Distortion artifacts in MRI and
their correction
Vibha Chaswal, Ph.D.
MRI Gradient fields
ā€¢ The gradient fields are superimposed over the static
magnetic field. For transverse images:
ā€¢ Z gradientā€“selects the slice, and is applied whilst the 90Ā° RF
pulse is on
ā€¢ X gradientā€“creates a frequency change in emitted signal
producing lines (or columns) of data, and is applied when
the echo RF is being measured
ā€¢ Y gradientā€“produces a change in the phase of rotation in
different voxels, and is applied between the 90Ā° and 180Ā°
RF pulses at a different strength during each cycle. The
number of phase steps (e.g. 256) affects scan time.
ā€¢ ā€œFrequency encodingā€ and ā€œPhase encodingā€ directions (X
and Y) are interchangeable.
Gradient spatial encoding
Pulse sequencing diagram
MRI distortion Artifacts
A whole rangeā€¦ā€¦..
Equipment-related
ā€¢ Non-uniformity of signal
ā€¢ Wraparound (phase wrap,
aliasing)
ā€¢ Peripheral signal artifacts
ā€¢ Slice overlap (cross-excitation)
ā€¢ Noise and RF interference
ā€¢ Truncation (ringing, Gibbs
artifact)
ā€¢ Phase smearing

Patient-related
ā€¢ Signal void due to freezing
ā€¢ Partial volume averaging
ā€¢ Chemical shift
ā€¢ Motion artifacts (blurring,
ghosting and pulsatility)
ā€¢ Flow artifacts
ā€¢ Susceptibility and image
distortion
ā€¢ Magic angle
B0 inhomogeneity and susceptibility
ā€¢ Extremely homogeneous static magnetic field is required
around isocenter of the magnet for MRI.
ā€¢ B0 homogeneity of the empty magnet:
typical value of 1ppm (0.001%) in a 50 cm diameter
spherical volume.
ā€¢ susceptibility = local changes in Bo due to a substanceā€™s
own magnetic properties
ā€¢ Artifacts occur at interfaces between substances of different
susceptibilities, e.g. air/bone/metal/ calcification/hemorrhage;
implants and previous surgery, tooth fillings, mascara, ear-rings,
hair-slides
ā€“ worse at higher field strengths
B0 inhomogeneity
ā€¢ Decreased frequency-selective fat-saturation efficacy of pulses
ā€¢ Effect more pronounced the further the image slice is positioned from
the isocenter of the magnet

(a) off-center acquisition with insufficient fat
suppression (arrows)

(b) Identical acquisition at isocenter

Reduction:
improve field shimming
Optimal positioning of the patient in z-direction
Dietrich et. al., Artifacts in 3-T MRI: Physical background and reduction strategies, EJR, 65, 2008
B0 inhomogeneity
ā€¢

Relatively small frequency shifts lead to substantial reduction of the
transversal steady-state magnetization and, thus, to band-shaped signal losses
in the image

Reduction:
Move the artifact out of the region of interest (apply frequency offset to nominal
resonant frequency)
Dietrich et. al., Artifacts in 3-T MRI: Physical background and reduction strategies, EJR, 65, 2008
Susceptibility: Metal artifacts
ā€¢

An important manifestation of a susceptibility-related artifact is the signal
loss in gradient-echo images around metallic implants or surgical clips

Right hip prosthesis is seen as region of signal drop
out (closed arrows)
Figure from: Mechlin et al, AJR 143, December 1984

STIR axial MRI image of a patient, arrows B point to
artifacts due to right hip prosthesis.
Figure from: Dr. Geller et al, Malignant Peripheral Nerve
Sheath Tumors (MPNST), ESUN, june 2006, 3(3)
http://sarcomahelp.org/learning_center/mpnst.html

Reduction:
Decrease Echo time, TE, of pulse sequence
Increase receiver band-width
Susceptibility: Geometric distortion
artifact
ā€¢
ā€¢

Usually in echo-planar imaging (EPI)
At interfaces between soft-tissue and bone/air e.g., in the
base of the skull or in the head-and-neck region

Reduction:
a) decrease the echo-spacing of the
read out train e.g., by increasing
the receiver bandwidth or by
parallel imaging techniques
b) Use fast-spin echo technique

(b) EPI acquisition without parallel imaging, severe distortion
artifacts are visible (arrows) and (c) EPI acquisition with
acceleration factor 2, distortion artifacts are still present but
considerably reduced (arrows).

Dietrich et. al., Artifacts in 3-T MRI: Physical background and reduction strategies, EJR, 65,
2008
B1 inhomogeneity
ā€¢ Spatial inhomogeneity of the B1 field results in
flip-angle deviations depending on the spatial
position
ā€¢ Reduced signal intensity in these areas or to
altered contrast particularly in FLASH sequences
whose T1-weighting depends on the flip angle
Dielectric resonance effects/ RF
interference/ standing wave effect
ā€¢

Constructive or destructive interferences of the transmitted RF field may be observed
resulting in either regional (e.g., central) brightening or regional signal loss,
respectively.

