Radiation dose modulation and dose
reduction techniques
Introduction to presentation
CT Dose Index
CTDI100
CTDIvol
DLP
Effective dose
CT dose reduction techniques
AEC
Filter
Reconstruction
CT Dose Index
CT dose index (CTDI) is the standardized
measure of the radiation output of a CT system,
Currently used measures are the CTDI100 and
CTDIw, modified to CTDIvol for modern helical
scanners. 
CTDI phantoms(PMMA)
Measurement of CTDI
• Ionisation pencil chamber
– Air-filled chamber
– X-rays cause ionisations in
chamber, causing current to
flow, proportional to exposure
– Instant readings, good
accuracy, easy to use
100 mm
Measurement of CTDI
• Single slice measurement
(z-axis)
Dose
T
Nominal
beam width
( )dzzD
T
1
CTDI
-
∫
+∞
∞
=
( )dzzD
T
1
CTDI
50
50-
100
∫
+
=
-50 +50
∫
+
−
=
50
50
D(z)dz
n.T
1
CTDI100
n = no. slices imaged simultaneously
T = nominal imaged width
Davge
T
100Davge
CTDI100
×
=
CTDI100
• CTDI100 is a measure of radiation dose on a 100mm
long pencil ionization chamber.
• Centre
• Periphery - 1 cm depth
(mean of 4 positions)
C
P1
P2
P3
P4
320 mm
‘Body’
‘Head’
160 mm
140
mm
Measurement of MSAD(Multiple Scan
Average Dose)
CTDI ≈ ‘Series’ dose
Multi-scan
‘Series’ dose
CTDI×
×
=
I
TN
MSAD
T
T is the nominal scan width (mm)
I is the distance between scans (mm)
I
N is the number of scans
N × T is the total nominal scan width
CTDI100
• Typical CTDI100 values (mGy)
40
Head
Body
40
4040
40
20
2020
20
10
Periphery : centre ≅ 1:1 Periphery : centre ≅ 2:1
Derivatives of CTDI100
• CTDIW
• Weighted average CTDI: represents the average
dose for contiguous irradiation
• CTDIw = 1/3 CTDI100,C + 2/3 CTDI100,P
CTDIC CTDIP
Derivatives of CTDI100
• CTDI100 & CTDIw refer to a pitch of 1
• Average absorbed dose dependent on pitch
Contiguous Extended Overlapping
CTDIvol
• Volume CTDI: Considers contiguous scanning
• CTDIvol = CTDIw
Pitch
Pitch = 1
CTDIvol = CTDIw
Pitch = 2
CTDIvol = CTDIw
2
Pitch = 0.5
CTDIvol = 2 x CTDIw
Radiation risk: DLP
• CTDIvol is a measure of absorbed dose, energy
absorbed per unit mass
• To measure radiation risk from stochastic effects
the total energy absorbed must be considered
• In CT this can be estimated using the Dose Length
Product, DLP
DLP = CTDIvol . L (mGy.cm)
DLP1 ≅ DLP2
where L = scan length
Dose length product
L1
T
Pitch 1
8 rotations
Pitch 2
8 rotations
L2
T
as CTDIvol2 = CTDIvol1
2
Estimates of effective dose (E)
• Can be obtained from DLP
• Effective dose = DLP. CF (mSv)
•
•
•
•
•
•
Conversion factors not scanner specific or location
specific
Region of body Conversion factor,EDLP
(mSv mGy-1
cm-1
)
Head & neck 0.0031
Head 0.0021
Neck 0.0059
Chest 0.014
Abdomen& pelvis 0.015
Trunk 0.015
CT dose reduction techniques
AEC systems have a number of
potential advantages, including better
control of patient radiation dose,
avoidance of photon starvation
artifacts, reduced load on the x-ray
tube, and the maintenance of image
quality in spite of different attenuation
values on CT scans
Types of AEC
Patient-Size AEC
Z-axis AEC
Rotational or Angular AEC
Operation of AEC Systems on
Different Multidetector CT Scanners
Exposure 3D
CARE Dose 4D
Patient-Size AECPatient size AEC: the tube current is adjusted
based on the overall size of the patient to reduce the
variation in image quality between small patients and
large patients. For a given patient size, the
appropriate milliamperage is selected and is used for
the entire examination or scan series
Hi-mA
Lo-mA
Z-axis AEC: tube current is modulated
according to patient attenuation along the z-
axis.The goal is to reduce the variation in image
quality of images from the same series.
