SlideShare a Scribd company logo
1 of 107
ULTRASOUND
TRANSDUCERS AND
RESOLUTION
Dr V S R Bhupal
 Ultrasound

is produced and detected
with a transducer, composed of one or
more ceramic elements with
electromechanical (piezoelectric)
properties.

• The ceramic element converts electrical

energy into mechanical energy to produce
ultrasound and mechanical energy into
electrical energy for ultrasound detection.


Over the past several decades, the transducer
assembly has evolved considerably in design,
function, and capability, from a single-element
resonance crystal to a broadband transducer
array of hundreds of individual elements.

•

A simple single-element, plane-piston source
transducer has major components including the

• piezoelectric material,
• matching layer,
• backing block,
• acoustic absorber,
• insulating cover,
• sensor electrodes, and
• transducer housing.
Piezoelectric Materials
A

piezoelectric material (often a crystal
or ceramic) is the functional component
of the transducer.

• It converts electrical energy into mechanical

(sound) energy by physical deformation of the
crystal structure.
 ConverseIy,

mechanical pressure
applied to its surface creates electrical
energy.

• Piezoelectric materials are characterized by a
well-defined molecular arrangement of
electrical dipoles.
 An

electrical dipole is a molecular entity
containing positive and negative electric
charges that has no net charge.

• When mechanically compressed by an

externally applied pressure, the alignment of
the dipoles is disturbed from the equilibrium
position to cause an imbalance of the charge
distribution.
A

potential difference (voltage) is created
across the element with one surface
maintaining a net positive charge and
one surface a net negative charge.

• Surface electrodes measure the voltage,
which is proportional to the incident
mechanical pressure amplitude.
 Conversely,

application of an external
voltage through conductors attached to
the surface electrodes induces the
mechanical expansion and contraction of
the transducer element.
 There

are natural and synthetic
piezoelectric materials.

• An example of a natural piezoelectric material

is quartz crystal, commonly used in watches
and other time pieces to provide a mechanical
vibration source at 32.768 kHz for interval
timing.

• This is one of several oscillation frequencies of
quartz, determined by the crystal cut and
machining properties.


Ultrasound transducers for medical imaging
applications employ a synthetic piezoelectric
ceramic, most often lead-zirconate-titanate
(PZT).

•

The piezoelectric attributes are attained after a
process of

• Molecular synthesis,
• Heating,
• Orientation of internal dipole structures with an applied
external voltage,
• Cooling to permanently maintain the dipole orientation,
and
• Cutting into a specific shape.


For PZT in its natural state, no piezoelectric
properties are exhibited; however, heating the
material past its “Curie temperature” (i.e., 3280
C to 3650 C) and applying an external voltage
causes the dipoles to align in the ceramic.

•

The external voltage is maintained until the material
has cooled to below its Curie temperature.

• Once the material has cooled, the dipoles retain their
alignment.
 At

equilibrium, there is no net charge on
ceramic surfaces.

• When compressed, an imbalance of charge
produces a voltage between the surfaces.

• Similarly, when a voltage is applied between

electrodes attached to both surfaces, mechanical
deformation occurs.
 The

piezoelectric element is composed
of aligned molecular dipoles.
 Under

the influence of mechanical
pressure from an adjacent medium
(e.g., an ultrasound echo), the element
thickness

• Contracts (at the peak pressure amplitude),
• Achieves equilibrium (with no pressure) or
• Expands (at the peak rarefactional pressure),

• This causes realignment of the electrical dipoles to
produce positive and negative surface charge.
 Surface

electrodes measure the
voltage as a function of time.
 An

external voltage source applied to the
element surfaces causes compression or
expansion from equilibrium by
realignment of the dipoles in response to
the electrical attraction or repulsion
force.
Resonance Transducers


Resonance transducers for pulse echo
ultrasound imaging are manufactured to
operate in a “resonance” mode, whereby a
voItage (commonly 150 V) of very short
duration (a voltage spike of ≈1 µsec) is
applied, causing the piezoelectric material to
initially contract, and subsequently vibrate at a
natural resonance frequency.

•

This frequency is selected by the “thickness cut,” due
to the preferential emission of ultrasound waves
whose wavelength is twice the thickness of the
piezoelectric material.
 The

operating frequency is determined
from the speed of sound in, and the
thickness of, the piezoelectric material.

• For example, a 5-MHz transducer will have a

wavelength in PZT (speed of sound in PZT is
≈ 4,000 m/sec) of

c 4000 m / sec
λ= =
= 8 × 10 −4 meters = 0.80 mm
f
5 × 106 / sec


A short duration
voltage spike causes
the resonance
piezoelectric element
to vibrate at its
natural frequency, fo,
which is determined
by the thickness of
the transducer equal
to 1/A.
 To

achieve the 5-MHz resonance
frequency, a transducer element
thickness of ½ X 0.8 mm = 0.4 mm is
required.

• Higher frequencies are achieved with thinner
elements, and lower frequencies with thicker
elements.

• Resonance transducers transmit and receive
preferentially at a single “center frequency.”
Damping Block


The damping block, layered on the back of the
piezoelectric element, absorbs the backward
directed ultrasound energy and attenuates
stray ultrasound signals from the housing.

•

This component also dampens the transducer
vibration to create an ultrasound pulse width and short
spatial pulse length, which is necessary to preserve
detail along he beam axis (axial resolution).
 Dampening

of the vibration (also known
as “ring-down”) lessens the purity of the
resonance frequency and introduces a
broadband frequency spectrum.

• With ring-down, an increase in the bandwidth

(range of frequencies) of the ultrasound pulse
occurs by introducing higher and lower
frequencies above and below the center
(resonance) frequency.


The “Q factor” describes the bandwidth of the
sound emanating from a transducer as
fo
Q=
Bandwidth



where fo is the center frequency and the
bandwidth is the width of the frequency
distribution.
A

“high Q” transducer has a narrow
bandwidth (i.e., very little damping) and a
corresponding long spatial pulse length.

• A “low Q” transducer has a wide bandwidth
and short spatial pulse length.
 Imaging

applications require a broad
bandwidth transducer in order to achieve
high spatial resolution along the direction
of beam travel.

• Blood velocity measurements by Doppler

instrumentation require a relatively narrowband transducer response in order to
preserve velocity information encoded by
changes in the echo frequency relative to the
incident frequency.
 Continuous-wave

ultrasound transducers
have a very high Q characteristic.

• While the Q factor is derived from the term

quality factor, a transducer with a low Q does
not imply poor quality in the signal.
Matching Layer


The matching layer provides the interface
between the transducer element and the tissue
and minimizes the acoustic impedance
differences between the transducer and the
patient.

