GDx Explained: Scanning Laser Polarimetry for RNFL Assessment
1. GDx
DR SHYLESH B DABKE
GLAUCOMA FELLOW
ARAVIND EYE HOSPITAL
Download and watch in slideshow mode, online viewing will be disgusting
2. INTRODUCTION
• GDX measures thickness of RNFL based on birefringent property of
the tissue.
• GDX is:
- Simple to use and easy for both the patient and operator.
- Near Infra-red wavelength(780 nm).
- Measurement time is 0.7 seconds.
- Total chair time less than 3 minutes for both eyes.
- Undilated pupils work best.
- Painless & safe procedure.
3. • The GDx :
- Maps the RNFL and compares them to a database of healthy,glaucoma-free
patients.
- Analyses the RNFL thickness around the optic disc.
• Sensitivity of 89% and a Specificity of 98%.
4. Why RNFL Assessment??
• RNFL defects - earliest sign of glaucoma.
• 88% of ocular hypertensives who converted to glaucoma had
RNFL defects along with VF defects. 60%of these converters
had RNFL defects 6 years prior to VF defects (Sommer et al).
• RNFL changes are detected more frequently than ONH changes
in eyes converted from ocular hypertension to glaucoma
(Quigley et al).
• RNFL damage occurs earlier than VF defects and ONH damage.
5. PRINCIPLE - Scanning Laser Polarimetry
• Scanning laser polarimetry is an imaging technology that is utilized to measure
peripapillary RNFL thickness.
• Based on the Principle of Birefringence.
• Main birefringent intraocular tissues are the cornea, lens and the retina.
• In the retina, the parallel arrangement of the microtubules in retinal ganglion
cell axons causes a change in the polarization of light passing through them.
• The change in the polarization of light is called Retardation.
• The retardation value is proportionate to the thickness of the RNFL
6. RNFL - highly ordered parallel axon bundles
Axons - microtubules, cylindrical intracellular organelles with
diameters smaller than the wavelength of light.
Highly ordered structure of microtubules are the source of
RNFL birefringence
8. These components travel at different velocities
which creates a relative phase shift
2
Retardation
1
Amount of phase shift is ≃ RNFL thickness
9. Polarized light passes through the eye and is reflected off the
retina. Because the RNFL is a birefringent, the two components of
the polarized light are phase shifted relative to each other
(retarded). The amount of retardation is captured by a detector,
and converted into thickness (microns).
10. Scanning Laser Polarimetry
• Amount of retardation is captured by a detector and converted into
thickness (microns)
• Raster scan - 40° horizontally and 20° vertically and includes both
peripapillary region and macular region
11. • Anterior segment birefringence - cornea and lens.
• The total retardation is the sum of the cornea, lens and the retina.
• Compensation of anterior segment birefringence is needed to isolate
RNFL birefringence.
• GDx VCC(variable corneal compensation) measures and individually
compensates for each eye.
12. • VCC stands for variable corneal compensator, which was created to account for
the variable corneal birefringence in patients
• Uses the birefringence of Henle’s layer in the macula as a control for
measurement of corneal birefringence
16. Patients information
• Patient data and quality score:
- Patient’s name
- Date of birth
- Gender
- Ethnicity
- An ideal quality score is from 7 to 10
17. FUNDUS IMAGE
• The fundus image is useful to check for
image quality:
• Every image has a Q Score representing the
overall quality of the scan
• The Q ranges from 1-10, with values 8-10
representing acceptable quality.
• This score is based on a number of factors
including :
-Well focused,
- Evenly illuminated,
- Optic disc is well centered,
- Ellipse is properly placed around the ONH.
19. • The fixed sized default calculation circle is a centred on ONH.
• Inner diameter of the band is 2.4mm, outer diameter of the band is
3.2mm & the band is 0.4mm wide.
• The calculation circle is the area found between the two concentric
circles, which measure the temporal-superiornasal-inferior-temporal
(TSNIT) and nerve fiber indicator (NFI) parameters
• By Resizing The Calculation Circle And Ellipse, the Operator
is able To Measure Beyond A Large Peripapillary Atrophy Area
20. RNFL Thickness Map
• The thickness map shows the RNFL thickness in a
color-coded format from blue to red.
• Hot colors like red and yellow mean high
retardation or thicker RNFL
• Cool colors like blue and green mean low
retardation / thinner RNFL
21. Healthy RNFL Glaucomatous RNFL
• A healthy eye has yellow and red colors in the superior and inferior regions
representing thick RNFL regions and blue and green areas nasally and temporally
representing thinner RNFL areas.
• In glaucoma, RNFL loss will result in a more uniform blue appearance
22. Deviation Maps
• Reveals the location & magnitude of RNFL
defects over the entire thickness map
• RNFL thickness of patient is compared to
the age-matched normative database
• Each square represents a “Super Pixel”
• RNFL thickness at each super pixel is
compared to age matched normative
database
• Super pixel that fall below normal range
is flagged by color squares based on
probability of normality
23. • Dark blue squares RNFL thickness is below the 5th
percentile of the normative database
• Light blue squares deviation below the 2% level
• Yellow deviation below 1%
• Red deviation below 0.5%.
24.
25. TSNIT Map • Displays the RNFL thickness values along the
calculation circle
• In a normal eye the TSNIT plot follows the
typical ‘Double Hump’ pattern, with thick RNFL
measures superiorly and inferiorly and thin
RNFL values nasally and temporally
• In a healthy eye, the TSNIT curve will fall
within the shaded area which represents the
95% normal range for that age
• When there is RNFL loss, the TSNIT curve will
fall below this shaded area, especially in the
superior and inferior regions
26. • In the center of the printout at the
bottom, the TSNIT graphs for both eyes
are displayed together.
• Healthy eye there is good symmetry
between the TSNIT graphs of the two
eyes and the two curves will overlap
• In glaucoma, one eye often has more
advanced RNFL loss and therefore the
two curves will have less overlap
27. Parameters Table
• The TSNIT parameters are summary
measures based on RNFL thickness
values within the calculation circle
• Normal parameter values are displayed
in green
• Abnormal values are color-coded based
on their probability of normality.colours
are similar to deviation maps.
28.
29. • TSNIT SD : Captures the amount of modulation (peak to trough difference)
in TSNIT graph.
In normal eye there is high modulation.
In glaucoma eye there is RNFL loss in Sup
& Inf region which results in low
modulation & a low TSNIT SD value.
31. The Nerve Fiber Indicator (NFI)
• Global measure based on the entire RNFL thickness map
• Calculated using an advanced form of neural network, called a
Support Vector Machine (SVM)
• Not colour coded
• Output values range from 1 –100
– 1-30 -> low likelihood of glaucoma
– 31-50 -> glaucoma suspect
– 51+ -> high likelihood of glaucoma
Clinical research has shown that the
NFI is the best parameter for
discriminating normal from glaucoma
36. Glaucoma is Detected Early with GDx VCC
Normal SAP
GDx VCC Abnormal OS
Normal SAP
37. CONCLUSIONS
• The ability to detect early glaucomatous structural changes has great
potential value in delaying and avoiding progression of the disease
• Should not be regarded as replacing the skilled ophthalmologist’s capacity
to evaluate all aspects of the patient’s diagnosis.
• But they can definitely aid in the complicated decision-making process