His bundle pacing can lead to complications that depend on lead positioning and hardware. Possible issues include high capture thresholds that may worsen over time, disease progression distal to the pacing site, atrial oversensing, and poor ventricular sensing. Careful evaluation at implant and thoughtful programming can prevent many complications, though some patients still experience issues like increased thresholds. Most patients do well long-term with His bundle pacing when complications are managed appropriately.
1. His Bundle Pacing: Which are
the Reported Complications &
How to Avoid Them
Sergio L. Pinski, MD, FHRS
Cleveland Clinic Florida
Weston, FL, USA
@SergioPinski
2. Complications of His Bundle Pacing
• Dependent on the hardware used (ie, Medtronic 3830 lead)
• Dependent on the lead position in the annulus (e.g, large A signal and
small V signal)
3. B
TV
CS
T
V
R
A HBP
RV
Correa de Sa et al. Circ Arrhythm Electrophysiol 2012;5;244
TV
M
SV
His Lead Generally Does Not Cross the Tricuspid Valve
8. Bhatt et al. JACC Clin Electrophysiol 2018;4:1397
9. Possible Complications
• Higher capture threshold, trade-off between safety margin and
battery longevity
• Possible progression of disease distal to the site of pacing resulting in
block
• Atrial oversensing
• Poor ventricular sensing
• Atrial capture
• No specific devices or algorithms, need to adapt.
16. High threshold
• A high threshold at implant > a higher threshold at follow-up. Do not
accept a threshold at implant > 2.5 V at 1 ms
• Exception is His current injury. Threshold may be initially high, but drops
during the case. Predicts excellent long-term outcome
• Red flags during implant:
• Large difference between unipolar and bipolar threshold
• Lead that is too positional, threshold changes a lot with more or less slack
• Too low impedance
• Loss of 1:1 His bundle pacing with rapid pacing
• Even so, some patients with good implant thresholds develop high
threshold in follow up. Fibrosis, microdislodgment, excessive slack
22. High Threshold
• > 2.5 V at 1 ms 20-35% at follow-up
• Lead revision in 5-10 %
23. Management of High Threshold
• Minimize V pacing
• Turn off, use back up lead
• Accept high output and accelerated battery drain
• Accept septal pacing (high His, low myocardial threshold)
• Lead revision
24.
25.
26. Back up RV lead
• At initial implant: more likely to use in infranodal block, selective HBP,
simultaneous AV nodal ablation, higher threshold, early on in experience.
Use a CRT device, His in LV channel.
• During upgrade from conventional dual-chamber pacing. Better to keep the
lead, but you don’t have to
• Always with defibrillator. If His lead to atrial channel (ie, permanent atrial
fibrillation with complete heart block) you can only pace bipolar.
• Connecting the His lead in the LV port of a CRTD allows tip-to-coil pacing
which has lower threshold
• No need to worry about sensing in the His lead, which will occur via RV
lead
27. Programming with back-up RV lead
• If 3 leads, His lead in LV channel. Program DDD(R) LV 80 ms ahead, the
back-up RV pulse will fall in refractory. With current MDT RV pacing
cannot be turned off. Chronically, output can be minimized to save
battery.
• If 2 leads (i.e., permanent atrial fibrillation with complete heart
block), His lead in atrial channel. Program DVI(R) (or DDI(R) with low
atrial sensitivity). Program AV delay long enough to allow inhibition
and disable safety pacing.
28.
29.
30.
31.
32. Progression of Disease
• Septal inflammation/scarring (e.g, sarcoidosis)
• Future TAVR or SAVR
• Future AV nodal ablation (+++ if needed from the left side)
38. Sensing (applies to DDD or VVI
pacemaker)
• Best assessed at the time of implant
• Unipolar or bipolar configuration
• Acceptable R wave amplitude 1 - 3mV
• More important is size of A signal in relationship to the V signa;
• May vary with patient position. Enough slack minimizes changes in
thresholds with position
• Different conducted beat vs. ectopic
• Use pacemakers with high sensitivity/autogain
43. A
H
V
A
H
V
BIPOLAR Sensitivity 0.5 mV
H
V
H
V
UNIPOLAR Sensitivity 1.0 mV
Far-field Atrial Oversensing in Sinus Rhythm
Courtesy Dr. Pugal Vijayaraman
53. Conclusions
• His bundle pacing creates unique challenges due to the lead location at
the AV annulus
• Complications are still more common than with standard RV pacing.
Most can be prevented by careful evaluation at implant.
• Complications not expected at implant can often be circumvented by
thoughtful programming
• Majority of patients with permanent HBP do well with no significant
issues
• Understanding the nuances and limitations associated with HBP is
crucial for long term success
54. •Slides will be available for download at
www.slideshare.net/sergiopinski
•I recommend following the Twitter hashtag
#dontdisthehis to learn, present, comment
about His bundle pacing
Editor's Notes
Selective HB capture with correction of LBBB followed by continued selective capture but loss of LB recruitment, total loss was at 0.75 V. Changes occur in evoked response morphology as well.
Document BBB corection 1.5V @ 1 ms, total loss 1 V @ 1 ms. Programmed 2.5 V @ 1 ms.
Sensing is best assessed at the time of implant prior to fixing the lead to avoid far field atrial electrograms or near field his electrograms that can resulting in oversensing. When patients seen in follow up, testing can be done in unipolar (if not dependent) or bipolar configuration, using caution to assess for oversensing of myopotentials when in unipolar configuration.
Acceptable R wave amplitude is 1 to 3 mV. Final programming should include the usual 2:1 safety margin. Need to pay close attention to the size of the atrial electrogram especially in relation to the size of the His electrogram to avoid oversensing.