•

It consists of layers of materials with acoustic
impedances that are intermediate to those of soft
tissue and the transducer material.

• The thickness of each layer is equal to one-fourth the

wavelength, determined from the center operating
frequency of the transducer and speed of sound in the
matching layer.
 For

example, the wavelength of sound in
a matching layer with a speed of sound
of 2,000 m/sec for a 5-MHz ultrasound
beam is 0.4 mm.

• The optimal matching layer thickness is equal
to ¼λ = ¼ x 0.4 mm = 0. 1 mm.

• In addition to the matching layers, acoustic

coupling gel (with acoustic impedance similar to
soft tissue) is used between the transducer and the
skin of the patient to eliminate air pockets that
could attenuate and reflect the ultrasound beam.
Nonresonance (BroadBandwidth) “Multifrequency”
Transducers


Modern transducer design coupled with digital
signal processing enables “multifrequency or
“multihertz” transducer operation, whereby rhe
center frequency can be adjusted in he
transmit mode.

•

Unlike the resonance transducer design, the
piezoelectric element is intricately machined into a
large number of small “rods,” and then filled with an
epoxy resin to create a smooth surface.
 The

acoustic properties are closer to
tissue than a pure PZT material, and
thus provide a greater transmission
efficiency of the ultrasound beam without
resorting to multiple matching layers.

• Multifrequency transducers have bandwidths
that exceed 80% of the center frequency.
 Excitation

of the multifrequency
transducer is accomplished with a short
square wave burst of 150 V with one to
three cycles, unlike the voltage spike
used for resonance transducers.

• This allows the center frequency to be

selected within the limits of the transducer
bandwidth.
 Likewise,

the broad bandwidth response
permits the reception of echoes within a
wide range of frequencies.

• For instance, ultrasound pulses can be

produced at a low frequency, and the echoes
received at higher frequency.
 “Harmonic

imaging” is a recently
introduced technique that uses this
ability;

• lower frequency ultrasound is transmitted into

the patient, and the higher frequency
harmonics (e.g., two times the transmitted
center frequency) created from the interaction
with contrast agents and tissues, are received
as echoes.
 Native

tissue harmonic imaging has
certain advantages including greater
depth of penetration, noise and clutter
removal, and improved lateral spatial
resolution.
Transducer Arrays
 The

majority of ultrasound systems
employ transducers with many individual
rectangular piezoelectric elements
arranged in linear or curvilinear arrays.

• Typically, 128 to 512 individual rectangular

elements compose the transducer assembly.

• Each element has a width typically less than half

the wavelength and a length of several millimeters.


Two modes of
activation are used
to produce a beam.

•

These are the “linear”
(sequential) and
“phased”
activation/receive
modes.
Linear Arrays
 Linear

array transducers typically contain
256 to 512 elements; physically these
are the largest transducer assemblies.
 In

operation, the simultaneous firing of’ a
small group of ≈ 20 adjacent elements
produces the ultrasound beam.

• The simultaneous activation produces a

synthetic aperture (effetive transducer width)
defined by the number of active elements.
 Echoes

are detected in the receive mode
by acquiring signals from most of the
transducer elements.

• Subsequent “A-line” acquisition occurs by

firing another group of transducer elements
displaced by one or two elements.
A

rectangular field of view is produced
with this transducer arrangement.

• For a curvilinear array, a trapezoidal field of
view is produced.
Phased Arrays
A

phased-array transducer is usually
composed of 64 to 128 individual
elements in a smaller package than a
linear array transducer.

• All transducer elements are activated nearly

(but not exactly) simultaneously to produce a
single ultrasound beam.


By using time delays in the electrical activarion
of the discrete elements across the face of the
transducer, the ultrasound beam can be
steered and focused electronically without
moving the transducer.

•

During ultrasound signal reception, all of the
transducer elements detect the returning echoes from
the beam path, and sophisticated algorithms
synthesize the image from the detected data.
BEAM PROPERTIES
 The

ultrasound beam propagates as a
longitudinal wave from the transducer
surface into the propagation medium,
and exhibits two distinct beam patterns:

• a slightly converging beam out to a distance
•

specified by the geometry and frequency of
the transducer (the near field), and
a diverging beam beyond that point (the far
field).


For an unfocused,
single-element
transducer, the
length of the near
field is determined
by the transducer
diameter and the
frequency of the
transmitted sound.
 For

multiple transducer element arrays,
an “effective” transducer diameter is
determined by the excitation of a group
of’ transducer elements.

• Because of the interactions of each of the

individual beams and the ability to focus
and steer the overall beam, the formulas
for a single-element, unfocused transducer
are not directly applicable.
The Near Field
 The

near field, also known as the
Fresnel zone, is adjacent to the
transducer face and has a converging
beam profile.

• Beam convergence in the near field occurs
because of multiple constructive and
destructive interference patterns of the
ultrasound waves from the transducer
surface.
 Huygen’s

principle describes a large
transducer surface as an infinite
number of point sources of sound
energy where each point is
characterized as a radial emitter.

• By analogy, a pebble dropped in a quiet pond
creates a radial wave pattern.


As individual wave
patterns interact, the
peaks and troughs from
adjacent sources
constructively and
destructively interfere,
causing the beam profile
to be tightly collimated in
the near field.
 The

ultrasound beam path is thus largely
confined to the dimensions of the active
portion of the transducer surface, with
the beam diameter converging to
approximately half the transducer
diameter at the end of the near field.
 The

near field length is dependent on the
transducer frequency and diameter:
d 2 r2
Near field length =
=
4λ λ

• where d is the transducer diameter, r is the

transducer radius, and λ is the wavelength of
ultrasound in the propagation medium.
soft tissue, λ = 1.54mm/f(MHz), and
the near field length can be expressed
as a function of frequency:

 In

(

)

d2
mm 2 ( MHz )
Near field length =
( mm)
4 × 1.54


A higher transducer
frequency (shorter
wavelength) will
result in a longer
near field, as will a
larger diameter
element.
 For

a 10-mm-diameter transducer, the
near field extends 5.7 cm at 3.5 MHz and
16.2 cm at 10 MHz in soft tissue.

• For a 15-mm-diameter transducer, the

corresponding near field lengths are 12.8 and
36.4 cm, respectively.
 Lateral

resolution (the ability of the
system to resolve objects in a direction
perpendicular to the beam direction) is
dependent on the beam diameter and is
best at the end of the near field for a
single-element transducer.