attenuation
Patient-Size AECRotational AEC: The tube current is decreased
and increased rapidly (modulated) during the course
of each rotation to compensate for differences in
attenuation between lateral (left-right) and A-P
(anterior-posterior) projections .
highattenuationlow attenuation
Combination of AEC functions
• Tube current is adjusted during scanning to compensate for
attenuation differences – dose applied to patient only where
needed, avoiding dose where it isn’t
mA
position
On line’ modulation–uses attenuation data from previous rotation–
adapts tube current to patient attenuation ‘on the fly’
Scan projection radiographs (SPRs, known as scout,
scanogram or topogram views) are the main way that
AEC systems assess the attenuation of the patient in
order to set the tube current
To obtain correct attenuation data from SPR
always centre the patient carefully
Patient is positioned in the
isocenter –
optimal dose and image quality
Patient is positioned too high -
increased mAsPatient is positioned too low -
reduced mAs and increased noise
Toshiba:defining image quality
requirements
Specify s.d. level (or ‘image quality level)–patient
mAcalculated to achieve this noise level at any
scan parameter settings
Set min & max mA
Benefits of CT scanner
AEC
Consistent image quality
Potential for dose reduction
through
exposure optimisation
Reduced tube loading
Extended scan runs(OLP)
Reduction in photon starvation
artifact
30
• Images along length of phantom (no AEC)
•
•
•
•
•
•
•
•
Constant mA
Testing the AEC
31
Testing the AEC
• Measure noise with AEC off and on
• Monitor mA, CTDIvol
0
4
8
12
16
20
24
28
-150 -100 -50 0 50 100 150
Z-position (mm)
Noise(%)
automA off
Noise Index 12
Increased
mA
Decreased
mA
Constant mA
Same
mA
Increased
mA
Decreased
mA
32
Coronal view Sagittal view
z-axis
AEC off
z-axis
AEC on
Noise
increases
Constant
noise
Testing the AEC – Viewing with MPR
photon starvation artifact
without angular mA modulation with angular mA modulation
Bowtie Filter
Longer bowtie path lines up with shorter
patient path
Reduce x-ray scatter (noise and artifact)
Maintain uniform x-ray at detector
Reduce surface dose by 50%
• Retrospectively gated CCTA(dose modulation )
~15 mSv100%
20% ~30% reduction
Image obtained in middiastole
(75% of R-R interval)
reconstructed using 1.5-mm slice
thickness at high tube current
shows low-noise
Image obtained in midsystole
(30% of R-R interval)
reconstructed using 1.5-mm slice
thickness shows higher-noise
image obtained at low radiation
dose
• Prospectively gated CCTA
~ 1 mSv
Prospective Gating of CTA(snapshot plus)
40mm
5mm overlap
Scan range
Conventional technology
without Dose Shield
SOMATOM Definition AS+ with
Adaptive Dose Shield
Scan range
Dynamic Collimation
• In helical scanning extra rotations are needed at end of imaged
volume
– Significant extra dose: wide beam widths and short scans
• Dynamic collimation - collimator blades open and close
asymetrically at start and end of scan
(a) conventional and (b) adaptive section collimation CT scanning protocols. For
adaptive section collimation, shape of x-ray cone beam at beginning and end of spiral
acquisition is controlled by two collimators made of absorbent material.
Image reconstruction(FBP)
attenuation
detectorposition
90°2 projections4 projections8 projections16 projections
attenuation detector position
0°
0(3) 0(9)
0(7) 0(5)
6 6
6 6
10 14
12
12
12 12
5 7
5 7
3 9
7 5
3
5
9
7
816
12 12
+1-1
-2+2
Iterative reconstruction
Iterative reconstruction
120 kVp and 300 mA FBP
120 kVp and 300 mA ASiR 100%
120 kVp and 300 mA ASiR 50%
Dose reduction
Dose reduction

