Introduction
The analysis of chlorine has become more and more important in the recent
years. The use of substitute fuel in cement plants, brickyards and other
industries with high demand for energy requires easy, low maintenance
chlorine analysis.
Traditionally chlorine determination is done using Schöninger digestion,
which is very time consuming and difficult to automate. Therefore, high
temperature combustion followed by coulometric detection of the formed
HCl has gained more and more interest. Still widely used, however, the
coulometric method has several major drawbacks. It requires a substantial
amount of maintenance effort due to the electrolyte solution needing to be
replaced very frequently, and the electrodes are susceptible to wear out.
Moreover, the dynamic range is limited to lower concentrations with this
approach. Ion chromatography in combination with conductivity detection
on the other hand is a more stable technique and allows the separate
determination of all halogens. The drawback of this solution is the very
high investment costs.
Elementar as the pioneer of elemental analysis has developed a simpler
alternative. The combination of the well-proven, state-of-the-art elemental
analyzers with high performance solid state electrochemical detectors (ECD)
makes chlorine analysis much easier, more reliable and more precise than
ever before. It allows chlorine measurements over a larger concentration
range with low maintenance effort.
Unique high precision chlorine analysis
for a large concentration range
TECHNICAL NOTE
UNIQUE CHLORINE OPTION
trace SN cube
rapid CS cube
vario MICRO cube
vario EL cube
vario MACRO cube
Measuring principle
The sample is transferred from the sample carousel into
the combustion tube through a blank-free ball valve.
Unlike systems with a horizontal furnace this does not
require difficult mechanics and/or compressed air to feed
the samples into the combustion tube.
The sample then enters the high-temperature combustion
zone, which ensures 100% transformation of all chlorine to
HCl. In the next step water is efficiently removed from the
carrier gas to avoid condensation and loss of HCl. The carrier
gas finally passes the solid-state electrochemical detector
to detect the HCl (see Figure 1). This default configuration is
available for the rapid CS cube, vario MICRO cube, vario EL
cube and vario MACRO cube models.
In some cases the ECD cell shows a slight cross sensitivity to
CO2, resulting in a small negative peak at the beginning of
the integration. For higher Cl concentrations (> 100 ppm)
this does not significantly contribute to the resulting
chlorine concentration. For low concentrations as in
gasoline this small negative peak may influence the results.
For this reason the setup of the chlorine option of the trace
SN cube, which is optimized for low Cl contents, is slightly
different (see Figure 2). After removing the moisture from
the carrier gas, the HCl is trapped on a special trap column.
After the combustion process and removal of interferences,
the trap column heats to release HCl, which is then
detected in the solid-state electrochemical detector. The
Advanced Purge and Trap technology leads to peak focusing
that significantly improves the signal-to-noise ratio and
guarantees highest sensitivity.
Evaluation of the detector signal is done on a standard
PC. The entire process is fully automated for easy use and
overnight operation.
Combustion
Drying Tube Detection
Sample Carousel
Blank-Free Ball Valve
CombustionTube
Furnace
ECD
Combustion
Detection
Sample Carousel
Blank-Free Ball Valve
CombustionTube
Furnace
ECD
High Sens
Gas Separation
Figure 1. Functional principle of the chlorine detection
option in the default configuration
Figure 2. Functional principle of the trace SN cube with the
chlorine detection option for low concentrations
Figure 5. HCl peak of a diesel sample (+ 5 ppm Cl) analyzed
using the trace SN cube with chlorine detection option and
20 ppm measuring cell.
Electrochemical detection
The following reactions occur in the electrochemical cell:
working electrode: Ag + HCl → AgCl + H+ + e
reference electrode: O2 + 4e + 4H+ → 2H2O
Counter Electrode
Reference Electrode
Sensing Electrode
Hydrophobic Membrane
Electrolyte
Capillary Diffusion Barrier
Calibration
There are several standards that can be used for calibration.
If no capsule press is available ammonium chloride (Cl
content 65.97%) is the substance of choice. Moreover, the
high solubility of chlorine in water is ideal for preparing
standard solutions. Another possible calibration standard
could be oil with a defined amount of Cl or if available
matrix specific chlorine standards. Usually these standards
can be used without accelerants. Figure 6, for example,
shows a typical calibration curve, obtained by a coal
standard with 0.11% Cl on the rapid CS cube with Cl-option.
Figure 3. Functional principle of the EC cell
Figure 4. HCl peak of a 1.4 mg NH4Cl sample analyzed using
the rapid CS cube with chlorine detection option and
5000 ppm measuring cell.
The electrons produced in this reaction lower the current
between working and reference electrode, which is a
measure for the HCl concentration in the carrier gas. This
detection method can be described as amperometric
detection. The capillary diffusion barrier is basically a hole
whose diameter determines the sensitivity of the cell.
The electrolyte is a concentrated, non-oxidizing acid. A
schematic diagram of the electrochemical cell is shown in
Figure 3. Chemistry dictates that the ECD cell can also be
affected by high concentrations of HBr, but this is rarely
seen in practice.
Cl (µg)
44.3
0.0
0.0 Peakarea (AU) 20745
y = 5.5571.10⁻⁴*x + 2.1391.10⁻⁶
r = 0.996
Figure 6. Calibration curve EC cell for chlorine
0 100 200 300 400 500 600
24000
20000
16000
12000
8000
4000
0
EC/Cl
[unit]
[sec.]
EC/Cl
[unit]
HCl column
[°C]
- 300
- 0
- 50
- 100
- 150
- 200
- 250
0 100 200 300 400 500 600
- 2000
- 1600
- 1200
- 800
- 400
- 0
[sec.]
Typical applications
The chlorine detection option is optimized for applications in the area of fuel, residual fuel, biomass, waste, soil, along
with others. Due to limitations in precision and accuracy, it is not recommended to use the Cl-option for classical elemental
analysis (e.g. to determine the Cl content in pure chemicals). The following table shows some typical applications for the
Cl-option.
Elementar Analysensysteme GmbH | Donaustraße 7 · 63452 Hanau (Germany)
Phone: +49 (0) 6181 9100-0 | info@elementar.de | www.elementar.de
Art.-No.05005032,01/2016A
SUBSTANCE WEIGHT (mg) DETECTOR Cl (W-%) REL. SD (%)
ANTHRACIT-1 34 200 0.151 1.52
ANTHRACIT-2 33 200 0.021 6.1
WOOD COAL 60 200 0.0045 8.9
BIOMASS-1 15 200 0.133 4.2
BIOMASS-2 30 200 0.0086 8.9
ORGANIC WASTE 20 200 0.142 4.3
RESIDUAL FUEL 31 5000 0.893 6.1
PLASTIC FIBER 10 200 0.354 11.3
DIESEL SPIKED WITH 1 PPM Cl 35 20 1.07 8.8
Available EC cells
Elementar provides three different types of electrochemical
cells depending on the concentration range. The names
of the cells refer to the maximum HCl concentration in
the gas stream which can be analyzed. Note that this
concentration cannot directly be transferred to the Cl
content of the sample.
The characteristics of the different electrochemical cells are
presented in the following table.
CELL TYPE LOWER LIMIT
(µg Cl abs.)
UPPER LIMIT
(µg Cl abs.)
REL. SD Cl
(%)*
20 PPM†
0.05 2 8
200 PPM 0.5 20 5
5000 PPM 15 1200 5
*: obtained on Conostan standards in the middle of the measuring range
†
: available for the trace SN cube only
Elementargroup – your partner for elemental analysis
Elementargroup is the world leader in high performance analysis of organic ele-
ments. Continuous innovation, creative solutions and comprehensive support
form the foundation of the Elementar and Isoprime brands ensuring our pro-
ducts continue to advance science across agriculture, chemical, environmental,
energy, materials and forensics markets in more than 80 countries.