• Lateral resolution is worst in areas close to
and far from the transducer surface.
 Pressure

amplitude characteristics in the
near field are very complex, caused by
the constructive and destructive
interference wave patterns of the
ultrasound beam.

• Peak ultrasound pressure occurs at the end

of the near field, corresponding to the
minimum beam diameter for a single-element
transducer.
 Pressures

vary rapidly from peak
compression to peak rarefaction several
times during transit through the near
field.

• Only when the far field is reached do the
ultrasound pressure variations decrease
continuously.
 The

far field is also known as the
Fraunhofer zone, and is where the beam
diverges.

• For a large-area single-element transducer,

the angle of ultrasound beam divergence, 0,
for the far field is given by

λ
sin θ = 1.22
d

• where d is the effective diameter of the

transducer and λ is the wavelength; both must
have the same units of distance.
 Less

beam divergence occurs with highfrequency, large-diameter transducers.

• Unlike the near field, where beam intensity

varies from maximum to minimum to
maximum in a converging beam, ultrasound
intensity in the far field decreases
monotonically with distance.
Transducer Array Beam
Formation and Focusing
 In

a transducer array, the narrow
piezoelectric element width (typically less
than one wavelength) produces a
diverging beam at a distance very close
to the transducer face.

• Formation and convergence of the ultrasound
beam occurs with the operation of several or
all of the transducer elements at the same
time.


Transducer elements in a linear array that are
fired simultaneously produce an effective
transducer width equal to the sum of the
widths of the individual elements.

• Individual beams interact via constructive and

destructive interference to produce a collimated
beam that has properties similar to the properties
of a single transducer of the same size.
 With

a phased-array transducer, the
beam is formed by interaction of the
individual wave fronts from each
transducer, each with a slight difference
in excitation time.

• Minor phase differences of adjacent beams
form constructive and destructive wave
summations that steer or focus the beam
profile.
COMMON TRANSDUCERS USED
IN CLINICAL SETTING
STRAIGHT LINEAR ARRAY
PROBE
The straight linear array probe is designed
for superficial imaging.
The crystals are aligned in a linear fashion
within a flat head and produce sound
waves in a straight line.
The image produced is rectangular in
shape.
 This

probe has higher frequencies (5–13
MHz), which provides better resolution
and less penetration.
 Therefore, this probe is ideal for imaging
superficial structures and in ultrasoundguided procedures.
Vascular access
Evaluate for deep venous thrombosis
Skin and soft tissue for abscess, foreign
body
Musculoskeletal—tendons, bones,
muscles
CURVILINEAR ARRAY
PROBE
 The

curvilinear array or convex probe is
used for scanning deeper structures. The
crystals are aligned along a curved
surface and cause a fanning out of the
beam, which results in a field of view that
is wider than the probe’s footprint.
 The

image generated is sector shaped.
These probes have frequencies ranging
between 1 and 8 MHz, which allows for
greater penetration, but less resolution.
These probes are most often used in
abdominal and pelvic applications.
 They are also useful in certain
musculoskeletal evaluations or
procedures when deeper anatomy needs
to be imaged or in obese patients.
 Abdominal

aorta
 Biliary/gallbladder/liver/pancreas
 Abdominal portion of FAST exam
 Kidney and bladder evaluation
 Transabdominal pelvic evaluation
ENDOCAVITARY PROBE
 The

endocavitary probe also has a
curved face, but a much higher
frequency (8–13 MHz) than the
curvilinear probe.
 This probe’s elongated shape allows it to
be inserted close to the anatomy being
evaluated.
 The

curved face creates a wide field of
view of almost 180° and its high
frequencies provide superior resolution .
This probe is used most commonly for
gynecological applications, but can also
be used for intraoral evaluation of
peritonsillar abscesses.
 Transvaginal ultrasound
 Intraoral
PHASED ARRAY PROBE
 Phased

array probes (Fig. 4-4a) have
crystals that are grouped closely
together.
 The timing of the electrical pulses that
are applied to the crystals varies and
they are fired in an oscillating manner.
 The

sound waves that are generated
originate from a single point and fan
outward, creating a sector-type image.
This probe has a smaller and flatter
footprint than the curvilinear one, which
allows the user to maneuver more easily
between the ribs and small spaces.
These probes have frequencies between
2 and 8 MHz.
IVUS PROBE
 IVUS

is a miniature ultrasound probe
positioned at the tip of a coronary
catheter.
 The probe emits ultrasound frequencies,
typically at 20-45 MHz, and the signal is
reflected from surrounding tissue and
reconstructed into a real-time
tomographic gray-scale image.
Spatial Resolution
 In

ultrasound, the major factor that limits
the spatial resolution and visibility of
detail is the volume of the acoustic pulse.


The axial, lateral,
and elevational (slice
thickness)
dimensions
determine the
minimal volume
element.
 Each

dimension has an effect on the
resolvability of objects in the image.
Axial Resolution
 Axial

resolution (also known as linear,
range, longitudinal, or depth resolution)
refers to the ability to discern two closely
spaced objects in the direction of the
beam.

• Achieving good axial resolution requires that
the returning echoes be distinct without
overlap.
 The

minimal required separation
distance between two reflectors is onehalf of the spatial pulse length (SPL) to
avoid the overlap of returning echoes, as
the distance traveled between two
reflectors is twice the separation
distance.


Objects spaced
closer than ½ SPL
will not be resolved.
 The

SPL is the number of cycles emitted
per pulse by the transducer multiplied by
the wavelength.

• Shorter pulses, producing better axial

resolution, can be achieved with greater
damping of the transducer element (to reduce
the pulse duration and number of cycles) or
with higher frequency (to reduce wavelength).
 For

imaging applications, the ultrasound
pulse typically consists of three cycles.

• At 5 MHz (wavelength of 0.31 mm), the SPL

is about 3 x 0.31 0.93 mm, which provides an
axial resolution of /2(0.93 mm) = 0.47 mm.
 At

a given frequency, shorter pulse
lengths require heavy damping and low
Q, broad-bandwidth operation.

• For a constant damping factor, higher

frequencies (shorter wavelengths) give better
axial resolution, but the imaging depth is
reduced.

• Axial resolution remains constant with depth.
Lateral Resolution
 Lateral

resolution, also known as
azimuthal resolution, refers to the ability
to discern as separate two closely
spaced objects perpendicular to the
beam direction.


For both single
element transducers
and multielement
array transducers,
the beam diameter
determines the
lateral resolution.
 Since

the beam diameter varies with the
distance from the transducer in the near
and far field, the lateral resolution is
depth dependent.

• The best lateral resolution occurs at the near
field—far field face.
 At

this depth, the effective beam
diameter is approximately equal to half
the transducer diameter.