More Related Content

What's hot

CT Scan Image reconstruction
CT Scan Image reconstructionCT Scan Image reconstruction
CT Scan Image reconstruction
Gunjan Patel
 

What's hot (20)

CT RADIATION DOSE REDUCTION
CT RADIATION DOSE REDUCTION CT RADIATION DOSE REDUCTION
CT RADIATION DOSE REDUCTION
 
magnetic resonance in angiography
magnetic resonance in  angiography magnetic resonance in  angiography
magnetic resonance in angiography
 
Dual Energy CT
Dual Energy CTDual Energy CT
Dual Energy CT
 
MDCT Principles and Applications- Avinesh Shrestha
MDCT Principles and Applications- Avinesh ShresthaMDCT Principles and Applications- Avinesh Shrestha
MDCT Principles and Applications- Avinesh Shrestha
 
MRI Parameters for Imaging
MRI Parameters for ImagingMRI Parameters for Imaging
MRI Parameters for Imaging
 
ct image quality
ct image qualityct image quality
ct image quality
 
BASIC MRI SEQUENCES
BASIC MRI SEQUENCESBASIC MRI SEQUENCES
BASIC MRI SEQUENCES
 
Computed Tomography Dose Index
Computed Tomography Dose IndexComputed Tomography Dose Index
Computed Tomography Dose Index
 
Dose reduction technique in ct scan
Dose reduction technique in ct scanDose reduction technique in ct scan
Dose reduction technique in ct scan
 
Quality Assurance Programme in Computed Tomography
Quality Assurance Programme in Computed TomographyQuality Assurance Programme in Computed Tomography
Quality Assurance Programme in Computed Tomography
 
Basic Pulse Sequences In MRI
Basic Pulse Sequences In MRIBasic Pulse Sequences In MRI
Basic Pulse Sequences In MRI
 
Ct protocols of thorax
Ct protocols of thoraxCt protocols of thorax
Ct protocols of thorax
 
Physics of Multidetector CT Scan
Physics of Multidetector CT ScanPhysics of Multidetector CT Scan
Physics of Multidetector CT Scan
 
MRI Coil and Gradient power-point slide pk
MRI Coil and Gradient power-point slide pkMRI Coil and Gradient power-point slide pk
MRI Coil and Gradient power-point slide pk
 
Ct head protocols
Ct head protocolsCt head protocols
Ct head protocols
 
Image Quality, Artifacts and it's Remedies in CT-Avinesh Shrestha
Image Quality, Artifacts and it's Remedies in CT-Avinesh ShresthaImage Quality, Artifacts and it's Remedies in CT-Avinesh Shrestha
Image Quality, Artifacts and it's Remedies in CT-Avinesh Shrestha
 
Presentation1. ct physics.
Presentation1. ct physics.Presentation1. ct physics.
Presentation1. ct physics.
 
Encoding and image formation
Encoding and image formationEncoding and image formation
Encoding and image formation
 
CT image quality and image resolution.pptx
CT image quality and image resolution.pptxCT image quality and image resolution.pptx
CT image quality and image resolution.pptx
 
CT Scan Image reconstruction
CT Scan Image reconstructionCT Scan Image reconstruction
CT Scan Image reconstruction
 

Viewers also liked

European guidance on estimating population doses from medical xray
European guidance on estimating population doses from medical xrayEuropean guidance on estimating population doses from medical xray
European guidance on estimating population doses from medical xray
Ana Pires
 
Rsna2012 pediatric dosemapvstdphantom
Rsna2012 pediatric dosemapvstdphantomRsna2012 pediatric dosemapvstdphantom
Rsna2012 pediatric dosemapvstdphantom
JeffYanof
 
Implementation of an audit and dose reduction program for ct matyagin
Implementation of an audit and dose reduction program for ct matyaginImplementation of an audit and dose reduction program for ct matyagin
Implementation of an audit and dose reduction program for ct matyagin
Leishman Associates
 
Optimization of ct scan protocol in acute abdomen 2003 revised aa
Optimization of ct scan protocol in acute abdomen 2003 revised aaOptimization of ct scan protocol in acute abdomen 2003 revised aa
Optimization of ct scan protocol in acute abdomen 2003 revised aa
Hisham Khatib
 
iMinds The Conference: Johan de Mey
iMinds The Conference: Johan de MeyiMinds The Conference: Johan de Mey
iMinds The Conference: Johan de Mey
imec
 
