Onsite Oil Analysis with ASTM
Compliant FTIR Instruments
Dr Steve Dye
Business Development Manager
Parker Kittiwake
Why use FTIR in Oil Analysis?
§ One of the most widely used laboratory tools for oil analysis
§ Fast - capable of detecting multiple oil analysis parameters simultaneously
§ e.g. water, glycol, soot, oxidation, nitration…
§ Easy to use - no extensive sample preparation or wet chemistry required
§ Inexpensive – after capital outlay, per measurement costs are minimal
§ Quick - replaces tedious and time-consuming physical and chemical methods
Infrared Radiation§ Part of the Electromagnetic Spectrum
§ Longer wavelength than Visible light – “heat” energy
§ Absorbed by molecules (not atoms)
§ Gives molecular information, not elemental
§ Absorption increases molecular vibrations and/or rotations
§ associated with the internal bonds between the atoms of the molecule
The Electromagnetic Spectrum
Visible
Light
Wavelength
(m)
Frequency
(Hz)
Wavenumber
(cm-1)
10-11 10-10 10-9 10-8 10-7 10-6 10-410-5 10-3 10-2 10010-1 101 102 103
1019 1018 1017 1016 1014 1013 1012
1011 1010 109 108 107 106 105
1Å 1nm 10nm 1μm 10μm 100μm 1mm 1cm
Hard
X-ray
Soft
X-ray
UV IR
Microwave
VHF Short Mid Long
Radiowave
Mid FarNear
Near
Far
Spectral Regions
Gamma
rays
1000 100 105000
Infrared Absorption
§ Molecules have several types of vibrations. (3n-5 for linear, 3n-6 for non-linear)
§ Associated with the internal bonds, some (not all) are IR active.
§ IR radiation absorbed at a characteristic energy (frequency / wavelength).
§ Qualitative analysis – “fingerprint” identifies which molecules are present
§ Amount of energy absorbed per unit volume depends on concentration.
§ Quantitative analysis – signal intensity reflects how much present (ppm or %)
H
O
H
H
O
HH
O
H
H
O
H
H
O
H
H
O
H
H
O
H
H H
O
H H
O
IR Radiation
Stretching
Bending,
In Plane
Bending,
Out of Plane
A
wavelength (µm)
An example; Water (H2O):
Water Molecule Vibrations
Traditional scanning IR spectroscopy
§ First developed in the 1950s
§ Step wise scanning of a dispersive element and recording of signal
§ Time consuming
§ Drift issues
§ Need a reference to obtain the “instrument function”
§ Virtually all instruments are now of the Fourier Transform Interferometer design
FTIR Spectroscopy
§ Development started in the late 1960s by analytical chemists borrowing from early work of
physicists.
§ Michelson Morley Interferometer was the classic example
§ First commercial FT instrument released in 1969.
§ Uses an interferometer to generate an interference pattern of infrared radiation.
§ Divide infrared radiation into two paths.
§ Vary the path length along one of the arms
§ Recombine the radiation.
§ Record signal intensity as a function of change in pathlength difference (speed, time)
§ Interference pattern (interferogram) is in the time domain.
§ We need frequency (wavenumber, Hz).
§ Conversion is by Fourier Transform.
§ Extremely quick these days with computer FFT1 algorithms
FTIR Spectrometer Principle
Infrared Light Sample
Moving
Mirror
Fixed Mirror
Detector
Interferogram
Beamsplitter
Moving mirror speed ->
Signal
-
>
0
0
Principle of FTIR Spectroscopy
Sample Interferogram
Background Interferogram
Ratio
A = -logT
Data
Fast Fourier
Transform
Algorithm
Transmittance Spectrum
Transmittance Spectrum Absorbance Spectrum
Single Beam Spectra
Advantages of FTIR Spectroscopy
§ All wavelengths scanned simultaneously (Fellget’s Advantage)
§ Reduction in data collection time by a factor of N
§ Increase in signal-to-noise ratio (SNR) by a factor of N 1/2 by multiple scanning
§ Greater amount of source energy to detector (Jacquinot’s Advantage)
§ No rectangular slit aperture
§ Much greater throughput compared to conventional IR
§ Fixed internal wavelength reference source
§ Typically a small HeNe laser
§ Provides high wavelength accuracy
Transmission (Sample) Cell
§ Fixed pathlength: Typically between 50-200μm
§ Note: ASTM standard mandates 100µm cell pathlength
§ Other cell pathlengths allow for different concentrations
§ Short pathlengths for strong absorbers / high concentration
§ Longer pathlengths for weak absorbers / low concentration
§ Cell window material choices:
§ Water insoluble ZnSe or CaF2
§ Water soluble KBr, NaCl or KCl
Sample: Fixed Pathlength
IR Transparent Cell Windows
IR Radiation To Detector
FTIR Spectrum of an Oil Sample
§ Load cell with sample and collect sample spectrum
§ The FTIR oil sample spectrum will reflect the molecular composition of the oil
§ For example, an ester based oil:
Wavenumber (cm-1)
Absorbance
3500 3000 2500 2000 1500 1000
e.g. Ester Based Oil
C
-
H Stretch
C=O Stretch
C
-
O Stretch
CH
2
, CH
3
Bending
CH
2
Rocking
IR
How can I use FTIR for oil analysis?
§ Most commonly used as a screening tool
§ FTIR is a rapid method to identify samples that have problems
§ Provides general information about the identity and condition of the oil
§ Degradation by-products, additive depletion, contaminants
§ FTIR supplements other methods to help diagnose any problems
§ For example, elemental analysis methods, viscosity measurements, chemical analyses.
Collecting an FTIR Spectrum
§ Specify FTIR instrument parameters:
§ Resolution (usually 2, 4 or 8 cm-1)
§ Lower wavenumber means higher resolution
§ Number of scans
§ Increasing the scans improves the signal-to-noise ratio by a factor of N1/2
§ Apodization
§ Corrects for interferogram side lobes and affects spectral band shapes
§ Collect background single beam spectrum
§ Air or empty cell reference
Sampling Oil for FTIR Analysis
§ Oil should be filtered to remove large particles.
§ Shake well to give a representative sample.
§ Most common sample cell is a transmission cell
– Manual loading or continuous flow pumping system
Example of Used Oil FTIR Spectrum
0.1
0.2
0.3
0.4
0.5
0.6
0.7
Absorbance
100015002000250030003500
Wavenumber (cm-1)
Water
Soot
Oxidation
Nitration
Antiwear
Sulfate
Fuel
Qualitative or Quantitative?
FTIR measurements can be either:
§ Qualitative
§ Identify constituents
§ Observe changes in peak intensities or absorbance – better/worse assessment
OR
§ Quantitative
§ Obtain a definite result in units of concentration (ex: ppm, g/L, %w/v)
§ Calibration using standards of known concentrations of the constituent or
parameter of interest
§ Correlate concentrations with absorbance to obtain regression coefficients
§ Prediction of concentrations in unknown sample using these coefficients
Quantitative Analysis - Example
§ Concentration of water in crankcase lubricants:
0.10 0.20 0.30 0.40
0.0
0.1
0.2
0.3
0.4
0.5
Water (%w/w)
Absorbance
.1
0
.2
.3
3800 3600 3400 3200
0.5%
0.4%
0.3%
0.2%
0.1%
0.05%
0.0%
Wavenumber (cm-1)
Absorbance
Considerations of Quantitative Analysis
§ Some methods are considered semi-quantitative:
§ Difficult to obtain calibration samples with reliable concentration values
§ Reference method only gives approximate values
§ Sample to sample variability
§ Actual samples can differ from calibration samples