• In the far field, the beam diverges and

substantially reduces the lateral resolution.
 The

typical lateral resolution for an
unfocused transducer is approximately 2
to 5 mm.

• A focused transducer uses an acoustic lens

(a curved acoustic material analogous to an
optical lens) to decrease the beam diameter
at a specified distance from the transducer.
 With

an acoustic lens, lateral resolution
at the near field-far field interface is
traded for better lateral resolution at a
shorter depth, but the far field beam
divergence is substantially increased.

• The lateral resolution of linear and curvilinear
array transducers can be varied.
Elevational Resolution
 The

elevational or slice-thickness
dimension of the ultrasound beam is
perpendicular to the image plane.

• Slice thickness plays a significant part in

image resolution, particularly with respect to
volume averaging of acoustic details in the
regions dose to the transducer and in the far
field beyond the focal zone.


Elevational
resolution is
dependent on the
transducer element
height in much the
same way that the
lateral resolution is
dependent on the
transducer element
width.
 Slice

thickness is typically the worst
measure of resolution for array
transducers.

• Use of a fixed focaI length lens across the

entire surface of the array provides improved
elevational resolution at the focal distance.
 Unfortunately,

this compromises
resolution due to partial volume
averaging before and after the
elevational focal zone (elevational
resolution quality control phantom image
shows the effects of variable resolution
with depth.
 Multiple

linear array transducers with five
to seven rows, known as 1.5dimensional (1.5-D) transducer arrays,
have the ability to steer and focus the
beam in the elevational dimension.

More Related Content

What's hot

Contrast media in CT
Contrast media in CTContrast media in CT
Contrast media in CT
Kyle Rousseau
 
Basic physics of ultrasound.JH
Basic physics of ultrasound.JHBasic physics of ultrasound.JH
Basic physics of ultrasound.JH
hari baskar
 

What's hot (20)

Basics Physics of ultrasound
Basics Physics of ultrasoundBasics Physics of ultrasound
Basics Physics of ultrasound
 
Principle of usg imaging, construction of transducers
Principle of usg imaging, construction of transducersPrinciple of usg imaging, construction of transducers
Principle of usg imaging, construction of transducers
 
Ultrasound contrast agents
Ultrasound contrast agentsUltrasound contrast agents
Ultrasound contrast agents
 
Fluroscopy
FluroscopyFluroscopy
Fluroscopy
 
Doppler Ultrasonography And Advancements in USG
Doppler Ultrasonography And Advancements in USGDoppler Ultrasonography And Advancements in USG
Doppler Ultrasonography And Advancements in USG
 
Doppler Physics
Doppler PhysicsDoppler Physics
Doppler Physics
 
Ultrasound transducer doppler ppt pdf pk
Ultrasound transducer doppler ppt pdf pkUltrasound transducer doppler ppt pdf pk
Ultrasound transducer doppler ppt pdf pk
 
Doppler Effect - Ultrasound
Doppler Effect - UltrasoundDoppler Effect - Ultrasound
Doppler Effect - Ultrasound
 
Types of transducers in ultrasound diagnostic
Types of transducers in ultrasound diagnosticTypes of transducers in ultrasound diagnostic
Types of transducers in ultrasound diagnostic
 
Usg diplay modes
Usg diplay  modesUsg diplay  modes
Usg diplay modes
 
Ultrasound instrumentation practical applications
Ultrasound instrumentation practical applicationsUltrasound instrumentation practical applications
Ultrasound instrumentation practical applications
 
Ultrasound imaging system
Ultrasound imaging system Ultrasound imaging system
Ultrasound imaging system
 
MDCT Principles and Applications- Avinesh Shrestha
MDCT Principles and Applications- Avinesh ShresthaMDCT Principles and Applications- Avinesh Shrestha
MDCT Principles and Applications- Avinesh Shrestha
 
Contrast media in CT
Contrast media in CTContrast media in CT
Contrast media in CT
 
Artifacts usg, ct, mri radiology ppt,pdf pk
Artifacts usg, ct, mri radiology  ppt,pdf pkArtifacts usg, ct, mri radiology  ppt,pdf pk
Artifacts usg, ct, mri radiology ppt,pdf pk
 
Basic physics of ultrasound.JH
Basic physics of ultrasound.JHBasic physics of ultrasound.JH
Basic physics of ultrasound.JH
 
CT Physics
CT PhysicsCT Physics
CT Physics
 
CT angiography
CT angiographyCT angiography
CT angiography
 
Usg artifacts
Usg artifactsUsg artifacts
Usg artifacts
 
Ultrasound artifacts
Ultrasound artifactsUltrasound artifacts
Ultrasound artifacts
 

Viewers also liked

Physics of Ultrasound Imaging
Physics of Ultrasound ImagingPhysics of Ultrasound Imaging
Physics of Ultrasound Imaging
u.surgery
 
Ultrasound imaging
Ultrasound imagingUltrasound imaging
Ultrasound imaging
NIVETA SINGH
 
Introduction to electrical engineering
Introduction to electrical engineeringIntroduction to electrical engineering
Introduction to electrical engineering
Rajneesh Budania
 
Physical Principles Of Ultrasound
Physical Principles Of UltrasoundPhysical Principles Of Ultrasound
Physical Principles Of Ultrasound
u.surgery
 
Cardiac Electrophysiology
Cardiac ElectrophysiologyCardiac Electrophysiology
Cardiac Electrophysiology
TeleClinEd
 
ELECTROENCEPHALOGRAM/ ELECTROENCEPHALOGRAPHY
ELECTROENCEPHALOGRAM/ ELECTROENCEPHALOGRAPHYELECTROENCEPHALOGRAM/ ELECTROENCEPHALOGRAPHY
ELECTROENCEPHALOGRAM/ ELECTROENCEPHALOGRAPHY
RoxanneMae Birador
 

Viewers also liked (20)

Physics of Ultrasound Imaging
Physics of Ultrasound ImagingPhysics of Ultrasound Imaging
Physics of Ultrasound Imaging
 
Ultrasound imaging
Ultrasound imagingUltrasound imaging
Ultrasound imaging
 
Discograpgy (intradiscal procedure)
Discograpgy (intradiscal procedure)Discograpgy (intradiscal procedure)
Discograpgy (intradiscal procedure)
 
Ultrasound artifacts and contrast enhanced ultrasound
Ultrasound artifacts and contrast enhanced ultrasoundUltrasound artifacts and contrast enhanced ultrasound
Ultrasound artifacts and contrast enhanced ultrasound
 
LCR Electronics Inc.
LCR Electronics Inc.LCR Electronics Inc.
LCR Electronics Inc.
 
