall be
assessed by examining 20 reference
individuals, from our subject population, and
comparing the obtained test results to those
of the manufacturer/ Literature.
ØThose 20 individuals should be selected in
such a way that will satisfy the exclusion and
partitioning criteria.
ØThe test results should be examined to
make sure that none of the results appears
to be an outlier.
Reference intervals
ØThe manufacturer's / Literature reference
intervals are considered verified if no more
than two of the 20 tested subjects' values (or
10% of the test results) fall outside those
ranges.
Reference intervals
ØExclusion / partitioning criteria include:
age, sex, fasting status, disease history,
drug history, previous surgeries, and time of
the cycle / pregnancy for females.
Method Comparison
Ø According to CLSI, at least 40 samples
should be assayed on both methods under
examination (two field methods), or between
one tested method and a reference method.
Ø Several statistical approaches can be
used, one of them is to calculate the
correlation coefficient “r”
Ø“r” should be more than or equal 0.95
Total Error
It is the summation of both Random and
Systematic error.
It is calculated as follow:
TE = Bias (%) + 2 CV
It is compared to Biological Variation (or any
other specifications) for Total Allowable
Error
Uncertainty
Ø It is an interval around a reported
laboratory result that specifies the location of
the true value with a given probability
Ø It takes into consideration both the
imprecision (SD), and the inaccuracy (Bias)
Ø It is calculated from the data of 6 month
minimum
What performance characteristics
are usually validated?
ØReportable range (Linearity)
ØPrecision (or imprecision)
ØAccuracy (or inaccuracy, bias)
ØReference interval
Laboratory Method Verification, March 2017
Laboratory Method Verification, March 2017

More Related Content

What's hot

What's hot (20)

Basics of laboratory internal quality control, Ola Elgaddar, 2012
Basics of laboratory internal quality control, Ola Elgaddar, 2012Basics of laboratory internal quality control, Ola Elgaddar, 2012
Basics of laboratory internal quality control, Ola Elgaddar, 2012
 
Laboratory Internal Quality Control presentation master revision, 2014
Laboratory Internal Quality Control presentation master revision, 2014Laboratory Internal Quality Control presentation master revision, 2014
Laboratory Internal Quality Control presentation master revision, 2014
 
Quality control in clinical laboratories
Quality control in clinical laboratoriesQuality control in clinical laboratories
Quality control in clinical laboratories
 
Method verification
Method verificationMethod verification
Method verification
 
Medical Laboratory Quality Assurance
Medical Laboratory Quality AssuranceMedical Laboratory Quality Assurance
Medical Laboratory Quality Assurance
 
Quality assurance in a medical laboratory
Quality assurance in a medical laboratoryQuality assurance in a medical laboratory
Quality assurance in a medical laboratory
 
External Quality Control Lecture MD General 2014 Course, Clin Path Ain Shams ...
External Quality Control Lecture MD General 2014 Course, Clin Path Ain Shams ...External Quality Control Lecture MD General 2014 Course, Clin Path Ain Shams ...
External Quality Control Lecture MD General 2014 Course, Clin Path Ain Shams ...
 
Quality assurance is necessary in blood banking
Quality assurance is necessary in blood bankingQuality assurance is necessary in blood banking
Quality assurance is necessary in blood banking
 
Laboratory audit ls
Laboratory audit lsLaboratory audit ls
Laboratory audit ls
 
Method validation
Method validationMethod validation
Method validation
 
Method validation
Method validationMethod validation
Method validation
 
Internal quality control
Internal quality controlInternal quality control
Internal quality control
 
Quality Control in Clinical Chemistry
Quality Control in Clinical ChemistryQuality Control in Clinical Chemistry
Quality Control in Clinical Chemistry
 
Internal quality control (IQC) in coagulation lab
Internal quality control (IQC) in coagulation labInternal quality control (IQC) in coagulation lab
Internal quality control (IQC) in coagulation lab
 
2. Quality Control Notes
2. Quality Control Notes2. Quality Control Notes
2. Quality Control Notes
 
Haemoglobin quality control by maintaining levey jennings chart
Haemoglobin quality control by maintaining levey jennings chartHaemoglobin quality control by maintaining levey jennings chart
Haemoglobin quality control by maintaining levey jennings chart
 
Quality control lecture CPath master 2014 Ain Shams
Quality control lecture CPath master 2014 Ain ShamsQuality control lecture CPath master 2014 Ain Shams
Quality control lecture CPath master 2014 Ain Shams
 
Quality assurance in blood banking
Quality assurance in blood bankingQuality assurance in blood banking
Quality assurance in blood banking
 
Quality assurance
Quality assuranceQuality assurance
Quality assurance
 
Quality assurance in medical laboratory
Quality assurance in medical laboratoryQuality assurance in medical laboratory
Quality assurance in medical laboratory
 

Viewers also liked

Viewers also liked (20)