More Related Content

What's hot

6. precipitation titrations
6. precipitation titrations6. precipitation titrations
6. precipitation titrationsNikithaGopalpet
 
Measurement of exhaust emissions
Measurement of exhaust emissionsMeasurement of exhaust emissions
Measurement of exhaust emissionsDENNY OTTARACKAL
 
Estimation of Chlorides in any water body ( Pond water)
Estimation of Chlorides in any water body ( Pond water)Estimation of Chlorides in any water body ( Pond water)
Estimation of Chlorides in any water body ( Pond water)Sadia Rahat
 
Detectors used in UV spectophotometry
Detectors  used  in UV spectophotometryDetectors  used  in UV spectophotometry
Detectors used in UV spectophotometryNimmiRoy
 
Determination of chloride content in water by Volhard method
Determination of chloride content in water by Volhard methodDetermination of chloride content in water by Volhard method
Determination of chloride content in water by Volhard methodPrasad Madushanka
 
Gas chromatography detectors
Gas chromatography detectorsGas chromatography detectors
Gas chromatography detectorsparnavi kadam
 
Precipiatation titration- Pharmaceutical Analysis
Precipiatation titration- Pharmaceutical AnalysisPrecipiatation titration- Pharmaceutical Analysis
Precipiatation titration- Pharmaceutical AnalysisSanchit Dhankhar
 
Aqueous Speciation in Hydrochemistry
Aqueous Speciation in HydrochemistryAqueous Speciation in Hydrochemistry
Aqueous Speciation in Hydrochemistryaqion
 
IA on effect of NaOH and KOH on the rate of hydrogen production, using alumin...
IA on effect of NaOH and KOH on the rate of hydrogen production, using alumin...IA on effect of NaOH and KOH on the rate of hydrogen production, using alumin...
IA on effect of NaOH and KOH on the rate of hydrogen production, using alumin...Lawrence kok
 
Precipitation titrations
Precipitation titrationsPrecipitation titrations
Precipitation titrationsSuresh Selvaraj
 
Precepitation titration mohrs method
Precepitation titration mohrs methodPrecepitation titration mohrs method
Precepitation titration mohrs methodHemn Gardy
 
J. Electrochem. Soc.-2003-Frank-C244-50
J. Electrochem. Soc.-2003-Frank-C244-50J. Electrochem. Soc.-2003-Frank-C244-50
J. Electrochem. Soc.-2003-Frank-C244-50Aaron Frank
 
Gas chromatography and its instrumentation
Gas chromatography and its instrumentationGas chromatography and its instrumentation
Gas chromatography and its instrumentationArgha Sen
 
pondus Hydrogenii-pH
pondus Hydrogenii-pHpondus Hydrogenii-pH
pondus Hydrogenii-pHAmitava Datta
 
IA Design and interesting reseach questions
IA Design and interesting reseach questionsIA Design and interesting reseach questions
IA Design and interesting reseach questionsLawrence kok
 
Argentometric titration
Argentometric titrationArgentometric titration
Argentometric titrationZainab&Sons
 
Karl Fischer titration,principle,apparatus, titration types,Endpoint detection
Karl Fischer titration,principle,apparatus, titration types,Endpoint detectionKarl Fischer titration,principle,apparatus, titration types,Endpoint detection
Karl Fischer titration,principle,apparatus, titration types,Endpoint detectionvarinder kumar
 

What's hot (20)

6. precipitation titrations
6. precipitation titrations6. precipitation titrations
6. precipitation titrations
 
Measurement of exhaust emissions
Measurement of exhaust emissionsMeasurement of exhaust emissions
Measurement of exhaust emissions
 
Estimation of Chlorides in any water body ( Pond water)
Estimation of Chlorides in any water body ( Pond water)Estimation of Chlorides in any water body ( Pond water)
Estimation of Chlorides in any water body ( Pond water)
 
Detectors used in UV spectophotometry
Detectors  used  in UV spectophotometryDetectors  used  in UV spectophotometry
Detectors used in UV spectophotometry
 