Q meter ppt
Q meter pptQ meter ppt
Q meter ppt
 
Strain guage
Strain guageStrain guage
Strain guage
 
Basic electrical engineering
Basic electrical engineering Basic electrical engineering
Basic electrical engineering
 
D.C. Circuits
D.C. CircuitsD.C. Circuits
D.C. Circuits
 
Best Electronics Measurement Equipments Manufacturers In Pune, India
Best Electronics Measurement Equipments Manufacturers In Pune, IndiaBest Electronics Measurement Equipments Manufacturers In Pune, India
Best Electronics Measurement Equipments Manufacturers In Pune, India
 
Network Theorems.ppt
Network Theorems.pptNetwork Theorems.ppt
Network Theorems.ppt
 
Transducer
Transducer Transducer
Transducer
 
ULTRASOUND IMAGING PRINCIPLES
ULTRASOUND IMAGING PRINCIPLESULTRASOUND IMAGING PRINCIPLES
ULTRASOUND IMAGING PRINCIPLES
 
Introduction to electrical engineering
Introduction to electrical engineeringIntroduction to electrical engineering
Introduction to electrical engineering
 
Physical Principles Of Ultrasound
Physical Principles Of UltrasoundPhysical Principles Of Ultrasound
Physical Principles Of Ultrasound
 
Lvdt
LvdtLvdt
Lvdt
 
Arrhythmia
ArrhythmiaArrhythmia
Arrhythmia
 
Cardiac Electrophysiology
Cardiac ElectrophysiologyCardiac Electrophysiology
Cardiac Electrophysiology
 
ELECTROENCEPHALOGRAM/ ELECTROENCEPHALOGRAPHY
ELECTROENCEPHALOGRAM/ ELECTROENCEPHALOGRAPHYELECTROENCEPHALOGRAM/ ELECTROENCEPHALOGRAPHY
ELECTROENCEPHALOGRAM/ ELECTROENCEPHALOGRAPHY
 
Electroencephalogram
ElectroencephalogramElectroencephalogram
Electroencephalogram
 

Similar to ultrasound transducers and resolution

usgphysics-151102111226-lva1-app6892.pdf
usgphysics-151102111226-lva1-app6892.pdfusgphysics-151102111226-lva1-app6892.pdf
usgphysics-151102111226-lva1-app6892.pdf
littlealphonsa
 
Presentation1
Presentation1Presentation1
Presentation1
poonam
 

Similar to ultrasound transducers and resolution (20)

usgphysics-151102111226-lva1-app6892.pdf
usgphysics-151102111226-lva1-app6892.pdfusgphysics-151102111226-lva1-app6892.pdf
usgphysics-151102111226-lva1-app6892.pdf
 
Ultrasound imaging system
Ultrasound imaging system Ultrasound imaging system
Ultrasound imaging system
 
Ultrasound Physics Jr.pptx
Ultrasound Physics  Jr.pptxUltrasound Physics  Jr.pptx
Ultrasound Physics Jr.pptx
 
Presentation1
Presentation1Presentation1
Presentation1
 
Usg physics
Usg physicsUsg physics
Usg physics
 
Ultrasoundtherapy 150201220726-conversion-gate02 (1)
Ultrasoundtherapy 150201220726-conversion-gate02 (1)Ultrasoundtherapy 150201220726-conversion-gate02 (1)
Ultrasoundtherapy 150201220726-conversion-gate02 (1)
 
Ultrasound and Doppler physics
Ultrasound and Doppler physicsUltrasound and Doppler physics
Ultrasound and Doppler physics
 
Ultrasound Imaging_2023.pdf
Ultrasound Imaging_2023.pdfUltrasound Imaging_2023.pdf
Ultrasound Imaging_2023.pdf
 
Ultrasound therapy
Ultrasound therapyUltrasound therapy
Ultrasound therapy
 
Physics of echo i.tammi raju
Physics of echo i.tammi rajuPhysics of echo i.tammi raju
Physics of echo i.tammi raju
 
Ultrasound physics
Ultrasound physicsUltrasound physics
Ultrasound physics
 
factors_affecting_xray_production.pdf
factors_affecting_xray_production.pdffactors_affecting_xray_production.pdf
factors_affecting_xray_production.pdf
 
Transducers 2020
Transducers 2020Transducers 2020
Transducers 2020
 
USphysics.pptx
USphysics.pptxUSphysics.pptx
USphysics.pptx
 
Presentation GROUP TWO.pptx
Presentation GROUP TWO.pptxPresentation GROUP TWO.pptx
Presentation GROUP TWO.pptx
 
Us transducers (2)
Us transducers (2)Us transducers (2)
Us transducers (2)
 
Bps xray mammo dsa
Bps xray mammo dsaBps xray mammo dsa
Bps xray mammo dsa
 
Acoustic sensors
Acoustic sensorsAcoustic sensors
Acoustic sensors
 
Ultrasound physics and image optimization1 (1)
Ultrasound physics and image optimization1 (1)Ultrasound physics and image optimization1 (1)
Ultrasound physics and image optimization1 (1)
 
BTech_Unit_I.pptx Ultrasonic sound ppt for students
BTech_Unit_I.pptx  Ultrasonic sound ppt for studentsBTech_Unit_I.pptx  Ultrasonic sound ppt for students
BTech_Unit_I.pptx Ultrasonic sound ppt for students
 