§ For example: For soot measurements, particle size may vary
§ For example: When measuring fuel, fuel composition may vary between locations
§ Interferences
§ Other oil sample constituents and parameters can interfere with the measurement
Interferences
§ Parameters with the same functional group will absorb at the same frequency,
interfering with one another
§ e.g. Water and Glycol
Water
Ethylene Glycol
Absorbance
Wavenumber (cm-1)
O-H O-H
Interferences from the Base Oil
§ The base oil or new oil additives may also interfere with the measurement of some
parameters, especially in the “fingerprint” region.
0.05
0.15
0.25
0.35
0.45
0.55
0.65
0.75
Absorbance
60080010001200140016001800
Wavenumber (cm-1)
Oxidation
Antiwear
Base Oil
Hydraulic Oil
Fuel
Sulfate
Fingerprint Region
Oil Additives
Spectral Subtraction
§ Spectral subtraction is used
to subtract out the
interferences from the base
oil and additives
§ The contaminants,
degradation by-products and
additive depletion can be
visualized and analyzed more
easily
Used Oil Spectrum
0.5
1.0
Absorbance
New Reference Oil Spectrum
0.5
1.0
Difference Spectrum
0.5
1.0
100015002000250030003500
Wavenumber (cm-1)
Water
Soot Oxidation
Sulfation
Antiwear
Considerations of Spectral Subtraction
§ There are several considerations to take into account when using the spectral subtraction
method:
§ New (reference) oil may not always be available
§ Batch to batch variation in both base oil and formulation for same oil brand
§ Need correct reference
§ Oil changes lead to dilution and/or carryover
§ Reference no longer good
§ Topping up with another batch or different brand
§ Again, reference is no longer good
Contamination/Blending with Different Oils
§ FTIR is also very useful for determining significant changes in oil chemistry
§ e.g. PAO mixed with an ester- or polyalkylglycol- (PAG) based oi
0.0
0.4
0.8
1.2
1.6
2.0
2.4
Absorbance
100015002000250030003500
Wavenumber (cm-1)
Ester
PAG
Parameter Frequency (cm-1) Traditional Method
Oxidation 1710 Acid number (AN) titration
Nitration 1630 None
Sulfation 1150 Base Number (BN) titration
Diesel Fuel 810 Flash Point, Gas Chromatography
Gasoline 750 Flash Point, GC
Water 3420 Karl Fisher
Glycol 1080,1040, 880 Colorimetry, GC
Soot 2000 Thermogravimetric
Antiwear 980 Elemental Analysis (Zn etc.)
(T)BN 1516,1152 BN by titration
Common Oil Parameters by FTIR
ASTM Standards for oil analysis
§ American Society for Testing and Materials (now ASTM International) defines standard
methods and practices for over 12,000 measurement methods for metals, petroleum,
construction, environmental and many more.
§ Seen by Industry as THE STANDARD that all measurements should be made in accordance
with.
§ ASTM have several published standards and work is ongoing to define more:
§ 1 Standard practice (E2412) which defines a total of 12 trend parameters for 3 types of
lubricating oils; Petroleum based, Extreme Pressure (EP) fluids and Synthetic Polyol
Esters.
§ 5 Standard methods – Oxidation (D7414), Nitration (7624), Phosphate Antiwear
(D7412), Sulphation (D7415), Oxidation in transmission oils (D7214)
§ Proposed methods (WIP) – Acid Number, Base Number, Water Content
ASTM Lubricant Practices & Methods
§ D7418-07 Standard Practice for Set-Up and Operation of Fourier Transform Infrared (FTIR)
Spectrometers for In-Service Oil Condition Monitoring.
§ E2412-10 Standard Practice for Condition Monitoring of in-service Lubricants by Trend Analysis
Using Fourier Transform Infrared (FT-IR) Spectrometry.
§ D7214-07a Standard Test Method for Determination of the Oxidation of Used Lubricants by FT-IR
Using Peak Area Increase Calculation.
§ D7414-09 Standard Test Method for Condition Monitoring of Oxidation in In-Service Petroleum
and Hydrocarbon Based Lubricants by Trend Analysis Using Fourier Transform Infrared (FT-IR)
Spectrometry.
§ D7412-09 Standard Test Method for Condition Monitoring of Phosphate Antiwear Additives in In-
Service Petroleum and Hydrocarbon Based Lubricants by Trend Analysis Using Fourier Transform
Infrared (FT-IR) Spectrometry.
§ D7415-09 Standard Test Method for Condition Monitoring of Sulfate By-Products in In-Service
Petroleum and Hydrocarbon Based Lubricants by Trend Analysis Using Fourier Transform Infrared
(FT-IR) Spectrometry.
§ D7624-10 Standard Test Method for Condition Monitoring of Nitration in In-Service Petroleum
and Hydrocarbon Based Lubricants by Trend Analysis Using Fourier Transform Infrared (FT-IR)
Spectrometry.
Field vs. Lab testing
§ Lab testing:
§ Lab testing can offer increased number of alternative tests on oil samples
(e.g XRF)
§ Expert advice may be offered on analysis
§ Time taken to ship samples from remote locations can be long
§ Ongoing cost per sample sent
§ Field Testing:
§ Immediate results available so action can be taken fast, before major
problems arise
§ No/Low ongoing costs after initial capital outlay
§ Some user intelligence required to set up parameters for alarm warning
levels etc
§ Subset of full lab tests available
Non ASTM compliant field instruments
Lab instruments unsuitable for the field
Parker Kittiwake FTIR3 Oil Analyser
ASTM Compliant, field deployable, all in a small, portable instrument
§ Simple to use
§ Software pre-loaded onto Netbook
§ Multiple equipment records
§ Trending of multiple parameters
§ Multiple users
Multiple equipment types
Results Screen – Spectral Mode
Historical data records on equipment
Time trending of parameters
What will the FTIR3 Measure
§ Sulphation – TO ASTM D7415-09
§ Oxidation – To ASTM D7414-09
§ Nitration – To ASTM D7624-10
§ Phosphate Antiwear – To ASTM D7412-09
§ E2412:
• Soot Water
• Diesel
• Fuel Contamination
• Ethyl Glycol Coolant
• Antioxidant depletion
§ Future published ASTM methods can be readily added to the instrument.
§ Customer specific methods can be developed and also added to the instrument.
Advantages and Limitations
§ Advantages of FTIR spectroscopy in oil analysis:
§ Rapid analysis with no extensive sample preparation
§ Can give information on multiple parameters simultaneously
§ Provides quantitative and qualitative data
§ Accurate and repeatable results
§ Limitations of FTIR spectroscopy in oil analysis:
§ Different base oils and additives can interfere with the measurement of
different parameters
§ For spectral subtraction, sometimes it can be difficult to find a suitable
reference oil
§ Difficulty measuring some components that are less than 0.1% w/w
§ No elemental analysis
In Summary….
§ FTIR provides a quick and accurate test method to measure multiple parameters
on in service oils
§ Advances in recent years have allowed reduced size, field deployable devices to
be developed
§ ASTM ratification of test Methods is an ongoing process, providing end users with
confidence in reliable and repeatable measurement techniques and results
§ More ASTM agreed methods are on the way for additional parameters
§ Field deployable, ASTM compliant devices are now available
Any questions?