Spectral x-ray photon counting
Spectral x-ray photon countingSpectral x-ray photon counting
Spectral x-ray photon counting
Gunnar Maehlum
 

Viewers also liked (20)

European guidance on estimating population doses from medical xray
European guidance on estimating population doses from medical xrayEuropean guidance on estimating population doses from medical xray
European guidance on estimating population doses from medical xray
 
Rsna2012 pediatric dosemapvstdphantom
Rsna2012 pediatric dosemapvstdphantomRsna2012 pediatric dosemapvstdphantom
Rsna2012 pediatric dosemapvstdphantom
 
Implementation of an audit and dose reduction program for ct matyagin
Implementation of an audit and dose reduction program for ct matyaginImplementation of an audit and dose reduction program for ct matyagin
Implementation of an audit and dose reduction program for ct matyagin
 
CT ITS BASIC PHYSICS
CT ITS BASIC PHYSICSCT ITS BASIC PHYSICS
CT ITS BASIC PHYSICS
 
Optimization of ct scan protocol in acute abdomen 2003 revised aa
Optimization of ct scan protocol in acute abdomen 2003 revised aaOptimization of ct scan protocol in acute abdomen 2003 revised aa
Optimization of ct scan protocol in acute abdomen 2003 revised aa
 
Qaulity Control of CT
Qaulity Control of CTQaulity Control of CT
Qaulity Control of CT
 
iMinds The Conference: Johan de Mey
iMinds The Conference: Johan de MeyiMinds The Conference: Johan de Mey
iMinds The Conference: Johan de Mey
 
Basic principles of CT scanning
Basic principles of CT scanningBasic principles of CT scanning
Basic principles of CT scanning
 
Computer Tomography (CT Scan)
Computer Tomography (CT Scan)Computer Tomography (CT Scan)
Computer Tomography (CT Scan)
 
Ct Basics
Ct BasicsCt Basics
Ct Basics
 
What is New in Cardiac CT? In Search of the Comprehensive and Conclusive Hear...
What is New in Cardiac CT? In Search of the Comprehensive and Conclusive Hear...What is New in Cardiac CT? In Search of the Comprehensive and Conclusive Hear...
What is New in Cardiac CT? In Search of the Comprehensive and Conclusive Hear...
 
Research ID's - University of Cape Town Libraries
Research ID's - University of Cape Town LibrariesResearch ID's - University of Cape Town Libraries
Research ID's - University of Cape Town Libraries
 
Spectral x-ray photon counting
Spectral x-ray photon countingSpectral x-ray photon counting
Spectral x-ray photon counting
 
Integrated Detector Electronics (IDEAS) ASIC product update
Integrated Detector Electronics (IDEAS) ASIC product updateIntegrated Detector Electronics (IDEAS) ASIC product update
Integrated Detector Electronics (IDEAS) ASIC product update
 
Computed Tomography
Computed TomographyComputed Tomography
Computed Tomography
 
Ct tube and detectors
Ct tube and detectorsCt tube and detectors
Ct tube and detectors
 
MR physics 101
 MR physics 101 MR physics 101
MR physics 101
 
Cardiac ct how i do it september 2011
Cardiac ct how i do it september 2011Cardiac ct how i do it september 2011
Cardiac ct how i do it september 2011
 
Mri physics-uk
Mri physics-ukMri physics-uk
Mri physics-uk
 
triple scan protocol (Dr. Gross) - a new and effective protocol for 3D planni...
triple scan protocol (Dr. Gross) - a new and effective protocol for 3D planni...triple scan protocol (Dr. Gross) - a new and effective protocol for 3D planni...
triple scan protocol (Dr. Gross) - a new and effective protocol for 3D planni...
 

Similar to Dose reduction

L10 Patient Dose
L10 Patient DoseL10 Patient Dose
L10 Patient Dose
lidgor
 

Similar to Dose reduction (20)

-Lecture 2.ppt
-Lecture 2.ppt-Lecture 2.ppt
-Lecture 2.ppt
 
4.pptx
4.pptx4.pptx
4.pptx
 
Dual energy imaging and digital tomosynthesis: Innovative X-ray based imaging...
Dual energy imaging and digital tomosynthesis: Innovative X-ray based imaging...Dual energy imaging and digital tomosynthesis: Innovative X-ray based imaging...
Dual energy imaging and digital tomosynthesis: Innovative X-ray based imaging...
 