ultrasound transducers and resolution

  • 2.  Ultrasound is produced and detected with a transducer, composed of one or more ceramic elements with electromechanical (piezoelectric) properties. • The ceramic element converts electrical energy into mechanical energy to produce ultrasound and mechanical energy into electrical energy for ultrasound detection.
  • 3.  Over the past several decades, the transducer assembly has evolved considerably in design, function, and capability, from a single-element resonance crystal to a broadband transducer array of hundreds of individual elements. • A simple single-element, plane-piston source transducer has major components including the • piezoelectric material, • matching layer, • backing block, • acoustic absorber, • insulating cover, • sensor electrodes, and • transducer housing.
  • 4.
  • 5. Piezoelectric Materials A piezoelectric material (often a crystal or ceramic) is the functional component of the transducer. • It converts electrical energy into mechanical (sound) energy by physical deformation of the crystal structure.
  • 6.  ConverseIy, mechanical pressure applied to its surface creates electrical energy. • Piezoelectric materials are characterized by a well-defined molecular arrangement of electrical dipoles.
  • 7.  An electrical dipole is a molecular entity containing positive and negative electric charges that has no net charge. • When mechanically compressed by an externally applied pressure, the alignment of the dipoles is disturbed from the equilibrium position to cause an imbalance of the charge distribution.
  • 8. A potential difference (voltage) is created across the element with one surface maintaining a net positive charge and one surface a net negative charge. • Surface electrodes measure the voltage, which is proportional to the incident mechanical pressure amplitude.
  • 9.  Conversely, application of an external voltage through conductors attached to the surface electrodes induces the mechanical expansion and contraction of the transducer element.
  • 10.  There are natural and synthetic piezoelectric materials. • An example of a natural piezoelectric material is quartz crystal, commonly used in watches and other time pieces to provide a mechanical vibration source at 32.768 kHz for interval timing. • This is one of several oscillation frequencies of quartz, determined by the crystal cut and machining properties.
  • 11.  Ultrasound transducers for medical imaging applications employ a synthetic piezoelectric ceramic, most often lead-zirconate-titanate (PZT). • The piezoelectric attributes are attained after a process of • Molecular synthesis, • Heating, • Orientation of internal dipole structures with an applied external voltage, • Cooling to permanently maintain the dipole orientation, and • Cutting into a specific shape.
  • 12.  For PZT in its natural state, no piezoelectric properties are exhibited; however, heating the material past its “Curie temperature” (i.e., 3280 C to 3650 C) and applying an external voltage causes the dipoles to align in the ceramic. • The external voltage is maintained until the material has cooled to below its Curie temperature. • Once the material has cooled, the dipoles retain their alignment.
  • 13.  At equilibrium, there is no net charge on ceramic surfaces. • When compressed, an imbalance of charge produces a voltage between the surfaces. • Similarly, when a voltage is applied between electrodes attached to both surfaces, mechanical deformation occurs.
  • 14.  The piezoelectric element is composed of aligned molecular dipoles.
  • 15.  Under the influence of mechanical pressure from an adjacent medium (e.g., an ultrasound echo), the element thickness • Contracts (at the peak pressure amplitude), • Achieves equilibrium (with no pressure) or • Expands (at the peak rarefactional pressure), • This causes realignment of the electrical dipoles to produce positive and negative surface charge.
  • 16.
  • 17.  Surface electrodes measure the voltage as a function of time.
  • 18.  An external voltage source applied to the element surfaces causes compression or expansion from equilibrium by realignment of the dipoles in response to the electrical attraction or repulsion force.
  • 19.
  • 20. Resonance Transducers  Resonance transducers for pulse echo ultrasound imaging are manufactured to operate in a “resonance” mode, whereby a voItage (commonly 150 V) of very short duration (a voltage spike of ≈1 µsec) is applied, causing the piezoelectric material to initially contract, and subsequently vibrate at a natural resonance frequency. • This frequency is selected by the “thickness cut,” due to the preferential emission of ultrasound waves whose wavelength is twice the thickness of the piezoelectric material.
  • 21.  The operating frequency is determined from the speed of sound in, and the thickness of, the piezoelectric material. • For example, a 5-MHz transducer will have a wavelength in PZT (speed of sound in PZT is ≈ 4,000 m/sec) of c 4000 m / sec λ= = = 8 × 10 −4 meters = 0.80 mm f 5 × 106 / sec
  • 22.  A short duration voltage spike causes the resonance piezoelectric element to vibrate at its natural frequency, fo, which is determined by the thickness of the transducer equal to 1/A.
  • 23.  To achieve the 5-MHz resonance frequency, a transducer element thickness of ½ X 0.8 mm = 0.4 mm is required. • Higher frequencies are achieved with thinner elements, and lower frequencies with thicker elements. • Resonance transducers transmit and receive preferentially at a single “center frequency.”
  • 24. Damping Block  The damping block, layered on the back of the piezoelectric element, absorbs the backward directed ultrasound energy and attenuates stray ultrasound signals from the housing. • This component also dampens the transducer vibration to create an ultrasound pulse width and short spatial pulse length, which is necessary to preserve detail along he beam axis (axial resolution).
  • 25.
  • 26.
  • 27.  Dampening of the vibration (also known as “ring-down”) lessens the purity of the resonance frequency and introduces a broadband frequency spectrum. • With ring-down, an increase in the bandwidth (range of frequencies) of the ultrasound pulse occurs by introducing higher and lower frequencies above and below the center (resonance) frequency.
  • 28.  The “Q factor” describes the bandwidth of the sound emanating from a transducer as fo Q= Bandwidth  where fo is the center frequency and the bandwidth is the width of the frequency distribution.
  • 29. A “high Q” transducer has a narrow bandwidth (i.e., very little damping) and a corresponding long spatial pulse length. • A “low Q” transducer has a wide bandwidth and short spatial pulse length.
  • 30.  Imaging applications require a broad bandwidth transducer in order to achieve high spatial resolution along the direction of beam travel. • Blood velocity measurements by Doppler instrumentation require a relatively narrowband transducer response in order to preserve velocity information encoded by changes in the echo frequency relative to the incident frequency.
  • 31.  Continuous-wave ultrasound transducers have a very high Q characteristic. • While the Q factor is derived from the term quality factor, a transducer with a low Q does not imply poor quality in the signal.
  • 32. Matching Layer  The matching layer provides the interface between the transducer element and the tissue and minimizes the acoustic impedance differences between the transducer and the patient. • It consists of layers of materials with acoustic impedances that are intermediate to those of soft tissue and the transducer material. • The thickness of each layer is equal to one-fourth the wavelength, determined from the center operating frequency of the transducer and speed of sound in the matching layer.
  • 33.  For example, the wavelength of sound in a matching layer with a speed of sound of 2,000 m/sec for a 5-MHz ultrasound beam is 0.4 mm. • The optimal matching layer thickness is equal to ¼λ = ¼ x 0.4 mm = 0. 1 mm. • In addition to the matching layers, acoustic coupling gel (with acoustic impedance similar to soft tissue) is used between the transducer and the skin of the patient to eliminate air pockets that could attenuate and reflect the ultrasound beam.