OLAP for Big Data (Druid vs Apache Kylin vs Apache Lens)
OLAP for Big Data (Druid vs Apache Kylin vs Apache Lens)OLAP for Big Data (Druid vs Apache Kylin vs Apache Lens)
OLAP for Big Data (Druid vs Apache Kylin vs Apache Lens)
 
Edital embasa n°_1_2017_concurso_público
Edital embasa n°_1_2017_concurso_públicoEdital embasa n°_1_2017_concurso_público
Edital embasa n°_1_2017_concurso_público
 
Introduccion a la Agronomia
Introduccion a la AgronomiaIntroduccion a la Agronomia
Introduccion a la Agronomia
 
Study pre-registration: Benefits and considerations
Study pre-registration: Benefits and considerationsStudy pre-registration: Benefits and considerations
Study pre-registration: Benefits and considerations
 
3Com 3C905CX-TX-M-19
3Com 3C905CX-TX-M-193Com 3C905CX-TX-M-19
3Com 3C905CX-TX-M-19
 
Meninges y Liquido Cefalorraquideo
Meninges y Liquido CefalorraquideoMeninges y Liquido Cefalorraquideo
Meninges y Liquido Cefalorraquideo
 
Presentacion roraima
Presentacion roraimaPresentacion roraima
Presentacion roraima
 
Improving the Quality and Value of the Laboratory Self-Inspection - Best of A...
Improving the Quality and Value of the Laboratory Self-Inspection - Best of A...Improving the Quality and Value of the Laboratory Self-Inspection - Best of A...
Improving the Quality and Value of the Laboratory Self-Inspection - Best of A...
 
Greek philosophy
Greek philosophyGreek philosophy
Greek philosophy
 
Marco de referencia
Marco de referenciaMarco de referencia
Marco de referencia
 
災害時における情報支援
災害時における情報支援災害時における情報支援
災害時における情報支援
 
Fatal Accidente en la vía Guayllabamba-Cusubamaba
Fatal Accidente en la vía Guayllabamba-CusubamabaFatal Accidente en la vía Guayllabamba-Cusubamaba
Fatal Accidente en la vía Guayllabamba-Cusubamaba
 
Maestriadiapositiva
MaestriadiapositivaMaestriadiapositiva
Maestriadiapositiva
 
Nietzsche, genio y figura
Nietzsche, genio y figuraNietzsche, genio y figura
Nietzsche, genio y figura
 
Ejemplo de actividades en Ardora
Ejemplo de actividades en ArdoraEjemplo de actividades en Ardora
Ejemplo de actividades en Ardora
 
Harmonization of Laboratory Indicators, 09 03-2017
Harmonization of Laboratory Indicators, 09 03-2017Harmonization of Laboratory Indicators, 09 03-2017
Harmonization of Laboratory Indicators, 09 03-2017
 
ELISA tips and tricks - March 2016
ELISA tips and tricks - March 2016ELISA tips and tricks - March 2016
ELISA tips and tricks - March 2016
 
What do clinicians need to know about lab tests?
What do clinicians need to know about lab tests?What do clinicians need to know about lab tests?
What do clinicians need to know about lab tests?
 
Standardization of Thyroid Function Tests, 05 05 - 2017
Standardization of Thyroid Function Tests, 05   05 - 2017Standardization of Thyroid Function Tests, 05   05 - 2017
Standardization of Thyroid Function Tests, 05 05 - 2017
 
Strategic planning, Ola Elgaddar, 12 12-2016
Strategic planning, Ola Elgaddar, 12 12-2016Strategic planning, Ola Elgaddar, 12 12-2016
Strategic planning, Ola Elgaddar, 12 12-2016
 

Similar to Laboratory Method Verification, March 2017

Related Substances-Method Validation-PPT_slide
Related Substances-Method Validation-PPT_slideRelated Substances-Method Validation-PPT_slide
Related Substances-Method Validation-PPT_slide
Bhanu Prakash N
 
Assay-Method validation-PPT _slide
Assay-Method validation-PPT _slideAssay-Method validation-PPT _slide
Assay-Method validation-PPT _slide
Bhanu Prakash N
 
International conference on harmonisation validation of analytical procedures
International conference on harmonisation validation of analytical proceduresInternational conference on harmonisation validation of analytical procedures
International conference on harmonisation validation of analytical procedures
noushin javidvand
 

Similar to Laboratory Method Verification, March 2017 (20)

Related Substances-Method Validation-PPT_slide
Related Substances-Method Validation-PPT_slideRelated Substances-Method Validation-PPT_slide
Related Substances-Method Validation-PPT_slide
 
Validation my ppt
Validation my pptValidation my ppt
Validation my ppt
 
Analytical Method Validation
Analytical Method Validation Analytical Method Validation
Analytical Method Validation
 