Determination of chloride content in water by Volhard method
Determination of chloride content in water by Volhard methodDetermination of chloride content in water by Volhard method
Determination of chloride content in water by Volhard method
 
Gas chromatography detectors
Gas chromatography detectorsGas chromatography detectors
Gas chromatography detectors
 
Precipiatation titration- Pharmaceutical Analysis
Precipiatation titration- Pharmaceutical AnalysisPrecipiatation titration- Pharmaceutical Analysis
Precipiatation titration- Pharmaceutical Analysis
 
Aqueous Speciation in Hydrochemistry
Aqueous Speciation in HydrochemistryAqueous Speciation in Hydrochemistry
Aqueous Speciation in Hydrochemistry
 
IA on effect of NaOH and KOH on the rate of hydrogen production, using alumin...
IA on effect of NaOH and KOH on the rate of hydrogen production, using alumin...IA on effect of NaOH and KOH on the rate of hydrogen production, using alumin...
IA on effect of NaOH and KOH on the rate of hydrogen production, using alumin...
 
Precipitation titrations
Precipitation titrationsPrecipitation titrations
Precipitation titrations
 
Gc
GcGc
Gc
 
Precepitation titration mohrs method
Precepitation titration mohrs methodPrecepitation titration mohrs method
Precepitation titration mohrs method
 
J. Electrochem. Soc.-2003-Frank-C244-50
J. Electrochem. Soc.-2003-Frank-C244-50J. Electrochem. Soc.-2003-Frank-C244-50
J. Electrochem. Soc.-2003-Frank-C244-50
 
Gas chromatography and its instrumentation
Gas chromatography and its instrumentationGas chromatography and its instrumentation
Gas chromatography and its instrumentation
 
Gas Chromatography
Gas ChromatographyGas Chromatography
Gas Chromatography
 
Gas chromatograpy
Gas chromatograpyGas chromatograpy
Gas chromatograpy
 
pondus Hydrogenii-pH
pondus Hydrogenii-pHpondus Hydrogenii-pH
pondus Hydrogenii-pH
 
IA Design and interesting reseach questions
IA Design and interesting reseach questionsIA Design and interesting reseach questions
IA Design and interesting reseach questions
 
Argentometric titration
Argentometric titrationArgentometric titration
Argentometric titration
 
Karl Fischer titration,principle,apparatus, titration types,Endpoint detection
Karl Fischer titration,principle,apparatus, titration types,Endpoint detectionKarl Fischer titration,principle,apparatus, titration types,Endpoint detection
Karl Fischer titration,principle,apparatus, titration types,Endpoint detection
 

Viewers also liked

Tutorial mercadolibre
Tutorial mercadolibreTutorial mercadolibre
Tutorial mercadolibreUFT
 
згубним звичкам скажем
згубним звичкам скажемзгубним звичкам скажем
згубним звичкам скажемKVtetjana
 
卒業研究(学部時)
卒業研究(学部時)卒業研究(学部時)
卒業研究(学部時)akikia2
 
Presentazione primo incontro pubblico
Presentazione primo incontro pubblicoPresentazione primo incontro pubblico
Presentazione primo incontro pubblicoProgettoxSpino2016
 
La soluzione SAP Mill nelle cartiere
La soluzione SAP Mill nelle cartiereLa soluzione SAP Mill nelle cartiere
La soluzione SAP Mill nelle cartiereSoftware Business Srl
 
Keyboard shortcuts in ms excel 4
Keyboard shortcuts in ms excel 4Keyboard shortcuts in ms excel 4
Keyboard shortcuts in ms excel 4thespotlesslives
 
Hand out George van Houtem - Geraffineerd Onderhandelen
Hand out George van Houtem - Geraffineerd Onderhandelen Hand out George van Houtem - Geraffineerd Onderhandelen
Hand out George van Houtem - Geraffineerd Onderhandelen Merlina Saric
 

Viewers also liked (15)

Presentacion Kyriakos
Presentacion KyriakosPresentacion Kyriakos
Presentacion Kyriakos
 
Llundato Christo Peters's CV
Llundato Christo Peters's CVLlundato Christo Peters's CV
Llundato Christo Peters's CV
 
29. 양재우성_2015 에너지자립마을 성과발표자료
29. 양재우성_2015 에너지자립마을 성과발표자료29. 양재우성_2015 에너지자립마을 성과발표자료
29. 양재우성_2015 에너지자립마을 성과발표자료
 
точка зору
точка зоруточка зору
точка зору
 
Tutorial mercadolibre
Tutorial mercadolibreTutorial mercadolibre
Tutorial mercadolibre
 
Revista Ayudeporte Nº 4
Revista Ayudeporte Nº 4Revista Ayudeporte Nº 4
Revista Ayudeporte Nº 4
 
згубним звичкам скажем
згубним звичкам скажемзгубним звичкам скажем
згубним звичкам скажем
 
卒業研究(学部時)
卒業研究(学部時)卒業研究(学部時)
卒業研究(学部時)
 
hashem alkharabsheh
hashem alkharabshehhashem alkharabsheh
hashem alkharabsheh
 
Karen Backer CV
Karen Backer CVKaren Backer CV
Karen Backer CV
 
Presentazione primo incontro pubblico
Presentazione primo incontro pubblicoPresentazione primo incontro pubblico
Presentazione primo incontro pubblico
 
25 raones per llegir
25 raones per llegir25 raones per llegir
25 raones per llegir
 
La soluzione SAP Mill nelle cartiere
La soluzione SAP Mill nelle cartiereLa soluzione SAP Mill nelle cartiere
La soluzione SAP Mill nelle cartiere
 
Keyboard shortcuts in ms excel 4
Keyboard shortcuts in ms excel 4Keyboard shortcuts in ms excel 4
Keyboard shortcuts in ms excel 4
 
Hand out George van Houtem - Geraffineerd Onderhandelen
Hand out George van Houtem - Geraffineerd Onderhandelen Hand out George van Houtem - Geraffineerd Onderhandelen
Hand out George van Houtem - Geraffineerd Onderhandelen
 

Similar to Chlorine measurements over a larger concentration range with low maintenance effort

Analysis of Impurities in Semiconductor-Grade Hydrochloric Acid with the NexI...
Analysis of Impurities in Semiconductor-Grade Hydrochloric Acid with the NexI...Analysis of Impurities in Semiconductor-Grade Hydrochloric Acid with the NexI...
Analysis of Impurities in Semiconductor-Grade Hydrochloric Acid with the NexI...PerkinElmer, Inc.
 