  • 34. Nonresonance (BroadBandwidth) “Multifrequency” Transducers  Modern transducer design coupled with digital signal processing enables “multifrequency or “multihertz” transducer operation, whereby rhe center frequency can be adjusted in he transmit mode. • Unlike the resonance transducer design, the piezoelectric element is intricately machined into a large number of small “rods,” and then filled with an epoxy resin to create a smooth surface.
  • 35.
  • 36.
  • 37.  The acoustic properties are closer to tissue than a pure PZT material, and thus provide a greater transmission efficiency of the ultrasound beam without resorting to multiple matching layers. • Multifrequency transducers have bandwidths that exceed 80% of the center frequency.
  • 38.  Excitation of the multifrequency transducer is accomplished with a short square wave burst of 150 V with one to three cycles, unlike the voltage spike used for resonance transducers. • This allows the center frequency to be selected within the limits of the transducer bandwidth.
  • 39.  Likewise, the broad bandwidth response permits the reception of echoes within a wide range of frequencies. • For instance, ultrasound pulses can be produced at a low frequency, and the echoes received at higher frequency.
  • 40.  “Harmonic imaging” is a recently introduced technique that uses this ability; • lower frequency ultrasound is transmitted into the patient, and the higher frequency harmonics (e.g., two times the transmitted center frequency) created from the interaction with contrast agents and tissues, are received as echoes.
  • 41.  Native tissue harmonic imaging has certain advantages including greater depth of penetration, noise and clutter removal, and improved lateral spatial resolution.
  • 42. Transducer Arrays  The majority of ultrasound systems employ transducers with many individual rectangular piezoelectric elements arranged in linear or curvilinear arrays. • Typically, 128 to 512 individual rectangular elements compose the transducer assembly. • Each element has a width typically less than half the wavelength and a length of several millimeters.
  • 43.  Two modes of activation are used to produce a beam. • These are the “linear” (sequential) and “phased” activation/receive modes.
  • 44. Linear Arrays  Linear array transducers typically contain 256 to 512 elements; physically these are the largest transducer assemblies.
  • 45.  In operation, the simultaneous firing of’ a small group of ≈ 20 adjacent elements produces the ultrasound beam. • The simultaneous activation produces a synthetic aperture (effetive transducer width) defined by the number of active elements.
  • 46.  Echoes are detected in the receive mode by acquiring signals from most of the transducer elements. • Subsequent “A-line” acquisition occurs by firing another group of transducer elements displaced by one or two elements.
  • 47. A rectangular field of view is produced with this transducer arrangement. • For a curvilinear array, a trapezoidal field of view is produced.
  • 48. Phased Arrays A phased-array transducer is usually composed of 64 to 128 individual elements in a smaller package than a linear array transducer. • All transducer elements are activated nearly (but not exactly) simultaneously to produce a single ultrasound beam.
  • 49.  By using time delays in the electrical activarion of the discrete elements across the face of the transducer, the ultrasound beam can be steered and focused electronically without moving the transducer. • During ultrasound signal reception, all of the transducer elements detect the returning echoes from the beam path, and sophisticated algorithms synthesize the image from the detected data.
  • 50. BEAM PROPERTIES  The ultrasound beam propagates as a longitudinal wave from the transducer surface into the propagation medium, and exhibits two distinct beam patterns: • a slightly converging beam out to a distance • specified by the geometry and frequency of the transducer (the near field), and a diverging beam beyond that point (the far field).
  • 51.  For an unfocused, single-element transducer, the length of the near field is determined by the transducer diameter and the frequency of the transmitted sound.
  • 52.  For multiple transducer element arrays, an “effective” transducer diameter is determined by the excitation of a group of’ transducer elements. • Because of the interactions of each of the individual beams and the ability to focus and steer the overall beam, the formulas for a single-element, unfocused transducer are not directly applicable.
  • 53. The Near Field  The near field, also known as the Fresnel zone, is adjacent to the transducer face and has a converging beam profile. • Beam convergence in the near field occurs because of multiple constructive and destructive interference patterns of the ultrasound waves from the transducer surface.
  • 54.  Huygen’s principle describes a large transducer surface as an infinite number of point sources of sound energy where each point is characterized as a radial emitter. • By analogy, a pebble dropped in a quiet pond creates a radial wave pattern.
  • 55.  As individual wave patterns interact, the peaks and troughs from adjacent sources constructively and destructively interfere, causing the beam profile to be tightly collimated in the near field.
  • 56.  The ultrasound beam path is thus largely confined to the dimensions of the active portion of the transducer surface, with the beam diameter converging to approximately half the transducer diameter at the end of the near field.
  • 57.  The near field length is dependent on the transducer frequency and diameter: d 2 r2 Near field length = = 4λ λ • where d is the transducer diameter, r is the transducer radius, and λ is the wavelength of ultrasound in the propagation medium.
  • 58. soft tissue, λ = 1.54mm/f(MHz), and the near field length can be expressed as a function of frequency:  In ( ) d2 mm 2 ( MHz ) Near field length = ( mm) 4 × 1.54
  • 59.  A higher transducer frequency (shorter wavelength) will result in a longer near field, as will a larger diameter element.
  • 60.  For a 10-mm-diameter transducer, the near field extends 5.7 cm at 3.5 MHz and 16.2 cm at 10 MHz in soft tissue. • For a 15-mm-diameter transducer, the corresponding near field lengths are 12.8 and 36.4 cm, respectively.
  • 61.  Lateral resolution (the ability of the system to resolve objects in a direction perpendicular to the beam direction) is dependent on the beam diameter and is best at the end of the near field for a single-element transducer. • Lateral resolution is worst in areas close to and far from the transducer surface.
  • 62.  Pressure amplitude characteristics in the near field are very complex, caused by the constructive and destructive interference wave patterns of the ultrasound beam. • Peak ultrasound pressure occurs at the end of the near field, corresponding to the minimum beam diameter for a single-element transducer.
  • 63.  Pressures vary rapidly from peak compression to peak rarefaction several times during transit through the near field. • Only when the far field is reached do the ultrasound pressure variations decrease continuously.
  • 64.  The far field is also known as the Fraunhofer zone, and is where the beam diverges. • For a large-area single-element transducer, the angle of ultrasound beam divergence, 0, for the far field is given by λ sin θ = 1.22 d • where d is the effective diameter of the transducer and λ is the wavelength; both must have the same units of distance.
  • 65.  Less beam divergence occurs with highfrequency, large-diameter transducers. • Unlike the near field, where beam intensity varies from maximum to minimum to maximum in a converging beam, ultrasound intensity in the far field decreases monotonically with distance.