Signal loss due to wavelength effect
Reduction:
a) Effect can be mitigated by positioning a dielectric cushion close to the ROI, or
b) Manually modify the RF-transmitter amplitude in order to reduce B1-inhomogeneityinduced signal loss
Chemical Shift Artifact
ā€¢ For MRI in vivo, two important groups of
molecules : protons in water and protons in
fat tissue.
ā€¢ Larmor frequencies difference of these
protons ~ 3.5 ppm.
ā€¢ chemical shift leads to a slight geometric shift
of the relative position of water and fat
protons in readout direction.
Chemical Shift Artifact Reduction

image-compromising superposition of
fat tissue onto other tissues
Reduction:
a)manually increase the RF-transmitter amplitude and apply image post-processing
filters to obtain more uniform image intensities.
b)Increase (~double) receiver bandwidth
c)Based on anatomy switch readout and phase encoding directions
Blood Flow related artifact
ā€¢ Magnetohydrodynamic effect: The flow of blood ions
perpendicular to the strong static magnetic field, B0, gives rise to
induced voltages and currents;

ā€œ(a) Signal loss in the pulmonary vessels in non-gated 3-T single-shot fast-spin-echo
(FSE) acquisition possibly related to the magnetohydrodynamic effect. (b) Non-gated
1.5-T single-shot FSE acquisition demonstrating visual signal in pulmonary vessels. (c)
ECG-gated 3-T single-shot FSE acquisition of the same volunteer as in (a) with data
acquisition during diastole; the vessel signal is restored.ā€ ā€¦ā€¦.Deitrich et al, EJR
65(2008)
SNR related artifacts
ā€¢ Artifacts to Noise ratio (ANR)
ā€¢ Artifacts are masked by increased statistical
noise.
ā€¢ Gibbs Ringing: caused by data clipping at the
edges of K-space (raw data readings are still over
noise level at the borders of acquired K-space)
Gibbs artifact

Reduction:
Increase spatial resolution or,
Apply reconstruction filters (Hanning filter)
Resources
Many more Artifacts but time restriction
ā‡’Resources for MR Artifacts ad their reduction
ā‡’MR-TIP: online database (open forum) from University
of Illinois, Excellent resource
ā‡’ Review article by Dietrich et.al,ā€Artifacts in 3-T MRI:
Physical Background and reduction strategiesā€ EJR, Vol
65, 2008
Thank You

More Related Content

What's hot

Contrast media in CT
Contrast media in CTContrast media in CT
Contrast media in CT
Kyle Rousseau
Ā 
CT Physics
CT PhysicsCT Physics
CT Physics
RMLIMS
Ā 
MRI CONTRAST AGENT.pdf
MRI CONTRAST AGENT.pdfMRI CONTRAST AGENT.pdf
MRI CONTRAST AGENT.pdf
KrishnaPrints
Ā 

What's hot (20)

Gradient Recalled Echo(GRE)
Gradient Recalled Echo(GRE)Gradient Recalled Echo(GRE)
Gradient Recalled Echo(GRE)
Ā 
Gradient echo pulse sequence and its application
Gradient echo pulse sequence and its applicationGradient echo pulse sequence and its application
Gradient echo pulse sequence and its application
Ā 
MRI Artefacts
MRI ArtefactsMRI Artefacts
MRI Artefacts
Ā 
Mri artifacts gamal mahdaly
Mri artifacts gamal mahdalyMri artifacts gamal mahdaly
Mri artifacts gamal mahdaly
Ā 
Contrast media in CT
Contrast media in CTContrast media in CT
Contrast media in CT
Ā 
CT Physics
CT PhysicsCT Physics
CT Physics
Ā 
MRI spin echo pulse sequences
MRI spin echo pulse sequencesMRI spin echo pulse sequences
MRI spin echo pulse sequences
Ā 
MRI Parameters for Imaging
MRI Parameters for ImagingMRI Parameters for Imaging
MRI Parameters for Imaging
Ā 
magnetic resonance in angiography
magnetic resonance in  angiography magnetic resonance in  angiography
magnetic resonance in angiography
Ā 
Mr hazard protection
Mr hazard protectionMr hazard protection
Mr hazard protection
Ā 
CT Artifacts
CT ArtifactsCT Artifacts
CT Artifacts
Ā 
MRI CONTRAST AGENT.pdf
MRI CONTRAST AGENT.pdfMRI CONTRAST AGENT.pdf
MRI CONTRAST AGENT.pdf
Ā 
Encoding and image formation
Encoding and image formationEncoding and image formation
Encoding and image formation
Ā 
Mri contrast
Mri contrastMri contrast
Mri contrast
Ā 
Mri sequences
Mri sequencesMri sequences
Mri sequences
Ā 
Mri spin echo pulse sequences its variations and
Mri spin echo pulse sequences its variations andMri spin echo pulse sequences its variations and
Mri spin echo pulse sequences its variations and
Ā 
Artifacts in MRI
Artifacts in MRIArtifacts in MRI
Artifacts in MRI
Ā 
Portable ct mobile ct
Portable ct mobile ctPortable ct mobile ct
Portable ct mobile ct
Ā 
MRI Artifact Radiology
MRI Artifact RadiologyMRI Artifact Radiology
MRI Artifact Radiology
Ā 
Ct quality control
Ct quality controlCt quality control
Ct quality control
Ā 

Viewers also liked

Special Procedures: TBI, TSET and IORT
Special Procedures: TBI, TSET and IORTSpecial Procedures: TBI, TSET and IORT
Special Procedures: TBI, TSET and IORT
Miami Cancer Institute
Ā 
Cardiac MRI
Cardiac MRICardiac MRI
Cardiac MRI
alisajjad20
Ā 
Exposure factors2
Exposure factors2Exposure factors2
Exposure factors2
mr_koky
Ā 

Viewers also liked (20)