Chlorination & measurement
Chlorination & measurementChlorination & measurement
Chlorination & measurementPraful Hanmante
 
Comparison of cycles and measurement of exhaust gases.pptx
Comparison of cycles and measurement of exhaust gases.pptxComparison of cycles and measurement of exhaust gases.pptx
Comparison of cycles and measurement of exhaust gases.pptxSAshwinDaniel
 
Experimental Investigation on Hypochlorous Acid Water Production using Electr...
Experimental Investigation on Hypochlorous Acid Water Production using Electr...Experimental Investigation on Hypochlorous Acid Water Production using Electr...
Experimental Investigation on Hypochlorous Acid Water Production using Electr...iosrjce
 
Determination of ascobic acid by coulometer
Determination of ascobic acid by coulometerDetermination of ascobic acid by coulometer
Determination of ascobic acid by coulometerchamith herath
 
Coulometry.pptx presentation assignment copy
Coulometry.pptx presentation assignment   copyCoulometry.pptx presentation assignment   copy
Coulometry.pptx presentation assignment copyKibetDerrick
 
Design Of Fuel Cell Membrane Test Stand
Design Of Fuel Cell Membrane Test StandDesign Of Fuel Cell Membrane Test Stand
Design Of Fuel Cell Membrane Test Standrebekah827
 
A_New_Method_Based_on_a_Zero_Gap_Electrolysis_Cell.pdf
A_New_Method_Based_on_a_Zero_Gap_Electrolysis_Cell.pdfA_New_Method_Based_on_a_Zero_Gap_Electrolysis_Cell.pdf
A_New_Method_Based_on_a_Zero_Gap_Electrolysis_Cell.pdfMuhammad Arif Qureshi
 
Scintillations principle, working, merits & demerits & applications
Scintillations  principle, working, merits & demerits & applicationsScintillations  principle, working, merits & demerits & applications
Scintillations principle, working, merits & demerits & applicationsAnas Kinia
 
Stamolys ca71 cod endress+hauser datasheet-cod analyzer
Stamolys ca71 cod endress+hauser datasheet-cod analyzerStamolys ca71 cod endress+hauser datasheet-cod analyzer
Stamolys ca71 cod endress+hauser datasheet-cod analyzerENVIMART
 
Treatment of Hospital Wastewater by Electro-Coagulation Technique
Treatment of Hospital Wastewater by Electro-Coagulation TechniqueTreatment of Hospital Wastewater by Electro-Coagulation Technique
Treatment of Hospital Wastewater by Electro-Coagulation TechniqueIRJET Journal
 
Ar31109115
Ar31109115Ar31109115
Ar31109115IJMER
 

Similar to Chlorine measurements over a larger concentration range with low maintenance effort (20)

Analysis of Impurities in Semiconductor-Grade Hydrochloric Acid with the NexI...
Analysis of Impurities in Semiconductor-Grade Hydrochloric Acid with the NexI...Analysis of Impurities in Semiconductor-Grade Hydrochloric Acid with the NexI...
Analysis of Impurities in Semiconductor-Grade Hydrochloric Acid with the NexI...
 
Chlorination & measurement
Chlorination & measurementChlorination & measurement
Chlorination & measurement
 
Comparison of cycles and measurement of exhaust gases.pptx
Comparison of cycles and measurement of exhaust gases.pptxComparison of cycles and measurement of exhaust gases.pptx
Comparison of cycles and measurement of exhaust gases.pptx
 
O0126291100
O0126291100O0126291100
O0126291100
 
O0126291100
O0126291100O0126291100
O0126291100
 
Experimental Investigation on Hypochlorous Acid Water Production using Electr...
Experimental Investigation on Hypochlorous Acid Water Production using Electr...Experimental Investigation on Hypochlorous Acid Water Production using Electr...
Experimental Investigation on Hypochlorous Acid Water Production using Electr...
 
Determination of ascobic acid by coulometer
Determination of ascobic acid by coulometerDetermination of ascobic acid by coulometer
Determination of ascobic acid by coulometer
 
gas chromatography
gas chromatographygas chromatography
gas chromatography
 
Coulometry.pptx presentation assignment copy
Coulometry.pptx presentation assignment   copyCoulometry.pptx presentation assignment   copy
Coulometry.pptx presentation assignment copy
 
varioelcube
varioelcubevarioelcube
varioelcube
 
Design Of Fuel Cell Membrane Test Stand
Design Of Fuel Cell Membrane Test StandDesign Of Fuel Cell Membrane Test Stand
Design Of Fuel Cell Membrane Test Stand
 
A_New_Method_Based_on_a_Zero_Gap_Electrolysis_Cell.pdf
A_New_Method_Based_on_a_Zero_Gap_Electrolysis_Cell.pdfA_New_Method_Based_on_a_Zero_Gap_Electrolysis_Cell.pdf
A_New_Method_Based_on_a_Zero_Gap_Electrolysis_Cell.pdf
 
Scintillations principle, working, merits & demerits & applications
Scintillations  principle, working, merits & demerits & applicationsScintillations  principle, working, merits & demerits & applications
Scintillations principle, working, merits & demerits & applications
 
Stamolys ca71 cod endress+hauser datasheet-cod analyzer
Stamolys ca71 cod endress+hauser datasheet-cod analyzerStamolys ca71 cod endress+hauser datasheet-cod analyzer
Stamolys ca71 cod endress+hauser datasheet-cod analyzer
 