Control of analytical quality using stable control materials postgrad
Control of analytical quality using stable control materials postgradControl of analytical quality using stable control materials postgrad
Control of analytical quality using stable control materials postgrad
 
analytical method validation and validation of hplc
analytical method validation and validation of hplcanalytical method validation and validation of hplc
analytical method validation and validation of hplc
 
Analytical Method Validation basics by Dr. A. Amsavel
Analytical Method Validation  basics by Dr. A. AmsavelAnalytical Method Validation  basics by Dr. A. Amsavel
Analytical Method Validation basics by Dr. A. Amsavel
 
Assay-Method validation-PPT _slide
Assay-Method validation-PPT _slideAssay-Method validation-PPT _slide
Assay-Method validation-PPT _slide
 
B.S 4- Class 1-Introduction to analytical chemistry
B.S 4- Class 1-Introduction to analytical chemistryB.S 4- Class 1-Introduction to analytical chemistry
B.S 4- Class 1-Introduction to analytical chemistry
 
Quality control.pptx
Quality control.pptxQuality control.pptx
Quality control.pptx
 
What is Nephelometry,and a fully automated Nephlometer analyzer for protein a...
What is Nephelometry,and a fully automated Nephlometer analyzer for protein a...What is Nephelometry,and a fully automated Nephlometer analyzer for protein a...
What is Nephelometry,and a fully automated Nephlometer analyzer for protein a...
 
Analytical Method Validations & Detailed Method Validation Parameters
Analytical Method Validations & Detailed Method Validation ParametersAnalytical Method Validations & Detailed Method Validation Parameters
Analytical Method Validations & Detailed Method Validation Parameters
 
Analytical method validation
Analytical method validationAnalytical method validation
Analytical method validation
 
Quality assurance part_2
Quality assurance part_2Quality assurance part_2
Quality assurance part_2
 
ANALYTICAL METHOD VALIDATION (PHARMACEUTICAL INDUSTRY).pdf
ANALYTICAL METHOD VALIDATION (PHARMACEUTICAL INDUSTRY).pdfANALYTICAL METHOD VALIDATION (PHARMACEUTICAL INDUSTRY).pdf
ANALYTICAL METHOD VALIDATION (PHARMACEUTICAL INDUSTRY).pdf
 
(IQC) - Internal Quality Control.pptx
(IQC) - Internal Quality Control.pptx(IQC) - Internal Quality Control.pptx
(IQC) - Internal Quality Control.pptx
 
Analytical Method Validation
Analytical Method ValidationAnalytical Method Validation
Analytical Method Validation
 
Method validation terms quality control and assurance
Method validation terms quality control and assurance Method validation terms quality control and assurance
Method validation terms quality control and assurance
 
Pharmacy Slideshow XL by Slidesgo.pptx
Pharmacy Slideshow XL by Slidesgo.pptxPharmacy Slideshow XL by Slidesgo.pptx
Pharmacy Slideshow XL by Slidesgo.pptx
 
HPLC validation.ppt
HPLC validation.pptHPLC validation.ppt
HPLC validation.ppt
 
International conference on harmonisation validation of analytical procedures
International conference on harmonisation validation of analytical proceduresInternational conference on harmonisation validation of analytical procedures
International conference on harmonisation validation of analytical procedures
 

More from Ola Elgaddar

More from Ola Elgaddar (20)

Enhancing Laboratory Leadership through Financial Management Skills.pptx
Enhancing Laboratory Leadership through Financial Management Skills.pptxEnhancing Laboratory Leadership through Financial Management Skills.pptx
Enhancing Laboratory Leadership through Financial Management Skills.pptx
 
Financial Tools, Basics for lab managers.pptx
Financial Tools, Basics for lab managers.pptxFinancial Tools, Basics for lab managers.pptx
Financial Tools, Basics for lab managers.pptx
 
Population Health Basic concepts - Ola Elgaddar .pdf
Population Health Basic concepts - Ola Elgaddar .pdfPopulation Health Basic concepts - Ola Elgaddar .pdf
Population Health Basic concepts - Ola Elgaddar .pdf
 
Strategies for Improving Laboratory Utilization
Strategies for Improving Laboratory Utilization Strategies for Improving Laboratory Utilization
Strategies for Improving Laboratory Utilization
 
Harmonization of Laboratory Testing, 08 04-2017
Harmonization of Laboratory Testing, 08 04-2017Harmonization of Laboratory Testing, 08 04-2017
Harmonization of Laboratory Testing, 08 04-2017
 
Are we using the correct quality goals?
Are we using the correct quality goals?Are we using the correct quality goals?
Are we using the correct quality goals?
 