More Related Content

What's hot

Comparison of sfc, gc and hplc
Comparison of sfc, gc and hplcComparison of sfc, gc and hplc
Comparison of sfc, gc and hplcVrushali Tambe
 
Combined Atomic Absorptoin Spectroscopy
Combined Atomic Absorptoin SpectroscopyCombined Atomic Absorptoin Spectroscopy
Combined Atomic Absorptoin Spectroscopyknowledge1995
 
gas chromatography
gas chromatographygas chromatography
gas chromatographyaqsa fatima
 
Flash chromatography
Flash chromatographyFlash chromatography
Flash chromatographyMahendra G S
 
Gas chromatography lecture
Gas chromatography lectureGas chromatography lecture
Gas chromatography lectureMehwish Nawaz
 
HIGH PERFORMANCE LIQUID CHROMATOGRAPHY.pptx
HIGH PERFORMANCE LIQUID CHROMATOGRAPHY.pptxHIGH PERFORMANCE LIQUID CHROMATOGRAPHY.pptx
HIGH PERFORMANCE LIQUID CHROMATOGRAPHY.pptxPallaviKumari112
 
Gas chromatography, an introduction.pdf
Gas chromatography, an introduction.pdfGas chromatography, an introduction.pdf
Gas chromatography, an introduction.pdfSherif Taha
 
Spectrofluorimetry (www.redicals.com)
Spectrofluorimetry (www.redicals.com)Spectrofluorimetry (www.redicals.com)
Spectrofluorimetry (www.redicals.com)Goa App
 
Atomic emission spectroscopy
Atomic emission spectroscopyAtomic emission spectroscopy
Atomic emission spectroscopyMahendra G S
 
FT-IR spectroscopy Instrumentation and Application, By- Anubhav singh, M.pharm
FT-IR spectroscopy Instrumentation and Application, By- Anubhav singh, M.pharmFT-IR spectroscopy Instrumentation and Application, By- Anubhav singh, M.pharm
FT-IR spectroscopy Instrumentation and Application, By- Anubhav singh, M.pharmAnubhav Singh
 
Hyphenated techniques- GCMS, LCMS, GCIR
Hyphenated techniques- GCMS, LCMS, GCIRHyphenated techniques- GCMS, LCMS, GCIR
Hyphenated techniques- GCMS, LCMS, GCIRSanket Shinde
 
Fluorescence Spectroscopy
Fluorescence SpectroscopyFluorescence Spectroscopy
Fluorescence SpectroscopyNizam Ashraf
 
Sample preparation for Chromatography
Sample preparation for Chromatography Sample preparation for Chromatography
Sample preparation for Chromatography latha13061996
 
Partition chromatography & partition paper chromatography
Partition chromatography & partition paper chromatographyPartition chromatography & partition paper chromatography
Partition chromatography & partition paper chromatographyArtina Aquitania
 

What's hot (20)

Comparison of sfc, gc and hplc
Comparison of sfc, gc and hplcComparison of sfc, gc and hplc
Comparison of sfc, gc and hplc
 
Infra red spectroscopy.
Infra red spectroscopy.Infra red spectroscopy.
Infra red spectroscopy.
 
SUPER CRITICAL FLUID CHROMATOGRAPHY
SUPER CRITICAL FLUID CHROMATOGRAPHY SUPER CRITICAL FLUID CHROMATOGRAPHY
SUPER CRITICAL FLUID CHROMATOGRAPHY
 
Combined Atomic Absorptoin Spectroscopy
Combined Atomic Absorptoin SpectroscopyCombined Atomic Absorptoin Spectroscopy
Combined Atomic Absorptoin Spectroscopy
 
FTIR
FTIRFTIR
FTIR
 
gas chromatography
gas chromatographygas chromatography
gas chromatography
 
Flash chromatography
Flash chromatographyFlash chromatography
Flash chromatography
 
Gas chromatography lecture
Gas chromatography lectureGas chromatography lecture
Gas chromatography lecture
 
HIGH PERFORMANCE LIQUID CHROMATOGRAPHY.pptx
HIGH PERFORMANCE LIQUID CHROMATOGRAPHY.pptxHIGH PERFORMANCE LIQUID CHROMATOGRAPHY.pptx
HIGH PERFORMANCE LIQUID CHROMATOGRAPHY.pptx
 
Errors
Errors Errors
Errors
 
Gas chromatography, an introduction.pdf
Gas chromatography, an introduction.pdfGas chromatography, an introduction.pdf
Gas chromatography, an introduction.pdf
 
Spectrofluorimetry (www.redicals.com)
Spectrofluorimetry (www.redicals.com)Spectrofluorimetry (www.redicals.com)
Spectrofluorimetry (www.redicals.com)
 
Atomic emission spectroscopy
Atomic emission spectroscopyAtomic emission spectroscopy
Atomic emission spectroscopy
 
FT-IR spectroscopy Instrumentation and Application, By- Anubhav singh, M.pharm
FT-IR spectroscopy Instrumentation and Application, By- Anubhav singh, M.pharmFT-IR spectroscopy Instrumentation and Application, By- Anubhav singh, M.pharm
FT-IR spectroscopy Instrumentation and Application, By- Anubhav singh, M.pharm
 
Hyphenated techniques- GCMS, LCMS, GCIR
Hyphenated techniques- GCMS, LCMS, GCIRHyphenated techniques- GCMS, LCMS, GCIR
Hyphenated techniques- GCMS, LCMS, GCIR
 
Atomic emission spectroscopy
Atomic emission spectroscopyAtomic emission spectroscopy
Atomic emission spectroscopy
 
Fluorescence Spectroscopy
Fluorescence SpectroscopyFluorescence Spectroscopy
Fluorescence Spectroscopy
 
Sample preparation for Chromatography
Sample preparation for Chromatography Sample preparation for Chromatography
Sample preparation for Chromatography
 
Partition chromatography & partition paper chromatography
Partition chromatography & partition paper chromatographyPartition chromatography & partition paper chromatography
Partition chromatography & partition paper chromatography
 
Recent advances in hplc and gc
Recent advances in hplc and gcRecent advances in hplc and gc
Recent advances in hplc and gc
 

Viewers also liked

Six reasons why oil analysis programs fail
Six reasons why oil analysis programs failSix reasons why oil analysis programs fail
Six reasons why oil analysis programs failTESTOIL
 
Pharmaceutical Solid Form
Pharmaceutical Solid FormPharmaceutical Solid Form
Pharmaceutical Solid FormSimon Curtis
 
Elf Training
Elf TrainingElf Training
Elf Trainingabdrazy
 
Le price training presentation
Le price training presentationLe price training presentation
Le price training presentationzainudinyahya
 
Lubrication fundamentals
Lubrication fundamentalsLubrication fundamentals
Lubrication fundamentalsFahad Aldhawi
 
Metals Tensile Testing Standards - ISO 6892-1 ASTM E8/8M - Strain Control
Metals Tensile Testing Standards - ISO 6892-1 ASTM E8/8M - Strain ControlMetals Tensile Testing Standards - ISO 6892-1 ASTM E8/8M - Strain Control
Metals Tensile Testing Standards - ISO 6892-1 ASTM E8/8M - Strain ControlDavid Fry
 