Cardiac MRI
Cardiac MRICardiac MRI
Cardiac MRI
Ā 
Gre&SE
Gre&SEGre&SE
Gre&SE
Ā 
Cardiac mri
Cardiac mriCardiac mri
Cardiac mri
Ā 
Mr spectroscopy
Mr spectroscopyMr spectroscopy
Mr spectroscopy
Ā 
Factors controlling the beam of x-ray
Factors controlling the beam of x-rayFactors controlling the beam of x-ray
Factors controlling the beam of x-ray
Ā 
Special Procedures: TBI, TSET and IORT
Special Procedures: TBI, TSET and IORTSpecial Procedures: TBI, TSET and IORT
Special Procedures: TBI, TSET and IORT
Ā 
Cardiac MRI principle
Cardiac MRI principleCardiac MRI principle
Cardiac MRI principle
Ā 
What is spatial Resolution
What is spatial ResolutionWhat is spatial Resolution
What is spatial Resolution
Ā 
Radiographic factor
Radiographic factorRadiographic factor
Radiographic factor
Ā 
Advances in Brachytherapy Treatment Planning and Delivery
Advances in Brachytherapy Treatment Planning and DeliveryAdvances in Brachytherapy Treatment Planning and Delivery
Advances in Brachytherapy Treatment Planning and Delivery
Ā 
Mri physics-uk
Mri physics-ukMri physics-uk
Mri physics-uk
Ā 
Contrast based artifacts
Contrast based artifactsContrast based artifacts
Contrast based artifacts
Ā 
Fluroscopy
Fluroscopy Fluroscopy
Fluroscopy
Ā 
Dark room equipments and entrance
Dark room equipments and entranceDark room equipments and entrance
Dark room equipments and entrance
Ā 
Cardiac MRI
Cardiac MRICardiac MRI
Cardiac MRI
Ā 
Mr fluoroscopy
Mr fluoroscopyMr fluoroscopy
Mr fluoroscopy
Ā 
Nguyen khoi viet cardiac mri for the evaluation of ischemic heart disease jfi...
Nguyen khoi viet cardiac mri for the evaluation of ischemic heart disease jfi...Nguyen khoi viet cardiac mri for the evaluation of ischemic heart disease jfi...
Nguyen khoi viet cardiac mri for the evaluation of ischemic heart disease jfi...
Ā 
Pns New
Pns NewPns New
Pns New
Ā 
076 cardiac magnetic resonance imaging
076 cardiac magnetic resonance imaging076 cardiac magnetic resonance imaging
076 cardiac magnetic resonance imaging
Ā 
Exposure factors2
Exposure factors2Exposure factors2
Exposure factors2
Ā 

Similar to Distortion Artifacts in MRI and their correction

ppMRIARTIFACTS.pptx
ppMRIARTIFACTS.pptxppMRIARTIFACTS.pptx
ppMRIARTIFACTS.pptx
Shubham661884
Ā 
MRIARTIFACTS.pptx
MRIARTIFACTS.pptxMRIARTIFACTS.pptx
MRIARTIFACTS.pptx
Shubham661884
Ā 
DIFFERENT TYPES OF RADIOGRAPHY CT_ARTIFACTS
DIFFERENT TYPES OF RADIOGRAPHY CT_ARTIFACTSDIFFERENT TYPES OF RADIOGRAPHY CT_ARTIFACTS
DIFFERENT TYPES OF RADIOGRAPHY CT_ARTIFACTS
MichaelAkortsu1
Ā 
Echo physics and instrumentation
Echo physics and instrumentationEcho physics and instrumentation
Echo physics and instrumentation
guru here
Ā 
artifactsusgctmriradiologypk-190830131406.pptx
artifactsusgctmriradiologypk-190830131406.pptxartifactsusgctmriradiologypk-190830131406.pptx
artifactsusgctmriradiologypk-190830131406.pptx
Shubham661884
Ā 
TGS Arcis- Canada Curvature Attributes and Applications Poster
TGS Arcis- Canada Curvature Attributes and Applications PosterTGS Arcis- Canada Curvature Attributes and Applications Poster
TGS Arcis- Canada Curvature Attributes and Applications Poster
TGS
Ā 
My Presentation @Ryerson University
My Presentation @Ryerson UniversityMy Presentation @Ryerson University
My Presentation @Ryerson University
Jagdish Bhatt
Ā 

Similar to Distortion Artifacts in MRI and their correction (20)

mri artefacts.pptx
mri artefacts.pptxmri artefacts.pptx
mri artefacts.pptx
Ā 
ppMRIARTIFACTS.pptx
ppMRIARTIFACTS.pptxppMRIARTIFACTS.pptx
ppMRIARTIFACTS.pptx
Ā 
Mri artifacts
Mri artifactsMri artifacts
Mri artifacts
Ā 
X-ray.ppt
X-ray.pptX-ray.ppt
X-ray.ppt
Ā 
ppMRIARTIFACTS.pptx
ppMRIARTIFACTS.pptxppMRIARTIFACTS.pptx
ppMRIARTIFACTS.pptx
Ā 
Swi seminar final
Swi seminar finalSwi seminar final
Swi seminar final
Ā 
Recent Advances In MRI
Recent Advances In MRIRecent Advances In MRI
Recent Advances In MRI
Ā 
MRIARTIFACTS.pptx
MRIARTIFACTS.pptxMRIARTIFACTS.pptx
MRIARTIFACTS.pptx
Ā 
Technical aspects of dental CBCT - Dr. Kavan Gandhi
Technical aspects of dental CBCT - Dr. Kavan GandhiTechnical aspects of dental CBCT - Dr. Kavan Gandhi
Technical aspects of dental CBCT - Dr. Kavan Gandhi
Ā 
DIFFERENT TYPES OF RADIOGRAPHY CT_ARTIFACTS
DIFFERENT TYPES OF RADIOGRAPHY CT_ARTIFACTSDIFFERENT TYPES OF RADIOGRAPHY CT_ARTIFACTS
DIFFERENT TYPES OF RADIOGRAPHY CT_ARTIFACTS
Ā 
Nuclear Magnification Resonance Imaging
Nuclear Magnification Resonance ImagingNuclear Magnification Resonance Imaging
Nuclear Magnification Resonance Imaging
Ā 
Tissue harmonic imaging
Tissue harmonic imaging Tissue harmonic imaging
Tissue harmonic imaging
Ā 
Echo physics and instrumentation
Echo physics and instrumentationEcho physics and instrumentation
Echo physics and instrumentation
Ā 
Ultasonic testing
Ultasonic testingUltasonic testing
Ultasonic testing
Ā 
Navneet.mr hardware & artifacts related to it
Navneet.mr hardware & artifacts related to itNavneet.mr hardware & artifacts related to it
Navneet.mr hardware & artifacts related to it
Ā 
Ct image artifacts
Ct image artifactsCt image artifacts
Ct image artifacts
Ā 
Physics basic
Physics basicPhysics basic
Physics basic
Ā 
artifactsusgctmriradiologypk-190830131406.pptx
artifactsusgctmriradiologypk-190830131406.pptxartifactsusgctmriradiologypk-190830131406.pptx
artifactsusgctmriradiologypk-190830131406.pptx
Ā 
TGS Arcis- Canada Curvature Attributes and Applications Poster
TGS Arcis- Canada Curvature Attributes and Applications PosterTGS Arcis- Canada Curvature Attributes and Applications Poster
TGS Arcis- Canada Curvature Attributes and Applications Poster
Ā 
My Presentation @Ryerson University
My Presentation @Ryerson UniversityMy Presentation @Ryerson University
My Presentation @Ryerson University
Ā 