Gas chromatography
Gas chromatographyGas chromatography
Gas chromatography
 
Treatment of Hospital Wastewater by Electro-Coagulation Technique
Treatment of Hospital Wastewater by Electro-Coagulation TechniqueTreatment of Hospital Wastewater by Electro-Coagulation Technique
Treatment of Hospital Wastewater by Electro-Coagulation Technique
 
Ar31109115
Ar31109115Ar31109115
Ar31109115
 
Idp presentation
Idp presentationIdp presentation
Idp presentation
 
Oxygen analyser
Oxygen analyserOxygen analyser
Oxygen analyser
 
Oxygen analyser
Oxygen analyserOxygen analyser
Oxygen analyser
 

More from Salm en Kipp bv Laboratoriumapparatuur

Helium-free, high performance N/protein analysis according to the Dumas princ...
Helium-free, high performance N/protein analysis according to the Dumas princ...Helium-free, high performance N/protein analysis according to the Dumas princ...
Helium-free, high performance N/protein analysis according to the Dumas princ...Salm en Kipp bv Laboratoriumapparatuur
 
Whitepaper: zekerheid over het juiste monster binnen de toenemende complexite...
Whitepaper: zekerheid over het juiste monster binnen de toenemende complexite...Whitepaper: zekerheid over het juiste monster binnen de toenemende complexite...
Whitepaper: zekerheid over het juiste monster binnen de toenemende complexite...Salm en Kipp bv Laboratoriumapparatuur
 
Elementar presentatie van het webinar: N / eiwitbepaling met behulp van de E...
 Elementar presentatie van het webinar: N / eiwitbepaling met behulp van de E... Elementar presentatie van het webinar: N / eiwitbepaling met behulp van de E...
Elementar presentatie van het webinar: N / eiwitbepaling met behulp van de E...Salm en Kipp bv Laboratoriumapparatuur
 
Note: snelle en nauwkeurige ethanol meting voor bierbrouwerijen met de YSI 29...
Note: snelle en nauwkeurige ethanol meting voor bierbrouwerijen met de YSI 29...Note: snelle en nauwkeurige ethanol meting voor bierbrouwerijen met de YSI 29...
Note: snelle en nauwkeurige ethanol meting voor bierbrouwerijen met de YSI 29...Salm en Kipp bv Laboratoriumapparatuur
 
Here are the slides of the application training on the vario MAX N/CN cube
Here are the slides of the application training on the vario MAX N/CN cubeHere are the slides of the application training on the vario MAX N/CN cube
Here are the slides of the application training on the vario MAX N/CN cubeSalm en Kipp bv Laboratoriumapparatuur
 

More from Salm en Kipp bv Laboratoriumapparatuur (14)

Brady GHS CLP Actiepakket-BBP12 verkrijgbaar bij @salmenkipp
Brady GHS CLP Actiepakket-BBP12 verkrijgbaar bij @salmenkippBrady GHS CLP Actiepakket-BBP12 verkrijgbaar bij @salmenkipp
Brady GHS CLP Actiepakket-BBP12 verkrijgbaar bij @salmenkipp
 
Helium-free, high performance N/protein analysis according to the Dumas princ...
Helium-free, high performance N/protein analysis according to the Dumas princ...Helium-free, high performance N/protein analysis according to the Dumas princ...
Helium-free, high performance N/protein analysis according to the Dumas princ...
 
Workshop osmolaliteit en pipet clinic
Workshop osmolaliteit en pipet clinicWorkshop osmolaliteit en pipet clinic
Workshop osmolaliteit en pipet clinic
 
Seminar cel- en weefselkweek
Seminar cel- en weefselkweekSeminar cel- en weefselkweek
Seminar cel- en weefselkweek
 
Snel en betrouwbare koolstof / stikstofbepaling
Snel en betrouwbare koolstof / stikstofbepaling Snel en betrouwbare koolstof / stikstofbepaling
Snel en betrouwbare koolstof / stikstofbepaling
 
Gids zuiver labwater
Gids zuiver labwaterGids zuiver labwater
Gids zuiver labwater
 
Whitepaper: zekerheid over het juiste monster binnen de toenemende complexite...
Whitepaper: zekerheid over het juiste monster binnen de toenemende complexite...Whitepaper: zekerheid over het juiste monster binnen de toenemende complexite...
Whitepaper: zekerheid over het juiste monster binnen de toenemende complexite...
 
Elementar presentatie van het webinar: N / eiwitbepaling met behulp van de E...
 Elementar presentatie van het webinar: N / eiwitbepaling met behulp van de E... Elementar presentatie van het webinar: N / eiwitbepaling met behulp van de E...
Elementar presentatie van het webinar: N / eiwitbepaling met behulp van de E...
 
Enclosure with laminar flow and inert atmosphere
Enclosure with laminar flow and inert atmosphereEnclosure with laminar flow and inert atmosphere
Enclosure with laminar flow and inert atmosphere
 
Note: snelle en nauwkeurige ethanol meting voor bierbrouwerijen met de YSI 29...
Note: snelle en nauwkeurige ethanol meting voor bierbrouwerijen met de YSI 29...Note: snelle en nauwkeurige ethanol meting voor bierbrouwerijen met de YSI 29...
Note: snelle en nauwkeurige ethanol meting voor bierbrouwerijen met de YSI 29...
 
