Successfully reported this slideshow.
We use your LinkedIn profile and activity data to personalize ads and to show you more relevant ads. You can change your ad preferences anytime.

IABP troubleshooting

21,606 views

Published on

various alarm in IABP, troubleshooting

Published in: Health & Medicine
  • Login to see the comments

IABP troubleshooting

  1. 1. IABP troubleshooting Dr Siva Subramaniyan PGIMER &Dr.RML Hospital New Delhi
  2. 2. IABP • The intra-aortic balloon pump (IABP) is a mechanical device that is temporarily used to improve cardiac function. • In many situations, the IABP is life-saving in its ability to stabilize patients as they await procedures such as heart transplant, coronary artery bypass grafting (CABG), or percutaneous coronary interventions (PCI) such as PTCA. • An IABP may be further indicated in the management of cardiogenic shock
  3. 3. Basic principles of counterpulsation • Counterpulsation (augmentation) is a term that describes balloon inflation in diastole and deflation in early systole. • Balloon inflation causes ‘volume displacement’ of blood within the aorta, both proximally and distally. This leads to a potential increase in coronary blood flow and potential improvements in systemic perfusion by augmentation of the intrinsic ‘Windkessel effect’, whereby potential energy stored in the aortic root during systole is converted to kinetic energy with the elastic recoil of the aortic root.
  4. 4. Physiological effects of IABP therapy • The primary goal of IABP treatment is to improve the ventricular performance of the failing heart by facilitating an increase in myocardial oxygen supply and a decrease in myocardial oxygen demand.
  5. 5. Expected changes with IABP support in hemodynamic profile in patients with Cardiogenic shock - Decrease in SBP by 20 % - Increase in aortic Diastolic Press. by 30 % ( raise coronary blood flow) - Increase in MAP - Reduction of the HR by 20% -Decrease in the mean PCWP by 20 % - Elevation in the COP by 20%
  6. 6. • Although these effects are predominately associated with enhancement of LV performance, IABP may also have favourable effects on right ventricular (RV) function by complex mechanisms including accentuation of RV myocardial blood flow, unloading the left ventricle causing reduction in left atrial and pulmonary vascular pressures and RV afterload. • IABP inflates at the onset of diastole, thereby increasing diastolic pressure and deflates just before systole, thus reducing LV afterload
  7. 7. The magnitude of these effects depends upon: (i) Balloon volume: the amount of blood displaced is proportional to the volume of the balloon. (ii) Heart rate: LV and aortic diastolic filling times are inversely proportional to heart rate; shorter diastolic time produces lesser balloon augmentation per unit time. (iii) Aortic compliance: as aortic compliance increases (or SVR decreases), the magnitude of diastolic augmentation decreases.
  8. 8. 14 IABP –instrumentation and techniques
  9. 9. 15 The IAB Counter pulsation system two principal parts A flexible catheter -2 lumen • first - for distal aspiration/flushing or pressure monitoring • second - for the periodic delivery and removal of helium gas to a closed balloon. A mobile console • system for helium transfer • computer for control of the inflation and deflation cycle
  10. 10. BALLOON IABP catheter: • 10-20 cm long polyurethane bladder • 25cc to 50cc capacity • Optimal 85% of aorta occluded (not 100%) • The shaft of the balloon catheter contains 2 lumens: - one allows for gas exchange from console to balloon - second lumen for catheter delivery over a guide wire for monitoring of central aortic pressure after installation.
  11. 11. 17 BALLOON SIZING  Sizing based on patients height  Four common balloon sizes  Balloon length and diameter increases with each larger size  40 cm³ balloon is most commonly used  Paediatric balloons also available : sizes 2.5, 5.0, 12.0 and 20 cm³ Balloon size Height 50 cm³ > 6 feet 40 cm³ 5 feet 4 inch to 6 feet 34 cm³ 5 feet to 5 feet 4 inch 25 cm³ < 5 feet
  12. 12. 18 INTRA AORTIC BALLOON
  13. 13. TIP
  14. 14. 24 - The end of the balloon should be just distal (1-2 cm) to the takeoff of the left subclavian artery - Position should be confirmed by fluoroscopy or chest x-ray POSITIONING
  15. 15. 25 Connecting to console: - Connect helium gas tube to the console via a long extender - Open helium tank. - The central lumen of the catheter is flushed and connected to pressure tubing with 3 way and then to a pressure transducer to allow for monitoring of central aortic pressure. - Zero the transducer Initial set-up: - Once connected properly the console would show ECG and pressure waveforms. - Check Basal mean pressure - Make sure the setting is at “auto” - Usually IABP started at 1:1 or 1:2 augmentation - Usually Augmentation is kept at maxim