Healthcare Governance and Patient Safety, Ola, 03 07-2014
Healthcare Governance and Patient Safety, Ola, 03 07-2014Healthcare Governance and Patient Safety, Ola, 03 07-2014
Healthcare Governance and Patient Safety, Ola, 03 07-2014
 
Laboratory diagnosis of H. Pylori infection, Ola Elgaddar
Laboratory diagnosis of H. Pylori infection, Ola ElgaddarLaboratory diagnosis of H. Pylori infection, Ola Elgaddar
Laboratory diagnosis of H. Pylori infection, Ola Elgaddar
 
4) Signaling, Ola Elgaddar
4) Signaling, Ola Elgaddar4) Signaling, Ola Elgaddar
4) Signaling, Ola Elgaddar
 
3) Cell Receptors, Ola Elgaddar
3) Cell Receptors, Ola Elgaddar3) Cell Receptors, Ola Elgaddar
3) Cell Receptors, Ola Elgaddar
 
Transport across cell membranes
Transport across cell membranesTransport across cell membranes
Transport across cell membranes
 
Cell structure, Ola Elgaddar
Cell structure, Ola ElgaddarCell structure, Ola Elgaddar
Cell structure, Ola Elgaddar
 
Medical Laboratories in Egypt, Ola Elgaddar, 30-12-2013
Medical Laboratories in Egypt, Ola Elgaddar, 30-12-2013Medical Laboratories in Egypt, Ola Elgaddar, 30-12-2013
Medical Laboratories in Egypt, Ola Elgaddar, 30-12-2013
 
Calcium, Ola Elgaddar, 25 11- 2013
Calcium, Ola Elgaddar, 25 11- 2013Calcium, Ola Elgaddar, 25 11- 2013
Calcium, Ola Elgaddar, 25 11- 2013
 
Body water and Osmolality
Body water and OsmolalityBody water and Osmolality
Body water and Osmolality
 
Laboratory diagnosis of muscle diseases
Laboratory diagnosis of muscle diseasesLaboratory diagnosis of muscle diseases
Laboratory diagnosis of muscle diseases
 
Pediatrics clinical chemistry
Pediatrics clinical chemistryPediatrics clinical chemistry
Pediatrics clinical chemistry
 
Concepts in molecular biology
Concepts in molecular biologyConcepts in molecular biology
Concepts in molecular biology
 
Study of the role of estrogen receptor alpha gene polymorphism in prostate ca...
Study of the role of estrogen receptor alpha gene polymorphism in prostate ca...Study of the role of estrogen receptor alpha gene polymorphism in prostate ca...
Study of the role of estrogen receptor alpha gene polymorphism in prostate ca...
 
Study of vascular endothelial growth factor (vegf) and bcl 2 protein levels ...
Study of vascular endothelial growth factor (vegf) and bcl 2 protein levels  ...Study of vascular endothelial growth factor (vegf) and bcl 2 protein levels  ...
Study of vascular endothelial growth factor (vegf) and bcl 2 protein levels ...
 

Recently uploaded

Ernakulam Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real Meet
Ernakulam Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real MeetErnakulam Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real Meet
Ernakulam Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real Meet
Call Girls Chandigarh
 
Nanded Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real Meet
Nanded Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real MeetNanded Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real Meet
Nanded Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real Meet
Call Girls Service
 
Chandigarh Escorts, 😋9988299661 😋50% off at Escort Service in Chandigarh
Chandigarh Escorts, 😋9988299661 😋50% off at Escort Service in ChandigarhChandigarh Escorts, 😋9988299661 😋50% off at Escort Service in Chandigarh
Chandigarh Escorts, 😋9988299661 😋50% off at Escort Service in Chandigarh
Sheetaleventcompany
 
Call Girls Service Anantapur 📲 6297143586 Book Now VIP Call Girls in Anantapur
Call Girls Service Anantapur 📲 6297143586 Book Now VIP Call Girls in AnantapurCall Girls Service Anantapur 📲 6297143586 Book Now VIP Call Girls in Anantapur
Call Girls Service Anantapur 📲 6297143586 Book Now VIP Call Girls in Anantapur
gragmanisha42
 
9316020077📞Goa Call Girls Numbers, Call Girls Whatsapp Numbers Goa
9316020077📞Goa  Call Girls  Numbers, Call Girls  Whatsapp Numbers Goa9316020077📞Goa  Call Girls  Numbers, Call Girls  Whatsapp Numbers Goa
9316020077📞Goa Call Girls Numbers, Call Girls Whatsapp Numbers Goa
russian goa call girl and escorts service
 
VIP Call Girls Noida Sia 9711199171 High Class Call Girl Near Me
VIP Call Girls Noida Sia 9711199171 High Class Call Girl Near MeVIP Call Girls Noida Sia 9711199171 High Class Call Girl Near Me
VIP Call Girls Noida Sia 9711199171 High Class Call Girl Near Me
mriyagarg453
 