AppNote Rapid bioprocess culture characterization
AppNote Rapid bioprocess culture characterizationAppNote Rapid bioprocess culture characterization
AppNote Rapid bioprocess culture characterization
 
Soot-free combustion of diesel, gasoline and other organic liquids
Soot-free combustion of diesel, gasoline and other organic liquids Soot-free combustion of diesel, gasoline and other organic liquids
Soot-free combustion of diesel, gasoline and other organic liquids
 
Here are the slides of the application training on the vario MAX N/CN cube
Here are the slides of the application training on the vario MAX N/CN cubeHere are the slides of the application training on the vario MAX N/CN cube
Here are the slides of the application training on the vario MAX N/CN cube
 
Salm en Kipp Labassortiment brochure
Salm en Kipp Labassortiment brochureSalm en Kipp Labassortiment brochure
Salm en Kipp Labassortiment brochure
 

Recently uploaded

Connecting the Dots for Information Discovery.pdf
Connecting the Dots for Information Discovery.pdfConnecting the Dots for Information Discovery.pdf
Connecting the Dots for Information Discovery.pdfNeo4j
 
React Native vs Ionic - The Best Mobile App Framework
React Native vs Ionic - The Best Mobile App FrameworkReact Native vs Ionic - The Best Mobile App Framework
React Native vs Ionic - The Best Mobile App FrameworkPixlogix Infotech
 
Testing tools and AI - ideas what to try with some tool examples
Testing tools and AI - ideas what to try with some tool examplesTesting tools and AI - ideas what to try with some tool examples
Testing tools and AI - ideas what to try with some tool examplesKari Kakkonen
 
Digital Tools & AI in Career Development
Digital Tools & AI in Career DevelopmentDigital Tools & AI in Career Development
Digital Tools & AI in Career DevelopmentMahmoud Rabie
 
Design pattern talk by Kaya Weers - 2024 (v2)
Design pattern talk by Kaya Weers - 2024 (v2)Design pattern talk by Kaya Weers - 2024 (v2)
Design pattern talk by Kaya Weers - 2024 (v2)Kaya Weers
 
JET Technology Labs White Paper for Virtualized Security and Encryption Techn...
JET Technology Labs White Paper for Virtualized Security and Encryption Techn...JET Technology Labs White Paper for Virtualized Security and Encryption Techn...
JET Technology Labs White Paper for Virtualized Security and Encryption Techn...amber724300
 
Microsoft 365 Copilot: How to boost your productivity with AI – Part one: Ado...
Microsoft 365 Copilot: How to boost your productivity with AI – Part one: Ado...Microsoft 365 Copilot: How to boost your productivity with AI – Part one: Ado...
Microsoft 365 Copilot: How to boost your productivity with AI – Part one: Ado...Nikki Chapple
 
Bridging Between CAD & GIS: 6 Ways to Automate Your Data Integration
Bridging Between CAD & GIS:  6 Ways to Automate Your Data IntegrationBridging Between CAD & GIS:  6 Ways to Automate Your Data Integration
Bridging Between CAD & GIS: 6 Ways to Automate Your Data Integrationmarketing932765
 
Accelerating Enterprise Software Engineering with Platformless
Accelerating Enterprise Software Engineering with PlatformlessAccelerating Enterprise Software Engineering with Platformless
Accelerating Enterprise Software Engineering with PlatformlessWSO2
 
Kuma Meshes Part I - The basics - A tutorial
Kuma Meshes Part I - The basics - A tutorialKuma Meshes Part I - The basics - A tutorial
Kuma Meshes Part I - The basics - A tutorialJoão Esperancinha
 
So einfach geht modernes Roaming fuer Notes und Nomad.pdf
So einfach geht modernes Roaming fuer Notes und Nomad.pdfSo einfach geht modernes Roaming fuer Notes und Nomad.pdf
So einfach geht modernes Roaming fuer Notes und Nomad.pdfpanagenda
 
A Framework for Development in the AI Age
A Framework for Development in the AI AgeA Framework for Development in the AI Age
A Framework for Development in the AI AgeCprime
 
QMMS Lesson 2 - Using MS Excel Formula.pdf
QMMS Lesson 2 - Using MS Excel Formula.pdfQMMS Lesson 2 - Using MS Excel Formula.pdf
QMMS Lesson 2 - Using MS Excel Formula.pdfROWELL MARQUINA
 
Genislab builds better products and faster go-to-market with Lean project man...
Genislab builds better products and faster go-to-market with Lean project man...Genislab builds better products and faster go-to-market with Lean project man...
Genislab builds better products and faster go-to-market with Lean project man...Farhan Tariq
 
[Webinar] SpiraTest - Setting New Standards in Quality Assurance
[Webinar] SpiraTest - Setting New Standards in Quality Assurance[Webinar] SpiraTest - Setting New Standards in Quality Assurance
[Webinar] SpiraTest - Setting New Standards in Quality AssuranceInflectra
 
Infrared simulation and processing on Nvidia platforms
Infrared simulation and processing on Nvidia platformsInfrared simulation and processing on Nvidia platforms
Infrared simulation and processing on Nvidia platformsYoss Cohen
 
Email Marketing Automation for Bonterra Impact Management (fka Social Solutio...
Email Marketing Automation for Bonterra Impact Management (fka Social Solutio...Email Marketing Automation for Bonterra Impact Management (fka Social Solutio...
Email Marketing Automation for Bonterra Impact Management (fka Social Solutio...Jeffrey Haguewood
 
Generative AI - Gitex v1Generative AI - Gitex v1.pptx
Generative AI - Gitex v1Generative AI - Gitex v1.pptxGenerative AI - Gitex v1Generative AI - Gitex v1.pptx
Generative AI - Gitex v1Generative AI - Gitex v1.pptxfnnc6jmgwh
 
Français Patch Tuesday - Avril
Français Patch Tuesday - AvrilFrançais Patch Tuesday - Avril
Français Patch Tuesday - AvrilIvanti
 

Recently uploaded (20)

Connecting the Dots for Information Discovery.pdf
Connecting the Dots for Information Discovery.pdfConnecting the Dots for Information Discovery.pdf
Connecting the Dots for Information Discovery.pdf
 
React Native vs Ionic - The Best Mobile App Framework
React Native vs Ionic - The Best Mobile App FrameworkReact Native vs Ionic - The Best Mobile App Framework
React Native vs Ionic - The Best Mobile App Framework
 
Testing tools and AI - ideas what to try with some tool examples
Testing tools and AI - ideas what to try with some tool examplesTesting tools and AI - ideas what to try with some tool examples
Testing tools and AI - ideas what to try with some tool examples
 
Digital Tools & AI in Career Development
Digital Tools & AI in Career DevelopmentDigital Tools & AI in Career Development
Digital Tools & AI in Career Development
 
Design pattern talk by Kaya Weers - 2024 (v2)
Design pattern talk by Kaya Weers - 2024 (v2)Design pattern talk by Kaya Weers - 2024 (v2)
Design pattern talk by Kaya Weers - 2024 (v2)
 
JET Technology Labs White Paper for Virtualized Security and Encryption Techn...
JET Technology Labs White Paper for Virtualized Security and Encryption Techn...JET Technology Labs White Paper for Virtualized Security and Encryption Techn...
JET Technology Labs White Paper for Virtualized Security and Encryption Techn...
 