More from Miami Cancer Institute

IGRT: MVCBCT Calibration and Acceptance testing Procedure
IGRT: MVCBCT Calibration and Acceptance testing ProcedureIGRT: MVCBCT Calibration and Acceptance testing Procedure
IGRT: MVCBCT Calibration and Acceptance testing Procedure
Miami Cancer Institute
Ā 
Ion chambers Acceptance Testing and Evaluation
Ion chambers Acceptance Testing and EvaluationIon chambers Acceptance Testing and Evaluation
Ion chambers Acceptance Testing and Evaluation
Miami Cancer Institute
Ā 
Linear Accelerator Acceptance, Commissioning and Annual QA
Linear Accelerator Acceptance, Commissioning and Annual QALinear Accelerator Acceptance, Commissioning and Annual QA
Linear Accelerator Acceptance, Commissioning and Annual QA
Miami Cancer Institute
Ā 
Stereotactic Radiosurgery at UIHC
Stereotactic Radiosurgery at UIHCStereotactic Radiosurgery at UIHC
Stereotactic Radiosurgery at UIHC
Miami Cancer Institute
Ā 
TPS modeling and Beam commissioning at UIHC Radiation Oncology
TPS modeling and Beam commissioning at UIHC Radiation OncologyTPS modeling and Beam commissioning at UIHC Radiation Oncology
TPS modeling and Beam commissioning at UIHC Radiation Oncology
Miami Cancer Institute
Ā 
Radiotherapy Treatment Simulation
Radiotherapy Treatment SimulationRadiotherapy Treatment Simulation
Radiotherapy Treatment Simulation
Miami Cancer Institute
Ā 
Image Contrast, Noise, Resolution
Image Contrast, Noise, ResolutionImage Contrast, Noise, Resolution
Image Contrast, Noise, Resolution
Miami Cancer Institute
Ā 
Mutual Information Algorithm applied to rigid registration
Mutual Information Algorithm applied to rigid registrationMutual Information Algorithm applied to rigid registration
Mutual Information Algorithm applied to rigid registration
Miami Cancer Institute
Ā 
Imaging for Radiotherapy delivery and verification
Imaging for Radiotherapy delivery and verificationImaging for Radiotherapy delivery and verification
Imaging for Radiotherapy delivery and verification
Miami Cancer Institute
Ā 
TG-51 and Clinical Monitor Unit Calculations
TG-51 and Clinical Monitor Unit CalculationsTG-51 and Clinical Monitor Unit Calculations
TG-51 and Clinical Monitor Unit Calculations
Miami Cancer Institute
Ā 

More from Miami Cancer Institute (14)

IGRT: MVCBCT Calibration and Acceptance testing Procedure
IGRT: MVCBCT Calibration and Acceptance testing ProcedureIGRT: MVCBCT Calibration and Acceptance testing Procedure
IGRT: MVCBCT Calibration and Acceptance testing Procedure
Ā 
Ion chambers Acceptance Testing and Evaluation
Ion chambers Acceptance Testing and EvaluationIon chambers Acceptance Testing and Evaluation
Ion chambers Acceptance Testing and Evaluation
Ā 
ArcCheck 3D-Diode array evaluation, commissioning, testing for VMAT QA
ArcCheck 3D-Diode array evaluation, commissioning, testing for VMAT QAArcCheck 3D-Diode array evaluation, commissioning, testing for VMAT QA
ArcCheck 3D-Diode array evaluation, commissioning, testing for VMAT QA
Ā 
Linear Accelerator Acceptance, Commissioning and Annual QA
Linear Accelerator Acceptance, Commissioning and Annual QALinear Accelerator Acceptance, Commissioning and Annual QA
Linear Accelerator Acceptance, Commissioning and Annual QA
Ā 
Shielding Design
Shielding DesignShielding Design
Shielding Design
Ā 
Stereotactic Radiosurgery at UIHC
Stereotactic Radiosurgery at UIHCStereotactic Radiosurgery at UIHC
Stereotactic Radiosurgery at UIHC
Ā 
TPS modeling and Beam commissioning at UIHC Radiation Oncology
TPS modeling and Beam commissioning at UIHC Radiation OncologyTPS modeling and Beam commissioning at UIHC Radiation Oncology
TPS modeling and Beam commissioning at UIHC Radiation Oncology
Ā 
Radiotherapy Treatment Simulation
Radiotherapy Treatment SimulationRadiotherapy Treatment Simulation
Radiotherapy Treatment Simulation
Ā 
Image Contrast, Noise, Resolution
Image Contrast, Noise, ResolutionImage Contrast, Noise, Resolution
Image Contrast, Noise, Resolution
Ā 
Mutual Information Algorithm applied to rigid registration
Mutual Information Algorithm applied to rigid registrationMutual Information Algorithm applied to rigid registration
Mutual Information Algorithm applied to rigid registration
Ā 
Causes of Noise in PET imaging
Causes of Noise in PET imagingCauses of Noise in PET imaging
Causes of Noise in PET imaging
Ā 
Imaging for Radiotherapy delivery and verification
Imaging for Radiotherapy delivery and verificationImaging for Radiotherapy delivery and verification
Imaging for Radiotherapy delivery and verification
Ā 
Clinical Reference Dosimetry
Clinical Reference DosimetryClinical Reference Dosimetry
Clinical Reference Dosimetry
Ā 
TG-51 and Clinical Monitor Unit Calculations
TG-51 and Clinical Monitor Unit CalculationsTG-51 and Clinical Monitor Unit Calculations
TG-51 and Clinical Monitor Unit Calculations
Ā 