VIP Call Girls Noida Jhanvi 9711199171 Best VIP Call Girls Near Me
VIP Call Girls Noida Jhanvi 9711199171 Best VIP Call Girls Near MeVIP Call Girls Noida Jhanvi 9711199171 Best VIP Call Girls Near Me
VIP Call Girls Noida Jhanvi 9711199171 Best VIP Call Girls Near Me
mriyagarg453
 
ooty Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real Meet
ooty Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real Meetooty Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real Meet
ooty Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real Meet
Call Girls Service
 
Muzaffarpur Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real Meet
Muzaffarpur Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real MeetMuzaffarpur Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real Meet
Muzaffarpur Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real Meet
Call Girls Service
 
Call Girls Service Faridabad 📲 9999965857 ヅ10k NiGhT Call Girls In Faridabad
Call Girls Service Faridabad 📲 9999965857 ヅ10k NiGhT Call Girls In FaridabadCall Girls Service Faridabad 📲 9999965857 ヅ10k NiGhT Call Girls In Faridabad
Call Girls Service Faridabad 📲 9999965857 ヅ10k NiGhT Call Girls In Faridabad
gragmanisha42
 
Bareilly Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real Meet
Bareilly Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real MeetBareilly Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real Meet
Bareilly Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real Meet
Call Girls Service
 

Recently uploaded (20)

Ernakulam Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real Meet
Ernakulam Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real MeetErnakulam Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real Meet
Ernakulam Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real Meet
 
Nanded Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real Meet
Nanded Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real MeetNanded Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real Meet
Nanded Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real Meet
 
Enjoyment ★ 8854095900 Indian Call Girls In Dehradun 🍆🍌 By Dehradun Call Girl ★
Enjoyment ★ 8854095900 Indian Call Girls In Dehradun 🍆🍌 By Dehradun Call Girl ★Enjoyment ★ 8854095900 Indian Call Girls In Dehradun 🍆🍌 By Dehradun Call Girl ★
Enjoyment ★ 8854095900 Indian Call Girls In Dehradun 🍆🍌 By Dehradun Call Girl ★
 
Chandigarh Escorts, 😋9988299661 😋50% off at Escort Service in Chandigarh
Chandigarh Escorts, 😋9988299661 😋50% off at Escort Service in ChandigarhChandigarh Escorts, 😋9988299661 😋50% off at Escort Service in Chandigarh
Chandigarh Escorts, 😋9988299661 😋50% off at Escort Service in Chandigarh
 
Call Girls Service Anantapur 📲 6297143586 Book Now VIP Call Girls in Anantapur
Call Girls Service Anantapur 📲 6297143586 Book Now VIP Call Girls in AnantapurCall Girls Service Anantapur 📲 6297143586 Book Now VIP Call Girls in Anantapur
Call Girls Service Anantapur 📲 6297143586 Book Now VIP Call Girls in Anantapur
 
Call Girls Hyderabad Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Hyderabad Just Call 9907093804 Top Class Call Girl Service AvailableCall Girls Hyderabad Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Hyderabad Just Call 9907093804 Top Class Call Girl Service Available
 
9316020077📞Goa Call Girls Numbers, Call Girls Whatsapp Numbers Goa
9316020077📞Goa  Call Girls  Numbers, Call Girls  Whatsapp Numbers Goa9316020077📞Goa  Call Girls  Numbers, Call Girls  Whatsapp Numbers Goa
9316020077📞Goa Call Girls Numbers, Call Girls Whatsapp Numbers Goa
 
Call Now ☎ 9999965857 !! Call Girls in Hauz Khas Escort Service Delhi N.C.R.
Call Now ☎ 9999965857 !! Call Girls in Hauz Khas Escort Service Delhi N.C.R.Call Now ☎ 9999965857 !! Call Girls in Hauz Khas Escort Service Delhi N.C.R.
Call Now ☎ 9999965857 !! Call Girls in Hauz Khas Escort Service Delhi N.C.R.
 
(Sonam Bajaj) Call Girl in Jaipur- 09257276172 Escorts Service 50% Off with C...
(Sonam Bajaj) Call Girl in Jaipur- 09257276172 Escorts Service 50% Off with C...(Sonam Bajaj) Call Girl in Jaipur- 09257276172 Escorts Service 50% Off with C...
(Sonam Bajaj) Call Girl in Jaipur- 09257276172 Escorts Service 50% Off with C...
 
VIP Call Girls Noida Sia 9711199171 High Class Call Girl Near Me
VIP Call Girls Noida Sia 9711199171 High Class Call Girl Near MeVIP Call Girls Noida Sia 9711199171 High Class Call Girl Near Me
VIP Call Girls Noida Sia 9711199171 High Class Call Girl Near Me
 
VIP Call Girls Noida Jhanvi 9711199171 Best VIP Call Girls Near Me
VIP Call Girls Noida Jhanvi 9711199171 Best VIP Call Girls Near MeVIP Call Girls Noida Jhanvi 9711199171 Best VIP Call Girls Near Me
VIP Call Girls Noida Jhanvi 9711199171 Best VIP Call Girls Near Me
 
(Deeksha) 💓 9920725232 💓High Profile Call Girls Navi Mumbai You Can Get The S...
(Deeksha) 💓 9920725232 💓High Profile Call Girls Navi Mumbai You Can Get The S...(Deeksha) 💓 9920725232 💓High Profile Call Girls Navi Mumbai You Can Get The S...
(Deeksha) 💓 9920725232 💓High Profile Call Girls Navi Mumbai You Can Get The S...
 