Microsoft 365 Copilot: How to boost your productivity with AI – Part one: Ado...
Microsoft 365 Copilot: How to boost your productivity with AI – Part one: Ado...Microsoft 365 Copilot: How to boost your productivity with AI – Part one: Ado...
Microsoft 365 Copilot: How to boost your productivity with AI – Part one: Ado...
 
Bridging Between CAD & GIS: 6 Ways to Automate Your Data Integration
Bridging Between CAD & GIS:  6 Ways to Automate Your Data IntegrationBridging Between CAD & GIS:  6 Ways to Automate Your Data Integration
Bridging Between CAD & GIS: 6 Ways to Automate Your Data Integration
 
Accelerating Enterprise Software Engineering with Platformless
Accelerating Enterprise Software Engineering with PlatformlessAccelerating Enterprise Software Engineering with Platformless
Accelerating Enterprise Software Engineering with Platformless
 
Kuma Meshes Part I - The basics - A tutorial
Kuma Meshes Part I - The basics - A tutorialKuma Meshes Part I - The basics - A tutorial
Kuma Meshes Part I - The basics - A tutorial
 
So einfach geht modernes Roaming fuer Notes und Nomad.pdf
So einfach geht modernes Roaming fuer Notes und Nomad.pdfSo einfach geht modernes Roaming fuer Notes und Nomad.pdf
So einfach geht modernes Roaming fuer Notes und Nomad.pdf
 
A Framework for Development in the AI Age
A Framework for Development in the AI AgeA Framework for Development in the AI Age
A Framework for Development in the AI Age
 
How Tech Giants Cut Corners to Harvest Data for A.I.
How Tech Giants Cut Corners to Harvest Data for A.I.How Tech Giants Cut Corners to Harvest Data for A.I.
How Tech Giants Cut Corners to Harvest Data for A.I.
 
QMMS Lesson 2 - Using MS Excel Formula.pdf
QMMS Lesson 2 - Using MS Excel Formula.pdfQMMS Lesson 2 - Using MS Excel Formula.pdf
QMMS Lesson 2 - Using MS Excel Formula.pdf
 
Genislab builds better products and faster go-to-market with Lean project man...
Genislab builds better products and faster go-to-market with Lean project man...Genislab builds better products and faster go-to-market with Lean project man...
Genislab builds better products and faster go-to-market with Lean project man...
 
[Webinar] SpiraTest - Setting New Standards in Quality Assurance
[Webinar] SpiraTest - Setting New Standards in Quality Assurance[Webinar] SpiraTest - Setting New Standards in Quality Assurance
[Webinar] SpiraTest - Setting New Standards in Quality Assurance
 
Infrared simulation and processing on Nvidia platforms
Infrared simulation and processing on Nvidia platformsInfrared simulation and processing on Nvidia platforms
Infrared simulation and processing on Nvidia platforms
 
Email Marketing Automation for Bonterra Impact Management (fka Social Solutio...
Email Marketing Automation for Bonterra Impact Management (fka Social Solutio...Email Marketing Automation for Bonterra Impact Management (fka Social Solutio...
Email Marketing Automation for Bonterra Impact Management (fka Social Solutio...
 
Generative AI - Gitex v1Generative AI - Gitex v1.pptx
Generative AI - Gitex v1Generative AI - Gitex v1.pptxGenerative AI - Gitex v1Generative AI - Gitex v1.pptx
Generative AI - Gitex v1Generative AI - Gitex v1.pptx
 
Français Patch Tuesday - Avril
Français Patch Tuesday - AvrilFrançais Patch Tuesday - Avril
Français Patch Tuesday - Avril
 

Chlorine measurements over a larger concentration range with low maintenance effort