Recently uploaded

Dehradun Call Girls Service {8854095900} ā¤ļøVVIP ROCKY Call Girl in Dehradun U...
Dehradun Call Girls Service {8854095900} ā¤ļøVVIP ROCKY Call Girl in Dehradun U...Dehradun Call Girls Service {8854095900} ā¤ļøVVIP ROCKY Call Girl in Dehradun U...
Dehradun Call Girls Service {8854095900} ā¤ļøVVIP ROCKY Call Girl in Dehradun U...
Sheetaleventcompany
Ā 
Ahmedabad Call Girls Book Now 8980367676 Top Class Ahmedabad Escort Service A...
Ahmedabad Call Girls Book Now 8980367676 Top Class Ahmedabad Escort Service A...Ahmedabad Call Girls Book Now 8980367676 Top Class Ahmedabad Escort Service A...
Ahmedabad Call Girls Book Now 8980367676 Top Class Ahmedabad Escort Service A...
Genuine Call Girls
Ā 
Premium Call Girls Nagpur {9xx000xx09} ā¤ļøVVIP POOJA Call Girls in Nagpur Maha...
Premium Call Girls Nagpur {9xx000xx09} ā¤ļøVVIP POOJA Call Girls in Nagpur Maha...Premium Call Girls Nagpur {9xx000xx09} ā¤ļøVVIP POOJA Call Girls in Nagpur Maha...
Premium Call Girls Nagpur {9xx000xx09} ā¤ļøVVIP POOJA Call Girls in Nagpur Maha...
Sheetaleventcompany
Ā 
ā¤ļøAmritsar Escorts Serviceā˜Žļø9815674956ā˜Žļø Call Girl service in Amritsarā˜Žļø Amri...
ā¤ļøAmritsar Escorts Serviceā˜Žļø9815674956ā˜Žļø Call Girl service in Amritsarā˜Žļø Amri...ā¤ļøAmritsar Escorts Serviceā˜Žļø9815674956ā˜Žļø Call Girl service in Amritsarā˜Žļø Amri...
ā¤ļøAmritsar Escorts Serviceā˜Žļø9815674956ā˜Žļø Call Girl service in Amritsarā˜Žļø Amri...
Sheetaleventcompany
Ā 
Independent Bangalore Call Girls (Adult Only) šŸ’ÆCall Us šŸ” 7304373326 šŸ” šŸ’ƒ Escor...
Independent Bangalore Call Girls (Adult Only) šŸ’ÆCall Us šŸ” 7304373326 šŸ” šŸ’ƒ Escor...Independent Bangalore Call Girls (Adult Only) šŸ’ÆCall Us šŸ” 7304373326 šŸ” šŸ’ƒ Escor...
Independent Bangalore Call Girls (Adult Only) šŸ’ÆCall Us šŸ” 7304373326 šŸ” šŸ’ƒ Escor...
Sheetaleventcompany
Ā 
Difference Between Skeletal Smooth and Cardiac Muscles
Difference Between Skeletal Smooth and Cardiac MusclesDifference Between Skeletal Smooth and Cardiac Muscles
Difference Between Skeletal Smooth and Cardiac Muscles
MedicoseAcademics
Ā 

Recently uploaded (20)