Principles of lubrication new
Principles of lubrication newPrinciples of lubrication new
Principles of lubrication newMamdouh Alhanafy
 
Lubrication Reliability by Lubretec : a 10 step approach to World Class Maint...
Lubrication Reliability by Lubretec : a 10 step approach to World Class Maint...Lubrication Reliability by Lubretec : a 10 step approach to World Class Maint...
Lubrication Reliability by Lubretec : a 10 step approach to World Class Maint...Toon Van Grunderbeeck
 
Lubrication Fundamentals: Lubricating Oil Basics
Lubrication Fundamentals: Lubricating Oil BasicsLubrication Fundamentals: Lubricating Oil Basics
Lubrication Fundamentals: Lubricating Oil BasicsTESTOIL
 
Fourier Transform Infrared Spectroscopy Ftir
Fourier Transform Infrared Spectroscopy FtirFourier Transform Infrared Spectroscopy Ftir
Fourier Transform Infrared Spectroscopy FtirGamal Abdel Hamid
 
Regression analysis ppt
Regression analysis pptRegression analysis ppt
Regression analysis pptElkana Rorio
 
Liquid Crystal and Liquid Crystal Polymer
Liquid Crystal and Liquid Crystal PolymerLiquid Crystal and Liquid Crystal Polymer
Liquid Crystal and Liquid Crystal PolymerSaurav Ch. Sarma
 

Viewers also liked (17)

Six reasons why oil analysis programs fail
Six reasons why oil analysis programs failSix reasons why oil analysis programs fail
Six reasons why oil analysis programs fail
 
Dispersive & FTIR
Dispersive & FTIRDispersive & FTIR
Dispersive & FTIR
 
Principles of ftir
Principles of ftirPrinciples of ftir
Principles of ftir
 
FTIR
FTIRFTIR
FTIR
 
Pharmaceutical Solid Form
Pharmaceutical Solid FormPharmaceutical Solid Form
Pharmaceutical Solid Form
 
Elf Training
Elf TrainingElf Training
Elf Training
 
Le price training presentation
Le price training presentationLe price training presentation
Le price training presentation
 
Lubrication fundamentals
Lubrication fundamentalsLubrication fundamentals
Lubrication fundamentals
 
Metals Tensile Testing Standards - ISO 6892-1 ASTM E8/8M - Strain Control
Metals Tensile Testing Standards - ISO 6892-1 ASTM E8/8M - Strain ControlMetals Tensile Testing Standards - ISO 6892-1 ASTM E8/8M - Strain Control
Metals Tensile Testing Standards - ISO 6892-1 ASTM E8/8M - Strain Control
 
Machinery Oil Analysis
Machinery Oil AnalysisMachinery Oil Analysis
Machinery Oil Analysis
 
Principles of lubrication new
Principles of lubrication newPrinciples of lubrication new
Principles of lubrication new
 
FTIR
FTIRFTIR
FTIR
 
Lubrication Reliability by Lubretec : a 10 step approach to World Class Maint...
Lubrication Reliability by Lubretec : a 10 step approach to World Class Maint...Lubrication Reliability by Lubretec : a 10 step approach to World Class Maint...
Lubrication Reliability by Lubretec : a 10 step approach to World Class Maint...
 
Lubrication Fundamentals: Lubricating Oil Basics
Lubrication Fundamentals: Lubricating Oil BasicsLubrication Fundamentals: Lubricating Oil Basics
Lubrication Fundamentals: Lubricating Oil Basics
 
Fourier Transform Infrared Spectroscopy Ftir
Fourier Transform Infrared Spectroscopy FtirFourier Transform Infrared Spectroscopy Ftir
Fourier Transform Infrared Spectroscopy Ftir
 
Regression analysis ppt
Regression analysis pptRegression analysis ppt
Regression analysis ppt
 
Liquid Crystal and Liquid Crystal Polymer
Liquid Crystal and Liquid Crystal PolymerLiquid Crystal and Liquid Crystal Polymer
Liquid Crystal and Liquid Crystal Polymer
 

Similar to Steve dye onsite oil analysis with astm compliant ftir intruments

Quality Measurements Using NIR/MIR Spectroscopy: A Rotten Apple Could Turn Yo...
Quality Measurements Using NIR/MIR Spectroscopy: A Rotten Apple Could Turn Yo...Quality Measurements Using NIR/MIR Spectroscopy: A Rotten Apple Could Turn Yo...
Quality Measurements Using NIR/MIR Spectroscopy: A Rotten Apple Could Turn Yo...TechRentals
 
2013 University of Virginia's W.M. Keck FRET Workshop: Basic Filter Considera...
2013 University of Virginia's W.M. Keck FRET Workshop: Basic Filter Considera...2013 University of Virginia's W.M. Keck FRET Workshop: Basic Filter Considera...
2013 University of Virginia's W.M. Keck FRET Workshop: Basic Filter Considera...Jeff Carmichael
 
Basic Concepts of Clinical Flowcytometry
Basic Concepts of Clinical FlowcytometryBasic Concepts of Clinical Flowcytometry
Basic Concepts of Clinical FlowcytometryPravin Amabade
 
Parker Kittiwake FTIR Infra-red Oil Analyser Brochure
Parker Kittiwake FTIR Infra-red Oil Analyser BrochureParker Kittiwake FTIR Infra-red Oil Analyser Brochure
Parker Kittiwake FTIR Infra-red Oil Analyser BrochureParker Kittiwake
 
Icpms basics and instrumentation
Icpms basics and instrumentationIcpms basics and instrumentation
Icpms basics and instrumentationAMOL SHINDE
 
UV-VIS SPECTRO final.ppt
UV-VIS SPECTRO final.pptUV-VIS SPECTRO final.ppt
UV-VIS SPECTRO final.pptJgdishrathi
 
High vacuum pump and sensors.ppt
High vacuum pump and sensors.pptHigh vacuum pump and sensors.ppt
High vacuum pump and sensors.pptFasial Ghazanfar
 
lab-oil-analysis
lab-oil-analysis lab-oil-analysis
lab-oil-analysis elantablee
 
Ieeepes boston xfmrdiagnosticssweetser
Ieeepes boston xfmrdiagnosticssweetserIeeepes boston xfmrdiagnosticssweetser
Ieeepes boston xfmrdiagnosticssweetserSelva Kumar
 
PPT - Electronics Industry Wastewater
PPT - Electronics Industry WastewaterPPT - Electronics Industry Wastewater
PPT - Electronics Industry WastewaterNitin Yadav
 
Wavelength Accuracy.pptx
Wavelength Accuracy.pptxWavelength Accuracy.pptx
Wavelength Accuracy.pptxsumitdekate1
 
Membrane bio reactor- Waste water treatment
Membrane bio reactor- Waste water treatmentMembrane bio reactor- Waste water treatment
Membrane bio reactor- Waste water treatmentakalankar
 
XRF Based Multi-Metals Continuous Water Analyzer
XRF Based Multi-Metals Continuous Water AnalyzerXRF Based Multi-Metals Continuous Water Analyzer
XRF Based Multi-Metals Continuous Water AnalyzerEuropean Tech Serv
 

Similar to Steve dye onsite oil analysis with astm compliant ftir intruments (20)

Quality Measurements Using NIR/MIR Spectroscopy: A Rotten Apple Could Turn Yo...
Quality Measurements Using NIR/MIR Spectroscopy: A Rotten Apple Could Turn Yo...Quality Measurements Using NIR/MIR Spectroscopy: A Rotten Apple Could Turn Yo...
Quality Measurements Using NIR/MIR Spectroscopy: A Rotten Apple Could Turn Yo...
 