  • 1. Introduction The analysis of chlorine has become more and more important in the recent years. The use of substitute fuel in cement plants, brickyards and other industries with high demand for energy requires easy, low maintenance chlorine analysis. Traditionally chlorine determination is done using Schöninger digestion, which is very time consuming and difficult to automate. Therefore, high temperature combustion followed by coulometric detection of the formed HCl has gained more and more interest. Still widely used, however, the coulometric method has several major drawbacks. It requires a substantial amount of maintenance effort due to the electrolyte solution needing to be replaced very frequently, and the electrodes are susceptible to wear out. Moreover, the dynamic range is limited to lower concentrations with this approach. Ion chromatography in combination with conductivity detection on the other hand is a more stable technique and allows the separate determination of all halogens. The drawback of this solution is the very high investment costs. Elementar as the pioneer of elemental analysis has developed a simpler alternative. The combination of the well-proven, state-of-the-art elemental analyzers with high performance solid state electrochemical detectors (ECD) makes chlorine analysis much easier, more reliable and more precise than ever before. It allows chlorine measurements over a larger concentration range with low maintenance effort. Unique high precision chlorine analysis for a large concentration range TECHNICAL NOTE UNIQUE CHLORINE OPTION trace SN cube rapid CS cube vario MICRO cube vario EL cube vario MACRO cube
  • 2. Measuring principle The sample is transferred from the sample carousel into the combustion tube through a blank-free ball valve. Unlike systems with a horizontal furnace this does not require difficult mechanics and/or compressed air to feed the samples into the combustion tube. The sample then enters the high-temperature combustion zone, which ensures 100% transformation of all chlorine to HCl. In the next step water is efficiently removed from the carrier gas to avoid condensation and loss of HCl. The carrier gas finally passes the solid-state electrochemical detector to detect the HCl (see Figure 1). This default configuration is available for the rapid CS cube, vario MICRO cube, vario EL cube and vario MACRO cube models. In some cases the ECD cell shows a slight cross sensitivity to CO2, resulting in a small negative peak at the beginning of the integration. For higher Cl concentrations (> 100 ppm) this does not significantly contribute to the resulting chlorine concentration. For low concentrations as in gasoline this small negative peak may influence the results. For this reason the setup of the chlorine option of the trace SN cube, which is optimized for low Cl contents, is slightly different (see Figure 2). After removing the moisture from the carrier gas, the HCl is trapped on a special trap column. After the combustion process and removal of interferences, the trap column heats to release HCl, which is then detected in the solid-state electrochemical detector. The Advanced Purge and Trap technology leads to peak focusing that significantly improves the signal-to-noise ratio and guarantees highest sensitivity. Evaluation of the detector signal is done on a standard PC. The entire process is fully automated for easy use and overnight operation. Combustion Drying Tube Detection Sample Carousel Blank-Free Ball Valve CombustionTube Furnace ECD Combustion Detection Sample Carousel Blank-Free Ball Valve CombustionTube Furnace ECD High Sens Gas Separation Figure 1. Functional principle of the chlorine detection option in the default configuration Figure 2. Functional principle of the trace SN cube with the chlorine detection option for low concentrations
  • 3. Figure 5. HCl peak of a diesel sample (+ 5 ppm Cl) analyzed using the trace SN cube with chlorine detection option and 20 ppm measuring cell. Electrochemical detection The following reactions occur in the electrochemical cell: working electrode: Ag + HCl → AgCl + H+ + e reference electrode: O2 + 4e + 4H+ → 2H2O Counter Electrode Reference Electrode Sensing Electrode Hydrophobic Membrane Electrolyte Capillary Diffusion Barrier Calibration There are several standards that can be used for calibration. If no capsule press is available ammonium chloride (Cl content 65.97%) is the substance of choice. Moreover, the high solubility of chlorine in water is ideal for preparing standard solutions. Another possible calibration standard could be oil with a defined amount of Cl or if available matrix specific chlorine standards. Usually these standards can be used without accelerants. Figure 6, for example, shows a typical calibration curve, obtained by a coal standard with 0.11% Cl on the rapid CS cube with Cl-option. Figure 3. Functional principle of the EC cell Figure 4. HCl peak of a 1.4 mg NH4Cl sample analyzed using the rapid CS cube with chlorine detection option and 5000 ppm measuring cell. The electrons produced in this reaction lower the current between working and reference electrode, which is a measure for the HCl concentration in the carrier gas. This detection method can be described as amperometric detection. The capillary diffusion barrier is basically a hole whose diameter determines the sensitivity of the cell. The electrolyte is a concentrated, non-oxidizing acid. A schematic diagram of the electrochemical cell is shown in Figure 3. Chemistry dictates that the ECD cell can also be affected by high concentrations of HBr, but this is rarely seen in practice. Cl (µg) 44.3 0.0 0.0 Peakarea (AU) 20745 y = 5.5571.10⁻⁴*x + 2.1391.10⁻⁶ r = 0.996 Figure 6. Calibration curve EC cell for chlorine 0 100 200 300 400 500 600 24000 20000 16000 12000 8000 4000 0 EC/Cl [unit] [sec.] EC/Cl [unit] HCl column [°C] - 300 - 0 - 50 - 100 - 150 - 200 - 250 0 100 200 300 400 500 600 - 2000 - 1600 - 1200 - 800 - 400 - 0 [sec.]
  • 4. Typical applications The chlorine detection option is optimized for applications in the area of fuel, residual fuel, biomass, waste, soil, along with others. Due to limitations in precision and accuracy, it is not recommended to use the Cl-option for classical elemental analysis (e.g. to determine the Cl content in pure chemicals). The following table shows some typical applications for the Cl-option. Elementar Analysensysteme GmbH | Donaustraße 7 · 63452 Hanau (Germany) Phone: +49 (0) 6181 9100-0 | info@elementar.de | www.elementar.de Art.-No.05005032,01/2016A SUBSTANCE WEIGHT (mg) DETECTOR Cl (W-%) REL. SD (%) ANTHRACIT-1 34 200 0.151 1.52 ANTHRACIT-2 33 200 0.021 6.1 WOOD COAL 60 200 0.0045 8.9 BIOMASS-1 15 200 0.133 4.2 BIOMASS-2 30 200 0.0086 8.9 ORGANIC WASTE 20 200 0.142 4.3 RESIDUAL FUEL 31 5000 0.893 6.1 PLASTIC FIBER 10 200 0.354 11.3 DIESEL SPIKED WITH 1 PPM Cl 35 20 1.07 8.8 Available EC cells Elementar provides three different types of electrochemical cells depending on the concentration range. The names of the cells refer to the maximum HCl concentration in the gas stream which can be analyzed. Note that this concentration cannot directly be transferred to the Cl content of the sample. The characteristics of the different electrochemical cells are presented in the following table. CELL TYPE LOWER LIMIT (µg Cl abs.) UPPER LIMIT (µg Cl abs.) REL. SD Cl (%)* 20 PPM† 0.05 2 8 200 PPM 0.5 20 5 5000 PPM 15 1200 5 *: obtained on Conostan standards in the middle of the measuring range † : available for the trace SN cube only Elementargroup – your partner for elemental analysis Elementargroup is the world leader in high performance analysis of organic ele- ments. Continuous innovation, creative solutions and comprehensive support form the foundation of the Elementar and Isoprime brands ensuring our pro- ducts continue to advance science across agriculture, chemical, environmental, energy, materials and forensics markets in more than 80 countries.