  • 66. Transducer Array Beam Formation and Focusing  In a transducer array, the narrow piezoelectric element width (typically less than one wavelength) produces a diverging beam at a distance very close to the transducer face. • Formation and convergence of the ultrasound beam occurs with the operation of several or all of the transducer elements at the same time.
  • 67.  Transducer elements in a linear array that are fired simultaneously produce an effective transducer width equal to the sum of the widths of the individual elements. • Individual beams interact via constructive and destructive interference to produce a collimated beam that has properties similar to the properties of a single transducer of the same size.
  • 68.  With a phased-array transducer, the beam is formed by interaction of the individual wave fronts from each transducer, each with a slight difference in excitation time. • Minor phase differences of adjacent beams form constructive and destructive wave summations that steer or focus the beam profile.
  • 69. COMMON TRANSDUCERS USED IN CLINICAL SETTING
  • 70. STRAIGHT LINEAR ARRAY PROBE The straight linear array probe is designed for superficial imaging. The crystals are aligned in a linear fashion within a flat head and produce sound waves in a straight line. The image produced is rectangular in shape.
  • 71.  This probe has higher frequencies (5–13 MHz), which provides better resolution and less penetration.  Therefore, this probe is ideal for imaging superficial structures and in ultrasoundguided procedures.
  • 72. Vascular access Evaluate for deep venous thrombosis Skin and soft tissue for abscess, foreign body Musculoskeletal—tendons, bones, muscles
  • 73.
  • 74. CURVILINEAR ARRAY PROBE  The curvilinear array or convex probe is used for scanning deeper structures. The crystals are aligned along a curved surface and cause a fanning out of the beam, which results in a field of view that is wider than the probe’s footprint.
  • 75.  The image generated is sector shaped. These probes have frequencies ranging between 1 and 8 MHz, which allows for greater penetration, but less resolution. These probes are most often used in abdominal and pelvic applications.  They are also useful in certain musculoskeletal evaluations or procedures when deeper anatomy needs to be imaged or in obese patients.
  • 76.  Abdominal aorta  Biliary/gallbladder/liver/pancreas  Abdominal portion of FAST exam  Kidney and bladder evaluation  Transabdominal pelvic evaluation
  • 77.
  • 78. ENDOCAVITARY PROBE  The endocavitary probe also has a curved face, but a much higher frequency (8–13 MHz) than the curvilinear probe.  This probe’s elongated shape allows it to be inserted close to the anatomy being evaluated.
  • 79.  The curved face creates a wide field of view of almost 180° and its high frequencies provide superior resolution . This probe is used most commonly for gynecological applications, but can also be used for intraoral evaluation of peritonsillar abscesses.  Transvaginal ultrasound  Intraoral
  • 80.
  • 81. PHASED ARRAY PROBE  Phased array probes (Fig. 4-4a) have crystals that are grouped closely together.  The timing of the electrical pulses that are applied to the crystals varies and they are fired in an oscillating manner.
  • 82.  The sound waves that are generated originate from a single point and fan outward, creating a sector-type image. This probe has a smaller and flatter footprint than the curvilinear one, which allows the user to maneuver more easily between the ribs and small spaces. These probes have frequencies between 2 and 8 MHz.
  • 83.
  • 84. IVUS PROBE  IVUS is a miniature ultrasound probe positioned at the tip of a coronary catheter.  The probe emits ultrasound frequencies, typically at 20-45 MHz, and the signal is reflected from surrounding tissue and reconstructed into a real-time tomographic gray-scale image.
  • 85.
  • 86.
  • 88.  In ultrasound, the major factor that limits the spatial resolution and visibility of detail is the volume of the acoustic pulse.
  • 89.  The axial, lateral, and elevational (slice thickness) dimensions determine the minimal volume element.
  • 90.  Each dimension has an effect on the resolvability of objects in the image.
  • 91. Axial Resolution  Axial resolution (also known as linear, range, longitudinal, or depth resolution) refers to the ability to discern two closely spaced objects in the direction of the beam. • Achieving good axial resolution requires that the returning echoes be distinct without overlap.
  • 92.  The minimal required separation distance between two reflectors is onehalf of the spatial pulse length (SPL) to avoid the overlap of returning echoes, as the distance traveled between two reflectors is twice the separation distance.
  • 93.  Objects spaced closer than ½ SPL will not be resolved.
  • 94.  The SPL is the number of cycles emitted per pulse by the transducer multiplied by the wavelength. • Shorter pulses, producing better axial resolution, can be achieved with greater damping of the transducer element (to reduce the pulse duration and number of cycles) or with higher frequency (to reduce wavelength).
  • 95.  For imaging applications, the ultrasound pulse typically consists of three cycles. • At 5 MHz (wavelength of 0.31 mm), the SPL is about 3 x 0.31 0.93 mm, which provides an axial resolution of /2(0.93 mm) = 0.47 mm.
  • 96.  At a given frequency, shorter pulse lengths require heavy damping and low Q, broad-bandwidth operation. • For a constant damping factor, higher frequencies (shorter wavelengths) give better axial resolution, but the imaging depth is reduced. • Axial resolution remains constant with depth.
  • 97. Lateral Resolution  Lateral resolution, also known as azimuthal resolution, refers to the ability to discern as separate two closely spaced objects perpendicular to the beam direction.
  • 98.  For both single element transducers and multielement array transducers, the beam diameter determines the lateral resolution.
  • 99.  Since the beam diameter varies with the distance from the transducer in the near and far field, the lateral resolution is depth dependent. • The best lateral resolution occurs at the near field—far field face.
  • 100.  At this depth, the effective beam diameter is approximately equal to half the transducer diameter. • In the far field, the beam diverges and substantially reduces the lateral resolution.
  • 101.  The typical lateral resolution for an unfocused transducer is approximately 2 to 5 mm. • A focused transducer uses an acoustic lens (a curved acoustic material analogous to an optical lens) to decrease the beam diameter at a specified distance from the transducer.
  • 102.  With an acoustic lens, lateral resolution at the near field-far field interface is traded for better lateral resolution at a shorter depth, but the far field beam divergence is substantially increased. • The lateral resolution of linear and curvilinear array transducers can be varied.
  • 103. Elevational Resolution  The elevational or slice-thickness dimension of the ultrasound beam is perpendicular to the image plane. • Slice thickness plays a significant part in image resolution, particularly with respect to volume averaging of acoustic details in the regions dose to the transducer and in the far field beyond the focal zone.
  • 104.  Elevational resolution is dependent on the transducer element height in much the same way that the lateral resolution is dependent on the transducer element width.
  • 105.  Slice thickness is typically the worst measure of resolution for array transducers. • Use of a fixed focaI length lens across the entire surface of the array provides improved elevational resolution at the focal distance.
  • 106.  Unfortunately, this compromises resolution due to partial volume averaging before and after the elevational focal zone (elevational resolution quality control phantom image shows the effects of variable resolution with depth.
  • 107.  Multiple linear array transducers with five to seven rows, known as 1.5dimensional (1.5-D) transducer arrays, have the ability to steer and focus the beam in the elevational dimension.