Kolkata Call Girls Shobhabazar šŸ’ÆCall Us šŸ” 8005736733 šŸ” šŸ’ƒ Top Class Call Gir...
Kolkata Call Girls Shobhabazar  šŸ’ÆCall Us šŸ” 8005736733 šŸ” šŸ’ƒ  Top Class Call Gir...Kolkata Call Girls Shobhabazar  šŸ’ÆCall Us šŸ” 8005736733 šŸ” šŸ’ƒ  Top Class Call Gir...
Kolkata Call Girls Shobhabazar šŸ’ÆCall Us šŸ” 8005736733 šŸ” šŸ’ƒ Top Class Call Gir...
Ā 
Call girls Service Phullen / 9332606886 Genuine Call girls with real Photos a...
Call girls Service Phullen / 9332606886 Genuine Call girls with real Photos a...Call girls Service Phullen / 9332606886 Genuine Call girls with real Photos a...
Call girls Service Phullen / 9332606886 Genuine Call girls with real Photos a...
Ā 
Dehradun Call Girls Service {8854095900} ā¤ļøVVIP ROCKY Call Girl in Dehradun U...
Dehradun Call Girls Service {8854095900} ā¤ļøVVIP ROCKY Call Girl in Dehradun U...Dehradun Call Girls Service {8854095900} ā¤ļøVVIP ROCKY Call Girl in Dehradun U...
Dehradun Call Girls Service {8854095900} ā¤ļøVVIP ROCKY Call Girl in Dehradun U...
Ā 
Call 8250092165 Patna Call Girls ā‚¹4.5k Cash Payment With Room Delivery
Call 8250092165 Patna Call Girls ā‚¹4.5k Cash Payment With Room DeliveryCall 8250092165 Patna Call Girls ā‚¹4.5k Cash Payment With Room Delivery
Call 8250092165 Patna Call Girls ā‚¹4.5k Cash Payment With Room Delivery
Ā 
Gastric Cancer: Š”linical Implementation of Artificial Intelligence, Synergeti...
Gastric Cancer: Š”linical Implementation of Artificial Intelligence, Synergeti...Gastric Cancer: Š”linical Implementation of Artificial Intelligence, Synergeti...
Gastric Cancer: Š”linical Implementation of Artificial Intelligence, Synergeti...
Ā 
Call Girls Rishikesh Just Call 9667172968 Top Class Call Girl Service Available
Call Girls Rishikesh Just Call 9667172968 Top Class Call Girl Service AvailableCall Girls Rishikesh Just Call 9667172968 Top Class Call Girl Service Available
Call Girls Rishikesh Just Call 9667172968 Top Class Call Girl Service Available
Ā 
šŸ’šReliable Call Girls Chandigarh šŸ’ÆNiamh šŸ“²šŸ”8868886958šŸ”Call Girl In Chandigarh N...
šŸ’šReliable Call Girls Chandigarh šŸ’ÆNiamh šŸ“²šŸ”8868886958šŸ”Call Girl In Chandigarh N...šŸ’šReliable Call Girls Chandigarh šŸ’ÆNiamh šŸ“²šŸ”8868886958šŸ”Call Girl In Chandigarh N...
šŸ’šReliable Call Girls Chandigarh šŸ’ÆNiamh šŸ“²šŸ”8868886958šŸ”Call Girl In Chandigarh N...
Ā 
Kolkata Call Girls Naktala šŸ’ÆCall Us šŸ” 8005736733 šŸ” šŸ’ƒ Top Class Call Girl Se...
Kolkata Call Girls Naktala  šŸ’ÆCall Us šŸ” 8005736733 šŸ” šŸ’ƒ  Top Class Call Girl Se...Kolkata Call Girls Naktala  šŸ’ÆCall Us šŸ” 8005736733 šŸ” šŸ’ƒ  Top Class Call Girl Se...
Kolkata Call Girls Naktala šŸ’ÆCall Us šŸ” 8005736733 šŸ” šŸ’ƒ Top Class Call Girl Se...
Ā 
Call Girl In Chandigarh šŸ“ž9809698092šŸ“ž JustšŸ“² Call Inaaya Chandigarh Call Girls ...
Call Girl In Chandigarh šŸ“ž9809698092šŸ“ž JustšŸ“² Call Inaaya Chandigarh Call Girls ...Call Girl In Chandigarh šŸ“ž9809698092šŸ“ž JustšŸ“² Call Inaaya Chandigarh Call Girls ...
Call Girl In Chandigarh šŸ“ž9809698092šŸ“ž JustšŸ“² Call Inaaya Chandigarh Call Girls ...
Ā 
Ahmedabad Call Girls Book Now 8980367676 Top Class Ahmedabad Escort Service A...
Ahmedabad Call Girls Book Now 8980367676 Top Class Ahmedabad Escort Service A...Ahmedabad Call Girls Book Now 8980367676 Top Class Ahmedabad Escort Service A...
Ahmedabad Call Girls Book Now 8980367676 Top Class Ahmedabad Escort Service A...