Determination of Common Counterions and Impurity Anions in Pharmaceuticals Us...
Determination of Common Counterions and Impurity Anions in Pharmaceuticals Us...Determination of Common Counterions and Impurity Anions in Pharmaceuticals Us...
Determination of Common Counterions and Impurity Anions in Pharmaceuticals Us...
 
Oil in water analyzer
Oil in water analyzerOil in water analyzer
Oil in water analyzer
 
Spectrolytic Corporate
Spectrolytic CorporateSpectrolytic Corporate
Spectrolytic Corporate
 
2013 University of Virginia's W.M. Keck FRET Workshop: Basic Filter Considera...
2013 University of Virginia's W.M. Keck FRET Workshop: Basic Filter Considera...2013 University of Virginia's W.M. Keck FRET Workshop: Basic Filter Considera...
2013 University of Virginia's W.M. Keck FRET Workshop: Basic Filter Considera...
 
Basic Concepts of Clinical Flowcytometry
Basic Concepts of Clinical FlowcytometryBasic Concepts of Clinical Flowcytometry
Basic Concepts of Clinical Flowcytometry
 
Parker Kittiwake FTIR Infra-red Oil Analyser Brochure
Parker Kittiwake FTIR Infra-red Oil Analyser BrochureParker Kittiwake FTIR Infra-red Oil Analyser Brochure
Parker Kittiwake FTIR Infra-red Oil Analyser Brochure
 
Icpms basics and instrumentation
Icpms basics and instrumentationIcpms basics and instrumentation
Icpms basics and instrumentation
 
UV-VIS SPECTRO final.ppt
UV-VIS SPECTRO final.pptUV-VIS SPECTRO final.ppt
UV-VIS SPECTRO final.ppt
 
Expanding Your High Performance Liquid Chromatography and Ultra High Performa...
Expanding Your High Performance Liquid Chromatography and Ultra High Performa...Expanding Your High Performance Liquid Chromatography and Ultra High Performa...
Expanding Your High Performance Liquid Chromatography and Ultra High Performa...
 
Expanding Your High Performance Liquid Chromatography and Ultra High Performa...
Expanding Your High Performance Liquid Chromatography and Ultra High Performa...Expanding Your High Performance Liquid Chromatography and Ultra High Performa...
Expanding Your High Performance Liquid Chromatography and Ultra High Performa...
 
HPLC
HPLCHPLC
HPLC
 
High vacuum pump and sensors.ppt
High vacuum pump and sensors.pptHigh vacuum pump and sensors.ppt
High vacuum pump and sensors.ppt
 
lab-oil-analysis
lab-oil-analysis lab-oil-analysis
lab-oil-analysis
 
Ieeepes boston xfmrdiagnosticssweetser
Ieeepes boston xfmrdiagnosticssweetserIeeepes boston xfmrdiagnosticssweetser
Ieeepes boston xfmrdiagnosticssweetser
 
PPT - Electronics Industry Wastewater
PPT - Electronics Industry WastewaterPPT - Electronics Industry Wastewater
PPT - Electronics Industry Wastewater
 
Wavelength Accuracy.pptx
Wavelength Accuracy.pptxWavelength Accuracy.pptx
Wavelength Accuracy.pptx
 
Validation of IR instrument
Validation of IR instrumentValidation of IR instrument
Validation of IR instrument
 
Membrane bio reactor- Waste water treatment
Membrane bio reactor- Waste water treatmentMembrane bio reactor- Waste water treatment
Membrane bio reactor- Waste water treatment
 
XRF Based Multi-Metals Continuous Water Analyzer
XRF Based Multi-Metals Continuous Water AnalyzerXRF Based Multi-Metals Continuous Water Analyzer
XRF Based Multi-Metals Continuous Water Analyzer
 

Recently uploaded

Unblocking The Main Thread Solving ANRs and Frozen Frames
Unblocking The Main Thread Solving ANRs and Frozen FramesUnblocking The Main Thread Solving ANRs and Frozen Frames
Unblocking The Main Thread Solving ANRs and Frozen FramesSinan KOZAK
 
08448380779 Call Girls In Diplomatic Enclave Women Seeking Men
08448380779 Call Girls In Diplomatic Enclave Women Seeking Men08448380779 Call Girls In Diplomatic Enclave Women Seeking Men
08448380779 Call Girls In Diplomatic Enclave Women Seeking MenDelhi Call girls
 
A Domino Admins Adventures (Engage 2024)
A Domino Admins Adventures (Engage 2024)A Domino Admins Adventures (Engage 2024)
A Domino Admins Adventures (Engage 2024)Gabriella Davis
 
Injustice - Developers Among Us (SciFiDevCon 2024)
Injustice - Developers Among Us (SciFiDevCon 2024)Injustice - Developers Among Us (SciFiDevCon 2024)
Injustice - Developers Among Us (SciFiDevCon 2024)Allon Mureinik
 
Enhancing Worker Digital Experience: A Hands-on Workshop for Partners
Enhancing Worker Digital Experience: A Hands-on Workshop for PartnersEnhancing Worker Digital Experience: A Hands-on Workshop for Partners
Enhancing Worker Digital Experience: A Hands-on Workshop for PartnersThousandEyes
 
Data Cloud, More than a CDP by Matt Robison
Data Cloud, More than a CDP by Matt RobisonData Cloud, More than a CDP by Matt Robison
Data Cloud, More than a CDP by Matt RobisonAnna Loughnan Colquhoun
 
Automating Business Process via MuleSoft Composer | Bangalore MuleSoft Meetup...
Automating Business Process via MuleSoft Composer | Bangalore MuleSoft Meetup...Automating Business Process via MuleSoft Composer | Bangalore MuleSoft Meetup...
Automating Business Process via MuleSoft Composer | Bangalore MuleSoft Meetup...shyamraj55
 
CNv6 Instructor Chapter 6 Quality of Service
CNv6 Instructor Chapter 6 Quality of ServiceCNv6 Instructor Chapter 6 Quality of Service
CNv6 Instructor Chapter 6 Quality of Servicegiselly40
 
Swan(sea) Song – personal research during my six years at Swansea ... and bey...
Swan(sea) Song – personal research during my six years at Swansea ... and bey...Swan(sea) Song – personal research during my six years at Swansea ... and bey...
Swan(sea) Song – personal research during my six years at Swansea ... and bey...Alan Dix
 
08448380779 Call Girls In Friends Colony Women Seeking Men
08448380779 Call Girls In Friends Colony Women Seeking Men08448380779 Call Girls In Friends Colony Women Seeking Men
08448380779 Call Girls In Friends Colony Women Seeking MenDelhi Call girls
 
Boost PC performance: How more available memory can improve productivity
Boost PC performance: How more available memory can improve productivityBoost PC performance: How more available memory can improve productivity
Boost PC performance: How more available memory can improve productivityPrincipled Technologies
 