❤️♀️@ Jaipur Call Girl Agency ❤️♀️@ Manjeet Russian Call Girls Service in Jai...
❤️♀️@ Jaipur Call Girl Agency ❤️♀️@ Manjeet Russian Call Girls Service in Jai...❤️♀️@ Jaipur Call Girl Agency ❤️♀️@ Manjeet Russian Call Girls Service in Jai...
❤️♀️@ Jaipur Call Girl Agency ❤️♀️@ Manjeet Russian Call Girls Service in Jai...
 
ooty Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real Meet
ooty Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real Meetooty Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real Meet
ooty Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real Meet
 
Muzaffarpur Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real Meet
Muzaffarpur Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real MeetMuzaffarpur Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real Meet
Muzaffarpur Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real Meet
 
Call Girls Thane Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Thane Just Call 9907093804 Top Class Call Girl Service AvailableCall Girls Thane Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Thane Just Call 9907093804 Top Class Call Girl Service Available
 
Call Girls Service Faridabad 📲 9999965857 ヅ10k NiGhT Call Girls In Faridabad
Call Girls Service Faridabad 📲 9999965857 ヅ10k NiGhT Call Girls In FaridabadCall Girls Service Faridabad 📲 9999965857 ヅ10k NiGhT Call Girls In Faridabad
Call Girls Service Faridabad 📲 9999965857 ヅ10k NiGhT Call Girls In Faridabad
 
Bareilly Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real Meet
Bareilly Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real MeetBareilly Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real Meet
Bareilly Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real Meet
 
❤️♀️@ Jaipur Call Girls ❤️♀️@ Jaispreet Call Girl Services in Jaipur QRYPCF ...
❤️♀️@ Jaipur Call Girls ❤️♀️@ Jaispreet Call Girl Services in Jaipur QRYPCF  ...❤️♀️@ Jaipur Call Girls ❤️♀️@ Jaispreet Call Girl Services in Jaipur QRYPCF  ...
❤️♀️@ Jaipur Call Girls ❤️♀️@ Jaispreet Call Girl Services in Jaipur QRYPCF ...
 
Call Girls Patiala Just Call 8250077686 Top Class Call Girl Service Available
Call Girls Patiala Just Call 8250077686 Top Class Call Girl Service AvailableCall Girls Patiala Just Call 8250077686 Top Class Call Girl Service Available
Call Girls Patiala Just Call 8250077686 Top Class Call Girl Service Available
 