Ā 
Chandigarh Call Girls Service ā¤ļøšŸ‘ 9809698092 šŸ‘„šŸ«¦Independent Escort Service Cha...
Chandigarh Call Girls Service ā¤ļøšŸ‘ 9809698092 šŸ‘„šŸ«¦Independent Escort Service Cha...Chandigarh Call Girls Service ā¤ļøšŸ‘ 9809698092 šŸ‘„šŸ«¦Independent Escort Service Cha...
Chandigarh Call Girls Service ā¤ļøšŸ‘ 9809698092 šŸ‘„šŸ«¦Independent Escort Service Cha...
Ā 
ā¤ļøCall Girl Service In Chandigarhā˜Žļø9814379184ā˜Žļø Call Girl in Chandigarhā˜Žļø Cha...
ā¤ļøCall Girl Service In Chandigarhā˜Žļø9814379184ā˜Žļø Call Girl in Chandigarhā˜Žļø Cha...ā¤ļøCall Girl Service In Chandigarhā˜Žļø9814379184ā˜Žļø Call Girl in Chandigarhā˜Žļø Cha...
ā¤ļøCall Girl Service In Chandigarhā˜Žļø9814379184ā˜Žļø Call Girl in Chandigarhā˜Žļø Cha...
Ā 
Premium Call Girls Nagpur {9xx000xx09} ā¤ļøVVIP POOJA Call Girls in Nagpur Maha...
Premium Call Girls Nagpur {9xx000xx09} ā¤ļøVVIP POOJA Call Girls in Nagpur Maha...Premium Call Girls Nagpur {9xx000xx09} ā¤ļøVVIP POOJA Call Girls in Nagpur Maha...
Premium Call Girls Nagpur {9xx000xx09} ā¤ļøVVIP POOJA Call Girls in Nagpur Maha...
Ā 
ā¤ļøAmritsar Escorts Serviceā˜Žļø9815674956ā˜Žļø Call Girl service in Amritsarā˜Žļø Amri...
ā¤ļøAmritsar Escorts Serviceā˜Žļø9815674956ā˜Žļø Call Girl service in Amritsarā˜Žļø Amri...ā¤ļøAmritsar Escorts Serviceā˜Žļø9815674956ā˜Žļø Call Girl service in Amritsarā˜Žļø Amri...
ā¤ļøAmritsar Escorts Serviceā˜Žļø9815674956ā˜Žļø Call Girl service in Amritsarā˜Žļø Amri...
Ā 
Independent Bangalore Call Girls (Adult Only) šŸ’ÆCall Us šŸ” 7304373326 šŸ” šŸ’ƒ Escor...
Independent Bangalore Call Girls (Adult Only) šŸ’ÆCall Us šŸ” 7304373326 šŸ” šŸ’ƒ Escor...Independent Bangalore Call Girls (Adult Only) šŸ’ÆCall Us šŸ” 7304373326 šŸ” šŸ’ƒ Escor...
Independent Bangalore Call Girls (Adult Only) šŸ’ÆCall Us šŸ” 7304373326 šŸ” šŸ’ƒ Escor...
Ā 
Bhawanipatna Call Girls šŸ“ž9332606886 Call Girls in Bhawanipatna Escorts servic...
Bhawanipatna Call Girls šŸ“ž9332606886 Call Girls in Bhawanipatna Escorts servic...Bhawanipatna Call Girls šŸ“ž9332606886 Call Girls in Bhawanipatna Escorts servic...
Bhawanipatna Call Girls šŸ“ž9332606886 Call Girls in Bhawanipatna Escorts servic...
Ā 
Chandigarh Call Girls Service ā¤ļøšŸ‘ 9809698092 šŸ‘„šŸ«¦Independent Escort Service Cha...
Chandigarh Call Girls Service ā¤ļøšŸ‘ 9809698092 šŸ‘„šŸ«¦Independent Escort Service Cha...Chandigarh Call Girls Service ā¤ļøšŸ‘ 9809698092 šŸ‘„šŸ«¦Independent Escort Service Cha...
Chandigarh Call Girls Service ā¤ļøšŸ‘ 9809698092 šŸ‘„šŸ«¦Independent Escort Service Cha...
Ā 
Difference Between Skeletal Smooth and Cardiac Muscles
Difference Between Skeletal Smooth and Cardiac MusclesDifference Between Skeletal Smooth and Cardiac Muscles
Difference Between Skeletal Smooth and Cardiac Muscles
Ā 
ā¤ļøChandigarh Escorts Serviceā˜Žļø9814379184ā˜Žļø Call Girl service in Chandigarhā˜Žļø ...
ā¤ļøChandigarh Escorts Serviceā˜Žļø9814379184ā˜Žļø Call Girl service in Chandigarhā˜Žļø ...ā¤ļøChandigarh Escorts Serviceā˜Žļø9814379184ā˜Žļø Call Girl service in Chandigarhā˜Žļø ...
ā¤ļøChandigarh Escorts Serviceā˜Žļø9814379184ā˜Žļø Call Girl service in Chandigarhā˜Žļø ...
Ā 
Cardiac Output, Venous Return, and Their Regulation
Cardiac Output, Venous Return, and Their RegulationCardiac Output, Venous Return, and Their Regulation
Cardiac Output, Venous Return, and Their Regulation
Ā 