CT Physics
CT PhysicsCT Physics
CT Physics
 
medical equipment lec 6 CT scan
medical equipment lec 6 CT scanmedical equipment lec 6 CT scan
medical equipment lec 6 CT scan
 
L10 Patient Dose
L10 Patient DoseL10 Patient Dose
L10 Patient Dose
 
CT dose reduction
CT dose reductionCT dose reduction
CT dose reduction
 
QUALITY ASSURANCE IN LINAC AND CYBERKNIFE.pptx
QUALITY ASSURANCE IN LINAC AND CYBERKNIFE.pptxQUALITY ASSURANCE IN LINAC AND CYBERKNIFE.pptx
QUALITY ASSURANCE IN LINAC AND CYBERKNIFE.pptx
 
Computed Tomography
Computed TomographyComputed Tomography
Computed Tomography
 
computed Tomography
computed Tomographycomputed Tomography
computed Tomography
 
QUALITY-ASSURANCE---MAMMOGRAPHY.pdf
QUALITY-ASSURANCE---MAMMOGRAPHY.pdfQUALITY-ASSURANCE---MAMMOGRAPHY.pdf
QUALITY-ASSURANCE---MAMMOGRAPHY.pdf
 
3.ppt
3.ppt3.ppt
3.ppt
 
Helical Tomotherapy
Helical TomotherapyHelical Tomotherapy
Helical Tomotherapy
 
Arc therapy [autosaved] [autosaved]
Arc therapy [autosaved] [autosaved]Arc therapy [autosaved] [autosaved]
Arc therapy [autosaved] [autosaved]
 
Tomo modulation
Tomo modulationTomo modulation
Tomo modulation
 
CT based Image Guided Radiotherapy - Physics & QA
CT based Image Guided Radiotherapy - Physics & QACT based Image Guided Radiotherapy - Physics & QA
CT based Image Guided Radiotherapy - Physics & QA
 
Quality assurance
Quality  assuranceQuality  assurance
Quality assurance
 
1 Medical Physics Radiation Risk Garringer and Aliison Dr. Muhammad Bin Zulfiqar
1 Medical Physics Radiation Risk Garringer and Aliison Dr. Muhammad Bin Zulfiqar1 Medical Physics Radiation Risk Garringer and Aliison Dr. Muhammad Bin Zulfiqar
1 Medical Physics Radiation Risk Garringer and Aliison Dr. Muhammad Bin Zulfiqar
 
Hrct i
Hrct iHrct i
Hrct i
 
Technical aspects of dental CBCT - Dr. Kavan Gandhi
Technical aspects of dental CBCT - Dr. Kavan GandhiTechnical aspects of dental CBCT - Dr. Kavan Gandhi
Technical aspects of dental CBCT - Dr. Kavan Gandhi
 

Recently uploaded

The basics of sentences session 3pptx.pptx
The basics of sentences session 3pptx.pptxThe basics of sentences session 3pptx.pptx
The basics of sentences session 3pptx.pptx
heathfieldcps1
 

Recently uploaded (20)

2024-NATIONAL-LEARNING-CAMP-AND-OTHER.pptx
2024-NATIONAL-LEARNING-CAMP-AND-OTHER.pptx2024-NATIONAL-LEARNING-CAMP-AND-OTHER.pptx
2024-NATIONAL-LEARNING-CAMP-AND-OTHER.pptx
 
HMCS Vancouver Pre-Deployment Brief - May 2024 (Web Version).pptx
HMCS Vancouver Pre-Deployment Brief - May 2024 (Web Version).pptxHMCS Vancouver Pre-Deployment Brief - May 2024 (Web Version).pptx
HMCS Vancouver Pre-Deployment Brief - May 2024 (Web Version).pptx
 
Exploring_the_Narrative_Style_of_Amitav_Ghoshs_Gun_Island.pptx
Exploring_the_Narrative_Style_of_Amitav_Ghoshs_Gun_Island.pptxExploring_the_Narrative_Style_of_Amitav_Ghoshs_Gun_Island.pptx
Exploring_the_Narrative_Style_of_Amitav_Ghoshs_Gun_Island.pptx
 