Transcript: #StandardsGoals for 2024: What’s new for BISAC - Tech Forum 2024
Transcript: #StandardsGoals for 2024: What’s new for BISAC - Tech Forum 2024Transcript: #StandardsGoals for 2024: What’s new for BISAC - Tech Forum 2024
Transcript: #StandardsGoals for 2024: What’s new for BISAC - Tech Forum 2024BookNet Canada
 
🐬 The future of MySQL is Postgres 🐘
🐬  The future of MySQL is Postgres   🐘🐬  The future of MySQL is Postgres   🐘
🐬 The future of MySQL is Postgres 🐘RTylerCroy
 
Scaling API-first – The story of a global engineering organization
Scaling API-first – The story of a global engineering organizationScaling API-first – The story of a global engineering organization
Scaling API-first – The story of a global engineering organizationRadu Cotescu
 
Finology Group – Insurtech Innovation Award 2024
Finology Group – Insurtech Innovation Award 2024Finology Group – Insurtech Innovation Award 2024
Finology Group – Insurtech Innovation Award 2024The Digital Insurer
 
Google AI Hackathon: LLM based Evaluator for RAG
Google AI Hackathon: LLM based Evaluator for RAGGoogle AI Hackathon: LLM based Evaluator for RAG
Google AI Hackathon: LLM based Evaluator for RAGSujit Pal
 
FULL ENJOY 🔝 8264348440 🔝 Call Girls in Diplomatic Enclave | Delhi
FULL ENJOY 🔝 8264348440 🔝 Call Girls in Diplomatic Enclave | DelhiFULL ENJOY 🔝 8264348440 🔝 Call Girls in Diplomatic Enclave | Delhi
FULL ENJOY 🔝 8264348440 🔝 Call Girls in Diplomatic Enclave | Delhisoniya singh
 
Breaking the Kubernetes Kill Chain: Host Path Mount
Breaking the Kubernetes Kill Chain: Host Path MountBreaking the Kubernetes Kill Chain: Host Path Mount
Breaking the Kubernetes Kill Chain: Host Path MountPuma Security, LLC
 
Strategies for Unlocking Knowledge Management in Microsoft 365 in the Copilot...
Strategies for Unlocking Knowledge Management in Microsoft 365 in the Copilot...Strategies for Unlocking Knowledge Management in Microsoft 365 in the Copilot...
Strategies for Unlocking Knowledge Management in Microsoft 365 in the Copilot...Drew Madelung
 
Presentation on how to chat with PDF using ChatGPT code interpreter
Presentation on how to chat with PDF using ChatGPT code interpreterPresentation on how to chat with PDF using ChatGPT code interpreter
Presentation on how to chat with PDF using ChatGPT code interpreternaman860154
 

Recently uploaded (20)

Unblocking The Main Thread Solving ANRs and Frozen Frames
Unblocking The Main Thread Solving ANRs and Frozen FramesUnblocking The Main Thread Solving ANRs and Frozen Frames
Unblocking The Main Thread Solving ANRs and Frozen Frames
 
08448380779 Call Girls In Diplomatic Enclave Women Seeking Men
08448380779 Call Girls In Diplomatic Enclave Women Seeking Men08448380779 Call Girls In Diplomatic Enclave Women Seeking Men
08448380779 Call Girls In Diplomatic Enclave Women Seeking Men
 
A Domino Admins Adventures (Engage 2024)
A Domino Admins Adventures (Engage 2024)A Domino Admins Adventures (Engage 2024)
A Domino Admins Adventures (Engage 2024)
 
Injustice - Developers Among Us (SciFiDevCon 2024)
Injustice - Developers Among Us (SciFiDevCon 2024)Injustice - Developers Among Us (SciFiDevCon 2024)
Injustice - Developers Among Us (SciFiDevCon 2024)
 
Enhancing Worker Digital Experience: A Hands-on Workshop for Partners
Enhancing Worker Digital Experience: A Hands-on Workshop for PartnersEnhancing Worker Digital Experience: A Hands-on Workshop for Partners
Enhancing Worker Digital Experience: A Hands-on Workshop for Partners
 
Data Cloud, More than a CDP by Matt Robison
Data Cloud, More than a CDP by Matt RobisonData Cloud, More than a CDP by Matt Robison
Data Cloud, More than a CDP by Matt Robison
 
Automating Business Process via MuleSoft Composer | Bangalore MuleSoft Meetup...
Automating Business Process via MuleSoft Composer | Bangalore MuleSoft Meetup...Automating Business Process via MuleSoft Composer | Bangalore MuleSoft Meetup...
Automating Business Process via MuleSoft Composer | Bangalore MuleSoft Meetup...
 
CNv6 Instructor Chapter 6 Quality of Service
CNv6 Instructor Chapter 6 Quality of ServiceCNv6 Instructor Chapter 6 Quality of Service
CNv6 Instructor Chapter 6 Quality of Service
 
Swan(sea) Song – personal research during my six years at Swansea ... and bey...
Swan(sea) Song – personal research during my six years at Swansea ... and bey...Swan(sea) Song – personal research during my six years at Swansea ... and bey...
Swan(sea) Song – personal research during my six years at Swansea ... and bey...
 
08448380779 Call Girls In Friends Colony Women Seeking Men
08448380779 Call Girls In Friends Colony Women Seeking Men08448380779 Call Girls In Friends Colony Women Seeking Men
08448380779 Call Girls In Friends Colony Women Seeking Men
 
Boost PC performance: How more available memory can improve productivity
Boost PC performance: How more available memory can improve productivityBoost PC performance: How more available memory can improve productivity
Boost PC performance: How more available memory can improve productivity
 
Transcript: #StandardsGoals for 2024: What’s new for BISAC - Tech Forum 2024
Transcript: #StandardsGoals for 2024: What’s new for BISAC - Tech Forum 2024Transcript: #StandardsGoals for 2024: What’s new for BISAC - Tech Forum 2024
Transcript: #StandardsGoals for 2024: What’s new for BISAC - Tech Forum 2024
 
🐬 The future of MySQL is Postgres 🐘
🐬  The future of MySQL is Postgres   🐘🐬  The future of MySQL is Postgres   🐘
🐬 The future of MySQL is Postgres 🐘
 
Scaling API-first – The story of a global engineering organization
Scaling API-first – The story of a global engineering organizationScaling API-first – The story of a global engineering organization
Scaling API-first – The story of a global engineering organization
 
Finology Group – Insurtech Innovation Award 2024
Finology Group – Insurtech Innovation Award 2024Finology Group – Insurtech Innovation Award 2024
Finology Group – Insurtech Innovation Award 2024
 
Google AI Hackathon: LLM based Evaluator for RAG
Google AI Hackathon: LLM based Evaluator for RAGGoogle AI Hackathon: LLM based Evaluator for RAG
Google AI Hackathon: LLM based Evaluator for RAG
 
FULL ENJOY 🔝 8264348440 🔝 Call Girls in Diplomatic Enclave | Delhi
FULL ENJOY 🔝 8264348440 🔝 Call Girls in Diplomatic Enclave | DelhiFULL ENJOY 🔝 8264348440 🔝 Call Girls in Diplomatic Enclave | Delhi
FULL ENJOY 🔝 8264348440 🔝 Call Girls in Diplomatic Enclave | Delhi
 
Breaking the Kubernetes Kill Chain: Host Path Mount
Breaking the Kubernetes Kill Chain: Host Path MountBreaking the Kubernetes Kill Chain: Host Path Mount
Breaking the Kubernetes Kill Chain: Host Path Mount
 
Strategies for Unlocking Knowledge Management in Microsoft 365 in the Copilot...
Strategies for Unlocking Knowledge Management in Microsoft 365 in the Copilot...Strategies for Unlocking Knowledge Management in Microsoft 365 in the Copilot...
Strategies for Unlocking Knowledge Management in Microsoft 365 in the Copilot...
 