Distortion Artifacts in MRI and their correction

  • 1. Distortion artifacts in MRI and their correction Vibha Chaswal, Ph.D.
  • 2. MRI Gradient fields ā€¢ The gradient fields are superimposed over the static magnetic field. For transverse images: ā€¢ Z gradientā€“selects the slice, and is applied whilst the 90Ā° RF pulse is on ā€¢ X gradientā€“creates a frequency change in emitted signal producing lines (or columns) of data, and is applied when the echo RF is being measured ā€¢ Y gradientā€“produces a change in the phase of rotation in different voxels, and is applied between the 90Ā° and 180Ā° RF pulses at a different strength during each cycle. The number of phase steps (e.g. 256) affects scan time. ā€¢ ā€œFrequency encodingā€ and ā€œPhase encodingā€ directions (X and Y) are interchangeable.
  • 3. Gradient spatial encoding Pulse sequencing diagram
  • 4. MRI distortion Artifacts A whole rangeā€¦ā€¦.. Equipment-related ā€¢ Non-uniformity of signal ā€¢ Wraparound (phase wrap, aliasing) ā€¢ Peripheral signal artifacts ā€¢ Slice overlap (cross-excitation) ā€¢ Noise and RF interference ā€¢ Truncation (ringing, Gibbs artifact) ā€¢ Phase smearing Patient-related ā€¢ Signal void due to freezing ā€¢ Partial volume averaging ā€¢ Chemical shift ā€¢ Motion artifacts (blurring, ghosting and pulsatility) ā€¢ Flow artifacts ā€¢ Susceptibility and image distortion ā€¢ Magic angle
  • 5. B0 inhomogeneity and susceptibility ā€¢ Extremely homogeneous static magnetic field is required around isocenter of the magnet for MRI. ā€¢ B0 homogeneity of the empty magnet: typical value of 1ppm (0.001%) in a 50 cm diameter spherical volume. ā€¢ susceptibility = local changes in Bo due to a substanceā€™s own magnetic properties ā€¢ Artifacts occur at interfaces between substances of different susceptibilities, e.g. air/bone/metal/ calcification/hemorrhage; implants and previous surgery, tooth fillings, mascara, ear-rings, hair-slides ā€“ worse at higher field strengths
  • 6. B0 inhomogeneity ā€¢ Decreased frequency-selective fat-saturation efficacy of pulses ā€¢ Effect more pronounced the further the image slice is positioned from the isocenter of the magnet (a) off-center acquisition with insufficient fat suppression (arrows) (b) Identical acquisition at isocenter Reduction: improve field shimming Optimal positioning of the patient in z-direction Dietrich et. al., Artifacts in 3-T MRI: Physical background and reduction strategies, EJR, 65, 2008
  • 7. B0 inhomogeneity ā€¢ Relatively small frequency shifts lead to substantial reduction of the transversal steady-state magnetization and, thus, to band-shaped signal losses in the image Reduction: Move the artifact out of the region of interest (apply frequency offset to nominal resonant frequency) Dietrich et. al., Artifacts in 3-T MRI: Physical background and reduction strategies, EJR, 65, 2008
  • 8. Susceptibility: Metal artifacts ā€¢ An important manifestation of a susceptibility-related artifact is the signal loss in gradient-echo images around metallic implants or surgical clips Right hip prosthesis is seen as region of signal drop out (closed arrows) Figure from: Mechlin et al, AJR 143, December 1984 STIR axial MRI image of a patient, arrows B point to artifacts due to right hip prosthesis. Figure from: Dr. Geller et al, Malignant Peripheral Nerve Sheath Tumors (MPNST), ESUN, june 2006, 3(3) http://sarcomahelp.org/learning_center/mpnst.html Reduction: Decrease Echo time, TE, of pulse sequence Increase receiver band-width
  • 9. Susceptibility: Geometric distortion artifact ā€¢ ā€¢ Usually in echo-planar imaging (EPI) At interfaces between soft-tissue and bone/air e.g., in the base of the skull or in the head-and-neck region Reduction: a) decrease the echo-spacing of the read out train e.g., by increasing the receiver bandwidth or by parallel imaging techniques b) Use fast-spin echo technique (b) EPI acquisition without parallel imaging, severe distortion artifacts are visible (arrows) and (c) EPI acquisition with acceleration factor 2, distortion artifacts are still present but considerably reduced (arrows). Dietrich et. al., Artifacts in 3-T MRI: Physical background and reduction strategies, EJR, 65, 2008
  • 10. B1 inhomogeneity ā€¢ Spatial inhomogeneity of the B1 field results in flip-angle deviations depending on the spatial position ā€¢ Reduced signal intensity in these areas or to altered contrast particularly in FLASH sequences whose T1-weighting depends on the flip angle
  • 11. Dielectric resonance effects/ RF interference/ standing wave effect ā€¢ Constructive or destructive interferences of the transmitted RF field may be observed resulting in either regional (e.g., central) brightening or regional signal loss, respectively. Signal loss due to wavelength effect Reduction: a) Effect can be mitigated by positioning a dielectric cushion close to the ROI, or b) Manually modify the RF-transmitter amplitude in order to reduce B1-inhomogeneityinduced signal loss
  • 12. Chemical Shift Artifact ā€¢ For MRI in vivo, two important groups of molecules : protons in water and protons in fat tissue. ā€¢ Larmor frequencies difference of these protons ~ 3.5 ppm. ā€¢ chemical shift leads to a slight geometric shift of the relative position of water and fat protons in readout direction.
  • 13. Chemical Shift Artifact Reduction image-compromising superposition of fat tissue onto other tissues Reduction: a)manually increase the RF-transmitter amplitude and apply image post-processing filters to obtain more uniform image intensities. b)Increase (~double) receiver bandwidth c)Based on anatomy switch readout and phase encoding directions
  • 14. Blood Flow related artifact ā€¢ Magnetohydrodynamic effect: The flow of blood ions perpendicular to the strong static magnetic field, B0, gives rise to induced voltages and currents; ā€œ(a) Signal loss in the pulmonary vessels in non-gated 3-T single-shot fast-spin-echo (FSE) acquisition possibly related to the magnetohydrodynamic effect. (b) Non-gated 1.5-T single-shot FSE acquisition demonstrating visual signal in pulmonary vessels. (c) ECG-gated 3-T single-shot FSE acquisition of the same volunteer as in (a) with data acquisition during diastole; the vessel signal is restored.ā€ ā€¦ā€¦.Deitrich et al, EJR 65(2008)
  • 15. SNR related artifacts ā€¢ Artifacts to Noise ratio (ANR) ā€¢ Artifacts are masked by increased statistical noise. ā€¢ Gibbs Ringing: caused by data clipping at the edges of K-space (raw data readings are still over noise level at the borders of acquired K-space)
  • 16. Gibbs artifact Reduction: Increase spatial resolution or, Apply reconstruction filters (Hanning filter)
  • 17. Resources Many more Artifacts but time restriction ā‡’Resources for MR Artifacts ad their reduction ā‡’MR-TIP: online database (open forum) from University of Illinois, Excellent resource ā‡’ Review article by Dietrich et.al,ā€Artifacts in 3-T MRI: Physical Background and reduction strategiesā€ EJR, Vol 65, 2008

Editor's Notes

  1. Spin echo
  2. Happens If the wavelength of the RF field is in the same order of magnitude as the geometric dimensions of the imaged object
  3. This is because the spatial position in MRI is encoded using varying Larmor frequencies induced by the applied magnetic gradient pulses.
  4. 1.5 T machines had high ANR and Gibbs was suppressed because of high noise. More in 3T
  5. Filter smooths the image and reduces the signal at the edges of K-space