Towards a code of practice for AI in AT.pptx
Towards a code of practice for AI in AT.pptxTowards a code of practice for AI in AT.pptx
Towards a code of practice for AI in AT.pptx
 
Unit 3 Emotional Intelligence and Spiritual Intelligence.pdf
Unit 3 Emotional Intelligence and Spiritual Intelligence.pdfUnit 3 Emotional Intelligence and Spiritual Intelligence.pdf
Unit 3 Emotional Intelligence and Spiritual Intelligence.pdf
 
Key note speaker Neum_Admir Softic_ENG.pdf
Key note speaker Neum_Admir Softic_ENG.pdfKey note speaker Neum_Admir Softic_ENG.pdf
Key note speaker Neum_Admir Softic_ENG.pdf
 
The basics of sentences session 3pptx.pptx
The basics of sentences session 3pptx.pptxThe basics of sentences session 3pptx.pptx
The basics of sentences session 3pptx.pptx
 
Holdier Curriculum Vitae (April 2024).pdf
Holdier Curriculum Vitae (April 2024).pdfHoldier Curriculum Vitae (April 2024).pdf
Holdier Curriculum Vitae (April 2024).pdf
 
FSB Advising Checklist - Orientation 2024
FSB Advising Checklist - Orientation 2024FSB Advising Checklist - Orientation 2024
FSB Advising Checklist - Orientation 2024
 
UGC NET Paper 1 Mathematical Reasoning & Aptitude.pdf
UGC NET Paper 1 Mathematical Reasoning & Aptitude.pdfUGC NET Paper 1 Mathematical Reasoning & Aptitude.pdf
UGC NET Paper 1 Mathematical Reasoning & Aptitude.pdf
 
SOC 101 Demonstration of Learning Presentation
SOC 101 Demonstration of Learning PresentationSOC 101 Demonstration of Learning Presentation
SOC 101 Demonstration of Learning Presentation
 
COMMUNICATING NEGATIVE NEWS - APPROACHES .pptx
COMMUNICATING NEGATIVE NEWS - APPROACHES .pptxCOMMUNICATING NEGATIVE NEWS - APPROACHES .pptx
COMMUNICATING NEGATIVE NEWS - APPROACHES .pptx
 
Python Notes for mca i year students osmania university.docx
Python Notes for mca i year students osmania university.docxPython Notes for mca i year students osmania university.docx
Python Notes for mca i year students osmania university.docx
 
On_Translating_a_Tamil_Poem_by_A_K_Ramanujan.pptx
On_Translating_a_Tamil_Poem_by_A_K_Ramanujan.pptxOn_Translating_a_Tamil_Poem_by_A_K_Ramanujan.pptx
On_Translating_a_Tamil_Poem_by_A_K_Ramanujan.pptx
 
Understanding Accommodations and Modifications
Understanding  Accommodations and ModificationsUnderstanding  Accommodations and Modifications
Understanding Accommodations and Modifications
 
General Principles of Intellectual Property: Concepts of Intellectual Proper...
General Principles of Intellectual Property: Concepts of Intellectual  Proper...General Principles of Intellectual Property: Concepts of Intellectual  Proper...
General Principles of Intellectual Property: Concepts of Intellectual Proper...
 
How to Create and Manage Wizard in Odoo 17
How to Create and Manage Wizard in Odoo 17How to Create and Manage Wizard in Odoo 17
How to Create and Manage Wizard in Odoo 17
 
Fostering Friendships - Enhancing Social Bonds in the Classroom
Fostering Friendships - Enhancing Social Bonds  in the ClassroomFostering Friendships - Enhancing Social Bonds  in the Classroom
Fostering Friendships - Enhancing Social Bonds in the Classroom
 
Beyond_Borders_Understanding_Anime_and_Manga_Fandom_A_Comprehensive_Audience_...
Beyond_Borders_Understanding_Anime_and_Manga_Fandom_A_Comprehensive_Audience_...Beyond_Borders_Understanding_Anime_and_Manga_Fandom_A_Comprehensive_Audience_...
Beyond_Borders_Understanding_Anime_and_Manga_Fandom_A_Comprehensive_Audience_...
 