Presentation on how to chat with PDF using ChatGPT code interpreter
Presentation on how to chat with PDF using ChatGPT code interpreterPresentation on how to chat with PDF using ChatGPT code interpreter
Presentation on how to chat with PDF using ChatGPT code interpreter
 

Steve dye onsite oil analysis with astm compliant ftir intruments

  • 1. Onsite Oil Analysis with ASTM Compliant FTIR Instruments Dr Steve Dye Business Development Manager Parker Kittiwake
  • 2. Why use FTIR in Oil Analysis? § One of the most widely used laboratory tools for oil analysis § Fast - capable of detecting multiple oil analysis parameters simultaneously § e.g. water, glycol, soot, oxidation, nitration… § Easy to use - no extensive sample preparation or wet chemistry required § Inexpensive – after capital outlay, per measurement costs are minimal § Quick - replaces tedious and time-consuming physical and chemical methods
  • 3. Infrared Radiation§ Part of the Electromagnetic Spectrum § Longer wavelength than Visible light – “heat” energy § Absorbed by molecules (not atoms) § Gives molecular information, not elemental § Absorption increases molecular vibrations and/or rotations § associated with the internal bonds between the atoms of the molecule The Electromagnetic Spectrum Visible Light Wavelength (m) Frequency (Hz) Wavenumber (cm-1) 10-11 10-10 10-9 10-8 10-7 10-6 10-410-5 10-3 10-2 10010-1 101 102 103 1019 1018 1017 1016 1014 1013 1012 1011 1010 109 108 107 106 105 1Å 1nm 10nm 1μm 10μm 100μm 1mm 1cm Hard X-ray Soft X-ray UV IR Microwave VHF Short Mid Long Radiowave Mid FarNear Near Far Spectral Regions Gamma rays 1000 100 105000
  • 4. Infrared Absorption § Molecules have several types of vibrations. (3n-5 for linear, 3n-6 for non-linear) § Associated with the internal bonds, some (not all) are IR active. § IR radiation absorbed at a characteristic energy (frequency / wavelength). § Qualitative analysis – “fingerprint” identifies which molecules are present § Amount of energy absorbed per unit volume depends on concentration. § Quantitative analysis – signal intensity reflects how much present (ppm or %) H O H H O HH O H H O H H O H H O H H O H H H O H H O IR Radiation Stretching Bending, In Plane Bending, Out of Plane A wavelength (µm) An example; Water (H2O):
  • 6. Traditional scanning IR spectroscopy § First developed in the 1950s § Step wise scanning of a dispersive element and recording of signal § Time consuming § Drift issues § Need a reference to obtain the “instrument function” § Virtually all instruments are now of the Fourier Transform Interferometer design
  • 7. FTIR Spectroscopy § Development started in the late 1960s by analytical chemists borrowing from early work of physicists. § Michelson Morley Interferometer was the classic example § First commercial FT instrument released in 1969. § Uses an interferometer to generate an interference pattern of infrared radiation. § Divide infrared radiation into two paths. § Vary the path length along one of the arms § Recombine the radiation. § Record signal intensity as a function of change in pathlength difference (speed, time) § Interference pattern (interferogram) is in the time domain. § We need frequency (wavenumber, Hz). § Conversion is by Fourier Transform. § Extremely quick these days with computer FFT1 algorithms
  • 8. FTIR Spectrometer Principle Infrared Light Sample Moving Mirror Fixed Mirror Detector Interferogram Beamsplitter Moving mirror speed -> Signal - > 0 0
  • 9. Principle of FTIR Spectroscopy Sample Interferogram Background Interferogram Ratio A = -logT Data Fast Fourier Transform Algorithm Transmittance Spectrum Transmittance Spectrum Absorbance Spectrum Single Beam Spectra
  • 10. Advantages of FTIR Spectroscopy § All wavelengths scanned simultaneously (Fellget’s Advantage) § Reduction in data collection time by a factor of N § Increase in signal-to-noise ratio (SNR) by a factor of N 1/2 by multiple scanning § Greater amount of source energy to detector (Jacquinot’s Advantage) § No rectangular slit aperture § Much greater throughput compared to conventional IR § Fixed internal wavelength reference source § Typically a small HeNe laser § Provides high wavelength accuracy
  • 11. Transmission (Sample) Cell § Fixed pathlength: Typically between 50-200μm § Note: ASTM standard mandates 100µm cell pathlength § Other cell pathlengths allow for different concentrations § Short pathlengths for strong absorbers / high concentration § Longer pathlengths for weak absorbers / low concentration § Cell window material choices: § Water insoluble ZnSe or CaF2 § Water soluble KBr, NaCl or KCl Sample: Fixed Pathlength IR Transparent Cell Windows IR Radiation To Detector
  • 12. FTIR Spectrum of an Oil Sample § Load cell with sample and collect sample spectrum § The FTIR oil sample spectrum will reflect the molecular composition of the oil § For example, an ester based oil: Wavenumber (cm-1) Absorbance 3500 3000 2500 2000 1500 1000 e.g. Ester Based Oil C - H Stretch C=O Stretch C - O Stretch CH 2 , CH 3 Bending CH 2 Rocking IR
  • 13. How can I use FTIR for oil analysis? § Most commonly used as a screening tool § FTIR is a rapid method to identify samples that have problems § Provides general information about the identity and condition of the oil § Degradation by-products, additive depletion, contaminants § FTIR supplements other methods to help diagnose any problems § For example, elemental analysis methods, viscosity measurements, chemical analyses.
  • 14. Collecting an FTIR Spectrum § Specify FTIR instrument parameters: § Resolution (usually 2, 4 or 8 cm-1) § Lower wavenumber means higher resolution § Number of scans § Increasing the scans improves the signal-to-noise ratio by a factor of N1/2 § Apodization § Corrects for interferogram side lobes and affects spectral band shapes § Collect background single beam spectrum § Air or empty cell reference
  • 15. Sampling Oil for FTIR Analysis § Oil should be filtered to remove large particles. § Shake well to give a representative sample. § Most common sample cell is a transmission cell – Manual loading or continuous flow pumping system
  • 16. Example of Used Oil FTIR Spectrum 0.1 0.2 0.3 0.4 0.5 0.6 0.7 Absorbance 100015002000250030003500 Wavenumber (cm-1) Water Soot Oxidation Nitration Antiwear Sulfate Fuel
  • 17. Qualitative or Quantitative? FTIR measurements can be either: § Qualitative § Identify constituents § Observe changes in peak intensities or absorbance – better/worse assessment OR § Quantitative § Obtain a definite result in units of concentration (ex: ppm, g/L, %w/v) § Calibration using standards of known concentrations of the constituent or parameter of interest § Correlate concentrations with absorbance to obtain regression coefficients § Prediction of concentrations in unknown sample using these coefficients
  • 18. Quantitative Analysis - Example § Concentration of water in crankcase lubricants: 0.10 0.20 0.30 0.40 0.0 0.1 0.2 0.3 0.4 0.5 Water (%w/w) Absorbance .1 0 .2 .3 3800 3600 3400 3200 0.5% 0.4% 0.3% 0.2% 0.1% 0.05% 0.0% Wavenumber (cm-1) Absorbance