Laboratory Method Verification, March 2017

  • 1. Laboratory Method Verification Ola H. Elgaddar MD, PhD, MBA, CPHQ, LSSGB, Lecturer of Chemical Pathology Alexandria University Ola.elgaddar@alexu.edu.eg
  • 2. Quality Doing the right thing right, from the first time and every time!
  • 3. Accreditation An institution or a program meets standards of quality set forth by an accrediting agency
  • 4.
  • 5. Validation / Verification?? Method validation:(Manufacturer concern) Establishing the performance of a new diagnostic tool. Confirmation, through the provision of objective evidence, that the requirements for a specific intended use or application have been fulfilled’ (doing correct test)...... ISO 9001:2005
  • 6. Validation / Verification?? Method verification: (Lab / user concern) A process to determine performance characteristics before a test system is utilized for patient testing. Confirmation, through the provision of objective evidence, that specified requirements have been fulfilled’ (doing test correctly)……ISO 9001:2005
  • 7.
  • 8. According to Westgard Ø The inner, hidden, deeper, secret meaning of method validation is error assessment. ØHow much error might be present in the test result within your laboratory ? ØCould this degree of error affect the interpretation and possibly patient care ? If the potential error is large enough to lead to misinterpretation, then the method is not acceptable.
  • 9.
  • 10. Ø Random error, RE, or imprecision is described as an error that can be either positive or negative, whose direction and exact magnitude cannot be predicted, where the distribution of results when replicate measurements are made on a single specimen. ØUsually, due to error in Pipetting
  • 11.
  • 12. Ø Systematic error, SE, or inaccuracy is an error that is always in one direction, displacing the mean of the distribution from its original value. ØIn contrast to random errors, systematic errors are in one direction and cause all the test results to be either high or low. ØEither constant or proportionate ØUsually, due to error in calibration
  • 13.
  • 14.
  • 15.
  • 16. Internal Quality Control (IQC) is used, on daily basis, in the decision to accept or reject results of patients samples and enables the lab to describe and monitor the quality of its work. -Usually it has 2 levels (Sometimes 3); representing the “Normal” and the “Pathological” analyte level. - Judged according to Westgard Multi-QC rules
  • 17. External Quality Control (EQC) = Proficiency test is used, on monthly or Bi- weekly (Or others) basis, where labs from all over the world join the program and send their used Method / Analyzer. - A statistical comparison is made and each lab result is compared to the result of its peer group in each analyte.
  • 18.
  • 19. The following items need verification ØAnalytical Specificity: Interference studies ØAnalytical Sensitivity: Calibration curve Detection limit ØReportable range: Linearity experiment ØPrecision: Replication study ØAccuracy: Bias / Recovery study ØReference Intervals
  • 20.
  • 21. Analytical Specificity The ability of an analytical method to detect “ONLY” the analyte of interest. Freedom from interference by any element or compound other than the analyte of interest
  • 22.
  • 23. Analytical Specificity - Analytical Specificity is verified using interference studies. - A validated method, known to be free of the interfering substance is used. A series of samples containing increased concentrations of the interfering substance are analyzed using that method, and the method under study, then both results are compared
  • 24.
  • 25. Some Automated systems have a “HIL” index
  • 26.
  • 27. Analytical Sensitivity - The ability of an analytical method to detect a low concentration of a given substance in a biological sample. The lower the detectable concentration, the greater the analytical sensitivity. - Detection limits studies
  • 28. Analytical Sensitivity OR, -The ability of an analytical method to detect (respond to) a change in concentration of the analyte. The smaller the detectable change (the change in concentration that can result in a definite change in the reported signal), the greater is the analytical sensitivity. - Slope of the Calibration Curve
  • 29.
  • 31. STANDARD • It is a solution of known concentration. • Formed by dissolving a known amount of an analyte in a specific volume of an aqueous solvent. • Ex: Dissolving 100 mg glucose in 100 ml D.W gives a standard of a 100 mg / dl concentration
  • 32. CALIBRATOR • A solution or a device of known quantitative or qualitative characteristics (eg: concentration, activity, intensity) • Used to calibrate or adjust a measurement procedure. • Matrix is preserved.
  • 33. • The calibrator material is reconstituted & introduced to the analyzer before using a new lot of reagent, and its concentration is assigned. • The calibrator is treated like samples and the absorbance of the developed color is determined.
  • 34. Using the provided concentration, the analyzer constructs a calibration curve 0 0.1 0.2 0.3 0.4 0.5 0.6 0.7 0.8 0.9 0 50 100 150 200 250
  • 35. • If the given concentration is 100 mg / dl, the absorbance of this point is determined. • By providing the analyzer with the range where this method is linear, the calibration curve is extended so that any sample concentration lying within that linearity range can be deduced from this curve
  • 36. Detection Limits 1.Lower Limit of Detection (LLD)= Limit of the blank (LOB) 2.Biologic Limit of Detection (BLD)= Limit of Detection (LOD) 3.Functional Sensitivity (FS)= Limit of Quantification (LOQ)
  • 37.
  • 38. Detection Limits 1.Lower Limit of Detection (LLD)= Limit of the blank (LOB) Most analytical instruments produce a signal even when a blank (reagent without analyte) is analyzed. This signal is referred to as the noise level. The LLD is the analyte concentration that is required to produce a signal greater than two (or three) times the standard deviation of the noise level.
  • 39. Detection Limits 1.Lower Limit of Detection (LLD)= Limit of the blank (LOB) Limit of Detection (LLD) is estimated as, the mean of the blank sample plus 2 or 3 times the SD obtained on the blank sample: LLD = mean(blk) + Zs(blk) where the Z-value is usually 2 or 3