Interdisciplinary_Insights_Data_Collection_Methods.pptx
Interdisciplinary_Insights_Data_Collection_Methods.pptxInterdisciplinary_Insights_Data_Collection_Methods.pptx
Interdisciplinary_Insights_Data_Collection_Methods.pptx
 

Dose reduction

  • 1. Radiation dose modulation and dose reduction techniques
  • 2. Introduction to presentation CT Dose Index CTDI100 CTDIvol DLP Effective dose CT dose reduction techniques AEC Filter Reconstruction
  • 3. CT Dose Index CT dose index (CTDI) is the standardized measure of the radiation output of a CT system, Currently used measures are the CTDI100 and CTDIw, modified to CTDIvol for modern helical scanners. 
  • 5.
  • 6. Measurement of CTDI • Ionisation pencil chamber – Air-filled chamber – X-rays cause ionisations in chamber, causing current to flow, proportional to exposure – Instant readings, good accuracy, easy to use 100 mm
  • 7. Measurement of CTDI • Single slice measurement (z-axis) Dose T Nominal beam width ( )dzzD T 1 CTDI - ∫ +∞ ∞ = ( )dzzD T 1 CTDI 50 50- 100 ∫ + = -50 +50 ∫ + − = 50 50 D(z)dz n.T 1 CTDI100 n = no. slices imaged simultaneously T = nominal imaged width Davge T 100Davge CTDI100 × =
  • 8. CTDI100 • CTDI100 is a measure of radiation dose on a 100mm long pencil ionization chamber. • Centre • Periphery - 1 cm depth (mean of 4 positions) C P1 P2 P3 P4 320 mm ‘Body’ ‘Head’ 160 mm 140 mm
  • 9. Measurement of MSAD(Multiple Scan Average Dose) CTDI ≈ ‘Series’ dose Multi-scan ‘Series’ dose CTDI× × = I TN MSAD T T is the nominal scan width (mm) I is the distance between scans (mm) I N is the number of scans N × T is the total nominal scan width
  • 10. CTDI100 • Typical CTDI100 values (mGy) 40 Head Body 40 4040 40 20 2020 20 10 Periphery : centre ≅ 1:1 Periphery : centre ≅ 2:1
  • 11. Derivatives of CTDI100 • CTDIW • Weighted average CTDI: represents the average dose for contiguous irradiation • CTDIw = 1/3 CTDI100,C + 2/3 CTDI100,P CTDIC CTDIP
  • 12. Derivatives of CTDI100 • CTDI100 & CTDIw refer to a pitch of 1 • Average absorbed dose dependent on pitch Contiguous Extended Overlapping
  • 13. CTDIvol • Volume CTDI: Considers contiguous scanning • CTDIvol = CTDIw Pitch Pitch = 1 CTDIvol = CTDIw Pitch = 2 CTDIvol = CTDIw 2 Pitch = 0.5 CTDIvol = 2 x CTDIw
  • 14. Radiation risk: DLP • CTDIvol is a measure of absorbed dose, energy absorbed per unit mass • To measure radiation risk from stochastic effects the total energy absorbed must be considered • In CT this can be estimated using the Dose Length Product, DLP
  • 15. DLP = CTDIvol . L (mGy.cm) DLP1 ≅ DLP2 where L = scan length Dose length product L1 T Pitch 1 8 rotations Pitch 2 8 rotations L2 T as CTDIvol2 = CTDIvol1 2
  • 16. Estimates of effective dose (E) • Can be obtained from DLP • Effective dose = DLP. CF (mSv) • • • • • • Conversion factors not scanner specific or location specific Region of body Conversion factor,EDLP (mSv mGy-1 cm-1 ) Head & neck 0.0031 Head 0.0021 Neck 0.0059 Chest 0.014 Abdomen& pelvis 0.015 Trunk 0.015
  • 17. CT dose reduction techniques AEC systems have a number of potential advantages, including better control of patient radiation dose, avoidance of photon starvation artifacts, reduced load on the x-ray tube, and the maintenance of image quality in spite of different attenuation values on CT scans
  • 18. Types of AEC Patient-Size AEC Z-axis AEC Rotational or Angular AEC Operation of AEC Systems on Different Multidetector CT Scanners Exposure 3D CARE Dose 4D
  • 19. Patient-Size AECPatient size AEC: the tube current is adjusted based on the overall size of the patient to reduce the variation in image quality between small patients and large patients. For a given patient size, the appropriate milliamperage is selected and is used for the entire examination or scan series Hi-mA Lo-mA