  16. 16. Modes of operation: • The trigger can be “automatic”, the computer sets trigger source, computer sets inflation and deflation times. Trigger automatically searches for most optimal source if initial trigger is not detected. • The Trigger can also be “semi automatic” the operator sets the source, operator sets initial inflation and deflation times. Computer will determine subsequent timing intervals.
  17. 17. 30
  18. 18. 31
  19. 19. 32
  20. 20. waveform is acceptable ? • First change from 1:1 to 1:2 augmentation • Check the dicrotic notch • As the balloon inflates at the onset of diastole, a sharp and deep ‘V’ is observed at the dicrotic notch. Balloon inflation causes augmentation of diastolic pressure and a second peak is observed. This peak is referred to as diastolic augmentation. Diastolic augmentation is ideally higher than the patient’s systolic pressure except when reduced stroke volume causes a relative decrease in augmentation
  21. 21. 2) Confirm if end diastolic wave following the augmented wave is less than an non augmented wave.
  22. 22. • The balloon aortic end-diastolic pressure is 5 mmhg to 10 mmhg less than the patient’s aortic end-diastolic pressure and • The assisted systolic pressure is 5 mmhg to 10 mmhg less than the patient’s previous systolic pressure
  23. 23. 41
  24. 24. BALLOON PRESSURE WAVEFORM
  25. 25. TROUBLESHOOTING OF IABP
  26. 26. TROUBLESHOOTING • Challenges can occur when IABP therapy is used in the CCU. The most frequent causes for IABP malfunction include the possibility that the - balloon may not adequately augment CO, - ECG trigger may not be functioning properly, or - autofill function may fail
  27. 27. 1) Augmentation below the set limit Reasons for low diastolic augmentation include the following issues: ● Incorrect timing of balloon inflation and deflation. This problem can be corrected by adjusting inflation and deflation on console. ● Leak in the balloon catheter. Occasionally balloons can rupture or tear. This problem warrants prompt changing of the balloon catheter. ● Leak in the gas circuit resulting in poor inflation. Check all connections from the catheter to the gas cylinder and tighten them as necessary.
  28. 28. ●Balloon not unwrapped completely. This issue results in poor or incomplete opening of the balloon. Check the position of the catheter with an x-ray. Attempt to inflate the balloon with an appropriate-sized syringe. If this measure is unsuccessful, the catheter will need to be replaced. ● Poor cardiac function or low vascular resistance. If appropriate, add positive inotropic agents and vasoactive agents. ● Dysrhythmias such as atrial fibrillation or ventricular tachycardia that result in low stroke volume. To correct these conditions, administer antiarrhythmics as ordered, and improve oxygen and electrolyte imbalances.
  29. 29. EARLY INFLATION
  30. 30. DELAYED INFLATION
  31. 31. EARLY DEFLATION
  32. 32. LATE DEFLATION
  33. 33. 2) NO TRIGGER • On occasion, the ECG trigger may not function properly. • Common causes of a faulty trigger include poor electrode placement, low ECG voltage, faulty electrode pads or cables, dysrhythmias, and other equipment’s interference with the ECG signal. • If the problem cannot be easily rectified, switching to the arterial pressure trigger will be necessary until the problem can be solved.
  34. 34. 3) Autofill Failure • The autofill feature on the IABP maintains the volume of gas within the balloon. if this feature is not functioning properly, an autofill alarm will sound. The cause of this problem could be an insufficient amount of gas in the tank or occlusion of the gas outlet. The amount of gas in the tank should be checked and the tank replaced as needed.
  35. 35. 4) Check IAB Catheter
  36. 36. 5) IAB Disconnected
  37. 37. 6) Prolonged Time in Standby
  38. 38. 3) Rapid Gas Loss or Leak in IAB Circuit
  39. 39. 59 Variation in balloon pressure wave forms Increased duration of plateau due to longer diastolic phase Decreased duration of plateau due to shortened diastolic phase
  40. 40. 60 Variation in balloon pressure wave forms Varying R-R intervals result in irregular plateau durations
  41. 41. 61 Variation in balloon pressure wave forms Increased height or amplitude of the waveform Decreased height or amplitude of the waveform
  42. 42. 62 Variation in balloon pressure wave forms Gas leak Leak in the closed system causing the balloon pressure waveform to fall below zero baseline.. - due to a loose connection - a leak in the IAB catheter - H2O condensation in the external tubing - a patient who is tachycardiac and febrile which causes increased gas diffusion through the IAB membrane
  43. 43. 63 CATHETER KINK Rounded balloon pressure waveform - Loss of plateau resulting from a kink or obstruction of shuttle gas - Kink in the catheter tubing - Improper IAB catheter position - Sheath not being pulled back to allow inflation of the IAB - IAB is too large for the aorta - IAB is not fully unwrapped
  44. 44. 64 “Balloon too large” syndrome
  45. 45. ACC/AHA 2013Practice Guidelines (ESC 2012- IIb) 67
  46. 46. ESC 2014 GUIDELINES 68
  47. 47. THANK YOU

×