  • 40. Detection Limits 2.Biologic Limit of Detection (BLD)= Limit of Detection (LOD) -It is the limit of blank, after the addition of the analyte of interest (Its lowest concentration) - BLD is estimated as the LLD plus 2 or 3 times the standard deviation obtained from the "spiked" sample BLD = LLD+ Zs(spk) where the Z-value is usually 2 or 3 - Results between BLD & LLD should be reported without quantitation
  • 41. Detection Limits 3.Functional Sensitivity (FS)= Limit of Quantification (LOQ) -The lowest concentration of target compounds that can be quantified confidently, that meets some pre-specified targets of imprecision, commonly CV=20% - Several spiked concentrations must be studied to determine the precision profile at the low concentration range and to select the lowest concentration at which a 20% CV is obtained.
  • 42.
  • 43. To Sum Up! 1.LoB is the highest apparent analyte concentration expected to be found when replicates of a blank sample containing no analyte are tested. 2.LoD is the lowest analyte concentration likely to be reliably distinguished from the LoB and at which detection is feasible. 3.LoQ is the lowest concentration at which the analyte can not only be reliably detected, but at which some predefined goals are met. The LoQ may be equivalent to the LoD or it could be at a much higher concentration.
  • 44.
  • 45. Reportable Range = Analytical Measurement Range It is the range of numeric results a method can produce without any special specimen pre-treatment, such as dilution.
  • 46. Ø It should be verified, for the manufacturer’s claim, using the linearity experiment. Ø It is performed using either calibrators, proficiency samples, or samples Ø Serial dilutions will be made covering the whole analytical measurement range, and reaching as close as possible to the claimed values of the manufacturer. ØEach dilution is to be processed in duplicate to remove the element of imprecision.
  • 47. Ø The observed measures (on the X-axis) are plotted against the expected measures (on the Y-axis), and a line point to point graph is constructed for each analyte. ØThe line is judged visually for its linearity and according to each experiment, the analytical measurement range of each method is verified, where any patient result obtained in the future, outside the verified range, cannot be released without further processing (Dilution or concentration).
  • 48. Dilution Expected conc Observed conc D1 8.27 8.27 D2 4.56 4.7 D4 2.5 2.6 D6 1.5 1.52 D7 1 1.07 D8 0.5 0.68 Correlation: r 0.999886844 Analytical range: fT4 : 0.1 - 12 ng / dl
  • 49. 0 1 2 3 4 5 6 7 8 9 0 2 4 6 8 10 Observed measure Expectedmeasure
  • 50.
  • 51.
  • 52.
  • 53. Precision Closeness of agreement between quantity values obtained by replicate measurements of a quantity, under specified conditions.
  • 54.
  • 55. Precision ØPrecision should be assessed using quality control material (A minimum of 2 levels), or pooled serum (Of minimum two concentration levels) ØEach level of the QC material / pooled serum is measured 5 times per day (Within run), for 5 days (in between runs)
  • 56. Precision ØThe measures obtained from this precision study are to be collected, and the mean, SD, and CV are calculated for each parameter, for the used QC levels / pooled serum. ØThe obtained CVs, are statistically compared to the manufacturer’s claim using ANOVA test of significance, to determine if there is a significant different between the obtained CVs (and their verification intervals), and the manufacturer’s claim at a certain CI; usually 95%
  • 57.
  • 58. Accuracy Closeness of the agreement between the result of a measurement and a true value of the measurand.
  • 59. Trueness Closeness of the agreement between the replicates of result of a measurement, and a true value of the measurand.
  • 60. ØThe difference between the mean of replicates of a measurement, and its true value is the BIAS. Ø The CLSI calculation of bias is based on the results of 7 – 11 PT samples; each is measured in duplicate, and then compared to the true value (Peer’s mean) using student T-test.
  • 61.
  • 65. Why is Z-score used in PT results??
  • 66. Bias is verified for the tested method when there is no significant difference between: q Mean Z-score of PT results (7 – 11) is not significantly different from Zero q Mean Z-score of PT results (7 – 11) is not significantly different from peer’s mean
  • 67.
  • 68. Reference intervals ØRemember that we are just “Verifying” the reference intervals stated by the manufacturer or published in the literature, and “transferring” them to the lab using the method under study. Ø “Establishment” of reference intervals is another issue.
  • 69.
  • 70.
  • 71. Reference intervals ØAcceptability of the transfer shall be assessed by examining 20 reference individuals, from our subject population, and comparing the obtained test results to those of the manufacturer/ Literature. ØThose 20 individuals should be selected in such a way that will satisfy the exclusion and partitioning criteria. ØThe test results should be examined to make sure that none of the results appears to be an outlier.
  • 72. Reference intervals ØThe manufacturer's / Literature reference intervals are considered verified if no more than two of the 20 tested subjects' values (or 10% of the test results) fall outside those ranges.
  • 73. Reference intervals ØExclusion / partitioning criteria include: age, sex, fasting status, disease history, drug history, previous surgeries, and time of the cycle / pregnancy for females.
  • 74.
  • 75.
  • 76. Method Comparison Ø According to CLSI, at least 40 samples should be assayed on both methods under examination (two field methods), or between one tested method and a reference method. Ø Several statistical approaches can be used, one of them is to calculate the correlation coefficient “r” Ø“r” should be more than or equal 0.95
  • 77.
  • 78. Total Error It is the summation of both Random and Systematic error. It is calculated as follow: TE = Bias (%) + 2 CV It is compared to Biological Variation (or any other specifications) for Total Allowable Error
  • 79.
  • 80. Uncertainty Ø It is an interval around a reported laboratory result that specifies the location of the true value with a given probability Ø It takes into consideration both the imprecision (SD), and the inaccuracy (Bias) Ø It is calculated from the data of 6 month minimum
  • 81. What performance characteristics are usually validated? ØReportable range (Linearity) ØPrecision (or imprecision) ØAccuracy (or inaccuracy, bias) ØReference interval