  • 20. Z-axis AEC: tube current is modulated according to patient attenuation along the z- axis.The goal is to reduce the variation in image quality of images from the same series. attenuation
  • 21. Patient-Size AECRotational AEC: The tube current is decreased and increased rapidly (modulated) during the course of each rotation to compensate for differences in attenuation between lateral (left-right) and A-P (anterior-posterior) projections . highattenuationlow attenuation
  • 22. Combination of AEC functions • Tube current is adjusted during scanning to compensate for attenuation differences – dose applied to patient only where needed, avoiding dose where it isn’t mA position
  • 23. On line’ modulation–uses attenuation data from previous rotation– adapts tube current to patient attenuation ‘on the fly’
  • 24.
  • 25. Scan projection radiographs (SPRs, known as scout, scanogram or topogram views) are the main way that AEC systems assess the attenuation of the patient in order to set the tube current
  • 26. To obtain correct attenuation data from SPR always centre the patient carefully Patient is positioned in the isocenter – optimal dose and image quality Patient is positioned too high - increased mAsPatient is positioned too low - reduced mAs and increased noise
  • 27. Toshiba:defining image quality requirements Specify s.d. level (or ‘image quality level)–patient mAcalculated to achieve this noise level at any scan parameter settings Set min & max mA
  • 28. Benefits of CT scanner AEC Consistent image quality Potential for dose reduction through exposure optimisation Reduced tube loading Extended scan runs(OLP) Reduction in photon starvation artifact
  • 29.
  • 30. 30 • Images along length of phantom (no AEC) • • • • • • • • Constant mA Testing the AEC
  • 31. 31 Testing the AEC • Measure noise with AEC off and on • Monitor mA, CTDIvol 0 4 8 12 16 20 24 28 -150 -100 -50 0 50 100 150 Z-position (mm) Noise(%) automA off Noise Index 12 Increased mA Decreased mA Constant mA Same mA Increased mA Decreased mA
  • 32. 32 Coronal view Sagittal view z-axis AEC off z-axis AEC on Noise increases Constant noise Testing the AEC – Viewing with MPR
  • 33.
  • 35. without angular mA modulation with angular mA modulation
  • 36. Bowtie Filter Longer bowtie path lines up with shorter patient path
  • 37. Reduce x-ray scatter (noise and artifact) Maintain uniform x-ray at detector Reduce surface dose by 50%
  • 38. • Retrospectively gated CCTA(dose modulation ) ~15 mSv100% 20% ~30% reduction
  • 39. Image obtained in middiastole (75% of R-R interval) reconstructed using 1.5-mm slice thickness at high tube current shows low-noise Image obtained in midsystole (30% of R-R interval) reconstructed using 1.5-mm slice thickness shows higher-noise image obtained at low radiation dose
  • 40. • Prospectively gated CCTA ~ 1 mSv Prospective Gating of CTA(snapshot plus) 40mm 5mm overlap
  • 41. Scan range Conventional technology without Dose Shield SOMATOM Definition AS+ with Adaptive Dose Shield Scan range Dynamic Collimation • In helical scanning extra rotations are needed at end of imaged volume – Significant extra dose: wide beam widths and short scans • Dynamic collimation - collimator blades open and close asymetrically at start and end of scan
  • 42. (a) conventional and (b) adaptive section collimation CT scanning protocols. For adaptive section collimation, shape of x-ray cone beam at beginning and end of spiral acquisition is controlled by two collimators made of absorbent material.
  • 43. Image reconstruction(FBP) attenuation detectorposition 90°2 projections4 projections8 projections16 projections attenuation detector position 0°
  • 44. 0(3) 0(9) 0(7) 0(5) 6 6 6 6 10 14 12 12 12 12 5 7 5 7 3 9 7 5 3 5 9 7 816 12 12 +1-1 -2+2 Iterative reconstruction
  • 45.
  • 47. 120 kVp and 300 mA FBP 120 kVp and 300 mA ASiR 100% 120 kVp and 300 mA ASiR 50%