  • 19. Considerations of Quantitative Analysis § Some methods are considered semi-quantitative: § Difficult to obtain calibration samples with reliable concentration values § Reference method only gives approximate values § Sample to sample variability § Actual samples can differ from calibration samples § For example: For soot measurements, particle size may vary § For example: When measuring fuel, fuel composition may vary between locations § Interferences § Other oil sample constituents and parameters can interfere with the measurement
  • 20. Interferences § Parameters with the same functional group will absorb at the same frequency, interfering with one another § e.g. Water and Glycol Water Ethylene Glycol Absorbance Wavenumber (cm-1) O-H O-H
  • 21. Interferences from the Base Oil § The base oil or new oil additives may also interfere with the measurement of some parameters, especially in the “fingerprint” region. 0.05 0.15 0.25 0.35 0.45 0.55 0.65 0.75 Absorbance 60080010001200140016001800 Wavenumber (cm-1) Oxidation Antiwear Base Oil Hydraulic Oil Fuel Sulfate Fingerprint Region Oil Additives
  • 22. Spectral Subtraction § Spectral subtraction is used to subtract out the interferences from the base oil and additives § The contaminants, degradation by-products and additive depletion can be visualized and analyzed more easily Used Oil Spectrum 0.5 1.0 Absorbance New Reference Oil Spectrum 0.5 1.0 Difference Spectrum 0.5 1.0 100015002000250030003500 Wavenumber (cm-1) Water Soot Oxidation Sulfation Antiwear
  • 23. Considerations of Spectral Subtraction § There are several considerations to take into account when using the spectral subtraction method: § New (reference) oil may not always be available § Batch to batch variation in both base oil and formulation for same oil brand § Need correct reference § Oil changes lead to dilution and/or carryover § Reference no longer good § Topping up with another batch or different brand § Again, reference is no longer good
  • 24. Contamination/Blending with Different Oils § FTIR is also very useful for determining significant changes in oil chemistry § e.g. PAO mixed with an ester- or polyalkylglycol- (PAG) based oi 0.0 0.4 0.8 1.2 1.6 2.0 2.4 Absorbance 100015002000250030003500 Wavenumber (cm-1) Ester PAG
  • 25. Parameter Frequency (cm-1) Traditional Method Oxidation 1710 Acid number (AN) titration Nitration 1630 None Sulfation 1150 Base Number (BN) titration Diesel Fuel 810 Flash Point, Gas Chromatography Gasoline 750 Flash Point, GC Water 3420 Karl Fisher Glycol 1080,1040, 880 Colorimetry, GC Soot 2000 Thermogravimetric Antiwear 980 Elemental Analysis (Zn etc.) (T)BN 1516,1152 BN by titration Common Oil Parameters by FTIR
  • 26. ASTM Standards for oil analysis § American Society for Testing and Materials (now ASTM International) defines standard methods and practices for over 12,000 measurement methods for metals, petroleum, construction, environmental and many more. § Seen by Industry as THE STANDARD that all measurements should be made in accordance with. § ASTM have several published standards and work is ongoing to define more: § 1 Standard practice (E2412) which defines a total of 12 trend parameters for 3 types of lubricating oils; Petroleum based, Extreme Pressure (EP) fluids and Synthetic Polyol Esters. § 5 Standard methods – Oxidation (D7414), Nitration (7624), Phosphate Antiwear (D7412), Sulphation (D7415), Oxidation in transmission oils (D7214) § Proposed methods (WIP) – Acid Number, Base Number, Water Content
  • 27. ASTM Lubricant Practices & Methods § D7418-07 Standard Practice for Set-Up and Operation of Fourier Transform Infrared (FTIR) Spectrometers for In-Service Oil Condition Monitoring. § E2412-10 Standard Practice for Condition Monitoring of in-service Lubricants by Trend Analysis Using Fourier Transform Infrared (FT-IR) Spectrometry. § D7214-07a Standard Test Method for Determination of the Oxidation of Used Lubricants by FT-IR Using Peak Area Increase Calculation. § D7414-09 Standard Test Method for Condition Monitoring of Oxidation in In-Service Petroleum and Hydrocarbon Based Lubricants by Trend Analysis Using Fourier Transform Infrared (FT-IR) Spectrometry. § D7412-09 Standard Test Method for Condition Monitoring of Phosphate Antiwear Additives in In- Service Petroleum and Hydrocarbon Based Lubricants by Trend Analysis Using Fourier Transform Infrared (FT-IR) Spectrometry. § D7415-09 Standard Test Method for Condition Monitoring of Sulfate By-Products in In-Service Petroleum and Hydrocarbon Based Lubricants by Trend Analysis Using Fourier Transform Infrared (FT-IR) Spectrometry. § D7624-10 Standard Test Method for Condition Monitoring of Nitration in In-Service Petroleum and Hydrocarbon Based Lubricants by Trend Analysis Using Fourier Transform Infrared (FT-IR) Spectrometry.
  • 28. Field vs. Lab testing § Lab testing: § Lab testing can offer increased number of alternative tests on oil samples (e.g XRF) § Expert advice may be offered on analysis § Time taken to ship samples from remote locations can be long § Ongoing cost per sample sent § Field Testing: § Immediate results available so action can be taken fast, before major problems arise § No/Low ongoing costs after initial capital outlay § Some user intelligence required to set up parameters for alarm warning levels etc § Subset of full lab tests available
  • 29. Non ASTM compliant field instruments
  • 30. Lab instruments unsuitable for the field
  • 31. Parker Kittiwake FTIR3 Oil Analyser ASTM Compliant, field deployable, all in a small, portable instrument § Simple to use § Software pre-loaded onto Netbook § Multiple equipment records § Trending of multiple parameters § Multiple users
  • 33. Results Screen – Spectral Mode
  • 34. Historical data records on equipment
  • 35. Time trending of parameters
  • 36. What will the FTIR3 Measure § Sulphation – TO ASTM D7415-09 § Oxidation – To ASTM D7414-09 § Nitration – To ASTM D7624-10 § Phosphate Antiwear – To ASTM D7412-09 § E2412: • Soot Water • Diesel • Fuel Contamination • Ethyl Glycol Coolant • Antioxidant depletion § Future published ASTM methods can be readily added to the instrument. § Customer specific methods can be developed and also added to the instrument.
  • 37. Advantages and Limitations § Advantages of FTIR spectroscopy in oil analysis: § Rapid analysis with no extensive sample preparation § Can give information on multiple parameters simultaneously § Provides quantitative and qualitative data § Accurate and repeatable results § Limitations of FTIR spectroscopy in oil analysis: § Different base oils and additives can interfere with the measurement of different parameters § For spectral subtraction, sometimes it can be difficult to find a suitable reference oil § Difficulty measuring some components that are less than 0.1% w/w § No elemental analysis
  • 38. In Summary…. § FTIR provides a quick and accurate test method to measure multiple parameters on in service oils § Advances in recent years have allowed reduced size, field deployable devices to be developed § ASTM ratification of test Methods is an ongoing process, providing end users with confidence in reliable and repeatable measurement techniques and results § More ASTM agreed methods are on the way for additional parameters § Field deployable, ASTM compliant devices are now available